Paddle Wheel Turbine Flow Meters Installation, Operating & Maintenance Manual

Similar documents
Paddle Wheel Turbine Flow Meters Installation, Operating & Maintenance Manual

TECHNICAL SPECIFICATIONS. Measuring Accuracy ±2% of full scale. Repeatability ±0.5% of full scale

Installation, Operation & Maintenance Manual

BASIC IN-LINE LIQUID FLOW RATE MONITORS. Data Sheet No. B-699. Select from aluminum, brass or stainless steel to meet system and liquid requirements.

Sanitary Turbine Flow Meter - TA3 Installation, Operating & Maintenance Manual

COMPANY. Pneumatic Flow Meters Installation, Operating & Maintenance Manual AW-Lake Company. All rights reserved. Doc ID:PNEUMAN082516

TR-QS Wafer-Style Turbine Flow Meters Installation, Operating & Maintenance Manual

200 Series Flow Meters

100 Series Flow Switches Technical Reference Manual

EX80 Series EX81 EX82. E X 8 0 S E R I E S E l e c t r o m a g n e t i C F l o w s e n s o r i n s t r u c t i o n s. The Leader in Flow Meter Value.

MPV1 INSTALLATION & MAINTENANCE INSTRUCTIONS

MPV1 INSTALLATION & MAINTENANCE INSTRUCTIONS

ARCHIVED. EX80 Series. (Includes Dates 6/28/05 to 10/06) EX81 EX82

RotorFlow Sensors Provide Visual Indication, Continuous Sensing and Accurate Switching

RotorFlow Sensors Provide Visual Indication, Continuous Sensing and Accurate Switching

COMPANY. Liquid Flow Meters Installation, Operating & Maintenance Manual AW-Lake Company. All rights reserved. Doc ID:FLOWLIQUIDMETERMAN082516

MM1/MM2 INSTALLATION & MAINTENANCE INSTRUCTIONS

RotorFlow Sensors Provide Visual Indication, Continuous Sensing and Accurate Switching

2100 Turbine Flow Sensor

PRODUCT INFORMATION BULLETIN

500 SERIES METERING FLOW SWITCHES TECHNICAL REFERENCE MANUAL. Proteus Industries Inc. Proteus 5-Year Warranty

100 Series Flow Switches

M Series. Feature rich flow meter and switch with in-line. flow

700 Series FluidVision

EZ-View Flow Meters. General Design Features OPERATING PRINCIPLE

TX80-Series FLOW SENSOR INSTRUCTIONS TX81 TX :2008 ISO CERTIFIED COMPANY TX80-SERIES FLOW SENSOR INSTRUCTIONS

IP80-Series FLOW SENSOR INSTRUCTIONS IP81 IP :2008 ISO CERTIFIED COMPANY IP80-SERIES FLOW SENSOR INSTRUCTIONS

FAIRCHILD T5420 DIGITAL-PNEUMATIC TRANSDUCER Installation, Operation and Maintenance Instructions

COMPANY. Flow Rate Alarms Installation, Operating & Maintenance Manual

BULLETIN NO.ELEC IM121/10A Replaces IM121/09A

DW PADDLE-BELLOWS FLOWMETER & SWITCH

Insertion turbine INSTRUCTION SHEET. TECHNICAL PRODUCT

BI-DIRECTIONAL INSERTION FLOW TRANSDUCER

PSF50 Paddle flow switch

Electromagnetic Flow Monitor magphant

F-4600 INLINE ULTRASONIC FLOW METER Installation and Operation Guide

LTX RF LEVEL SENSOR. Instruction Manual

IP800-Series FLOW SENSOR INSTRUCTIONS IP810 IP :2008 ISO CERTIFIED COMPANY IP800-SERIES FLOW SENSOR INSTRUCTIONS

Hydro-Flow Model 2200 Fixed Insertion Vortex Flowmeter Installation and Operation Manual

DW PADDLE-BELLOWS FLOWMETER & SWITCH

1100 Series Piston Type Differential Pressure Gauges

IP800-Series FLOW SENSOR INSTRUCTIONS IP810 IP :2008 ISO CERTIFIED COMPANY IP80-SERIES FLOW SENSOR INSTRUCTIONS

Model 1100 Turbine Flow Meter

CONNECTOR AMPLIFIER FOR PROPORTIONAL VALVES (0-10VDC Input Version)

