Dental Vacuum System. Installation Overview

Similar documents
Microprocessor Controlled Dental Vacuum System Installation Manual

Section. Vacuums. 866-DTE-INFO 147

PRE-INSTALLATION GUIDE

PowerAir Compressor Installation

Dry Vacuum System. Part Numbers: V15, 2V15, 3V15 and 4V15 USER S MANUAL. Recommend Daily use of Monarch CleanStream

ClassicSeries Compressor Installation

Installation Guide Smart-UPS On-Line SRT1000/1500 UXI-NCLI, SRT1000/1500 UXI-LI, Tower/Rack-Mount

User s and Installation Manual

Start UP Guide. Symmetra LX Tower Rack-Mount. UPS Models 200 V, 4-8 kva 208/240 V, 4-8 kva 220/230/240 V, 4-8 kva

Installation Guide Smart-UPS On-Line SRT1000/SRT1500 XLA Tower/Rack-Mount

Patient Care Facility Multiple Room Configuration Style 3 W/ Ver 4.41 pcb & Pulse Relay pcb(s)

Matala. VersiFlow Series. Instruction and Maintenance Manual

Operating Manual. High Performance Vacuum Pump Models and 15600

PowerMax Surgical Suction Producers

PRE-PLUMBED SEWAGE SYSTEM

Ground Mount / Piping / Electrical Quick Reference ECS Membrane Processor: PMC and ISD

ACC Series Power Conditioner OPERATION & INSTALLATION MANUAL

Air Handling Unit Construction Checklist

Procenter Delivery, Console Mounted on an Elevance Chair Installation Guide

Dental Dry Suction. Installation/User Guide

INSTALLATION INSTRUCTIONS

Electric Vehicle Charging Station

Users Manual Certified Series Direct Drive Pump 1-7 LPM

Installing Power Components

A-dec 311 Dental Chair and Related Systems

Cylindrical Silencers

Advantage-D. Operating Instructions and Maintenance Manual. Central Vacuum Systems (Expandable/Modular Models) (Ver.

High Frequency SineWave Guardian TM

GeyserMax-Flow Series

Active Evacuation System (AES)

INSTALLATION GUIDE AND USER MANUAL

SineWave Guardian TM 380V 600V INSTALLATION GUIDE. Quick Reference. ❶ How to Install Pages 6 17 ❷ Startup/Troubleshooting Pages WARNING

Instruction & Owner s Manual

Inst all a ti o n and Oper a tin g Instru c ti o n s

INSTALLATION INSTRUCTIONS

Generator Set Configurations Low Fuel Consumption, EU Stage IIIA Nonroad Emission Standards, China Nonroad III Emission Standards

SOS SERIES SOS1 SOS2. Spares On Site Battery Cabinet Installation Guide rEV3

XPC-EBP64 External Battery Pack User & Installation Manual Xtreme Power Conversion Corporation. All rights reserved.

Installation Manual. Q Series Quadplex Grinder Systems. Contents

Installation Manual. D3600- Series Duplex Grinder Systems. Contents

Electric Reciprocating Compressor. Installation Guide

1200W INVERTER GENERATOR

GE Monogram. Installation. Instructions. Professional Vent Hoods. 48" Models ZV881WSS, ZV891YSS ZV880WSS, ZV890YSS. 36" Models ZV671WSS, ZV681YSS

dv Sentry TM 208V 600V INSTALLATION GUIDE Quick Reference ❶ How to Install Pages 6 14 ❷ Startup/Troubleshooting Pages WARNING

SlimRack Bed Lift System OEM INSTALLATION MANUAL

27.6 Vdc 1 Amp Switch Mode Power Supply for Fire EN54-4:1997 +A1 +A2

User Manual. T6 Tachometer. Online: Telephone: P.O. Box St. Petersburg, Florida 33736

Engine Specifications 3512C HD ATAAC, V-12, 4-Stroke Water-Cooled Diesel Compression Ratio 14.7:1 Aspiration

Asymmetrical Installation Instructions. Components: i2cove Asymmetrical LED Light Fixtures. 12 [305mm] [918mm] 48.

