PERFORMANCE OF THE NEW PRIMARY TORQUE STANDARD MACHINE OF INMETRO, BRAZIL

Similar documents
INFLUENCE OF CROSS FORCES AND BENDING MOMENTS ON REFERENCE TORQUE SENSORS FOR TORQUE WRENCH CALIBRATION

Hybrid torque standard machine for 1 kn m developed in CENAM

LEVER OPTIMIZATION FOR TORQUE STANDARD MACHINES

A PORTABLE SYSTEM FOR THE CALIBRATIONOF TRANSDUCERS AND TORQUE WRENCHES: THE CALIBRATION BELL

A GERMAN TORQUE COMPARISON FROM 20 N m TO 200 N m AND ITS RESULTS

ANALYTICAL STUDY ON THE UNCERTAINTY OF LOAD CELLS CALIBRATED WITH DEADWEIGHT-FORCE-STANDARD MACHINE AND FORCE-COMPARATOR MACHINE

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF TORQUE METROLOGY (Torque Generating Device Calibration)

IMGC-CNR Activities to Improve the Force Chain in Italy. C. Ferrero and C. Marinari Istituto di Metrologia G. Colonnetti (IMGC-CNR) Torino - ITALY

THE DESIGN OF 1 N M TORQUE STANDARD MACHINE AT NIM

AVL PLU 131U FLOW METER

RENOVATION OF NMIJ'S 5 MN HYDRAULIC AMPLIFICATION TYPE FORCE STANDARD MACHINE

The Design, Development and Commissioning of a 2 knm Torque Standard Machine

TRACEABILITY OF THE CALIBRATION OF TEST CAR FOR ROLL BRAKE TESTER Aimo Pusa Raute-MIKES Finland, by Raute Precision Oy, Force and Mass Laboratory

Tension and Compression Load Cell Model 8435

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF FORCE METROLOGY:- CALIBRATION OF UNIAXIAL TESTING MACHINES

Measuring equipment for the development of efficient drive trains using sensor telemetry in the 200 C range

Add-on of the 2-MN force standard machine for testing load cells

A New verification system of mobile scales

Inspekt. Static electromechanical universal testing machines. Competency generates results

UK Weighing Federation Technical Articles

Load Cell for Manually Operated Presses Model 8451

MODELING SUSPENSION DAMPER MODULES USING LS-DYNA

Accurate measurement of compressed air consumption and detection of leaks. Measuring the individual consumption per customer / cost centre

FEASIBILITY STYDY OF CHAIN DRIVE IN WATER HYDRAULIC ROTARY JOINT

Permanent Multipath Clamp-On Transit Time Flow Meter

Multi capacity load cell prototype

Torque transducers the complete range for all tasks. Sensors and transducers

Publishable Summary for 14IND14 MNm Torque Torque measurement in the MN m range

Axial Piston Fixed Pump A17FNO Series 10

Industry 4.0 Is Unthinkable Without Calibration!

Small hydrocarbon flow calibration facilities at NMIJ

K3D300 ±200kN. Description

K3D300 ±50kN. Description

Shock tube based dynamic calibration of pressure sensors

White paper: Pneumatics or electrics important criteria when choosing technology

Digital torque transducers. T40 family. For high-precision torque measurement technology

Verifying the accuracy of involute gear measuring machines R.C. Frazer and J. Hu Design Unit, Stephenson Building, University ofnewcastle upon Tyne,

SPECIFICATION SHEET Pulse tool / Impact Wrench Testing System MODEL : IWT-101

AVL SHOT TO SHOT PLU 131 UREA FLOW MEASURING SYSTEM

Dead-weight tester in compact design Model CPB3800

Institute for Design of Electrical Measuring Instruments, Swatantryaveer Tatya Tope Marg, Chunabhatti, Sion, Mumbai, Maharashtra

PowerLink JC Series Torque Transducer User Manual

National comparison on verification of fuel dispensers

STIFF TORQUE TRANSDUCER WITH HIGH OVERLOAD CAPABILITY AND DIRECT FREQUENCY OUTPUT

Small Tool Instruments and Data Management

For the National Voluntary Laboratory Accreditation Program

Development of Dynamic Calibration Machine for Pressure Transducers

Preparing the New EuroLoop Liquid Flow Facility for Accreditation. Jos van der Grinten - NMi EuroLoop

