OPERATOR'S MANUAL AND MAINTENANCE INFORMATION MODEL T-2000 TEXTURE TEST SYSTEM

Similar documents
OPERATOR'S MANUAL AND MAINTENANCE INFORMATION MODEL TM-2 & TM-DX TEXTURE TEST SYSTEM

Tridak Model 3200 User Guide Piston Inserter

958A Embedded Linear Displacement Transducer. THEINSIDER the most trusted information comes from INSIDE

VOLTAGE CONNECTOR CORCOM S VOLTAGE SELECTING AND FUSED CONNECTOR

MPT-200A SPECIFICATIONS AND OPERATING INSTRUCTIONS

OFI MODEL 130 HTHP BENCHTOP CONSISTOMETER INSTRUCTIONAL MANUAL

MOREHOUSE INSTRUMENT COMPANY, INC. 60,000 LBS CAPACITY AIRCRAFT PART NUMBER SCALE S FORCE CALIBRATION PRESS PART NO.

ACROMARK HOT STAMP PRESS

Track Drive Circuit - General System. Simplified Travel Circuit Diagrams: Neutral Controls Forward Travel Reverse Travel

RATE CONTROLLED TORQUE WRENCH TESTER

Kit. Hydraulic Kit Installation Guide REV A. Item Component Part Number Qty

Larson Systems Inc. Digital Hand Tester User Manual

ADHESION/RELEASE TESTER

MPT-250B SPECIFICATIONS AND OPERATING INSTRUCTIONS

Setup and Programming Manual

PROCESS ELECTRONICS CORPORATION

three different ways, so it is important to be aware of how flow is to be specified

KILOTECH. KWD 500 Water tight. Portion/Weighing Scale

Larson Systems Inc. V200, V750 and V1500 Digital Valve Spring Tester User Manual. Current for Software Version 6.9

MicroGuard 586 Retrofit Rated Capacity Indicator System. Calibration and Testing for:

OPERATING INSTRUCTIONS & SERVICE MANUAL BLUE MAX II HYDROSTATIC TEST PUMP

The RCS-6V kit. Page of Contents. 1. This Book 1.1. Warning & safety What can I do with the RCS-kit? Tips 3

SC-Series. User Manual. American Weigh Scales. SC-501, SC-501A (500g x 0.01g) SC-2KG, SC-2KGA (2000g x 0.1g)

12 Series Linear Electric Actuator Installation, Operation & Maintenance Manual

P3 Electronic Brake Control

Hydraulic Hand Pallet Trucks

GENERAL INFORMATION AVOID PLACING THE SCALE IN LOCATIONS THAT MAY AFFECT ACCURACY:

ESCONDIDO FIRE DEPT TRAINING MANUAL Section DRIVER OPERATOR Page 1 of 13 Pumps and Accessory Equipment Revised

K10 Intrinsically Safe Electro-Pneumatic Positioner Operating Manual

958A Compact Housing Linear Displacement Transducer

MODEL AP/W HIGH VOLTAGE HOLIDAY DETECTOR (NEW PRODUCT RELEASE 2015)

SECTION G2: CABLE PROCESSOR MODULE MAINTENANCE

ITA / ITB - AS / AP / AP SERIES Electronic Table Top Weighing Balance

Operation Guide. Hydraulic Leveling Systems #2000, #2010, #3000, and # Table of Content. Introduction

JBI Docupunch P33 Automatic Punch

500Z Owner's Manual & Troubleshooting

Section 7.1. Wheel Drive Circuit - General System

MAGPOWR Spyder-Plus-S1 Tension Control

SECTION 25 FRONT AXLE. Chapter 2 Terraglide (Front Axle Suspension)

uline.com

TM-1620 OPERATING INSTRUCTIONS AND OWNERS MANUAL

Open Center Compact Valve Custom Installation Guide Rev A

IN-LINE HYDRAULIC FLOW TESTER

Electronic Brake Control

Bruker BioSpin NMR CASE. Installation Manual. Version. NMR Spectroscopy. think forward

Control Valves Positioner

Control Valves Positioner

HD 7700 Setup & Operator Manual

DIGITAL TORQUE GAUGE MODEL BTGE-G

CENTAC Inlet and Bypass Valve Positioners

USER INSTRUCTIONS FOR OPERATION & MAINTENANCE 4068 Pneuma c / Hydraulic Installa on Tool

! Warning, refer to accompanying documents.

