KAM IAS ISOKINETIC AUTOMATIC SAMPLER. User Manual IASMANUAL-0513 KAM CONTROLS, INC Ann Arbor Drive Houston, Texas USA

Similar documents
KAM E-IAS ELECTRIC ISOKINETIC AUTOMATIC SAMPLER

KAM CHA COLORIMETER/HAZE ANALYZER. User Manual API COMPLIANT PTB 04 ATEX 1027

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER SAMPLE / INJECTION PUMP

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER INLOOP CRUDE OIL SAMPLER

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER INFLOW ACE CRUDE OIL SAMPLER TELESCOPING INSERTION

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER INJECTION PUMP

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER INFLOW ACE CRUDE OIL SAMPLER FIXED INSERTION

Welker Sampler. Installation, Operation, & Maintenance Manual. Model GSS-4HP

HIGH PRESSURE CONTROL VALVE PISTON BALANCED

Welker Automatic Insertion Diffusing Probe Model AIP-3DP

Hydraulics. Part B, Section 1. This section covers the following unit configurations. 3700V 3800V 3900V

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER CRUDE OIL COLLECTION ASSEMBLY HIGH PRESSURE

Installation, Operation, and Maintenance Manual

Fisher 657 Diaphragm Actuator Sizes and 87

HYDRAULICS. TX420 & & lower. Hydraulic Tandem Pump Removal. 4. Remove the LH side panel (Fig. 0388).

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER AUTOMATIC INSERTION PROBE

Installation, Operation, and Maintenance Manual. Welker Automatic Insertion Corrosion Coupon Device Model AID-1CC

Material Specifications

6200 Series. Specifications. Fluid End Power End Models 6211, 6212, 6221, & 6222 Models 6241 & 6242 Part Material Part Material Part Material

INSTALLATION AND MAINTENANCE OF TOP LOADING ARM

Welker Adjustable Probe with Check Valve and Welker Automatic Insertion Probe with Check Valve

Operation and Maintenance Manual

MODEL 5540 CONTENTS. Installation, Operation and Maintenance Instructions 1.0 GENERAL

GH-BETTIS SERVICE INSTRUCTIONS DISASSEMBLY & REASSEMBLY FOR MODELS HD521-M4, HD721-M4 AND HD731-M4 DOUBLE ACTING SERIES PNEUMATIC ACTUATORS

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER INFLOW ACE CRUDE OIL SAMPLER WITH AI CONTROL

3M Overhaul Service Kit

KAM OID OPTICAL INTERFACE DETECTOR PTB 08 ATEX User Manual OIDMANUAL An ISO 9001 certified company

BETTIS SERVICE INSTRUCTIONS DISASSEMBLY AND REASSEMBLY FOR CB-SR-S SEISMIC SPRING RETURN SERIES PNEUMATIC ACTUATORS

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER INFLOW CRUDE OIL SAMPLER WITH AI CONTROL

AIRGO MISTING SYSTEM MAINTENANCE

INSTALLATION INSTRUCTIONS FOR THE TRUCK MOUNTED VIPER ADDITIVE INJECTION SYSTEM GTP-8776C

Cylinder and Valve: AirHawk II Air Mask

3.2 DRIVE TORQUE HUB. Roll, Leak and Brake Testing SECTION 3 - CHASSIS & TURNTABLE. 3-2 JLG Lift

BETTIS ACTUATOR & CONTROLS SERVICE INSTRUCTIONS FIELD CONVERSION OF SPRING CARTRIDGE FAIL DIRECTION CLOCKWISE TO COUNTER-CLOCKWISE OR THE INVERSE

METERING VALVE 2" STEM GUIDED

Pipe fi ttings MUST be installed by a certifi ed welder only. Signet will not assume liability of any kind for improper fi tting installations.

Welker Gas Sampler Model MPS-2

Paint/Solvent/Dump Valve and Flow-Through Valve

SD Bendix E-10PR Retarder Control Brake Valve DESCRIPTION. OPERATION - Refer to Figure 2

MEMORY SEAL BALL VALVES MAINTENANCE MANUAL 2 12, Class 150 & 300, Regular Port, Flanged Unibody

AVK SAUDI VALVES MANUFACTURING COMPANY

Welker Heated Instrument Regulator

INSTALLATION & MAINTENANCE MANUAL

Identify and Assemble the Pneumatic Components

Manifold Isolation Valves

OPERATING INSTRUCTIONS & SERVICE MANUAL BLUE MAX II HYDROSTATIC TEST PUMP

Hydraulics. Part B, Section 1. This section covers the following unit configurations. 3400V 3500V 3700V

Operation and Maintenance Manual for BS and BH Hydraulic Torque Wrenches

CAB TILT HYDRAULIC SYSTEM

OPERATION AND PARTS MANUAL

ASSEMBLY MANUAL SMC 2491 LOADER VALVE AND PLUMBING KIT INSTRUCTIONS KUBOTA TRACTORS

Type 657 Diaphragm Actuator Sizes and 87

Installation, Operation, and Maintenance Manual

Bettis M11 Manual Hydraulic Override System Operating Instructions for HD, T, and G Series Pneumatic and Hydraulic Actuators

FRP Ball Valves INSTALLATION & MAINTENANCE MANUAL

Calibration and Switching Module. User s Manual. A Swagelok Pre-Engineered. Subsystem. Pre-engineered subsystems available in weeks, not months.

