Installing a Cool On-Column Inlet

Similar documents
Installing a PTV Inlet

Installing a PTV Inlet

Installing a Purged Packed (PP) Inlet

Installing the CO 2 Cryogenic Oven Cooling Kit

Installing a Purged Packed Inlet

Installing an Auxiliary Pressure Control Module

Installing an Auxiliary Pressure Control Module

Agilent G1554A On-Column Capillary Inlet with EPC

Installation Guide. Purged Packed Inlet, Electronic Pneumatics Control on 6890 GC Accessory G1551A

Installation Guide. Thermal Conductivity Detector on a 6850 GC Accessory G2623A

Installation Guide. Flame Ionization Detector on a 6850 GC Accessory G2621

Installation Guide. Nitrogen-Phosphorus Detector on 6890 GC Accessories G1575A, G1576A, G1594A, G1599A

Installation Guide. Flame Ionization Detector on a 6890 GC. Accessories G1561A, G1562A, G1591A, G1598A

Installation Guide. Volatiles Interface for 6890 Gas Chromatograph Accessory G2319A

Installation Guide. Heater Interface Cable for Mass Sensitive Detector Accessory G1590A

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Installation Guide. Auxiliary Pressure Control Manifold on 6890 GC Accessory G1570A

Installing the 6850 Direct Valve Column Connector Kit

Agilent G2738A Upstream Capillary Interface

Agilent G3510A/G3511A Multimode Inlet

Nickel Catalyst Tube on a 6820 GC, Accessory G4318A

Installation Guide Thermal Conductivity Detector Accessory 19232E

Installing the Flame Ionization Detector EPC Flow Control Manifold

Valve Box and Valve Accessories G4315A, G4316A, G4317A

1310 Inlets. 1 of 36. Jun Illustrated Parts Breakdown Agilent 6890 Gas Chromatograph Service Manual

Nickel Catalyst Tube Accessory 19205A

Installation Guide NPD/ECD Chemical Filters Accessory 19199N

260 Solvent Vapor Exit Accessory

Agilent 6890 Gas Chromatograph

Agilent G2647A (Single) and G2648A (Dual) Flame Photometric Detector Update Kits

Agilent G2333B (Single) and G2334B (Dual) Flame Photometric Detectors

Agilent G3348B Flame Photometric Detector

1350 Valves. 1 of 14. Jun Illustrated Parts Breakdown Agilent 6890 Gas Chromatograph Service Manual

Agilent 6850 Series II Network GC System G2630A Performance Specifications

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

Agilent 490 Micro GC Internal and External Mount Genie Filter Brackets

SAP Simulator Installation

Hot Injection Adapter Kit - Installation Guide

Installing the G2743A/G2745A Valve Actuator Kits

INSTALLATION OF OPTIC 3 INJECTOR SYSTEM ONTO AN SHIMADZU GC-2010 GAS CHROMATOGRAPH.

6890 Installing ¼ Glass Columns

Installation Note for Solvent Selection Valve Upgrade Kit (G1360A)

In addition to this instruction sheet, the following are included in this kit (some additional parts are included for future maintenance):

Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application

SAP Simulator Installation

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

29 Installation Step 1. Unpacking the GC Cable diagrams Step 2. Placing the GC system on the benchtop Step 3. Turning the power on

Agilent 7820A Gas Chromatograph

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

GC Inlets. An Overview. Simon Jones GC Applications Engineer

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)

Technical Procedure for Gas Chromatography (GC-FID)

Technical Tips (April 2017) Product Information

6890 CHECKOUT PROCEDURE TCD (THERMAL CONDUCTIVITY) DETECTOR

For more information about Agilent Technologies services please visit our web site using the following URL

Electrically Held to Mechanically Latched Contactor

Deans Switch Microfluidics

Agilent 7697A Headspace Sampler

If there is additional labor or other costs above the 4 hrs labor you must call VMAC for approved coverage

Installation Instructions. PowerFlex 700 Drive - Frame 8 Components Replacement

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

A g i l e n t A G a s C h r o m a t o g r a p h Service Manual

HP Modular Cooling System 200/100 Options Installation Guide

Installation Guide. Agilent 7683 Automatic Liquid Sampler

Mustang 5.0L Cold Air Intake Installation Instructions P/N:

Lamp House with 2.5 kw lamp

SPEED CONTROL 4 AND 6 CYL. JEEP WRANGLER. Read entire instructions thoroughly before starting. INSTALLATION INSTRUCTIONS TOOLS REQUIRED:

ENC82 Environmental Enclosure Installation Instructions

Agilent G4367A Auxiliary Electronic Pressure Control (Aux EPC) Module

OPERATORS MANUAL/INSTRUCTIONS/PARTS & SERVICE FOR MANUAL & ELECTRIC

SPEED CONTROL 4 AND 6 CYL. JEEP WRANGLER. Read entire instructions thoroughly before starting. INSTALLATION INSTRUCTIONS TOOLS REQUIRED:

HP Adjustable Depth Fixed Rail Kit Installation

5000TOC Sensor Service Manual

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

Agilent 5975T. Mobile Lab Hardware Installation. Agilent Technologies

CrossLab. Supplies for Thermo Scientific GC Systems. work for you non-agilent GCs. Put Agilent innovation and quality to. Agilent

3.4L V6 SUPERCHARGER 7 TH INJECTOR KIT

Alcatel-Lucent 7310 DC Metro Cell AWS (B4) Kit MC-DBR1, MC-DBR2, MC-DBR3, MC-DBR4, MC-DBR5. Installation Manual

RENA AF371Feeder Operating Manual. Feeder. Operating Manual. Manual Part #: M AF371 Operations Rev

Genuine Restek Replacement Parts For Agilent 5890 GCs

18SP680Rev3 EPA04 MBE 4000 Car Hauler Low Pressure Fuel Lines

N-084M and N-069M-DV Series Heat Exchanger Replacement

Agilent Multimode Inlet for Gas Chromatography

H280-series Valve Overheat / Relocation Kit Instructions: Standard Oven. Instructions for Applying Kit #

Rear-upper cover (duplex products). See Rear-upper cover (duplex product) on page 111.

1501 Industrial Way N., Toms River, NJ Fax: PACKING LIST

AIR CLEANER GENERAL REMOVAL. 1CAUTION Do not run engine without filter element in place. Debris could be drawn into the engine causing damage.

Torque Module Installation and User Guide for model 250i Motorcycle Dynamometers.

2018 Mustang 5.0L Performance PAC1 Installation Instructions P/N: ( PERFPK1)

2014+ FORD FIESTA ST DIRECT-FIT OIL COOLER KIT PARTS LIST AND INSTALLATION GUIDE CAUTION PARTS INCLUDED NOTE TOOLS NEEDED INSTALL PROCEDURE

FD 120 Card Cutter MAINTENANCE MANUAL. MyBinding.com 5500 NE Moore Court Hillsboro, OR Toll Free: Local: /2011

2018 Mustang 5.0L Cold Air Intake Installation Instructions P/N: ( )

5. Refer to GTR Workshop Manual, section: and disconnect battery ground cable.

DETECTORS Photo Ionization Detector - PID. Overview

B-026INT October 6, 1999 FUEL FILTER BRACKET RETROFIT (JAPAN ONLY) GENERAL RETURN THIS TO: INITIAL HERE TECHNICIAN NO.

Ford 6.0L Powerstoke Regulated Fuel Return

w w w. h d o n l i n e s h o p. d e CRUISE CONTROL KIT GENERAL INSTALLATION -J04064 REV Kit Number Models Additional Parts Required

Agilent G2855A Deans Switching System

GC Injector Liners. Bruker Varian Agilent (HP) Perkin Elmer Finnigan. Phone: Page 4

Method Development for Capillary GC Systems. Slide 1

Transcription:

Agilent 6850 Series II Network GC System Accessory G3344B This kit contains: Description Quantity Machine screws, M4 x 0.7 12 mm 6 Cable ties,.062.625 diameter 6 Ship kit* 1 T-20 Torx screw, M4 x 8 mm 1 Cool on-column inlet assembly (with covers) 1 Nut and ferrule set, 1/8-inch 1 Disposable wrist strap 1 Cavity sleeve 1 Syringe kit* 1 Seal insulation 1 CO 2 CryoBlast weldment** 1 Post-drawn tube** 1 Tubing union, 1/8-inch 1 Nut, stainless steel, for 1/16-inch tubing 1 Ferrule set, stainless steel, for 1/16-inch tubing 1 Installation sheet (this document) 1 * Component items are listed on the next page. ** Used only with CryoBlast

The Ship kit contains: Description Quantity Insert, 250 microns 1 Inserts for 530-µm columns, no rings 1 Insert springs 2 Lowbld septa, 5 mm, 5/pk 1 PCOC 530 Septum nut 1 Needle support 250/320 1 Septa, 5-mm through hole, 25/pk 1 Needle, 0.25-mm on-column/5 µl syringe, 3/pk 1 The Syringe kit contains: Description Quantity Syringe plunger 1 Syringe, 5-µL on-column RN, barrel only 1 Needle, 0.32-mm on-column/5-µl syringe, 3/pk 1 The inlet is factory-assembled. Do not disassemble it during installation. 2

