USE, CARE & STERILIZATION INSTRUCTIONS. Consumable Accessories Burs Saw Blades K-Wires Steinmann Pins Twist Drills Rasps Ortho Pins/Drills Guide Wires

Similar documents
USE, CARE & STERILIZATION INSTRUCTIONS. Consumable Accessories Burs Saw Blades K-Wires Steinmann Pins Twist Drills Rasps Ortho Pins/Drills Guide Wires

SURGICAL POWER & ACCESSORIES. Rx Only PM-X00-710/PM-X IR-12.13

SURGICAL POWER & ACCESSORIES. Rx Only IR-12.13

IR SURGICAL POWER & ACCESSORIES

PM-X IR SURGICAL POWER & ACCESSORIES

PM-X IR SURGICAL POWER & ACCESSORIES

Instructions for use TKR POSITIONERS TKR TKR. Contents

Consensus Orthopedics Inc. Bespoke ACO Polyamide Acetabular Cup Orientation (ACO) guides and ACO bone models

Series 5000 Electric Modular Instruments. Instructions for Use

Snowden-Pencer. Single-Piece Laparoscopic Devices

Stryker Visum 300 Ceiling-Mounted Exam Light Assembly, Operations, and Maintenance Manual

OPERATION AND MAINTENANCE INSTRUCTION MANUAL. AEU-707A & AEU-707AV2 Implant / Surgery Systems

Rechargeable laryngoscope handles Directions for use

OPERATION AND MAINTENANCE INSTRUCTION MANUAL. AEU-17B & AEU-17BV2 Implant / Surgery Systems

Model 2310 Series. Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

SurePower Battery Pack Guide

SurePower TM II Battery Pack Guide

HEINE NT4 MED V

Operating Instructions. International English. CCS Dust Extractor

COLIBRI II INSTRUCTIONS FOR USE. Universal and powerful battery system for a variety of applications

REF 6643 SmartDriver Duo. Instructions for Use

Rev C /08

Series 7000 REF 7505, REF 7506, REF 7507 Instructions for Use

System 8. power tools. A family of power tools built for the whole hospital

Slimax LED Curing Light System

Handheld LED Light Source (HLS) Operator s Manual

LED Headlight User s Manual

Electric Pen Drive. Compact system for a wide range of applications.

COOPER POWER. SERIES CMU Medium voltage power fuse installation instructions. Fusing Equipment MN132032EN

Quickdraw. Model 2400 Series. Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

INSTALLATION INSTRUCTIONS UNDERMOUNT SINKS

Wire/Drill Driver FEATURES AND BENEFITS

Blades with the MatrixPRO Driver may result in damage to the MatrixPRO Driver and/or screwdriver blades, and should

SYNTHES and DYONICS. Synthes Blades for Large-Bone Power Systems. Synthes Twist Drills + Taps C-4. Dyonics Blades for Large-Bone Power Systems C-2 C-6

DCell Suction. Model DM Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

INSTRUCTIONS FOR USE. Rx Only

AIM LED LIGHT (wall mount)

INSTALLATION INSTRUCTIONS TOP MOUNT SINKS

UV-AIRE AIR PURIFYING SYSTEM

PRM-2 PHASE SEQUENCE TESTER PRM-3 PHASE SEQUENCE & MOTOR ROTATION TESTER User s Manual

Instructions for use of S5 Endo Motor

830RLA M [EX-26] 835 M [SF-41] 835 M [SF-31] 836 M [SF-21] 837 M [SF-11] 837 M

INSTALLATION INSTRUCTIONS

Instructions for use. Quick couplings RQ-03, RQ-04, RQ-14, RQ-24, RQ-34

SECTION 6 5 SERVICE PROCEDURES AND SPECIFICATIONS. Body

USER S GUIDE HIGH INTENSITY ULTRASONIC PROCESSOR

Technical Data Monograph

Instructions for use GENTLEmini LUX 4500 BR - REF

HP190SL Series Slimline Radon Fans

INSTALLATION INSTRUCTIONS TOP MOUNT SINK CH366

Instructions for Use Reprocessed HARMONIC ACE Shears without Adaptive Tissue Technology. Reprocessed Device for Single Use.