GH-BETTIS OPERATING & MAINTENANCE INSTRUCTIONS DISASSEMBLY & ASSEMBLY FOR THE T80X-M4-S DOUBLE ACTING SERIES HYDRAULIC ACTUATORS

Steam/Water Washdown Units Safety and Operation Installation and Maintenance Instructions

Pipe fittings MUST be installed by a certified welder only. +GF+ SIGNET will not assume liability of any kind for improper fitting installations.

Product Manual. S&S V-100 Ball Valves 2 though 8 Inch Designs. Introduction: Installation:

MAINTENANCE INSTRUCTIONS

Signet 2517 Brass Paddlewheel Flow Sensor * *

DeZURIK " BAW AWWA BUTTERFLY VALVES WITH EPOXY-RETAINED SEAT

Welker Probe Instrument Regulator with Liquid Eliminator Cartridge

Cleaning and Rotor Replacement Instructions

M Installation Guide. PRECISION Manifold Series. Introduction. A. Assemble Manifold Components

3M Overhaul Service Kit

Anderson Greenwood Series 800 POSRV Installation and Maintenance Instructions

2005 Toyota RAV AUTOMATIC TRANSMISSIONS U240E & U241E Overhaul

High Turn Down Valve Model SF10000HTD Operations and Maintenance Manual. P.O. BOX 1728 Woodinville, WA USA FAX 425.

Operation, Model MPS-2DA. Installation, responsibility

Model 306B H&J Transfer Case Service Manual

70 Series 8700 Coupler 2-pc Hub & Shaft Three Phase Brake Instructions IP43 & IP56 (NEMA 2 & 4) Housing

Type 1051 and 1052 Size 33 Diaphragm Rotary Actuator

Escape Cylinder and Valve

Metso High Consistency Sampler Metso Nove H. Installation & Owner s Manual K10196 V1.1 EN

Sachs shock manual. ( ) 2 & 4 Stroke RR Enduro. ( ) RS Dual Sport

ONYX VALVE CO MODEL DAO-PFC Installation & Maintenance

Pneumatic Chemical Metering Pump

Crispin Valves Operating Guide. Crispin

M-8100EP. Installation Guide ENGINEERED PLASTIC MANIFOLD SERIES. Introduction. A. Assemble Manifold Components

Maintenance & Repair Manual

Swing-Flex Check Valve

VSI INDUSTRIAL BALL VALVES SERIES 7100 SERIES 8100 SERIES 7200 SERIES 8300 SERIES 7400 SERIES 8400 SERIES 8000 SERIES 8500

351GF-15. Hydrant Coupler INSTALLATION / OPERATION / MAINTENANCE MODEL

SERIES 90/91 R PNEUMATIC ACTUATOR OPERATION AND MAINTENANCE MANUAL. The High Performance Company

DISCONTINUED. Installation. Aerada 900 Series Futura Faucet. With Accu-Zone (AZ) Infrared Control

Spring Return and Double Acting Pneumatic Quarter-turn Actuators Operations Manual

INSTALLATION, OPERATION, & MAINTENANCE MANUAL Models # MM-30, MM-30HT, MM-50, MM-50HT, & MM-50 EXL with Special HP Tool and Special HP EXL Tool

TWO-STAGE HYDRAULIC PUMP. RWP55-IBT-Air

ARMAGEDDON tm BY AIR AMERICA, INC.

STRAIGHT FITTING ORIFICE DISK MUFFLER MANIFOLD MOUNTING PLATE MULTI-FUNCTION IMPACT SEAL CYLINDER TUBE 1/4 BSPP ELBOW FITTING HEAD ASSEMBLY

OPERATION AND PARTS MANUAL

Welker Jet Top Entry Control Valve

Service Manual. Balanced Piston First Stage

DK Series Pneumatic Diaphragm Valves Product Data Sheet

INSTALLATION INSTRUCTIONS

SM64155 Maintenance & Repair Manual

Anderson Greenwood Series 400 Piston Pilot POPRV Installation and Maintenance Instructions

Dethatcher Kit Greensmaster 3000 Series

Transcription:

