Navigator ADS551 Low level dissolved oxygen Replacement / Upgrade procedures wet section spares

Similar documents
Navigator 550 Sodium wet-section

Endura AZ10 Oxygen probe Sensor sub-assembly tested spares kit

Shut-off valves and accessories for pressure measuring instruments

The perfect solution for wellhead flow measurement with CoriolisMaster, ProcessMaster and VortexMaster

ABB MEASUREMENT & ANALYTICS DATA SHEET. Series and Katharometer power supply units

ABB MEASUREMENT & ANALYTICS DATA SHEET series ph and Redox (ORP) industrial dip electrode systems

AP300 Series Combination ph/redox (ORP) Sensors. Measurement made easy. A reliable economic solution for a wide range of applications

ABB MEASUREMENT & ANALYTICS DATA SHEET Hydrazine monitor

ABB measurement & analytics data sheet. Model M26 Manifold line for 266 pressure transmitters

Moving Coil Indicators with direct temperature sensor input

Valve Blocks. For direct flanging to pressure and differential pressure transmitters. Drop-forging housing. Replaceable valve seat

Control Valve 23 / 24 Single-seat straight-way valve with exchangeable seat ring

In addition to this instruction sheet, the following are included in this kit (some additional parts are included for future maintenance):

5000TOC Sensor Service Manual

Models NAA, NAB, NAD Pneumatic differential pressure transmitters. Deltapi N Series A complete range of pneumatic pressure transmitters

Stem Assembly Supplement

Electric Diesel Pump EDP/12, EDP/24

ABB ME ASURE ME NT & AN AL YT ICS D AT A SHE ET. Electronic unit and accessories Retrofitting on Contrac

DB4604 GMR-SD and GMR40-SD Disc Brake Caliper - Spring Applied, Air Released

RSD100 (Contrac) Electrical Linear Actuator

ABB MEASUREMENT & ANALYTICS DATA SHEET. AC200 Series Industrial 2-electrode conductivity cells with rapid temperature response

FLEXIBLE PENDANTS with Axis MEDICAL GAS TERMINAL UNITS

FPD201 Single-chamber orifice fitting

RIGID PENDANT. Phoenix Pipeline Products Limited. Installation, Operation and Maintenance Manual

Model WEP Electro-pneumatic positioner. WEP positioner A valve control without compromise

BCV30 DN20 - Blowdown Control Valve Installation and Maintenance Instructions

BRIDGE-BOY Load limitation electronics with 1 or 3 set points

Welker Adjustable Probe with Check Valve and Welker Automatic Insertion Probe with Check Valve

2013 IMPCO Technologies, Inc. Page 1 of 20 EPR Repair Kit Instructions (PPI-122, Rev-B)

SLM - Sealing Liquid Monitor INSTRUCTIONS FOR USE

MAINTENANCE DESCRIPTION. Specifications Information and Repair Parts Manual &

I & M Mark 78 Series. Ideal Installation. Start-Up. Installation & Maintenance Instructions for Mark 78 Control Valves (1-1/2-2 )

R Series Electromagnetic flowmeter Reduced-bore flow sensors. The smart solution for remote applications

Solar Thermal. Grant Solar Pump Station Installation & Servicing Instructions

2600T Series Pressure Transmitters Custom Linearization Table with ABB Asset Vision Basic

I & M Mark 78 Series. Ideal Installation. Start-Up. Installation & Maintenance Instructions for Mark 78 Control Valves (1/2-1 )

ABB MEASUREMENT & ANALYTICS DATA SHEET. TEIP11 I/P signal converter for standard signals

EAS822 (Contrac) Power Electronic Unit

Instruction Manual. E1M40, E1M80, E2M40 and E2M80 E2M40S and E2M80S Rotary Vacuum Pumps. Item Number

II 2GD. xxxx. Westlock 2200 Series Quantum - ATEX certified Installation & Operating Instructions

Sub Section Title Page No. 1 Introduction 3. 2 Significant Equipment Detail 3. 3 Installation Guidelines 4. 4 Routine Maintenance 4

EQUALIZER International Limited 10T(I) Integral Hydraulic Spreading Wedge Repair Instruction Manual

C15C C15C. Page 1 of 20

SERVICE MANUAL OF SERVOMOTORS HD/HR RANGE

ABB cable glands SCG series Stainless Steel 316L, general purpose cable glands

HOT WASHER MODEL NO: KING 125 OPERATION & MAINTENANCE INSTRUCTIONS PART NO: LS1009

Relay Retrofit Program for REX 521 Cutting Tool Safety Guide

Technical Description. Wireless Automation Accessories

INSTALLATION/OPERATION/MAINTENANCE INSTRUCTIONS FOR ARCHON MODELS WD2010L, WD2010, WD2010H WASHDOWN STATIONS. ARCHON Industries, Inc.

