Isolation Transformer for Medical Applications

Similar documents
INSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS.

INSTALLATION AND OPERATING INSTRUCTIONS OF THE INTERNATIONAL ISOBOX SERIES ISOLATION TRANSFORMERS.

OPERATING AND MAINTENANCE MANUAL. Primary Current Injection Test Set. 750ADM-H mk2

BATTERY CHARGER-STARTER

362 Clamp Meter. Users Manual

PRM-4. Phase Sequence and Motor Rotation Tester. Users Manual. For detailed specifications and ordering info go to

Contents. Introduction Inspection Notes on Safety Precautions. Chapter 1 Overview Product Overview Names of Parts 2

Phoenix Inverter

EPS/ELA-Series User Manual EPS/ELA 250W

EC200 ELECTRONIC CONTROL SYSTEM

Declaration of Conformity

LDG6000SA DIESEL GENERATOR OWNERS MANUAL

Chargestar P24 / P32 Battery Charger Startmaster P300 Starter / Charger

ISO-TECH IPS1810H and IPS1603D DC power supply ISO-TECH IPS 1810H et IPS 1603D Alimentation électrique c.c.

Dr. Mach Lamps and Engineering

ACROBAT SWING STAND MODEL

& HIGH CURRENT DC POWER SUPPLIES INSTRUCTION MANUAL

5.5KVA GENERATOR MODEL NO: PG6500DVES OPERATION & MAINTENANCE INSTRUCTIONS PART NO: LS0616

Compact System NRGS 11-2 NRGS Original Installation Instructions English

MAKING MODERN LIVING POSSIBLE. Quick Setup VLT FCM 300 Series. Phone: Fax: Web: -

Go to to register your product and find more information.

3KVA DUAL VOLTAGE GENERATOR MODEL NO: PG3800DV

1200W INVERTER GENERATOR

DC POWER SUPPLY ALIMENTATION C.C.

Appendix: Safety and application notes for... 15

Simply reliable. ABB Single-phase Transformers. When transformation means safety and control.

GENERATOR MODEL NO: FG2500 OPERATION & MAINTENANCE INSTRUCTIONS PART NO: LS0114

Table of Contents. Classifications.. 5. Installation.,12. Focusing..14. Operating Instructions..15. Charts Description Maintenance...

Smart Battery Charger

Installation manual wall-mounted distributor

FREQUENCY INVERTER INSTALLATION & USER GUIDE

Switching DC Power Supply

User s Manual Power Supply IM E. 5th Edition

1100W PORTABLE GENERATOR

Whad 0.8, 1, 1.5 kva. Manuel d installation Installation manual. Part. LE05733AA-07/12-01 GF

PY-11 Voltage & Continuity Tester

Dycon D2430 EN54-4 Fire Alarm Power Supply Series

WARNING: Read these instructions before using the machine GENERATOR MODEL NO: IG3500F PART NO: OPERATION & MAINTENANCE INSTRUCTIONS

PHASE CONVERTERS OPERATING & MAINTENANCE INSTRUCTIONS. MODEL NO: PC40 and PC60. PART Nos:

PURE SINE WAVE DC TO AC POWER INVERTER

TIC 300 PRO. Users Manual Mode d emploi Bedienungshandbuch Manual d uso Manual de uso. Non Contact AC Voltage Detector. High Energy Tic Tracer

JANDS DD-8. Technical Manual. Version 1.0

Dycon D1532SM. EN50131/PD6662 Grade 3, 12V 2A Power Supply. Technical Description Installation and Operating Manual DYCON POWER SOLUTIONS LTD

233 True-rms Remote Display Digital Multimeter Safety Information

SFset with VIP 400/410 built-in control unit Medium Voltage Circuit Breaker

PRM-2 PHASE SEQUENCE TESTER PRM-3 PHASE SEQUENCE & MOTOR ROTATION TESTER User s Manual

MAKING MODERN LIVING POSSIBLE. UniLynx Indoor Installation Manual. ULX 1800i ULX 3000i ULX 3600i ULX 5400i SOLAR INVERTERS

USE AND INSTALLATION HANDBOOK

Mod: KLD6-12/35XLAS-N

Rescue Pac. Please read and fully understand the instructions in this manual before operation. Keep this manual safe for future reference

HEINE NT4 MED V

User s Manual Power Supply. IM E 3rd Edition. Yokogawa Electric Corporation

HIB Enclosed Inverter User Guide

Operating Instructions. International English. CCS Dust Extractor

Pro Booster 802Li. Please read and fully understand the instructions in this manual before operation. Keep this manual safe for future reference.

1 This instrument must only be used by a competent and trained person and operated in strict accordance with the instructions.

