Model G14SD Vapor Management System

Similar documents
Operator's Manual. Storage System. Ultrasound Probe Cabinet. Manufactured by:

NEUTRALIZATION SYSTEM FOR OPA & GLUTARALDEHYDE MEDI-NEWT. User Manual. Endoscope reprocessing the the way way it it should be be

Easy/Tran TF STANDBY POWER FURNACE SWITCH TOOLS NEEDED FOR INSTALLATION. PARTS LIST for Easy / Tran TF FURNACE TRANSFER SWITCH

Instruction Manual. Fudge Puppy Display Case

TXH TOTAL EXHAUST HOOD. Installation and Operations Manual

EXT DUCTLESS FUME HOOD

Contents. Overview. Safety Instructions. Before Installation. Installation. Operation. Maintenance. Replacement Parts. System Specifications 8 01

INSTRUCTION MANUAL IM245. RGS2012E1 S or E Retrofit Kit SUBMERSIBLE GRINDER PUMP INSTALLATION, OPERATION AND TROUBLESHOOTING MANUAL

READ AND SAVE THESE INSTRUCTIONS. ComfortBreeze UV360-1 SYSTEM 24V Ultra-Violet Air Cleaner. Trion

CANARY AUDIO. Vacuum Tube Preamplifier CA-800 OWNER S MANUAL. Handcrafted in California MADE IN USA

Installation Instructions Electric Heaters 5 20 kw

82V LITHIUM-ION BATTERY CHARGER GC 400

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL PSMR1

Bio-Pump Plus. Operating Instructions. Product # ZBP-200. Description. Specifications

CLEAN AIR WORKSTATION REVERSE FLOW. Installation and Operations Manual

ESE Series Cast Iron Sewage Pumps

OPERATION AND MAINTENANCE

TS69 TS65 TS55 TS45 TS5768 TS SERIES INSTALLATION/OWNER'S MANUAL

CANARY AUDIO CA-400. Vacuum Tube MM/MC Phono Preamplifier OWNER S MANUAL. Handcrafted in California MADE IN USA

SUBMERSIBLE SUMP PUMPS

SUBMERSIBLE SUMP PUMPS

Countertop B.I.B. Butter Dispenser Instruction Manual Model #2496

Operations & Maintenance Manual DFH-1000 DUCTLESS FUME HOOD

EXT HEPA/CARBON FUME HOOD

Automatic Burnout Furnaces 115 to 230-volt Models OPERATOR S MANUAL

60V RECHARGEABLE LITHIUM-ION BATTERY

Fiber Optic Illuminator. Model 21AC Model 21DC

VCM LAMINAR FLOW HOOD VERTICAL

1 AMP CURRENT SOURCE

Installation Instructions Electric Heaters 5 20 kw

Quickdraw. Model 2400 Series. Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

DESIGN-AIRE SL-RSR-EC. Installation and Operations Manual

Model 2310 Series. Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

DESIGN-AIRE SL-RSR-EC. Installation and Operations Manual

HL5-7K PLATFORM LIFT INSTRUCTIONS

Model 1672, 1673 Triple Output Power Supply

Electromagnetic Particle Brakes Model: PRB-H

EBS PRODUCTS


HL4K PLATFORM LIFT INSTRUCTIONS

The BVX-100 Fume Extraction System

SLA Battery Capacity Analyzer

AUTO-PUMP 120 VAC AIR COMPRESSOR SYSTEM INSTRUCTION MANUAL. 170 Cherry Avenue West Sayville, NY

HA/HAB Fiberglass Wall Mount Ventilators

DCell Suction. Model DM Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

User s Guide. Oreck Air Purifier with HEPA Filtration. Important! Read this manual carefully, and keep for future reference.