A Issue F. Instruction Manual. EMF3, EMF10 and EMF20 EMF20 Oil Mist Filter

Sentinel 250 Fire Hydrant

SP2 Options Installation and Maintenance Instructions


Fitting Instructions

18SP680Rev3 EPA04 MBE 4000 Car Hauler Low Pressure Fuel Lines

1 ½-Inch High Pressure Pumps Refer to pump manual for General Operating and Safety Instructions.

RSDE10 / RSDE20 (Contrac) Electrical Linear Actuator

I & M 8000 Series. Ideal Installation Schematic. Preferred Installation. Trouble Shooting

With regards to safety it is not necessary to check for correct operation.

ABB i-bus KNX Magnetic Contact EnOcean, 868 MHz MKE/A , 2CDG R0011

User Guide IM/TORBAR-EN Rev. D. Averaging pitot tubes

Fast-Flo Pump INSTRUCTIONS-PARTS LIST. Table of Contents 1:1 RATIO

Redundancy unit CP-C.1-A-RU Accessory for CP-C.1 range power supplies

Model 3051 Sensor Module Replacement

INSTALLATION INSTRUCTIONS SERVICE KIT - MONSOON EXTRA S1.4 bar SINGLE Part No MONSOON EXTRA U1.4 bar SINGLE Part No.

ABB Automation. Variable Area Dial Flow Indicator Series 10A2227. Series 10A2227 Variable Area Dial Flow Indicator

IOM IR-110-X. Solenoid Control Valve

KEYSTONE FIGURE 79 PNEUMATIC ACTUATOR OPERATING AND MAINTENANCE INSTRUCTIONS

INSTRUCTION, HALDEX PUMP LVS KIT P/N

FPD220 meter 2 di Dual-chamber orifice fitting with double isolation

TM-VE436MC.KIT. VE436MC Roll Grooving Tool WARNING EXPLODED VIEW OF SLIDE UPGRADE KIT SLIDE UPGRADE KIT INSTRUCTIONS

CHESTERTON FLOW GUARDIAN S50 AND SP50 SINGLE FLOWMETER INSTALLATION INSTRUCTIONS

GMR-S and GMR40-S Disc Brake Caliper - Spring Applied, Air Released

PN9000 Series Pneumatic Actuators Installation and Maintenance Instructions

BA440 PISTON. Hydraulic Air Compressor

Fisher V250 Ball Valve

Section /4200 Series Printers

EH Mechanical Booster Pumps

Rosemount 402 and 402VP

SW14.5TI INTEGRAL HYDRAULIC FLANGE SPREADING WEDGE. Repair Manual INNOVATION IN ITS MOST FUNCTIONAL FORM

Kit. Hydraulic Kit Installation Guide REV B Table 1 Component List. Item Component Part Number Qty

SITRANS F. Ultrasonic Flowmeters. Hi-Precision Mounts. Operating Instructions. Answers for industry.

WPS500X 500 psi /34.5 bar Automotive Pressure Transducer. User s Guide

Operating and Maintenance Instructions for: Figure 79 Pneumatic Actuators (U/E options)

ASNU Battleship MK2 Fitting Instructions

Removal and Installation of Fuel Injection Pumps

Operation and Maintenance Instructions

1100W PORTABLE GENERATOR

CV Master Cylinders. Publication CB 006. Typical Assembly

Mini contactors B6, BC6 with 3 pole

300 Series INSTRUCTION MANUAL. Compressed Air Filters Models 302 (grade) through 317 (grade)

Fisher 644 and 645 Differential Pressure Pump Governors

Breezair TBA 550 Installation Manual

20 TONNE HYDRAULIC PRESS MODEL NO: CSA20FBT

Page 1 of 19. Part# /10/2006

Escape Cylinder and Valve

MKIV Universal Hopper Service Manual TSP053.doc Issue 2.1 June 2004

Air Operated 1:1 Oil Ratio Pump OP-11

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER SAMPLE / INJECTION PUMP

High Pressure Leak Diagnostics System Part No & NA USER MANUAL LEAK DETECTION SERVICE

