User's Manual Digital High Voltage Insulation Tester Model

Similar documents
USER MANUAL. 10KV Digital High Voltage Insulation Tester. Model MG500. Additional User Manual Translations available at

User Guide. Model Insulation Tester / Megohmmeter

User Guide. Model Insulation Tester / Megohmmeter

Model Insulation Tester / Megohmmeter. User Manual

User Guide. Model Insulation Tester / Megohmmeter. Introduction

TES-1601 INSTRUCTION MANUAL AUTORANGING INSULATION TESTER SAFETY INFORMATION TES ELECTRICAL ELECTRONIC CORP. Warnings and Safety symbols:

Model ST Instruction Manual. Digital Insulation Tester. reedinstruments. www. com

User's Manual. High Voltage Megohmmeter. Analog Insulation Tester plus AC Voltage and Continuity Tests. Model

Digital/Analog Megohmmeter Model 1026

USER MANUAL. Insulation Tester + DMM. Model MG320

Non-Contact Phase Rotation Tester

MAX. AC 600V HOLD V M H.V. LIVE CIRCUIT DIGITAL INSULATION TESTER DO NOT OPERATE IF LIT POWER DATA HOLD TEST ON / OFF 250V 500V 1000V OFF CAT.

NEW 15KV HIGH VOLTAGE INSULATION RESISTANCE TESTER MODEL KM 7015 IN FEATURES : CAT IV 600V. Short Circuit Current 5mA.

PHASE ROTATION INDICATOR with OPEN PHASE CHECKER

ANALOGUE INSULATION-CONTINUITY and VOLTAGE METER

Mini Multimeter with Non-Contact Voltage Detector (NCV)

BM kv Digital Insulation Tester USER MANUAL

OPERATING INSTRUCTION

99 Washington Street Melrose, MA Fax TestEquipmentDepot.com. Instruction Manual. Model 2831D 4 ½ Digit True RMS Digital Multimeter

Digital/Analog Megohmmeter Model 1026 USER MANUAL

DIGITAL INSULATION TESTER RS-5500/5505 INSTRUCTION MANUAL

OPERATING INSTRUCTION

1.Safe testing IMPORTANT:

Model Battery Powered Milliohm Meter

Non-Contact Phase Rotation Tester

Mini Multimeter with Non-Contact Voltage Detector (NCV)

True RMS Autoranging Multimeter

Table of Contents. Title Page UT501/UT502 OPERATING MANUAL. Introduction. Unpacking the Meter. Safety Information. International Electrical Symbols

Safety design conforming to the following provisions

Autoranging Multimeter

EST Electrical Service Technician Answer Schedule

Instruction manual. High voltage insulation resistance tester KEW3121B/3122B

Autoranging Industrial Multimeter

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL PSMR1

Safety design conforming to the following provisions

DIGITAL RCD(ELCB) TESTER

Instruction Manual. 99 Washington Street Melrose, MA Fax TestEquipmentDepot.com. Model 308A Digital Insulation & Continuity Meter

INSTRUCTION MANUAL PORTABLE APPLIANCE TESTER

Digital Insulation Resistance Testers. you can depend on

Table of Contents Title Page No.

l The Battery Tester is designed for measuring the l AC four-terminal method to measure the internal

Declaration of Conformity

Chek-A-Cell TM TestEquipmentDepot.com. Instruction Manual Rev A 8/12

MEGGER BM15/MJ15 MEGGER. Insulation Testers. User Guide. MeterCenter (800)

User's Manual. Earth Ground Resistance Tester. Model Introduction

EZYPAT / EZYPAT+ / SMARTPAT PORTABLE APPLIANCE TESTER INSTRUCTION MANUAL

POWERED BY KT71 / Portable Appliance Tester. Instruction Manual

USER MANUAL. 200A AC Mini Clamp Meter. Model MA130. Please visit for user manual translations

MY /2-DIGIT DIGITAL MULTIMETER Users Manual

DIGITAL INSULATION RESISTANCE TESTER (5KV, 1T )

OPERATING INSTRUCTIONS

Users Manual. MegOhmMeter

Model (Circuit Tester) Model (GFCI Tester)

A.W. SPERRY INSTRUMENTS INC.

