Installation Guide. Auxiliary Pressure Control Manifold on 6890 GC Accessory G1570A

Similar documents
Installation Guide. Nitrogen-Phosphorus Detector on 6890 GC Accessories G1575A, G1576A, G1594A, G1599A

Installation Guide. Flame Ionization Detector on a 6890 GC. Accessories G1561A, G1562A, G1591A, G1598A

Installation Guide. Purged Packed Inlet, Electronic Pneumatics Control on 6890 GC Accessory G1551A

Installing an Auxiliary Pressure Control Module

Installation Guide. Heater Interface Cable for Mass Sensitive Detector Accessory G1590A

Installing the Flame Ionization Detector EPC Flow Control Manifold

Installing an Auxiliary Pressure Control Module

Installation Guide. Flame Ionization Detector on a 6850 GC Accessory G2621

Installation Guide NPD/ECD Chemical Filters Accessory 19199N

Installation Guide. Thermal Conductivity Detector on a 6850 GC Accessory G2623A

Installation Guide. Volatiles Interface for 6890 Gas Chromatograph Accessory G2319A

Installation Guide Thermal Conductivity Detector Accessory 19232E

Installing a Purged Packed (PP) Inlet

Nickel Catalyst Tube Accessory 19205A

Agilent G1554A On-Column Capillary Inlet with EPC

Installing a Cool On-Column Inlet

Installing a PTV Inlet

Installing a PTV Inlet

Installing the CO 2 Cryogenic Oven Cooling Kit

Valve Box and Valve Accessories G4315A, G4316A, G4317A

Installing a Purged Packed Inlet

Agilent G3510A/G3511A Multimode Inlet

Agilent G2738A Upstream Capillary Interface

Nickel Catalyst Tube on a 6820 GC, Accessory G4318A

Installing the 6850 Direct Valve Column Connector Kit

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Model Fixed Rack-Mount Kit

Switching DC Power Supply

PET UNIT DOSE CABINET

Replacing the Batteries in the Fortress LI 660

Installation Guide. Agilent 7683 Automatic Liquid Sampler

Deans Switch Microfluidics

Installation Instructions. PowerFlex 700 Drive - Frame 8 Components Replacement

Legacy Combustion Chamber Replacement Kit

Model 4430 Photoionization Detector (PID) Operator s Manual

Installation Instructions 300 A Current Sensor Kit for D1h/D3h/D5h/D6h Enclosure Sizes VLT Series FC 102, FC 103, FC 202, and FC 302

260 Solvent Vapor Exit Accessory

Agilent G2855A Deans Switching System

29 Installation Step 1. Unpacking the GC Cable diagrams Step 2. Placing the GC system on the benchtop Step 3. Turning the power on

SOS SERIES SOS1 SOS2. Spares On Site Battery Cabinet Installation Guide rEV3

F-4600 INLINE ULTRASONIC FLOW METER Installation and Operation Guide

FOG LAMPS INSTALL KIT

Triple Output Power Supply

5000TOC Sensor Service Manual

SITRANS F. Ultrasonic Flowmeters. Hi-Precision Mounts. Operating Instructions. Answers for industry.

Calibration and Performance Check: Agilent Dual Plasma Controller

N/S= Not Shown Bold = Part available for purchase. Italic = Part not available for purchase, listed and shown for reference only.

PET UNIT DOSE TABLE INSTALLATION MANUAL FN: Rev A 1/18

Digitrip Retrofit System for ITE K-3000, K-3000 S, K-4000 and K-4000 S Breakers

Catalyser Introduction. 3. Hazard Information. 2. Area of Application. 3.1 Symbology

Agilent G4367A Auxiliary Electronic Pressure Control (Aux EPC) Module

Prodigy HDLV Generation II Pump Panel

Siemens Drives & PLCs

If there is additional labor or other costs above the 4 hrs labor you must call VMAC for approved coverage

Installation Instructions 500 A Current Sensor Kit for D2h/D4h/D7h/D8h Enclosure Sizes VLT Series FC 102, FC 103, FC 202, and FC 302

