EM101 LOW FLOW MAGNETIC FLOWMETER INSTRUCTIONS

Similar documents
EM 101 EM EM EM GENERAL INFORMATION. SPECIFICATIONS MATERIALS Wetted Materials - FLOW RANGE

Instructions EM 101 Low Flow Electromagnetic Flow Meters

WMX-series MUNICIPAL/INDUSTRIAL MAGMETER INSTRUCTIONS

EX80 Series EX81 EX82. E X 8 0 S E R I E S E l e c t r o m a g n e t i C F l o w s e n s o r i n s t r u c t i o n s. The Leader in Flow Meter Value.

ARCHIVED. EX80 Series. (Includes Dates 6/28/05 to 10/06) EX81 EX82

TX800-Series FLOW SENSOR INSTRUCTIONS. TX81x TX82x 9 001:2008 ISO CERTIFIED COMPANY TX800-SERIES FLOW SENSOR INSTRUCTIONS

WMP-Series PLASTIC-BODIED MAGMETER INSTRUCTIONS 9001:2008 ISO CERTIFIED COMPANY WMP-SERIES PLASTIC-BODIED MAGMETER INSTRUCTIONS. 3 inch.

PE202. Low Flow Magmeter Instructions. Precision Flow Measurement An ONICON Brand PROUDLY MADE IN THE USA

EX80-Series ELECTROMAGNETIC FLOW SENSOR INSTRUCTIONS EX81 EX82 EX :2008 ISO CERTIFIED COMPANY

TX80-Series FLOW SENSOR INSTRUCTIONS TX81 TX :2008 ISO CERTIFIED COMPANY TX80-SERIES FLOW SENSOR INSTRUCTIONS

F-2000 MODEL RT INSTRUCTION MANUAL. Blue-White. Industries, Ltd.

EX100/200-Series INSERTION MAGNETIC FLOW METER INSTRUCTIONS EX110 EX210 EX150 EX :2008 ISO CERTIFIED COMPANY

SES. Stainless Single-Jet Meter Instructions PROUDLY MADE IN THE USA

IP800-Series FLOW SENSOR INSTRUCTIONS IP810 IP :2008 ISO CERTIFIED COMPANY IP80-SERIES FLOW SENSOR INSTRUCTIONS

IP80-Series FLOW SENSOR INSTRUCTIONS IP81 IP :2008 ISO CERTIFIED COMPANY IP80-SERIES FLOW SENSOR INSTRUCTIONS

imag 4700 SERIES FLANGED MAGMETER

ISTEC MAGNETIC METERS

IP800-Series. Insertion Paddlewheel Flow Sensor Instructions PROUDLY MADE IN THE USA

SERIES FLMG MUNICIPAL/INDUSTRIAL MAGMETER INSTRUCTIONS

AG2000 IRRIGATION MAGMETER INSTRUCTIONS

Electromagnetic Flowmeters. Product Overview.

2100 Turbine Flow Sensor

Rosemount 8750WB Magnetic Flowmeter System

Rosemount 8750WA Magnetic Flowmeter System For Water and Wastewater Industries

S-Series. Low Flow Meter Instructions SPX SPT. Precision Flow Measurement An ONICON Brand PROUDLY MADE IN THE USA

MFM220 & MFM220DC Plastic-Bodied Magmeter

RCM - Direct Reading Flowmeters

TX101/201 Insertion Turbine Instructions

IP800-Series. Insertion Paddlewheel Flow Sensor Instructions PROUDLY MADE IN THE USA

