Operators Manual. Model 3370 Air Cooled Recirculator rev.8/98

Similar documents
Operators Manual. Recirculating Chiller /06/08

Operators Manual. Microprocessor Controlled Water Baths /98

HTC-1200 Recirculating Heater. NESLAB Manual P/N Rev. 04/22/98. Operation and Maintenance Manual

ESE Series Cast Iron Sewage Pumps

MRC 150/300 USER MANUAL. 1 Laird Technologies

PowerFLO 7800 Series 12 Volt DC Motor-Driven Diaphragm Pumps

PowerFLO 5800/5900 Series

LYTRON MODULAR COOLING SYSTEM

TS400Q TS800Q. Auto Pressure Control Pump Instruction Manual. ISO 9001 Certified Walrus America Inc

GARDEN HOSE UTILITY PUMP

Users Manual Certified Series Direct Drive Pump 1-7 LPM

Penguin Gentoo 600B Cordless Sprayer User s Manual


16 Inch Hanging Pendant Barn Light

INSTALLATION AND OPERATION MANUAL

ProFlo FatBoy

Air-Operated Waste Oil Drainer

HR-20P Pneumatically Controlled Pressure Regulator

TPAK Series COOLANT PUMP

TPHK Series Immersible Pump

LUBRICATOR GUN INSTRUCTIONS-PARTS LIST. 10,000 psi (700 bar) Maximum Delivery Pressure. Detachable-type

Open Display Merchandiser Installation and Operation Manual Please read this manual completely before attempting to install or operate this equipment!

Model F822 thru F834 Mulsifyre Directional Spray Nozzles, Open, High Velocity General Description

Model CP420/CP425. Triplex Ceramic Plunger Pump Operating Instructions/ Repair and Service Manual

INSTALLATION INSTRUCTIONS, PARTS LIST AND CONFIGURATION GUIDE FOR THE CARBONATOR. Large Model No

Model AM2 M40 Panel Installation, Operation, and Maintenance Manual

SUPERPLUS WATER FILTRATION SYSTEM

4400 Airless Product Manual

Automotive Air Conditioning Manifold Gauge Sets

Use/Installation Guide

Installation, Operation and Maintenance Manual

Watts Series CSM-61. Flow Measurement/Balancing Valves Sizes: 1 1 4", 1 1 2", 2", 2 1 2", and 3" (32, 40, 50, 65 and 80mm) Installation Instructions

4600 SP Airless Product Manual

STOP CITY PRESSURE BOOSTER PUMP INSTRUCTION MANUAL

SAE Engine Mount. Selection For Pumping Application. Matching Pump End To Engine. Type B Centrifugal Pumps Close coupled pump ends.

Model BP6150. Triplex Ceramic Plunger Pump Operating Instructions/ Manual

60 Watt Industrial LED Low Bay Light

Instruction Manual for Operation & Maintenance

Easy/Tran TF STANDBY POWER FURNACE SWITCH TOOLS NEEDED FOR INSTALLATION. PARTS LIST for Easy / Tran TF FURNACE TRANSFER SWITCH

SAVE FOR CONSUMER INSTALLATION / OWNER'S MANUAL. Single Handle Kitchen Faucet Model # LK2500. Important! You choose the flow rate!

Type D0800. Bantam Cartridge Holder OPERATION MANUAL AND PARTS LIST SERIES 494. Mo de l # Vo ltag e D D /12/08

TPHK Series Immersible Pump

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL PSMR1

INSTRUCTION MANUAL INSTALLATION, OPERATION, AND MAINTENANCE ACTUATOR MANIFOLD

PolyStat Immersion Circulators

High Velocity Series Air Curtains

25 GALLON PORTABLE OIL LIFT

Active Controlled Cooling System

Cleaning System. Instructions/Parts A

Illumifalls Patio Pond

Operations and Parts Manual

MANUAL. Use / Installation. Outdoor Access Doors, Dry Storage, and Storage Drawer

TS Series. MultiStage Centrifugal Pump Instruction Manual. Walrus America Inc. ISO 9001 Certified

Dual Flow Manifold Systems Instruction Manual

Model , Series A 9 in. (23 cm) roller frame with 45 angle and 12 in. reach 1/2 in. (13 mm) nap roller cover

Environmental Series Air Curtains

BroadBand PowerShield. 20 AHr Battery. User Manual

1416 Cabinet Unit. Installation Operation Maintenance Troubleshooting Version Dec/17

COOL DRAFT INSTRUCTIONS OWNER S MANUAL CONTENTS. Cool Draft MiniMister 14 Misting Fan CDMP1420BLU

