Boosting Trace Detection Performance with the Vanquish Diode Array Detector and High-Sensitivity LightPipe Flow Cell

Similar documents
TECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure

The Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over

Determination of Sudan Dyes I IV in Curry Paste

Technical Overview. Introduction

High-throughput protein aggregate analysis of monoclonal antibodies using a novel dual-channel UHPLC instrument

The Vanquish Platform: Improved Throughput and Resolution of Xanthones in Mangosteen Pericarp

Stability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas

Analysis of Petroleum Fractions by ASTM D2887

Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution

Thermo Scientific Dionex UltiMate 3000 Thermostatted Column Compartment

High Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell

A Comparison of Boat Introduction and Direct Injection using the Thermo Scientific ipro 5000 Series Combustion Analyzer

Application Note. Author. Abstract. Energy & Chemicals - Petrochemicals. Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC

Thermo Scientific Dionex UltiMate 3000 Pumps

HyperShear TM HPLC Mixers

HyperShear TM HPLC and UHPLC Mixers

S 1125 HPLC Pump System

HPLC SYSTEM S 500 SERIES

HyperShear TM HPLC Mixers

HPLC System S 600 Series

HyperShear TM HPLC and UHPLC Mixers

C196-E081C. Ultra High Performance Liquid Chromatograph Nexera X2. Specifications

Online sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve

Performance Characteristics of the Agilent 1260 Infinity II Bio-Inert LC

Using the PSD for Backflushing on the Agilent 8890 GC System

Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve

HIGH PRESSURE SWITCHING VALVES

Setting up SilFlow for BackFlush in your GC

Information Technology and Economic Development: An Introduction to the Research Issues

STABILITY AND RELIABILITY: NEW APPROACHES IN PREPARATIVE HPLC COLUMN DESIGN

VERITY Purification Systems Build the perfect system to fit your research

Agilent InfinityLab LC Purification Solutions SELECTION GUIDE

breakthrough versatility

Analyte Focusing in Elevated Temperature HPLC

NEW HPLC - SYSTEM LINE

Thermo Scientific Finnpipette Performance Specifications

Neueste LC Entwicklungen

Valving Schemes for the BioLogic Duo-Flow Chromatography System

SFC-FID ASTM System. SFC-4000 Series SFC. Performance Innovation Reliability

Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE

CASE STUDY. LINK Engineering Company Inc. Resolving Customers Brake Noise Issues

QuickSplit TM Flow Splitters

Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105

Made UNDER high pressure

Application Note. Authors. Abstract. Energy & Chemicals

Liquid Chromatography

A Single Method for the Direct Determination of Total Glycerols in All Biodiesels Using Liquid Chromatography and Charged Aerosol Detection

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

Combustion ion chromatography with a Dionex Integrion HPIC System

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

Application Note. Abstract. Authors. Environmental Analysis

Plus Dodeca Cell System

Fixed Flow Splitter *

Evaluating Battery Run-Down with the N6781A 2-Quadrant Source/Measure Unit and the 14585A Control and Analysis Software.

STI OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Simultaneous Determination of Fatty Acid Methyl Esters Contents in the Biodiesel by HPLC-DAD Method

Agilent G5632 Valve Kit Instructions

Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography

DYNAS3 LEADING SOLUTIONS FOR DYNAMIC ENGINE TESTING. Explore the future

Agilent 1290 Infinity II Ultra Low Dispersion Kit

Thermo Scientific Heraeus Biofuge Stratos High-Speed Benchtop Centrifuges

Optimized Method for Analysis of Commercial and Prepared Biodiesel using UltraPerformance Convergence Chromatography (UPC 2 )

The unique operating principle of the TREIBMATIC / 7

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

Electrophoresis + Blotting + Imaging. V3 Western Workflow Visualize. Verify. Validate.