AG2000 Irrigation Magmeter

FT430/440/450 RATE/TOTAL INDICATOR INSTRUCTIONS

TX115/215 Hot-tap Insertion Turbine Instructions

Compu-Flow. Electromagnetic Insertion Flow Meter. Compu-Flow EXM MAG

ACSI MODEL 1406BB-04-AO POWER SUPPLY INSTALLATION INSTRUCTIONS

Insertion magnetic inductive flowmeter

Distributed by: M&M Control Service, Inc MICRO-FLO MICRO-FLO

G2 Industrial Grade Meters

TX101/201 Insertion Turbine Instructions

PMG MAGNETIC FLOWMETER

F-2700 SERIES INSERTION VORTEX FLOW METER

WMX-SERIES Flanged Magmeter

IP800-Series FLOW SENSOR INSTRUCTIONS IP810 IP :2008 ISO CERTIFIED COMPANY IP800-SERIES FLOW SENSOR INSTRUCTIONS

Appendix. Clearance. Skewed profiles. Application Schematics

RESISTIVITY MONITOR/CONTROLLERS

FPI Mag FULL PROFILE INSERTION ELECTROMAGNETIC FLOW METER MODELS 394L AND 395L MAG METER SUBMITTAL

Magnetic Inductive Flowmeter EP / UMF2 (B)

User Manual. T6 Tachometer. Online: Telephone: P.O. Box St. Petersburg, Florida 33736

Installation and Maintenance Instructions. World Leader in Modular Torque Limiters. PTM-4 Load Monitor

CENTRIFEEDER ELECTRONICS

Tracer VMA with AutoReg

Versa II SLR Lubricator

The Polaris MA1 is the first liquid flow meter offered by Magnetrol, expanding the already diverse level and flow measurement portfolio that has been

MFM300 Plastic-Bodied Magmeter

Electromagnetic flowmeters Series FLOMAT Insertion electromagnetic flowmeter for conductive liquids

F-1500 SERIES INSERTION TURBINE FLOW METER

Hydro-Flow 2200 Fixed Insertion Vortex Flow Meter

MICRO-FLO MICRO-FLO. Digital Paddlewheel Flow meter Operating Manual. Blue-White. Industries, Ltd.

PanaFlow MV80. In-line Multivariable Flowmeter for Mass, Temperature, and Pressure. GE Sensing. Applications. Key Benefits

TX 115/215 Hot-tap Insertion Turbine Instructions

A1 Commercial Grade Meters

ARCHITECTURAL CONTROL SYSTEMS, INCORPORATED ST. LOUIS, MISSOURI

DPT TARGET TYPE FLOWMETERS

DW PADDLE-BELLOWS FLOWMETER & SWITCH

FPI-X FULL PROFILE INSERTION ELECTROMAGNETIC FLOW METER MODEL 395X MAG METER SUBMITTAL

Operations Manual. Zero Speed Switch Sensor Model ZS09P

Softstarters. Softstarters Type SSM Medium voltage ,800V 1

Signet 2551 Magmeter Flow Sensor

Ultra Mag SUBMITTAL PACKAGE

TX100/200-SERIES INSERTION TURBINE INSTRUCTIONS TX101/201 TX115/215 ISO 9001:2008 CERTIFIED COMPANY

PRODUCT INFORMATION BULLETIN

Iron Gate Air Pump System Manual

F-2700 SERIES INSERTION VORTEX FLOW METER

Tracer VM with User Interface

DW PADDLE-BELLOWS FLOWMETER & SWITCH

Electromagnetic flowmeters Series FLOMID Electromagnetic flowmeter for conductive liquids

Product Guide: Series III Pump Control Board Set (RoHS)

Softstarters. Softstarters Power electronic brakes Type PEB

Hydro-Flow Model 3100

From. McCROMETER, INC WEST STETSON AVENUE HEMET, CA Phone (951) Toll Free (800) Fax (951)

F-1500 SERIES INSERTION TURBINE FLOW METER FEATURES Belcher Road South, Largo, FL USA Tel +1 (727) Fax +1 (727)

Paddle Wheel Turbine Flow Meters Installation, Operating & Maintenance Manual

Iron Gate Feed Pump Manual

EGM5300C MAGMAX. Compact Electromagnetic Flowmeter GENERAL FEATURES STANDARD SPECIFICATION. General Specification

