Technical Procedure for Gas Chromatography (GC-FID)

Similar documents
Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS)

Technical Procedure for the Examination of Arson Evidence

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

Analysis of Petroleum Fractions by ASTM D2887

Using the PSD for Backflushing on the Agilent 8890 GC System

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

High-Temperature Simulated Distillation System Based on the 6890N GC Application

Model 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE

Simulated Distillation Analyzers, Software, Standards, Consumables, Training

GC-101. Testing of Biodiesel Using Gas Chromatography. by Bob Armantrout. Presented at: Local Biodiesel: A Biodiesel Coop Conference

MET-Biodiesel Capillary GC Columns

GC Method Compliance and Large Valve Oven Application

Agilent 7693A Automated Liquid Sampler

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.

Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584

Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System

Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector

Agilent Multimode Inlet for Gas Chromatography

Agilent Multimode Inlet

Agilent 7693A Automated Liquid Sampler

Application Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER

Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis

SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS

Application. Gas Chromatography June 1995

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography

Agilent G2855A Deans Switching System

Analysis of Glycerin and Glycerides in Biodiesel (B100) Using ASTM D6584 and EN Application. Author. Abstract. Introduction

Performing ASTM 6584 free and total glycerin in BioDiesel using an SRI Gas Chromatograph and PeakSimple software

Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC

The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector

6890 CHECKOUT PROCEDURE TCD (THERMAL CONDUCTIVITY) DETECTOR

Application Note. Abstract. Authors. Environmental Analysis

Agilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass. James D. McCurry, Ph.D. Senior Scientist

Practical Steps in GC Troubleshooting

Choose for Your Instrument. GC Supplies - Agilent Instrument Quick Pick Guide

GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method

Method Development for Capillary GC Systems. Slide 1

Beverage Grade Carbon Dioxide

Fast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC

DETERMINATION OF N-BUTANOL AND ISOBUTANOL IN GASOLINE USING GAS CHROMATOGRAPHY (GC-FID)

Refinery Gas. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Agilent 6850 Series II Network GC System G2630A Performance Specifications

Techniques for Improving the Reproducibility of GC Analysis

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)

White Paper. Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Introduction. Background Information

Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter

Application Note. Author. Introduction. Energy and Fuels

DANI Transformer Oil Gas Analyzer

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585

SPECIFICATION. Bench top triple quadrupole Mass Spectrometry System coupled to a gas performance Chromatography (GC-MS/MS)

Agilent and ASTM. Update on Recent Activities. Page 1

Application Note. Author. Abstract. Energy & Chemicals - Petrochemicals. Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

Optimized Method Development of Large Volume Injection for GC/MS/MS of Food Pesticides

Analysis of Fatty Acid Methyl Esters (FAMES), and Examination of Biodiesel Samples for these Components, by GCxGC-FID

Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application

Choose for Your Instrument

Setting up SilFlow for BackFlush in your GC

Forensic Blood Alcohol Determination with the Intuvo 9000 GC

Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105

AppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS

Large Volume Injection of Polycyclic Aromatic Hydrocarbons

Antek Gas Chromatography Products: A Guided Tour Through Industry

Application Note. Authors. Abstract. Energy & Chemicals

Oil Dries Used in Arson Scenes: New Absorbent Material

2-D GC/MS and the Deans Switch: Theory and Applications. Fred Feyerherm, Agilent Technologies Inc. Houston, Tx

Typical Gas Chromatographic System

Characterization of Tiki Torch Fuels

ASTM D 6730 Detailed Hydrocarbon Analysis

ANALYSIS OF GASOLINE RANGE HYDROCARBONS ON BP1-PONA

Techniques for Making Your GC Analysis More Repeatable, Reproducible and Robust

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS

Determination of Sudan Dyes I IV in Curry Paste

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

Choose for Your Instrument. GC Supplies - Bruker (Varian) Instrument Quick Pick Guide

ASTM D for Denatured Fuel Ethanol Automating Calculations and Reports with Empower 2 Software