OPERATION and MAINTENANCE INSTRUCTION MANUAL. AEU-10C/SS Power-Tip Systems

General warnings and cautions

Nimbus Flex Ambulatory Infusion Pump

Installation and Operation Manual

MatrixPRO Driver Instructions for Use

MIDAS REX ELECTRIC BONE MILL BM110 & BM120

INSTALLATION INSTRUCTIONS

FADE Series Wall Axial Fans

Model: UV-16R/24 READ THESE INSTRUCTIONS CAREFULLY AND COMPLETELY BEFORE PROCEEDING WITH THE INSTALLATION.

OPERATING MANUAL POWER DRIVE SYSTEM. Cat. No RevH

OPERATION and MAINTENANCE INSTRUCTION MANUAL. AEU-03SS Porta-Tip Systems

NEUTRALIZATION SYSTEM FOR OPA & GLUTARALDEHYDE MEDI-NEWT. User Manual. Endoscope reprocessing the the way way it it should be be

USER GUIDE LED LIGHTED MIRROR W/BUILT-IN MAGNIFIER. Model: , , ,

Assistant s Instrumentation T

Digital Apex Locator. ipex OPERATION MANUAL OM-E0285E

The instruments are powered by nitrogen or compressed air and require no lubrication. The instruments and hose are steam autoclavable.

INSTALLATION DATA MANUAL MODEL 148/248/348

AIM LED LIGHT (ceiling mount)

FISO Catheter Life Sciences series

Installation Instructions

Installation and Operation Manual

PAD-MOUNTED SWITCHGEAR INSPECTION & MAINTENANCE RECOMMENDATIONS TYPE PSI/II. 15kV 25kV. Qualified Persons

Copyright 2004 Alcatel. All rights reserved.

Installation, Operation, and Maintenance Manual

Endodontic Handpiece (GT) USER MANUAL

Carbon Fiber Overlay for Patient Positioning for GE Healthcare Global Tables 1700, 2000, and PET/CT. Installation and Setup Manual

VARIFUEL MAINTENANCE AND REPAIR INSTRUCTIONS

DENTAL LIGHT 300LED DENTAL LIGHT OPERATING INSTRUCTIONS IMPORTANT

BATTERY POWER LINE II

Operating Instruction INTRAmatic Prophy Handpiece 19 ES

Sentinel Enteral Feeding Pump. Model: S-1000-SI. Operating Manual

Model P772 Conductivity Meter. User Manual

Invacare. Battery Charger User Manual. Model: POC1 115, POC1 115 AZ, POC1 115 C, POC1 115 EU

Instructions for use GENTLEsilence 6500B - REF

Operator's Manual. Storage System. Ultrasound Probe Cabinet. Manufactured by:

Accessory ACCESSORY User Manual

UNIVERSAL REMOVABLE REAR CROSSBAR CONVERSION INSTRUCTIONS FOR HAULER RACK II & SERVICE BODY SERIES RACKS

SMOKE EVACUATION SYSTEM SAFETY PRECAUTIONS

TO AVOID DOING IRREPARABLE DAMAGE TO DRIVE CIRCUITRY NEVER APPLY AC POWER DIRECTLY TO LED LIGHTBOXES!

VARIFUEL MAINTENANCE AND REPAIR INSTRUCTIONS

SECTION 6 5 SERVICE PROCEDURES AND SPECIFICATIONS. Body

Cleaning & Sanitizing Materials Used in the Transport of Animals

21", 33", 45", & 61" Installation & Operation Manual

Introduction. Pressed Shaft Flowmeter. Flowmeter Service Kits. Disassembly. Idler Shaft Removal

TABLE OF CONTENTS GENERAL INFORMATION

WARNING Risk of property damage, injury, or death.

Installation Instructions for KPS and KPH Circuit Breakers and Molded Case Switches

1000 Emergency Care Stretcher 1500 PACU Stretcher OPERATIONS MANUAL. For Parts or Technical Assistance

QA4 Surgery Trolley System Manual Function Operating Instructions

Transcription:

USE, CARE & STERILIZATION INSTRUCTIONS Consumable Accessories Burs Saw Blades K-Wires Steinmann Pins Twist Drills Rasps Ortho Pins/Drills Guide Wires Rx Only

EXPLANATION OF SYMBOLS 2 g h i N B D L M d IK H Batch code. Catalog number. Consult instructions for use. Date of manufacture. Do not discard. Dispose of product or recycle in accordance with local laws and regulations. Do not re-sterilize. Do not reuse. Do not use if package is damaged. Manufacturer. Non-sterile. Sterilized using irradiation. Use by.