Parts identification Parts identification CO 2 CryoBlast weldment Inlet assembly Cavity sleeve Seal insulation Post-drawn tube Nut and ferrule sets Cable ties and machine screws Disposable wrist strap Required tools T-20 Torx screwdriver Flat-blade screwdriver Inlet wrench Forceps 3

Overview Overview Caution Before starting, review the safety information listed at the end of this document. 1. Disconnect the GC. 2. Remove the lid top cover. 3. Remove the existing inlet. 4. Install the new inlet. 5. Install the cryogenic oven cooling kit, if applicable. 6. Restore the GC to operating condition. 7. Zero the pressure sensor. 4

Disconnect the GC Disconnect the GC WARNING Hydrogen gas is flammable and potentially explosive. Before replacing the flow module, turn off the hydrogen gas at the source. 1. Turn off the GC and unplug the power cord. Allow time for all heated zones to cool. 2. Open the lid. If a column is installed, disconnect it at the detector end. Remove the nut warmer, insulation, column hanger and capillary adapter, if present. Then, remove the bottom cover plate and close the lid. 3. Turn all gases off at their sources. Disconnect the carrier and detector gas tubing from the back panel of the instrument. Disconnect gas supply tubing from these fittings 5

Remove the lid top cover Remove the lid top cover Remove vent tubing, if present Lid top cover, without valve box accessory Remove T-20 Torx screws (8 places) Remove detector cover, if desired 6

Remove the existing inlet Remove the existing inlet If the GC currently has an inlet installed, you must remove it. 1. Loosen the screws holding the connector cover plate next to the inlet flow module. Disconnect Remove Loosen 2. Slide the cover plate off and disconnect the cable from the flow module. 3. Disconnect the gas fitting on the back of the flow module. 4. Remove the three screws on top of the flow module. 5. To continue this procedure, see the following section corresponding to the type of inlet you are removing. Cool On-Column inlet 1. Use a T-20 Torx screwdriver to loosen the three captive screws that attach the inlet weldment to the top of the inlet carrier. 2. Slide the inlet up and out of the carrier. If necessary, you can also slide the insulation sleeve off of the bottom of the inlet. 7

Remove the existing inlet Purged Packed inlet 1. Trace the heater/sensor cable from the inlet to the wiring harness connector. Disconnect it. 2. Remove the three screws holding the inlet. 3. Lift the inlet and flow module out of the lid. Remove the insulation in the hole under the inlet. Programmable Temperature Vaporization inlet 1. Lift the PTV inlet up. 2. Remove the heater connector from the side panel. 3. Disconnect the power cable, thermocouple connector, and cryo connector from the flow module and board. Disconnect the module ribbon cable from the pneumatics board. 4. Slide the flow module out of the chassis, remove the chemical trap assembly from the mounting bracket, and remove the PTV assembly from the GC. 5. Slide the cryo assembly out of the open-top slot next to the PTV pneumatics module. Unwind and remove the tubing that is coiled up inside the cryo assembly. 6. Liquid N 2 Cryo Assembly Only. Slide the valve bracket into the opentop slot next to the PTV pneumatics module. Secure from the front with a short (M4 x 12) screw. 7. Remove the chemical trap and tubing. 8

Remove the existing inlet Split/Splitless inlet 1. Use the inlet wrench to release the large nut on top of the inlet. Large nut Insert weldment (includes nut) 2. If you do not have a gas or liquid sample valve Remove the insert weldment and the flow module. If you do have a gas or liquid sample valve The insert weldment is part of the valve assembly. Move it out of the way. Disconnect the gang block fitting (also part of the valve assembly), from the flow module. Remove the flow module. 3. Use forceps to remove the liner and O-ring. 4. Disconnect the split vent trap. Disconnect Top of liner 5. Trace the heater/sensor cable to the wiring harness connector. Disconnect it. 6. Remove the three screws holding the inlet body. 7. Lift the inlet assembly out of the lid. Remove the insulation in the hole under the inlet. 9

Install the new inlet Install the new inlet Caution Board components can be damaged by static electricity. Use a properly grounded static control wrist strap when installing the flow module. 1. Move the lid to the normal open position and slide the stop plate all the way up into position behind the roller. Tighten the stop plate screw. 2. Insert the CO 2 CryoBlast weldment into the inlet base as far the bend in the wire (until there is resistance). CO 2 CryoBlast weldment (top) Inlet base CO 2 CryoBlast weldment (bottom - pushed through) 10

Install the new inlet 3. Place the insulation on the inlet base so the CO 2 CryoBlast weldment fits through the slit in the insulation. Slit in insulation 4. Place the inlet and CO 2 CryoBlast weldment so it lines up with each screw hole in the oven top and the heater/sensor cable sits in the trough in the inlet chassis. Start the screws in the slots, but leave them loose. Open the lid to check that the items are positioned correctly, and then close the lid. Heater/ Sensor cable Insulation 11