Stability Axial Compression Technology Application Ordering Information

One-Minute Essential Oils Characterization by Gas Chromatography through Nanovolume Injection

System Manual. Agilent 1260 Infinity Capillary LC System

Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis

Over time consistency of PPP results in the OECD countries

Agilent Valve Kit G4231A/B and G4232A/B/C

STEVAL-ISV012V1. Up to 5 W solar battery charger for single-cell Li-ion and Li-Pol batteries based on the SPV1040 and L6924D.

INFINEUM WORLDWIDE WINTER DIESEL FUEL QUALITY SURVEY

ACQUITY UPLC I-Class System with 2D Technology

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

UView Mini Transilluminator

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System

Selecting DC Sources for Telecommunications Equipment Test Systems

Instrumentation of HPLC. HPLC Solvent Pumping Systems

PRODUCT MANUAL. CARBONATE REMOVAL DEVICE 200 CRD 200 (4 mm) (P/N ) CRD 200 (2 mm) (P/N )

Agilent 1100 Series Capillary LC System. System Manual

Smart design has its rewards. NEW! Thermo Scientific MSC-Advantage Class II Biological Safety Cabinet. Part of Thermo Fisher Scientific

Infinite F50 family INNOVATIVE EIGHT CHANNEL ABSORBANCE READERS WITH ADVANCED LED TECHNOLOGY. 21 CFR Part 11* 98/79/EC IVD-D*

Dionex CRD Carbonate Removal Device

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

Fork through-beam sensors

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column

AppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS

Choose for Your Instrument. GC Supplies - Agilent Instrument Quick Pick Guide

Thermo Scientific Heraeus Primo and Primo R Centrifuges. Flexible and efficient, compact high-performance centrifuges

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

Detection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology

NI PS-15/16/17 Side Mount Brackets

FG Series, 230V ac Toroidal Single Primary Transformers

BioLogic DuoFlow Valving Guide

Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D

Transcription:

Boosting Trace Detection Performance with the Vanquish Diode Array Detector and High-Sensitivity LightPipe Flow Cell Alexandra Manka and Holger Franz Thermo Fisher Scientific, Germering, Germany Technical Note 165 Key Words Column ID, Dispersion, HPLC, LightPipe Flow Cell, Noise, Sensitivity, Signal-to-Noise Ratio Goal To demonstrate how the Thermo Scientific Vanquish Diode Array Detector (DAD) HL with a 6 mm high-sensitivity LightPipe flow cell boosts UV trace detection performance of HPLC applications. Detection side effects such as refractive index effects and peak dispersion are explained and used to evaluate the data from this flow cell. The peak dispersion data further leads to recommendations which column formats are best suited for this type of flow cell. The Vanquish DAD is available for the Vanquish system platform and the Thermo Scientific Dionex UltiMate 3 UHPLC system platform. Introduction One of the most important parameters for any detection technology is sensitivity. More sensitivity enables the detection of analyte traces, and therefore, in many cases a successful analysis. Different generations of UV-Vis absorption detectors for high-performance liquid chromatography (HPLC) have made evolutionary improvements in trace detection performance, through optimized optics, electronics, and dispersion characteristics. Although UV-Vis absorption detector technology is well established and mature, the Vanquish Diode Array Detector elevates absorption detection performance to a new industry-leading level. It is based on a completely new design, improving all aspects relevant for best trace detection performance. Most importantly, it uses recently developed LightPipe technology in the flow cell. According to Beer s law, the absorbance of light, and thus the signal height, is directly proportional to the interaction length between light and sample. LightPipe technology is designed to combine a long flow cell light path with minimum peak dispersion and noise. To achieve this, it relies on total internal reflections at the interface between the pipe and air on the outside 1, as shown in Figure 1. Light Beam Air Fused Silica Pipe Analyte Inside Pipe Figure 1. Simplified scheme and operating principle of a LightPipe flow cell.

