ECscan40 Pocket Conductivity/TDS/Salinity Tester Instruction Manual

Similar documents
ECscan30 Pocket Conductivity/TDS Tester Instruction Manual

ECscan20 Pocket Conductivity Tester Instruction Manual

ECscan10H Pocket Conductivity Tester. Instruction Manual BANTE INSTRUMENTS LIMITED

TDSscan10M Pocket TDS Tester Instruction Manual

PHscan30S Pocket ph Meter Instruction Manual

PHscan10S Pocket ph Tester Instruction Manual

ORPscan20 Pocket ORP Tester Instruction Manual

Bante220 Portable ph Meter Instruction Manual

Operation Manual MODEL EC331. Microcomputer Based Conductivity /Resistivity & Temperature Pocket Meter EC331

EC1 Pocket Conductivity Tester Instruction Manual

EC20 Pocket Conductivity Tester Instruction Manual

PC60 Premium Multi-Parameter Tester (ph/ec/tds/salinity/temp.) Instruction Manual. APERA INSTRUMENTS, LLC V6.4

Salt 1 Pocket Salinity Tester Instruction Manual

COM-100 EC / TDS / TEMPERATURE METER

PH60 Series Premium ph Testers

Instruction Manual. PHscan30 Pocket ph Tester

PH20 Value ph Tester Instruction Manual

USER S GUIDE WARRANTY. This manual and all information contained herein is copyright HM Digital, Inc. ONE YEAR LIMITED WARRANTY

SERVICE MANUAL. STARTER Portable Meter ST300 ST300C ST300D ST400D. Ohaus Corporation, 7 Campus Drive, Suite 310, Parsippany, NJ (973)

Large Display ph Pens , , Instruction Manual

TRACER. POCKETESTER ä ORP CODE 1742

HI Portable Waterproof Skin ph Meter INSTRUCTION MANUAL

Sunleaves TDS Expert Plus

SoilStik ph Meter PRODUCT MANUAL. Item # 2105

WalkLAB Digital Dissolved Oxygen meter HD9030

Care and use guide Guía de uso y cuidado Guide d utilisation et d entretien.

Tracer ORP PockeTester TM Code 1742

Instruction Manual Portamess 911 Cond

STARTER 2000 Bench ph Meter Instruction Manual

8,1,8,1,2,7,2,7,6,3,6,3,4,5,4,5

Model P772 Conductivity Meter. User Manual

Care and use guide. Contents. page

HI Ethylene Glycol Refractometer INSTRUCTION MANUAL

User Manual Pocket Pro, Pro + ph

HanTron Model HT-3P. Manual. Installation Maintenance Repair Manual

OPERATION MANUAL PEN TYPE. Model: 8690 DESIGN PATENTED: US D S

SERVICE MANUAL STARTER 2100 Bench ph Meter STARTER 3100 Bench ph Meter STARTER 3100C Bench ph Meter

ACCURACY RELIABILITY SIMPLICITY

Visit us at

MODEL IQ150 ph Meter Getting Started TWO-POINT CALIBRATION--QUICK REFERENCE GUIDE

THREE RANGE DS METERS Operation Instructions

4&5 RANGE CONDUCTIVITY METERS Operation Instructions

KILOTECH. KWD 500 Water tight. Portion/Weighing Scale

DMG200. Digital Micron Gauge. 99 Washington Street Melrose, MA Phone Toll Free

Digital Formaldehyde Meter

Marin Control Digital

Instruction Manual. Single-Input Thermistor Thermometer (Model )

Moisture meter. Operators manual

l The Battery Tester is designed for measuring the l AC four-terminal method to measure the internal

Rally computer 3 Rally computer 3.GPS *

Temp-14 Single Input Thermistor Thermometer

testo 206 ph/ Temperature Measuring Instrument Bedienungsanleitung Instruction Manual

operating manual DL7 101 DIGITAL THERMOMETER

INSTRUCTION MANUAL MARTINDALE CM52 CLAMP METER. ELECTRIC Trusted by professionals. Other products from Martindale:

INSTRUCTION MANUAL. Handy Polaris

PT100 Temperature Sensor User Manual

SC-Series. User Manual. American Weigh Scales. SC-501, SC-501A (500g x 0.01g) SC-2KG, SC-2KGA (2000g x 0.1g)

