Benchtop HTHP Curing Chamber. Part No Instruction Manual. Updated 11/7/2017 Ver. 3.0

Similar documents
Atmospheric Consistometer # (115V) # (220V) Instruction Manual. Updated 1/25/2018 Ver. 6.0

Model 20 Constant Speed Blender : 115 Volt : 230 Volt. Instruction Manual. Updated 11/14/2017 Ver. 4.0

Model 12BL Filter Press, Multi Unit. # Unit # Unit. Instruction Manual. Updated 08/25/2015 Ver. 2.0

HTHP Filter Press for Cement Testing with 175-mL, Double-Capped Test Cell and CO 2 # : (115 V) # : (230 V) Instruction Manual

HTHP Filter Press for Cement Testing with 175-mL, Double-Capped Test Cell and N 2 # : (115 V) # : (230 V) Instruction Manual

Electrical Stability Meter. Part No Instruction Manual. Updated 12/10/2015 Ver. 3.0

Digital Resistivity Meter. Part No Instruction Manual. Updated 09/01/2015 Ver. 3.0

PHPA Polymer Concentration Kit. Part No. # Instruction Manual. Updated 1/23/2017 Ver. 2.0

4-Unit HTHP Filter Press With Load Cell Option. # S VAC # S VAC # S-LC Load Cell Option. Instruction Manual

Laboratory Gas Flow Model # VAC # VAC. Instruction Manual. Updated 11/18/2014 Ver. 1.16

Methylene Blue Test Kit (115 Volt) (230 Volt) Instruction Manual. Updated 2/6/2017 Ver. 4.0

HTHP Filter Press with Threaded Cells for Drilling Fluids Testing. # : 115 Volt # : 230 Volt. Instruction Manual

Twin Cell Ultrasonic Cement Analyzer. Part No Instruction Manual

Ultrasonic Cement Analyzer. Single Cell: # Dual Cell: # Instruction Manual

CRF-840 Rock-Fluid Centrifuge. Part No. # Instruction Manual. Updated 5/18/2018 Ver. 3.0

Dynamic HTHP Filter Press With Magnetic Drive

Model 130 Benchtop Consistometer. # Standard # DAS - With Computer for Data Acquisition and Control. Instruction Manual

Permeability Plugging Tester - P.P.T. 2,000 PSI (13,800 kpa) F (260 C) Part No (115V) Part No (230V) Instruction Manual

Permeability Plugging Tester. # : 115 Volt # : 230 Volt. Instruction Manual

Features. HTHP Filter Press for Drilling Fluids, 500 ml DRILLING FLUIDS EQUIPMENT

Instruction Manual Updated 8/3/2009 Ver. 1.1

Updated 12/13/2013 Ver. 2.6

Instruction Manual Updated 12/17/2013 Ver. 1.1

HTHP Consistometer with Data Acquisition System Model 100 (25,000 PSI, 400 F): DAS (Single Cell) / DAS (Dual Cell)

Features. HTHP Filter Press for Drilling Fluid Testing DRILLING FLUIDS EQUIPMENT

OFI MODEL 130 HTHP BENCHTOP CONSISTOMETER INSTRUCTIONAL MANUAL

CRD600 Automatic Fitting Inserter

CRD610 Automatic Fitting Inserter

OFI MODEL 60 ATMOSPHERIC CONSISTOMETER INSTRUCTIONAL MANUAL

CLEAN ROOM DEVICES, LLC "WHERE TUBING AND FITTINGS COME TOGETHER"

Operator's Manual. Models (¾") (1") (1½ ) (2 ) (3") (4") (6") (8")

EP (Extreme Pressure) and Lubricity Tester

7.3L POWERSTROKE BANJO BOLT KIT Fits L Powerstroke Diesel. Installation Guide

HALLMARK INDUSTRIES INC

CLEAN ROOM DEVICES, LLC "WHERE TUBING AND FITTINGS COME TOGETHER"

HR-20P Pneumatically Controlled Pressure Regulator

Adjustable Angled Incline Conveyor Owners Manual with Operating Instructions

DeZURIK AFR3 Filter Regulator Used on Pneumatic Cylinder Actuators

Read this entire manual before operation begins.