Transcription:

Instruction replacement procedure INS/ANAINST/025 EN Rev. A Navigator ADS551 Low level dissolved oxygen Replacement / Upgrade procedures wet section spares Kit references: flowcell, tundish, flowmeter, drain valve, Modbus cable Measurement made easy 1 Introduction This publication details replacement / upgrade procedures for the following Navigator ADS551 wet section spares: Flowcell PCB assembly, PCB housing cover, PCB housing gasket (replacement page 4): AW502 228 flowcell PCB housing gasket AW502 227 flowcell PCB housing cover AW502 225 PCB assembly Tundish assembly (replacement page 5): AW502 065 tundish assembly Flowmeter assembly (replacement page 5 / upgrade page 9): AW502 250 flowmeter assembly (upgrade kit) AW502 060 flowmeter assembly (replacement kit) Drain valve assembly (replacement page 6 / upgrade page 10): AW502 240 drain valve assembly (upgrade kit) AW502 056 drain valve assembly (replacement kit) Temperature sensor assembly (replacement page 6): AW502 220 temperature sensor assembly Modbus cable assembly (replacement page 7): AW502 090 1.5 m (4.9 ft. cable) AW502 091 5 m (16.4 ft. cable) AW502 092 10 m (32.8 ft. cable) Flow control valve (replacement page 8 / upgrade page 11): AW502 270 flow control valve assembly (upgrade kit imperial) AW502 275 flow control valve assembly (upgrade kit metric) AW502 068 flow control valve assembly (replacement kit imperial) AW502 069 flow control valve assembly (replacement kit metric) Before carrying out any procedures, read Section 3 Health & Safety, page 2. These procedures must be carried out by a suitably-trained technician. 2 Tools / Items required Pozidrive screwdriver Slot-head terminal screwdriver Anti-static strap Spanners (adjustable)

3 Health & Safety WARNING Bodily injury These procedures must be carried out by a trained technician. Chemical Ensure personal protective equipment (PPE) such as gloves and eye protection are worn during any maintenance. Observe all health and safety procedures for handling chemicals. To familiarize yourself with handling precautions, dangers and emergency procedures, always review the Material Safety Data Sheets prior to handling containers, reservoirs and delivery systems that contain chemical reagents and standards. Take care if cleaning any spillages and observe all relevant safety instructions. Wipe up any spillages using clean water. Electrical Isolate all high voltage supplies to the transmitter before performing replacement procedures. The wet-section is vulnerable to electrostatic damage. Wear an anti-static strap or dismantle the wet-section on an anti-static workbench. Ensure all electrical connections are kept dry at all times. General Shut off the external sample supply to the wet-section and drain the flowcell - refer to the wet-section Operating instructions (OI/ADS550-EN) for flowcell drainage options. When a procedure is complete, restore power to the transmitter and sample to the wet-section at the correct flow rate. If necessary, calibrate the wet-section refer to the wet-section Operating Instructions (OI/ADS550-EN) for calibration instructions. Perform general cleaning of the wet section using a damp cloth only mild detergent can be used as a cleaning aid. Do not use Acetone or any organic solvents. 2 INS/ANAINST/025 EN Rev. A Navigator ADS551 Low level dissolved oxygen Replacement / Upgrade procedures wet section spares

4 Overview Flowmeter assembly Section 5.3, page 5 (replacement) Section 6.1, page 9 (upgrade) Tundish assembly Section 5.2, page 5 (replacement) Flowcell PCB assembly Section 5.1, page 4 (replacement) Temperature sensor assembly Section 5.5, page 6 (replacement) Flow control valve assembly Section 5.7, page 8 (replacement) Section 6.3, page 11 (upgrade) Modbus cable assembly Section 5.6, page 7 (replacement) Drain valve assembly Section 5.4, page 6 (replacement) Section 6.2, page 10 (upgrade) Fig. 4.1 ADS551 Wet section spares item locations Navigator ADS551 Low level dissolved oxygen Replacement / Upgrade procedures wet section spares INS/ANAINST/025 EN Rev. A 3

