Conductor Strength Tests on ACCR 1622TW T13 Pecos Conductor 3M HCC Testing Laboratory August 5, 2011 Phu Trac Sr. Quality Engineer: Phu Trac Technical Manager: Dr. Herve Deve Lead Development Specialist: Dr. Colin McCullough Approved Date 8/8/2011 c{â gütv 8/8/2011 Herve Deve 8/8/2011 Colin McCullouogh
3M ACCR 1622TW-T13 Conductor Strength Tests Laboratory: 3M HCC Testing Laboratory, St Paul, MN, USA Date of Tests: 2/11/11 to 4/12/11 Reference Documents: See Appendix A Test Data: This report contains conductor strength test data for 3M ACCR 1622TW- T13. Test reports for conductor with deadend, conductor with T-Tap and conductor with splice are available in separate reports. Reel Ref Number - 128002 Manufacturing date: 7/2/2009 Requestor: 3M Test Ref Number. Test Numbers 572, 573, 623 Summary of Results Three conductor samples were tested for tensile strength. The conductor tests averaged 107.4% RBS (Rated Breaking Strength) and fully met specification requirements. Summary of Test Procedure: The conductor was tested as a straight load to failure. Testing Equipment Equipment used for tensile testing is a Roberts Tensile Tester machine model number: RTE-LC13 (Serial Number 08142). The tester is capable of testing samples up to 670 kn. The load cell was recalibrated on 3/28/2011 by Instron Calibration Laboratory. The load cell has accuracy better than 1%.
1) Picture of Roberts Tensile Tester 2) Picture of the calibration certificate. Test Method The conductor samples was terminated using cast resin terminations on both ends. For more information on test methods, please see document Testing of Conductor, Full- Tension Compression Dead-ends, and Splices/Joints for ACCR at Room Temperature. Results The RBS (Rated Breaking Strength) of 1622TW-T13 conductor is 246,876 N. The SFML (Sample Minimum Failure Load) for the conductor is 95% of RBS (234,532 N).
Table 1: Table showing all three test results and sample information. Sample Test Number Description of Method Max Load (N) %RBS Failure Mode Conductor 572 Conductor 573 Conductor 623 a) Conductor with direct ramp to failure 261,400 105.9 Gauge break a) Conductor with direct ramp to failure 268,837 108.9 Gauge break a) Conductor with direct ramp to failure 265,403 107.5 Gauge break Analysis / Conclusions Test result show that the 3M ACCR 1622TW-T13 samples has a load to failure that exceeded the 95% SMFL. The average conductor strength result was 107.4% RBS. The failure modes for all tests were gauge breaks. Figure 1: Load versus Time for conductor strength tests. 1622TW-13 Conductor Strength Tests 300000 250000 Load (N) 200000 150000 100000 50000 Test 572 Test 578 Test 623 0 0 50 100 150 200 250 Time (s)
APPENDIX A I. Reference Documents: EN50182 IEC 61284 Testing of Conductor, Full-Tension Compression Dead-ends, and Splices/Joints for ACCR at Room Temperature (attached at bottom, section II). II. Test Method Testing of Conductor, Full-Tension Compression Dead-ends, and Splices/Joints for ACCR at Room Temperature 1.0 Goal: This document covers the sample geometry and test procedure for strength testing of ACCR conductor and compression fittings at room temperature. 2.0 Background: Tensile testing of ACCR compression fittings should avoid sample configurations that use multiple fittings within a single tensile test. For example a length of conductor with a compression dead-end at each end, or a length of conductor with a compression dead-end at each end AND a splice in the center. This test configuration is NOT advisable for testing with ACCR. When a compression accessory is pressed onto a conductor sample, there is some loosening of the aluminum strands at the end of the fitting. In ACSR, this can even be quite severe. In an actual transmission line, this is not usually a cause for concern, because in time the loose aluminum will distribute over the length of the span, and looseness is not significant. During testing of ACSR, there is extra cause for concern, since at the site of the aluminum looseness, the steel core will initially carry all the load. Only after the steel has stretched sufficiently to match the extra length of the aluminum looseness, will the aluminum begin to load. Since the aluminum and steel both have a high ductility and so can stretch sufficiently, the full strength of the aluminum and steel will eventually be accessed and the conductor and fittings can be expected to approach the full strength of the conductor. However, with ACCR, the core has a limited strain to failure, and so if there is sufficient looseness in the aluminum from pressing of say two or three fittings in the short length of a tensile sample rather than one, then the core may reach the failure strain before the aluminum reaches it s peak load. Thus the test is a poor representation of the intended use. The preferred method of testing with ACCR is to isolate just one compression fitting in a test length, and apply resin terminations for the loading points, as outlined in section 3.0 below.