An ISO 900 certified company TEL + 73 784-0000 FAX + 73 784-000 Email Sales@Kam.com KAM IAS ISOKINETIC AUTOMATIC SAMPLER PER API 8.2, ASTM D477 AND ISO 37 User Manual IASMANUAL-053 3939 Ann Arbor Drive Houston, Texas 77063 USA www.kam.com

TABLE OF CONTENTS SECTION TITLE PAGE Introduction 2 Available Models and Mounting Options 2 Theory of Operation 3 Features 4 Applications 4 Sampler Sequence 4 2 Specifications 5 Specifications 5 3 Installation 6 Main Line 6 Field Air Connections Removal 2 4 Maintenance 3 Seal Kit and Replacement Schedule 3 Sampler Disassembly 4 Replacing Seals and O-rings 8 Reassembly 22 Testing 26 5 Troubleshooting 27 CAUTION: When installing the IAS sampler in a pipeline containing petroleum products, petro-chemicals, waste waters with the presence of pressure & temperature, and high-pressure steam refer to the Pipeline Operators Health, Safety and Environmental Policy Procedures to ensure safe installation. reserves the right to make changes to this document without notice.

INTRODUCTION AVAILABLE MODELS and MOUNTING OPTIONS FIG. - Option : Retractable IAS on a main pipe, with 2, 3, or 4 flanged seal housing Full-opening Ball Valve Recommended KAM SMS Static Mixing Spool Q FIG. -2 Option 2: Retractable IAS on a main pipe, with /4" or 2" MNPT seal housing Full-opening Ball Valve Recommended KAM SMS Static Mixing Spool Q 2

INTRODUCTION CONTINUED FIG. -3 Option 3: IAS FT Flow Through with 2" 50# flanges THEORY OF OPERATION Employing the simplest design in the industry, the KAM IAS Isokinetic Automatic Sampler is the ideal instrument for pipeline sampling during custody transfer. The IAS sampler is designed to extract a representative sample of the pipeline into a KAM SR Sample Receiver. With a KAM CSM Circulating Sample Mixer, you may then mix and analyze the sample for custody transfer, API gravity measurements, shrinkage, chemical composition, etc. FIG. -2 Sampling Chamber Locking Collar Air Inlets Seal Housing Actuator Housing Sample Outlet Pressure Relief Valve 3

INTRODUCTION CONTINUED FEATURES APPLICATIONS User friendly Positive displacement design Pneumatic, hydraulic Insertable and extractable through full-opening ball valve Fast loop and main line models Variable insertion Simple design Isokinetic Pressure relief valve Custody transfer Marine loading, unloading, lightering Pipeline Truck loading, unloading Production Refinery Power plant fuel oil Research Quality control in SAMPLER SEQUENCE FIG. -3 4

SPECIFICATIONS Media: Wetted parts: Crude oil, refined products, chemicals, water and wastewater 36 stainless steel shaft and probe cage 304 stainless steel sample piston Teflon sample chamber seal Fluid temperature: Power: Sample size: -40º to 350ºF (-40º to 77ºC) Air pneumatic, hydraulic 0.5 ml - 3 ml (Please specify) Repeatability: Exceeds API 8.2 Mounting: Main line - ¼" MNPT, 2" MNPT seal housing, or 2", 3", or 4" flanged seal housing Fast loop - ¼" FNPT Pressure ratings: ANSI 50, 300, 600, 900 Probe dimensions: Shaft length: Ø.25" x 3." (32mmh x 79mm) 30" (762mm) Other sizes available REQUIREMENTS Air: Consumption: Minimum Sample Time: 70 to 25 psi 20 cubic inches per stroke.5 seconds 5

INSTALLATION PRIOR TO INSTALLATION Before installing the IAS Sampler, make sure that it was not damaged during transit. KAM CONTROLS recommends that the IAS Sampler be installed 2 to 4 pipe diameters downstream of a KAM static mixer or flow conditioner. At the time of installation the pipeline pressure needs to be reduced to under 00 psi in order to be able to install the IAS Sampler. The IAS Sampler should be installed horizontally at 3 or 9 o clock. Fig. 3-. The IAS Sampler Probe should be inserted so that the middle of the sampling window is in the 20% center area of the pipe. Fig. 3-. MAIN LINE INSTALLATION Before mounting the IAS Sampler on the Full-Opening Ball Valve, determine the insertion length required. The IAS Sampler should be installed according to Fig. 3-. The Full-Opening Ball Valve is used to isolate the IAS Sampler from the pipeline during installation or removal. The Seal Housing of the IAS Sampler allows the Sampler Probe to be inserted in and out of the pipe under pressure and flow conditions up to 00 psi. It is the user s responsibility to ensure that the IAS Sampler is placed in the most representative point in the flow profile. Prior to installation, make sure you have the correct Pressure Relief Valve. The IAS Sampler is normally shipped with a PRV designed for pressures up to 750 psi. If the line pressure is greater than 750 psi, please replace the spring inside the PRV with the appropriate Spring. See Table 3-2. FIG. 3- END VIEW OF A KAM IAS INSTALLED ON A MAIN PIPE Sampler Probe Seal Housing Locking Collar Full-opening Ball Valve Hex Socket Cap Screw TABLE 3-2 PRV Spring Cracking Pressure Range (psig) 50 to 350 350 to 750 750 to 500 500 to 2250 Spring Color Blue Yellow (included with IAS Sampler) Purple Orange 6