5 Replacement procedures 5.1 Replacing the flowcell PCB housing gasket, PCB housing cover and PCB Part numbers: AW502 228 flowcell PCB housing gasket AW502 227 flowcell PCB housing cover AW502 225 PCB assembly Referring to Fig. 5.1: 1. Remove and retain 4 flowcell PCB cover fixing screws and washers A and remove PCB cover B. 2. Remove flowcell PCB housing gasket C. If replacing the gasket, discard. 3. If replacing PCB assembly D, disconnect all cables from the terminal block on the PCB assembly, loosen all cable glands then remove cables. It is necessary to partially remove the glands, blanking plugs and the LLDO connector to remove PCB assembly D. 4. Remove and retain 2 PCB fixing screws and fibre washers E and remove PCB assembly D. Take care not to lose 2 spacers F fitted on the screw threads behind the PCB assembly. 8. Refit, or for new items, fit flowcell PCB housing gasket C and cover B using 4 PCB cover fixing screws and washers A. (Ensure flowcell PCB gasket C is located correctly.) F A D B E Fig. 5.1 Replacing the flowcell PCB housing gasket, PCB housing and PCB C To secure the new PCB into the flowcell PCB housing, pass 2 PCB fixing screws and fibre washers E through the PCB mounting holes, fit the spacers F onto the screw threads. 5. Carefully offer PCB assembly D up into the flowcell PCB housing and tighten 2 PCB fixing screws. 6. Remake cable connections (disconnected at step 3) refer to Fig. 5.2 and Table 5.1 for terminal connections. 7. Tighten the cable glands, plugs and LLDO connector into the flowcell PCB housing to ensure a watertight seal. TB1 TB3 TB5 TB4 TB2 White Screen Red Black Screen Blue Red Brown Black Green Black Red White Red Serial (Tx) cable Drain valve cable Temp sensor cable Fig. 5.2 Cable connections to transmitter ADS551 Flowmeter cable LLDO cable Cable Serial Drain valve Temp sensor LLDO sensor Flowmeter Color White Green Screen Black Red Red Black White Screen Red Blue Red Brown Black Terminal block ID TB1 TB3 TB5 TB4 TB2 Description Data ve Data +ve Screen 0 V 24 V +ve ve Pt1000 Pt1000 +ve LLDO sensor ve LLDO sensor +ve GND ve Table 5.1 Cable connections to transmitter ADS551 4 INS/ANAINST/025 EN Rev. A Navigator ADS551 Low level dissolved oxygen Replacement / Upgrade procedures wet section spares

5.2 Replacing the tundish assembly AW502 065 tundish assembly Referring to Fig. 5.3: 1. Rotate sample outlet tube A until the open end is free of tundish assembly B. 2. Disconnect drain tube C from tundish assembly B. 3. Unscrew and remove 2 tundish retaining screws D taking care to retain spacers E located between the rear of the tundish assembly and the main case. 4. Remove old tundish assembly B. 5. Fit the new tundish assembly by reversing the removal procedure. 5.3 Replacing the flowmeter assembly AW502 060 Referring to Fig. 5.4: 1. Depress manual override button A on the drain valve assembly to drain the flowcell. 2. Remove the flowcell PCB cover as detailed in step 1 of Section 5.1, page 4. 3. Loosen cable gland B at the flowcell PCB housing C. 4. Disconnect flowmeter cable D from wet-section PCB terminals TB2 see Fig. 5.7, page 7 and remove cable. 5. Depress the QD coupling collar E on QD coupling F and lift the top QD coupling F complete with sample outlet tube G away from the flowmeter H. 6. Depress the QD coupling collar K on QD coupling L and lift the flowmeter away. 7. Fit the replacement flowmeter into the QD coupling L and refit QD coupling F to the flowmeter outlet. Ensure orientation of the flow arrow on the flowmeter is upwards. Fig. 5.3 Replacing the tundish assembly 8. Pass the flowmeter cable through the cable gland B at the flowcell PCB housing C and terminal connections to PCB terminals TB2 Fig. 5.7, page 7. 9. Tighten cable gland B and refit the PCB cover as detailed in step 1 of Section 5.1, page 4. 10. Ensure the sample outlet tube G is located correctly in the tundish assembly J. Fig. 5.4 Replacing the flowmeter assembly Navigator ADS551 Low level dissolved oxygen Replacement / Upgrade procedures wet section spares INS/ANAINST/025 EN Rev. A 5