Instruction Manual PORTABLE ph METER D-71

DMG100 Digital Micron Gauge Limited Warranty DMG100 DMG100 Digital Micron Gauge USA CANADA COPYRIGHT 2011 UEi

Exterior Digital Load Scale 201-EDG-01(B) Installation and Operation Manual Please read carefully before installation

NPS Series. Pallet Jack Scale Operator s Guide (V1611)

User s Manual. Model PH71 Personal ph Meter IM 12B03D01-01E. IM 12B03D01-01E 3rd Edition

INSTRUCTION MANUAL. Mi490 Peroxide value. Milwaukee Lab Photometer. Test Equipment Depot Washington Street Melrose, MA 02176

Care and use guide. Contents. page

INSTALLATION INSTRUCTIONS

Valor 1000W Series Instruction Manual. Serie Valor 1000W Manual de Instrucciones. Série Valor 1000W Guide de I utilisateur

MoistureMatch A next generation grain tester

American Weigh Scales SG-300. User Manual. SG-300 (300g x 0.1g) Copyright 2012 American Weigh Scales, Inc. All rights reserved. Rev. 2.

Visit us at

American Weigh Scales LS-110. User Manual. LS-110 (110lb x 0.2lb) Copyright 2010 American Weigh Scales, Inc. All rights reserved. Rev. 2.

RAPID OIL TEST GAUGE

Care and use guide. Contents. page

SSP-Series Portable Balances Operation Manual

USER MANUAL. White LED Light Meter. Model LT40. Additional User Manual Translations available at

Troubleshooting. CryoSmart 1 - CryoSmart 20

SMT-AZC INSTALLATION / OPERATION MANUAL

HI Free Chlorine Meter

Inlet Controller TC5-ITA USER'S MANUAL. M rev. 02 K rev. 00

Patient Lift Scale Indicator Owner s Manual

TABLE OF CONTENTS YOUR MISTBOX S SERIAL NUMBER

Short manual testo 610

IAQ-CALC TM Indoor Air Quality Meter

User's Guide. CO 2 Meter. Model CO250

OPERATING INSTRUCTIONS

Grain Moisture Meter. Operators manual

INTRODUCTION. Backlight for operating in dark area Low battery indicator Auto power off function. CO levels and its effect

Centrifuge Operator / Service Manual

TMD90A Dual Input Digital Thermometer

Instruction Manual. ph Meter D-51

INSTRUCTION MANUAL DT173 DT175 DIGITAL THERMOMETERS ALWAYS READ THESE INSTRUCTIONS BEFORE PROCEEDING

Manual KREBS Viscometer PCE-RVI 5

ortable ph, Conductivity & DO2 Meters

MDX-300 Series. For 12-volt automotive starting batteries and starting/charging systems INSTRUCTION MANUAL

Indoor Air Quality Meter

Conductivity Measurement CLM 280

User s Manual. Model PH72 Personal ph/orp Meter IM 12B03D02-01E. IM 12B03D02-01E 3rd Edition

IAQ-CALC INDOOR AIR QUALITY METER MODEL 7515

WPT Installation Manual. Wireless Pneumatic Thermostat- WPT-800 Series. 1. Overview. 1.1 Parts Included. 1.2 Prerequisites for Installation

Installation and Operation Manual

IAQ-CALC TM Indoor Air Quality Meter

Transcription:

ECscan40 Pocket Conductivity/TDS/Salinity Tester Instruction Manual BANTE INSTRUMENTS LIMITED Technology to create the future...

ECSCAN40 POCKET CONDUCTIVITY/TDS/SALINITY TESTER 1 Introduction Thank you for selecting the ECscan40 pocket conductivity tester. This manual provides a step-by-step guide to help you operate the meter, please carefully read the following instructions before use. Unpacking the Meter The following list describes the standard accessories of the tester. After the unpacking, please check all accessories are complete. If any are damaged or missing, please contact nearest distributor. ACCESSORIES: Conductivity Calibration Solutions (84µS/cm, 1413µS/cm, 12.88mS/cm) AAA Batteries Carrying Case Keypad ECscan40 tester has a succinct membrane keypad, names and symbols describe the each function key controls. INDEX: KEY MEAS/HOLD CAL ENTER FUNCTION Power the tester ON/OFF. Freezes the measured values on the display, press the key again to resume measuring. In the calibration mode, exits calibration and returns to measurement mode. Press the key to enter the calibration mode. Press and hold the key to enter setup menu. In the calibration mode, press the key to set calibration values. In the setup mode, press the key to select default option. Confirms the calibration or displayed option. Toggles between conductivity, TDS and salinity measurement modes.