Model 1100B CHG Terminator. Installation Instructions

AMERICAN AVK COMPANY AVK SERIES 24 - HIGH PRESSURE, WET BARREL HYDRANT FIELD MAINTENANCE AND INSTRUCTION MANUAL TABLE OF CONTENTS

StormPro BA Series Sump Pump

AMERICAN AVK COMPANY AVK SERIES 67 - HIGH PRESSURE, POST/FLUSHING HYDRANT FIELD MAINTENANCE AND INSTRUCTION MANUAL TABLE OF CONTENTS

HYDRAULIC LEVELING SYSTEMS OPERATIONS MANUAL (For systems with touch pad part number , , , , or no number at all)

Parts Manual Vermeer/ McLaughlin Mud System 230P

PRYCO, INC. P. O. BOX 108 Mechanicsburg, IL OPERATIONS AND MAINTENANCE MANUAL For PUMP SETS

OWNER S MANUAL SELF-PRIMING PORTABLE UTILITY PUMP

Table of Contents. Technical Information Warning Statement

Parts Manual Vermeer/ McLaughlin Mud System 230P

Operations & Maintenance Manual DFH-1000 DUCTLESS FUME HOOD

Model Series. PumpAgents.com - buy pumps and parts online ELECTRIC BILGE PUMPS. Model Series FEATURES SPECIFICATIONS STANDARD MODELS

8.1.4 The 175 ml, 250 ml, or 500 ml unit may be used for static filtration testing up to and including 175 C (350 F).

TMCo,, Inc. The Measurement Company

Model Speed Viscometer. Features DRILLING FLUIDS EQUIPMENT

OWNERS MANUAL INSTALLATION AND OPERATING INSTRUCTIONS REPAIR PARTS LIST. Centrifugal Pump Primer MODEL 6D, HAN-DEE PRIMER

HIGH PRESS. - HIGH TEMP. (HPHT) FILTER PRESS 4 - UNIT ASSEMBLY 175 ml Capacity - OFI No

GRINNELL Model B302 and Model BN302 Grooved End Butterfly Valves with Gear Operators or Lever-Lock Operators General Description

APCO CRF-100A RUBBER FLAPPER SWING CHECK VALVES

4" ENVIRONMENTAL E-SERIES PUMPS OWNER'S MANUAL. DANGER warns about hazards that will cause. WARNING warns about hazards that can cause

INSTALLATION AND MAINTENANCE MANUAL FORM #PM-126 REV A 12/09

MAINFRAME HOT RUNNER TEMPERATURE CONTROL SYSTEMS. Instruction Manual

Instruction Manual Updated 5/15/2013 Ver. 2.2

Colt Series C400, C500

Warning Statement Read Before Riding

Model NTX7 Series Automatic Battery Charger User s Manual Rev. 1.0 October 17, 2006

Series 957, 957N, 957Z, 957RPDA, 957NRPDA, 957ZRPDA

HTHP Filter Press for Ceramic Disks with 175-mL, Double-Capped Test Cell and CO 2 # : (115 V) # : (230 V) Instruction Manual

FR1118- P10. User s Manual

Pay For Water System DU 1A046A

PERMEABILITY PLUGGING TESTER - P.P.T. OFI Part No

TL-105-D COLUMN HEATER

Pneumatic Cylinder 14 Bore X 22 Stroke Part No. R (Formerly P )

FAN POWERED HEPA FILTER UNIT

Innovatech User Manual. Predator 2400 T H E S U R F A C E P R E P A R A T I O N S P E C I A L I S T S

Read this entire manual before operation begins.

Power Float Manifold. Installation and Operations Manual Module 11A

OPERATION MANUAL Variable Speed Pump Controller Dated: 06/04/2013 Pump Down Application. Document No.: LMSII_V100_OM Page 1 of 8 Model-V100 LMS II

Analog Condensate Monitor Instruction & Maintenance Manual

AIR CLEANERS Model MC 3000 OWNER S MANUAL CAUTION Read complete instructions before operating. Please file for future reference.