4-WIRE EARTH RESISTANCE TESTER C1 ES P1 S. Max. AC 200V MICROPROCESSOR CONTROLLED TEST. Switch ON, TEST, SCROLL RESULTS 20Ω 200Ω 2000Ω OFF

Instruction Manual. Model 307A Analog Insulation & Continuity Meter

& HIGH CURRENT DC POWER SUPPLIES INSTRUCTION MANUAL

AMB V Insulation Resistance Tester. Users Manual. For detailed specifications and ordering info go to

3 PHASE TESTER & MOTOR ROTATION TESTER

AC/DC CURRENT TRANSDUCER INSTRUCTION MANUAL

MOdel No: MM19.V3 WARNING CAUTION 1. SAFETY INSTRUCTIONS

INSTRUCTION MANUAL ANALOGUE INSULATION TESTER KEW 3131M KYORITSU ELECTRICAL INSTRUMENTS WORKS,LTD., TOKYO JAPAN

KMP Insulation-Continuity Tester. Users Manual

MM300. INSTRUCTION MANUAL Manual-Ranging Digital Multimeter

INSTRUCTION MANUAL MARTINDALE CM52 CLAMP METER. ELECTRIC Trusted by professionals. Other products from Martindale:

User Guide. Panel Mount Digital Tachometer. Model Washington Street Melrose, MA Phone Toll Free

Users Manual. Accessories Motor and Phase Rotation Indicator ships with the following items: 3 test leads 3 test probes 3 alligator clips 9 V battery

1000FLT. Users Manual. Fluorescent Light Tester

BATTERY BOOSTER/CHARGER MODEL NO: DIGICAR 600

WARNING MANUFACTURER ASSUMES NO LIABILITY IF UNIT OPERATED IN AN UNSAFE MANNER.

Automatic taper of charge rate for superior battery life through good equalization of cells and low water use rate.

BATTERY BOOSTER/CHARGER MODEL NO: DIGICAR 900

BATTERY OPERATED DIGITAL MILLI-OHMMETER

Digital Multimeter User's Manual

Tracer ORP PockeTester TM Code 1742

MegOhmMeter. Users Manual

Advanced Test Equipment Rentals ATEC (2832)

Vol-Con XL (61-086) Voltage/Continuity Tester and Vol-Test (61-085) Voltage Tester Instruction Manual WARNING

Distributor. Kewtech Corporation Limited 76 St. Catherine s Grove Lincoln LN5 8NA KEWTECH

SafetyPAT Plus A B P/N:

OPERATING INSTRUCTIONS

TRIPLETT. Model 310 Type C Type 9. Test Equipment Depot Washington Street, Melrose, MA TestEquipmentDepot.

KTD30 Digital Insulation and Continuity Tester. User Manual. kewtechcorp.com

INSULATION RESISTANCE TESTER FOR PHOTOVOLTAIC SYSTEMS MSEI-200C INSTRUCTION MANUAL

Phase Sequence Tester

MIT515, MIT525, MIT1025, MIT1525

OPERATING INSTRUCTION

1 This instrument must only be used by a competent and trained person and operated in strict accordance with the instructions.

362 Clamp Meter. Users Manual

DC POWER SUPPLY ALIMENTATION C.C.

1662/1663/1664 FC. Electrical Installation Tester. Safety Information. Warnings

BATTERY CHARGER RS-1000 Instruction Manual

MIT300 Series Insulation and Continuity Testers

90W DC Power Supply. User Manual. 99 Washington Street Melrose, MA Phone Toll Free

Attention ETCR8600 Leakage Protector Tester To read this user manual carefully. To comply with the operating cautions presented in this manual.

KD1E Range Potential Indicators

Digital Low Resistance Ohm Meter

Dawson DDM190D. Digital Multimeter User s Manual

8-Bay Rapid Charger. for AA/AAA NiMH & NiCd Rechargeable Batteries. Owner's Manual. Please read before using this equipment.

KTD30 Digital Insulation and Continuity Tester

Transcription:

99 Washington Street Melrose, MA 02176 Phone 781-665-1400 Toll Free 1-800-517-8431 Visit us at www.testequipmentdepot.com Back to the Extech 380375 Product Page User's Manual Digital High Voltage Insulation Tester Model 380375 Test Equipment Depot, 99 Washington Street, Melrose, MA 02176 781.665.1400 800.517.8431 Fax: 781.665.0780