Hardware Testing of Photovoltaic Inverter Loss of Mains Protection Performance

Operation Manual for Torque Sensors

Global Flow Measurement Calibration Consistency

ROTARY MODULES. Rotary modules

Figure 1 Linear Output Hall Effect Transducer (LOHET TM )

Technical Datasheet. HM R Series. Turbine Flow Meters. with Female Inch Threads

Pressure balance Differential pressure version Model CPB5600DP

Alcalab Private Limited, 3rd Floor, Ashiana Trade Centre, Adityapur, Jamshedpur, Jharkhand. Discipline Mechanical Calibration Issue Date

Technical Datasheet. HM AC Series. Turbine Flow Meters. for High Pressure Applications up to 1,400 bar

6 Options and Accessories

OPTIMISING CHASSIS ALIGNMENT USING VEHICLE SENSORS

Storvik HAL Compactor

Proven to be better. Development trends in industrial rolling bearings

Mass Comparators Advanced Radwag solutions for traceability of measurement

Technical Datasheet. HM U Series. Turbine Flow Meters. for lubricating liquids

SOLUTIONS FOR SAFE HOT COIL EVACUATION AND COIL HANDLING IN CASE OF THICK AND HIGH STRENGTH STEEL

Flow Research and Test Center

Highly dynamic control of a test bench for highspeed train pantographs

Hydraulic Deadweight Tester Model CPB5800

DFX Series Liquid Ultrasonic Flow Meters

Flow Research and Test Center. Creating resource technology for today and tomorrow.

Inline Flow meter for compressed air and gases DN15 (1/2 ) - DN80 (3 )

inter.noise 2000 The 29th International Congress and Exhibition on Noise Control Engineering August 2000, Nice, FRANCE

Config file is loaded in controller; parameters are shown in tuning tab of SMAC control center

Contents. Page. 1. Product description. 2. The AXC line of linear axes. 3. AXLT line of linear tables. AXC and AXS product overview...

S-SERIES DISPLACEMENT TRANSDUCERS

JJG Translated English of Chinese Standard: JJG Wayne Zheng et. al

LUCAS ACE - System Overview

Skid against Curb simulation using Abaqus/Explicit

SOME INTERESTING ESTING FEATURES OF TURBOCHARGER ROTOR DYNAMICS

Building instructions Lounger M

Application of ABAQUS to Analyzing Shrink Fitting Process of Semi Built-up Type Marine Engine Crankshaft

STIFFNESS CHARACTERISTICS OF MAIN BEARINGS FOUNDATION OF MARINE ENGINE

Test Report No

THERMOELECTRIC SAMPLE CONDITIONER SYSTEM (TESC)

Config file is loaded in controller; parameters are shown in tuning tab of SMAC control center

Legal Metrology Control of a Liquefied Petroleum Gas Transfer Standard and Field Practice

The Dynamic Characteristics of the Torque Sensor by Bearing Interference Fit

EMC-HD. C 01_2 Subheadline_15pt/7.2mm

Autocal Solutions Pvt. Ltd., Plot No. BG-74, Gala No. 1 & 2, Jai Tulja Bhavani Indl. Premises, Telco Road, MIDC Bhosari, Pune, Maharashtra

DESIGN SOLUTION OF SEMI-AUTOMATIC LINEAR LOADING MEASUREMENT LINE

Application of Steering Robot in the Test of Vehicle Dynamic Characteristics

Selecting Weights and Certificates

FUNDAMENTALS OF INSERTION TURBINE METERS Les Bottoms Thermo Electron Corporation, Flow Systems

Bicycle Hardware in the Loop Simulator for Braking Dynamics Assistance System

22. Modern trends in diagnostics of technical condition of material handling equipment drives

Heating chamber with mechanical convection

An overview of the dynamic calibration of piezoelectric pressure transducers

More Precision. mainsensor Magneto-inductive displacement sensor

Orbipore CPS91 and CPS91D

Laboratory Methods for Testing Actuators

Heating chamber with mechanical convection

Transcription:

