APPLICATIONS TN Determining the Effects of Harmful LC Solvent Vapors on GC Analysis

Similar documents
LC SOLVENT SAFETY PRODUCTS. Mobile Phase and Waste Starter Kit. Instructions

Best Selling 30 % OFF* GC Accessories. Expires: September 27, 2013

Supported Liquid Extraction (SLE)

APPLICATION TN Quantitative Analysis of Cocaine Metabolites and Adulterants in Oral Fluid

Stability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

C L O S E D S Y S T E M S VAPOR AND SPILL CONTROL FOR LABORATORY SOLVENTS

vision safety caps Manufacturer: VisionEC-Lab Ltd. Hungary

CLOSED SYSTEMS FOR SOLVENT CONTAINMENT VAPOR AND SPILL CONTROL FOR LABORATORY SOLVENTS

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

HPLC TROUBLESHOOTING GUIDE

Comparison of LC/MS/MS and GC/MSD Analyses of Pesticide Residues in Food Products when Using the QuEChERS Sample Preparation Technique

This document is a preview generated by EVS

Safety Products brochure 11 Int INDEX. Exhaust Filters 2. Air Inlet Valve 5. GL45 Caps 6. DIN50 Caps 8. GL38 Caps 10. DIN60 Caps 11.

UView Mini Transilluminator

Safety Products brochure 11 Int INDEX. Exhaust Filters 2. Air Inlet Valve 5. GL45 Caps 6. DIN50 Caps 8. GL38 Caps 10. DIN60 Caps 11.

SFC-FID ASTM System. SFC-4000 Series SFC. Performance Innovation Reliability

Large Volume Injection of Polycyclic Aromatic Hydrocarbons

Infographics on Electromobility (January 2019)

Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve

Setting up SilFlow for BackFlush in your GC

DANI Transformer Oil Gas Analyzer

Liquefied Gas Injector. Solution for the Sampling and Analysis of Liquefied Gases

breakthrough versatility

APPLICATIONS TN Automated Sample Preparation for a Comprehensive Drug Research Panel from Oral Fluid Using Quantisal Device.

C OMPONENT S F O R A C L O S E D - S YS T E M HPL C

Technical Procedure for Gas Chromatography (GC-FID)

Valving Schemes for the BioLogic Duo-Flow Chromatography System

Determination of Sudan Dyes I IV in Curry Paste

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

Thermo Scientific Dionex UltiMate 3000 Thermostatted Column Compartment

I. Ježek et al. Correspondence to: I. Ježek and G. Močnik

Comparison of techniques for the gas phase analysis of ammonia and nitrogen oxides in Diesel exhaust gases

The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector

Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution

VERITY Purification Systems Build the perfect system to fit your research

Global Valuation Summary - SAMPLE PAGES

High Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell

A Comparison of Boat Introduction and Direct Injection using the Thermo Scientific ipro 5000 Series Combustion Analyzer

Analysis of Petroleum Fractions by ASTM D2887

Solvent Safety Bottle Caps

SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

Sectoral Profile - Services

TECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure

Beverage Grade Carbon Dioxide

Photo courtesy of NZTA

Over time consistency of PPP results in the OECD countries

Phenomenex Exclusive! Inlet Base Seals. Agilent 5000, 6000 and. Guaranteed fit for Series GCs. Easy Seals

The Stability of Sulfur Compounds, Low Molecular Weight Gases, and VOCs in Four Air Sample Bag Materials