M SERIES Multi-Jet Cold Water Meter

Distributed by: M&M Control Service, Inc

Positive Displacement Flowmeters

OMP Positive Displacement Flowmeters

DW PADDLE-BELLOWS FLOWMETER & SWITCH

ProSeries NEMA 4X SONIC-PRO. Engineering and Technical Data. Transit Time Ultrasonic Flowmeter. Industries, Ltd. s Ultrasonic Flow Meters

KEL DIFFERENTIAL PRESSURE FLOWMETERS

Hydro-Flow Model 2200 Fixed Insertion Vortex Flowmeter Installation and Operation Manual

XCITE Owner s Manual. Reso-not TM Damping System XCITE 1502C HYDRAULIC POWER SUPPLY

Product Data Sheet 2012 Rev 5 FLOWMETRIX SA SAFMAG Electromagnetic Flowmeter Doc Date

Paddle Wheel Turbine Flow Meters Installation, Operating & Maintenance Manual

Troubleshooting Bosch Proportional Valves

Hydro-Flow Model 3100 Retractable Insertion Vortex Flowmeter Installation and Operation Manual

FLEX-PRO. Peristaltic Metering Pump. SERIES A3V and A4V. Operating Manual. ProSeries. by Blue-White Ind.

Hydro-Flow Model 2200 Installation and Operation Guide B

Full Profile Insertion

Temperature Controllers

Transcription:

Spec Tech Industrial 203 Vest Ave. Valley Park, MO 63088 Phone: 888 SPECTECH Email: sales@spectechind.com www.spectechind.com EM101 EM101 LOW FLOW MAGNETIC FLOWMETER INSTRUCTIONS LOW FLOW MAGNETIC FLOWMETER INSTRUCTIONS The Leader in Flow Meter Value

Table of Contents General Information Features, Specifications, Flow Range... Page 1 Installation and Operation Mounting, Connections, Grounding,, Outputs, Mounting Diagrams... Page 2 Connections and Grounding Flow Meter, Board, Ideal Grounding with Metallic Piping, Standard Grounding... Page 3 FT520 Batch Control... Page 4 Settings Set-Up... Page 5 Troubleshooting Problems, Probable Causes, Things to Try... Back TableS, DIAGRAMS & Charts Features... Page 1 Specifications... Page 1 Flow Range... Page 1 Mounting Diagrams... Page 2 Standard Connections and Grounding Diagrams... Page 3 FT520 Connection Diagram... Page 4 Settings Diagrams... Page 5 Troubleshooting Problems, Probable Causes, Things to Try... Back

GENERAL INFORMATION The EM101 is a small electromagnetic low-flow flowmeter with chemically-resistant plastic wetted parts and a choice of Hastelloy or platinum electrodes suitable for use with a variety of chemicals. Capable of measuring pulsating flows from diaphragm-type metering pumps, it is designed primarily for electrically-conductive chemical injection applications. The 1/4, 3/8, and 1/2 sizes monitor maximum flows of 1, 3, and 8 gallons per minute (or 4, 11, and 30 L/min), respectively. FEATURES Captive plastic cover screw (4) (mounting screws inside) Barb or NPT fittings are available. The EM101 is 115 Vac, 60 Hz powered and includes a rate and total display as well as a variety of outputs. For continuous transmission of a flow signal, there is a 4-20 ma output, a 0-5 volt signal, and a frequency signal proportional to flow. In addition, there are relay alarm outputs for user-set low and high flow levels. Backlit LCD display (16 characters x 2 lines) Keypad switches, tactile PVDF housing Power cord Connector (2) for optional output cables 6 ft. control cables (2) out to display Fusion coated aluminum housing Connector (2) for sensor cables PVDF plastic fittings (2) Male NPT or barb SPECIFICATIONS Flow Meter 10-24 grounding screws (on bottom of meter) Removable bracket for mounting Materials Meter Housing Electrodes O-Ring Housing Rate Units Total Units Power Accuracy Max Fluid Temperature Maximum Pressure Minimum Conductivity Outputs Fusion coated aluminum Platinum or Hastelloy Viton (EPDM optional) Kynar PVDF Gal/Minute, Liters/Second, Gal/Hour Gallons, Liters 115 Vac, 60 Hz /-1% of reading, plus.005 gpm (0.02 lpm) 185 F (85 C) 150 psi (@ 75 F) 20 microsiemens 4-20 ma, 0-5 Volts, both isolated Frequency to 9999 pulses per gallon, isolated (10 ma, 30 Vdc) High alarm, low alarm relay, isolated (100 ma, 110 Vac/Vdc) Flow Range SIZE LOW FLOW CUTOFF MINIMUM FLOW for spec accuracy MAXIMUM FLOW L/MIN GAL/MIN GAL/HR L/MIN GAL/MIN GAL/HR L/MIN GAL/MIN GAL/HR -025-038 -050.04.01.60.26.07 4.2 3.8 1 60.11.03 1.80.76 0.2 12 11 3 180.26.07 4.20 1.9 0.5 30 30 8 480 Page 1