DW PADDLE-BELLOWS FLOWMETER & SWITCH

MODEL 1100 TURBINE FLOW METER

Electronic Dynamo Regulator INSTRUCTION MANUAL. COPYRIGHT 2014 CLOVER SYSTEMS All Rights Reserved

QB4 PRESSURE CONTROL VALVE INSTALLATION & MAINTENANCE INSTRUCTIONS

Hydro-Flow 2200 Fixed Insertion Vortex Flow Meter

T21 Thermal Differential Switch

ELECTRONIC CONTROLLER

CP (Vdc) CP (Milliamp) CP (Slidewire)

User's Manual: Series 270I Model 270I Process Current Loop-Powered Isolator

LEVEL AND FLOW MEASUREMENT WS131 FLOW METER USER MANUAL.

Hydro-Flow Model 2200 Installation and Operation Guide B

IP800-Series. Insertion Paddlewheel Flow Sensor Instructions PROUDLY MADE IN THE USA

ACTIVE ELECTRODELESS SENSORS Instruction Manual

Carbon Monoxide Transmitter

Hydro-Flow Model 3100

Hydro-Flow Model 3100 Retractable Insertion Vortex Flowmeter Installation and Operation Manual

Paddle-wheel flow controller for On/Off control

Operating Instruction for Plastic Vortex Flow Transmitter. Model: UV

FM200 FLOW/NO-FLOW MONITOR INSTALLATION AND TECHNICAL MANUAL 120 VAC MODEL 6/1/2017

Model 124 Filter Minder & Model 124

EZ-View. Flow Meters ISO CERTIFIED 9001

KEL DIFFERENTIAL PRESSURE FLOWMETERS

Hydro-Flow Model 2300 Installation and Operation Guide B

EZ-View Flow Meters. With Flow-Alert TM Flow Switch REDUCE MAINTENANCE COSTS TEL

Ultrasonic Two-Wire Point Level Control

IP800-Series. Insertion Paddlewheel Flow Sensor Instructions PROUDLY MADE IN THE USA

Electronic Dynamo Regulator INSTRUCTION MANUAL. COPYRIGHT 2014 CLOVER SYSTEMS All Rights Reserved

Liquid Turbine Flowmeter

Series BFM Bulk Flow Monitor 1A FUSE RED LED RELAY. Part Number BFM-1 BFM-2 BFM-3 BFS-1

KOBOLD DF-MA Flow Transmitter. User Instructions

Condensed Version of Operator s Manual For The VI/VP700 Series Flowmeter and VI800 Series Flowmeter

Condensed Version of Operator s Manual For The VL600 Series Flowmeters

RUN ACCUM. TOTAL STOP BAT LOW HIGH

BI-DIRECTIONAL INSERTION METER

GTFM IV. Operation / Installation Manual. Gas Turbine Flow Monitor. Computer Weld Technology, Inc. Manual Part Number: A8M5010 Revised: June 26, 2008

GE Sensing & Inspection Technologies

Troubleshooting Bosch Proportional Valves

Table of Contents. Proportional & Precision Regulator Instrumentation

Figure 1: WPR2 Dimensions and Hardware

MODEL 106F TEFLON LIQUID FLOW SENSOR

Low Profile Insertion Vortex

SIGNET 7002 Vortex Flow Sensor ENGLISH

CP-8551 & CP Application. Features. Applicable Literature. Electronic to Pneumatic Transducers General Instructions

MODEL U801 PTFE Flo-Controller for UHP Liquid Applications. uhp

Chapter Page. 1 Introduction Description 1

Paddle-wheel flow controller for On/Off control

DPT TARGET TYPE FLOWMETERS

CP-8301-xxx Series. Application. Features. Applicable Literature. 1 to 15 Vdc Input Electronic Actuator Drive General Instructions

Signet 7000/7001 Vortex Flow Sensors * *

CP-1TT Remote System

G2 Industrial Grade Meters

BOILER BLOWDOWN CONDUCTIVITY CONTROLLER (CONTINUOUS METHOD)

Hot Tap Insertion Vortex

Precision Regulators.