CAUTIONS & WARNINGS Do not re-use or reprocess used consumable accessories. Follow the instructions and warnings issued by the suppliers of any cleaning and disinfection agents and equipment used. Follow universal precautions and protective apparel when handling and cleaning contaminated consumable accessories. Do not clean or re-sterilize cannulated drills, high speed burs, laser welded or brazed two piece saw blades. Do not process consumable accessories in automated washer-sterilizer. General to all Consumable Accessories: The product is intended for use only by authorized persons fully trained in their safe and effective use. In case of unexpected product anomaly, it is recommended to have back-up consumable accessories to reduce any surgical delays and to prevent prolonged or additional anesthesia exposure. Consider potential patient reactions to contact with a particular metal to avoid possible allergic reaction. Inspect product package for any damage that may affect sterility prior to use. Inspect for debris on product in pouch prior to use. If contaminants can be seen do not use. Always verify sterile product is within its labeled expiration date to ensure sterility prior to use. Only use with appropriate power system. See applicable power system instructions for use. Do not use consumable accessory at speeds exceeding the handpiece manufacturer s specifications. Irrigation is recommended during use as necessary to prevent tissue or bone necrosis. Do not come into contact with other metallic objects such as retractors during use to prevent consumable accessory failure and/or patient injury. Failure to do so may result in excessive heat generation or metal shavings in the surgical site. Do not use excessive, lateral, twisting or bending forces to prevent consumable accessory failure such as bending or breakage. Exercise extra caution when used with alignment guides or cutting fixtures. 3

Dispose of product or recycle in accordance with local laws and regulations. All product should be stored in an environment that prevents premature degradation. The product should be protected from prolonged exposure to direct UV light, excessive heat, and humidity. Burs, Rasps and Twist Drills: Use proper bur guard with bur per the handpiece manufacturer s specifications. Do not use bur/twist drill to bend, pry or come into contact with metallic objects. Failure to do so can lead to breakage or metal shavings in the surgical site. Saw Blades and Rasps: Cutting in metal guides or accidental contact with metal retractors may damage the saw blade and may necessitate its replacement. Insert saw blade into alignment guide or cutting fixture prior to activating handpiece. Saw blades become hot from friction. Irrigation of saw blades is recommended during uset to prevent bone or tissue necrosis and is required when using an alignment guide or cutting fixture. K-Wires, Steinmann Pins and Guide Wires: Immobilize the patient s area of fixation to prevent excessive forces on the K-Wire or pin. Excessive forces on the K-Wire or pin may cause loosening or metal fatigue and subsequent device fracture. K-Wires or pins should be removed as soon as adequate healing has occurred. Magnetic Resonance Imaging (MRI) Safety Statement: The effects of the Magnetic Resonance environment on the K-Wires and Steinmann Pins have not been determined. These medical devices have not been tested for magnetic field/migration, artifacts or heating in the MRI environment. 4

CLEANING INSTRUCTIONS: Do not re-use or reprocess used consumable accessories. Unused consumable accessory removed from original non-sterile package: 1. Sterilize as directed on the table on page 8. Unused consumable accessory contaminated during surgery: 1. Remove consumable accessories from handpiece and or attachment. 2. Clean consumable accessories as soon as possible to avoid drying of contaminants. Disinfecting the consumable accessories may be needed to protect personnel handling the instrument before cleaning and sterilization. If cleaning must be delayed, immerse the consumable accessory in a neutral-ph cleaning solution to prevent drying of contaminants on the instrument. Burs, Saw Blades, K-Wires, Steinmann Pins and Guide Wires: 1. Rinse soiled consumable accessory thoroughly with a large amount of lukewarm (target lower end of the 72-110 F/22-43 C temperature range) running tap water for a minimum of one (1) minute. 2. Make an enzymatic cleaning solution such as STERIS Prolystica 2x Concentrate Enzymatic Presoak and Cleaner [1/8 ounce per one (1) gallon of warm tap water] (target lower end of the 91-109 F/33-43 C temperature range). 3. Soak the consumable accessory in the enzymatic cleaning solution for two (2) minutes, making sure the consumable accessory is completely immersed to prevent aerosolization of contaminants. 4. Manually clean the consumable accessory using a soft-bristled brush for one (1) minute to remove soil and debris. 5. Rinse soiled consumable accessory by immersing the consumable accessory in a sufficient volume (enough to cover the device) of lukewarm, tap water (target lower end of the 72-110 F/22-43 C temperature range) for one (1) minute. 6. Repeat Step 5 two (2) more times for a total of three (3) rinses. Visually inspect consumable accessories for the presence or absence of residual soil and/or cleaner. 5

6 7. Dry immediately with a clean lint-free soft cloth. 8. Inspect for damage and replace with new consumable accessory if damaged. 9. Sterilize as directed on the table on page 8.

