C H I L L E D M I R R O R HYGROMETERS HUMIDITY MEASUREMENT YOU CAN COUNT ON.
1
2
3
4
5
6
7
8
9
M O D E L E 4 The Model E4 is a general purpose chilled mirror dew point hygrometer which provides optional temperature, relative humidity, and pressure measurements. Housed in a gasketed steel NEMA 12 enclosure, the E4 is resistant to the dust and moisture found in many industrial environments. The basic model, Hygro E4-DP, measures dew/frost point. Two additional models are available: the Hygro-E4-RH and the Hygro-E4-PR. Both have dual analog outputs, an assignable alarm relay and the option of one or two displays. The E4-RH measures dew point and temperature, and calculates %RH. All three parameters can be displayed or Model E4 output. The E4-PR measures dew point and pressure, and calculates ppmv, grains per pound, and other parameters. The Hygro E4 can be mated with any of General Eastern s five models of chilled mirror sensors to provide overall dew point measurement range of -80 C to +50 C (-112 F to +122 F). Every E4 is standardly equipped with PACER, General Eastern s patented circuitry for providing long, unattended operation with minimal cleaning. FEATURES State-of-the-art industrial dew/frost point monitor which requires no calibration Multiparameter measurement and display capability; specify only what you need Choice of five, interchangeable chilled mirror sensors NIST Traceability Four-wire PRT measurement for dew/frost point and air temperature 4-20 ma and 0-5 VDC analog outputs with user programmable scaling PACER contaminant error reduction circuitry 10
M O D E L E 4 SPECIFICATIONS PERFORMANCE Accuracy: Complete system at 25 C (77 F) Dew/Frost point: ±0.2 C (±0.36 F) Temperature (optional): ±0.2 C (±0.36 F) Pressure (optional): ±0.5% of full scale Relative Humidity (optional): ±0.5% at 40% RH, ±1.25% at 95% RH (nom) Sensitivity: > 0.05 C (0.09 F) Repeatability: ±0.05 C (±0.09 F) Hysteresis: None Measurement Range: Available chilled mirror sensors: 1111H, D-2, SIM-12H, 1311DR, 1311XR Range: -80 C to +50 C (-112 F to +122 F) dew/frost point, depending on sensor used Required Sample Flow: Approx. 2.0 scfh (1L/min) Response Time: Dew/Frost Point Cooling Rate: 1.5 C (2.7 F)/sec (typical above 0 C [32 F]) Temperature Response: <7 sec (+25 to +70 C [+77 to 158 F] step change) Pressure Response: 1 sec to 90% of steady state (10% to 90% change) Display Resolution: 0.1 (dew/frost point temperature), 0.1% (RH) Update Time: 3 sec FUNCTIONAL Outputs Analog: 4-20 ma DC, 500 ohm max., 0-5 VDC, 5 ma max Analog Scaling (standard): Field or factory programmable Standard Dew/Frost Point Scaling: Sensor C F 111H, D-2-40 to +60-40 to +140 SIM-12H -25 to +100-13 to 212 1311DR -80 to +20-112 to +68 Digital: RS-232C Alarm: Relay (optional): Form C, SPDT, 5A, 250 VAC (resistive load) Service Alarm: TTL compatible. Service flag also available on RS-232C output Balance Status: TTL compatible. Balance (Pacer cycle) flag also available on RS-232C output Displays: 1st Display: Standard: 3-1/2 digit LED, Dew Point Optional: %RH, Grains/Lb, PPMv, Grams/Kg 2nd Display: Standard: 3-1/2 digit LED, Temperature, Pressure Power: 100, 115 or 230 VAC (±10%), 50-60 Hz 50 watts max (specify when ordering) Operating Ranges: Dew Point Sensors: Ambient Temperature: -15 to +95 C (+5 to +203 F), depending on sensor Pressure: 0-300 psia (0 to 22 bar), depending on sensor Electronics: Ambient Temperature: 0 to +50 C (+32 to +122 F), depending on sensor Relative Humidity: 90% max PHYSICAL Dimensions: See Fig. 4 Weight: 11 lbs (5.0kg) Shipping Weight: 12 lbs (5.5kg) approx. Environmental: NEMA-12 water and dust resistant, wall mount Figure 4: Model E4 Dimensions Configurations: Analog Alarm Config. Outputs Displays Relays E4-DP 1 1 0 E4-RH 2 1 1 E4-RH-2D 2 2 1 E4-PR 2 1 1 E4-PR-2D 2 2 1 Accessories T - 100E Temperature Sensor PT-30A/PT-300A Pressure Transducer SSM-1 Sampling System Module BF-10DX Sample line filter (See pages 25-31 for accessory specifications) 11
12
13
14
15
16
17
18
CHILLED MIRROR SENSOR C O M P A R I S O N C H A R T Model 1111H Model D-2 Model SIM-12H Model 1311DR Model 1311XR PERFORMANCE Accuracy (complete system at 25 C (77 F)) Standard +/- 0.2 C +/- 0.2 C +/- 0.2 C +/- 0.2 C +/- 0.15 C Optional +/- 0.15 C +/- 0.15 C +/- 0.15 C +/- 0.15 C Cooling stages 1 2 2 4 5 Depression (at 25 C (77 F), 1 atm, air) 45 C 65 C 65 C 95 C, air 100 C Typical measurement range (at 25 C (77 F), 1 atm, air) (at 85 C) Dew/Frost point -15 to +25 C -35 to +25 C -10 to +85 C -65 to +25 C, air -80 to 25 C -75 to +25 C, liq. RH (equivalent) 6% to 100% 1.5% to 100% 1% to 100% 0.03% to 100%, air 0.0003% to 100% 0.007% to 100%, liq FUNCTIONAL Power From elect. From elect. 115/230 VAC 115/230 VAC 115/230 VAC 75 watts 300 watts 700 watts Operating ranges Ambient temperature -15 to +80 C -25 to +85 C -15 to +50 C 0 to +35 C 0 to +35 C Pressure (psig) -3 to +200-14.7 to +150-3 to +50 0 to +300 0 to +100 Auxiliary cooling No No No Std.-air or liq. Std.-liq. SENSOR BODY MATERIAL Epoxy-coated Stainl. steel Anod. Alum. Stainl. Steel Stainl. Steel Alum. 19
20
21
22
23
24
25
26
27
28
29
30
31
General Eastern Instruments 20 Commerce Way, Woburn, MA 01801 Telephone: 800-225-3208 / 781-938-7070 Fax: 781-938-1071 A40191181/C.00 (ECO #4342) http://www.geinet.com 01/01