5.4 Replacing the drain valve assembly and drain valve cable assembly AW502 056 Referring to Fig. 5.5: 1. Depress manual override button A on drain valve B to drain the wet-section. 2. Disconnect clip-on cable connector C from the end of the drain valve coil housing by depressing clip D. 3. Disconnect the drain tube from barbed outlet connector E on the bottom of drain valve B. 4. Lock hex coupling G and rotate nut H clockwise to release couplings. Remove the valve assembly, complete with coupling G. Remove coupling G from the valve assembly and fit to new valve, complete with new bonded seal I. 5. If fitting a new drain valve cable assembly, remove the wet-section cover as detailed in step 1 of Section 5.1, page 4, loosen the drain valve cable gland in the wet-section PCB housing and proceed with steps 6, 7and 8 (below). If using the existing cable, re-connect the clip-on cable connector C to the end of the drain valve coil housing. 6. Disconnect the drain valve cable from the terminal block see Fig. 5.7, page 7 and discard the cable. 7. Feed the replacement drain valve cable through the cable gland and make connections to PCB wet-section terminals TB3 (see Fig. 5.3, page 5) and at the transmitter (see Operating instruction OI/ADS550-EN). 8. Connect the drain valve clip-on cable connector C to the end of the drain valve coil housing. 9. Reconnect the drain tube to barbed outlet connector E on bottom of drain valve B. A I H F 5.5 Replacing the temperature sensor assembly AW502 220 Referring to Fig. 5.5: 1. Depress manual override button A on drain valve B to drain the wet-section. Referring to Fig. 5.6, page 7: 2. Remove and retain 4 flowcell PCB cover fixing screws and washers C and remove PCB cover D. 3. Remove flowcell PCB housing gasket E. 4. Loosen cable gland F. 5. Loosen terminal screws G and disconnect temperature sensor cables H from TB5. 6. Withdraw sensor cables H from flowcell PCB housing I through cable gland F. 7. Unscrew the knurled nut on temperature sensor assembly J and carefully withdraw the temperature sensor from flowcell housing K. Check existing small O-ring L fitted on the end of the temperature sensor is withdrawn along with temperature sensor assembly J. 8. Ensure a new O-ring L is securely located on the new temperature sensor, insert the sensor into flowcell housing K and tighten the knurled nut on temperature sensor assembly J finger-tight. 9. Feed cables from the new temperature sensor through cable gland F, make connections correctly at TB5 and tighten screws G. 10. Tighten cable gland F. 11. Refit flowcell PCB housing gasket E ensuring it is located correctly. 12. Refit PCB flowcell cover D and secure with 4 flowcell PCB cover fixing screws and washers C. D C B E G Fig. 5.5 Replacing the drain valve assembly and drain valve cable assembly 6 INS/ANAINST/025 EN Rev. A Navigator ADS551 Low level dissolved oxygen Replacement / Upgrade procedures wet section spares

TB1 TB3 TB5 TB4 TB2 White Screen Red Black Screen Blue Red Brown Black Green Black Red White Red 5.6 Replacing the Modbus cable assembly Part numbers: AW502 090 / 1.5 m (4.9 ft.) AW502 091 / 5 m (16.4 ft.) AW502 092 / 10 m (32.8 ft.) AW502 093 / 20 m (65.6 ft.) Referring to Fig. 5.1, page 4: 1. Remove the PCB cover. E F Referring to Fig. 5.7: 2. Loosen cable gland A at the wet-section PCB housing B and disconnect the Modbus cable C from terminal block connections D (TB1) marked white, green, screen, black and red. 3. Withdraw Modbus cable C from the PCB housing and flowcell case assembly and discard. 4. Feed the replacement Modbus cable through cable gland A and remake connections to the (TB1) white, green, screen, red and black terminal connections at the wet section PCB and at the transmitter. 5. Refit the PCB cover by reversing the removal procedure. TB1 TB3 TB5 TB4 TB2 White Screen Red Green Black Black Screen Blue Red Brown Black Red White Red Fig. 5.6 Replacing the temperature sensor assembly Fig. 5.7 Replacing the Modbus cable assembly Navigator ADS551 Low level dissolved oxygen Replacement / Upgrade procedures wet section spares INS/ANAINST/025 EN Rev. A 7

