UVR-Mi UV-air flow Cleaner - Recirculator

Similar documents
UV-cabinet DNA/RNA UVC/T-M-AR

Orbital Shaker-Incubator ES-80

BT403. A Geno Technology, Inc. (USA) brand name. BT-300 Power Supply. Cat. No. BT

NANOPAC-300 & 500 Power Supply. Instruction manual NANOPAC-300 & NANOPAC-500

nual k Voltage Stabilizer N) manual New Brunswick Voltage Stabilizer Operating manual

Switching DC Power Supply

UV Transilluminator Dual Wavelength (Large)

Triple Output Power Supply

Centrifuge. Model XC-1000 Operating Manual

Source Capture Air Purification System

Operating Instructions

4-Hour. Rapid Charger. for AA/AAA NiMH & NiCd Rechargeable Batteries. Owner's Manual

Snapshot LX5 USER MANUAL. OK on Dimmer Outdoor OK Sound Activated DMX512 Master/Slave 115V/230V Switch Replaceable Fuse User Serviceable Duty Cycle

BroadBand PowerShield. 20 AHr Battery. User Manual

OPERATORS MANUAL FOR USE WITH MAGNETIC/HEATED STIRRER MG200

Operating Instructions for PAC800 Battery Charger

Instruction Manual. Labnet Spectrafuge 16M Microcentrifuge

MINE CHARGER INSTALLATION AND OPERATION MANUAL MODEL NUMBERS: MSP MSQ MSR MSS

INSTALLATION AND SERVICE INSTRUCTIONS FOR 2000 SERIES FLAMEPROOF XENON AND FLUORESCENT BEACONS

Xenon Fiberoptic Lightsource

Model 1672, 1673 Triple Output Power Supply

EXT DUCTLESS FUME HOOD

Diaphragm Vacuum Pumps and Compressors

Diaphragm Vacuum Pumps and Compressors

UView Mini Transilluminator

Operating Instructions. International English. CCS Dust Extractor

Register your instrument! HeatSealer S100. Operating manual

VCM LAMINAR FLOW HOOD VERTICAL

Automatic Burnout Furnaces 115 to 230-volt Models OPERATOR S MANUAL

OPERATIONS MANUAL. Mini Centrifuge Model MCF Certified

Pulsar Evolution 1500 / 1500 Rack 1100 / 1100 Rack 800 / 800 Rack 500 Rack

Operators Manual. Recirculating Chiller /06/08

EXT HEPA/CARBON FUME HOOD

Broadband PowerShield. External Battery Pack 20 Ah 12 V Battery. User Manual

Installation manual wall-mounted distributor

Installation manual portable distributors

Swing Piston Compressors and Vacuum Pumps

4 - SETTING UP THE MACROMAN CHARGING POSITION 4 Wall mounted 4 Workbench set-up 4

PURE SINE WAVE DC TO AC POWER INVERTER

RT Series Step Down Transformer for RT Series UPS 6-10kVA UL Input Vac Output Vac User Guide

Triple Output High Resolution Power Supply. User Manual

9120 EBCU Battery Charger MODELS: 9120EBCU96V 9120EBCU240V

DCell Suction. Model DM Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

Lumitester PD-20 Control Kit

Model 605 Portable Aspirator

Diaphragm Vacuum Pump

DC to AC Power Inverters

SafeFAST Top. Class II Microbiological Safety Cabinets PROTECTION, SAFETY, RELIABILITY. AND MORE.

C.fm Page 1 Thursday, February 21, :28 AM Raptor user manual

PolyStat Immersion Circulators

PRM-2 PHASE SEQUENCE TESTER PRM-3 PHASE SEQUENCE & MOTOR ROTATION TESTER User s Manual

DIGITAL BATTERY TORQUE WRENCH (BC-RAD SELECT) USER GUIDE

P. D. Q. Automatic Burnout Furnaces. 115 and 230-volt Models OPERATOR S MANUAL

Uninterruptible Power System

Cytotoxic Safety Cabinets ENVAIR eco safe Plus

EX Series User Manual

Bio Lion Table Top Centrifuge XC-L5

Pulsar EXtreme CLA 1500C

Instruction Manual. Vibration Calibrator VC20. Manfred Weber. Metra Mess- und Frequenztechnik in Radebeul e.k. Meissner Str D Radebeul

BRAVER UPS. (Uninterruptible Power System) User s Manual

V1606 3U RACK ENCLOSURE

Type MZ15. Operating Instructions. Bedienungsanleitung Manuel d utilisation

The Efficient Solution for Pipetting up to 100 ml at once! USER'S GUIDE

Hybex Microsample Incubator

Register your instrument! MiniSpin /MiniSpin plus. Operating manual

USER MANUAL. Blazer Vista 1000/1400/2000. Uninterruptible Power System

DC Master 24/ A

MODEL A96 SERIES. 130Vdc Switchmode Utility Rectifier / Battery Charger. Used with LaMarche Power Cage ECN/DATE

Power Inverter 400 MW Owner s Manual

Record the serial number and date of purchase in your manual for future reference.