PL900 ILLUMINATOR. Fiber-Lite. Operation Manual. Setup. Fiber Optic Connection A PRODUCT OF DOLAN-JENNER INDUSTRIES. Voltage Selection:

INSTALLATION AND MAINTENANCE MANUAL FORM #PM-122 REV A 12/09

TK08 TUNING KIT TECHNICAL MANUAL FOR 16/26/27/28 SERIES SHOCKS Revised 6/17/14

Compressor Clutch Replacement Procedure

PROwatt 150 DC to AC Power Inverter. Owner s Guide

Parkit360 Force. Owner s Manual V2.6.2

DUSTTRAK AEROSOL MONITOR SOLAR POWER KIT MODEL

INTRODUCTION INSTALLATION

Reach ins, Freeezers & Refrigerators Installation & Operation Manual

TK08 TUNING KIT TECHNICAL MANUAL FOR 16/26/27/28 SERIES SHOCKS Revised 6/25/14 Tech Line: (952) Fax (952)

TRINETICS CSD SERIES OIL SWITCH INSTALLATION INSTRUCTIONS

SPH-400 i Series General Manual

INSTALLATION, OPERATION AND MAINTENANCE MANUAL

ULTRAVIOLET WATER DISINFECTION SYSTEM OPERATION & MAINTENANCE MANUAL

AT Clutch Major Service Sizes 25, 55, 115

Plastic Tube Flar ing Syste m

US Patent 0540, 400 Other Patents Pending. Read this manual completely before assembling and riding our PET PRO-FLEX 500. Always wear a Helmet!


The Da-Lite Difference.

Model 2008 I Battery Operated Irrigation Timer with 3/4 in. Anti-Siphon Valve

INSTALLATION INSTRUCTIONS

Transcription:

Benchtop HTHP Curing Chamber Part No. 120-55 Instruction Manual Updated 11/7/2017 Ver. 3.0 OFI Testing Equipment, Inc. 11302 Steeplecrest Dr. Houston, Texas 77065 U.S.A. Tele: 832.320.7300 Fax: 713.880.9886 www.ofite.com Copyright OFITE 2015

Table of Contents Intro...2 Description...2 Components...2 Specifications...3 Setup...4 Preparing the Slurry...6 Operation...7 Termination...9 Maintenance...10 Appendix... 11 Fuses... 11 Temperature Controller...12 Auto Cool...15 Rupture Disc...16 Warranty and Return Policy...17 OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com 1

Intro The OFITE Benchtop HTHP Curing Chamber is designed to prepare well cement specimens for compressive strength tests. It is necessary to determine the amount of time required for a cement to develop compressive strength so that drilling/production operations can be resumed as quickly as possible. The goal is to design a slurry that can quickly develop compressive strength so that the waiting on cement time may be minimized. The OFITE HTHP Curing Chambers provide a means of curing cement specimens under typical downhole temperatures and pressures. Description Cement is poured into a special mold that produces specimens measuring 2" 2" 2". The mold is placed into the test cell and the pressure is increased via an air-driven hydraulic pump. Test temperature is governed by a PID temperature controller, which actuates the heater. After a predetermined amount of time, the temperature of the test cell is reduced by the cooling system. Specimens are removed and the compressive strength is determined as outlined in API Specification 10. Components #120-55-005 High-Pressure Filter #120-55-006 Heater; Thinband; 1,650-Watt; 230-Volt; Qty: 2 #122-052 Rupture Disk; 5,500 PSI (37.9 MPa) #122-083-1 Cement Mold Assembly, 4 Specimen #152-38 AC Power Cord; 3-Conductor; International (Continental European) Optional: #120-55-SP Spare Parts for Benchtop Curing Chamber #120-55-003 Breaker, 16 Amp#120-55-006 Heater, Qty: 2 #122-034 Needle Valve, 15K, Qty: 2 #122-052 Rupture Disk, 5,500 PSI, Qty: 2 #122-072 Fuse, 1 Amp, 5 mm 20 mm, Qty: 8 #122-083-1 Cement Mold Assembly, 4 Specimen OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com 2

Specifications Air Supply Maximum 150 PSI (1035 kpa) Cooling Water Supply 40 PSI (276 kpa) Readout Safety Features Digital Pressure Relief Valve Safety Head with Rupture Disk Power Requirements 220 Volts, 50 / 60 Hz, 40 Amp Weight 215 lb (94.6 kg) Dimensions 25" 16" 20" (63.5 40.6 50.8 cm) Shipping Weight 255 lbs (115.8 kg) Shipping Dimensions 30" 20" 24" (76.2 50.8 61 cm) Industry Standards API Specification 10, ASTM Standard C-109 Maximum Temperature / Pressure Combinations Temperature F ( C) Pressure PSI (kpa) 125 (51.7) 5,250 (36,200) 190 (87.8) 4,800 (33,096) 260 (126.7) 4,400 (30,338) 375 (191) 4,000 (27,600) 400.0 (204.4) 3,800 (26,201) OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com 3