OPERATION and MAINTENANCE INSTRUCTION MANUAL

Installation Instructions

BC-9000 OPERATIONS MANUAL BATTERY CHARGER COFKO ELECTRONICS LLC COPYRIGHT 2014 P/N

CTFRP Series Power Supplies

DWS404 DWS524 DWS654 DWS684 DWS694. DWS SERIES INSTALLATION/OWNER'S MANUAL Car Audio Speakers

BlinkerBeam Solar Controller

DM1016S INSTALLATION/OWNER'S MANUAL 10" Marine DVC Subwoofer

1250 LB. CAPACITY MECHANICAL WHEEL DOLLY

OPERATING and MAINTENANCE INSTRUCTIONS MAX PUNCH TM T HYDRAULIC PUNCH (MP360SDPRO)

4" ENVIRONMENTAL E-SERIES PUMPS OWNER'S MANUAL. DANGER warns about hazards that will cause. WARNING warns about hazards that can cause

Installation Instructions

INSTALLATION INSTRUCTIONS

User s Manual Power Supply. IM E 2nd Edition. Yokogawa Electric Corporation

HL4K PONTOON LIFT INSTRUCTIONS

INSTALLATION/OWNER'S MANUAL DP " Woofer in Enclosure

Nature Power Inverters. True Sinewave Inverter Modified Sinewave Inverter. Owner s Manual

FLEETWOOD TRAVEL TRAILER SLIDEOUT SYSTEM OWNER S MANUAL

Environmental Series Air Curtains

PL900 ILLUMINATOR. Fiber-Lite. Operation Manual. Setup. Fiber Optic Connection A PRODUCT OF DOLAN-JENNER INDUSTRIES. Voltage Selection:

Installation & Maintenance of the Fresh-Aire UV Airborne Duct System for Commercial HVAC Systems. Parts Included. Optional Parts

Instruction Manual. 99 Washington Street Melrose, MA Fax TestEquipmentDepot.com. Model 308A Digital Insulation & Continuity Meter

Owner s Manual & Safety Instructions

HDE36B BATTERY POWERED CRIMP TOOL

TX21U Temperature/Humidity Sensor. Instruction Manual

Power Inverter 400 MW Owner s Manual

OPERATION AND MAINTENANCE INSTRUCTION MANUAL. AEU-17B & AEU-17BV2 Implant / Surgery Systems

Model NTX7 Series Automatic Battery Charger User s Manual Rev. 1.0 October 17, 2006

12.0-Volt Lithium-Ion Power Source

ECSS. Electric Chain Saw Chain Sharpener Assembly & Operating Instructions

This Manual Provides Installation and Operation Instructions for the following models:

Owner s Manual & Safety Instructions

TRAILER WINCH MODELS ST315 AND ST712. General Safety (Continued) Description. Unpacking. General Safety Information.

MODEL GS1-100 Air Filtration System. Operations Manual

MODEL 950C-NP-UV Installation and Owners Manual

User s Manual and Operating Instructions

Model 1693, 1694 Switching DC Power Supply

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com BVX-100. Fume Extraction System _B

Rechargeable Inflator

FUNNEL CAKE FRYER Instruction Manual Models: 8078 / FC-4, 8082 / FC-6, and 8090 / FC-4

Air Bear Supreme and Air Bear Cub. Media Air Cleaners READ AND SAVE THESE INSTRUCTIONS UNITS NOT FOR COMMERCIAL USE. TRION

ME310B BATTERY POWERED CRIMP TOOL

Amarr SuperFlex. Installation Instructions and Owner's Manual. Entrematic 165 Carriage Court Winston Salem, NC

Installation Guide Smart-UPS 1200/1500 VA 100/120/230 Vac Rack-Mount 1U

AFE8B BATTERY POWERED CRIMP TOOL

Podiatry Procedures Chair. For Models: Barrier-Free. User s Guide

Megohmmeter Test Probe

SUNC1200 / ITEM #40882 SUBMERSIBLE UTILITY PUMP OPERATIONS MANUAL

User s Manual and Operating Instructions

Miracle State Of The Art Air Filtration System!