Cummins 6.7 Litre 2500/3500 Diesel EGR System. Always wear gloves and safety glasses when performing this service

SHORT-STOP. Electronic Motor Brake Type G. Instructions and Setup Manual

PREPOINT CHAMFER GRINDER MODEL Section 1. Set up and operation

SSP-Series Portable Balances Operation Manual

Operator s Manual. Fairbanks FH Series by Fairbanks Scales, Inc. All rights reserved. . Revision 1 06/2017

Carburetor Instructions

ALTERNATOR - BOSCH 35/75-AMP & 40/90-AMP

B-RAD Select USER MANUAL TABLE OF CONTENTS

2007 Mercury Marine SmartCraft Gauges

BMW/Isuzu 3.0 Liter Diesel EGR Cleaning Instructions

PVP Field Calibration and Accuracy of Torque Wrenches. Proceedings of ASME PVP ASME Pressure Vessel and Piping Conference PVP2011-

Cyclone 600 Upcut Cut Off Saw

Welker Sampler. Installation, Operation, & Maintenance Manual. Model GSS-4HP

Kit. Hydraulic Kit Installation Guide REV B Table 1 Component List. Item Component Part Number Qty

CI 3000 Coil Inserter

Installation & Calibration

Advanced Torque Products

Kit. Hydraulic Kit Installation A. Item Component Part Number Qty

POWERS CONTROL RC 195 Multiple Input Receiver- Controller

Installation, Operation and Maintenance Manual

All images and other materials are protected by United States copyright law and may not be reproduced, distributed, transmitted, displayed, published

G E N E S I S S T O R M SPECIALIZED TOP OFF AND REPLENISHMENT MODULE

FOUR-WHEEL ANTI-LOCK BRAKE SYSTEM (4ABS)

All of the control valves that we install the Mk. 16IQ positioner on are factory calibrated for proper operation prior to shipment.

BRAKE SYSTEM, HYDRAULICALLY ACTUATED - 631G TRACTOR Cat Tractors with standard shoe/drum brakes

American Weigh Scales LS-110. User Manual. LS-110 (110lb x 0.2lb) Copyright 2010 American Weigh Scales, Inc. All rights reserved. Rev. 2.

Tri-roll External Thread Gage

Operation Guide. Operation Guide. Winnebago Hydraulic Leveling Systems by Kwikee. Introduction. Table of Content WARNINGS

ATS-9600/9700 TABBER/LABELER

BMS-LiFePower. 123SmartBMS. Instruction manual

American Weigh Scales SG-300. User Manual. SG-300 (300g x 0.1g) Copyright 2012 American Weigh Scales, Inc. All rights reserved. Rev. 2.

MODEL #303 PLUNGING MACHINE

TECHNICAL NOTE #4 Revised May 24, BOGART ENGINEERING Two Bar Road, Boulder Creek, CA (831)

TERMINATOR User Manual

RD712 & RD712XL Remote Displays. Model 615 / 615XL Indicator User s Manual

SECTION B: SYSTEM OPERATING COMPONENTS

Advantage-D. Operating Instructions and Maintenance Manual. Central Vacuum Systems (Expandable/Modular Models) (Ver.

AU DU PONT DE LUTTRE BRUSSELS BELGIUM PHONE: FAX: OPERATING MANUAL. Electric Fully Automatic Floor Saw FS 1218 EX

Adjusting brake shoes for AutoPark parking brake

Techcon Systems TS6500 CIM Automatic Techkit Mixer

Torque Test Stands Series TS MODELS TST / TSTH & TSTM / TSTMH. User s Guide

LIPPERTCOMPONENTS, INC.