2 The Vanquish DAD can be equipped with a high-sensitivity flow cell of 6 mm length for maximum interaction between light and analyte. This technical note focuses on the use of this flow cell. It demonstrates how much the sensitivity can be improved, discusses side effects, and leads to recommended column formats. Experimental Instrumentation First setup for detector and flow cell comparison UltiMate 3 RS UHPLC system equipped with: HPG-34RS Pump (P/N 54.46) WPS-3TRS Thermostatted Autosampler (P/N 584.2) TCC-3RS Thermostatted Column Compartment (P/N 573.) SRD-34 Solvent Rack with Degasser (P/N 535.9245) Either a Vanquish Diode Array Detector HL (P/N VH-D1-A) equipped with a LightPipe Flow Cell, high sensitivity, 13 µl volume, 6 mm path length (P/N 683.2), or conventional Diode Array Detector equipped with a 13 µl flow cell of 1 mm path length Chromatographic Conditions Column: Thermo Scientific Acclaim, C18, 4.6 15 mm, 3 µm (P/N 59133) Mixer volume: 2 µl Solvent A: Water Solvent B: Acetonitrile (ACN) Gradient: Time (min) %B 5 5. 1 5.4 1 5.5 5 8.5 5 Flow rate: 2.2 ml/min Temperature: 35 C Injection volume: 1 µl Detection: Slit width: 4 nm slit (both detectors) Wavelength: 254 nm Bandwidth: 4 nm Data collection rate: 2 Hz Response time:.2 s Analytes: 1. Uracil 2. Acetanilide 3. Acetophenone 4. Propiophenone 5. Butyrophenone 6. Benzophenone 7. Valerophenone 8. Hexanophenone 9. Heptanophenone 1. Octanophenone 5 µg/ml each, except for uracil: 25 µg/ml

Second setup to measure the influence of peak volume on dispersion Vanquish UHPLC system equipped with: 3 Vanquish System Base (P/N VH-S1-A) Vanquish Binary Pump H (P/N VH-P1-A) Vanquish Split Sampler HT (P/N VH-A1-A) Vanquish Column Compartment H (P/N VH-C1-A) UltiMate 3 VWD-34RS Variable Wavelength Detector (P/N 574.1) equipped with Ultra-Low Dispersion UV-Monitor, 45 nl volume (P/N 674.285) and Vanquish Diode Array Detector HL (P/N VH-D1-A) equipped with high-sensitivity LightPipe Flow Cell, 13 µl volume, 6 mm pathlength (P/N 683.2) Chromatographic conditions Column: Thermo Scientific Hypersil GOLD, 1.9 µm, Hypersil GOLD, 1.9 µm, Hypersil GOLD, 1.9 µm, 1. x 1 mm 2.1 x 1 mm 3. x 1 mm (P/N 252-113) (P/N 252-1213) (P/N 252-133) Flow rate:.12 ml/min.53 ml/min 1.8 ml/min Injection volume: 1. µl 4.4 µl 9. µl Mobile phase: 5 % acetonitrile in water, premixed Mixer volume: 35 µl Temperature: 35 C Detection: VWD-34RS Vanquish DAD HL Wavelength: 254 nm Wavelength: 254 nm Data collection rate: 5 Hz Slit width: 4 nm slit Bandwidth: 4 nm Data collection rate: 5 Hz Response time:.1 s Analytes 1. Uracil 2. Acetanilide 3. Acetophenone 4. Propiophenone 5. Butyrophenone 6. Benzophenone 7. Valerophenone 8. Hexanophenone 5 µg/ml each, except for uracil: 25 µg/ml Data Processing Thermo Scientific Dionex Chromeleon Chromatography Data System software, version 7.2 SR 2 or 6.8 SR 14 (or later)