Setup The Benchtop Curing Chamber consists of two modules. The Cell Module houses the test cell, heaters, and pressure fittings. The Control Module houses all of the electronic controls and the timer. 1. Carefully remove both modules from the crate. 2. Make sure the Fill Cell, Air to Cylinder, and Pressure Release valves are closed (turned fully clockwise). 3. Connect a Nitrogen supply (690 1,035 kpa or 100 150 PSI), a water supply, and a drain line to the designated ¼" (6.35 mm) ports on the back of the Cell Module. 4. Plug the data cable into the designated ports on the two modules. 5. Plug the thermocouple into the Thermocouple Port on the top of the Cell Module. Plug the temp data cable (male on both ends) into the Temp Data Port on the top of the Cell Module. Plug the other end of the cable into the designated port on the back of the Control Module. Thermocouple Port Temp Data Port 6. Make sure all of the switches on the control module are off. Plug the unit into a grounded electrical outlet. 7. Turn the Main switch on. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com 4

Temperature Controller Timer Control Module Control Module Control Module: Main Switch - Provides power to the entire unit. Pump Switch - Provides power to the pump. Heat Switch - Provides power to the heaters. Timer Switch - Provides power to the timer. Auto Cool Switch - Automatically begins cooling at the end of the test. Sonalert Switch - During an alarm condition, the unit will beep only if this switch is on. Cool Switch - Manually begins cooling. Alarm Switch - If this switch is on, the unit will enter an alarm condition when the temperature exceeds 400 F (200 C). Temperature Controller - Controls the temperature in the test. (See page 12 for instructions) Timer - Displays the current duration of the test and controls the Auto Cool function. (See page 15 for instructions) Cell Module: Pressure Gauge - Displays the pressure inside the test cell. Regulator - Controls the pressure inside the test cell. Turn it clockwise to open and counter-clockwise to close. Fill Cell Valve - Allows the cell to fill with water, which adds pressure. Air to Cylinder Valve - Pumps air into the cell to displace the water and allow it to drain. Pressure Release Valve - Releases pressure from the cell. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com 5

Setup Preparing the Slurry It is very important to begin a test as soon as possible after preparing the cement slurry. Before the slurry is prepared, make sure the unit is ready to begin a test. If a custom temperature profile is being used, program the temperature controller (see page 12 for instructions) first before mixing the cement. 1. Spread a thin coat of grease on all interior surfaces of the cement molds (#122-083-1). This will make it easier to remove the cement after the test. 2. Prepare the cement slurry according to API Specification 10. 3. Fill half of each mold with the prepared slurry. 4. With a puddling rod, tap the bottom inside of the cement mold to remove any trapped air bubbles. 5. Stir the remaining slurry and fill each mold to overflowing. 6. Puddle each specimen to remove any trapped air bubbles. 7. Using a straight edge, scrape the excess slurry from the top of each specimen to make them level. 8. Attach the top plate and secure it in place with the screws provided. Cement Molds OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com 6

Operation Before beginning a test, make sure all switches are in the off position and all valves are closed. 1. Loosen all eight set screws on the cell cap with a torque wrench. 2. Unscrew the cell cap and remove it from the cell. 3. Prepare the cement slurry as described on page 6. Carefully lower the cement molds into the test cell. 4. Remove the handle from the mold. 5. Lubricate the threads of the cell cap with high-temperature thread lubricant (#165-44). 6. Place the cell cap back onto the cell and screw it in place hand tight. Be careful not to damage the seal rings and the cap itself. 7. Tighten all the set screws on the cell cap by hand. 8. Adjust the supplied torque wrench to 90 inch-pounds. Tighten the first screw until the torque wrench clicks. Skip four screws and tighten the fifth until the torque wrench clicks. Continue in this manner, tightening every fifth screw until all eight have been tightened. 9. Now adjust the torque wrench to 180 inch-pounds and tighten all of the set screws as described in the previous step. The cap must be tightened consistently and uniformly every time to create a good seal. 10. Loosen the thermocouple compression fitting in the center of the cell cap. Insert the thermocouple through the compression fitting. Tighten the fitting finger tight and then loosen it ⅛ of a turn. 11. Make sure the Air to Cylinder and Pressure Release valves are closed. Turn the Regulator completely counter-clockwise. 12. Turn the Pump switch on. 13. Open the Fill Cell valve. Water will flow into the test cell and the displaced air will be forced out through the loosened thermocouple compression fitting. 14. Keep a ⅝" wrench handy and watch the top of the cell. When water begins to flow from the compression fitting, tighten it with the wrench. This will ensure that no air remains in the cell. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com 7