Operations and Service Manual. X30208 Load Bank

Instruction Manual AVTM for. Strip Chart Recorder Catalog Nos and

Upgrading the Sure Coat Modular Gun Control System

Replacing the Battery in the Patriot SPS 250, SPS 300 and SPS 450

VMC - Hydraulic Counterbalance - Replacement

Installation Instructions

Endura AZ10 Oxygen probe Sensor sub-assembly tested spares kit

Alcatel-Lucent 7310 DC Metro Cell AWS (B4) Kit MC-DBR1, MC-DBR2, MC-DBR3, MC-DBR4, MC-DBR5. Installation Manual

VISOR AUDIO KIT P/N , APPLICATION BEFORE YOU BEGIN KIT CONTENTS. Verify accessory fitment at Polaris.com.

<THESE INSTRUCTIONS MUST BE GIVEN TO THE END USER> B&W

INSTALLATION INSTRUCTIONS

Manual No: Revision: B. CCVP Rain Shield. Installation Guide

Model 1693, 1694 Switching DC Power Supply

HP Modular Cooling System 200/100 Options Installation Guide

SL-10/20/30, ST-10/10Y - Motor Coupling - Alignment

Installation Instructions

SCION xb AUTO-DIMMING MIRROR Preparation

Section 7 - Troubleshooting Guide

DASH AUDIO KIT P/N APPLICATION BEFORE YOU BEGIN KIT CONTENTS. Instr Rev 01 04/16 Page 1 of 6. Verify accessory fitment at Polaris.com.

Removable Applied Panel Mounting System

INSTALLATION INSTRUCTIONS PART NUMBER AEM C (GUN METAL GRAY FINISH)

TOYOTA TUNDRA BIG BRAKE KIT Section I - Installation Preparation

DC POWER SUPPLY ALIMENTATION C.C.

TOYOTA Yaris Hatchback EC REARVIEW MIRROR Preparation

Control Panel Interface Upgrade Installation Guide For Model 200i and 250i Motorcycle Dynamometers Serial Number 202xxxx.

Part Number: PT

JODALE PERRY. Parts List & Mounting Instructions. Jacobsen HR9016 JDP BUILT FOR LIFE

JEEVES. JEEVES Installation Manual. Installation Manual The Easiest Do-It-Yourself Dumbwaiter on the Market

Exhaust System Installation Turbo Downpipe Audi MKII TT-RS 2.5T CD100028

TRAVEL'R DIRECT RESPONSE

Scion xb EC Rearview Mirror w/ Shift Light Preparation

Read these instructions thoroughly before attempting to install this option.

An ISO 9001 Company. BOP 1KW-MG FIRMWARE RETROFIT KIT

INSTALLATION INSTRUCTIONS

CRECOMZR054B00 CRECOMZR055B00. VERTICAL AND HORIZONTAL ECONOMI$ER2 ACCESSORY MEDIUM ROOFTOP UNITS 15 to 27 1/2 TONS WARNING TABLE OF CONTENTS

RCA PLUG KIT P/N APPLICATION BEFORE YOU BEGIN KIT CONTENTS. Verify accessory fitment at Polaris.com.

nual k Voltage Stabilizer N) manual New Brunswick Voltage Stabilizer Operating manual

Exchange of rollers from the XTS-Mover

PFadvantage Metalfor Araus 1360

Forbidden Diesel Performance Ford 6.7L Powerstroke EGR Delete Instructions 6.7L EGR DELETE BY FORBIDDEN DIESEL PERFORMANCE. Page 1

HP 6890Series Gas Chromatograph. Installation Guide. Flame Ionization Detector. Accessories HP G1561A HP G1562A HP G1591A HP G1598A

850 TETHER SWITCH KIT

Sentronic LP IO-Link CLASS A 617 Series. 617 Series IO-Link CLASS A with Display and Controls Installation Manual

2-row and All-row systems included.