5.7 Replacing the flow control valve assembly AW502 068 (imperial, for 1 /4 in. pipe) AW502 069 (metric, for 6 mm pipe) Ensure the correct kit has been ordered (6 mm pipe or ¼ in. pipe). Before proceeding, shut off flow to the flowcell assembly. Referring to Fig. 5.8: 1. Depress manual override button A on drain valve B to drain the wet-section. 2. Unscrew coupling nut C at the flow control valve inlet and remove sample feed pipe D. 3. Unscrew coupling nut E and remove old flow control valve F. 4. Offer up replacement flow control valve F to inlet stub G on flowcell assembly H and with the flow control valve in the vertical position, lightly nip up coupling nut E do not tighten at this stage. 5. Push sample feed pipe D into the base of replacement flow control valve F and lightly nip up coupling nut C do not tighten at this stage. 6. Check that sample feed pipe D and flow control valve F are in the correct orientation. 7. Tighten coupling nut E (on the valve outlet side) and coupling nut C (on the sample inlet pipe) finger-tight, then turn another 1 to 1 1 /4 turns using a spanner. 8. Turn the flow control valve F to the shut position; reintroduce the sample upstream at a low flow rate initially and check for leakage. If any leaks are detected, slowly tighten coupling nuts E and C until no further leakage is detected. 9. Open the feed line upstream to the required volume flow rate and check for leaks. If no leaks are detected, open flow control valve F until the required flow rate is present. 1 to 1 1 /4 turns Fig. 5.8 Replacing the flow control valve assembly 1 to 1 1 /4 turns 8 INS/ANAINST/025 EN Rev. A Navigator ADS551 Low level dissolved oxygen Replacement / Upgrade procedures wet section spares

6 Upgrade procedures 6.1 Upgrading the flowmeter AW502 250 Referring to Fig. 6.1: 1. Depress manual override button A on the drain valve assembly to drain the flowcell. 2. Remove outlet tube B from QD coupling C by depressing collar D on QD coupling C and pulling tube B up. 3. Unscrew and remove QD coupling C and wipe up any water spillage. 4. Screw new QD coupling C in and fit: flowmeter E (upper) QD coupling F sample outlet tube G Ensure sample outlet tube G is located correctly over tundish assembly H. 5. Remove the flowcell PCB cover as detailed in step 1 of Section 5.1, page 4. 6. Remove blanking plug I from front of flowcell PCB housing J and replace with supplied cable gland (and O-ring) K. 7. Pass flowmeter cable L through cable gland K at flowcell PCB housing J and make connections to PCB terminal block TB2 refer to Fig. 5.7, page 7. 8. Tighten cable gland K to form a watertight seal. 9. Check sample outlet tube G is located correctly over tundish assembly H and restart sample flow. Referring to Operating instruction OI/ADS550-EN: 10. Configure the transmitter to read the output from the flowmeter by selecting Sensor Setup, scrolling to Flowmeter and selecting Enabled. (This enables auto calibrations to be configured.) Upgraded flowmeter fitted and connected Fig. 6.1 Upgrading the flowmeter assembly Navigator ADS551 Low level dissolved oxygen Replacement / Upgrade procedures wet section spares INS/ANAINST/025 EN Rev. A 9

