Online sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve

Similar documents
The Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over

High Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell

Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution

Technical Overview. Introduction

Application Note. Author. Abstract. Energy & Chemicals - Petrochemicals. Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany

Totally Automated Method for the Determination of Sudan Dyes in Food via On-Line Filtration, SPE and HPLC Analysis

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

Determination of Sudan Dyes I IV in Curry Paste

Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve

Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC

TECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure

Evaluation of Total Petroleum Hydrocarbon in Soil Using LC with Fraction Collector and GC/MS

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585

HIGH PRESSURE SWITCHING VALVES

Application Note. Authors. Abstract. Energy & Chemicals

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

Agilent G5632 Valve Kit Instructions

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

STABILITY AND RELIABILITY: NEW APPROACHES IN PREPARATIVE HPLC COLUMN DESIGN

Using the PSD for Backflushing on the Agilent 8890 GC System

Accuracy Starts Here. Agilent Captiva Filtration Selection Guide

Agilent Multimode Inlet for Gas Chromatography

Agilent Valve Kit G4231A/B/C and G4232B/C/D

Large Volume Injection of Polycyclic Aromatic Hydrocarbons

Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems

Agilent Valve Kit G4231A/B and G4232A/B/C

Phase Distribution of Ethanol, and Water in Ethyl Esters at K and K

Agilent InfinityLab LC Purification Solutions SELECTION GUIDE

n multi-residue analysis of pesticides olfgang Schwack, Claudia Oellig niversity of Hohenheim

Pipetted Liquid Sampling. Automated sample preparation platform for online LC-MS/MS bioanalysis

2-D GC/MS and the Deans Switch: Theory and Applications. Fred Feyerherm, Agilent Technologies Inc. Houston, Tx

Optimized Method Development of Large Volume Injection for GC/MS/MS of Food Pesticides

Performance Characteristics of the Agilent 1260 Infinity II Bio-Inert LC

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)

Setting up SilFlow for BackFlush in your GC

High-throughput protein aggregate analysis of monoclonal antibodies using a novel dual-channel UHPLC instrument

A Single Method for the Direct Determination of Total Glycerols in All Biodiesels Using Liquid Chromatography and Charged Aerosol Detection

Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS)

C196-E081C. Ultra High Performance Liquid Chromatograph Nexera X2. Specifications

Method Development for Capillary GC Systems. Slide 1

URB '-carbamoylbiphenyl-3-yl cyclohexylcarbamate. DEA Reference Material Collection. Form Chemical Formula Molecular Weight Melting Point ( o C)

Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE

Fast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC

S 1125 HPLC Pump System

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis

Quantify with confidence and screen for the unexpected

Technical Procedure for Gas Chromatography (GC-FID)

Boosting Trace Detection Performance with the Vanquish Diode Array Detector and High-Sensitivity LightPipe Flow Cell

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

Stability Axial Compression Technology Application Ordering Information

Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut

Agilent 1290 Infinity Autosampler

Installation of the 2 position/10 port valve into the 1100 series thermostatted column compartment

The Only No Compromise

Neueste LC Entwicklungen

GC Analysis of PCBs in Fish Oil by Direct Injection Without Any Sample Cleanup. Philip L. Wylie, Ph.D. Sr. Applications Chemist Agilent Technologies

Application Note. Author. Introduction. Energy and Fuels

Agilent 1290 Infinity II Ultra Low Dispersion Kit

NEW HPLC - SYSTEM LINE

Optimized Method for Analysis of Commercial and Prepared Biodiesel using UltraPerformance Convergence Chromatography (UPC 2 )

ACQUITY UPLC I-Class System with 2D Technology

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS

APPLICATION OF SOLID PHASE MICROEXTRACTION (SPME) IN PROFILING HYDROCARBONS IN OIL SPILL CASES

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

Comparison of LC/MS/MS and GC/MSD Analyses of Pesticide Residues in Food Products when Using the QuEChERS Sample Preparation Technique

Application Note. Abstract. Authors. Environmental Analysis

Combustion ion chromatography with a Dionex Integrion HPIC System

The Vanquish Platform: Improved Throughput and Resolution of Xanthones in Mangosteen Pericarp

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

DANI Transformer Oil Gas Analyzer

Agilent Online-SPE Direct Inject Kit G4744A

HPLC System S 600 Series

Quality-by-Design in Method Development. Mijo Stanic General Manager and Technical Director

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC

Agilent 7693A Automated Liquid Sampler

C2, C3, C4 Monomer Analysis

HyperShear TM HPLC Mixers

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System

Set Your Sights on Superior Performance

Agilent 1200 Series High Performance Autosamplers & Micro Well Plate Autosampler

mono-layer ACR High purity Reinforced 1-10 Excellent Fair mono-layer DD High purity Reinforced Excellent Excellent