Model 700 Microarray Oven

6L Oil-less Air Compressor 53103

Bio Lion Table Top Centrifuge XC-H165

TXH TOTAL EXHAUST HOOD. Installation and Operations Manual

ACROBAT SWING STAND MODEL

DC to AC Power Inverters

Solar Hybrid Power Generating System CPS1200EOH12SC CPS2200EOH24SC CPS3000EOH24SC. User s Manual K01-C

Model 1693, 1694 Switching DC Power Supply

OWNER S MANUAL. Model YUA2AMPCH 2 AMP Dual-Bank Automatic Battery Charger & Maintainer READ ENTIRE MANUAL BEFORE USING THIS PRODUCT

RC 4/36. English. Printed: Doc-Nr: PUB / / 000 / 02

Quickdraw. Model 2400 Series. Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

Operating Instructions CYR52

Nature Power Inverters. True Sinewave Inverter Modified Sinewave Inverter. Owner s Manual

User Manual Industrial Power Storage System (ips) ips-m Series Battery System

Contents. Overview. Safety Instructions. Before Installation. Installation. Operation. Maintenance. Replacement Parts. System Specifications 8 01

FL 10 DIAPHRAGM PUMP INSTALLATION INSTRUCTIONS. Before operating the pump, please read the Installation Instructions and safety precautions.

Centrifuge. Model XC-2000 Operating Manual

ADI-125/750 ADI-125/1500 ADI-125/2500

PNA FISH Workstation Operating Instructions

Class II biological safety cabinets

System 3 ZB1PS Device Chassis. Operator s Manual

Battery Charger Series EASY Charger

Snapshot. Use on Dimmer Outdoor Use Sound-Activated DMX Master/Slave 120 V/230 V Switchable Replaceable Fuse User-Serviceable.

D.fm Page 1 Wednesday, March 5, :21 PM Raptor user manual

Contents. Introduction Inspection Notes on Safety Precautions. Chapter 1 Overview Product Overview Names of Parts 2

Reach ins, Freeezers & Refrigerators Installation & Operation Manual

INSTRUCTION MANUAL FOR THE CLINICAL 50 CENTRIFUGE

OPK-300N OPTIMAX Multi-Lite

Battery Charger. Series EASY Charger. User s manual

USER MANUAL. Uninterruptible power supply MODEL: 53965,

Transcription:

UVR-Mi UV-air flow Cleaner - Recirculator Medical Biological Research & Technologies Operating Manual Certificate for versions: V.2AA V.2AB

2

Contents 1. Safety Precautions 2. General Information 3. Getting Started 4. Operation 5. Specifications 6. Maintenance 7. Warranty and Claims 8. Declaration of Conformity 3

1. Safety Precautions The following symbols mean: Caution: Caution: GENERAL SAFETY Make sure you have fully read and understood the present Manual before using the equipment. Please pay special attention to sections marked by this symbol. Do not switch on the unit with the cover removed or without filters! UV-lamp must be covered at all times during operation. Otherwise it can expose the operator and other people to a dangerous level of UV light. Use only as specified in the Operating Manual provided. The unit should not be used if dropped or damaged. After transportation or storage keep the unit under room temperature for 2-3hrs before connecting it to the electric circuit. Use only cleaning and decontamination methods recommended by the manufacturer. The unit is designed only for air decontamination. Do not make modifications to the design of the unit. The unit is not moisture-resistant. ELECTRICAL SAFETY Connect only to electric circuit with voltage corresponding to that on the serial number label. Ensure that the switch and the plug are easily accessible during use. Do not plug the unit into an ungrounded power socket, and do not use an ungrounded extension lead. Disconnect the unit from the electric circuit before moving. If liquid penetrates into the unit, disconnect it from the electric circuit and have it checked by a repair and maintenance technician. UV recirculator must be used only in clean rooms because dust accumulation on the electrical parts of the device can cause a short circuit. Do not operate the unit in premises where condensation can form. Operating conditions of the unit are defined in the Specifications section. 4

DURING OPERATION Do not operate the unit in environments with aggressive or explosive chemical mixtures. Please contact manufacturer for possible operation of the unit in specific atmospheres. Do not operate the unit without dust filters installed. Do not operate the unit if it is faulty or has been installed incorrectly. Do not use outside laboratory rooms. BIOLOGICAL SAFETY It is the user's responsibility to carry out appropriate decontamination if hazardous material is spilt on or penetrates into the equipment. 5