4 Results and Discussion Comparison of Signal Response To show the effect of the extra-long light path of the high-sensitivity LightPipe flow cell, experiments were conducted with an UltiMate 3 system coupled either to a conventional DAD with a conventional 1 mm flow cell, or to the Vanquish DAD HL with a 6 mm highsensitivity LightPipe flow cell (see first experimental setup). The illuminated channels of both flow cells have the same volume of 13 µl. 4 1 ACN: 5.% 2 3 Compared to the conventional 1 mm flow cell, the 6 mm high-sensitivity LightPipe flow cell offers a six-fold increased light path. According to Beer s law, the absorbance of light, and thus the signal height, is proportional to the length of the light path. An overlay of chromatograms obtained from measurements with both flow cell types is shown in Figure 2. As expected, peaks measured with the 6 mm flow cell are almost six times higher. 4 5 6 7 Vanquish DAD with 6 mm high-sensitivity flow cell Conventional DAD with 1 mm analytical flow cell 8 9 1. 1 4 6 2.9 min 3.5 5. Although the nominal volumes of the 6 mm LightPipe and the 1 mm conventional flow cells are the same, peaks measured with the long LightPipe flow cell seem to be wider. In fact, peak widths (both peak width at baseline and peak width at half height) measured with these flow cells only differ by about 4% as a result of the different geometries of the cells flow paths. Noise Performance Comparison The noise is as important for the signal-to-noise (S/N) performance as the signal height. Figure 3 shows blank injections corresponding to the chromatograms shown in Figure 2 with two zooms. Zoom A shows a part of the linear gradient directly influencing the peaks eluting in this region. Zoom B magnifies a peak-like positive deflection caused by the sudden 1% to 5% acetonitrile step. When changing quickly between different eluents or eluent-mixtures with different refractive indexes, light is focused or dispersed by the boundary layer between these eluents/eluent-mixtures. This gradient step causes a so-called dynamic refractive index (dynamic RI) effect. Depending on the characteristics of the flow cell, this can result in positive or negative deflections (focusing or dispersion of light within the optics) of varying extent, as shown in Figure 4. Zoom B of Figure 3 demonstrates that the dynamic RI effect is pronounced for the conventional flow cell. Although analytes typically elute prior to this baseline artifact, with more complex gradient profiles such a dynamic RI effect may adversely affect the separation or may be misinterpreted as an analyte peak. It is therefore beneficial to reduce dynamic RI effects as much as possible. -5. min 8.5 Figure 2. Overlay of chromatograms measured with Vanquish DAD and 6 mm high-sensitivity LightPipe flow cell (blue) and conventional DAD with 1 mm analytical flow cell (red). 25 Vanquish DAD with 6 mm high-sensitivity flow cell Conventional DAD with 1 mm analytical flow cell 1. 25 Zoom B 2.2 Zoom A ACN: 5. % 5..7 2.3 min 3.4-3 6.2 min 6.3-2. min 8.5 Figure 3. Overlay of blank injections corresponding to the chromatograms in Figure 2 measured with the Vanquish DAD and 6 mm LightPipe flow cell (blue) and conventional DAD with 1 mm analytical flow cells (red). Zoom A shows the baseline in the area of analyte peak elution and Zoom B magnifies the dynamic RI effect caused by a quick gradient change.