Introduction Congratulations on your purchase of Extech s High Voltage Insulation Tester. This device has four voltage test ranges (up to 5kV) and measures insulation resistance to 250GΩ. This professional meter, with proper care, will provide years of safe reliable service. Safety 1. Circuits under test must be de-energized and isolated before connections are made. 2. Circuit connections must not be touched during a test. 3. After insulation tests, capacitors must be discharged. 4. Test leads (including alligator clips) must be in good working order, clean and without broken or cracked insulation. 5. When servicing, use only specified replacement parts. 6. Environmental conditions: Indoor use only; Installation category II Pollution degree 2 Altitude up to 2000 meters Relative Humidity 80% max.; Temperature 0 to 40 o C International Safety Symbols Caution, refer to this manual before using this meter Dangerous Voltages; risk of electric shock Meter is protected throughout by double or reinforced insulation Test Equipment Depot, 99 Washington Street, Melrose, MA 02176 781.665.1400 800.517.8431 Fax: 781.665.0780 2

Specifications General specifications Display Range selection Bargraph Automatic discharge Low battery indicator Power source Auto Power off Enersave TM mode Operating conditions Storage conditions Dimensions Weight 2 x 16 character alphanumeric multi-function dot matrix Automatic ranging Displays voltage ramp, soak, and decay After automatic & manual stop or upon completion of test "Replace battery" displayed when battery voltage is low Eight 1.5V AA cells; Battery life 40 hrs (no load at 5kV) After 30 minutes Shorter test time to conserve battery life 32 to 104 o F (0 to 40 o C); < 80% RH 14 to 140 o F (-10 to 60 o C); < 80% RH 4.7 x 6.7 x 3.7 (120 x 170 x 95mm) 1.94 lbs (880g) Range Specifications Test Voltage Resistance Ranges Short Circuit Current Accuracy Resolution 500V DC 0 to 25 GΩ 1mA 1000V DC 0 to 50 GΩ 1mA 2500V DC 0 to 125 GΩ 1mA 5000V DC 0 to 250 GΩ 1.6 ma 0 to 50GΩ: ± 3% reading 50 to 250GΩ: ± 5% reading 1kΩ Meter Description 1. Test lead jacks compartment 2. Dot Matrix Display 2 3. Voltage test buttons 4. ON / TEST button 5. Battery compartment is located on bottom of unit 3 4 1 \ 5 3

Operation Note: Ensure that the circuit under test does not include devices or components that can be damaged by 5000VDC; such devices include power factor correction capacitors, low voltage mineral insulated cables, electronic light dimmers, ballasts and starters for fluorescent lamps. Connecting the Test Leads to the meter The Black (Earth) and Red (Line) test leads connect to the meter in the lead compartment located on the back of the meter (the meter jacks are color coded). NOTE: The Green (Guard) lead is typically not used. Do not connect this test lead to the meter unless absolutely certain it is going to be used. For more information, continue reading this Operation section and the Applications section to follow. Connecting the Test Leads to the device under test The Black (Earth) lead is connected to the conductive material. The Red (Line) lead is connected to the insulating material. The Green (Guard) lead is at the same potential as the negative terminal (0V) and, when required, is used to prevent erroneous readings by removing parallel leakage paths from the measurement. Insulation Resistance Measurements 1. Connect the test leads to the meter and the device under test. 2. Press the ON / TEST button to turn the unit on. The Select Test unit will display the turn-on screens and then stop at 2.5kV, 1kV, the Select Test display. 3. Select the desired test voltage by pressing the 500V, 1KV, 2.5KV, or 5KV button. The screen will briefly display the selected test voltage and then will proceed to the Connect Leads Testing for LIVE display. If the circuit is live the display will indicate Live Warning Circuit Live and the meter will beep. If this occurs, remove the test leads and press the ON / TEST button. 4. Press the ON / TEST button to apply the voltage and begin the test. The meter will beep every 2 seconds which indicated high voltage is on the test leads and the meter will intermittently buzz while the high voltage is being generated. 5. On the display, elapsed test time appears on the upper right, the resistance reading appears on the upper left, the selected voltage is shown on the lower right and the bar graph indicates the actual voltage present on the test leads. 5kV 500V Connect Leads Testing for LIVE R= 108.2MΩ 6.5s 0 500V 6. At the completion of the test the meter will discharge R= 128.4MΩ 7.4s the device under test and the display HOLD. Do not 0 HOLD 500V remove the test leads until the bargraph disappears (complete discharge) and HOLD appears in the display. 7. The meter automatically discharges the system at the end of the test. The test ends when one of the following occurs. a. After approximately 10 seconds in the Enersave TM mode. b. After approximately 100 seconds in the continuous mode. c. When the red ON / TEST button is pressed. 4