2. General Information UV DNA/RNA Recirculators are compact airflow exchange chambers with builtin UV lamps and a fan unit equipped with dust filters and a control unit. UV DNA/RNA Recirculators provide active constant airflow in close vicinity to UV lamps, thus ensuring maximum efficiency of disinfection (see Figure 1). In this version ozone-free 25 W G13 lamps with 9000 hours long life time are used. The software of the control unit allows: -programming a time of switching-on; -switching off the device in a real time mode; -estimation of lamp overall operating time and condition. 20 2 UV intensity, mw/cm 15 10 5 0 50 100 150 200 250 300 Distance to UV source, cm Fig. 1. Dependence of UV intensity on distance to UV source (25W lamp 254 nm) UV radiation affects viability of microorganisms by causing photochemical reactions in the structure of DNA and RNA. Adjacent pyrimidine molecules form dimers and block the reproduction of microorganisms, as a result, causing their death. The diagram (fig.2) shows the process of formation of pyrimidine dimers using thymine as an example. Fig. 2. Photochemical reaction [the formation of pyrimidine dimers; as an example taken thymine] (source http://www.photobiology.info) 6

3. Getting started 3.1. Unpacking. Remove packing materials carefully and retain them for future shipment or storage of the unit. Examine the unit carefully for any damage incurred during transit. The warranty does not cover in-transit damage. 3.2. Complete set. Package contents: Standard set -UVR-Mi UV-air flow Cleaner - Recirculator...1 pce. -spare dust filters...4 pcs. -spare fuse...1 pce. -screw with dowel...1 pce. -Operating Manual, Certificate...1 copy Optional accessories -UVR-S support stand...on request 3.3. Set up: -Choose recirculator location so that the air intake and discharge operate unimpeded, providing the best air circulation in the room. Caution! UV recirculator must be used only in clean rooms because dust accumulation on the electrical parts of the device can cause a short circuit. -Fix the unit on the wall with the screw and dowel, or on a UVR-S movable support stand (Fig.3). -Position the unit so that there is easy access to the power switch and the power plug; attachment point on the unit method of attachment on the wall attachment points on the UVR-S support stand Fig. 3. Unit attachment methods method of attachment on the UVR-S support stand 7

4. Operation 4.1. Plug the power cord to a grounded power socket. 4.2. Switch ON the power switch (fig.4/9) situated on the side of the unit (position I). 4.3. The following parameters will be shown on the display: -Air flow recirculation start time in the automatic mode - Time On (fig.4/1);. -Air flow recirculation end time in the automatic mode - Time Off (fig.4/2); -UV-recirculator operating indicator - Uv lamp (fig.4/3); -Real time - Clock (fig.4/4). 4.4. Hold the Set key (Fig. 4/5) pressed for more than 6 s to activate timer and clock setting mode. Choose the Clock parameter (the active parameter is flashing; each pressing of Set key consecutively activates the parameters). 4.4.1. Use the p (Fig.4/6) and q keys (Fig.4/7) to set required time (note: pressing the key for more than 2 s will increase the values change speed.). 4.4.2. If no key is pressed for more than 10 s, the changes will be saved. 4.5. The unit can operate in automatic or manual modes. 4.6. Operating in manual mode. Press the Run/Stop key (fig.4/8) to start or stop recirculation (automatic mode time interval has priority over the manual mode). During recirculator operation # symbol indication is shown on the display (fig.4/3). 4.7. Operating in automatic mode. Set recirculation start and end time, as mentioned below. 1 4 3 UVR-Mi UV Cleaner-Recirculator 2 Time On UV lamp Time Off 09:00 ^ 12:00 11:35 Clock Set Run/Stop 9 5 6 7 8 Fig.4 Control panel 8

4.7.1 Holding the Set key (Fig. 4/5) pressed for more than 6 s choose the Time On (fig.4/1) parameter (recirculation start time) and, using the p and q keys, set the required start time value in hours, then pressing the Set key switch to the minute value setting. 4.7.2 When the start time is set, switch to setting the Time Off parameter (Fig. 4/2) and input the required values in a similar way. 4.7.3 If no key is pressed for more than 10 s, the changes will be saved. 4.8. The unit saves the changes, and the set parameters will not be lost after switching off the unit. 4.9. The recirculation can be started or stopped at any time by pressing the Run/Stop key. 4.10. Press the Set key to indicate the lamp total run time (Fig. 5). Caution! Holding the Set key (Fig 4/5) pressed in the lamp run time mode for more than 10 s will reset the lamp run time timer. 4.11. If the ERROR UV LAMP FAILURE readout (Fig. 6) is shown on the display or the indicator (Fig. 8/2) is not lit, the lamp or lamps are not functioning. It is necessary to replace UV-lamp (follow the instructions in the paragraph 6.7). Note: If the lamps replacement does not resolve the failure message, contact the service department. 4.12. After 9000 hours operation the"end OF UV LAMP LIFE TIME" readout (Fig. 7) will be shown on the display. UV lamps are to be replaced, follow the instructions in the paragraph 6.7. 4.13. After finishing the operation switch OFF the unit with the switch (position O) and unplug the power cord from electric circuit. Caution! Bactericidal lamp properties reduce by ~12% after 5000 hours of operation (data given by the manufacturer), so lamp replacement or air recirculation time extension by 12% is recommended in such a case in order to achieve the desired air cleaning result. UV lamp TOTAL UV LAMP Time On Clock Time Off RUN TIME 24 HR Fig. 5 Display in lamp total run time mode Time On UV lamp Clock Time Off ERROR UV LAMP FAILURE Time On UV lamp Clock Time Off END OF UV LAMP LIFE TIME Fig. 6 UV-lamps failure Fig. 7 End of UV lamp(s) message life time message 9