Analysis and. Separation of Oxygenates in Hydrocarbon Matrices. Simon Jones Application Engineer Folsom, CA

4001 Transesterification of castor oil to ricinoleic acid methyl ester

Method Detection Limits for EPA Method 8015 Diesel Range Organics using Fully Automated Extraction and Concentration

Techniques for Making Your GC Analysis More Repeatable, Reproducible and Robust

Liquefied Gas Injector. Solution for the Sampling and Analysis of Liquefied Gases

Detailed Hydrocarbon Analysis Featuring Rtx -1 PONA Columns

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

Productivity Improvements for Dual Column Applications: USP <467> and Blood Alcohol Analysis

Restek Corporation 110 Benner Circle, Bellefonte, Pa Barry L. Burger, Neil Johansen, Valerie Gamble, Donald Rhoades

AZ Chrom s.r.o.

GC Inlets. An Overview. Simon Jones GC Applications Engineer

Fausto Munari e Andrea Cadoppi ThermoFisher - Italy

The Stability of Sulfur Compounds, Low Molecular Weight Gases, and VOCs in Four Air Sample Bag Materials

Installing a Cool On-Column Inlet

Transcription:

Technical Procedure for Gas Chromatography (GC-FID) 1.0 Purpose This technical procedure shall be followed for the operation of the gas chromatograph (GC- FID). 2.0 Scope This procedure applies to all gas chromatograph-flame ionization detectors set up for liquid samples in the Trace Evidence Section. 3.0 Definitions N/A 4.0 Equipment, Materials, and Reagents 4.1 Equipment Agilent 6890 Gas Chromatograph with a Flame Ionization Detector Agilent 7890 Gas Chromatograph with a Flame Ionization Detector 4.2 Materials Auto sampler vials, 100 µl inserts, and crimp seals Vial crimper Vial de-crimper 10 micro-liter autosampler syringe A non-polar capillary column with a (5 %-Phenyl)-methylpolysiloxane stationary phase such as a DB-5 or HP-5 50 % Residual Gasoline reference material 100 % #2 fuel oil/diesel fuel reference material 4.3 Reagents 5.0 Procedure Acetone Carbon disulfide - Reagent A.C.S. grade Petroleum ether Optima Grade Hydrogen gas Air Zero grade Nitrogen UHP grade 5.1 Start-up and Calibration 5.1.1 The GC-FID shall be kept on at all times. 5.1.2 Monthly GC-FID Performance Check 5.1.2.1 To ensure performance, the following samples shall be analyzed and the chromatographs electronically archived. Petroleum ether blank. Page 1 of 7

Carbon disulfide adsorption elution blank. 50 % residual gasoline reference material. 100 % # 2 fuel oil/diesel fuel reference material. 5.1.2.2 If there is a shift in peak retention time by 0.2 minute or more or a change in peak resolution, maintenance shall be performed or a service engineer must be called. Once maintenance has been performed, the samples shall be reexamined. If the resulting chromatographs are within the limits, the instrument may be used for casework. 5.1.3 New Instrument Setup and Performance Verification: 5.2 Casework Analysis 5.1.3.1 New GC-FID instruments shall be installed by a certified engineer according to the manufacturer s guidelines. 5.1.3.2 The samples used in the monthly GC-FID performance check shall be analyzed on the instrument and the resulting chromatographs compared to the same samples acquired on similar instruments. The resulting chromatographs shall have similar peak ratios and component separation in order for the new instrument to be utilized for case work. 5.2.1 Samples shall be prepared for examination according to the appropriate technical procedure. 5.2.2 Once the samples are prepared, they shall be analyzed using one of the GC-FID programs located in 5.3 and 5.4. 5.2.3 Once data has been collected, the resulting data file shall be loaded in the Data Analysis program. The resulting chromatograph shall be included in the Case Record. 5.2.4 After the method has run, allow the instrument to return to the set program and leave the GC-FID on. 5.3 Agilent 6890 Gas Chromatograph (Raleigh Laboratory) Program parameters 5.3.1 ARSONL.M 5.3.1.1 Injection Using the autosampler with a 10 µl syringe injection, perform eight solvent washes from each wash bottle. Inject a 1.0 µl sample with a fast plunger speed, three sample washes, and three sample pumps. 5.3.1.2 Inlet Run in split mode at 240 C with a 50:1 split ratio. 5.3.1.3 Carrier gas Hydrogen. 5.3.1.4 Oven Run a temperature program starting at 80.0 C for 0.00 minutes. Page 2 of 7