15. Set the pressure in the test cell by turning the Regulator clockwise until the gauge reads the desired pressure. If the pressure rises too high, slowly open the Pressure Release valve. When the pressure has dropped, close the valve. Always release the pressure very slowly to avoid pulling cement into the plumbing. 16. Set the temperature controller on the Control Module to the desired temperature. The temperature controller features both manual and automatic operating modes. To use the manual mode, simply use the up and down arrows to set the test temperature. To use the automatic mode, refer to page 12 for instructions. 17. Turn on the Heater switch. 18. Turn the Sonalert on to be notified of an alarm condition. 19. Program the set points on the timer and press the R button. See page 15 for more information. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com 8

Termination When the test is complete, the cement slurry will be fully cured. The system must be shut down very carefully to avoid injury or damage to equipment. 1. Turn the Heat switch off. If the timer is in automatic mode, press and hold the Run/Hold button until the Run indicator goes off. If the timer is in manual mode, use the arrows to bring the temperature down to 0. 2. If Auto Cool is off, turn on the Cool switch. This will begin circulating water through the test cell. 3. When the temperature in the cell drops below 180 F (80 C), turn off the pump and slowly open the Pressure Release valve. Do not attempt to release the pressure until the temperature is below 180 F (80 C). Always release the pressure very slowly to avoid pulling cement into the plumbing. 4. Close the Fill Cell valve. 5. Slowly open the Air to Cylinder valve. This will force the water out of the cell and into the drain line. When air starts escaping from the drain, close the Air to Cylinder valve. 6. Make sure the pressure gauge reads 0. Slowly loosen the thermocouple compression fitting and remove the thermocouple. 7. Loosen the screws on the cell cap. Unscrew the cap and remove it from the cell. Remove the cement molds from the cell and allow them to cool as directed in API Specification 10. 8. Put the cell cap back on the cell to prevent dust and debris from settling inside. Close all valves and turn all switches off. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com 9

Maintenance 1. One of the most important elements of trouble free operation is keeping the interior of the test cell as clean as possible. Never insert a cement covered cement mold into the test cell and periodically examine the test cell to ensure that it is clean. High-pressure valves wear quickly when exposed to fluids containing cement and other particulate matter. Also make sure the air used for the air supply is clear. 2. OFITE uses a high pressure filter to protect valve stems. It is recommended that filtered water be used in the curing chamber to prevent particulate matter from entering the pump and possibly causing damage. Clean or replace these filters when fluid flow is reduced from that of a clean system. Remember: Filters are inexpensive when compared to the costs of replacing the components they were designed to protect. 3. The test cell cap threads have been lubricated prior to shipment and periodically should be re-lubricated. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com 10

Appendix Fuses The Benchtop Curing Chamber uses several fuses and circuit breakers to protect the electrical components. Each of these should be checked periodically. The fuses are located inside the Control Module behind the left side panel. The circuit breakers are located inside the Cell Module behind the back panel. 1. Main Circuit Breaker - Check if there are any problems with power to the unit. 2. Heater Circuit Breaker - Check if the unit s heaters are not functioning. 3. The three fuses in the Control Module are all 1-amp (#122-072). They each protect different systems within the unit. - F3: Instruments and Cooling - F4: Pump, Heat, and Sonalert - F5: Main Relay Main Circuit Breaker Heater Circuit Breaker Fuses OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com 11