Owner s Manual and Assembly Instructions

2-Pack Indoor Simulated Dome Security Cameras

ACC Series Power Conditioner OPERATION & INSTALLATION MANUAL

SUBMERSIBLE MINI-PUMP

HATCHGRIP Installation Instructions/Operation and Maintenance Manual

Transcription:

Instruction Manual Model G14SD Vapor Management System (For Models: DSD and SSD 110/115v) Manufactured in USA for Medivators Reprocessing Systems by: PCI Medical Inc. 12 Bridge Street, PO Box 188 Deep River, CT 06417 USA Tel: (800) 862-3394 Fax: (866) 271-5982 e-mail: info@pcimedical.com web: www.pcimedical.com

50096-273 Rev. 01 1

Contents Page Labels and Symbols... 3 Environmental Conditions... 2 Supply Rating... 3 Important Information Please Read Before Use... 4 Intended Use... 4 Instruction Manual... 4 Repair and Modification... 3 Signal Words... 4 Chapter 1 Functions... 5 1.1 General Information... 5 1.2 Switch/Fuse Location... 5 1.3 Specifications... 6 Chapter 2 Installation and Setup... 7 2.1 Upon Delivery... 7 2.2 Component List... 7 2.3 Installation of Adaptor Box and G14SD... 7 2.4 Overview... 7 2.5 Installation... 7 Chapter 3 Operation... 9 Chapter 4 System Maintenance...10 4.1 Changing the Filters... 10 4.2 Care... 10 Chapter 5 Troubleshooting Guide...11 Warranty...12 Filter Guarantee...12 Parts List...13 50096-273 Rev. 01 2

Labels and Symbols The meanings of the labels and symbols that appear on the packaging and/or the instrument are as follows: Dangerous electrical voltage Hazardous area Ground point Environmental Conditions For indoor use only. Altitude up to 2,000 meters. Temperature range: 5 C to 40 C. Maximum relative humidity: 80% for temperatures up to 31 C, decreasing linearly to 50% relative humidity at 40 C. Main supply voltage fluctuations not to exceed ± 10% of the nominal voltage. Insulation over voltage category II. Pollution degree 2. Supply Ratings AC 110-120V 0.5A - 1.0A 50/60 Hz Fuse: 1.0 amp externally mounted 50096-273 Rev. 01 3

Important Information Please Read Before Use Intended Use The G14SD has been designed to minimize exposure to chemical fumes from the Medivators Reprocessing Systems/Olympus DSD and SSD automated washer/disinfectors. The unit s exhaust fan and carbon filter remove fumes of high-level disinfectant solutions during the disinfection cycle. The cabinet design allows for simple attachment to the back of the DSD and SSD machines. Do not use the system for any purpose other than its intended use. Instruction Manual This instruction manual contains essential information on using this system safely and effectively. Before use, thoroughly review this manual and the manuals of all equipment which will be used during the procedure, and use all equipment as instructed. Keep this and all related instruction manuals in a safe, accessible location. If you have any questions or comments regarding any information in this instruction manual, please contact PCI medical. Repair and Modification This system contains no user-serviceable parts. Do not modify or attempt to repair it; contact PCI Medical Inc. to purchase replacement filters and parts. Equipment damage can result from user attempts to repair or modify system components. Before attempting to perform routine maintenance on the system, thoroughly review and understand this manual for complete familiarization with the system and its operation. Signal Words The following signal words are used throughout this manual: WARNING CAUTION NOTE Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices or potential equipment damage. Indicates additional helpful information. 50096-273 Rev. 01 4

Chapter 1 Functions 1.1 General Information Figure 1. GUS Model G14SD Vapor Management System Installed on Olympus DSD 1.2 Switch/Fuse Location Power switch The switch is illuminated red when the system is ON. Power cord Connect to a hospital-grade wall mains outlet. Fuse Type FSF 5 x 20 mm, 1 amp, 250 VAC. Return vent Filtered air is returned to the room through the bottom vent. Ratings plate The model name, electrical information and serial number are shown. Figure 2. Switch/fuse location Fuse Ratings Plate and Serial Number On/Off Switch Power Cord 50096-273 Rev. 01 5