INSTALLATION and OPERATION INSTALLATION Mounting. Mount the display housing to a secure surface with screws or bolts. Remove the front cover to gain access to the mounting holes, directly under the front cover screws. Attach the flow meter to a secure surface using the foot bracket. Alternatively, the unit can be supported by the piping and the foot bracket removed. See mounting diagrams below. CAUTION: Although this meter has an empty pipe detection function, under certain conditions of empty or partially-full pipe the meter may read a flow when there is none. If this is a hazardous condition, mount the meter in such a way as to ensure the meter will always be full of liquid. FLOW METER.218 DIA 3.75 3.00 2.50 3.00 3.50 CONNECTIONS The meter ships with the coil activation and signal leads already connected to the display housing. To connect optional output signal or alarm relay leads, remove the front cover. See the Connections diagram, page 3. Power connection uses a standard power cord. If conduit connection is required, remove the cord and strain relief and use the strain relief hole for a conduit connector. Grounding. For proper operation, one or both of the ground lugs must be well connected to a good quality earth ground. (The ground lugs also retain the foot bracket). See the Grounding diagrams, page 3. The meter comes with a 12 foot ground wire attached. OPERATION. The flow rate is displayed in the time and volume units that have been selected in Set-Up (page 5) (liters/min, for example). The cumulative total flow is displayed in the chosen units, up to eight digits. When the total reaches its maximum, the display resets to zero and begins again. 3.75 6 ft. cables to display Outputs. The analog output varies continuously with the flow. If the output is too jumpy (changes too frequently), it can be damped in one of two ways: 1) increase the averaging time (under the Fast Analog Output setting); or 2) select Disabled (the default condition) under Fast Analog Output and increase the amount of damping using the Low Pass Filter setting. DISPLAY 4.40 2.30 The pulse output will produce a 50% duty cycle pulse at the volume intervals for which it is set - one pulse per liter, for example. Note: Since each pulse consists of equal times on and off, if the interval between pulses is large, the pulse may remain in the on condition for several seconds. 4.93 2.92 4.93 The relay alarm output will only energize if the flow goes above (high) or below (low) the flow alarm setting. The alarm relay will remain energized until the flow exceeds the setpoint by.25% (hysteresis). 1.00.71.59 Page 2 1.25 1.25 1.25

CONNECTIONS and GROUNDING DISPLAY BOARD FLOW METER WHITE GREEN BLACK RED Cable To Sensor To (UNUSED) Power Supply - - - Analog Device 4-20mA Output (Consult Factory for Other Ranges) Fuse,.25 AMP, 5 X 20mm 110 Vac 110 Vac G Pulse Output Low Alarm High Alarm Strain Reliefs and Housing Front are Not Shown IDEAL GROUNDING WITH METALLIC PIPING STANDARD GROUNDING To Earth Ground Mounting Bracket Bare End Under Clamp To Earth Ground Mounting Bracket Removed Connect Ground to Either Screw Page 3