If aluminum slack does occur after sample harvesting and / or after pressing of the accessories, the slack must be removed prior to applying the resin end terminations. This is done by applying the appropriate size pulling grips to the open ends of the sample length, then bringing the sample up to a load of about 3000 lbs (13.3 kn). Take some rectangular wood blocks and gently tap the conductor sample, working your way down from the accessory towards the pulling grips. Finally, use three hose clamps to clamp down on each open end. Once this step is completed, the sample is now ready for resin end terminations. 3.0 Preferred test configurations for use with ACCR: (a) Dead-end Strength Test Apply one compression dead-end fitting at one end of the sample, and a resin end at the other end. Dead-end Resin (b) Splice/Joint Strength Test Apply one compression splice fitting at the center of the sample, and a resin socket-ends at the two free ends. Resin Splice Resin (c) Conductor Only Strength Test Apply resin socket-ends at the two free ends. Resin Resin 3.1 Method for Attaching Resin End-Sockets Details for preparing resin terminations for ACCR can be found in the Test Method Document entitled, Preparation of ACCR Samples using Resin End-Terminations. This is a relatively conventional procedure with care taken not to spread the core wires excessively and impart too much bending strain. 3.2 Fractures Near Terminations
Fractures are considered to be affected by the end termination if they occur within 150 mm (6 inches) of the end termination. These fractures should be bound by a lower rating to account for the stress concentration at and near terminations, such as a 90% RBS requirement, rather than a 95% RBS requirement.. If the event that slipping of wires occurs in the resin terminations (e.g. due to poor wire cleaning, aged resin) and the test result is below 95% RBS, then the test shall be deemed invalid and will be repeated. If the fracture occurs within 150 mm (6 inches) or if slipping of the wires occurs but the test result is still greater then 95% RBS, then the test is deemed a valid test. No retesting will be necessary 4.0 Test Procedure 4.1 Conductor Strength Tests Resin terminations shall be used at all free ends. Sample Gauge Length 10 to 12 ft (3 to 4 m). Conductor is tested as a straight load to failure. Loading shall be applied at a rate equivalent up to 1% strain per minute. This results in a test that takes approximately 40 seconds to reach failure. This is an exception to EN 50182 because the loading rate is faster. In load control, the loading rate to produce this strain rate is calculated as: Conductor RBS 0.6 and given in units of lbs/min or N/min Thus for 1622TW-T13 ACCR conductor, the rate is: 55,500/0.6 = 92,500 lbs/min 246,876/0.6 = 411,460 N/min Install sample in test frame and preload to approximately 1000 lbs (4448 N). Setup strain measuring equipment. Check sample straightness and alignment of ends (visual check) and data acquisition software is running. The force and crosshead displacement are measured throughout the test. The temperature of the test laboratory shall be measured and noted. Measure gauge length and zero stress strain and crosshead displacement. Start to record the test. Perform test as indicated. At the end of the test, note; (i) the breaking load of the conductor (ii) location of fracture (iii) ambient temperature during test (iv) Photographs of test set-up (v) photographs of failure location
4.2 Deadend strength tests (Room Temperature) Firstly a deadend termination shall be applied at one end of the conductor, and then secondly a resin termination shall be applied at the other end. Sample Gauge Length 7.1 to 8.2 ft (2.16 to 2.5 m) SMFL (sample minimum failure load) = 95% RBS. RBS for 1622TW-T13 conductor is 55,500 lbs (246,876 N). SMFL would equal 0.95*55,500 lbs = 52,725 lbs (234,533 N). One sample shall be tested per IEC 61284 /11.5.1. Loading shall be applied at a rate up to 1% strain per minute. This is an exception to IEC 61284 because the loading rate is faster. Install sample in test frame. The temperature of the test laboratory shall be measured and noted. Measure the gauge length. Start recording the test. Increase load to 60% SMFL (31,635 lbs / 140,720 N) Hold for 60 minutes. Note any slip distance at the mouth of the deadend. Steadily increase the load to bring to failure. At the end of the test, note; (i) the breaking load of the conductor (ii) location of fracture (iii) slip of the conductor in the DE (iv) photographs of failure location 4.3 Splice/Joint strength tests Firstly a splice is installed at the center of the intended gauge length, joining two pieces of conductor together. Resin terminations are then applied at the two free ends. Sample Gauge Length 14.2 to 16.4 ft (4.3 to 5.0 m) SMFL (sample minimum failure load) = 95% RBS One sample shall be tested per IEC 61284 /11.5.1. Loading shall be applied at a rate up to 1% strain per minute. This is an exception to IEC 61284 because the loading rate is faster.