NT-500 Technical Support Guide. For use by qualified company approved Technicians

Installation & User Instructions

DKG-114 MANUAL AND REMOTE START UNIT

Installation manual portable distributors

Uninterruptible power supply Supply MEg101.4

WELDING INVERTER. PEGAS 160 E Smart PEGAS 200 E Smart OPERATING MANUAL. ALFA IN a.s. PEGAS E Smart Manual EN 04

Model 1672, 1673 Triple Output Power Supply

BT508/BT510/BT520/BT521 Battery Analyzer Safety Sheet

User Manual Rittal PMC UPS 6kVA

Domae. The right solution for distribution boards

1587 FC/1587/1577. Insulation Multimeter. Safety Information. Warnings

DC Master 24/ A

AC50A. Leakage Clamp Meter. User Manual. For detailed specifications and ordering info go to

Triple Output Power Supply

Instruction Manual LAMI CORPORATION INC.

RS-3 PRO RS-1007 PRO. CAT IV Analog Clamp meter Series. Users Manual. For detailed specifications and ordering info go to

STAND MODEL FOR MACH M2

superstart starter/chargers

These operating instructions apply for: NCX 380 NCZ 300 NCX 480 NCZ 370 NCX 580 L NCZ 480 NCX 660 K NCZ 560 NCZ 660 NCZ 800

GENERAL INFORMATION. INSTRUCTIONS FOR 500A Lithium Jump Starter

90W DC Power Supply. User Manual. 99 Washington Street Melrose, MA Phone Toll Free

BATTERY STARTER/CHARGER MODEL NO: BC125, BC190

1-50 KVA SINGLE-PHASE SERVO REGULATOR INSTRUCTIONS FOR USE SERVO SINGLE-PHASE VOLTAGE REGULATOR

1.Safe testing IMPORTANT:

Range 16A to 63A. Sensitivity 30mA, 100mA, 300mA, 500mA. Execution Double Pole, Four Pole. Specification IEC / IS : 2000

Equipment model: auxiliary control. Pack Cat.Nos 10/16 A V± P + T shuttered module module

Installation and Operating Instructions. Solar System Controller ISC3030

ELECTRICAL INSTALLER EXAMINATION 18 November 2017 QUESTION AND ANSWER BOOKLET

Battery Charger JCB- SCH20LI.2

Battery Charger JCB-FCH12Li

PSU1 Power Supply Unit

OPERATION MANUAL MODELS TWE-250 TWE-321 TWE-375 TRU WELD EQUIPMENT COMPANY 6400 N. HONEYTOWN ROAD SMITHVILLE, OHIO (330)

Instructions for use. Quick couplings RQ-03, RQ-04, RQ-14, RQ-24, RQ-34

SYMBOL LEGEND DANGER WARNING NOTE THIS INDICATES DANGER TO THE LIFE AND HEALTH OF THE USER IS APPROPRIATE PRECAUTIONS ARE NOT TAKEN

Assembly instructions PRORUNNER mk1 2

Differential pressure sensor

Differential Pressure Transmitter

GENERATOR MODEL NO: FG3005 OPERATION & MAINTENANCE INSTRUCTIONS PART NO: LS0413

Calibrator Capabilities Function Range Resolution Measure V dc 28 V 1 mv Measure ma dc Source ma dc

GENERATOR MODEL NO: FG3000 OPERATION & MAINTENANCE INSTRUCTIONS PART NO: LS0609

Contents. Overview. Safety Instructions. Before Installation. Installation. Operation. Maintenance. Replacement Parts. System Specifications 8 01

Operator's Manual. Storage System. Ultrasound Probe Cabinet. Manufactured by:

Installation, Operation and Maintenance Manual. EVC Controller (from February 2007)

KEWTECH. KT56 digital multi function tester. Instruction manual

Transcription:

User Instructions Isolation Transformer for Medical Applications Type: Isolation Transformers REOMED 300VA; REOMED 600VA; REOMED 1000VA REO (UK) Limited Units 2 4 Callow Hill Road, Craven Arms Business Park Craven Arms, Shropshire SY7 8NT Tel. 01588 673411 Fax 01588 672718 www.reo.co.uk email:main@reo.co.uk