INSTALLATION CONTINUED. 2. 3. 4. Lay the IAS Sampler on the ground or a table. Loosen Hex Socket Cap Screws on the Locking Collar. Fig. 3-. This will allow the IAS Sampler Shaft to slide through the Seal Housing. Push the IAS Sampler Shaft though the Seal Housing until the middle of the sampler window chamber is aligned with the end of the Seal Housing. Fig. 3-3 or Fig. 3-4. Use a Sharpie or other permanent marker to mark the shaft at the edge of the Locking Collar. (Do not use anything sharp to mark the shaft. This will create grooves that will damage the O-rings in the Seal Housing.) FIG. 3-3 MNPT SEAL HOUSING FIG. 3-4 FLANGED SEAL HOUSING Align Here Mark Here Align Here Mark Here 5. Pull Shaft back until the Probe is all the way in the Seal Housing and tighten the Hex Socket Cap Screws on the Locking Collar. This will prevent the IAS Sampler Shaft from sliding and the Probe will be protected inside the Seal Housing. 6. Measure the distance (D) from the top of the main pipe to the end of the connection where the IAS Sampler is going to be installed. Fig. 3-5. FIG. 3-5 D D 7

INSTALLATION CONTINUED 7. Calculate the insertion distance using the following formula: TID Total Insertion Distance D Distance from the top of the valve to the pipe WT Pipe Wall Thickness TD Threaded Depth ID Inside Diameter For Flanged Seal Housing Total Insertion Distance (TID)= D + WT + (Pipe ID x 7/6) + Seal Thickness Example: D= 4" WT=.25" ID = 8.25" Seal Thickness=/8" TID=4 +.25 + (8.25 x 7/6) + /8 TID=4 +.25 + 3.55 +.25= 7.93 For 2" MNPT Seal Housing TID cannot be calculated until the Seal Housing is screwed into place. Bolt or Screw the IAS Sampler to the Valve or designated installation location. (KAM CONTROLS recommends liquid thread sealant and not teflon tape for the threaded IAS Sampler.) You must then measure the Threaded Depth (TD) into the Valve or connection in order to calculate TID. You can do this by measuring the distance from the edge of the Valve or female connection to the top of the Seal Housing body and subtracting that distance from 5.25". Fig 3-6. Total Insertion Distance (TID)= D + WT + 7/6 Pipe ID -Threaded Depth (TD) FIG. 3-6 Threaded Depth Measuring Points For example: If the measured distance from the top of the Valve and the top of the Seal Housing body is 4.65" you should calculate the threaded depth (TD) by subtracting 4.65" from 5.25". (5.25-4.65=0.6) In this case the threaded depth (TD) would be.6" 8

INSTALLATION CONTINUED You are now ready to calculate the TID. TID= D + WT + (Pipe ID x 7/6) - TD Example: D= 4" WT=.25" ID = 8.25" TD=.6" TID=4 +.25 + 3.55 -.6=7.2 8. Use the Calculated TID and make another mark on the shaft, measuring from First Mark. Fig 3-7. FIG. 3-7 TID First Mark Second Mark 9. If you have an IAS Sampler with a Flanged Seal Housing, you may now attach it to the Valve on the pipeline. 0. Slowly open Full Opening Valve and check for leaks.. Loosen Socket Head Screw on the Locking Collar. 2. Align Window of the IAS Sampler to face the flowing stream. This can be done by aligning the Flow Indicators in parallel with the main pipe. The Flow Indicators are located at the bottom of the Actuator Housing. 3. Push the IAS Sampler in until the Second Mark is at the top edge of the Locking Collar. Fig. 3-8. FIG. 3-8 Second Mark 9