1.3 Specifications Airflow: Volume of air treated: CFM...50 Average Face Velocity: FPM....60 Electrical: Voltage... 110V 50/60hz Current-Running... 19 watts Current-Max...0.5 amps Filters: Carbon filter... activated carbon Fan/Blower:...backward curved impeller blower Construction: Fume Hood... powder coated carbon steel & ABS plastic Dimensions: G14SD (Height x Width x Depth)... 14.3 x 20.5 x 8 Weight 24 lbs 50096-273 Rev. 01 6

Chapter 2 Installation and Setup 2.1 The 110 Volt Charcoal Fume Removal System Kit, MK01-0055, consists of the following items: Item Description Quantity Medivators Part Number 1 2-inch Hole Saw 1 MB03-0503 2 Vent Louver 2 MC13-0007 3 Screws 10 MS05-0045 4 Drill Template Side A and B 1 MT06-0506 5 Pilot Drill 0.144 (#27) 1 MB03-0504 6 Charcoal Fume Removal System 1 78399-219 2.2 Installation of adaptor box and G14SD Tools needed: Power Hand Drill Philips Screwdriver Tape Vacuum 2.3 Installation Instructions NOTE: 1. These instructions are for mounting the Charcoal Fume Removal System (CRFS) to the DSD cabinet. 2. If the DSD already has the vent louver fitted proceed to step 2.3.8. 3. Wear proper protective gear during this procedure. 4. If the unit is already hard plumbed, call the facility maintenance personnel to disconnect the unit. 5. Take extra care while moving the unit if the backside is hard to reach. 6. Unplug the unit before installation. 2.3.1 Open both lids of the unit and mask around the floating lids and the basin openings, as much as possible, to prevent debris and plastic chips from falling inside the basin. 2.3.2 Remove the six center screws holding the hinge to the back plate of the unit, as shown in figure 1. Remove center screws Figure 1. 50096-273 Rev. 01 7

2.3.3 Line up the template, MT06-0506, with the DSD cabinet by matching the six small holes on the top of the template to the six center screw holes that the screws were removed from in step 1.3.2. Refer to figure 2. Fasten Template Here Fasten Template Here Template 2-inch Hole Locations Figure 2. 2.3.4 Attach the template to the DSD cabinet by fastening two screws through the two screw holes on each end of the template. Refer to figure 2. 2.3.5 Mark the center of 2-inch holes using the provided pilot drill bit. Refer to figure 2. 2.3.6 Remove the template and discard. 2.3.7 Drill both 2" holes using the provided hole saw. It is important to keep the hole saw level to insure the proper fit of the vent louvers. 2.3.8 Unpack the Fume Removal System and make sure the filter is unwrapped and properly in place. NOTE: The charcoal particles in the filter may have settled during shipment. Shake the charcoal filter horizontally to evenly distribute the charcoal particles. 2.4.9 Mount the CFRS to the six screw holes at the back of the DSD disinfector using the provided screws (MS05-0045). Refer to figure 3. 2.4.10 Use the pilot drill to pre-drill, the bottom screw holes, 3/16-inch (4.7mm) deep, into the back plate of the DSD disinfector. Be careful not to drill through the cabinet wall. Refer to figure 3. 2.4.11 Fasten the bottom side of the CFRS using 3 screws (MS05-0045). 2.4.12 Vacuum all chips and debris from behind the unit and from the basin compartments. 2.4.13 Insert the vent louvers (MC13-0007) into both 2-inch holes with the openings in the louvers facing down. 2.4.14 Attach the power plug to the power cord of the CFRS. (If applicable) 2.4.15 Plug the power cord to a proper outlet. 2.4.16 Turn on the main switch at the right hand side of the CFRS. 50096-273 Rev. 01 8

Chapter 3 Operation The G14SD is designed to run continuously to prevent any fume build up inside the unit. The red light in the on/off switch will be illuminated when system is operating. 50096-273 Rev. 01 9

