Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis

Similar documents
High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

Application Note. Abstract. Authors. Environmental Analysis

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

White Paper. Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Introduction. Background Information

Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System

Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

Using the PSD for Backflushing on the Agilent 8890 GC System

Fast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis

High-Temperature Simulated Distillation System Based on the 6890N GC Application

Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE

Analysis of Petroleum Fractions by ASTM D2887

ASTM D 6730 Detailed Hydrocarbon Analysis

Method Development for Capillary GC Systems. Slide 1

Restek Corporation 110 Benner Circle, Bellefonte, Pa Barry L. Burger, Neil Johansen, Valerie Gamble, Donald Rhoades

Simulated Distillation Analyzers, Software, Standards, Consumables, Training

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector

Refinery Gas. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Analysis of Fatty Acid Methyl Esters (FAMES), and Examination of Biodiesel Samples for these Components, by GCxGC-FID

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column

Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584

AppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS

The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector

Application. Gas Chromatography June 1995

Same Column and Gas Type Try Different/Faster Velocities. Same Column, Switch He to H2 Carrier Then Try Faster Velocities

Setting up SilFlow for BackFlush in your GC

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC

REFINED RESTEK. Detailed Hydrocarbon Analysis. innovative petrochemical solutions. Featuring Rtx -DHA Columns

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

Detailed Hydrocarbon Analysis Featuring Rtx -1 PONA Columns

Technical Procedure for Gas Chromatography (GC-FID)

Application Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER

Group-Type Analysis (PiPNA) in Diesel and Jet Fuel by Flow Modulated GCxGC FID.

Live Crude Oil Volatility

The next level of True Blue Performance

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS

SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

Live Crude Oil Volatility

ANALYSIS OF GASOLINE RANGE HYDROCARBONS ON BP1-PONA

Antek Gas Chromatography Products: A Guided Tour Through Industry

Application Note. Author. Introduction. Energy and Fuels

Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)

CONSTITUTION OF COAL TAR PITCH AND ITS EFFECT ON PROPERTIES

Application Note. Authors. Abstract. Energy & Chemicals

GC Method Compliance and Large Valve Oven Application

High Temperature Simulated Distillation. Terry Thompson. Intertek Testing Services St. Rose, Louisiana

Agilent Multimode Inlet for Gas Chromatography

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

DANI Transformer Oil Gas Analyzer

Journal of KONES Powertrain and Transport, Vol. 21, No ISSN: e-issn: ICID: DOI: /

Optimized Method for Analysis of Commercial and Prepared Biodiesel using UltraPerformance Convergence Chromatography (UPC 2 )

High-Performing Restek PAL SPME Fibers

Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC

Sulfur Detection at ppb Levels in Light Hydrocarbon Streams

Agilent Multimode Inlet

GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method

Analysis of Glycerin and Glycerides in Biodiesel (B100) Using ASTM D6584 and EN Application. Author. Abstract. Introduction

Agilent and ASTM. Update on Recent Activities. Page 1

Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS)

TECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure

Rapid Qualitative GC-TOFMS Analysis of a Petroleum Refinery Reformate Standard

Analytical and Testing Instruments for the HPI Market Shimadzu s Total Solutions for the Hydrocarbon Processing Industry

Ultrafast GC Performance in the Real World: Multi Lab Studies for Repeatability & Reproducibility

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

Practical Steps in GC Troubleshooting

MET-Biodiesel Capillary GC Columns

Correlating TBP to Simulated Distillations. COQA Long Beach, CA

Operation and Applications of Differential Flow Modulation

The Advantage of Real Atmospheric Distillation using D7345 Test Method. Presented by Jonathan Cole, PAC

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585

Stability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas

S-PRO 3200 GC System for Sulfur Analysis

Liquefied Gas Injector. Solution for the Sampling and Analysis of Liquefied Gases

Application Note. Author. Abstract. Energy & Chemicals - Petrochemicals. Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany

Characterization of Gulf Region Tar Balls Following the Deepwater Horizon Oil Spill

Application of High-Temperature Simulated Distillation to the Residuum Oil Supercritical Extraction Process in Petroleum Refining

Fausto Munari e Andrea Cadoppi ThermoFisher - Italy

On-Line NIR Analysis of Crude Distillation Unit

A Comparison of Boat Introduction and Direct Injection using the Thermo Scientific ipro 5000 Series Combustion Analyzer

White Paper.