Operation and Maintenance Instructions Manual BOOSTER. Volume amplifier

Transcription:

Installation Guide Auxiliary Pressure Control Manifold on 6890 GC Accessory G1570A

Agilent Technologies 2001 All Rights Reserved. Reproduction, adaptation, or translation without permission is prohibited, except as allowed under the copyright laws. Part number G1570-90307 First Edition, MAY 2000 Second Edition, APR 2001 Printed in USA Safety Information The Agilent Technologies 6890 Gas Chromatograph meets the following IEC (International Electrotechnical Commission) classifications: Safety Class 1, Transient Overvoltage Category II, and Pollution Degree 2. This unit has been designed and tested in accordance with recognized safety standards and designed for use indoors. If the instrument is used in a manner not specified by the manufacturer, the protection provided by the instrument may be impaired. Whenever the safety protection of the Agilent 6890 has been compromised, disconnect the unit from all power sources and secure the unit against unintended operation. Refer servicing to qualified service personnel. Substituting parts or performing any unauthorized modification to the instrument may result in a safety hazard. Disconnect the AC power cord before removing covers. The customer should not attempt to replace the battery or fuses in this instrument. The battery contained in this instrument is recyclable. Safety Symbols Warnings in the manual or on the instrument must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions violates safety standards of design and the intended use of the instrument. Agilent Technologies assumes no liability for the customer s failure to comply with these requirements. WARNING A warning calls attention to a condition or possible situation that could cause injury to the user. CAUTION A caution calls attention to a condition or possible situation that could damage or destroy the product or the user s work. # Indicates a hot surface " Indicates earth (ground) terminal Sound Emission Certification for Federal Republic of Germany Sound pressure Lp < 68 db(a) During normal operation At the operator position According to ISO 7779 (Type Test) Schallemission Schalldruckpegel LP < 68 db(a) Am Arbeitsplatz Normaler Betrieb Nach DIN 45635 T. 19 (Typprüfung) Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USA

Installing an Auxiliary Pressure Control Manifold Accessory G1570A The Auxiliary Pressure Control Manifold adds three electronicallycontrolled gas streams to an Agilent 6890 Series Gas Chromatograph (GC). This kit can be used to replace an existing Auxiliary Pressure Control Manifold or to install a new one. This kit contains: Kit G1570 64000 Qty. Auxiliary EPC flow control manifold 1 Mounting bracket, Agilent 6890 1 Top rear panel 1 Installation sheet (this document) 1 Blank label 1 Mounting bracket 1 Screw, M4 x 0.7 x 10 mm 1 Hex nut, 5/16-inch 20 thread 3 Auxiliary EPC flow restrictor kit 1 Aux EPC label 1 1

Mounting bracket Gang fitting installs here ID tag Ribbon cable Auxiliary EPC flow control manifold Slot for label tag Hex nuts Blank label Top rear panel PCB bracket Auxiliary EPC flow restrictor kit and Aux EPC label not shown Figure 1. Auxiliary EPC flow control manifold replacement kit Tools required 7/16-inch open-ended wrench T-20 Torx driver Needle-nosed pliers Safety information Before continuing, read the safety information in your GC Operating Manual. 2

Removing the covers Removing the covers WARNING Hydrogen gas is flammable and potentially explosive. Before replacing the manifold, turn off the hydrogen gas at the source. WARNING Before proceeding, turn off the oven and any heated zones and let them cool down. Turn off all detector gases at their supply, then turn off the main power switch and unplug the power cord. 1. Remove the pneumatics cover, the RFI shield under it, and the top rear panel. Pneumatics cover RFI shield Remove screw Top rear panel 2. Raise the gray plastic top cover (with the holes and ventilation slots) to the vertical position. Examine the hinge in the right rear corner. Early 6890 models. The hinge is a metal bracket attached to the oven top. Pull the clip at its top toward you to release the hinge pin. 3

Removing the covers Push the pin to the left to release the cover. Raise the right side of the cover and remove it. Discard the cover. Current 6890 models. Raise the right side of the cover and remove it. Discard the cover. 3. Remove the electronics cover and the right side cover. Detector cover Electronics cover Right side cover 4