INSTALLATION CONTINUED 4. Re-tighten the Hex Socket Cap Screw. 5. Tighten the Hex Nuts holding the Locking Collar from ¼ to ½ of a turn. The Hex Nuts holding down the Locking Collar should never be over tightened. Their major function is to apply light pressure on the Chevron Packing to ensure a seal between the Seal Housing Body and the Insertion Shaft. Fig. 3-9. FIG. 3-9 Locking Collar Hex Nuts Hex Socket Cap Screw 6. The PRV arrives set at the maximum cracking pressure for the spring installed (normally a yellow spring for 750 psi). If the maximum pressure does not exceed 350 psi, set the cracking pressure to the lowest cracking pressure for the PRV by twisting the cap 8 turns counter-clockwise. The PRV is going to have the highest cracking pressure when the cap is screwed all the way in and the lowest cracking pressure when the cap is almost all the way out. If your line pressure is within the range of the cracking pressure of the PRV, take the following steps to set the correct cracking pressure: a) Slowly turn the cap counter-clockwise until liquid starts to come out of the PRV outlet. b) Turn the cap two turns clockwise to stop the leak and set the PRV cracking pressure 200psi above the current pressure. c) Tighten lock nut against the cap. d) Lock wire cap and body together to maintain set pressure. Once you have set the PRV cracking pressure you are ready to make the field air connections. 0

INSTALLATION CONTINUED FIELD AIR CONNECTIONS. 2. 3. Using Stainless Steel Tubing or Stainless Steel Braided Hose, connect the normally open port (usually marked with a letter "A" or "NO") of the Solenoid Valve to the bottom of the Actuator Housing. Connect the normally closed port (usually marked "B" or "NC") to the top of the Actuator Housing. Fig. 3-0. Connect the Solenoid Valve Coil to a KAM SC Sampler Controller according to the manual for the Solenoid Valve. The SC Controller should be programmed in a way that it energizes the coil for at least.5 seconds. Every time the coil is energized and de-energized the IAS Sampler will take a sample. Using a combination of ¼" Stainless Steel Tubing and Stainless Steel Braided Hose, connect the IAS Sampler PRV outlet to the inlet of a Sample Container. FIG. 3-0 Solenoid Coil to Sampler Controller Computer EA A P EB B Regulated Air Supply Sample Outlet to Sample Container

INSTALLATION CONTINUED REMOVING THE IAS SAMPLER. 2. 3. Turn off the air supply going to the Solenoid Valve. Disconnect the ¼" air connections going to the IAS Sampler Actuator Housing. Disconnect the ¼" connection going to the IAS Sampler sample outlet. It is very important to depressurize the pipeline to below 00 psi. Ideally the pressure needs to be close to 0 psi. Failure to do so could result in injury or damage to the sampler. 4. 5. 6. 7. Once all the connections going to the IAS Sampler are removed, slowly loosen the Locking Collar by turning the Hex Socket Cap Screws counter-clockwise. If there is pressure on the line, it will push the IAS Sampler out when the Locking Collar is loose. If there isn t enough pressure on the line, manually pull the IAS Sampler until the end of the probe stops inside the Seal Housing. Tighten the Locking Collar by turning the Socket Cap Screws clockwise to prevent the IAS Sampler from sliding back in the pipeline or in the way of the Isolation Valve. Close the Isolation Valve. Remove the IAS Sampler from the Isolation Valve. FIG. 3- Seal Housing Locking Collar Actuator Housing Isolation Valve Hex Socket Cap Screw 2

MAINTENANCE During normal operation of the KAM IAS Sampler, some seals will wear. In order to ensure continuous accurate sampling, KAM CONTROLS recommends replacing the seals on your IAS Sampler every six months. A complete kit with all the IAS Sampler O-rings, seals, etc. can be order by contacting KAM CONTROLS at sales@kam.com and requesting Seal Kit 9000 or by calling + 73-784-0000 or faxing your request to + 73-784-000. FIG. 4-23 3 2 9 8 5 7 6 8 9 5 6 7 22 2 4 20 5 2 4 8 9 0 3 TABLE 4-2 SEAL KIT 9000 ITEM 2 3 4 5 6 7 8 9 0 2 3 4 5 6 7 8 9 20 2 22 23 DESCRIPTION 2-04 Viton O-ring 2-04 Teflon Back-up ring Conical Check Check Spring 2-0 Viton O-ring 2-020 Viton O-ring 2-020 Teflon Back-up ring 2-4 Viton O-ring 2-4 Teflon Back-up ring 2-022 Viton O-ring 2-022 Teflon Back-up ring 4300 Series Polypack PSP-338A Seal 2-234 Neoprene O-ring 465SH959 Wiper Seal 2-08 Neoprene O-ring 2-024 Viton O-ring 2-24 Viton O-ring 2-24 Teflon Back-up ring Teflon Chevron Packing Suction Relief Spring Suction Relief Valve Bottom Stop Teflon Seal PART # 9005 902 90785 90786 9004 902 903 907 908 904 905 90230 9020 909 902 906 900 9007 95 90 90789 90788 9000 QTY 2 2 4 4 4 3 2 2 REPLACEMENT FREQUENCY Biannually or every time maintenance is performed Biannually or every time maintenance is performed Only if damaged Yearly or when damaged Yearly Yearly Yearly Biannually or every time maintenance is performed Biannually or every time maintenance is performed Biannually or every time maintenance is performed Biannually or every time maintenance is performed Biannually or every time maintenance is performed Biannually or every time maintenance is performed Yearly Biannually or every time maintenance is performed Every two years or if damaged Biannually or every time maintenance is performed Biannually or every time maintenance is performed Biannually or every time maintenance is performed Biannually or every time maintenance is performed Yearly Only if damaged Biannually or every time maintenance is performed 3