Pre-Installation Guide Performer

ESE Series Cast Iron Sewage Pumps

SECTION MOTOR REQUIREMENTS for HVAC

HANGKAI GROUP HOYMILES MICRO-INVERTER MI-250

Matrix AP 400V 690V INSTALLATION GUIDE. Quick Reference. ❶ How to Install Pages 6 20 ❷ Startup/Troubleshooting Pages WARNING

Self Cleaning Hood System Installation, Operation, and Maintenance Manual

Flow Max Fluid Pump. by Duraself OEM INSTALLATION MANUAL

Enclosed Electric Rotary Screw Compressor Installation Guide

EVoInnovate EVSE Model Numbers: EVO & EVO User Manual & Installation Guide. Revision 1.0

BATTERY SAVER LOW RIPPLE HO

SEWAGE PUMP MODEL # Zoeller is a registered trademark of Zoeller Co. All Rights Reserved. Español p. 14

Instruction Manual. Blink HQ Charger. Charge on. a CarCharging Company

High Performance Vacuum Pump Model 15120A/15121A Operating Manual...

User Manual. Solar Charge Controller 3KW

BroadBand PowerShield. 20 AHr Battery. User Manual

INSTRUCTION & INSTALLATION

FLEX VANE IMPELLER MODEL 3000-XXX Replacement Parts Kit Instructions

F-4600 INLINE ULTRASONIC FLOW METER Installation and Operation Guide

Quick Engineering Facts Decimal Metric

Raritan PX. Power Cord Installation Guide. Safety Guidelines. Safety Instructions

SUBMERSIBLE PUMP HSE RANGE

FieldView External Charger. User Guide

Mini-Roto TM Paddle Wheel Installation & Operation Manual

CV41. Vehicle Mounting Kit. Reference Guide

AUTOMATIC BEST BATTERY SELECTOR INSTALLATION & OPERATION BBS-4800 BBS-4800E

Electrical Installation

Section 02: Pre-Installation Procedures

SUBMERSIBLE PUMP HSE RANGE

Podiatry Procedures Chair. For Models: Barrier-Free. User s Guide

Pneumatic Isolation Module

Utility Distribution Systems Installation, Operation, and Maintenance Manual

STANDBY 200 ekw 250 kva

Max-Upender/Tilt Series Service Manual

C T h e A d va n t a g e

Infinitybox Express Road Race Car Kit Installation Guide

2460 / 2472 / Series Simplex Grinder Systems

Models

SECTION ENCLOSED SWITCHES AND CIRCUIT BREAKERS

Specifications. C13 Generator Set Electric Power

INSTALLATION GUIDE AND USER MANUAL

Model 8144NC Fresh Air Ventilator

Installation Operation and Maintenance Manual

AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST

DRAIN PUMP SYSTEMS MODELS 104*, 105, 110, 115, 120, 131

CALTRAP INSTALLATION AND OPERATIONS MANUAL

WARNING: FAILURE TO FOLLOW THESE RULES MAY RESULT IN SERIOUS PERSONAL INJURY CAUTION: INSTALLATION LOCATION:

1500 or 1800 RPM. Engine Specifications C18 ATAAC, I-6, 4-Stroke Water-Cooled Diesel 145 mm (5.71 in) 183 mm (7.2 in) Compression Ratio 14.

GENERAL INFORMATION. H-1649, H-1650, H-1651 H-1653, H-1654 Easy-Count. uline.com. that may hurt accuracy:

220 to Volts. Generator Set Configurations U.S. EPA Certified for Stationary Emergency Use Only (Tier 2 Nonroad Equivalent Emission Standards)

Bore mm 4.35 in Displacement 8.8 L 535 in³ Stroke mm 4.5 in Compression Ratio 10.1:1

Technical Specification

Transcription:

Dental Vacuum System Installation Overview This guide contains site preparation and installation instructions for RAMVAC Bulldog QT and Bison Dental Vacuum Systems with S1 Electrols. It is recommended that this equipment only be installed by licensed electrical and plumbing contractors. Step 1 Site Requirements Step 2 Tank Plumbing Step S-Type Exhaust Filter Standard on Bison 7 and 9. Optional on Bison 3, 5 and all Bulldog models. 3 Exhaust Step 4 Add OIl Step 5 Remote Switching Step 6 Electrical Power Step 7 Installation Checklist (Separate Document) A Brand of the RAMVAC 212 N. Main Street Spearfish, SD 57783 (800) 5-RAMVAC (800) 572-6822 (866) DTE-INFO Fax: (605) 642-3776 ramvac@ramvac.com www.ramvac.com www.dentalez.com Effective: 17-Sep-13