Contents 1.0 Safety instructions 1.1 Technical Description 1.2 Intended use 1.3 Operating Instructions 1.4 1.5 1.6 2.0 3.0 Transport/ Storage/ Disposal Construction and Models Symbols Technical Data (Appendix) Guidelines and Manufacturer s Declaration (Appendix) 1.0 Safety Instructions The following instructions are provided for the personal safety of the operating personnel and the protection of the described products and also any attached devices.! Warning! Hazardous Voltages. Failure to observe these instructions can cause death, serious injury or damage to equipment. Isolate mains supply before installing, uninstalling, fuse replacement or system changes. Observe accident prevention and safety regulations that apply specially to the application. Before start-up check whether the rated voltage of the equipment agrees with the local mains voltage. Use of the equipment in highly inflammable or combustible environments is prohibited. The penetration of water into the equipment can cause fire and electrical shock hazards to occur. No vases or water filled containers are to be placed on the equipment. Select a safe location for the equipment. Before start-up of the equipment the operating instructione must be read and applied correctly. WARNiNG: UNDER NO CIRCUMSTANCES MUST OTHER SUPPLY VOLTAGES AND FREQUENCIES BE CONNECTED AND APPLIED TO THE EQUIPMENT. REO INDUCTIVE COMPONENTS AG 2

1.1 Technical Description Toroidal core transformers for the medical industry with multiple output sockets (to suit local requirements), in a sturdy metal housing with a starting current limiter and primary micro fuses (L+N). Product features: Sturdy metal housing Leakage current < 500 µa Short-circuit and overload protected Built-in thermal sensor Fused primary side ON/OFF Switch Switch-on current limiter either electronic or MCB Optional over voltage protection and mains filter 2m mains cable with plug and cold connector Proof of conformance: Product Class I Cold connection secondary output sockets (IEC320) Conformance with EN 60 601 (Medical Safety Standard) Conformance and test report according to Regulations for Medical Products 93/42/EWG Technical Data (Appendix) Guidelines and Manufacturer s Declaration (Appendix) 1.2 Intended Use For use with medical electrical devices and non-medical electrical devices in patient environments and/or areas for medical use, to provide safe galvanic separation between primary and secondary (double and strengthened isolation), a very low leakage current and the required strict air and leakage clearances, by using a special, toroidal-core insulating transformer, according to the EEC guidelines for medical products 93/42/EWG. The equipment fully complies with the specified electrical limit values for the employment in patient environment and/or in areas for medical use. The equipment also fully complies with the strict requirements for a very small leakage, as well as a reduction of the sum of the leakage currents for several single devices in a power supply circuit, in a patient environment and/or in areas for medical use. To provide a method of connection of combinations of medical electrical devices and not-medical electrical devices in the patient environment and/or in areas for medical use the equipment the medically electrical devices and/or the not-medical electrical devices, according to the EEC guideline for medicine products 93/42 EEC (e.g.: electrical data processing accessories such as PC, screen, printer, plotter, interfacing with medically electrical devices as well as video devices, diagnostic and monitoring cameras, diagnosis -, measuring and testing sets and their combinations as well as many other devices). 1.3 Operating Instructions: Please check whether a safe mains connection has been made and that the whole is system is switched (OFF), before further devices are plugged into the output sockets of the equipment. Please double check the earthing connections for the mains input socket of the REOMED unit. Make sure that all devices are in the switched (OFF) condition, before these are connected to the REOMED unit. Please check the output voltage of the REOMED and confirm that the devices will operate correctly at this voltage. Additionally the isolating transformer has a standard earth connection (POAG plug for equi-potential earth connection according to DIN 42801), which can be used for connection to appropriate equi-potential earthing equipment terminals. The insulating transformer can then be connected to other medical instruments, providing these comply with the Medical Safety Standard EN 60601-1. REO INDUCTIVE COMPONENTS AG 3

Next connect the required devices to the REOMED (output sockets). When the REOMED is switched on there will be power at the output sockets. The devices connected with the REOMED can now be switched on (please observe all regulations and requirements for the input and connection cables) for the inlets and connecting cables), check also that all input and output cables are fully restrained and securely terminated.. The total power of the REOMED corresponds to the sum of the individual devices connected to the output sockets and are used simultaneously. It is most essential to check that the total output power does not exceed the total rated power, which is stated on the rating label of the equipment. The relevant rating data for the series of REOMED transformers can be found in a table below. When overloaded the REOMED isolation transformer switches off automatically. overloaded switches off these. When the temperature of the isolation transformer cools to 55 C cooled down it can resume operation. Note: Only for devices with NTC s as starting current limiters After switching the equipment off, a "waiting period" of approximately 1 minute is recommended before switching on again, otherwise the fuses may blow or the circuit breaker may trip. Fuse changing: Switch off the equipment and remove the mains cable! Remove the fuse-holder cap using a fine screwdriver. The faulty fuse can now be removed and replaced. Ensure that the fuse is correctly rated depending on the voltage setting. Use only Type TT fuses. Mains voltage setting (for switchable versions only): Switch off the equipment and remove the mains cable! For 300VA units the voltage can be changed in the base, between V/V with a screwdriver. For REOMED 600VA and 1000VA use the mains selector above the mains switch. After removal of the fuse holder (see fuse changing) the voltage selector can be removed from its holder and replaced after the required voltage has been selected (readable externally). Attention: Check the fuse values after changing switch settings! Cleaning: Before cleaning remove the mains cable. Use a damp cloth to clean the equipment. Do not use a liquid cleaner or sprays. Water and Liquids Do not use the equipment in the vicinity of water e.g. near sinks, wash basins, baths, or in a damp cellar. Routine inspection: The equipment is maintenance free Technical safety inspection: At least every 24 months a visual inspection should be made to ensure that there is no damage to the mains cables or housing and also the earth connections and lerakage current should be measured and recorded. REO INDUCTIVE COMPONENTS AG 4

