Installation and Maintenance Manual. Medical Gas Outlet. British Standard Compatible

Similar documents
Installation and Maintenance Manual. Medical Gas Outlet. British Standard Slim Style Compatible

PRECISION UK Ltd. Medical Gas Outlet BS Standard

FLEXIBLE PENDANTS with Axis MEDICAL GAS TERMINAL UNITS

Outlets and Accessories

Axis MEDICAL GAS TERMINAL UNITS

Axis AGSS TERMINAL UNITS

RIGID PENDANT. Phoenix Pipeline Products Limited. Installation, Operation and Maintenance Manual

Service Manual Gemini, Diamond II and Diamond III Medical Gas Outlets for Concealed Piping and Console Applications

Gemini, Diamond II and Diamond III Medical Gas Outlets for Concealed Piping and Console Applications

At Amico Labs we are dedicated to continued growth while producing products of outstanding quality, and providing excellent service.

Pressure Switch with Gauge for Pressure

Technical Datasheet. SP Terminal Units. Classification. Services. Terminal Block. Check Valve Assembly. Tailpipe Options

Condor Manual. Corporation

PRECISION UK Ltd DUPLEX AGSS PLANT

CEGA FORANO BY GREGGERSEN

RETRACTABLE CEILING COLUMN

Installation Instructions. ARS (Amico Rail Systems)

Technical Specification

Servicing Instructions

Operating & Maintenance Manual. Automatic Medical Gas Manifold Digital Dome Loaded Regulator NFPA v1.5

Automatic Medical Gas Manifold Dome Loaded Regulator NFPA v1.1

Spectra EA Series. the right connection. product benefits. Horizontal positioning radius of 81 (205 cm) Vertical positioning range of 27 (68 cm)

6 Bay Battery Charger Operation Manual

Installation Operation Maintenance

LLS-2054BE. Smart Battery Pack Operator Manual. ST Technologies 6018 Bowdendale Avenue Jacksonville, FL USA

OPERATING MANUAL POWER DRIVE SYSTEM. Cat. No RevH

The world's smallest 700 lb patient lift... Weighs less than 10 lbs!

Hospital Bed Collection. Amico Care Series Mobility - Simplicity - Versatility

Installation Operation Maintenance. LSSN Butterfly Valve AGA Approved 50MM - 150MM. QAD#IM6055.REVA

V4043 and V4044 Motorized Valves

FkcoqpfEctg Eqpuqng Qwvngv

Maintenance Information

DCell Suction. Model DM Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA USA Tel: Ref.

A-320 Balanced 1st Stage

Stretchers. Titan Transport and Titan Bariatric

2 CYLINDER MOBILE CART ASSEMBLY INSTRUCTIONS

MEDICAL IRRIGATION K PUMP OPERATING MANUAL: Manufactured for: K.M.I 3185 Palisades Dr Corona CA USA Model No.

TECHNICAL DATA. Viking Technical Data may be found on Style: 90 Degree Pattern (inlet to outlet)

Welker Adjustable Probe with Check Valve and Welker Automatic Insertion Probe with Check Valve

2200 North Main Street Washington, PA PH: FX:

MAINTENANCE PROCEDURE FOR MK17

TECHNICAL DATA 1-1/2 (dn40)

REGULATORS DELTA 3 SECOND STAGE TROUBLE SHOOTING SYMPTOM POSSIBLE CAUSE TREATMENT

MODEL 905V OPERATING INSTRUCTIONS

Welker Sampler. Installation, Operation, & Maintenance Manual. Model GSS-4HP

1 - S-2251 Showerhead 6 - S-2288 Showerhead AP - Anchor Plate (S-2280 Showerhead Only)

No Infrared Refrigerant Leak Detector

Table of Contents Next FLO-SAFE Manifold System Pre-Installation Kit Matrx Toll-Free: Technical Support: Fax: Rev.

Quickdraw. Model 2400 Series. Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

Welker Gas Sampler Model MPS-2

Installation, Operation, and Maintenance Manual

2013 IMPCO Technologies, Inc. Page 1 of 20 EPR Repair Kit Instructions (PPI-122, Rev-B)

SERVICE MANUAL. DiamondCare Recessed Wall Outlet and Electrical. MAN Rev. B

Steam/Water Washdown Units Safety and Operation Installation and Maintenance Instructions

Welker Probe Instrument Regulator with Liquid Eliminator Cartridge

Fluid-O-Tech ROTOFLOW ROTARY VANE PUMP REBUILD MANUAL

CYLINDER VALVES. Assembly & Maintenance Guide

Supplies Required. Teflon Tape Allen Key Wrenches Screwdriver 1/2" NPT Brass Pipe Plug Adjustable Wrench IMPORTANT!