Install sample in test frame. The temperature of the test laboratory shall be measured and noted. Measure the gauge length. Start recording the test. Increase load to 60% SMFL (31,635 lbs / 140,720 N) Hold for 60 minutes. Note any slip distance at the mouth of the deadend. Increase load to SMFL. Hold for 1 minute. Steadily increase the load to bring to failure. At the end of the test, note; (i) the breaking load of the conductor (ii) location of fracture (iii) slip of the conductor in the splice (iv) photographs of failure location 5.0 Criteria 5.1 Conductor - gauge length failure and > 95% RBS = Passed - within 150 mm of resin ends and >95% RBS = Passed, no retest required - within 150 mm of resin ends, and < 95% RBS retest required 5.2 Deadend termination - no slip of the conductor in the deadend < 95% RBS - no failure of the conductor or the deadend < 95% RBS 5.3 Splice/Joint termination - no slip of the conductor in the splice < 95% RBS - no failure of the conductor or the splice < 95% RBS 6.0 Reference Documents IEC 61284 EN 50182
7.0 Disclaimer TECHNICAL INFORMATION AND DATA; PRODUCT USE. Technical information and data, recommendations, and other statements provided by 3M are based on information, tests, or experience which 3M believes to be reliable, but the accuracy or completeness of such information is not guaranteed. Some of the standard industry tests conducted by 3M were originally designed for conventional conductors and were adapted by 3M for use with composite conductors; references to such tests should be considered in light of this fact. Such technical information and data are intended for persons with knowledge and technical skills sufficient to assess and apply their own informed judgment to the information. No license under any 3M or third party intellectual property rights is granted or implied with this information. Buyer is responsible for evaluating and determining whether the 3M product is suitable and appropriate for Buyer s particular use and intended application. The conditions of evaluation, selection, and use of a 3M product can vary widely and affect the use and intended application of a 3M product. Because many of these conditions are uniquely within Buyer s knowledge and control, Buyer must evaluate and determine whether the 3M product is suitable and appropriate for a particular use and intended application. 3M products are sold or provided to Buyer solely for Buyer s internal manufacturing use or consumption and may not be repackaged, resold, or redistributed in any manner without the prior written consent of 3M. WARRANTY AND LIMITED REMEDY. Unless stated otherwise in 3M s product literature, packaging inserts or product packaging for individual products, 3M warrants that each 3M product meets the applicable specifications at the time 3M ships the product. Individual products may have additional or different warranties as stated on product literature, package inserts or product packages. 3M MAKES NO OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING, BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE OR ANY IMPLIED WARRANTY ARISING OUT OF A COURSE OF DEALING, CUSTOM OR USAGE OF TRADE. User is responsible for determining whether the 3M product is fit for a particular purpose and suitable for user s application. If the 3M product is defective within the warranty period, your exclusive remedy and 3M s and seller s sole obligation will be, at 3M s option, to repair the product, replace the product, or refund the purchase price. LIMITATION OF LIABILITY. Except where prohibited by law, 3M and seller will not be liable for any loss or damage arising from the 3M product, whether direct, indirect, special, incidental or consequential, regardless of the legal theory asserted, including warranty, contract, negligence or strict liability.