TX101/201 Insertion Turbine Instructions

Similar documents
TX101/201 Insertion Turbine Instructions

TX115/215 Hot-tap Insertion Turbine Instructions

TX 115/215 Hot-tap Insertion Turbine Instructions

TX100/200-SERIES INSERTION TURBINE INSTRUCTIONS TX101/201 TX115/215 ISO 9001:2008 CERTIFIED COMPANY

SERIES TBS INSERTION TURBINE TBS-0XX TBS-1XX

TX100/200-SERIES INSERTION TURBINE. T X / S E R I E S I N S E R T I O N T U R B I N E instru c tions TX101/201 TX115/215

TX80-Series FLOW SENSOR INSTRUCTIONS TX81 TX :2008 ISO CERTIFIED COMPANY TX80-SERIES FLOW SENSOR INSTRUCTIONS

IP800-Series FLOW SENSOR INSTRUCTIONS IP810 IP :2008 ISO CERTIFIED COMPANY IP80-SERIES FLOW SENSOR INSTRUCTIONS

IP80-Series FLOW SENSOR INSTRUCTIONS IP81 IP :2008 ISO CERTIFIED COMPANY IP80-SERIES FLOW SENSOR INSTRUCTIONS

IP800-Series FLOW SENSOR INSTRUCTIONS IP810 IP :2008 ISO CERTIFIED COMPANY IP800-SERIES FLOW SENSOR INSTRUCTIONS

IP800-Series. Insertion Paddlewheel Flow Sensor Instructions PROUDLY MADE IN THE USA

IP800-Series. Insertion Paddlewheel Flow Sensor Instructions PROUDLY MADE IN THE USA

TX800-Series FLOW SENSOR INSTRUCTIONS. TX81x TX82x 9 001:2008 ISO CERTIFIED COMPANY TX800-SERIES FLOW SENSOR INSTRUCTIONS

EX100/200-Series INSERTION MAGNETIC FLOW METER INSTRUCTIONS EX110 EX210 EX150 EX :2008 ISO CERTIFIED COMPANY

SeaMetrics Flow Meters and Flow Computers

ARCHIVED. EX80 Series. (Includes Dates 6/28/05 to 10/06) EX81 EX82

S-Series. Low Flow Meter Instructions SPX SPT. Precision Flow Measurement An ONICON Brand PROUDLY MADE IN THE USA

EX80 Series EX81 EX82. E X 8 0 S E R I E S E l e c t r o m a g n e t i C F l o w s e n s o r i n s t r u c t i o n s. The Leader in Flow Meter Value.

SES. Stainless Single-Jet Meter Instructions PROUDLY MADE IN THE USA

EX100/200 Series. Insertion Magnetic Flow Meter Instructions. EX11x EX21x. EX15x EX25x PROUDLY MADE IN THE USA

EX80-Series ELECTROMAGNETIC FLOW SENSOR INSTRUCTIONS EX81 EX82 EX :2008 ISO CERTIFIED COMPANY

TURBINE METERS - WP Series

PSF50 Paddle flow switch

Sidel Therm & CO 2 Counter Instruction & Installation Manual

EM 101 EM EM EM GENERAL INFORMATION. SPECIFICATIONS MATERIALS Wetted Materials - FLOW RANGE

Pipe fi ttings MUST be installed by a certifi ed welder only. Signet will not assume liability of any kind for improper fi tting installations.

WMX-series MUNICIPAL/INDUSTRIAL MAGMETER INSTRUCTIONS

M SERIES Multi-Jet Cold Water Meter

Pipe fittings MUST be installed by a certified welder only. +GF+ SIGNET will not assume liability of any kind for improper fitting installations.

FIXED DEPTH INSERTION METER INSTRUCTIONS

TR-QS Wafer-Style Turbine Flow Meters Installation, Operating & Maintenance Manual

Insertion turbine INSTRUCTION SHEET. TECHNICAL PRODUCT

Signet 2517 Brass Paddlewheel Flow Sensor * *

Pipe fi ttings MUST be installed by a certifi ed welder only. Signet will not assume liability of any kind for improper fi tting installations.

G2 Industrial Grade Meters

EM101 LOW FLOW MAGNETIC FLOWMETER INSTRUCTIONS

Sanitary Turbine Flow Meter - TA3 Installation, Operating & Maintenance Manual

WMP-Series PLASTIC-BODIED MAGMETER INSTRUCTIONS 9001:2008 ISO CERTIFIED COMPANY WMP-SERIES PLASTIC-BODIED MAGMETER INSTRUCTIONS. 3 inch.