NEUTRALIZATION SYSTEM FOR OPA & GLUTARALDEHYDE MEDI-NEWT. User Manual. Endoscope reprocessing the the way way it it should be be

Installation Manual DIAPHRAGM WELL TANK

D Instructions/Parts. Siphon Feed Detail Spray Gun D

TS69 TS65 TS55 TS45 TS5768 TS SERIES INSTALLATION/OWNER'S MANUAL

4400 Airless Product Manual

CDS-JOHN BLUE COMPANY

PARTS MANUAL FOR TWO STAGE AIR COMPRESSOR

F PRO. 12 Volt Electric Sprayer. Part #AS76 Copyright 2014 by Hydro-Force Rev. 1/14

StormPro BA Series Sump Pump

Submersible Pond Pump with UV 800 Gallons Per Hour

J. & M. Mfg. Co., Inc. 284 Railroad Street - P.O. Box 547 Fort Recovery, OH Ph: (419) Fax: (419)

VSP SERIES Installation, Operation and Maintenance Manual

INSTALLATION & OPERATION MANUAL

Intended Use: Explanation of Signal Word Consequences

SUBMERSIBLE SUMP PUMPS

Premium Pond Pump 330 Gallons Per Hour

P SERIES PUMPS. 18mm Versions Nickle-Aluminum Bronze Models: P , P , P , P , P , P , P

General Information - Inspection, Delivery, Service and Warranty Forms. Delivery and Inspection Information

MODEL 900 IMPELLER-TYPE FLOW METER

Matala. VersiFlow Series. Instruction and Maintenance Manual

PARTS AND INSTALLATION MANUAL

Pressure Roller with 24-inch Fixed Extension - For application of architectural paints and coatings -

S33 Sump Pump INSTRUCTIONS AND SERVICE MANUAL VERTICAL FLOAT SWITCH S33V1 & S33V1C AUTOMATIC S33P1 & S33PC-1 (CONTROL WITH SERIES PLUG) NOT SHOWN

INSTRUCTIONS AND SERVICE MANUAL WITH PARTS LIST

User Guide IGD Series

Model GP Triplex Ceramic Plunger Pump Operating Instructions/ Manual

OWNER'S MANUAL. Royal Sovereign International Inc. RET Desk Series

Installation, Operation and Maintenance Manual

Universal Bevel Drives Service Manual

SAVE FOR CONSUMER INSTALLATION / OWNER S MANUAL. Includes: Tools: Mystic - Pull-Out Kitchen Faucet LKMY1041CR LKMY1041NK LKLFMY1041CR LKLFMY1041NK D 1

Model 2008 I Battery Operated Irrigation Timer with 3/4 in. Anti-Siphon Valve

6-TON DOUBLE LOCKING JACK STANDS OWNER S MANUAL

Blue Air. Commercial Refrigeration Inc. Installation & Operation Manual Glass Door Countertop Refrigerator

HALLMARK INDUSTRIES INC

PEDESTAL SUMP PUMP. MODEL # Español p. 11. Zoeller is a registered trademark of Zoeller Co. All Rights Reserved.

IMPORTANT OPERATING CONDITIONS. Failure to comply with any of these conditions invalidates the warranty. STANDARD CONFIGURATIONS

INSTALLATION/OWNER'S MANUAL DP " Woofer in Enclosure

OWNER S MANUAL REMOTE CONTROL, DUAL HEAD FOGGING SYSTEMS

Z100A Clinical Centrifuge

RipTide. Brewing Pump. Assembly, Operation, & Maintenance

Telescoping Pressure Washer Wand 18'

Transcription:

Operators Manual Model 3370 Air Cooled Recirculator 110-080 rev.8/98

Table of contents Section 1. General Information 1.1 Warranty 1.2 Unpacking Section 2. Product Information 2.1 Description 2.2 Specification Chart Section 3. Operation and Set Up 3.1 Location 3.2 Plumbing 3.3 Filling the Reservoir 3.4 Operation 3.5 Adjusting By-Pass Relief Valve Section 4. Maintenance 4.1 Air Filter 4.2 Inlet Water Filter 4.3 Positive Displacement Pump Motor 4.4 Periodic Inspection