Dental Vacuum System Installation TABLE OF CONTENTS Installation Overview 1 Table of Contents 2 Technical Description 3 Step 1 Bulldog Site Requirements 4 Step 1 Bison Site Requirements 5 Step 2 Tank Plumbing 6 Step 3 Exhaust Plumbing 7 Step 4 Add Oil 8 Step 5 Remote Switching Options 9 Step 6 Electrical Ratings 10 Step 6 Single Phase Installation 11 Step 6 Three Phase Startup 11-12 Step 7 Installation Checklist 13 SEE USER MANUAL FOR: Safety and Regulatory Information Operating and Shipping Conditions EMC Information WARNING No modification of this equipment is allowed. Modification will void the manufacturers warranty. Page 2

Dental Vacuum System Installation Technical Description The RAMVAC Bulldog QT and Bison Vacuum Systems utilizes an oil lubricated, rotary vane, positive displacement pump to provide a reliable vacuum source for the Dental office. This vacuum unit is intended solely for the removal of dental waste material from the oral cavity during dental procedures. The pump is powered via a belt drive system by an electric motor that is controlled by the S2 Control System. The S2 Control is an electronic, smart control that utilizes a microprocessor and sensors to monitor the vacuum system and provide alarms or reminders of system needs. A drip oiling system meters lubricating oil into the pump and is then removed from the airflow by filters inside the oil reservoir, creating a closed loop system. Page 3

Bulldog QT Site Requirements Comply with Local and National Electrical and Plumbing Codes. Follow NFPA 99c Edition, Level 3 Vacuum System Recommendations. Ambient Temperature 32 to 104 F (0 to 40 C) Protect from Water. Controls are NOT waterproof. Motors are NOT Waterproof. Humidity 0% to 95% No Condensing Moisture Common Installations Shown Contact RAMVAC for more options It is recommended that this equipment only be installed by licensed electrical and plumbing contractors. Bulldog - 15 Gallon Otter Tank stacked Bulldog - 15 Gallon Otter Tank suspended Bulldog -15-Gallon Otter Tank side by side Product Dimensions 15-Gallon Otter Tank Bulldog QT 1 & 2 Bulldog-15 Gallon Otter Tank Stacked Nominal Width 25 (640 mm) 21 (520 mm) 25 (640 mm) Nominal Depth 19 (490 mm) 17 (430 mm) 19 (490 mm) Nominal Height 24 (610 mm) 28 (640 mm) 44 (1120 mm) Nominal Height with S-Type Exhaust Filter N/A 48 (1220 mm) 67 (1700mm) Drain valve centerline to floor 8 (203 mm) N/A 8 (203 mm) Add 6 (155 mm) to each dimension for working space Page 4

Bison Site Requirements Comply with Local and National Electrical and Plumbing Codes. Follow NFPA 99c Edition, Level 3 Vacuum System Recommendations. Ambient Temperature 32 to 104 F (0 to 40 C) Protect from Water. Controls are NOT Waterproof. Motors are NOT Waterproof. Humidity 0% to 95% No Condensing Moisture Common Installations Shown Contact RAMVAC for more options It is recommended that this equipment only be installed by licensed electrical and plumbing contractors. S-Type Exhaust Filter * Standard on Bison 7 and 9. Optional on Bison 3, 5, and all Bulldog models. Bison - 50 Gallon Tank side by side Bison - 30 gallon Otter Tank side by side Bison - 30-Gallon Otter Tank stacked using Bison Platform Product Dimensions 30-Gallon Otter Tank Bison 3 & 5 Bison - 30 gallon Otter Tank Stacked Bison 7 & 9 50-Gallon Cylindrical Tank Nominal Width 29 (740 mm) 26 (660 mm) 32 (820 mm) 26 (660 mm) 22 (560mm) Nominal Depth 21 (540 mm) 19 (490 mm) 21 (540 mm) 22 (560 mm) 22 (560mm) Nominal Height 36 (920 mm) 30 (760 mm) 70 (1780 mm) 30 (765 mm) 62 (1580mm) Drain valve centerline to floor 8 (203 mm) N/A 8 (203 mm) N/A 5 (130mm) * For S-type Exhaust Filters add 26 inches to height of Power Unit. Add 6 (155 mm) to each dimension for working space. Page 5