5. Specification The unit is designed for operation in cold rooms, incubators and closed laboratory rooms at ambient temperature from +4 C to +40 C in a non-condensing atmosphere and maximum relative humidity 80% for temperatures up to 31 C decreasing linearly to 50% relative humidity at 40 C. 5.1. UV light source...ozone-free 2 x 25 W G13 UV-C 5.2. Wave length...253.7 nm 5.3. 2 Emission intensity (averaged)...36 mw/cm 5.4. Lamp service life...9000 h 5.5. 3 Flow speed with standard filters...14 m /hr 5.6. Dimensions...130x110x660 mm 5.7. Operating voltage / power consumption...230 V, 50 Hz, 110 W (0,5 A) 5.8. Weight*...3.4 kg * Accurate within ±10%. Optional accessories Description Catalogue number UVR-S Movable support stand BS-040105-AK Replacement parts Dust Filter Catalogue number BS-000000S31 Biosan is committed to a continuous programme of improvement and reserves the right to alter design and specifications of the equipment without additional notice. 10

6. Maintenance 6.1. If the unit requires maintenance, disconnect the unit from the electric circuit and contact Biosan or your local Biosan representative. 6.2. All maintenance and repair operations must be performed only by qualified and specially trained personnel. 6.3. Fuse replacement. Disconnect the device from electric circuit. Open the fuse holder located on the side of the unit (fig.8/1) by turning its cover counterclockwise using a screwdriver. Replace with the appropriate fuse according to the unit operating voltage, see table below (type M - time lag: Medium). Unit version V.2AA V.2AB Operating voltage 230 V, 50 Hz 230 V, 50 Hz Fuse rate M 2.0 A M 2.0 A 6.4. Dust filter control/replacement. The dust filters on either end of the UV-recirculator should be checked monthly and cleaned or replaced when they become clogged. Fig. 8. Dust filters location To check, replace or clean the filters, simply unclip the covers (fig.8/), if it is necessary fit a new or rinse in water, dry and set up existing filters. Clip covers back in place. This maintenance operation is performed by the user. 6.5. Cleaning of the outside parts. Standard ethanol (75%) or other cleaning agents recommended for cleaning of laboratory equipment can be used for cleaning and decontamination of the unit. This maintenance operation is performed by the user. 6.6. Cleaning of the inside parts. Cleaning if the inside parts must be performed only by qualified and specially trained personnel. In order to ensure proper efficiency in the long run it is recommended to clean the UV-recirculator once a month. 6.7. UV-lamp replacement Replacement must be performed only by qualified and specially trained personnel. UV- lamp replacement is necessary after lamp stops functioning or at the end of manufacturer specified life time. Use the indicator in the center of recirculator (fig. 8/2) to check operation of UV lamp inside the recirculator. If the indicator is alight while the switch is ON, then the UV lamp is functioning. If it is not, replace the lamp. 2 1 11

7. Warranty and Claims 7.1. The Manufacturer guarantees the compliance of unit with the requirements of Specifications, provided the Customer follows the operation, storage and transportation instructions. 7.2. The warranted service life of unit from date of delivery to the Customer is 24 months (exclude UV-lamps and dust filters). Contact your local distributor to check availability of extended warranty. 7.3. If any manufacturing defects are discovered by the Customer, an unsatisfactory equipment claim shall be compiled, certified and sent to the local distributor address. Please visit www.biosan.lv, Technical support section to obtain the claim form. 7.4. The following information will be required in the event that warranty or postwarranty service comes necessary. Complete the table below and retain for your records. Model Serial number Date of sale UVR-M UV-airflow Cleaner - Recirculator 12

13

14 8. Declaration of Conformity

15

Biosan SIA Ratsupites 7, build. 2, Riga, LV-1067, Latvia Phone: +371 6742 6137 Fax: +371 6742 8101 http://www.biosan.lv Version 2.04 - January 2015