Performing ASTM 6584 free and total glycerin in BioDiesel using an SRI Gas Chromatograph and PeakSimple software

STI OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Crude Evaluation Best Practices

Dual Plasma Sulfur and Nitrogen Chemiluminescence Detectors. Unsurpassed Stability, Selectivity, and Sensitivity for your GC Analysis

Agilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass. James D. McCurry, Ph.D. Senior Scientist

Model 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors

Set Your Sights on Superior Performance

Mineral Turpentine Adulterant in Lubricating Oil

Optimized Supercritical Fluid Chromatographic Instrumentation for the Analysis of Petroleum Fractions

Large Volume Injection of Polycyclic Aromatic Hydrocarbons

ASTM D for Denatured Fuel Ethanol Automating Calculations and Reports with Empower 2 Software

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

Agilent 6850 Series II Network GC System G2630A Performance Specifications

Transcription:

Introduction Petroleum & Petrochemical Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis By Katarina Oden, Barry Burger, and Amanda Rigdon Crude oil consists of thousands of different hydrocarbons that span a very wide volatility range. In order to efficiently produce quality finished products, refineries depend on simulated distillation (SimDist) analyses to characterize the composition of their crude feedstock prior to processing. SimDist is a gas chromatography (GC) technique that allows fast, reproducible, and easily automated determination of boiling point distribution of crude oil samples. Until recently, helium has been the only accepted gas for ASTM SimDist analysis. The ASTM SimDist committee now has approved the use of alternative carrier gases nitrogen and hydrogen for SimDist analyses. This article will explore alternative carrier gas options for SimDist analyses and discuss their various benefits and drawbacks. Nitrogen or Hydrogen: Both are Sustainable Alternatives Alternative carrier gases for SimDist analysis are explored in this article in the context of ASTM Method D7213. This method was developed to cover boiling range distribution of petroleum products and fractions such as medium and heavy naphtha, kerosene, jet fuels, and diesel, which have initial boiling points greater than 1 C and final boiling points less than 615 C, ranging from C5- C6. ASTM D7213 specifies that in order to successfully conduct the analysis, resolution between C5 and C52 must be greater than 1 and not more than 1. Additionally, peak symmetry or skewness must be greater than.5 and must not exceed 2. A traditional ASTM D7213 analysis using helium carrier gas is shown in Figure 1. Figure 1: All ASTM Method D7213 requirements are met using an MXT -1HT SimDist Column under typical conditions using helium as the carrier gas. pa 4 2 C4 C5 skew =.9 R (C5/C52) = 3.1 38 pa 38 C Column MXT -1HT SimDist, 5 m,.53 mm ID,.88 µm (cat.# 7131) Retention time standard for D7213 (C5-), Separation Systems, Inc. Liner: Restek Premium 1.7 mm ID on-column inlet liner (cat.# 2343.1) Inlet Temp. Program: 9 C to 38 C at 2 C/min Temp.: 4 C to 38 C at 1 C/min (hold 1 min) Carrier Gas He, constant flow Flow Rate: 12 ml/min Detector FID @ 39 C Make-up Gas Flow Rate: 18 ml/min Make-up Gas Type: N2 Hydrogen flow: 43 ml/min 4 ml/min Instrument Agilent 789B GC 1 2 3 4 GC_P85 Pure Chromatography