Chapter 4 System Maintenance 4.1 Changing the Filters The GUS System has a patented carbon filter that should last six months, even during continuous use. The pleated design of the filters eliminates the need for a pre-filter. PCI Medical Inc. will notify your facility 30 days prior to each filter change and arrange for timely delivery of a replacement filter. The 4-year filter program automatically ships and bills at each change interval, saving your facility time and money. With the four-year program, the prices for the filters are fixed, and if purchased within six months of receipt of your station, the warranty is extended to four years. Please see warranty information for specifics. NOTE Contact Olympus to order replacement filters. CAUTION The system s filter must be replaced every six (6) months. Wear personal protective equipment such as eyewear, facemask, moisture-resistant clothing and chemical-resistant gloves. Disconnect the system s power supply before removing the filter. 1. Turn the system s power switch OFF, and disconnect the power cord from the wall mains outlet. 2. Unlatch the top lid. 3. Remove the old filter by pulling up on the strap located on top of the filter. Place in plastic bag and discard with other non-hazardous waste. 4. Replace with a new filter. Verify that the direction of the arrows on the filter is pointing to the back of the GUS system. 5. Close and latch lid. 4.2 Care Cleaning the System 1. Turn the power switch OFF and disconnect the power cord from the wall-mains outlet. 2. To clean metal surfaces: Use a clean, lint-free cloth moistened with a mild detergent solution and water. A mild solvent may be used on these surfaces for difficult-to-clean areas 3. Make sure that the unit is completely dry before use. 50096-273 Rev. 01 10

Chapter 5 Troubleshooting Guide If the GUS System is not running, check the following: 1. Ensure that the unit is plugged into a live wall outlet. 2. Make sure the red On/Off switch is on. The switch should be illuminated when on. 3. In the event that the system is still not working, you will need to check the fuse, which is located next to the On/Off switch. 4. Disconnect the plug from wall outlet and remove the fuse. Check to see if the filament is damaged. If so, replace with a standard 1-amp fuse. 5. If the system is still not working, contact PCI Medical Inc. at: 1-800-862-3394 50096-273 Rev. 01 11

Warranty FULL ONE-YEAR WARRANTY PCI Medical Inc. warrants the GUS Vapor Management System against defects in material or workmanship for a period of one year from the date of delivery. PCI Medical Inc. will repair or replace, at its option, with evidences, a defect in material or workmanship within the warranty period. The warranty does not include filters. PCI Medical Inc. assumes no liability for consequential damage of any kind as a result of the use or misuse of the system by the purchaser, their employees or others. LIMITED FOUR-YEAR WARRANTY This warranty is only available to those customers who have issued a purchase order for the four-year filter replacement program within six months of receiving their system. During the period beginning with the date of original purchase, and continuing through the fourth year, PCI Medical Inc. will repair or replace, at its option, with evidences, any part of the GUS Vapor Management System within the warranty period if defective in material or workmanship. This warranty does not cover filters or acrylic unless it can be determined that failure is due to defects in material or workmanship. TERMS APPLICABLE TO BOTH WARRANTIES What is not covered by these warranties: 1. Conditions and damages resulting from any of the following: Improper installation, delivery within the facility, or maintenance Any repair, modification, alteration or adjustment not authorized by the manufacturer Misuse, abuse, accidents or unreasonable use Incorrect electric current, voltage or supply Failure to follow proper procedures for use outlined in this manual 2. Consequential or incidental damages sustained by any person as a result of any breach of these warranties. Some states do not allow the exclusion or limitation of consequential or incidental damages, so the above exclusion may not apply. These warranties give you specific legal rights, and you may also have other rights, which may vary from state to state. Filter Guarantee PCI Medical Inc. guarantees that your filter will adsorb fumes from high-level disinfectants for a minimum of six months under normal use. If a breakthrough should occur during this period, PCI Medical Inc. will replace the filter at no charge. 50096-273 Rev. 01 12

PARTS LIST Part Number MF01-0037 Item Replacement Carbon Filter REGISTRATION CARD Please complete and return the attached Warranty and Filter Replacement Card to PCI Medical Inc., to ensure proper notification for filter replacements. You may complete this warranty card online at: www.pcimedical.com GUS is a registered trademark of PCI medical inc. GUS Model G14SD Instruction Manual 05/2004 Printed in the United States of America No part of this publication may be reproduced without the express written permission of PCI medical inc. Vs.10/14/04 50096-273 Rev. 01 13