Tank Plumbing Facility Vacuum Lines Build Like a Drain Line: - Use DWV Fittings - Minimum slope ¼ per 10 ft (2 mm per 1m) Up to 5 rooms: 1-1/2 (or larger) PVC Schedule 40 or equivalent 5 to 15 rooms: 2 PVC (or larger) Schedule 40 or equivalent If an inline low spot is unavoidable, place it in a known location and incorporate a clean-out. For overhead piping, contact RAMVAC For nitrous oxide scavenging, contact RAMVAC Follow NFPA 99c Edition, RAMVAC Instructions, and local codes. Tank to Vacuum Power Unit Bulldog - 1 1/2 PVC Schedule 40 with DWV fittings or equivalent. Bison - 2 PVC Schedule 40 with DWV fittings or equivalent. Use new, clean pipe to connect tank to Vacuum Power Unit. Do NOT reuse old or exsisting pipe. Vent Trap Vent Drain Branch To Sewer Trap Drain Line Options NFPA 99c Edition Slide Valve Drain Check Valve Waste Pipe Bulldog or Bison (Bison Shown) To Sewer Vent Drain Branch Trap Trap Vent Multiple Tank Setup Guide is packed with Tanks Tank Drain Line Waste Pipe, Slide Valve, and Drain Check Valve - 1-1/2 PVC Schedule 40 with DWV fittings. - Min. slope ¼ per 10 ft (2 mm per 1m). - Use only RAMVAC supplied 1 1/2 drain check valve. Trap and Drain Branch - Trap and branch line 2 sizes larger than waste pipe. - Trap seal minimum 4 deep. Vents - Trap vent size no less than half the trap and drain branch size. - Vents no smaller than drain check valve. - Vents must extend minimum of 6 above tank top before turning horizontal. Follow NFPA 99c Edition, RAMVAC Instructions, and local codes. Page 6

Exhaust Plumbing BIOHAZARD Breathing Dental Vacuum System Exhaust May Be Harmful. Exhaust to Outside of Building. Follow NFPA 99c, RAMVAC Instructions, and local requirements. Minimum Exhaust Length = 15 feet from RAMVAC pump to outside end. Contact RAMVAC for shorter distances. Outside End (Installer supplied) Turn outside end down toward ground and screen to prevent entry of water, debris, and creatures. Locate inconspicuously for sight and sound. Locate to prevent exhaust from entering building. Outside Wall Roof Line Run Dedicated Exhaust Lines for each RAMVAC unit installed: DO NOT connect to plumbing vents. Size: 2 Type: Bulldog QT - PVC Sch 40, Sch 80, CPVC, or metal (NOT tar coated). Bulldog QT with vacuum set stronger than 9 Hg, use Bison type. Bison - Sch 80, CPVC, or metal (NOT tar coated). (Option: PVC Sch 40 or equivalent after first 10 feet of run). Support Loosely: DO NOT allow piping to touch ductwork. DO NOT make solid attachment to building structure. Seal: Slope: All joints must be oil tight. All horizontal lines toward the vacuum unit, a standard plumbing slope. Rubber Roofs: Terminating an exhaust run over a rubber roof is not a recommended practice. If exhaust termination above a rubber roof is necessary, protect the roof from contact with the exhaust oil vapor or droplets. 1. Terminate exhaust pipe a sufficient distance from rubber roof, or 2. Cover the rubber roof surface with material not sensitive to oil, and 3. Install a RAMVAC S-Type Exhaust filter. Page 7

Add Oil BULLDOG Remove oil filler cap. Add 3 quarts Mobil 1, 15w50. Check Oil Fill Level at sightglass and add or drain oil accordingly. (see image below). Use only recommended oil, available locally and also available from RAMVAC. Securely install oil fill cap. Check oil drip rate. See Check Oil Drip Rate Dispose of used oil at a gas station or lubricant recycling station. 53542 To read correctly, get eye level with sight glass while unit is sitting level, and check that oil level is even with the head of the red arrow. Add or drain oil accordingly. BISON Remove oil filler cap. Add 5 quarts Mobil 1, 15w50. Check Oil Fill Level at sightglass and add or drain oil accordingly. (see image below). Use only recommended oil, available locally and also available from RAMVAC. Securely install oil fill cap. Check oil drip rate. See Check Oil Drip Rate Dispose of used oil at a gas station or lubricant recycling station. To read correctly, get eye level with sight glass while unit is sitting level, and check that oil level is even with the head of the red arrow. Add or drain oil accordingly. Page 8