HyperShear TM HPLC and UHPLC Mixers

Analyte Focusing in Elevated Temperature HPLC

SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS

HPLC SYSTEM S 500 SERIES

Beverage Grade Carbon Dioxide

SPECIFICATION. Bench top triple quadrupole Mass Spectrometry System coupled to a gas performance Chromatography (GC-MS/MS)

Set Your Sights on Superior Performance

ACQUITY UPLC I-Class System Specifications

GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method

Quantification and Characterization of Sulfur in Low-Sulfur Reformulated Gasolines by GC-ICP-MS Application

Supplementary Material

Application. Gas Chromatography June 1995

Analysis of Petroleum Fractions by ASTM D2887

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column

Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis

Transcription:

Online sample cleanup on the Agilent 129 Infinity LC using a built in 2-position/6-port valve Analysis of Sudan red compounds in paprika powder Application Note Food Authors Angelika Gratzfeld-Huesgen Agilent Technologies Waldbronn, Germany Abstract Sudan dyes are Azo-dyes and classified as Group 3 of the potential carcinogenic compounds based on findings of the IARC (International Agency for Research on Cancer). These compounds have been banned for food products in the EU, Japan, and the United States since 24. In this Application Note it is shown how column switching using a 2-position/6-port valve can be used for online sample cleanup. The paprika extract was injected onto the cleaning column, and compounds eluting before and after the peaks of interest were discarded. Sudan red compounds in paprika powder were extracted and sensitively detected by selective ion monitoring (SIM) using an Agilent single quadrupole mass spectrometer. Online cleaning significantly increases sensitivity by almost 6% for Sudan 1 and 2% for Sudan 2. Sudan dyes can be detected down to approximately 1 µg/kg using the applied method.

Introduction Sudan dyes are Azo-dyes and classified as Group 3 of the potential carcinogenic compounds based on findings of the IARC (International Agency for Research on Cancer). These compounds have been banned for food products in the EU, Japan and the United States since 24. If these compounds are detected in spices or products containing spices such as worcestershire sauce, the food products have to be recalled and destroyed 1. In this Application Note we show the use of column switching with a 2-position/6-port valve for online sample cleanup. In addition, it is illustrated how Sudan red compounds in paprika powder can be sensitively detected by selective ion monitoring (SIM) using an Agilent 614 single quadrupole LC/MS system. Experimental The instrument used was an Agilent 129 Infinity LC system, equipped with the following modules: Two Agilent 129 Infinity Binary Pumps with built-in degassing units Agilent 129 Infinity Autosampler An additional filter was installed between the injector valve and column preheating device to protect the cleaning column from clogging. Agilent 129 Infinity Thermostatted Column Compartment Agilent 129 Infinity DAD SL for 16-Hz operation 2-position/6-port valve for online sample cleaning Agilent 614 single quadrupole LC/MS system Agilent ZORBAX RRHD Eclipse Plus C-18 columns, packed with 1.8-µm particles 1 8 6 4 2 Results and discussion Sample preparation of Sudan dyes in spices such as chili or paprika powder is relatively simple and straightforward. The powder is extracted with acetonitrile in an ultrasonic bath. The liquid phase is then filtered and introduced onto an HPLC column. Nevertheless, the extract contains a substantial amount of coloring and non-coloring compounds (Figure 1). 2 4 6 8 1 12 14 Figure 1 Paprika powder extract analyzed using MSD in scan mode. Chromatographic conditions Columns: Analytical column: Agilent ZORBAX RRHD Eclipse Plus C-18, 2.1 mm 1 mm, 1.8 µm Cleaning column: Agilent ZORBAX RRHT Eclipse Plus C-8, 2.1 mm 3 mm, 1.8 µm Mobile phases: A = water, B = acetonitrile For the analytical pump the water was modified with 4 µl TFA Flow analytical pump:.5 ml/; stop time 15 ; post time 3 Flow cleaning pump:.3 ml/; stop time no limit Gradient for analytical pump: At 5% B, at 5 95% B Gradient cleaning pump: At 5% B Column oven: 4 C on both sides, valve switch = next run Valve switch: At.2 position B DAD: 22/2 nm, Ref = off, 45/2, PW >.12, 2 Hz, slit width 4 nm Injector: 1 µl injection volume, needle wash for 6 s 2