Series 957, 957N, 957Z, 957RPDA, 957NRPDA, 957ZRPDA

Service Handbook High-Pressure Washer Pump

DECLARATION OF CONFORMITY FOR THE MEDICAL DEVICE FAMILY OUTLETS FOR COMPRESSED MEDICAL GASES AND VACUUM AND FOR ANAESTHETIC GAS SCAVENGING SYSTEMS

Proposition 65 Warning This product contains chemicals known to the State of California to cause cancer or birth defects or other reproductive harm.

Q= C. v P S Refer to Table 1 for part numbers and shipping weights. Accessories: Table 1: Valve Part Numbers and Specifications

Welker Automatic Insertion Diffusing Probe Model AIP-3DP

4400 TwinHybrid Gas Seal

T60 REGULATOR REPAIR KIT INSTRUCTIONS

Wastewater Combination Air Valve. Operation, Maintenance and Installation Manual

REGULATORS ALPHA 8 SECOND STAGE OCEANIC ALPHA 8 SERVICE PROCEDURE

ABS Dry Installed Waste Water Pumps Series FR

Service Manual Air Tech Second Stage

SAFETY MANUAL READ FIRST!

Installation Instructions

INSTALLATION, OPERATION AND MAINTENANCE MANUAL

AIR/HYDRAULIC INJECTION GUN MODEL INSTRUCTIONS

Installation Guidelines

Owner s Manual WARNING

250L Dual Cartridge Seal

250L Cartridge Dual Seal

Installation Instructions

Second Stage Regulator - 1/4 Turn

Service Manual Air Plus Second Stage

REGULATORS DELTA 4 SECOND STAGE OCEANIC DELTA 4 SERVICE PROCEDURE

MEMORY SEAL BALL VALVES MAINTENANCE MANUAL 2 12, Class 150 & 300, Regular Port, Flanged Unibody

Cordless Endodontic Handpiece. Manual Revision Date: July 2005 OPERATION MANUAL

Troubleshooting the Transmission Hydraulic System

Installation Instructions


Centrifugal Pumps (Part Nos. PS2SS PS73SS) PS2SS

INSTRUCTION MANUAL (original instructions) K2, K3

www. IndustrialPartFinder.com

LC149 SERIES BUTTERFLY VALVE INSTALLATION, OPERATION, & MAINTENANCE

INSTALLATION, OPERATION AND MAINTENANCE MANUAL

Installation and Operation Instruction Manual FWINSMAN303. Yardney - Filtaworx FW06EX - FW10

5020 Series Injector

SERVICE INSTRUCTIONS. Transfer Pump

1/2" AIR DRIVEN DIAPHRAGM PUMP

Prospenser. User Manual Mode d emploi Bedienungsanleitung Manual Usuario. For capacities of 2.5 ml 5 ml 10 ml 30 ml and 50 ml

Model 2310 Series. Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

OPERATION AND MAINTENANCE INSTRUCTION MANUAL. AEU-707A & AEU-707AV2 Implant / Surgery Systems

Transcription:

Installation and Maintenance Manual Medical Gas Outlet British Standard Compatible w w w. a m i c o. c o m

Table of Contents Product Description 3 Cleaning and Lubrication 3 Inspection and Testing 4 Installation & Dimensions 4-7 Flush Mount 5 Console Mount 6 Surface Mount 7 Service 8-12 Preventive Maintenance 8 Latch-Valve Assembly 9 AGSS Latch-Valve Assembly 10 AGSS Flow Control Adjustment 11 Rough-In Assembly-Flush, Console and Surface Mount 12 Model Numbers 13 Latch-Valve Assembly 13 Rough-in Assembly 13 Complete Outlets 13 Gas Indexing 14 Replacement Components 15-17 Latch-Valve Assembly 15 AGSS Replacement Components 16 Rough-In 17 2 Amico Corporation