Remote Switching Options The RAMVAC can be run continuously throughout the workday. To avoid wasting electricity, turn off the RAMVAC if vacuum will not be needed for an hour or more. Note: The tank will drain only when no vacuum is present Vacuum must be turned off at least once per day! Illuminated Remote Panel RAMVAC OWL Touch Non-Illuminated remote Switches Switch light is steady-on when system is running normally. Switch light flashes for maintenance or one of the heads has been disabled by the disable button on the C2 Control. Recommended Switching Touch Pad illuminates while equipment is running. OWL gives complete breakdown of data on selected equipment Non-illuminated switches provide no indication for system status. RAMVAC Remote Panel or equivelent with illuminated 24VDC Switch 1. 2. 3. F. G. H. S2 Control Terminals OWL Connection Com Low Voltage Switch Connection FGH OWL Hub RAMVAC OWL Touch S2 Control Alternative Switching Toggle Switch or other non-illuminated switch (option) Air Switch Air Light Common Illuminated Remote Switch F. H. F - +24 VDC G - +24 VDC H - DC Common S2 Control Terminals Note: Maximum wire length for low voltage 18 gauge wire : 500 feet Note: High Voltage switching is an option but not recommended. Contact RAMVAC. Note: OWL to S2 connection must be made with Cat 6 shielded cable using RJ45 connectors. Page 9

Electrical Ratings WARNING TO AVOID RISK OF ELECTRIC SHOCK, THIS EQUIPMENT MUST ONLY BE CONNECTED TO A SUPPLY MAINS WITH PRETECTIVE EARTH. 60 Hz Models MODEL MOTOR THHN Wire Size 1 PROTECTION 2,3 Hp (kw) Phase Voltage 4 Max Amps 50 ft. 100 ft. Breaker 2 Fusetron 3 Bison 3, 5, 7 60 Hz 3.0 (2.2) 1 230v 16.0 A 10 ga. 10 ga. 20 amp FRN 25 230v 8.4 A 12 ga. 12 ga. 15 amp FRN 12 3 460v 4.2 A 12 ga. 12 ga. 15 amp FRS 5 Bison 9 60 Hz Bulldog QT 1, 2 60 Hz 50 Hz Models 5.0 (3.7) 3 1.0 (0.7) 1 230v 13.2 A 12 ga. 12 ga. 20 amp FRN 25 460v 6.6 A 12 ga. 12 ga. 15 amp FRS 10 115v 13.4 A 12 ga. 10 ga. 20 amp FRN 25 230v 6.7 A 12 ga. 12 ga. 15 amp FRN 8 MODEL MOTOR THHN Wire Size 1 PROTECTION 2,3 Hp (kw) Phase Voltage 4 Max Amps 50 ft. 100 ft. Breaker 2 3.0 (2.2) 1 220v 16.0 A 10 ga. 10 ga. 20 amp Bison 3, 5, 7 50 Hz Bison 9 50 Hz Bulldog QT 1, 2 50 Hz 5.0 (3.7) 3 5.0 (3.7) 3 1.5 (1.1) 1 220v 14.4 A 12 ga. 12 ga. 15 amp 380v 8.3 A 12 ga. 12 ga. 15 amp 440v 7.2 A 12 ga. 12 ga. 15 amp 220v 14.4 A 12 ga. 12 ga. 20 amp 440v 7.2 A 12 ga. 12 ga. 15 amp 110v 19 A 12 ga. 10 ga. 20 amp 220v 10 A 12 ga. 12 ga. 15 amp NOTES: 1. Recommendation only. Ensure compliance with National and Local codes. 2. Fuses must have time delays or be otherwise suitable for motor circuit. 3. Disconnect must be fusible and sized per UL 98, UL 489, or UL 508. Must be supplied and installed by a licensed electrician. 4. Motors are NEMA rated. See NEMA guidelines for applicable voltage and frequency range. Page 10