PERFORMANCE OF THE NEW PRIMARY TORQUE STANDARD MACHINE OF INMETRO, BRAZIL R. Oliveira 1, L. Cabral 1, U. Kolwinski 2, D. Schwind 2 1 INMETRO National Institute of Metrology, Xerem, Brazil 2 GTM Gassmann Theiss Messtechnik, Bickenbach, Germany ABSTRACT A new design of primary torque standard machines (TSM) with feedback-control of the lever support was described at IMEKO XVI. World Congress in Vienna [1]. The purpose of this paper is to present the performance of a hand operated option model in the wide range from 2 N m up to 3000 N m. Influences of hand operation are estimated as well as ambient influences like temperature changes and air flow. A protection chamber against the ambient influences was built and the improvement is shown. The TSM has three mass stacks to cover all transducer capacities from 20 N m up to 3000 N m in each ten steps. The masses are calibrated within 5 10-6 by German DKD. The lever arms were precisely adjusted and also DKD-calibrated within 5 10-6. The adjustment procedure is described. The measurement uncertainty is calculated by the design data according GUM. In addition the best measurement capability of the TSM and the relative deviations are calculated by the results of several torque transfer standards, former calibrated at PTB. Both calculations are compared. The next steps of the upgrade concept from a hand-operated model to a full-automated model are illuminated regarding the requirements in Brazil. 1. INTRODUCTION The increasing level of industrialisation in Brazil demands an improvement in the National Torque Standard of INMETRO regarding the reduction of uncertainty and a better traceability. In order to reach an immediate implementation and to be in line with the budget, the decision was made to acquire a hand operated dead-weight-tsm with the possibility of a subsequent automation. The target relative uncertainty was fixed to 1 10-4 (k=2). From a metrological side of view the advantages of a completely automated standard machine are the possibilities to cover the lever and the mass handling against disturbing influences and to guarantee an asymptotically load change, independent of the operator. In case of a hand-operated TSM, these points can t be realised in the same way. Therefore the influences have to be investigated. 2. DESCRIPTION The TSM consists of a T-formed base frame with a horizontal measuring axis. The test space may be varied from 10 mm to 1000 mm by moving the manual operated counter torque drive. The doublesided lever is supported by strain-controlled cross-hinges and wears on each end a strain-controlled mass-hinge to connect the masses. The masses are manually placed on one of the two scale pans, depending on the desired torque direction. The measured strain-sum of all hinges is scaled in unit N m and indicated with a precision amplifier. The sensitivity is about 0,525 mv/v per N m and the indicated resolution is +/- 0,00002 N m. This enables to measure a torque of 2 N m with a relative uncertainty of +/- 1 10-5 (k=2). With the hand-wheel on the counter torque drive the moment of the hinges will be adjusted to zero while loading. For this reason the influence of the lever support is negligible or the indicated very small value may be taken into account. 1

To avoid side forces and parasitic bending moments, the device under test is coupled to both the lever and the counter torque drive via elastic Rexnord-couples. For the lower torque range < 100 N m some smaller couples will be added and the big couples were blocked. Mass hinge Hand wheel for counter torque Scale pan Figure 1: Torque Standard Machine 3000 N m 3. LEVER ARM ADJUSTMENT The lever length has been calibrated at a German DKD-laboratory. The distance between the two mass axes is 1999.995 mm with an uncertainty of 7 µm. To achieve this exact value, an adjustment procedure was developed at GTM, particularly with regard to the fact, that GTM itself is not DKD-accredited in the physical quantity of length and has no hightech measuring devices. A similar procedure was described in [2]. The mass-hinge thickness is 1 mm. The distance between the inner concave surfaces is about 1999 mm, the distance of the outer concave surfaces is about 2001 mm. To measure the real distance, GTM used a profile with two µm-indicating callipers on its ends. With this profile the mass-hinges were touched on each four points and the measured values were compared with gauge blocks of the desired length. Figure 2 shows the touching of the mass-hinge in one point. After calculating the real lever length, one mass-hinge was moved by usage of gauge blocks to adjust the length exactly. The adjustment of the centre of the lever is checked with a balance test. This balance test can be performed by the operator at any time. So it is possible to have a monitoring of the lever ratio. 2