CONNECTIONS FT520 - Batch Control FT520 Pulse Outputs (for metering pumps) Pulse 1 Pulse 2 RESUME COM BATCH Resume Start/Stop Remote Controls Batch Outputs Relay 1 Relay 2 Main power fuse 250mA part # 26926 NO COM NC NO COM NC - 5V 10V 4-20mA A P Active s1 s2 Jumpers Passive A B An Output 0-5 V or 0-10 V or 4-20 ma AC Power Line Neutral Ground Analog Switch Analog Header EM101 Cable - To (UNUSED) Power Supply - - Analog Device 4-20mA Output (Consult Factory for Other Ranges) 110 Vac 110 Vac G Low Alarm High Alarm Page 4

TINGS D I S P L A Y -UP. Use the instructions below to set up your control. Use the key to move from one operation to the next, and use the and keys as described to change settings within operations. Rate Tot This is the power-up display. Rate/total display (shows two seconds after power-up), Press to begin programming. Volume Units G Use the key to choose the volume units: milliliters (ml) liters (L) or gallons (G) Time Units M Use the key to choose the time base for rate: per second (S), per minute (M), per hour (H), per day (D). Tot Dec Point Format 0.0 Use the key to choose number of decimal places: 0, 0.0, 0.00. An. output Type Fast An. Output Rate For Analog Output: Rate For High Alarm: Rate For Low Alarm: Leave Analog Output Type unchanged unless you intend to use the 0-5 V output. Standard setting is for 4-20 ma. Consult factory for 0-5 V output. Leave default ( Fast Analog Output ) unchanged unless you have a demanding closed-loop application. Standard setting is disabled. If you enable this output, use the key to select a response time, 50 to 1000 milliseconds. This will control how rapidly the analog output tracks flow. Use the and key to set the percentage of full scale at which peak analog output (20 ma or 5 V) occurs. On a max 10 gpm meter, 60% would be 6 gpm, for instance. Use the and key to set the percent of full scale at which the high flow alarm will energize. On a 10 gpm meter, 90% would equal 9 gpm, for instance. Use the and key to set the percent of full scale at which the low flow alarm will energize. On a 10 gpm meter, 10% would equal 1 gpm, for instance. Pulse Output Low Pass Digital Filter: Totalizer Reset Use the and key to set the number of pulses (0000 to 9999) per unit. If you have selected liters, it will be pulses per liter, etc. This setting controls the smooth/responsive trade off of the display. Increasing the time makes the display smoother, but slower to respond to change. Use the to set 1, 2, 5, 10, or 20 seconds. If using the meter with pulsing flows, it may be necessary to increase the time. Use the key to switch between totalizer reset Enabled and Disabled. If the reset is enabled, every time the key is pressed during normal operation, the totalizer will reset to zero. Settings finished. Page 5

TROUBLESHOOTING Problem Probable Causes Try... No No power Blown fuse Loose ribbon cable Check for 110 Vac Test fuse, replace if bad Check ribbon cable connections Flow rate always reads 0 Reversed flow direction Missing ground wire Empty pipe Flow rate below minimum Loose wiring or incorrect wiring Fluid conductivity <20 microsiemens/cm Reverse flow connections Install ground wire Install meter in vertical position Use next smaller flow meter Check connections on display board Select another flow meter Flow rate incorrect Missing or incorrect ground wire Fluid conductivity <20 microsiemens/cm Empty pipe Check for proper grounding Select another flow meter Install meter in vertical position SeaMetrics Incorporated 19026 72nd Avenue South Kent, Washington 98032 USA (P) 253.872.0284 (F) 253.872.0285 1.800.975.8153 www.seametrics.com LT-11314-B 11/15/06