Online sample cleanup is performed by a precolumn, or cleaning column, which is acccessed by valve-switching between the analytical and the cleaning columns. This process eliates severe contaation of the analytical column from oils and other residues present in spices. The online cleaning procedure uses two columns, a 2-position/6-port valve, and a second pump to provide flow through both columns simultaneously. The workflow used for cleaning the paprika powder extract online is as follows (Figure 2): Valve in Position A Wasting early eluting compounds Water/ACN=8/2 1 Figure 2 Workflow of online sample cleaning. Valve in Position B Analytical run Gradient 2 to 95% in 5 Run time 9 6.5 Valve in Position A Cleaning of precolumn with 95% ACN and equilibration with start conditions Stop 9 Equilibration of analytical column 3 1. The sample is introduced to the shorter cleaning column (valve position A). All compounds, eluting within 1 ute are wasted. 2. The valve is switched after one ute to operate the cleaning column and analytical column in series (valve position B). The analytical gradient starts and the peaks of interest elute. 3. The valve is switched again to the waste position and all remaining compounds are discarded (valve position A). The acetonitrile percentage for the cleaning column is increased to 95% to flush out all remaining compounds. Then the column is equilibrated to the start conditions. 4. Equilibration of the analytical column to the start conditions takes place after the analytical run. Figure 3 illustrates a scheme of the two valve positions and two different flow directions. Figure 3 Scheme of the two valve positions. 3

Sample preparation of paprika powder The following procedure was used for sample preparation: 1. A 2.5 g amount of paprika powder was extracted with 2 ml of acetonitrile and left for 45 utes in the ultrasonic bath. 2. The liquid phase was filtered through a.45-µm Econofilter from Agilent. 3. A 1 ml amount of filtered liquid was spiked with 5 ng Sudan 2, 3 ng Sudan orange G and 4 ng Sudan. 4. A 5 µl amount of the prepared extract was injected. 5. A 5 µl sample contained 5 pg/µl of Sudan 2. This procedure can detect approximately 1 µg/kg in a Paprika powder sample. Deteration of Sudan red compounds A standard was injected according to the above workflow. The resulting chromatograms are shown in Figure 4. mau 3 25 2 15 1 5 Sudan 2, Target mass 277, 5 ng Sudan 1, Target mass 249, 4 ng Sudan orange G, Target mass 215, 3 ng 1 2 3 4 5 6 7 8 Chromatographic conditions MSD TIC DAD signal at 45 nm Figure 4 Overlay of DAD signal, MSD TIC and extracted ion chromatograms of a standard sample. Columns: Analytical column: Agilent ZORBAX RRHD Eclipse Plus C-18, 2.1 mm 1 mm, 1.8 µm, Cleaning column: Agilent ZORBAX RRHT Eclipse Plus C-8, 2.1 mm 3 mm, 1.8 µm Mobile phases: A= water, B= acetonitrile Analytical pump Water +4 µl TFA Flow analytical pump:.5 ml/; stop time 9 ; post time 3 Flow cleaning pump:.5 ml/; stop time no limit Gradient for analytical pump: At 2% B; at 1 2% B; at 5 95% B Gradient cleaning pump: At 2% B; at 6.5 2% B; at 6.6 95% B; at 8.1 2% B Column oven: 4 C on both sides, valve switch = next run Valve switching: At utes position A; at 1 position B; at 6.5 position A DAD: 22/2 nm, Ref = off, 45/2, PW >.12, 2 Hz, slit width 4 nm Injector: 1 µl injection volume, needle wash for 6 sec MSD: Peak width.3, Positive SIM parameters for Mass 215, 249 and 277, fragmentor = 1, Actual dwell 45 Gas temp = 35, drying gas = 12 L/, Neb Pres = 35 psig, V cap positive = 3V 4

The Agilent 614 single quadrupole MS was used in SIM mode to analyze the Sudan red compounds with highest sensitivity. The target masses were 215, 249 and 277. 3 2 1 5 μl injection volume Sudan orange G, Target mass 215, 3 pg/μl 1 2 3 4 5 6 7 8 The detection conditions were evaluated by injecting 5 µl of the standard sample. The resulting chromatograms are shown in Figure 5. The max pressure achieved was approximately 825 bar. It can be seen that the extract contains a certain amount of compounds detected at the selected parameters. Sufficient resolution using a 1 mm length column can confidently detect the Sudan dyes. A 5 µl injection contained 5 pg/µl of Sudan 2, 3 pg/µl Sudan orange G and 4 pg/µl Sudan 1. Sudan dyes in a paprika powder sample can be detected down to approximately 15 1 5 8 4 8 4 Chromatographic conditions Sudan 1, Target mass 249, 4 pg/μl 1 2 3 4 5 6 7 8 Sudan 2, Target mass 277, 5 pg/μl 1 2 3 4 5 6 7 8 Standard 1 2 3 4 5 6 7 8 Figure 5 Overlay of standard and sample extract chromatograms using Agilent 614 single quadrupole MS in SIM mode. Columns: Analytical column: Agilent ZORBAX RRHD Eclipse Plus C-18, 2.1 mm 1 mm, 1.8 µm, Cleaning column: Agilent ZORBAX RRHT Eclipse Plus C-8, 2.1 mm 3 mm, 1.8 µm Mobile phases: A= water, B= acetonitrile Analytical pump Water +4 µl TFA Flow analytical pump:.5 ml/; stop time 9 ; post time 3 Flow cleaning pump:.5 ml/; stop time no limit Gradient for analytical pump: At 2% B; at 1 2% B; at 5 95% B Gradient cleaning pump: At 2% B; at 6.5 2% B; at 6.6 95% B; at 8.1 2% B Column oven: 4 C on both sides, valve switch = next run Valve switching: At utes position A, at 1 position B at 6.5 position A DAD: 22/2 nm, Ref = off, 45/2, PW >.12, 2 Hz, slit width 4 nm Injector: 1 (standard) or 5 µl (spiked extract) injection volume, needle wash for 6 sec MSD: Peak width.3 ; Positive SIM parameters for Mass 215, 249 and 277, fragmentor = 1; Actual dwell 45; Gas temp = 35, drying gas = 12 L/; Neb Pres = 35 psig; V cap positive = 3 V 5