ECSCAN40 POCKET CONDUCTIVITY/TDS/SALINITY TESTER 2 Display The tester is equipped with a clear LCD display that used to show measured value, mode indicators and unit indicators. mus ppm ppt INDEX: INDICATOR MEAS CAL SETUP HOLD ATC DESCRIPTION Indicates tester is in the measurement mode Indicates tester is in the calibration mode Indicates tester is in SETUP mode Indicates the measured value has been frozen Indicates temperature compensation is enabled Prior to Use Remove the electrode protective cap from tester. Soak the electrode for a few minutes in tap water to remove dirt and oil stains on the electrode. Power On/Off Press MEAS key to turn on the tester, the display shows measured values. Press and hold the MEAS key for 5 seconds, the tester will turn off. If you do not press any key for 8 minutes, the tester will automatically turn off to conserve energy. To disable the auto-off function, please read the SETUP MENU section.

ECSCAN40 POCKET CONDUCTIVITY/TDS/SALINITY TESTER 3 Setup Menu ECscan40 tester contains an integrated setup menu that allows you to customize each option to meet measurement requirements. PARAMETER DESCRIPTION OPTIONS DESCRIPTION DEFAULT Calibration Point 1 point 2 points 3 points TDS Factor Adjustable range: 0.1 to 1.0 0.5 Temperature Unit Temperature Calibration Auto-Hold Auto-Off Reset Degrees Celsius Degrees Fahrenheit s the temperature calibration mode Enable Disable Enable Disable Restores the parameters back to default settings Do not use this function SETTING THE DEFAULT PARAMETERS: 1. Press and hold the CAL key for 3 seconds to enter setup menu, the tester shows current selected number of calibration points. 2. Press CAL key to select 1 or 2 or 3 point calibration. Press ENTER key to confirm, the tester goes to TDS factor setting mode, the display shows 0.5/TDS. 3. Press CAL key to set the TDS conversion factor. Press ENTER key to confirm, the tester goes to temperature unit selection mode, the display shows o C/UNIT.

ECSCAN40 POCKET CONDUCTIVITY/TDS/SALINITY TESTER 4 4. Press CAL key to select the temperature unit ( o C or o F). Press ENTER key to confirm, the tester goes to temperature calibration mode, the display shows o C/CAL. 5. If you do not want to perform temperature calibration, press ENTER key, the tester goes to next option. If needed, press CAL key to enter the temperature calibration mode, the display shows current temperature reading. Cal 6. Press CAL key to set temperature value. Press ENTER key to confirm, the tester shows YES/HOLD indicating that the auto-hold function is enabled. If the auto-hold function is enabled, the tester will automatically sense a stable endpoint reading and freeze it. If you disable this function, the tester allows user to freeze the reading by pressing the HOLD key. 7. Press CAL key to enable or disable the auto-hold function. Press ENTER key to confirm, the tester shows YES/OFF indicating that the auto-off function is enabled. When the auto-off function is enabled, if you do not press any keys for 8 minutes, the tester will automatically turn off to conserve energy.