Total credit to the non-financial sector (core debt), % of GDP Table F1.1

Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis

16 Ideal for standard laboratory chromatography. 19 High temperature columns ideal for flow chemistry

HIGH PRESSURE SWITCHING VALVES

testing to perfection MultiTest-i Technical Datasheet

NI PS-15/16/17 Side Mount Brackets

Using the PSD for Backflushing on the Agilent 8890 GC System

Agilent G5632 Valve Kit Instructions

Spain s exports by oil product group and country of destination

Application. Gas Chromatography June 1995

Specifications for Refrigerants

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

Successful policies and measures in Europe

VICI JOUR novinky říjen 2011

LIABLE RESULTS AND RELIABLE RESULTS AND EXCEPTIONAL REPRODUCIBILITY


1. INTERNATIONAL OVERVIEW. 1.0 Area and population. population (1,000) area

Spare Parts List Series VP04 Pneumatic Remote Control Valve

RISK DASHBOARD ANNEX CREDIT RISK PARAMETERS Q2 2017

RISK DASHBOARD ANNEX CREDIT RISK PARAMETERS Q4 2017

260 Solvent Vapor Exit Accessory

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

Bright prospects Solvents for spectroscopy Uvasol

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

Gear Pumps/Motors Series PGP/PGM 500

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC

This convenient system is made entirely of inert parts: glass, Baseline drift

APPLICATION OF SOLID PHASE MICROEXTRACTION (SPME) IN PROFILING HYDROCARBONS IN OIL SPILL CASES

Taxing Petrol and Diesel

Electrophoresis + Blotting + Imaging. V3 Western Workflow Visualize. Verify. Validate.

Application Note. Authors. Abstract. Energy & Chemicals

Bio-Rad Laboratories. Liquichek Hematology Controls

Set Your Sights on Superior Performance

Set Your Sights on Superior Performance

Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter

A multi-model approach: international electric vehicle adoption

INSTALLATION GUIDE NI 9921/9922 Outdoor IP Enclosure For NI WLS-9000 Series Devices

This document is a preview generated by EVS

Allegro 2D Standard Systems

Technical Procedure for the Examination of Arson Evidence

Runwithit! The Mini-PROTEAN. 3 Mini Vertical Electrophoresis System is So Easy to Use, You Won t Want to Run With Anything Else.

AUSTRIA. Table 1. FDI flows in the host economy, by geographical origin. (Millions of US dollars)

Antek Gas Chromatography Products: A Guided Tour Through Industry

Information Technology and Economic Development: An Introduction to the Research Issues

Online sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve

Stern-Wywiol Gruppe GmbH + Co. KG

Application Note. Abstract. Authors. Environmental Analysis

ASTM D for Denatured Fuel Ethanol Automating Calculations and Reports with Empower 2 Software

Phase Distribution of Ethanol, and Water in Ethyl Esters at K and K

Transcription:

APPLICATIONS Determining the Effects of Harmful LC Solvent Vapors on GC Analysis Ramkumar Dhandapani, Zeshan Aqeel, and Nicole Blood Phenomenex, Inc., 411 Madrid Ave., Torrance, CA 90501 USA Ramkumar Dhandapani Global Product Manager, GC Ramkumar loves to write poems, and to watch & read Shakespeare s plays. Abstract Losses of volatile mobile phase components to the atmosphere can result in poor chromatography. Using GC-MS, the changes are observed with and without a SecurityCAP product. Introduction The SecurityCAP mobile phase and solvent waste safety caps prevent dangerous vapors and gases from leaving LC solvent reservoirs and solvent waste containers. Over time, if not properly contained, these harmful solvents can have a negative impact on the health of employees and visitors in the lab. OSHA, ICH, and USP have placed allowable limits for harmful solvents in the workplace or in pharmaceutical ingredients that are listed in Table 1. Also, labs that do not have a dedicated area for both HPLC and GC instruments will suffer from cross contamination of HPLC mobile phase solvents in their GC analyses. False positive GC peaks can be detected on sensitive instruments, such as GC-Headspace or GC-MS due to escaped LC solvent vapors. To examine the presence of cross contamination by LC solvent vapors, an experiment was conducted using INFICON Micro GC Fusion Gas Analyzer. This experiment was designed to determine solvent contamination for low boiling solvents like methanol, acetonitrile, and dichloromethane in GC analysis with and without the use of SecurityCAP products. The SecurityCAP products not only protects HPLC/UHPLC results by preventing airborne particulates and contaminants from entering the solvent bottle, causing irreproducible results, bacterial growth, and ghost peaks, but it also prevents cross contamination of vapors. When laboratory safety and dependable results are a priority, SecurityCAP is an essential component to laboratory safety. Experimental Conditions The experimental design consisted of placing the portable INFI- CON Micro GC Fusion Gas Analyzer device inside a closed environment such as a 2 x 2 box to monitor the presence of solvent vapor. A pictorial representation of the experiment is shown in Figure 1. First, this experiment was performed in an empty box. A mini fan was kept in the corner of the box in order to circulate the vapors. Then an open, 1 L glass solvent bottle was kept inside the box. The box was then closed for 3-minutes to let the space equilibrate with the vapors, then a reading was taken at 3 minute intervals, in order for the vapors to fill in the space. Every reading requires box evacuation for 1 minute, 3 minutes of equilibration, and recording the solvent vapor using INFICON Micro GC-TCD. Box evacuation was performed for 1 minute by fanning out the vapors which prevented carry over from one experiment to another. Common reversed phase HPLC solvents such as methanol, acetonitrile, tetrahydrofuran, and normal phase HPLC solvents like dichloromethane and ethyl acetate were individually analyzed in a 1 L solvent bottle to monitor solvent contamination. The experiment was repeated for solvent bottles with and without SecurityCAP installed. All these experiments were performed to mimic real lab environment, where open solvent bottles can cause health hazard and false positive results. GC-MS/MS Conditions Column: 100 % PDMS Dimensions: 20 m x 0.15 mm x 0.15 µm Temperature Program: 60 ºC hold for 65 ºC/sec, 160 ºC @ 0.8 ºC/sec and hold for 10 sec at 160 ºC Column Pressure: 35 psi Injection Time: 30 sec Sampling Time: 15 sec (through a tube/straw) Injection Temperature: 90 ºC Sample Inlet Heater Temperature: 50 ºC Detector: TCD Table 1. Allowable Limits for Harmful Solvents Solvent OSHA Permitted Exposure Limit (ppm) Boiling Point (ºC) ICH/USP Residual Solvent Limits (ppm) Acetonitrile 40 82 410 Methanol 200 64.7 3000 Tetrahydrofuran 200 66 720 Dichloromethane 25 39.6 600 Ethyl Acetate 500 77.1 5000 For additional technical notes, visit www.phenomenex.com Page 1 of 8

Figure 1. Set-up for Open Mobile Phase Bottle and SecurityCAP Installed Bottle FAN FAN Micro GC Fusion Micro GC Fusion Figure 2. Empty Box Open Mobile Phase Bottle SecurityCAP 51. 50. 49. 49k 48. 1 Figure 3. Acetonitrile with Open Mobile Phase Bottle 52k 51. Acetonitrile Peak 50. 49. 1 Page 2 of 8

Figure 4. Acetonitrile with SecurityCAP Installed 51. No Acetonitrile Peak 49. 1 Figure 5. Methanol with Open Mobile Phase Bottle 51. 50. Methanol Peak 49. 1 Figure 6. Methanol with SecurityCAP Installed 53k 52. 52k No Methanol Peak 51. 50. 1 For additional technical notes, visit www.phenomenex.com Page 3 of 8

Figure 7. Tetrahydrofuran with Open Mobile Phase Bottle 50. Tetrahydrofuran Peak 49. 49K 48. 47. 0 1 Figure 8. Tetrahydrofuran with SecurityCAP Installed 51. No Tetrahydrofuran Peak 50. 49. 49.k 48. 1 Figure 9. Dichloromethane (DCM) with Open Mobile Phase Bottle 50. DCM Peak 49. 49k 1 Page 4 of 8