MAINTENANCE CONTINUED TAKING THE IAS SAMPLER APART TO REMOVE THE SEALS. 2. 3. 4. 5. To perform maintenance the IAS Sampler first needs to be removed from the pipeline according to the instructions in Section 3, page. Once you ve removed your IAS Sampler from the pipeline, clean all surfaces as much as possible. Place the IAS Sampler on a clean work surface and loosen the Locking Collar by turning the Hex Socket Cap Screws counter-clockwise. Fig. 4-3. Slide the Seal Housing back until the Sampling Chamber is completely exposed. Remove the Sampling Chamber from the shaft using a /8 wrench on the wrench flat and turning it counterclockwise. Fig. 4-3. FIG. 4-3 Sampling Chamber Wrench Flat Locking Collar Sampling Chamber Seal Housing Hex Socket Cap Screws 6. Retighten the Hex Socket Cap Screws. Using the wrench flat on the on the seal housing, turn the Seal Housing and Outer Shaft in a counter clockwise direction to unscrew them from the Seal Chamber. WARNING: DO NOT use a pipe wrench on the Outer Shaft as this could damage the shaft and Sample Tube, prohibiting sampling motion. Slide Seal Housing and Outer Shaft off Sampling Tube. Fig. 4-4. FIG. 4-4 Seal Housing Outer Shaft 4

MAINTENANCE CONTINUED 7. Loosen Hex Socket Cap Screws once again and slide Seal Housing off Outer Shaft. Fig. 4-5. FIG. 4-5 Seal Housing Outer Shaft 8. Remove the (4) ¼ Socket Set Screws holding the Seal Chamber to the Actuator Housing using a 3/6" allen wrench. Fig. 4-6. FIG. 4-6 Seal Chamber Actuator Housing ¼" Socket Set Screws 9. Pull Seal Chamber away from the Actuator Housing and completely off the Sample Tube. Normally, Seal Bushing # will slide with the Seal Chamber down the Sample Tube until it hits the Sample Piston. Pull the Seal Chamber free of Seal Bushing #. Seal Bushing #2 will either remain against the Actuator Housing, or slide down the Sample Tube with Seal Bushing #. Fig. 4-7. FIG. 4-7 Seal Bushing # Seal Bushing #2 Seal Chamber Sample Piston Sample Tube Piston Actuator Tube 5

MAINTENANCE CONTINUED 9. Remove the Sample Piston by turning it counter-clockwise using a 5/8 wrench. Fig. 4-8. 0. Slide Seal Bushing # off the Sample tube. Fig. 4-8. If Seal Bushing # did not slide free of the Actuator Housing when you removed the Seal chamber, you can remove it when you disassemble the Actuator Housing. Fig. 4-9.. If Seal Bushing #2 slid free of the Actuator Housing when you removed the Seal chamber, slide it off the Sample Tube. If not, it will be removed when the Actuator Housing is disassembled. Fig. 4-9. FIG. 4-8 Sample Piston Seal Bushing # Sample Tube 6. Conical Check Check Spring 2. 3. Remove (6) ¼" Bolts from the Actuator Housing using a 7/6" nut driver/wrench, including the two bolts anchoring the Sample Outlet Plate to the Actuator Housing. Fig. 4-9. Using a /2" wrench, remove /8 NPT Male Adapter on the Sample Outlet Tube from the Piston Actuator Tube. Fig. 4-9. FIG. 4-9 Air Inlets (4) ¼ -20 x Bolts Piston Actuator Tube 8" NPT Male Adapter Sample Tube Sample Outlet Plate (2)¼"-20 x ¼" Bolts 6

MAINTENANCE CONTINUED 4. Remove Actuator Housing Lid by pushing the Sample Tube toward the Actuator Housing. The Actuator Housing Lid should lift off the housing body. If you are unable to push the lid off the Actuator Housing by hand, this can be done by forcing compressed air into the Air Inlet on the Actuator Housing. CAUTION: Stand clear of the end of the housing. Fig. 4-0. 5. 6. Slide Piston Disk Assembly and Sampling Tube completely free of the Actuator Housing. Fig. 4-0. Slide Seal Bushing # 2 from the inside of the Actuator Housing if it did not come out on step 0. Fig. 4-0. FIG. 4-0 Seal Bushing # 2 Sample tube Piston Actuator Tube Actuator Housing Lid Piston Disk Assembly 7. Remove (4) 6 3-2 * /2 Flat Head Phillips Screws in Piston Disk Spacer. Fig. 4-. Unscrew Piston Disk from Piston Actuator Tube and slide off. FIG. 4-6 3-2 * /2 Flat Head Phillips Screws Piston Disk Spacer Piston Disk 7