Table of Contents Introduction 3 Chapter 2: Operation 4 Chapter 3: Components 5 Chapter 4: Specifications 8

SpiritPFC Torque/Horsepower Comparison Dynamometer Test Date: 5/7/2006

PORT-A-WEIGH CRANE SCALES

INSTRUCTION MANUAL FOR. VOLTAGE REGULATOR Model: APR Part Number:

Model FMBC3000 Series Operating Instructions. Drum Pump Flow Meter

6.0 SPECIFICATIONS CONTENTS. Calibration. According to factory procedureeeeeeeeeeeeeee Accuracy*

Transcription:

OPERATOR'S MANUAL AND MAINTENANCE INFORMATION MODEL T-2000 TEXTURE TEST SYSTEM This Publication contains information proprietary To Food Technology Corporation The contents of this publication may not be disclosed to any other organization without written permission from Food Technology Corporation FOOD TECHNOLOGY CORPORATION 45921 Maries Road Suite 120 STERLING, VIRGINIA 20166 USA For additional information, accessories repair parts or service please call Food Technology Corporation (703) 444-1870 fax 703-444-9860 July 2004

SECTION ONE INTRODUCTION This manual contains the information necessary to install, operate and maintain the Food Technology T-2000 Texture Test System. These units are simple, trouble-free and rugged in operation. However, since they are highly accurate and precise measuring instruments they should be maintained and operated as outlined in this manual. GENERAL DESCRIPTION This FTC Texture Test System consists of four basic parts: 1. The Hydraulic System - consisting of the power unit, which supplies the hydraulic pressure and the ram or piston which transmits this pressure into mechanical force. 2. The Test Cell - which holds the sample and applies the "deformation" action to the food sample. 3. The Force Transducer - which mounts between the ram and the test cell and outputs an electrical signal proportional to the force applied. 4.The Texturegage - which converts the electrical signal from the transducer to a digital display in either Tenderometer Units, newtons, pounds of force, joules or inchpounds depending on the mode of operation.

SECTION TWO INSTALLATION 1. Examine the shipping containers for any sign of damage. Notify the carrier if damage to the containers or it s contents is evident. Otherwise proceed with unpacking and setup. 2. Remove the screws around the base of the shipping crate and lift the box off the base of the crate. 3. Remove the bolts that secure the Texturepress to the shipping skid from the bottom of the pallet and install the four leveling feet. Remove any packing material from around the test cell area. 4. Remove the back panel of the press by unscrewing the Phillips head screws around the perimeter of the rear of the unit. 5. If there is any sign of oil leakage, check the oil level. The oil level should be approximately two to three inches from the top of the reservoir. 6. With the back panel removed remove the plastic plug from the vent hole at the top of the oil reservoir and replace it with the metal breather plug. 7. Replace the back cover panel and place the Texturepress on the work surface and level the unit as accurately as possible by rotating the leveling feet. 8. Check that the Main Power switch is in the OFF position and plug the power cord into the AC power source required. Be certain that the power requirements are compatible with the local power supply.

SECTION THREE OPERATING INSTRUCTIONS CONTROLS AND INDICATORS MAIN POWER Applies AC power to the system and texturegage power supplies. This switch may be left on continuously to eliminate the electronic warm-up between testing sessions. If the system will not be used for more than 5 hours, it is advisable to turn off the system. PUMP POWER Applies AC power to the motor thereby activating the hydraulic system for testing. Between tests, that are more than a few minutes apart, it is recommended to switch off the pump power to conserve energy, reduce heat build up and undue wear on the system. FAST RETURN When this switch is in the "ON" position the speed control is bypassed on the up stroke or return stroke. When in the "OFF" position, the up stroke moves at the same speed as the down stroke. PRESS MOTION This controls the directional valve causing the ram to move UP or DOWN. This should always be in the stop position when the unit is idle or during normal testing. START Depressing the START button initiates the auto cycle of the press. This will start the ram down at the controlled speed, clear out the previous peak reading, and automatically retract the ram at the completion of the down stroke. If the Fast Return switch is on the ram will return quickly. The peak value and the work value will be held until the zero switch or the start button is pressed again. UP / DOWN This switch will increase or decrease the displayed value to adjust the calibration number. When calibrating the texturegage, hold the CAL / ZERO switch in the CAL position and compare the calibration number on the force transducer sticker to the displayed number. Use the UP/DOWN switch to increment or decrement to display