Section 1. General Information 1.1 Warranty Thank you for your purchase. We are confident it will serve you for a long time. Our warranty to you is as follows: The manufacturer agrees to correct for the original user of this product, either by repair, or at the manufacturer s election, by replacement, any defect in parts which develops within 24 months after delivery of this product to the original user. The warranty period for labor is 12 months after delivery. In the event of replacement, the replacement unit will be warranted for 90 days or for the remainder of the original unit s parts or labor warranty period, whichever is longer. If this product should require service, contact the manufacturer s office for instructions. When return of a product is necessary, a return authorization number will be assigned and the product should be shipped, (transportation charges pre-paid), to the indicated service center. To insure prompt handling, the return authorization number should be placed on the outside of the package and a detailed explanation of the defect enclosed with the item. This warranty shall not apply if the defect or malfunction was caused by accident, neglect, unreasonable use, improper service, or other causes not arising out of defects in material or workmanship. There are no warranties, expressed or implied, including, but not limited to, those of merchantability or fitness for a particular purpose which extends beyond the description and period set forth herein. The manufacturer s sole obligation under this warranty is limited to the repair or replacement of a defective product and the manufacturer shall not, in any event, be liable for any incidental or consequential damages of any kind resulting from use or possession of this product. Some states do not allow: (A) limitations on how long an implied warranty lasts or (B) the exclusion or limitation of incidental or consequential damages, so the above limitations or exclusions may not apply to you. This warranty gives you specific legal rights. You may also have other rights which vary from state to state. 1.2 Unpacking Your recirculator is shipped in a special carton. Retain the carton and all packing materials until the unit is completely assembled and working properly. Set up and run the unit immediately to confirm proper operation. Beyond one week, your unit may be warranty repaired, but not replaced. If the unit is damaged or does not operate properly, contact the transportation company, file a damage claim and contact the company where your unit was purchased. Read all instructions pertaining to safety, setup and operation. Proper operation is the users responsibility.

Section 2. Product Information 2.1 Description This air cooled unit continuously pumps near ambient temperature water from a reservoir to an external heat source, through a cooling radiator, and then returns cooler water to the reservoir. By blowing ambient air over the radiator, water flowing inside is cooled and the heat is exchanged into the atmosphere. 2.2 Specification Chart Cooling Capacity (Using Water @ 25 C) Dimensions (w x d x h) External Fluid Connections Reservoir Capacity Maximum Liquid Temperature Maximum Fluid Flow Rate By-Pass Relief Valve Shipping Weight 3700 Watts (12,600 BTU) 15 x 15 x 25 in. (38 x 38 x 63.5 cm) Inlet & outlet, 1/ 2 in. female NPT 1.3 gallons (4.8L) 170 F (77 C) 2 gpm (7.57 lpm) Factory pre-set to 50 psi. Adjustable 20 to 100psi. 69 Lbs (31.3kg) Electrical Specifications 120V, 60Hz, 1ph, 5.5A or 240V, 50Hz, 1ph, 2.75A Recommended Coolant Distilled Water Note: For 50Hz Models, derate cooling by 17% Section 3. Operation and Set Up 3.1 Location Locate your recirculator on a level surface free from drafts and direct sunlight. Do not place it where there are corrosive fumes, excessive moisture, high room temperatures, or excessively dusty areas. Recirculators must be 18 inches minimum away from vertical surfaces so air flow is not restricted. Install as close as possible to the heat source. Avoid voltage drops by using properly grounded power outlets wired with 14 gauge or larger diameter wire and if possible, be close to the power distribution panel. The use of extension cords is not recommended. This will avoid low line voltage problems. 3.2 Plumbing The supply inlet and outlet on the rear of the recirculator are plainly marked. Connect the recirculator outlet to the inlet of the heat source. The outlet of the heat source should be connected to the inlet of the recirculator. Use heavy wall tubing, preferably reinforced, to make connections. Minimize bends in the tubing as well as the length of tubing from the recirculator to the heat source. Do not run the connecting tubing near any type of heat source. Be sure all connections are secure and clamped before filling with water.