Product Description The Amico Medical Gas British Outlet is composed of two separated modules: the Rough-in Assembly and the Latch- Valve Assembly. The Rough-in Assembly consists of a brass machined body that incorporates a spring loaded check assembly. A 12.7 mm OD, 9.52 mm ID copper pipe is sliver brazed into the body for external pipeline connections. The brass body and pipe assembly are inserted into a gas specific plate. The Rough-in Assembly has a colour coded label on the front plate and the copper pipe, so that the installer can easily identify the gas that the copper pipe should be connected to. The Roughin Assembly incorporated a check valve that allows the Latch-Valve Assembly to be removed for service, without requiring the pipeline to be shut down. The Rough-in Assembly has a DUAL pin gas specific indexing arrangement to prevent the wrong Latch-Valve assembly from being plugged into the Rough-in Assembly (see page 17). The Latch-Valve Assembly is manufactured with a gas specific housing which accepts, retains and releases the probe, indexed on the back side to match gas specific indexing of the Rough-in Assembly. CAUTION: DO NOT over tighten the Latch-Valve mounting screws! Distortion of the Latch-Valve can occur. Outlets are available for oxygen, medical air, surgical air, nitrous oxide, medical vacuum and anesthetic gas scavenging system. These are designed in accordance with HTM 02-01 and ISO 9170. Cleaning and Lubrication The Amico British Outlets are factory cleaned for oxygen service. Exposed surfaces of the Outlet may be cleaned with a mild detergent solution or wiped with a disinfectant commonly used in patient rooms, which is compatible with plastic. Lubricate elastomer seals sparingly with silicone lubricant that is oxygen compatible. DO NOT USE OIL www.amico.com 3

Inspection and Testing Pressure drops across the Amico British Outlet shall comply with ISO 9170 Standards. Terminal Unit Nominal Distribution Pressure (kpa) Test Pressure (kpa) Test Flow (1/min) Maximum Pressure Drop Across a Terminal Unit (kpa) 400 to 500 320 40 15 400 to 500 320 200 70 700 to 1,000 560 350 70 * Absolute pressure Vacuum *40 25 15 AGSS Requirements for flow and pressure drop across terminal units with probe inserted ISO 9170 Standards. Terminal Type Test Pressure Test Flow (1/min) Maximum Pressure Drop Across a Terminal Unit (kpa) 1 Atmospheric pressure 90 15 Note: The Amico British Outlets meet and exceed the requirements at the time of manufacture. However piping source capacity, sizing and restrictions may prevent outlets form attaining these values. Installation and Dimensions The Amico British Outlet should be mounted at a height of between 900 mm and 1400 mm above finished floor level and not less than 200 mm from any obstruction. Where more than one Outlet unit is to be mounted in one location, these should be mounted at the following spacing; Two Outlet units - 150 mm centres Three or more Outlets units - 135 mm centres Outlet units mounted in a horizontal array shall be installed in the following sequence when viewed from the front, left to right; O₂; N₂O; 50%O₂/50%N₂O; Medical Air, Surgical Air, Vacuum; AGSS Please refer to HTM 02-01. 4 Amico Corporation

Installation and Dimensions Flush Mount Flush Mounted Terminal Unit 58mm 58mm 80mm 102mm 58mm 80mm 102mm 22mm 73mm www.amico.com 5

Installation and Dimensions Console Mount Console Mounted Terminal Unit 53mm 29mm 260mm 92mm 83mm 102mm 46mm 73mm 102mm 64mm 6 Amico Corporation

Installation and Dimensions Surface Mount Surface Mounted Terminal Unit 72mm 58mm 80mm 58mm 80mm 8mm 88mm www.amico.com 7

Preventative Maintenance Regular inspection and maintenance of the Outlets will prolong its life and reduce the possibility of sudden, inconvenient component failures. The use of damaged probes or faulty equipment may require further maintenance. Outlets should be subjected to regular inspection and testing as follows: Monthly; a. Visually inspect the Outlet for signs of damage. b. Check to see if the Front Latch Assembly (2nd Fix) operates smoothly with the gas specific probe. c. If there is any disagreement with the operation of the Front Latch Assembly (2nd Fix), remove the unit for closer inspection/repair/replacement if necessary. Annually; a. Remove Trim Plate and clean debris from the mounting box. b. Remove Front Latch Assembly (2nd fix) and inspect the barrel for any damage. c. Inspect the Rough-in Assembly (1st fix) for any damage. If damage or leaking replace components. d. Test Front Latch Assembly (2nd fix) for correct operation (connection and disconnection of the probe) using a blank probe. 8 Amico Corporation