Flow Monitors Excess Flow Valves Flow Meters. ChemTec.com

CP-9301 CP Application. Features. Applicable Literature. Electronic Actuator Drives General Instructions

Transcription:

COMPANY Paddle Wheel Turbine Flow Meters Installation, Operating & Maintenance Manual

Mechanical Specifications Flow Meter with Polypropylene Body (ES version) Maximum Operating Pressure: 150 PSIG Maximum Operating Temperature: 20-150 F Flow Measuring Range: 0.5-15 GPM (2-60 LPM) Turn Down Ratio: 10:1 Wetted Components Component Casing Cover Seal Turbine Bearing Shaft Materials Glass-Filled Polypropylene Clear Polycarbonate Buna-N (Other options available) Acetal Copolymer PEEK (Polyetheretherketone) Stainless Steel Non-Wetted Components Component Encapsulant Strain Relief Lock Ring Wire Insulation Materials Epoxy Nylon Glass-Filled Polypropylene High-Temperature PVC 2

Flow Meter with Stainless Steel Body Maximum Operating Pressure: With Clear Cover: to 200 PSIG (14 bar) With Optional Stainless Steel Cover: to 500 PSIG (34 bar) Fluid Temperature Range: 20-225 F (-7 to 107 ) Flow Measuring Range: 1/2 porting: 0.5 15 GPM (2 60 LPM) 3/4 1 porting: 1.5 50 GPM (60 200 LPM) Turn Down Ratio: 10:1 Wetted Components Component Materials Casing Stainless Steel 316 Cover Stainless Steel 316 (optional clear polycarbonate) Seal Turbine Bearing Shaft Buna-N (other options available) Acetal Copolymer PEEK (Polyetheretherketone) 316 Stainless Steel Non-Wetted Components Component Materials Encapsulant Epoxy Strain Relief Lock Ring Wire Insulation Nylon Stainless Steel High-Temperature PVC Note: Buna-N is a registered trademark of Chemische Werke Huls. 3

Electronic Specifications Power Requirements Load driving capacity Maximum Transmission Distance 4-20 ma version 12-35 VDC, loop powered Use the following equation to calculate maximum load resistance: Max Loop Load ( ) = 50 (Power supply volts 12). Limited only by wire resistance & supply voltage Response time Resolution Over-current limit Other protection 2 seconds to 90% (step change) Infinite Self limiting at 35 ma Reverse polarity Power Requirements Maximum Transmission Distance Switch Contact Hysteresis Set Point Repeatability Relay Output 12-35 VDC 200 feet recommended Form C, 5A max 120 or 240 VAC 5% of set point maximum 1% of full scale Power Requirements Maximum Current Minimum Load resistance Maximum Transmission Distance 0-5 VDC version 12-35 VDC 25 ma DC 1000 Ohms 200 feet recommended Resolution Response time Infinite < 5 seconds to 90% (step change) 4

Power Requirements Response Time Maximum Current Pulse Output Version 12-24 VDC <100 ms 25 ma DC Maximum Transmission Distance 200 feet recommended Minimum Load Resistance Protection 1000 Ohms Short circuit & reverse polarity Mechanical Dimensions + A F B C E D D* DIM 1/2 NPTF 3/4 NPTF - 1 NPTF A 1.94 (49mm) 3.06 (78mm) B 1.13 (29mm) 1.33 (34mm) C 2.00 (51mm) 2.46 (62mm) D 2.45 (62mm) 2.78 (71mm) D* 2.45 (62mm) 2.88 (73mm) E 3.70 (94mm) 5.25 (133mm) F 2.63 (67mm) 3.80 (97mm) *Dimensions with clear polycarbonate cover installed. 5

Introduction This manual is a service guide produced by the manufacturer. The manual provides specific procedures and/or illustrations for installation, inspection, cleaning and filtration of all KW Series flow meters. When properly followed, these procedures will keep your paddle wheel flow meter operating dependably for many years. It is important for operators and maintenance personnel to be safety conscience when operating or repairing equipment. Developing a thorough knowledge of the precautionary areas and following safe operating procedures, can prevent equipment damage and/or personal injury. Before making any repair, read all of the repair procedures to learn the correct method and all precautions. Description of Operation The paddle wheel flow meters integrate rugged tangential turbine technology with a precision electronic output signal hermetically encapsulated within the body of the meter. The flow transmitter is ideal for measuring flow rates in cooling circuits, HVAC systems and batching operations. Additional features are: Simple in-the-field serviceability of its moving components. The integral 4-20mA circuit design eliminates the need for separate signal conditioning modules. Units come factory calibrated to your system requirements. The low impedance 4-20mA circuit can transmit a clean signal over lowcost wire for several thousand feet without degradation. Accurately measures flow in both directions. Installation - Mechanical 1. The KW flow meter has NPT (National Pipe Thread) plumbing connections. Lake Monitors recommends that a paste type pipe sealant be used on these threads. [Teflon tape sealant can be used as long as it is applied in such a way that it will not enter the flow stream. Pieces of Teflon tape can wrap around the turbine andimpede its rotation.] 6