Appendix Temperature Controller The controller allows the temperature profile to be programed for each test. This profile will be divided into at least two segments. Each segment controls a portion of the test and determines the temperature, time, and behavior of the test cell during a specified time period. The four buttons along the bottom of the display provide access to the temperature profiles. Begin by pressing the PAGE button three times. The display will read Prog List. Now, press the SCROLL key repeatedly until the setting you wish to change is shown on the display. Then press either arrow key until the appropriate value is displayed. Once a value has been chosen, press the SCROLL key again to select a new setting. The first group of settings should read as follows: Setting Value Prg1 1 Hb OFF Hb.u 0.0 Rmp.u min Dwl.u min Cyc.n 1 These settings will be the same for every test. Do not change them. 1. Begin by defining the first segment of the test. a. Press the SCROLL key repeatedly until Seg appears on the display. b. Press either arrow key repeatedly until 1 appears on the display. This is now editing segment 1. 2. The first setting is Type. a. Press the SCROLL key until the word Type appears on the display. b. The available options are rmp.r, rmp.t, or dwell. Press either arrow key until the appropriate value appears on the display. Rmp.r programs the controller to steadily increase the temperature by a specified rate (degrees per minute). If this value is chosen, the next option will be Tgt, which is the target temperature and then Rate which is the desired rate the temperature will increase. Rmp.t increases the temperature over a specified time interval (minutes). If this valueis chosen, the next option will be Tgt (target temperature) and then Dur (duration in minutes). Dwell holds the temperature at its current setting for the duration specified. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com 12

Temperature Controller RUN/HOLD Page Scroll 3. Now define the second segment. a. Press the SCROLL key until Seg appears on the display. b. Press either arrow key until 2 appears on the display. This is now editing segment 2. 3. Continue this process with each segment in the test. 4. When you reach the last segment, set the Type to end. The next setting will be End.t. If sop is chosen, the heat will turn off and the test will end. If dwell is chosen, the heat will hold at the current temperature indefinitely. Example 1: Heat the sample at 2.5 per minute and stop at 150. Hold at 150 for 180 minutes and then stop the heat. Setting Value Description Prg 1 Hb OFF Hb.u 0.0 Rmp.u min Dwl.u min Cyc.n 1 OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com 13

Seg 1 Segment 1 Type rmp.r Increase temperature at a specified rate Tgt 150 Heat to 150 Rate 2.5 Increase temperature at 2.5 per minute Seg 2 Segment 2 Type dwell Hold on the current temperature Dur 180 Hold for 180 minutes Seg 3 Segment 3 Type end This is the last segment End.t sop Stop the heat Example 2: Heat the sample to 200 over a period of 90 minutes. Then increase the temperature to 300 at a rate of 3 per minute. Hold that temperature indefinitely. Setting Value Description Prg 1 Hb OFF Hb.u 0.0 Rmp.u min Dwl.u min Cyc.n 1 Seg 1 Segment 1 Type rmp.t Increase temperature for a specified time Tgt 200 Heat to 200 Dur 90 Increase temperature for 90 minutes Seg 2 Segment 2 Type rmp.r Increase temperature at a specified rate Tgt 300 Heat to 300 Rate 3 Increase temperature at 3 per minute Seg 3 Segment 3 Type end This is the last segment End.t dwell Hold at the current temperature indefinitely OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com 14

Appendix Auto Cool The timer on the front of the Control Module controls the automatic cooling system. When the Timer switch is turned on, the timer automatically starts counting upwards from 0. When the R button is pressed, the timer resets to 0 and resumes counting upwards. Two set points can be programmed into the timer. When the first set point is reached, the timer activates the water solenoid and begins circulating water through the test cell. When the second set point is reached, the timer deactivates the solenoid and stops the flow of water into the cell. For example, assume the test is being ran for 5 hours then the cell is cooled for 2 hours. The following steps will program the timer: 1. Press 1 to program setpoint 1. Use the 1 key to set the first digit to 0. Press the 2 key to set the hours to 5 and use the two blank keys to set the hours and minutes to 0. 2. Press E to save these settings. 3. Press 2 to program setpoint 2. Use the 1 key to set the first digit to 0. Press the 2 again to set up the hours to 7 and use the two blank keys to set the hours and minutes to 0. Be sure to program setpoint 2 with the total time from the beginning of the test, not from the last setpoint. In this example, the initial test time of 5 hours plus the cooling time of 2 hours equals a total of 7 hours for setpoint 2. 4. Press E to save these settings. Now when the R button is pressed, the timer will begin counting upwards. At five hours it will begin cooling and at seven hours it will stop. The Auto Cool switch must be on for this feature to work. Timer Front Panel Layout OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com 15