Install the new inlet 5. Place the flow module as shown. Start the screw in the back. Install the "C" tube to the flow module input fitting and connect the ribbon cable. Start this screw 6. Route tubing around the flow module and side of the lid. 7. Tighten the screws to secure the inlet body in place. 8. Open the lid and install cavity sleeve to keep the insulation in place. Close the lid when finished. Cavity sleeve 12

Install the new inlet 9. Feed the cryo tubing from the inlet around the front of the flow module and through the slot in the lid. Slot in lid 10. Open the lid and raise it to the service position. 11. Pull the tubing the rest of the way through the slot in the lid and wrap it in a lazy S. 12. Connect the other half of the tube. 13. Loosen the electronics cover. 13

Install the new inlet 14. Route the tube as shown. To avoid crushing the tubing, do not let it obstruct the cam and follower. 15. Use the tie wraps to secure the tubing. Tie wraps 14

Install the new inlet 16. Feed the lower half of the CryoBlast tubing into the correct slot in the electronics cover and out of the GC. If CryoBlast is already installed, loosen the mounting bracket to provide enough room to push the tubing through. 17. Install the nut and ferrule set onto the tubing. 18. From the left side of the GC, locate the heater/sensor cable and its corresponding connector. Tuck the cable underneath the clip at the side of the GC and connect it to the nearest square connector. Heater/Sensor cable Tubing 19. Route the stainless steel tubing along the side of the chassis, around the fan bracket, and into the flow module. 15

Install cryogenic oven cooling Install cryogenic oven cooling If you will be using cryogenic oven cooling, install the CO 2 Cryogenic Oven Cooling kit before you finish installing this inlet. There are some specific instructions for the cool on-column inlet when installing the chassis for the cryo oven. See the directions for that kit for more information. 1. Pull any inlet cryo tube or wires through the hole on the side of the GC. 2. There are four mounting holes for the chassis on the left side of the GC (two of them are in the row of ventilation slots at the top). Place the chassis over the holes. Insert screws through the brass eyelets and secure them loosely. The small black grommet behind the coolant outlet union should be over the smaller hole in the side panel. Mounting holes (one is behind the valve) Route any inlet cryo tube or wires through here Coolant outlet union 16

Restore the GC to operating condition Restore the GC to operating condition 1. Install the lid top cover. 2. Plug in the GC and turn it on. 3. Press Inlet. If the you have configured the GC correctly, the following screen appears. Note: A pressure value is displayed when the carrier gas is off or not connected. This is not an error. After the carrier gas is connected and the inlet is zeroed, the pressure value equals the setpoint value. 4. Restore carrier and other gases to the instrument. 5. Apply your normal operating pressures. Leak-check the flow module, back panel, and column fittings. 17

Zero the pressure sensor Zero the pressure sensor You must zero the pressure sensor of the inlet s flow module after installing it on the GC. Doing so ensures an accurate inlet pressure display. Do not connect the carrier gas to the flow module until you have zeroed the pressure sensor. 1. Plug in the GC and turn it on. 2. Wait 15 minutes. Doing so allows the GC to reach thermal equilibrium.display this screen. Status / Service / Calibration / Inlet Cal 3. Select Enable Auto Flow Zero to automatically recalibrate the zero at the end of every run. 4. To zero the flow sensor manually, press ZERO FLOW. The inlet flow will be momentarily interrupted. The process takes about 2 seconds. To zero the pressure sensor, turn the carrier gas off at the source. Separate one of the connections in the supply tubing to be certain that there is no pressure trapped in the plumbing. Press ZERO PRESS. 5. Restore normal carrier gas flow. 6. Turn off the GC and unplug the power cord. 7. Plumb the carrier gas to the flow module. 8. Plug in the GC and turn it on. 9. Configure the GC s column and carrier gas. 18

Agilent Technologies, Inc. 2004 All Rights Reserved. Reproduction, adaptation, or translation without permission is prohibited, except as allowed under the copyright laws. Part number G3344-90007 First Edition, June 2004 Printed in USA Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 Acknowledgements Torx is a U.S. registered trademark of Textron, Inc. Safety Symbols Warnings in the manual or on the instrument must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions violates safety standards of design and the intended use of the instrument. Agilent Technologies assumes no liability for the customer s failure to comply with these requirements. In the manual A warning calls attention to a condition or possible situation that could cause injury to the user. A caution calls attention to a condition or possible situation that could damage or destroy the product or the user s work. On the instrument See accompanying instructions for more information. Indicates a hot surface. Indicates hazardous voltages. Indicates earth (ground) terminal. Indicates explosion hazard. Indicates radioactivity hazard. Indicates electrostatic discharge hazard. Indicates pinch hazard. G3344-90007