Zoom A of Figure 3 shows differences in noise and drift of the two flow cell types. The drift of the baseline correlating with the gradient profile can be caused by absorption of the mobile phase or by static refractive index (static RI) effects. The mobile phase composition in this test is non-absorbing at the measuring wavelength of 254 nm, thus the drift is caused by static RI effects. These are the results of the optical characteristics of the flow cell changing with the optical density of the mobile phase. Figure 4 shows dynamic and static RI effects as result of two gradient steps (from % to 1% acetonitrile, and back to %) measured in a conventional DAD flow cell. The dynamic RI effects are a consequence of the sudden gradient step, leading to a negative deflection at the first gradient step and a positive deflection at the second step. The direction and appearance of these deflections depend on the flow cell design and are characteristic for each flow cell. The static RI effect between the two gradient steps, emerging as baseline shift corresponding to the gradient step, is a result of reflections at different interfaces inside the flow cell. Depending on the mobile phase s refractive index, light is reflected differently at the interfaces between mobile phase and flow cell, for example at the interface between the flow cell windows and the sample. Thus, with varying mobile phase composition, this reflection loss changes, leading to a baseline shift. With a linear gradient instead of gradient steps, the baseline shift of the static RI effect translates into a baseline drift. Zoom A of Figure 3 also exhibits a different noise performance for both flow cell types. The baseline is less noisy with the LightPipe flow cell. This can be attributed to the latest optical and electronic technology used in the Vanquish DAD and LightPipe flow cells. The noise difference may vary depending on isocratic or gradient elution, the type of gradient, and the efficiency of the mobile phase mixing. Signal-to-Noise Performance To achieve the best signal-to-noise ratios, the signal height should be maximized and the baseline noise minimized simultaneously. Over the course of a gradient, the baseline noise may vary. The Vanquish DAD with the high sensitivity LightPipe flow cell demonstrates increased signal heights (Figure 2) and excellent noise behavior (Figure 3). Depending on the peak and the corresponding baseline noise, this results in a S/N enhancement of factor 5-14 compared to a conventional DAD with 1 mm non-lightpipe flow cell. Figure 5 shows the S/N improvement as a ratio of the S/N value for the Vanquish DAD with high sensitivity LightPipe flow cell to the S/N value for the conventional DAD with 1 mm flow cell for all peaks of Figure 2. Noise values were obtained from respective blanks (as shown in Figure 3) in a window of five times peak width at half height around each peak. 5 6. 1. dynamic RI effect. static RI effect Acetonitrile:. % Flow: 1. ml/min dynamic RI effect -5.. min 15. Figure 4. Dynamic and static refractive index (RI) effects as consequence of two non-absorbing gradient steps.. Zoom A of Figure 3 demonstrates that the static RI effect (under given test conditions) is more distinct with the 6 mm LightPipe flow cell. However, this needs to be put in relation to the increased signal obtained with the long light path of this cell. The baseline slope is approximately four times steeper, but signals are typically six times higher. In proportion to the peak height, the baseline drift is therefore less distinct.

6 S/N improvement 16 14 12 1 8 6 4 2 1 2 3 4 5 6 7 8 9 1 Peak No. Figure 5. S/N improvement for the Vanquish DAD with 6 mm LightPipe flow cell compared to a conventional DAD with 1 mm flow cell. The S/N improvement with the Vanquish DAD is not the same for all peaks. The increase in signal height is not causing this difference as it is very similar for all analyte bands. Instead, the reason for the variations of the S/N is that the noise changes over the course of the gradient. The noise of the Vanquish DAD is always equal or better than for the conventional DAD resulting in a S/N improvement as high as a factor of 14. Limitations: Choosing the right column In general, extra column band broadening will be insignificant if the flow cell volume is no larger than approximately 1% of the (smallest) peak volume. 2,3 Although the volume of the high-sensitivity LightPipe flow cell is quite small compared to the length of its light path, a 13 µl volume is not ideal for UHPLC applications with peak volumes down to 2 µl or less. The smaller the ratio between peak volume and flow cell volume, the more extensive is the dispersion leading to more and more loss in chromatographic efficiency, smaller peaks, and pronounced peak tailing. The peak broadening effects on narrow peaks in the 13 μl flowcell were determined by two sequentially connected detectors with different flow cell volumes (also see second experimental setup). The first detector was a variable wavelength detector (VWD) with a 45 nl flow cell, creating only negligible dispersion. In the evaluation of this experiment, it is therefore treated as reference for narrowest peaks (original peak volume). Downstream of this detector, the 6 mm high sensitivity LightPipe flow cell was connected. To vary the peak volume, columns with identical packing material but different inner diameters were used. Different peak volumes were obtained as a consequence of the inner diameter (ID) change of the column: The smaller the ID, the smaller the peak volume. Furthermore, different peak volumes were obtained through isocratic separations increasing peak volumes with retention time. The ratio of Vanquish DAD peak volumes to VWD peak volumes is called peak broadening factor. Figure 6 shows the peak broadening in the high sensitivity LightPipe flow cell as a function of the original peak volume. Under isocratic conditions, within one chromatographic separation, the peak volume increases with increasing retention time or retention factors k, respectively. Peak Broadening Factor 3.5 3. 2.5 2. 1.5 Peak Broadening 1 2 4 6 8 1 12 14 Original Peak Volume (µl) column ID ID = 1 mm ID = 2.1 mm ID = 3 mm Figure 6. Peak broadening caused by the 6 mm high-sensitivity LightPipe flow cell compared to peak volumes obtained from a VWD with 45 nl flow cell as a function of original peak volume. Furthermore, Figure 6 demonstrates that larger column IDs lead to larger peak volumes and that with increasing peak volumes peak broadening decreases dramatically. In this set of experiments, acceptable peak broadening factors between 1.25 and 1.3 over the entire chromatographic separation were only obtained with the 3 mm ID column. The influence of the column geometry on the peak volume can be explained with the following equations. The peak volume V Peak is defined as V Peak = 4 column, (Eq. 1) where column is the dispersion generated by the column. The column dispersion can be also described as 2 column = V o2 (1 + k) 2 /N, (Eq. 2) with V o being the column void volume, k the retention factor and N the number of theoretical plates. Combining these two equations leads to 4V V Peak = o (1 + k). (Eq. 3) N Equation 3 expresses how the column geometry and the retention factor influence the peak volume. A larger column geometry increases its volume and consequently also its void volume, leading to increased peak volumes. A higher retention factor also leads to higher peak dispersion and consequently increased peak volumes.