so that the numbers match closely. PEAK / INTEGRAL When in the "PEAK" position the switch allows the Texturegage to hold the maximum reading encountered, when in the "INTEGRAL" position it displays the work integral during the down stroke. The center position is used during the zero and calibration process and displays the instantaneous value of force on the Force Transducer. CAL / ZERO This switch is used initially to calibrate the Texturgage and compensate for the "tare" or weight of the suspended part of the test cell. The zero position can also be used between tests to zero the texturegage's peak readings prior to starting a new test. THE DISPLAY This is the digital readout presented in Tenderometer Units, newtons, pounds of force, joules or inch-pounds depending on the mode of operation. The display has a few coded displays that have meanings. A zero " walking" across the display means that the force transducer is not connected properly. "]]]]" means that the unit needs to be zeroed and "[[[[" means that the unit needs to be calibrated. SPEED CONTROL This knob controls the speed of the hydraulic cylinder. To set the speed, time how long it takes for the cylinder to go from the full up to the full down position. Adjust the speed control knob to give the desired stroke time (speed). Turn the knob clockwise to slow the ram and counter clockwise to speed it up. This setting typically, does not change. PRESSURE CONTROL This knob controls the maximum pressure the hydraulic pump will generate and therefore the maximum force the ram can apply. There is a ten times mechanical advantage in the system so 100 psi on the gage indicates that 1000 pound of force can be generated by the hydraulic ram. To set the pressure, move the ram to either the full up or full down position and leave the motion switch in the up or down position so the ram is trying to continue it's movement but cannot because it is at the end of travel. Adjust the pressure control to the lowest pressure that will still give adequate force for the test. For example, if you expect to be measuring forces around 2500 pounds, set the pressure to at least 250 psi. This will give the unit sufficient pressure to generate the required forces.

PREPARATION AND CALIBRATION 1. Check that the PRESS MOTION SWITCH is in the center position and the force transducer is plugged into the right side of the press, then turn ON the MAIN POWER. 2. Allow the unit to warm up for a few minutes (the display will flash for 2 minutes, the duration of the warm up period). The system will lock in overload and the cylinder will not move during this period. 3. Attach the force transducer to the ram by means of the securing pin (the wire should be positioned to the rear), attach the top part of the test cell to the transducer and lock it by turning the small tab on the bottom front of the transducer. Slide the bottom "box" portion of the test cell into its holder in the lower end of the columns. 4. The box of the cell should slide into place easily. If it does not check the machine to see that it is level. An adjustment screw located on the backstop plate of the press columns controls how far the bottom part of the cell slides into the press. This is adjusted at the factory but may be reset if moved so that the top blades slide down the center of the box. 5. With the PEAK / INTEGRAL SWITCH in the center position and the cell and transducer in place, depress the CAL / ZERO SWITCH to ZERO and release. 6. Now depress and hold the CAL / ZERO SWITCH to CAL to check the force transducer's calibration. If the display is different than the CAL number indicated on the transducer calibration/serial number sticker, increment or decrement the display with the UP / DOWN SWITCH until it reads properly. 7. Depress the ZERO switch to clear the calibration number from the peak memory and place the waste tray into the unit below the cell.

TEST PROCEDURE 1. Remove the test cell and fill the cell by volume or weight with the product. 2. Slide the cell box into the cell guides of the Texturepress. 3. Turn the PEAK \ INTEGRAL switch to PEAK position. (Even if the PEAK is not turned on the TG4-E will remember the highest reading since the ZERO Switch was last depressed). If the reading is other than zero, depress the ZERO switch to zero the peak memory. Depending upon the load cell in use, the display may fluctuate somewhat at zero. This is normal. 4. Press the START switch 5. Record the reading obtained after completion of the test. 6. Remove the cell box to prepare for the next test. Empty the waste tray as needed. TESTING FOOD PRODUCTS Almost any food product may be tested for texture by using the T-2000. It is important to note the factors that will effect accurate and consistent reading however. 1. The samples must all be a consistent size, either by weight or volume. Variations will effect the readings. 2. The samples must be consistent in preparation. Temperature, exposure to air, cook time, and the location from where the sample is taken, can all have dramatic effects on the texture readings. 3. The speed of the test can also change the readings. Make sure that the down stroke time of the test stays consistent.