Figure 1 shows that Restek s 5 m x.53 mm x.88 µm MXT -1HT SimDist column (cat.# 7131) has excellent sample capacity, ensuring peaks do not exceed the method skewness requirements. Extremely low column bleed results in accurate and reproducible retention times for the later eluting hydrocarbons. Figure 2: Van Deemter curves for hydrogen, helium, and nitrogen. Optimum efficiency is obtained at the linear velocity corresponding to the minimum point on each curve. Van Deemter Plot N 2 Nitrogen Nitrogen is a very good alternative carrier gas because it is readily available in cylinders or from nitrogen generators and has no explosive properties. Additionally, nitrogen is the most efficient carrier gas in comparison to helium and hydrogen when operated at their respective optimum velocities. Efficiency directly relates to peak width more efficient separations will have narrower peaks with higher resolution. Figure 2 shows van Deemter curves for hydrogen, nitrogen, and helium. The y-axis measures height equivalent to a theoretical HETP 1 2 3 4 5 6 7 Average Linear Velocity (cm/sec) plate (HETP) and higher efficiency occurs at lower HETP values. Therefore, efficiency is maximized at the minimum point of each van Deemter curve. The x-axis measures carrier gas linear velocity, thus, Figure 2 shows that the separation efficiency with each carrier gas differs depending on linear velocity. Nitrogen has the lowest minimum van Deemter curve value; however, this minimum occurs at a very slow linear velocity. This means that even though very good separations can be achieved using nitrogen carrier gas, the analysis time will be longer compared to using helium or hydrogen when operated at their respective optimum velocities. Increasing the linear velocity in order to decrease the analysis time when using nitrogen will result in a rapid loss of column efficiency. The MXT -1HT SimDist column was tested for ASTM D7213 performance using nitrogen carrier gas. If nitrogen were employed at its optimum linear velocity, the resulting method would be much longer than the original method. Instead of translating methods to yield maximum efficiency, many nitrogen users simply replace helium with nitrogen, keeping all other method parameters the same (e.g., carrier gas linear velocity and oven program). Even though some efficiency is lost when employing nitrogen as a carrier gas without flow optimization, Restek s MXT -1HT SimDist column offers enough efficiency to easily meet ASTM requirements for resolution between C5 and C52. Due to the loss of efficiency, peak heights are lower, meaning some response may be lost, but the low-bleed performance of this column allows for detection of broader peaks at lower levels, and baseline resolution of high molecular weight hydrocarbons is possible (Figure 3). Figure 3: Even with nitrogen as the carrier gas, MXT -1HT SimDist columns provide adequate resolution to meet ASTM Method D7213 requirements. C4 C5 skew =.97 R (C5/C52 = 1.8) H e Points of Greatest Efficiency Column MXT -1HT SimDist, 5 m,.53 mm ID,.88 µm (cat.# 7131) Liner: Restek Premium 1.7 mm ID on-column inlet liner (cat.# 2343.1) Inlet Temp. Program: 9 C to 38 C at 2 C/min Temp.: 4 C to 38 C at 1 C/min (hold 1 min) Carrier Gas N2, constant flow Flow Rate: 11.9 ml/min Detector FID @ 39 C Make-up Gas Flow Rate: 18 ml/min Make-up Gas Type: N2 Hydrogen flow: 43 ml/min 4 ml/min Instrument Agilent 789B GC H 2 1 2 3 4 GC_P86 2

Hydrogen Hydrogen is comparable in price to nitrogen, and it has the additional advantage of being easy to generate. In contrast to nitrogen, hydrogen has a faster optimum linear velocity than helium (Figure 2). This means that with proper translation, methods can be shortened by approximately 25% when switching to hydrogen carrier gas. Restek s EZGC method translator (http:///ezgc-mtfc ) was used in custom mode to translate ASTM D7213 from helium to hydrogen carrier gas. Since the optimum linear velocity of hydrogen is faster than that of helium, changing the linear velocity to match the optimum for hydrogen will result in a faster analysis. However, the oven ramp rates must be changed proportionally in order to preserve the chromatography of the original method. Figure 4 shows the chromatogram generated using hydrogen carrier gas at 2 ml/min with a translated oven program. Elution time of was shortened by 15 minutes (compared to using helium in Figure 1) while maintaining very good resolution between C5 and C52. Figure 4: Overlay of chromatograms of a qualitative mixture of normal paraffins (C5-) and a reference gas oil comparing their boiling point distributions. 21 min elution of Red = Reference gas oil #1 Black = Retention time standard for D7213 C4 C5 skew =.94 R (C5/C52 = 2.7) 1 2 GC_P87 Column MXT -1HT SimDist, 5 m,.53 mm ID,.88 µm (cat.# 7131) Reference gas oil #1 Liner: Restek Premium 1.7 mm ID on-column inlet liner (cat.# 2343.1) Inlet Temp. Program: 9 C to 38 C at 2 C/min Temp.: 4 C to 38 C at 16.9 C/min (hold 5 min) Carrier Gas H2, constant flow Flow Rate: 2 ml/min Detector FID @ 39 C Make-up Gas Flow Rate: 25 ml/min Hydrogen flow: 25 ml/min 4 ml/min Instrument Agilent 789B GC 3