1 µg/kg. Influence on abundance with and without online cleaning The Sudan standard and the spiked matrix was injected without online cleaning on a 1 mm 2.1 mm Agilent ZORBAX RRHD Eclipse Plus C-18 column to measure the effect of cleaning on abundance results. The resulting standard chromatogram is shown (Figure 6). The injection volume for the Sudan standard was set to 3 µl and the spiked extract to 5 µl, to be comparable with the injection volume used with online cleaning. The spiked matrix was injected under the same chromatographic conditions as the standards without online cleaning. The influence on the abundance of the peaks of interest is significant for Sudan 1 and 2 (Figures 7 and 8). For Sudan orange, the same abundance with and without cleaning was observed (Figure 9). At the retention time of Sudan orange, the matrix is relatively clean, whereas many co-eluting peaks occur for Sudan 1 and 3 (Figure 1). mau 6 4 2 _ 2 2 4 6 8 1 mau 2 1 _ 1 6 4 2 Chromatographic conditions Columns: Mobile phases: Flow : Gradient: Column oven: DAD: Injector: 2 4 6 8 1 2 4 6 8 1 Figure 6 Analysis of Sudan standard without online cleaning. Agilent ZORBAX HD Eclipse Plus C-18, 2.1 mm 1 mm, 1.8 µm, A = Water +4 µl/l TFA, B = acetonitrile.5 ml/; stop time 15 ; post time 3 At 15% B; at 1 9% B 4 C on both sides, valve switch = next run 22/2 nm, Ref = off, 45/2, PW >.12, 2 Hz, slit width 4 nm 3 (standard) or 5 µl (spiked extract) injection volume, needle wash for 6 sec MSD: Peak width.3 ; Positive SIM parameters for Mass 215, 249 and 277, fragmentor = 1; Actual dwell 45; Gas temp = 35, drying gas = 12 L/; Neb Pres = 35 psig; V cap positive = 3 V Without cleaning = 148 Abundance 16 12 8 4 2 4 6 8 1 12 14 16 12 8 4 With cleaning = 16525 Abundance +57.6% 2 4 6 8 1 12 14 Figure 7 Influence of cleaning on abundance for Sudan 1. 6

Cleaning of the matrix before and after elution of the peaks of interest is advantageous if a more sensitive compound analysis is needed. Up to nearly 6% more abundance could be obtained for Sudan 1. Conclusion In this Application Note it is shown how column switching using a 2-position/6-port valve can be employed for online sample cleanup. The paprika extract was injected onto the cleaning column, and compounds eluting before and after the peaks of interest were discarded. Online cleaning can significantly increase sensitivity, close to 6% for Sudan 1 and 2% for Sudan 2. Sudan red compounds in paprika powder were extracted and were detected by SIM mode using an Agilent 614 single quadrupole LC/MS system. Approximately 1 µg/kg Sudan dyes in paprika powder can be detected. References 1. ASTA white paper on Sudan and Related Dyes, A publication of the American Spice Trade Association (http://www.astaspice.org/pubs/ ASTAWhitepaper.cfm) copyright 25 35 3 25 2 15 1 5 8 6 4 2 Without cleaning = 191 Abundance 2 4 6 8 1 12 14 With cleaning = 2341 Abundance = +22.5% 2 4 6 8 1 12 14 Figure 8 Influence of cleaning on abundance of Sudan 2. 35 3 25 2 15 1 5 35 3 25 2 15 1 5 With cleaning = 349 Abundance 2 4 6 8 1 12 14 Without cleaning = 3775 Abundance 2 4 6 8 1 12 14 Figure 9 Influence of cleaning on abundance for Sudan orange. 7

www.agilent.com/chem/129 Agilent Technologies, Inc., 21 February 1, 21 Publication Number 599-5255EN