Removing the existing manifold, if present Removing the existing manifold, if present 1. Remove the gas supply tubing from the present manifold. Disconnect gas lines OR Disconnect gas lines Manifold, before January 1999 Manifold, after January 1999 2. Remove the Torx T-20 mounting screw from the front of the manifold. Slot for tubing Mounting screw 3. Disengage the tubing from the slots in the chassis so that the gang fitting on the manifold can be removed easily. Caution Make sure you are properly grounded with an ESD strap before continuing. Caution Always hold the manifold by its support bracket to avoid damaging board components. 4. Unlock the manifold s ribbon cable from the pneumatics control board and detach the connector. The adjacent ribbon cable may have to be removed as well. 5

Removing the existing manifold, if present 5. Remove the one Torx T-20 screw holding the gang fitting on the manifold. Caution Do not lose the O-rings when you remove the gang fitting. Remove screw and gang fitting Manifold, before January 1999 Manifold, after January 1999 6

Replacing the PCB bracket Replacing the PCB bracket 1. Compare the PCB bracket under the pneumatics PC board with the one included in the accessory kit. Check for the differences in the three mounting screw tabs. If they are the same, proceed to the next section. If they are different, the existing bracket must be replaced with the new one. WARNING The rest of this procedure must be performed only by Agilent service personnel. Before proceeding, confirm that the main power switch is turned off and that the power cord is disconnected from the power source. 2. Remove the GC lower back panel. Caution Make sure you are properly grounded with an ESD strap before continuing. 3. If an MIO card is present, remove it and the jumper card. 4. Disconnect all cables from the pneumatics PC board. 5. Trace the ribbon cable on the end of the board to the main board and disconnect it there. Feed the cable up through the PCB bracket to the top. 6. Remove the screws holding the PCB bracket and slide it and the board out of the GC. 7. Remove the eight screws that fasten the board to the PCB bracket. Transfer the board to the new bracket and secure it with the eight screws. 8. Slide the board and bracket assembly back into the GC. Use the 10 mm screw provided in the kit for the left-most position, on the side of the GC. Use the screws removed earlier for the other positions. 9. Pass the ribbon cable down through the slot in the PCB bracket and connect it to the main board. Reconnect all cables to the pneumatics board. 10. Re-install the MIO card and the jumper card. 11. Re-install the lower back panel. 7

Installing the new manifold Installing the new manifold Caution Always hold the manifold by its support bracket to avoid damaging board components. 1. Slip the ID tag on the new manifold through the slot in the mounting bracket, then align the bracket holes over the gas fittings. Secure the bracket with three 7/16-inch hex nuts. Mounting bracket Gang fitting installs here Blank label Hex nuts Ribbon cable Label ID tag through slot in bracket 2. Peel the blank label from its backing and paste it on the mounting bracket over the screw heads. 3. If you need to change one or more auxiliary channel frits, do so now. See Changing the flow restrictors (optional) on page 13. 4. Insert the gang fitting through the cutout in the manifold bracket and install it onto the new manifold assembly so that the tubing runs back and away from the fitting. 8

Installing the new manifold Be sure that the three O-rings are in place. Be sure the left tube clears the inner edge of the bracket. Tighten the screw on the gang fitting until the gang fitting touches the manifold. Route tubing along this path Check for interference Bracket is flush with carrier rails Attach gang fitting 5. Route the ribbon cable to the right side of the manifold. Then slide the manifold and bracket assembly into the Aux slot until the bracket seats flush against the end of the rails. Ribbon cable 6. Route the gas tubing behind the manifold, over the top of the chassis, and through the slots. 7. Connect the ribbon cable to the mating connector on the pneumatics board. This is behind the connector for the back detector and faces up. 8. Secure the manifold in place using the Torx T-20 mounting screw. 9

Installing the new manifold 9. If the detector cable is in the way, remove it temporarily while you connect the Aux cable. Arrange the cable to keep it away from the valves and keep it from being pinched against the manifold. Auxiliary manifold connector Back detector manifold connector 10. Using a pair of needle-nosed pliers, remove the top rear panel cutout for Auxiliary. Also remove any cutouts needed to access other manifolds or accessories installed in the GC. Insert tip of pliers here Back inlet Front detector Front inlet Auxiliary Left-most screw slot Back detector 11. Place the new top rear panel on its left-most mounting screw. Use the screw as a hinge and angle the panel while sliding each manifold ID tag through its cutout in the panel, working from left to right. When all the tags are through the panel, finish installing the panel on the GC. 10