5.3.1.5 Column Use a HP-5MS or DB-5MS column which is 0.25 mm in diameter that is approximately 30 m long with a 0.25 µm film thickness. The column shall be kept at a constant flow of 1.6 ml/min. 5.3.1.6 Flame Ionization Detector The detector temperature shall be set at 300 C. 5.3.2 ARSON.M 5.3.2.1 Injection Using the autosampler with a 10 µl syringe injection, perform eight solvent washes from each wash bottle. Inject a 5.0 µl sample with a fast plunger speed, no sample washes, and three sample pumps. 5.3.2.2 Inlet Run in split mode at 240 C with a 50:1 split ratio. 5.3.2.3 Carrier gas Hydrogen. 5.3.2.4 Oven Run a temperature program starting at 80.0 C for 0.00 minutes. 5.3.2.5 Column Use a HP-5MS or DB-5MS column which is 0.25 mm in diameter that is approximately 30 m long with a 0.25 µm film thickness. The column shall be kept at a constant flow of 1.6 ml/min. 5.3.2.6 Flame Ionization Detector The detector temperature shall be set at 300 C. 5.4 Agilent 7890 Gas Chromatograph (Raleigh Laboratory) Program Parameters 5.4.1 ARSONL.M 5.4.1.1 Injection Using the autosampler with a 10 µl syringe injection, perform three pre-inject and eight post-inject solvent washes from each wash bottle. Inject a 1.0 µl sample with a 6000 µl/min plunger speed, three sample washes, and three sample pumps. 5.4.1.2 Inlet Run in split mode at 240 C with a 50:1 split ratio. 5.4.1.3 Carrier gas Hydrogen. 5.4.1.4 Oven Run a temperature program starting at 80.0 C for 0.00 minutes. Page 3 of 7

5.4.1.5 Column Use a HP-5 or DB-5 column which is 0.32 mm in diameter that is approximately 30 m long with a 0.25 µm film thickness. The column shall be kept at a constant flow of 1.6 ml/min. 5.4.1.6 Flame Ionization Detector The detector temperature shall be set at 300 C. 5.4.2 ARSON.M 5.4.2.1 Injection Using the autosampler with a 10 µl syringe injection, perform three pre-inject and eight post-inject solvent washes from each wash bottle. Inject a 5.0 µl sample with a 6000 µl/min plunger speed, no sample washes, and three sample pumps. 5.4.2.2 Inlet Run in split mode at 240 C with a 50:1 split ratio. 5.4.2.3 Carrier gas Hydrogen. 5.4.2.4 Oven Run a temperature program starting at 80.0 C for 0.00 minutes. 5.4.2.5 Column Use a HP-5 or DB-5 column which is 0.32 mm in diameter that is approximately 30 m long with a 0.25 µm film thickness. The column shall be kept at a constant flow of 1.6 ml/min. 5.4.2.6 Flame Ionization Detector The detector temperature is set at 300 C. The flow rate for hydrogen is 30.0 ml/min, for air is 400.0 ml/min, and the 5.5 Agilent 6890 Gas Chromatograph (Western Regional Laboratory) Program parameters 5.5.1 ARSONL.M 5.5.1.1 Injection Using the autosampler with a 10 µl syringe injection, perform eight solvent washes from each wash bottle. Inject a 1.0 µl sample with a fast plunger speed, three sample washes, and three sample pumps. 5.5.1.2 Inlet Run in split mode at 240 C with a 50:1 split ratio. 5.5.1.3 Carrier gas Hydrogen. 5.5.1.4 Oven Run a temperature program starting at 80.0 C for 0.00 minutes. 5.5.1.5 Column Use a (5 %-Phenyl)-methylpolysiloxane (such as HP-5MS or DB- 5MS) column which is 0.25 mm in diameter that is approximately 30 m long Page 4 of 7