ECSCAN40 POCKET CONDUCTIVITY/TDS/SALINITY TESTER 5 8. Press CAL key to enable or disable the auto-off function. Press ENTER key to confirm, the tester shows NO/RST indicating that do not reset the tester. Reset function will restore the tester back to factory default settings, all calibration values and selected parameters will be reset. 9. Press CAL key to enable or disable the reset function. Press ENTER key to confirm, the tester returns to measurement mode. Setting is completed. EXIT THE SETUP MENU: Press MEAS key in setup mode, the tester will return to measurement mode immediately. Conductivity Calibration ECscan40 tester allows up to 3 points calibration in the conductivity mode. To ensure higher accuracy, we recommend that you perform 3 points calibration or select a standard value close to the sample value you are measuring. The tester will automatically detect these conductivity standard solutions and prompt the user to calibrate the meter. When the calibration is done, all new calibration values will automatically override existing data. The following table shows acceptable conductivity range of calibration solution for each measuring range. MEASURING RANGE CALIBRATION SOLUTION RANGE DEFAULT 0~20µS/cm 7~17µS/cm 10µS/cm 20~200µS/cm 70~170µS/cm 84µS/cm 200~2000µS/cm 700~1700µS/cm 1413µS/cm 2~20mS/cm 7~17mS/cm 12.88mS/cm Ensure that you use fresh conductivity standard solution during the calibration. Do not reuse calibration solutions as it may be contaminated and affect the calibration and accuracy of measurement. SINGLE POINT CALIBRATION: 1.1 Rinse the conductivity electrode with distilled water, then rinse with a small amount of calibration solution. 1.2 Press CAL key, the tester enters the calibration mode. Cal

ECSCAN40 POCKET CONDUCTIVITY/TDS/SALINITY TESTER 6 1.3 Dip the electrode into the calibration solution, the tester immediately shows current calibration standard (e.g. 1413µS/cm). 1.4 Press ENTER key, the default calibration value begins flashing. us 1.5 Press CAL key to set each digit, press ENTER key to confirm. When the setting is done, ensure the displayed value matches your calibration standard. us 1.6 Press ENTER key to start the calibration. Wait for the reading to stabilize, the display shows END. The tester returns to measurement mode automatically. Single point calibration is completed. 2 POINTS CALIBRATION: 2.1 Make sure that you have selected 2 points calibration in the setup menu. 2.2 Repeat steps 1.2 to 1.6 above, when the first calibration point is done, the display will show CAL2. The tester prompts you to continue with second point calibration. 2.3 Dip the electrode into the calibration solution, the tester automatically shows current calibration standard (e.g. 84µS/cm). us

ECSCAN40 POCKET CONDUCTIVITY/TDS/SALINITY TESTER 7 2.4 Press ENTER key, the default calibration value begins flashing. 2.5 Press CAL key to set each digit, press ENTER key to confirm. When the setting is done, ensure the displayed value matches your calibration standard. 2.6 Press ENTER key to start the calibration. Wait for the reading to stabilize, the display shows END. The tester returns to measurement mode. Second point calibration is completed. 3 POINTS CALIBRATION: 3.1 If you have selected 3 points calibration in the setup menu, when second calibration point is done, the display will immediately show CAL3. The tester prompts you to continue with third point calibration. 3.2 Repeat the above steps until the display shows END, the tester returns to measurement mode. Calibration is completed. Temperature Calibration During the measurement, if the temperature reading displayed differs from that of an accurate thermometer, you need to calibrate the tester. 1. Press and hold the CAL key for 3 seconds to enter setup menu. 2. Press ENTER key until tester shows the o C/CAL or o F/CAL. 3. Press CAL key to enter the temperature calibration mode. 4. Press CAL key again to set temperature value (Resolution: 0.5 o C). 5. Press ENTER key to confirm, the display shows next option. 6. Press MEAS key, the tester returns to measurement mode. Calibration is completed.

ECSCAN40 POCKET CONDUCTIVITY/TDS/SALINITY TESTER 8 During the calibration process, if you want to exit the calibration, press MEAS key, the tester will return to measurement mode. Performing the conductivity calibration will simultaneously calibrate the corresponding TDS and salinity value. Switching the Measurement Mode 1. In the conductivity mode, press ENTER key until the display shows TDS. The tester is now enters to TDS measurement mode. ppm 2. Press ENTER key, the tester enters the salinity measurement mode. ppt 3. Press ENTER key again, the tester returns to conductivity measurement mode. us Measurement Rinse the conductivity electrode with distilled water. Dip the electrode into the sample solution. Stir the tester gently. Wait for the reading to stabilize, record the measured value on the display. Hold Function ECscan40 tester has two data hold modes. When the Auto-Hold function is enabled, the tester will automatically sense a stable endpoint measured value and freeze it, HOLD indicator will appear on the display. If the Auto-Hold function is disabled, press HOLD key, the tester will immediately freeze currently displayed value. Press the key again to release the measured reading, you can continue to take measurement. Meas Hold us