2. The recommended mounting orientation would be any plane that will place the axis of the turbine horizontal with respect to the ground. The unit will operate satisfactorily with its axis at angles other than horizontal, but side loading of the bearing surfaces will lead to premature wear of the rotating parts. 3. For optimum flow meter performance a minimum length of upstream and downstream piping is required. It is recommended that a minimum length equal to 10 pipe diameters of straight pipe be installed directly on the upstream side of the flow meter and 5 pipe diameters on the downstream side of the flow meter. This helps to eliminate turbulence in the fluid. Having shorter pipe lengths, other system components and elbows to close to the flow meter may adversely affect the accuracy and repeatability of the flow meter. Piping should be the same size as the meter bore or port size. Do not locate the flow meter or the connection cable close to electric motors, transformers, sparking devices, high voltage lines or place connecting cable in a conduit with wire supplying power for such devices. These devices can 7

induce false signals in the flow meter coil or cable, causing the meter to read inaccurately. 4. It is recommended that the meter be installed in such a position that the round access cover can be removed for cleaning and turbine servicing. It is also recommended that a union be placed near the sensor to allow easy removal. Installation - Electronic (4-20mA Only) The KW circuit is a two-wire loop-powered design that transmits a 4-20mA signal that is proportional to flow rate. The noise-immune current transmission from the sensor can be routed with low cost two conductor twisted-pair cable. The circuit operates on 12-35 Vdc and requires a source capable of supplying at least 24mA of current. The circuit has built in polarity protection and over-current limiting to protect both the transmitter and what the transmitter is connected to. 1. Connect the red wire (pin 1 for units with the electronic disconnect option) from the transmitter sensor to the positive 12-35 Vdc power supply output. 2. Connect the black wire (pin 2 for units with the electronic disconnect option) from the transmitter to the positive side of the loop load (resistor, chart recorder, data acquisition board, meter, etc.). This connection may be labeled 4-20mA Input or 4-20mA (+) on some devices. 3. If applicable, connect the negative side of the loop load to negative side of the power supply. 4. Apply power to the system. 5. If everything is operating correctly, the green LED on the meter will illuminate and 4mA will be flowing in the loop. As the flow rate increases, the intensity of the green LED will increase in proportion to the ma output signal for trouble shooting purposes. If the LED does not illuminate: Check wiring terminations for good connections Check wiring polarity Verify correct supply voltage Ensure that the load impedance is within allowable limits Apply the DC supply voltage directly across the sensor wires. If the LED does illuminate, the load is either: too great of impedance or an open circuit. If the LED does not illuminate, the sensor s lead wires or circuit are defective. 8

Installation - Electronic (0-5Vdc Output Only) 1. Connect the red wire (pin 1 for units with the electronic disconnect option) from the sensor to the positive terminal of the 12-35 VDC power supply. 2. Connect the black wire from the transmitter (pin 2 for units with the electronic disconnect option) to the negative terminal of the 12-35 VDC power supply. 3. Connect the green wire from the transmitter (pin 3 for units with electronic disconnect) to the (+) 0-5 Vdc input of the data acquisition device. This connection may be labeled Voltage Input or Analog Input on some devices. 4. If applicable, connect the negative side of the power supply to the negative side of the pulse input. 5. Apply power to the system. 6. If everything is operating correctly, the green LED will illuminate brightly and the data acquisition device should show an increase in flow rate as fluid starts flowing through the sensor. If data acquisition device does not show an increase in flow rate: Check wiring terminations for good connections Verify that the LED is illuminated Verify that the DC supply voltage is between 12 and 35 Vdc Installation Electronic (Pulse Output Only) The KW pulse output circuit is a three-wire DC-powered design that transmits a frequency proportional to flow rate. The circuit operates on 5-24 Vdc and will consume a maximum of 25 ma from the power supply. 1. Connect the red wire (pin 1 for units with the electronic disconnect option) from the sensor to the positive terminal of the 5-24VDC power supply. 2. Connect the black wire from the transmitter (pin 2 for units with the electronic disconnect option) to the negative terminal of the 5-24 Vdc power supply. 3. Connect the green wire from the transmitter (pin 3 for units with electronic disconnect) to the (+) pulse input of the data acquisition device. This connection may be labeled Pulse In or DC Input on some devices. 4. If applicable, connect the negative side of the power supply to the negative side of the pulse input. 5. Apply power to the system. 6. If everything is operating correctly, the data acquisition device should begin counting pulses when fluid starts flowing through the sensor. 9