1.4 Transportation, Storage and Disposal The following conditions apply for transportation and storage times of up to 15 weeks Temperature: - 10 C... + 50 C Relative Humidity: 10%... 90% Air Pressure: 500hPa...1060hPa and for operating conditions Temperature: + 10 C... + 40 C Relative Humidity: 30%...75% Air pressure: 700hPa...1060hPa Store in original packing in a closed or attic room (damp proof!) The equipment must not be subjected to strong vibration. Disposal Packaging is made from recyclable materials. Metal equipment parts should be disposed of through a metal recycling centre Plastics, electrical components and printed circuit boards should be disposed as electronic waste. Disposal must comply with local nation current regulations. Appropriate disposal enterprises are to be consulted. Please check with your town or general waste collection centres regarding local regulations. 1.5 Model Versions: Model Item Number + Option Power Supply Voltage Output Voltage Output Current Fuse Rated Current (VA) (V) (V) (A) (A) REOMED 300 65B5166Axx 300 / 1,30 TT3,15/TT1,60 REOMED 600 65B5167Axx 600 / 2,60 TT6,30/TT3,15 REOMED 1000 65B4132Axx 1000 / 4,35 TT10,0/TT5,0 REOMED 300 REOMED 600 REOMED 1000 65B5008Axx 65B5064Axx 65B5088Axx 300 600 1000 1,30 2,60 4,35 TT1,60 TT3,15 TT5,00 REOMED 300 REOMED 600 REOMED 1000 65B5161Axx 65B5162Axx 65B5163Axx 300 600 1000 1,30 2,60 4,35 TT3,15 TT6,30 TT10,0 REOMED 300 REOMED 600 REOMED 1000 65B5252Axx 65B5253Axx 65B5254Axx 300 600 1000 2,60 5,20 8,70 TT3,15 TT6,30 TT10,0 xx = Option: -10 NTC -01 Transient protection -20 NTC + Timer Relay -02 EMC Filter -30 NTC + Fuse -03 Transient Protection + EMC Filter -40 NTC + Timer Relay + Fuse -50 Electronic switch-on current limiter Example - Option: 65B5166A23 NTC + Timer Relay + Transient protection + EMC Filter REO INDUCTIVE COMPONENTS AG 5

Switch-on current limiter option The REOMED units can be fitted with either an NTC (Negative Temperature Coefficient Resistor) or an electronic switch-on current limiter. With the NTC It is important to allow a pause between switching the equipment off and back on again. Equipment input cable and device output cables Input: The mains supply cable supplied with the equipment is provided for connection to the mains supply network (mains socket). Output: All equipment connection cables must meet the relevant standards and regulations of the country in which the REOMED is used e.g. UL/CSA/VDE/SEMKO/CHAR etc. In the USA and Canada a special cable is necessary for use in hospitals. All cables must have a protective earth connection (3-pin!). Safety accessories: The REOMED series has been tested under strict safety conditions. The fuses in the primary circuit protect the REOMED against overload and short-circuit. If fuses other than those specified used, then this could expose patients or personnel to danger if they are connected to the equipment and also connected devices could be seriously damaged. The fuses must be UL/CSA approved for use in the North American market and VDE/CEn approved for the European market. Please use only super+slowacting micro fuses (TT marking) for the REOMED. The original equipment is supplied with spare Class G fuses. DO NOT USE FUSES WITH HIGHER CURRENT RATINGS! REO INDUCTIVE COMPONENTS AG 6

1.6 Symbols: O = Mains ON I = Mains OFF Equi-potential Earth Connection AC Current F Fuse T Delay Fuse Not domestic rubbish Conformity Sign! Attention: Read accompanying documents RQS REO-Quality Mark (Serial number) REO INDUCTIVE COMPONENTS AG 7