with a 0.25 µm film thickness. The column shall be kept at a constant flow of 1.6 ml/min. 5.5.1.6 Flame Ionization Detector The detector temperature shall be set at 300 C. makeup flow of helium is 25.0 ml/min. 5.5.2 ARSON.M 5.6 Standards and Controls 5.5.2.1 Injection Using the autosampler with a 10 µl syringe injection, perform eight solvent washes from each wash bottle. Inject a 5.0 µl sample with a fast plunger speed, no sample washes, and three sample pumps. 5.5.2.2 Inlet Run in split mode at 240 C with a 50:1 split ratio. 5.5.2.3 Carrier gas Hydrogen. 5.5.2.4 Oven Run a temperature program starting at 80.0 C for 0.00 minutes. 5.5.2.5 Column Use a (5 %-Phenyl)-methylpolysiloxane (such as HP-5MS or DB- 5MS) column which is 0.25 mm in diameter that is approximately 30 m long with a 0.25 µm film thickness. The column shall be kept at a constant flow of 1.6 ml/min. 5.5.2.6 Flame Ionization Detector The detector temperature shall be set at 300 C. makeup flow of helium is 25.0 ml/min. 5.6.1 There shall be at least one solvent blank run for each case. 5.6.1.1 If the sample being analyzed is a neat sample, petroleum ether or carbon disulfide shall be used for the blank. 5.6.1.2 If an extraction is performed using acetone, chloroform, carbon disulfide, petroleum ether, etc., the same solvent shall be used for the blank. 5.6.2 The resulting data from the blanks shall not have peaks that interfere with the analysis. 5.6.3 The wash bottles on the autosampler shall be filled with acetone. 5.7 Maintenance 5.7.1 The septa for the injection port shall be changed at least monthly. 5.7.2 The injection port liner shall be changed at least yearly. Page 5 of 7

5.8 Sampling and Sample Selection N/A 5.9 Calculations N/A 5.10 Uncertainty of Measurement N/A 6.0 Limitations N/A 7.0 Safety 7.1 Carbon disulfide may be toxic. Consult Material Safety Data Sheets for information on safe use. 7.2 Burns may result from contact with hot items such as containers, syringes, liners, and septa nuts. 7.3 Syringe needles are sharp and can easily puncture skin. 7.4 Care shall be exercised when using the above-listed items or procedures. 8.0 References ATF National Laboratory Center Class. Laboratory Detection and Identification of Accelerants Found in Arson Debris. Saferstein, R. Forensic Science Handbook. Volume I, 2 nd edition. Upper Saddle River, NJ: Prentice Hall, 2002. Chapter 9: Arson and Explosive Investigation. pp. 479-524. ASTM Guidelines ASTM Standard E1387, 2001, Standard Test Method for Ignitable Liquid Residues in Extracts from Fire Debris Samples by Gas Chromatography. ASTM International, West Conshohocken, PA, 2001, www.astm.org. 9.0 Records Maintenance log Use log Quality Control Check Chromatographs Request for Instrumental Examination of Evidence 10.0 Attachments N/A Page 6 of 7

Revision History Effective Date Version Number Reason 09/17/2012 1 Original ISO Document 02/01/2013 2 Request for Instrumental Examination of Evidence was added as a record. 5.5 Instrument parameters for the Western Regional Laboratory were added. 09/30/2013 3 Quality Control Check log was removed from 9.0; Quality Control Check Chromatograms are stored electronically. Page 7 of 7