SOME NOTES ABOUT THE TG4-E TEXTUREGAGE The TG4-E will hold a complete set of cell calibration numbers in memory. This memory is retained even when power is removed from the unit. One drawback to this type of memory is that there is a finite number of times that it can be erased and rewritten. Therefore to eliminate a failure in the cal number memory, any new calibration number will not be written to memory until it has been set and remained unchanged for ten minutes. Due to thermal electronic changes the displayed cal number may change a small amount as the system is left on. One or two points should not have an adverse affect on the reading and should not require resetting the calibration number. The display has a few coded displays that have meanings. A flashing display means that the unit is warming up. A zero " walking" across the display means that the force transducer is not connected properly. "]]]]" means that the unit needs to be zeroed and "[[[[" means that the unit needs to be calibrated. The unit also has a self-diagnostic program and if the display will show an error code if something is detected to be wrong by the internal computer. In this case record the display and contact Food Technology Corporation. THE AUTO CYCLE The AUTO CYCLE was developed to speed repetitive testing and makes the system easier to use. This option allows the operator to simply fill the test cell and push a button to complete a test. When the button is pushed, the hydraulic cylinder goes down, the reading is taken, and the unit returns to its original position after finishing the down stroke. The reading will be held until the test cell is filled and the button is pushed again to start the next test. At that time reading is automatically zeroed and the process is repeated. When using this feature the up/down ram motion switch must be in the center position or the option will not work. There are two sensors located beside the cylinder under the cylinder cover. These are magnetic switches and are actuated by flags that move up and down on a rod attached to the anti-rotation bar.

In the course of shipping or use the flags may move slightly or have to be readjusted to allow the full movement of the cylinder. This will be evident if the Auto cycle feature does not perform properly or the cylinder does not retract fully to allow the cell box to be removed. To adjust the flags, use the position switch to move the cylinder to the absolute top and bottom and mark those positions for reference (the anti rotation bar and the support columns work well for this). Remove the cylinder cover and then slightly loosen the setscrews that secure the flags to the rod. Adjust their positions up or down until the cylinder cycles as close to the full stroke as possible. The bottom sensor is not as critical as the top sensor since the last centimeter or two of movement usually reveals little data. The top sensor, however, needs to be set high enough so the cylinder returns close enough to the top to remove the test cell, but low enough so that the flag actuates it every time and resets. It may take a few trials to get it just right but once it is set retighten the setscrews to secure the new settings. Replace to cylinder cover. INTEGRAL MEASUREMENT The system has the ability to calculate the amount of work done on the sample during the down stroke of a test. The work done equals the multiplication of the force and the distance. Since the force varies during the distance, integration must occur. In English units the display reads in "inch-pounds", in metric units the display reads in "joules". Since the integral calculation is based on distance, the calculations are performed between two position points; when the ram leaves the top of its stroke and it reaches the bottom. There are two magnetic sensors located on the press columns that are use to locate the top and the bottom of the stroke. The texturegage must sense a departure from the top and an arrival at the bottom for an accurate integral to be calculated. Because of this the Integral will not be displayed until the unit has completed its down stroke. If the unit appears to not integrate it may be that the sensors need to be adjusted. Refer to the section on the AUTO CYCLE option for the adjustment procedure.

SECTION FOUR MAINTENANCE The FTC Texture Test System was designed to give a long and trouble-free life. However, as with any instruments there are some dos and don'ts that should be considered. The System is water-resistant but NOT waterproof. A water hose should never be used on the unit, especially the Texturegage section. Cleaning may be accomplished by simply wiping the unit down. If the area is to be hosed down, the System must be covered. The hydraulic power unit should give years of service without any need for repairs. Periodic checks of the oil level should be all that is required. The oil level should be approximately 2 to 3 inches from the bottom of the tank. If the unit should ever need oil it can be purchased from Food Technology. We use and recommend Exxon Esstic 33 Hydraulic oil. Do not mix other oils. The test cell is the heart of the instrument and a little care and special treatment will go a long way in extending its useful life. The cell should be carefully cleaned and dried after each testing session, particularly before it is stored for an extended period of time. The "sharpness" factor of the blades will gradually change with use, thereby affecting the readings. Depending on the use of the machine, we recommend that the cell be returned to Food Technology for calibration on an annual or BI-annual basis. They should be carefully packaged and sent to our plant. Thank you for purchasing Food Technology Corporation equipment. We are always available to help with any problems or questions you may have concerning food texture and quality control. FOOD TECHNOLOGY CORPORATION 45921 Maries Road Suite 120 STERLING, VIRGINIA 20166