Single Phase Installation For three phase power, see Three Phase Startup. RAMVAC Supplied Equipment Owner / Contractor Supplied Equipment Disconnect Power Panel (dedicated breakers) 6 ft. conduit and wires (connects to Disconnect) Bulldog or Bison (Bulldog shown) Three Phase Vacuum Power Unit Startup To verify proper pump rotation, use this procedure when starting a three phase vacuum unit for the first time. CAUTION: Continuous improper rotation can permanently damage pump. 1. Lift Vacuum Controller up and out of Filtrols. (See Illustration) 2. Remove Main Air Filter from inside Filtrols. (See Illustration) 3. Check for correct rotation: a. Place hand in Filtrols close to pump inlet pipe. b. Run Power Unit just long enough to feel if inlet pipe is blowing or sucking air; then turn the Power Unit off. 4. If air blows out of inlet pipe: a. Reverse any two motor leads on starter. (See bottom right illustration). b. Repeat step 3. 5. If air sucks into inlet pipe: a. Clean any oil or debris out of Filtrols. b. Reinstall Main Air Filter. c. Reinstall Vacuum Controller. Vacuum Controller Main Air Filter Filtrols Page 11

Three Phase Startup Notes: Contents inside motor starters will vary. For simplicity, not all starter box wiring is shown. Starter Box 3 phase and 50Hz models ONLY Power Panel (dedicated breakers) Disconnect Switch any two of these wires to reverse rotation. Page 12

INSTALLATION CHECKLIST Dentist or Institution Name: Dental Facility Name: Print or Type Mailing Address: (City) (State) (Zip) Phone: ( ) - Facility person responsible Installation Date: for RAMVAC maintenance: Email: Vacuum Unit Model No. 8 E... Tank Model No. 5 Complete by Installer Vacuum Unit Serial No. Tank Serial No. Vacuum Plumbing Electrical Number of Treatment Rooms / Chairs Breaker/Disconnect Amp Rating Number of Vacuum Sinks Line Voltage Number of Vacuum Cuspidors Running Voltage Piping Layout Amps Lead 1 (all motors) [ ] Overhead Lift Height (Floor to Trunk Line) Amps Lead 2 (all motors) Copy [ ] Below Grade Amps Lead 3 (3ph motors) [ ] Pipe Size (Trunk Line) Pump Lubrication [ ] Other [ ] Machine is installed level [ ] Oil Level OK Tank Drain Drip Rate [ ] Air Gap Drips/min - Right Dripper (All systems) [ ] No Air Gap, Vented per NFPA 99c Drips/min - Left Dripper (All systems) Exhaust Plumbing Drips/min - Center Dripper (Bison 7 & 9) Y N Dedicated exhaust plumbing per unit to outside of building Y N Exhaust pipes joints sealed Switching Y N Exhaust piping loosely supported to isolate vibration What on-off switching method will be used? Y N Outside end capped or turned down to prevent entry of rain or snow [ ] Switch on S1 Electrol Y N Outside end screened to prevent entry of animals Exhaust Pipe Material [ ] PVC Sch 40 [ ] PVC Sch 80 [ ] Copper [ ] Other [ ] Remote Switch Seperate Document (Shipped with Power Unit) [ ] RAMVAC Remote Control Panel [ ] Non-Illuminated Switch [ ] Non-RAMVAC Illuminated Switch I have demonstrated and explained Operation and Maintenance procedures, using the User Guide and hardware, to the facility person responsible for RAMVAC maintenance: (Printed Name of Installer) (Dealer) Free Gift Certificate for Installers! Just complete information on back of white copy. Complete by Owner Warranty Initialization I have a User Guide and I will comply with the operation and maintenance instructions or contact RAMVAC if I cannot. I have a completed copy of the RAMVAC Installation Checklist. I understand that RAMVAC warranty coverage depends on RAMVAC Corporation s receipt of this document. (Printed Name of Owner or Authorized Agent) (Signature of Owner or Authorized Agent) Page 13

(800) 5-RAMVAC FAX (605) 642-3776 212 N. MAIN STREET SPEARFISH, SD 57783 www.ramvac.com 2007 RAMVAC Dental Products All rights reserved. No part of this publication may be copied or distributed, transmitted or transcribed in any form or by any means without the expressed written permission of RAMVAC Dental Products, Spearfish, SD 57783 EXCLUSIVELY FROM ISO 13485:2003 certified facility 212 N. MAIN STREET SPEARFISH, SD 57783 TOLL FREE: 800-5-RAMVAC (800-572-6822) PHONE: (605) 642-4614 FAX: (605) 642-3776 e-mail: ramvac@ramvac.com website: www.ramvac.com A Brand of the The Integrated Supplier PHONE: (866-DTE-INFO) www.dentalez.com Effective date: 18-SEP-13