STERILIZATION INSTRUCTIONS WARNINGS: The use of disinfecting solutions for an exterior instrument wipe will not sterilize the consumable accessory. Inspect sterilized consumable accessories. If discoloration or rust is visible do not use. CAUTIONS: Do not process in automated washer-sterilizer. Do not re-use or reprocess used consumable accessories. Do not use cold sterilization solutions as these contain oxidizing agents which may cause damage to the device. Do not sterilize while connected to handpiece and or attachment. Wrap the instruments per AAMI recognized guidelines. Follow the sterilizer manufacturer s written instructions for cycle parameters, load configuration and AAMI guidelines for steam sterilization. Do not clean or re-sterilize cannulated drills, high speed burs, laser welded or brazed two piece saw blades. 7

PARAMETERS FOR STERILIZATION Saw Blades/K-Wires/Steinmann Pins/Twist Drills/Guide Wires Steam Sterilization Type Minimum Temperature Minimum Exposure Time Minimum Dry Time Pre-Vacuum (Wrapped) 270 F (132 C) 4 minutes 30 minutes Gravity (Wrapped) 270 F (132 C) 15 minutes 30 minutes Burs Stainless Steel, Carbide and Diamond Steam Sterilization Type Minimum Temperature Minimum Exposure Time Minimum Dry Time Pre-Vacuum (Pouch) 273 F (134 C) 3 minutes 30 minutes Gravity (Pouch) 275 F (135 C) 10 minutes 30 minutes Note: These processes have been validated for a one-time sterilization as being capable of cleaning and sterilizing Burs/Saw Blades/K-Wires/ Steinmann Pins/Twist Drills/Guide Wires. 8

RETURN GOODS POLICY If product needs to be returned, the following applies: Consumable/Disposable Product: Any unused consumable/disposable product may be returned for credit within ninety (90) days of invoice date if it qualifies for resale (unopened and unmarked package). Products may be subject to a restocking fee. Credit will not be issued for products that do not qualify for resale. Capital Equipment: Any unused capital equipment may be returned for credit within thirty (30) days of invoice date if it qualifies for resale (unopened and unmarked package). Products may be subject to a restocking fee. Credit will not be issued for products that do not qualify for resale. Packaging and Shipping: Returned items should be packaged in original packaging. Credit will not be issued for items damaged in return shipment due to packaging inadequacy. Canadian customers contact your distributor for return policies. Note: It is unlawful to ship contaminated non-sterilized products. 9

Contact a Customer Service Representative at Brasseler USA Surgical Instrumentation at 800.535.6638 to obtain a Return Merchandise Authorization (RMA) number. Please include the following information with the returned product(s): Catalog number, serial number and lot number (if applicable) of device. Return Merchandise Authorization (RMA) number noted with the return. Original invoice number or copy of original invoice. Original invoice date. Customer name, address and account number. Itemized packing list. Brief statement describing reason for product return. Return to: Brasseler USA Surgical Instrumentation 4837 McGrath Street Ventura, CA 93003 Canadian customers contact your distributor for return policies. 10

PRODUCT DISPOSAL Dispose of product or recycle in accordance with local laws and regulations. 11

M Brasseler U.S.A. Medical, LLC One Brasseler Boulevard Savannah, GA 31419 800.535.6638 912.921.7578 (fax) BrasselerUSAMedical.com Brasseler U.S.A. Medical, LLC, has implemented a quality management system that is certified under ISO 13485:2003. STERIS and Prolystica are trademarks of STERIS Corporation. Brasseler U.S.A. Medical, LLC, is not affiliated with the above and makes no claim to copyrights or trademarks which are the property of this company. Brasseler USA is a registered trademark of Peter Brasseler Holdings, LLC, or its affiliates. All other trademarks are trademarks of their respective owners or holders. Colors, specifications and product availability subject to change. Brasseler USA Surgical Instrumentation products are sold by Brasseler U.S.A. Medical, LLC, One Brasseler Boulevard, Savannah, Georgia 31419, United States. 2015 Brasseler U.S.A. Medical, LLC All rights reserved Printed in the U.S.A. BM-3880-REV3-10.23-2M-DXM