The Enersave TM Mode Enersave TM conserves battery life by reducing the test time from 100 seconds to approx. 10 seconds. Enersave TM is the default mode of operation. To disable Enersave TM press and hold the ON / TEST button for three seconds to begin a test. Bargraph Voltage Display The bargraph represents the voltage present on the test leads as it charges, soaks, and discharges. The bargraph appears on the lower portion of the display during tests. Automatic Low Resistance Detect If the display shows the message LOW MΩ, the test should be interrupted immediately by pressing the ON / TEST button. This message indicates that the insulation under test has broken down and the meter is trying to inject a high potential onto a short circuit. Elapsed Timer The test duration is indicated on the display. This is particularly useful in determining whether an insulating material under test will break down in a given amount of time. Manual Test STOP To stop a test in progress, press the ON / TEST button. The test will immediately end and the system will automatically discharge. Automatic Test STOP The test automatically stops after 100 seconds if the meter is not in the Enersave TM mode. When in the Enersave TM mode, the test automatically stops after approx. 10 seconds. Live Circuit Warning If the test leads are connected to a live circuit when testing, a warning beeper will sound and the meter will display Live Warning Circuit Live. In this case, correct the problem and retest. Automatic Discharge At the end of a test, the meter automatically discharges the high voltage. The automatic discharge status is reflected on the display. During discharge the beeper will sound and, when completely discharged, the HOLD icon appears on the display. Auto OFF The meter will automatically turn off after approximately 30 minutes of operation. Maintenance Battery Replacement When REPLACE BATTERY appears on the display, replace the eight 1.5V AA batteries. 1. Ensure that the meter is powered down and that the test leads are not connected 2. Close the instrument cover and turn the instrument upside down 3. Remove the phillips head screw located on the bottom of the meter 4. Remove the battery compartment cover 5. Replace the eight batteries ensuring proper polarity 6. Affix the rear cover and secure the rear screw Cleaning Periodically wipe the case with a dry cloth. Do not use solvents or abrasives to clean this instrument. 5

Applications Measuring Power Tools and Small Appliances For small appliances, connect the Black (EARTH) lead to conductors and the Red (LINE) lead to insulating material. For single or double insulated power tools, one lead should be connected to the device s chuck, blade, etc. and the other lead to one of the AC power cord conductors (test both conductors in turn). Refer to the power tool application diagram below. AC Plug For 2 or 3 prong plugs, connect the RED lead to one conductor at a time Connect the BLACK lead to the chuck Testing Motors AC Motors Disconnect the motor from the line by: a. Disconnecting the wires from the motor terminals or b. Opening the main switch If the main switch is opened, and the motor also has a motor-starter, then the starter must be held in the ON position. With the main switch opened, the measured resistance will include the resistance of the motor wire and all other components between the motor and the main switch. If a weakness is indicated, the motor and other components should be checked individually. If the motor is disconnected at the motor terminals, connect the RED lead to the grounded motor housing and the BLACK lead to one of the motor leads. DC Motors 1. Disconnect the motor from the line. 2. To test the brush rigging, field coils and armature, connect the RED lead to the grounded motor housing and the BLACK lead to the brush on the commutator. 3. If the resistance measurement indicates a weakness, raise the brushes off of the commutator and separately test the armature, field coils and brush rigging by connecting one lead to each individually, leaving the other connected to the grounded motor housing. This also applies to DC Generators. Testing Single Conductor Cables 1. Completely disconnect the cable under test from its source and destination. 2. Connect the test leads to the cable as shown at right. Connect BLACK lead to the motor leads Connect RED lead to grounded housing Core Insulation Sheath Covering EARTH GUARD LINE 6

Testing Multi Conductor Cables 1. Completely disconnect the cable under test from its source and destination. 2. Refer to the diagrams at right for three possible tests. EARTH LINE EARTH LINE GUARD Testing one conductor to all of the conductors Testing one conductor to earth GUARD LINE EARTH Testing one conductor to all conductors minus ground Transformer Testing Transformer tests are performed with the transformer completely disconnected from the line and the load. Note that the case ground should not be removed. The five tests listed below will completely test a single-phase transformer. Note that at least 1 minute should be allowed between each test. 1. High voltage winding to low voltage winding and ground 2. Low voltage winding to high voltage winding and ground 3. High voltage winding to low voltage winding (pictured below) 4. High voltage winding to ground 5. Low voltage winding to ground Connect the BLACK lead here Connect the RED lead here Test Equipment Depot, 99 Washington Street, Melrose, MA 02176 781.665.1400 800.517.8431 Fax: 781.665.0780