BI-DIRECTIONAL INSERTION FLOW TRANSDUCER

FLOWX3 F3.20 High Pressure Paddlewheel Flow Sensor MAIN FEATURES APPLICATIONS OPERATING PRINCIPLE

LEVEL AND FLOW MEASUREMENT WS131 FLOW METER USER MANUAL.

Flow Meter & Control Catalog

G2 SERIES INDUSTRIAL GRADE METERS

INSERTION flow sensor for continuous flow measurement

MODEL 1100 TURBINE FLOW METER

Hydro-Flow Model 2200 Installation and Operation Guide B

RUN ACCUM. TOTAL STOP BAT LOW HIGH

EX800-Series. Electromagnetic Flow Sensor Instructions PROUDLY MADE IN THE USA

INSERTION flowmeter with paddle wheel, ELEMENT design

FPD1001 thru FPD1003, FPD1102 and FPD1103, FPD1201 thru FPD1203 (includes -R Option) Positive Displacement Flowmeters

BI-DIRECTIONAL INSERTION METER

Model 1100 Turbine Flow Meter

Shop online at omega.com For latest product manuals: omegamanual.info

Paddle-wheel flow controller with optical principle for On/Off control

EZ-IN Series Turbine Flowmeter

INDUSTRIAL METER CATALOG version seven

MODEL 1100 TURBINE METER REPAIR KITS

FLS F6.60 MAGMETER FLOW SENSOR INSERTION FLOW SENSORS APPLICATIONS MAIN FEATURES

FLOW METERS FOR MINING

Compu-Flow. Electromagnetic Insertion Flow Meter. Compu-Flow EXM MAG

Paddle-wheel flow controller for On/Off control

Industrial Meter Catalog

ZDM Positive Displacement Flow Meter User Instructions

Hot Tap Insertion Vortex

SERIES FLMG MUNICIPAL/INDUSTRIAL MAGMETER INSTRUCTIONS

Turbine Flow Meter Series Turbine Meter USER MANUAL

Liquid Turbine Flowmeter

Low Profile Insertion Vortex

Operating Instruction for Plastic Vortex Flow Transmitter. Model: UV

Model TUR-200D Turbine Flowmeter

Model FVP General Specification. Flanged Vortex Plate. Product Features

AG2000 IRRIGATION MAGMETER INSTRUCTIONS

General Information... Temperature/Resistance Table... Part Numbering Examples for RTD s. High Temperature MgO Elements (RA)...

Appendix. Clearance. Skewed profiles. Application Schematics

Hydro-Flow Model 2200 Fixed Insertion Vortex Flowmeter Installation and Operation Manual

Paddle Wheel Turbine Flow Meters Installation, Operating & Maintenance Manual

ELECTRONIC CONTROLLER

A1 Commercial Grade Meters

Paddle-wheel flow controller for On/Off control

Installation & Operation Manual

Chapter Page. 1 Introduction Description 1

A1 SERIES COMMERCIAL GRADE METERS A1 SERIES COMMERCIAL GRADE METERS

Paddle Wheel Turbine Flow Meters Installation, Operating & Maintenance Manual

F-2000 MODEL RT INSTRUCTION MANUAL. Blue-White. Industries, Ltd.

Technical Bulletin. 1.0 Introduction

Flow Meters & Instruments. Insertion Paddle Wheel & Magnetic Flow Meters Flow & Anaytical Instruments. Technical Catalogue

Warnings and Cautions. Precision Flow Insertion Mass Flow Meter. Instruction Manual BULLETIN F-41-PFS FR# R

Hydro-Flow Model 3100

FPD1001B thru FPD1003B, FPD1102B and FPD1103B, FPD1201B thru FPD1203B SERIES LOW FLOW Positive Displacement Flowmeters

A1 Commercial Grade Meters

KOBOLD DF-MA Flow Transmitter. User Instructions

RPFS. ROTA PULSE FLOW SENSOR (Slip Insertion Paddlewheel) DATASHEET FEATURES

Condensed Version of Operator s Manual For The VI/VP700 Series Flowmeter and VI800 Series Flowmeter

Distributed by: M&M Control Service, Inc MICRO-FLO MICRO-FLO

MULTI-JET WATER METERS

Hydro-Flow Model 3100 Retractable Insertion Vortex Flowmeter Installation and Operation Manual

flow x3 flow monitoring

WJ-Series. Turbine Meter Instructions ISO. 6 Meter. 3 Meter. Precision Flow Measurement. An ONICON Brand 9001:2008. if ie

OVAL GEAR FLOWMETER ELECTRONIC MODEL 019 / ¾

Axial Turbine Flow Sensor Series Turbotron VTH 25 / VTI 25 / VTM 25

Transcription:

TX101/201 Insertion Turbine Instructions General Information Ruby bearings and a non-drag pick-off give these adjustable insertion turbine flow sensors a wide flow and long life. A sensor detects the passage of miniature magnets in the rotor blades. (Magnets are in 2 or 4 blades only.) The resulting square-wave signal can be sent for hundreds of feet without a transmitter, over unshielded cable. This signal can be connected directly to many PLC s and other controls without any additional electronics. Installation fittings are standard 1-1/2" NPT. A depth adjustment system allows two basic sizes to cover pipe sizes from 2" to 48". Fittings such as saddles and weldolets may be purchased locally, and are also available from SeaMetrics. A modular system of electronics can be installed directly on the flow sensor, in a splashproof cast housing. The FT415/420 provide digital rate and total display, as well as programmable pulse. The FT420 also provides a 4-20 ma analog output. The AO55 is a blind analog transmitter. Programmable pulse for pump pacing comes from the PD10. Each of these controls is also available in a wall mount housing. Specifications Sensor Hall-Effect Sensor Materials Sensor body Rotor Bearings Shaft 12 VDC current sinking pulse Brass, 316 SS Polypro, PVDF (optional) Ruby Nickel-bound tungsten carbide, zirconia ceramic optional 0.2-30 FPS (.006-9 M/s) +/ 1% FS 200 psi (14 bar) Range Accuracy Maximum Pressure Maximum Temperature 200 F (93 C) Pipe Size TX101 2" - 10" (50-250mm) TX201 10" - 48" (250-1200mm) Fitting Size 1-1/2" NPT Insertion Force 0.44 x pressure in pipe Power 5-24 VDC, 1.5 ma Cable #22 AWG 3-con, 18' (6m) Max. Cable Run 2,000 ft. (650m) Features Rugged cast housing Adjustable depth for wide range of pipe sizes 18 ft. cable Adapter fitting with standard NPT threads Modular electronics: Rate/total/pulse Blind 4-20 ma transmitter Pulse Divider Battery rate/total/pulse Removeable flow sensor Field Replaceable shafts and bearings Wide flow range LT-10615A Page 1 of 6

Installation Piping. For best results, the TX sensor should be installed with at least ten diameters of straight pipe upstream and five downstream. Certain extreme situations such as partially-opened valves are particularly difficult and may require more straight diameters upstream. All 100/200 Series sensors are supplied with a 1-1/2" male pipe thread fitting. Any fitting which provides the matching NPT female thread may be used. The installation procedure compensates for differences in height of the fitting. These water meters are not recommended for installation downstream of! the boiler feedwater pump where installation fault may expose the meter to boiler pressure and temperature. Maximum recommended temperature is 130ºF (Plastic), 200ºF (Metal). Immersion. The 100/200 Series standard sensors are not designed for continuous underwater operation. If this is a possibility, as in a flooded vault, a unit modified for immersion should be specified. Fitting Installation. First cut a minimum 1-3/4" hole in the pipe. If possible, measure the wall thickness and write it down for use in depth setting, below. Then install the threaded fitting (saddle, weldolet, etc.) on the pipe. Meter Installation. Remove the threaded adapter from the sensor. Using a thread sealant, install the adapter in the pipe fitting. Do not overtighten. Slide the meter through the adapter fitting. Engage the locking collar threads and turn until the collar stops, about 1-1/2 to 2 turns. Depth Setting. It is important for accuracy that the sensor be inserted the correct depth into the pipe. Follow these steps: "D" FLOW 10 x Dia. 5 x Dia. Horizontal is the preferred installation orientation, since it improves low-flow performance and avoids problems with trapped air. Bottom, top, and vertical pipe installations are all acceptable if required by the piping layout. Fair (unacceptable if air is present) 1. In Table 1, find the model of sensor which you have. Take Dimension C from this table for the pipe size. 2. Subtract wall thickness of your pipe to find Dimension D. Use Table 2 if you do not know the thickness. 3. Measuring from the outside of the pipe to the joint in the housing, as shown in the diagram, adjust the sensor to Dimension D and tighten enough to hold it there. 4. Line the conduit housing with the centerline of the pipe, as shown. 5. Check Dimension D one more time. 6. Tighten the compression nut fully. Best 2 of 6 Fair (unacceptable if fluid contains sediment)