If the data acquisition device is not counting : Check wiring terminations for good connections Verify correct supply voltage and current 7. The installation should be complete. The relationship between the frequency output and flow rate is shown in the graphs above. Installation Electronic (Relay Output Only) 1. Connect the red wire from the sensor to the positive terminal of the 12-35 Vdc power supply. 2. Connect the black wire from the transmitter to the negative terminal of the 12-35 Vdc power supply. 3. (Versions with stainless steel sensor body only) Be sure to properly ground the KW sensor housing by connecting a ground wire to the sensor body. This step may be unnecessary if the piping system that the flow sensor is plumbed into is already connected to Earth ground. 4. Wire the appropriate relay contacts to the load that is to be switched per Table 1 below. 5. If the unit is operating correctly, the green power indication LED and the red relay status LED should illuminate after the DC supply voltage is turned on. As fluid flow is increased above the factory-adjusted set point, the red relay status LED should turn off. If the LEDs do not illuminate when power is applied: Check wiring terminations for good connections Verify that the DC supply voltage is between 12 and 35 Vdc Table 1: Wiring Connections Wire Color Connection Red +12-35 Vdc Black DC Ground Green Relay Common White Relay Normally Closed Contact Brown Relay Normally Open Contact Note: Relay condition shown is with power on & flow rate above setpoint. 10

Set Point Adjustment (Relay Output Only) 1. Adjust the flow rate through the line in which the KW meter is installed to the rate that corresponds to the desired relay trip point. 2. If the red LED on the back side of the sensor is not illuminated, use a small flat bladed screwdriver to slowly turn the adjustment screw on the set point potentiometer counter-clockwise until the red LED illuminates. 3. If the red LED is already illuminated, turn the adjustment screw on the potentiometer clockwise until red LED turns off. Next, slowly rotate the adjustment screw counter-clockwise until the red LED illuminates. 4. Once the set point has been adjusted, the relay will operate as shown in the Truth Table illustrated in Table 2 below. Table 2: Truth Table - Relay Operation Condition Flow rate < set point Flow rate > set point Loss of power to sensor Relay NC Contact Relay NO Contact Red LED Green LED Open Closed Illuminated Illuminated Closed Open Not Illuminated Illuminated Open Closed Not Illuminated Not Illuminated Service and Maintenance The sensor is designed to provide years of low maintenance service in industrial environments. As with all mechanical rotating devices, the bearing surfaces will wear with use. The life of the parts will depend on factors such as cleanliness of the fluid, media, mounting orientation, temperature, fluid velocity and frequency of operation. The KW meter was designed with simple field-replacement of the rotating parts in mind. To inspect or replace the rotating components: 1. Relieve pressure in the piping system. 2. Remove the retainer ring that secures the turbine access cover. 3. Remove the access cover with pliers, taking care not to damage the o-ring seal. 4. Pull out the turbine assembly and the shaft. 5. Inspect the shaft for things that may have wrapped around it. 6. Inspect the turbine bearing surface for wear and elongation. Replace as 11

necessary. 7. Clean any rust off of the magnets that may have accumulated. 8. Reassemble the unit by placing the turbine into the body cavity with the two magnet pockets facing inward. Place the shaft into the turbine hole and guide it into the retaining hole in the body cavity. Lubricate the o-ring seal with some glycerin or other lubricant and press it into the pocket of the body. Replace the retaining ring securely before applying pressure to the system. Circuit recalibration (4-20mA version): 1. Place a milliamp meter into the current loop. 2. Turn off the flow going though the meter. Adjust the OFFSET control for a reading of 4mA on the milliamp meter. 3. Adjust the flow rate though the meter to full flow rate. Adjust the SPAN control for a reading of 20mA on the milliamp meter. The two controls are not interactive, so further adjustment should not be necessary. Circuit recalibration (0-5Vdc version): 1. Place a voltmeter across the black & green wires of the sensor pigtail. 2. Turn off the flow going though the sensor. Verify a reading of less than 50 mvdc on the voltmeter. 3. Adjust the flow rate though the meter to full-scale flow rate. Adjust the SPAN control for a reading of 5 Vdc on the voltmeter. 12

COMPANY 262.884.9800 www.aw-lake.com 8809 Industrial Drive, Franksville, WI 53126 13