MAINTENANCE CONTINUED REPLACING THE SEALS AND O-RINGS. 2. 3. 4. Remove the bottom Stop from the Sampling Chamber by unscrewing it. Fig. 4-. Clean as necessary. Remove the Top from the Bottom Stop. To do so, use a 5/32 hex wrench and turning it counter clockwise. Replace the Teflon Seal. Fig. 4-. Replace Top in Bottom Stop, then place a drop of Loctite 242 on Bottom Stop threads, and screw Bottom Stop back into the Sampling Chamber. Fig. 4-. FIG. 4- Bottom Stop Sampling Chamber Teflon Seal Top to Bottom Stop 5. Replace the PSP seal from the exterior of the Piston Disk. Replace 2-08 O-ring. Fig. 4-2. FIG. 4-2 O-ring (2-08) PSP Seal 8

MAINTENANCE CONTINUED 6. 7. On Seal Bushing # remove worn Wiper Seal, O-rings and Teflon Backups using a small pick or screwdriver. Insert the first replacement Teflon Backup by squeezing it into an oblong shape and pushing it through the center of the Seal Bushing until it catches inside the back groove. Continue pushing the Backup into the groove until it snaps into place. Then, using the same procedure, place an O-ring 2-4 directly in front of the Teflon Backup inside the same groove. Repeat the process for the second Teflon Backup and O-ring, then replace the Wiper Seal on the end of Seal Bushing #. Fig. 4-3. On Seal Bushing #2, remove worn Wiper Seal and Polypack Seals using a small pick or screwdriver. Replacing the center Polypack Seals (2 and 3) first. Squeeze the Polypack Seal into an oblong shape and insert through the end of the Seal Bushing until it catches on one of the two interior grooves. Polypack Seals are directional. It is very important to install the center Polypack Seals with the O-ring/grooved side facing the Wiper Seal end of the Seal Bushing according to Figure 4-3. Once you ve installed the two center Polypack Seals, replace their corresponding O-rings. Then replace the outer Polypack Seals ( and 4) and )-rings in the same way, making sure that Polypack Seal 4 on the end away from the Wiper Seal faces in the opposite direction from the other three. Replace the Wiper Seal and the two outer O-rings. FIG. 4-3 Wiper Seal () Wiper Seal () O-ring 2-024 (2) O-ring 2-4 (2) Teflon Backup 2-4 2 3 4 Polypack Seals 9

MAINTENANCE CONTINUED 8. Using the same method as in step 6 and 7, replace Teflon Backups, O-rings and Wiper Seal on the interior of the Actuator Housing Lid with the O-rings toward the interior side of the lid and the Wiper Seal toward the exterior. Then replace the large exterior O-ring (2-234) on the lid. Fig. 4-4. FIG. 4-4 Teflon Backups Wiper Seal Interior O-rings Exterior O-ring 9. To change the O-rings and Chevron Packing within the Seal Housing, first remove the ½" Locking Nuts on the end of the Seal Housing. Fig 4-5. Then remove the clamp and bushing. Replace the Teflon Back-up Ring, O- ring and Chevron Packing in the manner described in step 6. FIG. 4-5 Chevron Packing Clamp Teflon Back-up Ring 2-24 O-ring 2-24 Bushing ½" Locking Nuts 20

MAINTENANCE CONTINUED 0. Replace two back-ups and one O-ring (2-04) from Sample Piston. Fig. 4-6. FIG. 4-6 Sample Piston O-ring and back ups. Replace the two O-rings and two back-ups in the Seal Chamber using the method desribed in step 6. Fig. 4-6a. FIG. 4-6a Teflon Back-up Ring and O-ring 2-020 Teflon Back-up Ring and O-ring 2-022 2

MAINTENANCE CONTINUED REASSEMBLING THE IAS SAMPLER. Slide Piston Disk back over Piston Actuator Tube and screw into place. Slide Piston Disk Spacer over tube from the opposite end and secure in place with 6 3-2 * /2 Flat Head Phillips Screws. Fig. 4-7. FIG. 4-7 6 3-2 * /2 Flat Head Phillips Screws Piston Actuator Tube Piston Disk Spacer Piston Disk 2. 3. Slide Sample Tube and Piston Disk Assembly back into the Actuator Housing. Fig. 4-8. Slide Actuator Housing Lid onto Piston Actuator Tube and push into Actuator Housing until it snaps into place. Make sure you align the two Air Inlets in the appropriate positions to reattach to air supply. Fig. 4-8. FIG. 4-8 Actuator Housing Sample tube Piston Actuator Tube Actuator Housing Lid Piston Disk Assembly Piston Disk Spacer Piston Disk 22