In general, column and flow cell have to be chosen carefully to prevent disproportional peak broadening caused by a mismatch between peak volume and flow cell volume. As a general guideline, flow cells with illuminated volumes > 1 µl, no matter if of conventional or optical waveguide type, are recommended to be used with column IDs 3 mm. Using them with column IDs 2 mm may lead to pronounced loss of chromatographic efficiency. Conclusion The Vanquish DAD HL with the high-sensitivity LightPipe flow cell is a powerful instrument to improve detection performance. The data of this technical note demonstrates that this technology can increase the peak height by a factor of up to six and the signal-to-noise ratio by a factor of up to 14 compared to a conventional DAD. It also demonstrates and explains why the high-sensitivity LightPipe flow cell is best suited for column IDs of 3 mm. References 1. Thermo Fisher Scientific Product Spotlight: The Latest LightPipe Technology for the UltiMate 3 System, 215 [Online] http://www.thermoscientific.com/ content/dam/tfs/atg/cmd/cmd-documents/bro/bro/ chrom/lc/mod/sp-71543-hplc-dad-lightpipe- Technology-SP71543-EN.pdf (accessed April 1, 215) 2. Poole, C.F. The Essence of Chromatography; Elsevier Science: Amsterdam, The Netherlands, 23; pp 44 46. 3. Hanai, T.T. HPLC: A Practical Guide; The Royal Society of Chemistry: Cambridge, U.K., 1999; pp 25 26. Technical Note 165 www.thermofisher.com/chromatography 216 Thermo Fisher Scientific Inc. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Africa +43 1 333 5 34 Australia +61 3 9757 43 Austria +43 81 282 26 Belgium +32 53 73 42 41 Brazil +55 11 3731 514 Canada +1 8 53 8447 China 8 81 5118 (free call domestic) 4 65 5118 TN71674-EN 9/16S Denmark +45 7 23 62 6 Europe-Other +43 1 333 5 34 Finland +358 1 3292 2 France +33 1 6 92 48 Germany +49 613 48 114 India +91 22 6742 9494 Italy +39 2 95 591 Japan +81 6 6885 1213 Korea +82 2 342 86 Latin America +1 561 688 87 Middle East +43 1 333 5 34 Netherlands +31 76 579 55 55 New Zealand +64 9 98 67 Norway +46 8 556 468 Russia/CIS +43 1 333 5 34 Singapore +65 6289 119 Sweden +46 8 556 468 Switzerland +41 61 716 77 Taiwan +886 2 8751 6655 UK/Ireland +44 1442 233555 USA +1 8 532 4752