Although hydrogen is affordable and allows the opportunity to significantly shorten analysis time, it is an explosive gas, which causes concern for some laboratories. Additionally, hydrogen is a reducing gas, meaning it is reactive whereas nitrogen and helium are nonreactive. In general, concerns regarding the reactivity of hydrogen revolve around higher temperature analysis of active analytes, but poorly coated metal columns also can be prone to reduction reactions from hydrogen at high temperatures. The MXT -1HT SimDist column shows no decrease in performance over many temperature cycles and CS2 injections using hydrogen carrier gas, indicating a rugged, high-quality column coating (Figure 5, Table I). Figure 5: Excellent repeatability for C5- hydrocarbons is obtained using the MXT -1HT SimDist column, ensuring consistent performance from injection to injection. = 1 = 45 = 84 C4 C2 Bleed 22 pa 1 2 GC_P88 Column MXT -1HT SimDist, 5 m,.53 mm ID,.88 µm (cat.# 7131) PTV on column Liner: Restek Premium 1.7 mm ID on-column inlet liner (cat.# 2343.1) Inlet Temp. Program: 9 C to 38 C at 2 C/min Temp.: 4 C to 38 C at 16.9 C/min (hold 5 min) Carrier Gas H2, constant flow Flow Rate: 2 ml/min Detector FID @ 39 C Make-up Gas Flow Rate: 25 ml/min Make-up Gas Type: N2 Hydrogen flow: 25 ml/min 4 ml/min Instrument Agilent 789B GC 4

Table I: Retention time repeatability for selected carbon numbers taken from the GC chromatograms of the C5 to reference standard analyzed in Figure 5. Carbon No. Run 1 (min) Run 45 (min) Run 84 (min) % RSD (n=3) C8.328.322.322 1.69 C1 1.23 1.13 1.11.633 C14 3.668 3.658 3.654.197 C2 7.182 7.173 7.168.1 C28 1.654 1.646 1.641.62 C36 13.33 13.295 13.29.49 C4 14.419 14.41 14.411.34 C5 16.776 16.772 16.767.27 C6 18.696 18.691 18.685.3 C66 19.691 19.688 19.684.18 Conclusion Carrier gas changes for SimDist methods, such as ASTM Method D7213, can be easily implemented in your daily process. MXT -1HT SimDist columns offer excellent efficiency so adequate separations can be achieved even when operating at faster-thanoptimal linear velocities. In addition, these columns are robust enough to withstand high temperatures and extreme conditions and still provide the low bleed levels needed for accurate boiling point determination analyses. Using MXT -1HT SimDist columns and EZGC method translation software (http:///ezgc-mtfc ), helium-based SimDist methods can be easily translated to alternative carrier gases while maintaining ASTM Method D7213 requirements. Questions about this or any other Restek product? Contact us or your local Restek representative (/contact-us). Restek patents and trademarks are the property of Restek Corporation. (See /Patents-Trademarks for full list.) Other trademarks in Restek literature or on its website are the property of their respective owners. Restek registered trademarks are registered in the U.S. and may also be registered in other countries. 215 Restek Corporation. All rights reserved. Printed in the U.S.A. Lit. Cat.# PCAR2269-UNV