ECSCAN40 POCKET CONDUCTIVITY/TDS/SALINITY TESTER 9 Electrode Care and Maintenance After each use, the electrode module should be rinsed thoroughly in deionised water. If there is a build-up of solids inside the measurement area of the cell, these should be removed very carefully with a cotton bud soaked in solvent, taking care not to touch the metal parts of the inner cell. Electrode Replacement When the tester fails to calibrate or gives fluctuating readings for calibration standards, you need to replace the electrode module. Twist the electrode collar counter clockwise, pull the old electrode module away from the tester. Align the slot on the new electrode module, gently push the module into the tester. Twist the electrode collar clockwise until it is tight. Installation is completed. Replacing the Batteries If the battery indicator is disappeared during the use, you need to replace the batteries. Twist the electrode collar counter clockwise, pull the electrode module out from the tester. Insert two AAA batteries into the battery compartment, note polarity. + Align the slot on the electrode module, push the module into the tester. Twist the electrode collar clockwise until it is tight. Troubleshooting LCD DISPLAY CAUSE CORRECTIVE ACTION Electrode dried out Soak the conductivity electrode in tap water for 10 minutes Measured value is out of range Check the electrode whether clogged, dirty or broken Incorrect calibration solutions Use fresh conductivity standard solutions for calibration Setting value does not match calibration solution Reset the calibration value Electrode is broken Replace the conductivity electrode module

ECSCAN40 POCKET CONDUCTIVITY/TDS/SALINITY TESTER 10 Specifications Conductivity TDS Salinity Temperature Others MODEL ECSCAN40 Range 0~20.00µS/cm, 200.0µS/cm, 2000µS/cm, 20.00mS/cm Accuracy ±1% F.S Resolution 0.01, 0.1, 1 Calibration Points 1 to 3 points Calibration Solutions 10µS/cm, 84µS/cm, 1413µS/cm, 12.88mS/cm Range 0~10ppt (Max. 20ppt, depending on factor setting) Accuracy ±1% F.S Resolution 0.01, 0.1, 1 TDS Factor 0.1~1.0 (Default 0.5) Range 0~10ppt Accuracy ±1% F.S Resolution 0.01ppt Range 0~60 C, 32~140 F Accuracy ±1 C Resolution 0.1 C, 0.1 F Calibration Points 1 point Calibration Ranges Measured value ±10 C Temperature Compensation 0~60 C, 32~140 F Temperature Coefficient 2%/ C Cell Constant K=1 Normalization Temperature 25 C Sensor Type Order Code: E-ECSCAN-C1-10K Power Requirements 2 1.5V "AAA" Batteries Dimensions 185(L) 40(Dia.)mm Weight 100g

ECSCAN40 POCKET CONDUCTIVITY/TDS/SALINITY TESTER 11 Addendum: How to prepare the conductivity standard solutions Place AR potassium chloride reagent in a 50mL beaker and dry in an oven for 3 to 5 hours at 105 C, then cool to room temperature in desiccator. Calibration Solutions 84µS/cm 1413µS/cm 12.88mS/cm 111.8mS/cm Accurately weigh out 42.35mg of KCL, and dissolve in 1000mL de-ionised water Accurately weigh out 745.9mg of KCL, and dissolve in 1000mL de-ionised water Accurately weigh out 7.45g of KCL, and dissolve in 1000mL de-ionised water Accurately weigh out 74.5g of KCL, and dissolve in 1000mL de-ionised water Stir the solution until the solution has thoroughly mixed. Preparation is completed.

Hazardous Substance Statement Bante Instruments is committed to the reduction and eventual elimination of all hazardous substances in both the manufacturing process and finished products we supply. We have an active manufacturing and procurement program to minimize and eliminate the use of harmful heavy metals such as cadmium, lead, mercury and the like. New technologies and design parameters are also promoting these efforts and we expect to have little or no such materials in our product in the coming years. We welcome our customer suggestions on how to speed up these efforts. Warranty The warranty period for tester is one year from the date of shipment. Above warranty does not cover sensor and calibration solutions. For more information, please contact nearest authorized distributor. BANTE INSTRUMENTS LIMITED E-mail: banteinstruments@yahoo.com Phone: +8621-6404-1598 Website: www.bante-china.com The information in this document is subject to change without notice. Copyright Bante Instruments Limited, 2015. All rights reserved.