Installing the new manifold 12. Attach the Aux EPC label to the top of the pneumatics chassis. Attach label here Gas outlet block 13. Install the RFI shield, the pneumatics cover, and the detector top cover. 11

Restore the GC to operating condition Restore the GC to operating condition 1. Plug in the GC and turn it on. 2. Press [Aux #][3], [Aux #][4], or [Aux #][5] to access a channel. If your manifold is correctly installed, you will see this display: PNEUMATIC AUX 3 Pressure 0.x Off Init time 0.00 Rate 1 (off) An actual flow value is displayed when the gases are off or not connected. This is not an error. After the gases are connected and the detector is operational, the actual flow values will be equal to the setpoint values. 3. If the display reads Aux not installed, recheck your cable connections. 4. Zero the pressure sensors: a. Make certain that no gases are connected to your manifold. b. Press [Options] and scroll to Calibration Aux pressure. c. Scroll to Aux 3 zero and press [On]. d. Scroll to Aux 4 zero and press [On]. e. Scroll to Aux 5 zero and press [On]. 5. Connect the source gas lines to the manifold. 6. Restore gas pressures and leak check all fittings. 12

Changing the flow restrictors (optional) Changing the flow restrictors (optional) The auxiliary channels are controlled by a pressure setpoint. To work properly, there must be adequate flow resistance downstream of the pressure sensor. The auxiliary channel pneumatics manifold provides a frittype restrictor for each channel. Four frits are available: Frit marking Flow resistance Agilent Part Number Blue Dot High 19234-60660 Red Dot Medium 19231-60770 Brown Dot Low 19231-60610 None (brass tube) Zero G1570-20547 The Red Dot frit is in all three channels when the instrument is shipped. Figure 1 and Figure 2 show approximate pressure/flow relationships for the three Dot frits, assuming there is no significant additional resistance downstream of the frits. If the Zero resistance frit is installed, the user must provide flow resistance downstream and generate the pressure/flow relationships. WARNING When hydrogen is used, dangerously high flows are possible if insufficient flow resistance is provided downstream of the supply tube. Always use either the High (Blue Dot) or Medium (Red Dot) frit with hydrogen. 13

Changing the flow restrictors (optional) 80 (Ambient conditions: 25 C, 14.7 psia) 70 19234-60660 (blue dot) 60 Minimum source pressure (psig) 50 40 19231-60770 (red dot) 30 19231-60610 (brown dot) 20 10 0 0.1 1.0 10.0 100.0 1000.0 Gas Flow in ml/min Figure 2. Pressure requirements for AUX EPC flow restrictors with air, nitrogen, or helium 14

Changing the flow restrictors (optional) 80 (Ambient conditions: 25 C, 14.7 psia) 70 60 19234-60660 (blue dot) Minimum source pressure (psig) 50 40 19231-60770 (red dot) 30 20 19231-60610 (brown dot) 10 0 0.1 1.0 10.0 100.0 1000.0 Gas Flow in ml/min Figure 3. Pressure requirements for AUX EPC flow restrictors with hydrogen 15

Changing the flow restrictors (optional) Changing an auxiliary channel frit 1. Locate the block that connects the three gas outlet tubes for the auxiliary channels to the pneumatics module. Gang fitting block 2. Remove the screw that holds the block to the pneumatics module. Pull the block free of the module and rotate it so that the frits are on top. Aux 4 Gas restrictors Aux 5 Aux 3 Aux 3 Aux 5 Aux 4 Gas outlets 3. Pull the frit to be changed out of the block. Also remove the O-ring that seals it. 4. Place an O-ring on the new frit. Place the O-ring/frit combination in the block. 5. Reconnect the block to the pneumatics module. Tighten the screw firmly. 16

Printed on recycled paper. This product is recyclable. Agilent Technologies, Inc. Printed in USA APR 2001 G1570-90307