6.2 Upgrading the drain valve assembly Upgrade kit part number: AW502 240 Referring to Fig. 6.2: 1. Unscrew and remove blanking plug A (with O-ring) at base of flowcell assembly B. There will be sample left in the flowcell so it is advisable to have a beaker or similar receptacle to catch the sample. 2. Fit sealing washer C to male adaptor coupling D, screw this assembly into flowcell base B and tighten. 3. Fit Boded seal E onto thread of male stud coupling F, screw into the upstream port of the drain solenoid valve G and tighten note arrow position on valve body. 4. Wrap PTFE tape around the thread of barbed connector H and screw into the outlet port of the drain solenoid valve G. 5. Offer drain solenoid valve assembly G up to male adaptor coupling D and push onto spigot; orientate the valve (as shown in Fig. 6.2) and tighten couplings D and F together ensure couplings are fully tightened. 6. Fit the supplied length of drain tube I onto the barbed connector J on drain valve G and route as short and straight as possible to a suitable drain. 7. Remove blanking plug K from the underside of flowcell PCB housing L and replace with supplied cable gland and O-ring M. 8. Feed cable N from drain solenoid valve assembly G through the cable gland M and make connections to PCB terminal block TB3 refer to Fig. 5.7, page 7. 9. Tighten cable gland M to form a watertight seal. L K M C D F E G H J I B A Referring to Fig. 6.1, page 9: 10. Check sample outlet tube G is located correctly over tundish assembly H and restart sample flow. Referring to Operating instruction OI/ADS550-EN: 11. Configure the transmitter by selecting Sensor Setup, scrolling to Calibration Valve and selecting Enabled. (This enables auto calibrations to be configured.) Correct drain valve orientation Upgraded drain valve fitted and connected Fig. 6.2 Upgrading the drain valve assembly 10 INS/ANAINST/025 EN Rev. A Navigator ADS551 Low level dissolved oxygen Replacement / Upgrade procedures wet section spares

6.3 Upgrading the flow control valve assembly AW502 270 (imperial, for 1 /4 in. pipe) AW502 275 (metric, for 6 mm pipe) Ensure the correct kit has been ordered (6 mm pipe or ¼ in. pipe). Before proceeding, shut off flow to the flowcell assembly. Referring to Fig. 6.3: 1. Depress manual override button A on drain valve B to drain the wet-section. 2. Unscrew coupling nut C on sample inlet coupling D and remove sample feed pipe E. 3. Unscrew the remainder of coupling D from the side of flowcell assembly F and remove along with nylon washer G. 4. Fit new 3 /8 in. BSP x 6 mm or 1 /4 in. coupling (D) and new nylon washer G into flowcell inlet port block H and tighten. Take care not to overtighten as this may strip the thread in the inlet port block H. 5. Offer up flow control valve I to the inlet stub J of coupling D (above), and with the valve in the vertical position, lightly nip up coupling nut K do not tighten at this stage. 6. Push the sample feed pipe into the base of flow control valve I and lightly nip up coupling nut L it may be necessary to cut or bend the pipe to arrange it into the correct orientation for mating with the flow control valve. 7. Check that sample feed pipe E and flow control valve I are in the correct orientation. 8. Tighten coupling nuts K (on the valve outlet side) and L (on the sample inlet pipe) finger-tight and then turn another 1 to 1 1 /4 turns using a spanner. 9. Turn flow control valve I to the shut position; reintroduce the sample upstream at a low flow rate initially and check for leakage. If any leaks are detected, slowly tighten coupling nuts L and K until no further leakage is detected. 10. Open the feed line upstream to the required volume flow rate and check for leaks. If no leaks are detected, open flow control valve I until the required flow rate is present. 1 to 1 1 /4 turns 1 to 1 1 /4 turns Fig. 6.3 Upgrading the flow control valve assembly Navigator ADS551 Low level dissolved oxygen Replacement / Upgrade procedures wet section spares INS/ANAINST/025 EN Rev. A 11

Contact us ABB Limited Process Automation Howard Road, St. Neots Cambridgeshire, PE19 8EU UK Tel: +44 (0)870 600 6122 Fax: +44 (0)1480 213 339 Mail: enquiries.mp.uk@gb.abb.com ABB Inc. Process Automation 125 E. County Line Road Warminster, PA 18974 USA Tel: +1 215 674 6000 Fax: +1 215 674 7183 Note We reserve the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document. We reserve all rights in this document and in the subject matter and illustrations contained therein. Any reproduction, disclosure to third parties or utilization of its contents in whole or in parts is forbidden without prior written consent of ABB. Copyright 2016 ABB All rights reserved 3KXA494540R2001 INS/ANAINST/025 EN Rev. A 04.2016 www.abb.com/measurement Sales Service