3.3 Filling the Reservoir Distilled water is recommended. To prevent algae growth in the system, shield tubing from light. Insulation around the tubing will suffice. Use an algaecide in the cooling water to minimize algae growth (PolyScience Lab Algicide is recommended to prevent algae growth). Do NOT use chlorine bleach as an algaecide. Remove the back panel and unscrew the reservoir cap. Carefully fill the reservoir with distilled water. If possible, also fill the inlet and outlet lines. Although additional water may have to be added later to completely fill the system, replace the reservoir cap at this time. Do not replace the back panel until the reservoir and whole system are full. Remove back grill panel to access the reservoir cap. Wettable components of this recirculator are copper, brass, PVC, polyethylene, nylon and stainless steel. Any fluids used in the circulating system must be compatible with these materials. Do NOT use the following fluids: 1. Solutions containing automotive antifreeze 2. Hard tap water 3. Deionized water with a specific resistance > 1 megohm 4. Any flammable fluids 5. Any concentrations of acids or bases 6. Chemical solutions with halides, bleach or chromates Figure 3.3 WARNING: Do not use a flammable liquid as a fire hazard may result. 3.4 Operation Connect the AC power cord to a power source of the same voltage characteristics as indicated on the serial number label on the rear of the circulator. Be sure that the power outlet is properly grounded. Press the recirculator ON using the Power Switch on the front panel. If the unit does not start, press the Power Switch OFF, press the circuit breaker button on the rear, then turn the Power Switch ON again. The cooling fan and pump motor will begin to function. Power Switch Figure 3.4

Observe the liquid level in the reservoir. If it is low, or if the Low-Liquid Warning light is on, remove the reservoir cap and add water until the reservoir is full. The recirculator may be temporarily operated with the reservoir cap removed when filling. The reservoir cap must not be left open during normal operation. Check all plumbing connections for leaks. If leaks are found, do not continue to operate the recirculator until all connections are secure. The Low-Liquid Warning light on the front panel will turn on if the fluid in the reservoir is low. When this happens, immediately replace the fluid as necessary. NEVER LET THE RECIRCULATOR PUMP RUN DRY. 3.5 Safety By-Pass Relief Valve Adjustment Procedure Flat Head Screw Driver Adjustment Screw (Cover Nut Not Shown) Safety By-Pass Adjustment Valve Pump Head Figure 3.5 To protect the pump in case of blockage or shut off of fluid flow, a Safety By-Pass Relief Value is built into the pump head. This permits fluid to continue flowing within the pump and minimizes the possibility of damage due to a dry pump. Adjustment of the Safety By-Pass Pressure Relief (factory set at 50 psi): 1. A fluid-filled pressure gauge with a range of 0 to 150 psi (not supplied) must be attached to the recirculator outlet fitting so as to prevent fluid flow. 2. The reservoir must be filled as indicated in Section 3.4 3. Remove the side access panel from the recirculator to gain access to the Safety By-Pass Adjustment Screw on the Pump Head. (See Fig. 3.5) 4. Turn the unit ON and check for leaks. Be sure pressure gauge is not leaking. The gauge should display approximately 50 psi. 5. Remove the Cover Nut on the Pump Head to access the Adjustment Screw. 6. To INCREASE the Safety By-Pass Relief pressure setting, use a flat head screwdriver to turn the Adjustment Screw clockwise. While reading the pressure gauge, turn the screw until the desired setting is achieved. To DECREASE the pressure setting, turn the pressure Adjustment Screw counter-clockwise until the desired setting is achieved. CAUTION: Maximum pressure is 100psi. Care must be taken to limit adjustment of Safety By-Pass Relief Valve setting to 100psi or less. Damage may occur at settings exceeding 100 psi and will void warranty.

Section 4. Maintenance 4.1 Air Filter Set up a periodic maintenance schedule to clean the removable air filter at least monthly. Frequency of cleanings will depend on the atmospheric conditions around the recirculator location, as well as length of time the unit is operated. 4.2 Inlet Water Filter The precision, positive displacement pump can be damaged by mineral scale or other foreign particles in the cooling stream. Therefore the pump inlet is protected by a nylon filter holder containing an 80 mesh stainless steel screen. Clean this screen periodically. To remove any initial particles, clean the screen a few days after first starting the unit. 4.3 Positive Displacement Pump Motor The front and rear bearings of the 1 /3 HP pump motor require periodic oiling. Unplug the recirculator from the power source. To access the motor, remove the screws which hold the sides and front cover to the recirculator's frame. The pump is located under the fan and heat exchanger. Oiling instructions are on the motor. With heavy use, oil the motor once a year. With medium use, oil every 3 years. With light use, oil every 5 years. For each oiling, use 30 drops of non detergent SAE 20 oil in both bearings. 4.4 Periodic Inspection All components should be inspected at least once a month or more frequently if the recirculator is used continuously. Examine all fittings for leaks and also check the fluid level. If the level is low, check the heat source as well as the recirculator for leaks. Replace the recirculating fluid every three months. If algae growth is noted, replace more frequently.