CPM280. Technical Information. ph/orp measurement Battery-driven pocket meter for ph and ORP measurement

Similar documents
Conductivity Measurement CLM 280

Orbipore CPS91 and CPS91D

Orbisint CPS11 and CPS11D

Orbisint CPS12/CPS12D/CPS13

Technical Information Ceraliquid CPS41D and CPS41

Orbisint CPS11 and CPS11D

Ceraliquid CPS42/CPS42D/CPS43

Ceraliquid CPS41 and CPS41D

Flowfit CPA250. Technical Information. Flow assembly for ph and ORP sensors

Cleanfit W CPA450. Technical Information. Retractable assembly for 12 mm sensors for DO/pH/ORP measurement

Technical Information

Cleanfit P CPA472D. Technical Information. Retractable process assembly for ph, ORP and other industry sensors Heavy-duty version

Cleanfit CPA471. Technical Information. Compact retractable assembly for installation of ph, ORP electrodes in tanks or pipelines

Cleanfit H CPA475. Technical Information. Retractable assembly for ph, ORP measurement under sterile conditions

Technical Information Cleanfit CPA472D

Signet DryLoc ph/orp Electrodes

Signet ph/orp Electrodes

Signet DryLoc ph/orp Electrodes

Signet ph/orp Electrodes

Personal ph, ph/orp, Conductivity Meters

GHMsilverline Handheld Instruments

PH60 Series Premium ph Testers

Signet ph/orp Electrodes

Flowfit CYA251. Technical Information. Flow assembly for nitrate/sac, turbidity and oxygen sensors

Operating Instructions. Flowfit W CPA250. Flow assembly for ph/orp sensors. BA041C/07/en/

SPT- I / U PRESSURE TRANSMITTER for general applications

AP300 Series Combination ph/redox (ORP) Sensors. Measurement made easy. A reliable economic solution for a wide range of applications

Rod probe 11375Z. Technical Information. Conductive level limit detection Partially insulated rod probe for use in conductive liquids

* * Rev. 9 01/18 Product Manual

JUMO tecline HD ph and JUMO tecline HD Rd

Signet 2760 DryLoc ph/orp Preamplifier

Signet 3719 ph/orp Wet-Tap

General Purpose ph/orp Sensors

Soliswitch FTE30. Technical Information

Instruction Manual Portamess 911 Cond

Signet 3719 ph/orp Wet-Tap Assembly

Indumax P CLS50. Technical Information. Highly resistant, inductive conductivity sensor for standard, hazardous and high-temperature applications

HI 8510-HI 8512 HI 8710-HI 8711 HI 8720-HI HI HI

ortable ph, Conductivity & DO2 Meters

ABB MEASUREMENT & ANALYTICS DATA SHEET. AP200 Rugged ph/redox (ORP) sensor systems with rapid temperature response for critical processes

Bante220 Portable ph Meter Instruction Manual

ph-kompaktelektrodensystem sensopac CPA 320

Technical Information Cleanfit CPA451

SIGNET 2750 ph/orp Sensor ENGLISH

RTD Temperature Sensor omnigrad T -TR 470

Portable pressure calibrator, Wally Box III Model CPH7600

Technical Information Omnigrad TST90, TET90

Technical Information Cleanfit CPA871

Rugged ph/redox (ORP) Sensor Systems with Rapid Temperature Response for Critical Processes

General Purpose ph/orp Sensors

Analysis ph / ORP. Characteristics System. Applications Food industry Chemical and Pharmaceutical industry Water and Sewage treatment

DL 01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05

Digital ph transmitter

SensoLyt 700 SensoLyt 700 SW

Soliswitch FTE20. Technical Information. Point level switch for granular solids

Encapsulated Differential ph and ORP Sensors (available in LCP, foreground, or Ryton body materials)

Table of Contents 1. OVERVIEW OPERATION FUNCTION SPECIFICATIONS INSTRUMENT LAYOUT INSTALLATION...