Appendix Rupture Disc The Curing Chamber is assembled with a rupture disk (#122-052) which is rated for a maximum of 5,500 PSI. It is a solid metal preformed differential pressure relief device which is used to protect the system from an overpressure condition. The rupture disk provides instantaneous full-opening within milliseconds of an overpressure situation. The main symptom of a blown rupture disk that the system will not be able to build pressure. The rupture disk is located in the back side of the Curing Chamber. To replace the rupture disk: 1. Turn the Curing Chamber off and depressurize the air system. 2. Use a phillips head screwdriver to remove the rear panel. 3. Use a ⅝" wrench to remove the top and bottom tube fittings to the high pressure tee. 4. Use a 9/16" wrench to remove the high pressure tee assembly from the outlet tube fitting. 5. Orient the tee assembly so that the outlet port faces up. 6. Use two adjustable wrenches to remove the outlet fitting and gain access to the rupture disk. 7. Replace the rupture disk (#122-052), reassemble the housing, and reinstall the tee assembly to the tubing. 8. Replace the rear panel. High Pressure Tee ⅝" Adjustable Wrench Outlet Fitting 9/16" OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com 16

Warranty and Return Policy Warranty: OFI Testing Equipment, Inc. (OFITE) warrants that the products shall be free from liens and defects in title, and shall conform in all respects to the terms of the sales order and the specifications applicable to the products. All products shall be furnished subject to OFITE s standard manufacturing variations and practices. Unless the warranty period is otherwise extended in writing, the following warranty shall apply: if, at any time prior to twelve (12) months from the date of invoice, the products, or any part thereof, do not conform to these warranties or to the specifications applicable thereto, and OFITE is so notified in writing upon discovery, OFITE shall promptly repair or replace the defective products. Notwithstanding the foregoing, OFITE s warranty obligations shall not extend to any use by the buyer of the products in conditions more severe than OFITE s recommendations, nor to any defects which were visually observable by the buyer but which are not promptly brought to OFITE s attention. In the event that the buyer has purchased installation and commissioning services on applicable products, the above warranty shall extend for an additional period of twelve (12) months from the date of the original warranty expiration for such products. In the event that OFITE is requested to provide customized research and development for the buyer, OFITE shall use its best efforts but makes no guarantees to the buyer that any products will be provided. OFITE makes no other warranties or guarantees to the buyer, either express or implied, and the warranties provided in this clause shall be exclusive of any other warranties including ANY IMPLIED OR STATUTORY WARRANTIES OF FITNESS FOR PURPOSE, MERCHANTABILITY, AND OTHER STATUTORY REMEDIES WHICH ARE WAIVED. This limited warranty does not cover any losses or damages that occur as a result of: Improper installation or maintenance of the products Misuse Neglect Adjustment by non-authorized sources Improper environment Excessive or inadequate heating or air conditioning or electrical power failures, surges, or other irregularities Equipment, products, or material not manufactured by OFITE Firmware or hardware that have been modified or altered by a third party Consumable parts (bearings, accessories, etc.) Returns and Repairs: Items being returned must be carefully packaged to prevent damage in shipment and insured against possible damage or loss. OFITE will not be responsible for equipment damaged due to insufficient packaging. Any non-defective items returned to OFITE within ninety (90) days of invoice are subject to a 15% restocking fee. Items returned must be received by OFITE in original condition for it to be accepted. Reagents and special order items will not be accepted for return or refund. OFITE employs experienced personnel to service and repair equipment manufactured by us, as well as other companies. To help expedite the repair process, please include a repair form with all equipment sent to OFITE for repair. Be sure to include your name, company name, phone number, email address, detailed description of work to be done, purchase order number, and a shipping address for returning the equipment. All repairs performed as repair as needed are subject to the ninety (90) day limited warranty. All Certified Repairs are subject to the twelve (12) month limited warranty. Returns and potential warranty repairs require a Return Material Authorization (RMA) number. An RMA form is available from your sales or service representative. Please ship all equipment (with the RMA number for returns or warranty repairs) to the following address: OFI Testing Equipment, Inc. Attn: Repair Department 11302 Steeplecrest Dr. Houston, TX 77065 USA OFITE also offers competitive service contracts for repairing and/or maintaining your lab equipment, including equipment from other manufacturers. For more information about our technical support and repair services, please contact techservice@ofite.com. OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com 17