MAINTENANCE CONTINUED 4. Coat the ends of the (6) ¼" Actuator Housing Lid bolts with Loctite 242. Using a 7 6 " wrench, reattach the Actuator Housing Lid and Sample Outlet Plate, making sure to use the (2) ¼" bolts in the Sample Outlet Plate. Tighten the 6 bolts gradually and evenly. Fig. 4-9. 5. Reconnect the Sample Outlet Tube to the Piston Actuator Tube. This is easiest if you first remove the Male Adapter from the Compression Fitting. Wrap one end of the Male Adapter in teflon tape, then screw it into the Piston Actuator Tube using a ½" wrench and holding the Piston Actuator Tube Shaft in place with a 9 6 " wrench. Then attach the Compression Fitting at the end of the Sample Outlet Tube to the exposed end of the Male Adapter. Fig. 4-9. FIG. 4-9 (4) ¼ -20 x Bolts Piston Actuator Tube 8" NPT Male Adapter (2)¼"-20 x ¼" Bolts 6. Slide Seal Bushing # down into Seal Chamber, Wiper Seal end first. Then put Seal Bushing #2 into Seal Chamber, Wiper Seal end first. Fig. 4-20. FIG. 4-20a FIG. 4-20b Seal Bushing # Seal Bushing #2 Seal Chamber 23

MAINTENANCE CONTINUED 7. Slide Seal Chamber back onto the Sampling Tube until Seal Bushing #2 rests against the Actuator Housing. Reconnect the Seal chamber to the Actuator Housing using the (4)¼" Socket Screws tightened gradually and evenly. Fig. 4-2. FIG. 4-2 Actuator Housing Seal Chamber Sampling Tube 8. Put Loctite 242 on exposed threads of Outer Shaft, then slide Outer Shaft over Sampling Tube and screw into Seal Chamber, tightening by hand. Fig. 4-22. FIG. 4-22 Outer Shaft Loctite Air Inlet Sampling Tube Seal Chamber 9. Force compressed air into the Air Inlet on the Actuator Housing Lid (FIG. 4-22) until the Sampling Tube extends outside the end of the Outer Shaft and the threads are visible. Fig. 4-23a. 0. Reassemble Sample Piston, putting the Conical Check back in place and slipping a new Spring over the top. Fig. 4-2a. Wrap teflon tape on the threaded end of the Sampling Tube, then use a 9/6" wrench on the wrench flat located on the outside end of the Actuator Housing to hold the Sampling Tube in place, and screw the Sample Piston onto the Sampler Tube. Fig. 4-23b. FIG. 4-23a FIG. 4-23b Sample Piston Sampling Tube Sampling Tube Conical Check Sample Piston Wrench Flat Spring 24

MAINTENANCE CONTINUED. Force compressed air into the Air Inlet on the Actuator Housing to push the Sampler Tube back into place. Fig. 4-24. FIG. 4-24 Air Inlet 2. Slide the Seal Housing onto the Outer Shaft. Fig. 4-25. 3. Using Loctite 242 on the exposed threaded end of the Outer Shaft, reattach the Sampling Chamber to the Outer shaft. Fig. 4-25. FIG. 4-25 Outer Shaft Locking Collar Sampling Chamber Seal Housing Hex Socket Cap Screws 4. Slide the Seal Housing down until it rests against the Sampling Chamber and tighten into place issuing the Hex Socket Cap Screws in order to protect the sampling chamber during transport. Using the wrench flat on the Seal Housing, turn Seal Housing (and Outer Shaft) in a clockwise direction to fully tighten seal between Outer Shaft and Seal Chamber. Fig. 4-26. FIG. 4-26 Seal 25

MAINTENANCE CONTINUED You are now ready to test the reassembled IAS Sampler.. Place the sampling end into a bucket of water. Force air through the air inlets to create the sampling piston action. Capture multiple samples in a beaker and measure for consistency. FIG. 4-27 Air Inlet Sample Outlet Water 2. Reinstall your IAS using the steps outline in section 3. 26

TROUBLESHOOTING Problem: Poor sample draw or no sample Solution: Double check Spring inside Sample Piston and the O-ring on the Bottom Stop and replace if broken or worn. Problem: Oil in air lines Solution: Usually, this means that the Seals inside the Seal Bushings are worn out. Perform standard maintenance. If problem persists, the Piston Tube may be scratched and will need to be replaced. Problem: Piston s not moving Solution: Check air supply and solenoid valve Problem: Oil leaking through Seal Housing Solution: Perform routine maintenance replacing O-rings, Seals, etc. 27