ABB MEASUREMENT & ANALYTICS DATA SHEET series ph and Redox (ORP) industrial dip electrode systems

Conductivity Measuring Cells ConduMax W CLS 19

Analysis ph / ORP. Applications. Characteristics. Food industry Chemical and Pharmaceutical industry Water and Sewage treatment

PC60 Premium Multi-Parameter Tester (ph/ec/tds/salinity/temp.) Instruction Manual. APERA INSTRUMENTS, LLC V6.4

Flow Fittings. for Electrochemical Sensors. B Operating Manual /

SIGNET 2760 DryLoc ph/orp Preamplifier

ecotrans Lf 01/02 Microprocessor Transmitter / Switching Device for Conductivity

Digital ph transmitter

Conductivity Sensors ConduMax W CLS 19

Stamolys CA71SI. Technical Information

Flow Fittings. for Electrochemical Sensors. B Operating Manual /

SIGNET 3719 ph/orp Wet-Tap

Environmental/Analytical LabCivil Engineering Lab Equipment

JUMO ecotrans Lf 01/02

Instruction Manual. ph Meter D-51

DM01. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

JUMO tecline H2O2 JUMO tecline PAA Sensor for hydrogen peroxide and peracetic acid

HI Portable Waterproof Skin ph Meter INSTRUCTION MANUAL

User Manual Pocket Pro, Pro + ph

Stamolys CA71AM. Technical Information

Rosemount 8750WA Magnetic Flowmeter System For Water and Wastewater Industries

NEW QUICK CABLE-TO-SENSOR RELEASE,

ABB MEASUREMENT & ANALYTICS DATA SHEET Hydrazine monitor

General Specifications

Type MZ15. Operating Instructions. Bedienungsanleitung Manuel d utilisation

Screw-in resistance thermometers with plug connector

PHscan30S Pocket ph Meter Instruction Manual

Immersion fittings. Brief description. Installation options Standard. Series Page 1/7. Data Sheet

TITREX. One instrument for all your titration and dosing applications

Visit us at

General Specifications

Signet 2760 DryLoc ph/orp Preamplifiers & Connectors

Titrette. class A precision. The next generation Digital Burette!

Operating Instructions Dipfit CPA111

General Specifications

N E W! The next generation Digital Burette.

testo 206 ph/ Temperature Measuring Instrument Bedienungsanleitung Instruction Manual

to mv Redox measurement accuracy. 173 x 195 x 103mm / 1.5kg Display

ELPRO Lepenik & Co. d.o.o., Ob gozdu 7c, Rogoza, 2204 Miklavž na Dravskem polju, Slovenija tel:

ORPscan20 Pocket ORP Tester Instruction Manual

Combination ph/redox (ORP) Sensors

Optidew Vision Optical Dew-Point Meter

Combination ph/redox (ORP) Sensors

Transcription:

Technical Information CPM280 ph/orp measurement Battery-driven pocket meter for ph and ORP measurement Application CPM280 is a pocket meter for measuring ph value and ORP. It is suitable for the following fields of application: Wastewater treatment plants Drinking and surface water Laboratory and field measurement Your benefits Comfortable operation: Clear multi-function display with user guidance Precise measurement: Automatic temperature compensation via the ph combination electrode with built-in temperature sensor (NTC 30) Flexible calibration: Conventional calibration with two buffer solutions Fully automatic two-point or three-point calibration with pre-defined buffer solutions according to DIN 19266 Fully automatic two-point or three-point calibration with Endress+Hauser quality buffer solutions Safe and cost-effective operation: Long service life with four commonly used 1.5 V batteries: approx. 3000 h Automatic power-down Dust and water-proof housing (IP 66) TI346C/07/en/11.08 71084903

Function and system design Measuring principle Measuring system! Note! ph measurement The ph value is used as a unit of measurement for the acidity or alkalinity of a liquid medium. The membrane glass of the electrode supplies an electrochemical potential which is dependent upon the ph value of the medium. This potential is generated by the selective penetration of H + ions through the outer layer of the membrane. An electrochemical boundary layer with an electric potential forms at this point. An integrated Ag/AgCl reference system serves as reference electrode. The transmitter converts the measured voltage into the corresponding ph value using the Nernst equation. Redox measurement The redox potential is a unit of measurement for the state of equilibria between oxidising and reducing components of a medium. Redox potential is measured similarly to the ph value. A platinum or gold electrode is used instead of ph-sensitive membrane glass. Analog to the ph measurement, an integrated Ag/AgCl reference system is used as a reference electrode. A complete measuring system comprises: CPM280 pocket meter Norylab CPS51 ph combination electrode with built-in temperature sensor NTC 30, fixed cable and water-proof DIN plug Plastic bowl for sampling You can use the supplied stand as a sensor holder. 1 2 Measuring system 1 Pocket meter 2 Sensor cable 3 Sensor 4 Plastic bowl 4 3 a0008900 2 Endress+Hauser