Figure 10. Dichloromethane (DCM) with SecurityCAP Installed 50. No DCM Peak 49. 49k 48. 1 Figure 11. Standard Test for Ethyl Acetate 1 1 100k 90k Ethyl Acetate Peak 80k 70k 60k 1 2 30k Figure 12. Ethyl Acetate with Open Mobile Phase Bottle 52. 52k Ethyl Acetate Peak 51. 50. 1 2 30k For additional technical notes, visit www.phenomenex.com Page 5 of 8

Figure 13. Ethyl Acetate with SecurityCAP Installed 52k 51. No Ethyl Acetate Peak 50. 49. 1 2 30k Results and Discussion The results of the experiment are represented in the form of chromatograms. Figure 2 shows a blank chromatogram that shows three distinct peaks. The first peak represents atmospheric nitrogen, second peak represents carbon dioxide, and the third peak represents the moisture present in air. Since TCD is a universal detector, it may be able to detect even more permanent gases. Figure 3 shows the comparative chromatogram for experiment with the acetonitrile bottle without a Security- CAP and Figure 4 shows the chromatogram with the Security- CAP installed. The open bottle without a SecurityCAP showed contamination of acetonitrile while the SecurityCAP resulted in a chromatogram that displayed no acetonitrile peak. Figures 5-8 shows the same pattern where SafetyCAP installed bottles showed no solvent contamination with methanol, tetrahydrofuran respectively. Figure 9 shows a higher peak intensity for an open bottle of dichloromethane. This is because dichloromethane has a lower boiling point than other solvents considered for this study. So, it readily goes to the vapor phase at room temperature and contaminates the atmosphere easily. However, it is a chlorinated solvent and it has a very low permitted exposure limit of 25 ppm as per OSHA as shown in Table 1. Figures 12 and 13 show chromatograms for ethyl acetate with and without a SecurityCAP. Since ethyl acetate has a higher boiling point, a very low peak intensity was observed using the open bottle. Table 1 shows the permitted exposure limit as per OSHA for all the solvents that are considered for this study. Exceeding these limits will cause various health hazards and can even be fatal. Not only do open solvent bottles cause health hazards, but also the saturated solvent vapors can cause potential fire accidents in the laboratory. Aside from this, a contaminated lab environment with a mixture of solvents can show false positive and ghost peaks on complementary chromatographic techniques. For example, there are numerous labs that have HPLC and GC instruments side by side. In GC, the analytes of interest are commonly volatiles, while in HPLC, these are used as mobile phase. If the mobile phase bottles are kept open or if they are not sealed completely with a SecurityCAP, solvent contaminants will be observed in GC data. The table also shows the ICH limits for residual solvents as per guidelines Q3C(R6). USP has the same limits for these solvents in pharmaceutical ingredients. Prior to testing the pharmaceutical ingredients for residual solvents, it is highly recommended to run blanks on a GC instrument. If the lab environment is contaminated with solvent vapors from open mobile phase bottles, the contamination peaks for methanol, acetonitrile, and methylene chloride, etc. will be observed. The intensity of contaminated peaks in the blank will be enhanced if the adopted method is headspace or mass spec based detection as these techniques have very low detection limits. If a blank chromatogram was not analyzed, false positive results would show up on samples due to HPLC solvent contamination in the GC instrument. With SecurityCAP installation to the mobile phase reservoir and solvent waste reservoir, environmental contamination can be prevented. By adopting SecurityCAP mobile phase and solvent waste safety systems: health hazards due to solvent can be prevented fire accidents due to saturated solvent vapor can be prevented And ghost peaks or false positives for GC analysis can be prevented Conclusion The results demonstrate that solvent peak contamination can occur from open mobile phase bottles and can alter results due to vapor contamination. Using SecurityCAP mobile phase and solvent waste safety systems, lab environment contamination can be prevented, making SecurityCAP an essential component for safe laboratory practice. References 1. https://www.osha.gov/dts/chemicalsampling/data/ch_253450.html 2. http://www.ich.org/products/guidelines/quality/article/quality-guidelines.html 3. http://www.ich.org/fileadmin/public_web_site/ich_products/guidelines/ Quality/Q3C/Q3C R6 Step_4.pdf Page 6 of 8