Revised: 1-28-02 Table 1: Dimension "C" Nominal Pipe Size 2" 2-1/2" 3" 4" 5" 6" 8" 10" 12" 16" 24" 30" 36" 48" TX101 B,S (Brass or Stainless) 9.16 9.08 8.99 8.82 8.65 8.48 8.14 7.80 7.46 6.78 TX201 B,S (Brass or Stainless) 14.16 14.08 13.99 13.82 13.65 13.48 13.14 12.80 12.46 11.78 10.42 9.40 9.38 Table 2: Pipe Wall Thickness Nominal Pipe Size 2" 2-1/2" 3" 4" 5" 6" 8" 10" 12" 16" 24" 30" 36" IPS Steel Schedule(10S) 0.109 0.120 0.120 0.120 0.134 0.134 0.148 0.165 0.180 IPS Schedule (40) 0.154 0.203 0.216 0.237 0.258 0.280 0.322 0.365 0.406 0.500 0.687 0.750 IPS Schedule (80) 0.218 0.276 0.300 0.337 0.375 0.432 0.500 0.593 0.687 0.843 1.218 (Type K) 0.083 0.095 0.109 0.134 0.160 0.192 0.271 (Type L) 0.070 0.080 0.090 0.100 0.125 0.140 0.200 Copper Pipe 0.157 0.188 0.219 0.250 0.250 0.250 0.313 Class 52 Duct. Iron 0.280 0.290 0.310 0.330 0.350 0.370 0.400 3 of 6

Flow in gallons per minute at various velocities: Sched. 40 pipe Feet / sec. (0.2) Nominal pipe size 2" 2-1/2" 3" 4" 5" 6" 8" 10" 12" 16" 24" 36" 38" 48" 2.1 3.0 4.6 7.9 12.5 18 31.2 49.1 70 110 251 583 707 1127 (0.5) 5.2 7.5 11.5 19.8 31.2 45 78 123 174 275 627 1460 1770 2820 (1.0) 10.5 14.9 23 39.7 62.4 90 156 246 349 551 1250 2910 3530 5640 (2.0) 20.9 29.8 46.1 79.4 125 180 312 492 698 1100 2510 5830 7070 11280 (5.0) 52.3 74.6 115 198 312 450 780 1230 1740 2750 6270 14570 17670 28200 (10.0) 105 149 230 397 624 900 1560 2460 3490 5510 12530 29140 35350 56400 (20.0) 209 298 461 794 1250 1800 3120 4920 6980 11020 25060 58270 70700 112800 (30.0) 314 448 691 1190 1870 2700 4680 7370 10470 16520 37600 87410 106050 170000 TX100/200 K-factors for various pipe sizes PVC/Steel Sch. 40 PVC/Steel Sch. 80 Stainless Steel (10S) Stainless Steel (40S) Stainless Steel (80S) (Type K) (Type L) Copper Pipe Duct. Iron (Class 52) Nominal pipe size 2" 2-1/2" 3" 4" 5" 6" 8" 10" 12" 16" 24" 30" 36" 38" 42" 45.43 31.84 17.41 10.11 6.43 4.45 2.13 1.35 0.951 0.602 0.258 0.158 0.11 0.098 0.080 51.62 35.97 19.49 11.19 7.07 4.08 2.33 1.48 1.05 0.660 41.21 27.95 15.42 9.03 5.84 4.06 1.95 1.25 0.880 45.43 31.84 17.41 10.11 6.43 4.45 2.13 1.35 0.951 51.62 35.97 19.49 11.19 7.07 4.08 2.33 1.48 50.57 32.73 19.39 10.91 7.10 4.11 2.36 50.57 31.94 18.89 10.75 6.90 3.97 45.65 31.05 17.48 10.24 6.39 3.61 2.12 15.61 9.20 3.43 1.92 1.25 0.87 0.49 4 of 6