Input Measured variable Measuring range ph value or ORP [mv] ph value: -2 to 16 ORP: -1999 to +1999 mv Temperature: -5 to 100 C (20 to 212 F) Power supply Electrical connection via fixed cable and DIN plug Ref 1 2 3 Electrical connection 1 ph sensor connection 2 Temperature sensor connection 3 Reference electrode connection a0008105 Supply voltage Battery power Service life 4 x 1.5 V alkaline-manganese batteries, type AA approx. 3000 h Performance characteristics Reference operating conditions Reference temperature: Reference pressure: 25 C (77 F) 1013 hpa (15 psi) Maximum measured error Temperature compensation ph value: ORP: Temperature: -5 to 80 C (20 to 180 F) 0.01 ph ± 1 digit 0.3 mv ± 1 digit 0.1 K ± 1 digit Endress+Hauser 3

Environment Storage temperature 5 to +65 C (20 to 150 F) Ingress protection IP 66 Climate class climate class 2 Electromagnetic compatibility Interference emission acc. to DIN EN 50081-1 / 50081-2 Process Process temperature -5 to 80 C (20 to 180 F) Design, dimensions pocket meter Mechanical construction 80 (3.15) 38 (1.5) 80 (3.15) 172 (6.77) 38 (1.5) 35 (1.38) mm (inch) a0008905 4 Endress+Hauser

Design, dimensions case and sensor mm (inch) 355 (14.0) 115 (4.53) 305 (12.0) 120 (4.72) mm (inch) 12 (0.47) Case dimensions a0008909 Sensor dimensions a0008908 Weight Pocket meter Sensor Case, total weight 0.4 kg (0.9 lbs) 0.1 kg (0.2 lbs) 2.2 kg (4.9 lbs) Materials Pocket meter Housing: Keypad: Sensor Membrane glass: polypropylen (PP) silicone UNITEC Human interface Display elements 1 2 3 1 2 3 4 5 Measured value Status display Sensor symbol Temperature Displayed functions 5 4 LC display a0008484 Displayed functions LoBat TP AutoCal DIN AutoCal TEC CalError ARng AR Low battery Symbol of the temperature display Automatic calibration function, buffer solutions acc. to DIN 19266 Automatic calibration function, Endress+Hauser quality buffer ph 4, ph 7 Calibration error Autorange function on Autorange function off Endress+Hauser 5

Operating elements M CAL I AR M I CAL AR RUN/ ENTER Operation mode On/off Calibrate Autoread on/off (Hold function) Run/Enter RUN/ENTER Keypad a0008490 Ordering information Order number Scope of delivery ph/orp pocket meter CPM280 order no. 51510538 The measuring set comprises: 1 CPM280 pocket meter 1 Norylab CPS51 ph combination electrode with built-in NTC 30 temperature sensor, 1.5 m (4.9 ft) fixed cable and water-proof DIN plug 2 plastic bottles with buffer solutions ph 4 and ph 7, 50 ml (1.7 fl.oz) each 1 plastic bottle with electrolyte solution 3 mol/l KCl, 50 ml (1.7 fl.oz) 1 plastic bowl 50 ml (1.7 fl.oz) 1 stand used as electrode holder 4 alkali-manganese batteries 1.5 V 1 solid plastic case made from PP with integrated measuring point and separate pocket for sensor and accessories 1 Operating Instructions BA346C/07/b4 6 Endress+Hauser

Accessories and spare parts Sensor Calibration solutions Orbisint CPS51/CPS52 ph resp. ORP sensor with plastic shaft for pocket meter with built-in temperature sensor Order nos: CPS51-7EK2FDA (ph sensor) CPS52-0PK2FDA (ORP sensor) Technical buffer solutions for ORP electrodes +220 mv, ph 7.0, 100 ml (3.4 fl.oz.); order no. CPY3-0 +468 mv, ph 0.1, 100 ml (3.4 fl.oz.); order no. CPY3-1 High-quality buffer solutions of Endress+Hauser The secondary buffer solutions have been referenced to primary reference material of the PTB (German Federal Physico-technical Institute) and to standard reference material of NIST (National Institute of Standards and Technology) according to DIN 19266 by a DKD (German Calibration Service) accredited laboratory. ph value A ph 2.00 (accuracy ± 0.02 ph) C ph 4.00 (accuracy ± 0.02 ph) E ph 7.00 (accuracy ± 0.02 ph) G ph 9.00 (accuracy ± 0.02 ph) I ph 9.20 (accuracy ± 0.02 ph) K ph 10.00 (accuracy ± 0.05 ph) M ph 12.00 (accuracy ± 0.05 ph) Quantity 01 20 x 18 ml (0.68 fl.oz) only buffer solutions ph 4.00 and 7.00 02 250 ml (8.45 fl.oz) 10 1000 ml (0.26 US gal) 50 5000 ml (1.32 US gal) canister for Topcal S Certificates A Buffer analysis certificate Version 1 Standard CPY20- complete order code Endress+Hauser 7