SecurityCAP Ordering Information Starter Kits SecurityCAP Mobile Phase (Eluent) Safety Starter Kits Part No. Description AC2-1245 2-port GL45 Cap and 6-month Safety Filter AC2-4245 2-port GL45 Caps (x4) and 6-month Safety Filters (x4) AC2-4240 2-port Merck S40 Caps (x4) and 6-month Safety Filters (x4) AC2-1345 3-port GL45 Cap and 6-month Safety Filter AC2-4345 3-port GL45 Caps (x4) and 6-month Safety Filters (x4) AC2-4445 4-port GL45 Cap (x1) and 2-port Caps (3x) and 6-month Safety Filters (x4) AC2-1445 4-port GL45 Cap and 6-month Safety Filter AC2-1545 5-port GL45 Cap and 6-month Safety Filter AC2-1561 5-port S60/S61 Cap and 6-month Safety Filter SecurityCAP Waste Safety Starter Kits AC1-1245 2-port GL/DIN45 Cap and 6-month Exhaust Filter and Barbed Connector ea AC1-1545 5-port GL/DIN45 Cap and 6-month Exhaust Filter ea AC1-1551 5-port DIN51 Cap and 6-month Exhaust Filter ea AC1-1553 5-port B53 Cap and 6-month Exhaust Filter ea AC1-1561 5-port S61 Cap and 6-month Exhaust Filter ea Replacement Filters SecurityCAP Mobile Phase Safety Filters AC2-0161 6-month Capacity, 1/4 in.-28 Threads ea AC2-0961 6-month Capacity, 1/4 in.-28 Threads 10/pk Fittings and Accessories SecurityCAP Fittings AC3-1101 for 1/16 in. or 2.0 mm ID Tubing, 1/4 in.-28 Threads (POM), blue ea AC3-1201 for 2.3-2.6 mm ID Tubing, 1/4 in.-28 Threads (POM), white ea AC3-2101 for 1/8 in. ID Tubing, 1/4 in.-28 Threads (POM), black ea SecurityCAP Connectors AC3-1001 Barbed Connector, for 5-8 mm ID Tubing (PTFE), white ea AC3-1301 Y-connector, for 6-8 mm ID Tubing (POM), white ea POM = polyoxymethylene PTFE = polytetrafluoroethylene (Teflon ) SecurityCAP Waste Safety Filters AC1-0161 6-month Exhaust Filter for SecurityCAP, 1/4 in.-28 Threads ea AC1-0361 6-month Exhaust Filter for SecurityCAP, 1/4 in.-28 Threads 3/pk AC1-0162 6-month Exhaust Filter for Wide-port Caps, GL14 Threads ea AC1-0362 6-month Exhaust Filter for Wide-port Caps, GL14 Threads 3/pk SecurityCAP Adapter AC2-1138 Cap Thread Adapter, PTFE, GPI/GL 38 Female to GL45 Male ea AC3-1111 Waste Adapter for Male 1/4 in. NPT-port (PTFE) ea SecurityCAP Sealing Plug AC3-2001 1/4 in.-28 Threads (POM), white ea SecurityCAP Waste Safety Compatibility Table S Phenomenex SecurityCAP Filters Supplier ea 3/pk S.C.A.T SafetyWasteCaps AC1-0162 AC1-0362 AIT Smart Healthy Caps AC1-0162 AC1-0362 Agilent InfinityLab Stay Safe Caps AC1-0162 AC1-0362 VICI Jour Waste Caps AC1-0161 AC1-0361 Canary-Safe Safety Caps AC1-0162 AC1-0362 DURAN DG Safety Caps AC1-0162 AC1-0362 VapLock Safety Caps (with AC3-1111) AC1-0161 AC1-0361 If SecurityCAP Safety Products do not perform as well or better than your current solvent safety products of similar type, dimensions, and material, return the product with comparative data within 45 days for a FULL REFUND. For additional technical notes, visit www.phenomenex.com Page 7 of 8