Figure 2: Adjustment of the lever length 4. AMBIENT UNCERTAINTY CONTRIBUTIONS The following parameters make a contribution to the measurement uncertainty: masses, mass density, local gravity, lever length, lever support, temperature, air density, airflow, shocks and manual operation of the TSM. Some of the contributions are well known because of calibration results, e.g. the mass uncertainty. Others are estimated by additional tests, e.g. the influence of the air conditioning. While commissioning it was detected, that the airflow in the lab is notable for the torque. To minimise the influence, a chamber was built around the TSM. The chamber works well on minimising the airflow, but some additional points must be considered. Before and during a calibration it is important to have the temperature variation in the chamber completely under control, as the lever length depends on this element. Therefore some parts of the chamber can be opened during the calibration, what generates no influence to the stability of the lever. To avoid an undue influence, the mechanical end stops of the lever have to be adapted to the stiffness of the torque axis. With the use of this end stops there is no influence of the operator if he pushes the lever on the end stop. The quality of the load change sequence is equal to automatic load change systems. Before starting a calibration, the sequence for changing the masses must be defined according to the range it will work. For example: while calibrating a 3000 N m capacity transducer, the operator should have the last steps (after 2400 N m) with masses of 100 N and not with masses of 200 N or 300 N, because the mass stack is higher and it is easier to operate with smaller masses on that height. 5. MEASUREMENT UNCERTAINTY To verify the estimated uncertainty, the Standard Machine was calibrated with a set of torque transfer standards. The nominal values are 50 N m, 100 N m, 500 N m, 1000 N m and 5000 N m. The transfer standards were calibrated according DIN 51309 and the best measurement capability was calculated considering the calibration results of German PTB according [3]. Figure 3 shows exemplary the calculated bmc of the clockwise torque in the range from 100 N m to 3000 N m. Both results, clockwise and anti-clockwise torque, are comparable in bmc and relative deviation. The relative uncertainty of the TSM is 1 10-4 (k=2) in the range from 100 N m to 3000 N m. 3

Figure 4 and 5 show exemplary the deviations in intercomparison tests with PTB. In the range from 2 N m to 2000 N m the best measurement capability is in good accordance with the specified relative uncertainty of 1 10-4 (k=2). Torque Standard Machine 3000 N m best measurement capability 1E-04 9E-05 8E-05 7E-05 6E-05 5E-05 4E-05 3E-05 2E-05 1E-05 0E+00 0 500 1000 1500 2000 2500 3000 Torque / N m 500 N m 1000 N m 3000 N m Figure 3: Calculated Best Measurement Capability DEVIATION INMETRO-PTB 0,010 0,008 0,006 dev up or in % dev down or in % 0,004 rel. dev. in % 0,002 0,000-2000 -1500-1000 -500 0 500 1000 1500 2000-0,002-0,004-0,006-0,008-0,010 torque in N m Figure 4: Relative deviation to PTB (range 2000 N m) DEVIATION INMETRO-PTB 0,010 0,008 0,006 dev up or in % dev down or in % rel. dev. in % 0,004 0,002 0,000 4-20 -15-10 -5 0 5 10 15 20-0,002-0,004

6. EXTENSION CONCEPT Figure 5: Relative deviation to PTB (range 20 N m) Base frame, lever and counter torque as well as the coupling parts are constructed for a maximum load of 5000 N m. If the torque situation in Brazil demands for higher torque than 3000 N m, an upgrade of the masses will be realised. First step in automating the standard machine may be a closed loop controller and an electric drive for the counter torque. At present the machine is operated with two persons, one for changing masses and the other to operate the counter torque. With an automated counter torque only one person can operate the machine. An important point in traceability of torque is the possibility of calibrating torque transfer wrenches on the new torque standard machine. Side forces of considerable amount are working at torque wrenches. If these side forces are acting out of the measurement axis, a disturbing torque results and deviations of the applied and the indicated torque are detectable. For that, an extension of the standard machine will be used to bear the side forces without effect to the applied torque. 7. CONCLUSION The improvement in the National primary torque standard with the reduction of the relative uncertainty to the limit of 1 10-4 (k=2), reflects an improvement in the whole traceability of the torque unit in Brazil. The implementation of the Torque Standard Machine in INMETRO s Torque Laboratory (LAFOR), provides a reference structure that allows to attend the metrological demands and needs of national industry, research centres and accredited laboratories, with higher accuracy and great reliability. A compatible budget and possibility of upgrade for a higher range and automated operation made the decision of INMETRO in favour of this manual operated 3000 N m TSM, with simple construction and high availability. REFERENCES [1] H. Gassmann, T. Allgeier, U. Kolwinski, A new design of torque standard machines, Proceedings of the 16 th IMEKO World Conference, Wien 2000 [2] D. Roeske, K. Adolf, D. Peschel, Lever optimization for Torque Standard Machines, Proceedings of the 16 th IMEKO World Conference, Wien 2000 [3] A. Sawla, Uncertainty scope of the force calibration machines, Proceedings of the 16 th IMEKO World Conference, Wien 2000 5