Connection. Sensors are supplied with 18 ft. of cable. For sensors with no additional electronics, see diagram for color coding. For sensors with on-board electronics, see the manual accompanying the electronic module. Calibration ( K-Factor ). In order to properly process pulses from the flow sensor, a number must be entered into the control to which the sensor is connected. This number, called the K-factor, is the number of pulses the sensor puts out per unit of fluid passing through the pipe. It is normally provided for SeaMetrics sensors in pulses per gallon, and is given on the chart K-factors for Various Pipe Sizes. These numbers are based on extensive testing, which has shown close agreement between different TX sensors in the same installation. Typically, most K-factor error can be attributed to installation variables, such as depth setting and fitting configuration. It is occasionally possible to field calibrate a sensor by catching the fluid in a measured container and comparing with the number of pulses recorded. (To record individual pulses, set the K-factor on the control to 1.00.) This is especially desirable if the installation has less than the recommended length of straight pipe upstream of the sensor. Operation RED (+) 6-24 VDC WHITE (signal) BLACK (-) Power 18' cable standard Theory. In principle, an insertion flow sensor measures the velocity of flow at one point in the pipe, and flow rate and total can be inferred from this one point. Accuracy is decreased by any factor which makes the flow at the measured point unrepresentative of the entire flow stream. This of course includes distorted flow patterns caused by upsream fittings too close to the sensor. The worst offenders are fittings which increase the flow on one side of the pipe, such as partially-opened gate or butterfly valves. Fluid moving in a pipe does not flow at the same velocity. Toward the center of the pipe, fluid moves faster than at the wall, and the relationship between the two changes as overall flow rate increases. This change in the velocity profile can result in nonlinearity, which means that the K-factor which is correct for one flow rate may be incorrect for another. The recommended depth settings have been carefully chosen to minimize this source of error, and should be followed carefully, especially in the smaller pipe sizes. Flow Range. These sensors are designed to operate at flow velocities of 0.2 to 30 feet per second. (See chart for conversion to gallons per minute.) If erratic readings are encountered at low flows, check the chart to see if flow is below minimum for the pipe size. The standard shaft and bearings should have a long life at continuous high flow. Repair! Caution! Never attempt to remove a flow sensor when there is pressure in the pipe. Loosen the locking collar slowly to release any trapped pressure. If fluid sprays out when removing the sensor, stop turning and depressurize the pipe. Failure to do so could result in the sensor being thrown from the pipe, resulting in damage or serious injury. Rotor Replacement.* Rotors are easily field-replaced. Shaft and rotor are a single unit, and are not replaced separately. If replacement is due only to normal shaft wear, bearing replacement is probably not necessary. If the rotor has been damaged by impact, the bearings should also be replaced. Rotor and bearings can be ordered as a kit (see parts listing, pg 6). Follow these steps: 1. Unscrew the threaded bearing housings to expose the shaft ends. If bearings are being replaced, back them completely out. 2. Remove the rotor. Put the new rotor in its place. 3. Thread in one bearing housing part way, then the other. Take care to start the end of the shaft into the bearing hole before tightening further. 4. Screw in bearing housings until they bottom. Note: Do not use excessive force. 5. Check for free spin. Blowing lightly on the rotor should result in it spinning rapidly and coasting to a smooth stop. Turbine Rotor Bearing Housing Shaft * Note that on ceramic shaft models the shafts are in the screws and the bearings are in the rotor. Follow the same basic procedure above. (#25945) 5 of 6

Troubleshooting. The flow sensor has only one moving part, the rotor. If this is turning properly and there is no signal, the Hall-effect sensor is not operating properly. To check the signal, apply 12 VDC power to the red (+) and black (-) leads. Set a multimeter to voltage reading. Put the positive multimeter lead on the red wire and the negative lead on the white wire. Slowly turn the rotor. Voltage reading should swing between +12 Volts and 0 Volts as the rotor turns. If it does not, the Hall effect sensor is not working properly. Checking for continuity is not a useful test of these sensors. 1 2 3 13 14 10 11 4 9 12 7 5 6 All SeaMetrics flow sensors are repairable, and can be returned to the factory or distributor for repair. Please first obtain a Return Material Authorization (RMA) number. TX Parts Listing 1 Upper housing 26181 2 Gasket 26211 3 Lower housing 29930 4 Housing screw 26229 5 Plug, steel 26073 6 Plug, plastic 26079 7 Strain relief 7655 8 Sensor w/cable 26310 (Standard - FT420, FT520) 8 Sensor w/cable 29953 (Micropower - FT415) 9 Tube 14130 (Brass - TX101B) 15130 (S/S - TX101S) 14131 (Brass - TX201B) 15131 (S/S -TX201S) 10 Compression nut 14199 (Brass) 15064 (S/S) 11 Ferrule 26065 12 Locking collar 14190 (Brass) 15070 (S/S) 13 Adapter fitting o-ring 26029 14 Adapter fitting 14180 (Brass) 15060 (S/S) 15 Rotor housing o-ring 16454 16 Rotor housing 25977 (Brass) 25978 (S/S) 17 Jewel bearing assy. 25901 (2 req'd) Ceramic shaft holder 16710 (2 req'd) 18 Turbine Rotor assy. 25946 (Polypro/Carbide) 25947 Kynar (for ceramic) 8 19 Rotor repair kit (consists of #17 & #18) 25930 Standard (Polypro/ Carbide) 25945 (Ceramic/Kynar) 15 16 17 17 18 6 of 6 20419 80th Ave. So., Kent WA. 98032 USA Phone: 253-872-0284 Fax: 253-872-0285 www.seametrics.com 1-800-975-8153