APPLICATIONS Australia t: +61 (0)2-9428-6444 f: +61 (0)2-9428-6445 auinfo@phenomenex.com Austria t: +43 (0)1-319-1301 f: +43 (0)1-319-1300 anfrage@phenomenex.com Luxembourg t: +31 (0)30-2418700 f: +31 (0)30-2383749 nlinfo@phenomenex.com Mexico t: 01-800-844-5226 f: 001-310-328-7768 tecnicomx@phenomenex.com Belgium t: +32 (0)2 503 4015 (French) t: +32 (0)2 511 8666 (Dutch) f: +31 (0)30-2383749 beinfo@phenomenex.com Canada t: +1 (800) 543-3681 f: +1 (310) 328-7768 info@phenomenex.com China t: +86 400-606-8099 f: +86 (0)22 2532-1033 phen@agela.com Denmark t: +45 4824 8048 f: +45 4810 6265 nordicinfo@phenomenex.com Finland t: +358 (0)9 4789 0063 f: +45 4810 6265 nordicinfo@phenomenex.com France t: +33 (0)1 30 09 21 10 f: +33 (0)1 30 09 21 11 franceinfo@phenomenex.com Germany t: +49 (0)6021-58830-0 f: +49 (0)6021-58830-11 anfrage@phenomenex.com India t: +91 (0)40-3012 2400 f: +91 (0)40-3012 2411 indiainfo@phenomenex.com Ireland t: +353 (0)1 247 5405 f: +44 1625-501796 eireinfo@phenomenex.com Italy t: +39 051 6327511 f: +39 051 6327555 italiainfo@phenomenex.com The Netherlands t: +31 (0)30-2418700 f: +31 (0)30-2383749 nlinfo@phenomenex.com New Zealand t: +64 (0)9-4780951 f: +64 (0)9-4780952 nzinfo@phenomenex.com Norway t: +47 810 02 005 f: +45 4810 6265 nordicinfo@phenomenex.com Portugal t: +351 221 450 488 f: +34 91-413-2290 ptinfo@phenomenex.com Spain t: +34 91-413-8613 f: +34 91-413-2290 espinfo@phenomenex.com Sweden t: +46 (0)8 611 6950 f: +45 4810 6265 nordicinfo@phenomenex.com Switzerland t: +41 61 692 20 20 f: +41 61 692 20 22 swissinfo@phenomenex.com United Kingdom t: +44 (0)1625-501367 f: +44 (0)1625-501796 ukinfo@phenomenex.com USA t: +1 (310) 212-0555 f: +1 (310) 328-7768 info@phenomenex.com All other countries Corporate Office USA t: +1 (310) 212-0555 f: +1 (310) 328-7768 info@phenomenex.com www.phenomenex.com Phenomenex products are available worldwide. For the distributor in your country, contact Phenomenex USA, International Department at international@phenomenex.com Terms and Conditions Subject to Phenomenex Standard Terms and Conditions which may be viewed at www.phenomenex.com/termsandconditions. Trademarks SecurityCAP is a trademark of Phenomenex. INFICON and Fusion are registered trademarks of INFICON GmbH. S.C.A.T. is a registered trademark of S.C.A.T. Europe GmbH. Agilent is a registered trademark of Agilent Technologies, Inc. VICI Jour is a registered trademark of VICI AG International. Canary-Safe is a trademark of Analytical Sales and Services, Inc. DURAN is a registered trademark of DWK Life Sciences GmbH. VapLock is a trademark of Western Fluids, LLC. Teflon is a registered trademark of The Chemours Company FC, LLC. Disclaimers Phenomenex is not affiliated with the above companies. FOR RESEARCH USE ONLY. Not for use in clinical diagnostic procedures. 2018 Phenomenex, Inc. All rights reserved. TN57651217_W Page 8 of 8