Latch-Valve Assembly 10 11 9 5 6 7 8 4 1 3 2 1. Unscrew the two Trim Plate screws (1), and remove Trim Box Plate (2) and the Trim plate (3). 2. Unscrew the two Retaining screws (5) until the Latch-Valve Assembly (4) can be removed from the Outlet. 3. Remove the two Retaining washer screws (11), and remove the Retaining washer (10). 4. Remove the gas connector (9) 5. Remove the Body Seal (6) and Poppet (7) from the front of the Gas Connector (9). Inspect the items for wear or damage and replace the Body Seal (6). 6. Remove the Gas or Vacuum Spring (8, to inspect if wear id damage. Re-install the Gas or Vacuum Spring. 7. Re-install all internal components into the Gas Connector (9). Check that the Gas or Vacuum Spring (8), Body Seal (6) and Poppet (7) are in place. Re-install the round Retaining washer (10) and secure with two Retaining washer screws (11), do not over tighten. 8. Re-install the Latch-Valve Assembly into the outlet. Coat the connector (9) with a thin coat of oxygen compatible silicone lubricant to aid insertion. Tighten down the Retaining screws (5), DO NOT over tighten, as this could damage the Latch-Valve Assembly. 9. Re-install Trim Plate (3) and Trim Box Plate (2) and tighten down Trim Plate screws (1). 10. Connect gas specific adapter into the outlet. The connection should be smooth and the adapter should lock and remain in place allowing gas to flow. If not, replace the entire Latch-Valve Assembly (4). www.amico.com 9

AGSS Latch-Valve Assembly 10 11 12 9 5 6 7 8 4 1 3 2 1. Unscrew the two Trim Plate screws (1), and remove Trim Box Plate (2) and the Trim plate (3). 2. Unscrew the two Retaining screws (5) until the Latch-Valve Assembly (4) can be removed for the Outlet. 3. After removing the Latch-Valve Assembly (4) from the Outlet, you can then pull the Check Valve Body Assembly from the back side of the Latch-Valve Assembly (4). Remove the Check Valve Body Seal (10) and check for any abnormal wear or Damage and Replace, coat the O-Ring with a thin coat of oxygen compatible silicone lubricant. 4. Disassembly of the Check Valve Body Assembly: Hold the Check Valve Body Assembly in one hand, and push down the Check Valve (6) with your thumb of the same hand, compressing the Check Valve Spring (9). Remove the Check Valve O-Ring (12). Release the Check Valve (6) slowly; the Check Valve Body Assembly will fall apart. 5. Reassembly of the Check Valve Body Assembly: Re-install the Check Valve Spring (9) onto the Check Valve (6), and then insert them into the Check Valve Body (11). Again, hold the Check Valve Body Assembly in one hand, and push down the Check Valve (6) with your thumb of the same hand, compressing the Check Valve Spring (9). Before re-installing the Check Valve O-Ring (12), coat the O-Ring with a thin coat of oxygen compatible silicone lubricant. 6. Insert the Check Valve Body Assembly into the back of the Latch-Valve Assembly. 7. Re-install the Latch-Valve Assembly into the outlet. Coat the outer portion of the Check Valve Body Assembly with a thin coat of oxygen compatible silicone lubricant, to aid insertion. Tighten down Retaining Screws (5), DO NOT over tighten, as this could damage the Latch-Valve. 8. Re-install Trim Plate (3) and Trim Box Plate (2) and tighten down Trim Plate Screws (1). 9. Connect gas specific adapter into the outlet. The connection should thread in smooth and the adapter should remain in place allowing gas to flow. If not, replace the entire Latch-Valve Assembly (4). 10 Amico Corporation

AGSS Flow Control Adjustment Reference Part Numbers: O-BSQ-REC-AGS O-BSQ-CON-AGS O-BSQ-SUR-ARS A.G.S.S. Threaded Nose Valve Flow Control Screw (3mm Allen Key) Surface Mount Box British Standard A.G.S.S. 1. Insert a 3 mm Allen Key into the front of the Check Valve. 2. Adjust the Flow Control Screw to desired setting. www.amico.com 11

Rough-In Assembly CAUTION: Ensure that the supply pressure is shut off before performing service. Inside the Rough-in Assembly is a secondary check valve whose function is to shut off gas flow when the Latch-Valve Assembly is removed. This seat/seal also prevents leakage around the latch-valve connector. As the secondary seal is only a static seal, it will rarely need replacement. However, if the seat/seal does need replacement, follow the following procedure: NOTE: Suction inlets (Vacuum and AGSS) are not supplied with Secondary Check Valves. The Secondary Check Valve (4) and Secondary Check Valve Spring (5) are not required. 1. Ensure that no pressure exists in the line by depressing the secondary check (4). 2. Remove the retaining ring (1) from the inside of the outlet body. Use a small screwdriver to pull the end of the ring toward the center and then pull the ring up and out. 3. Remove the washer (2), seat/seal (3), secondary check valve (4) and the secondary check valve spring (5). Inspect items for wear or damage and replace the seat/seal (3). 4. Re-install the spring (5), secondary check valve (4), seat/seal (3) and the washer (2). Insert the retaining ring (1) into the slot and ensure that the whole ring is seated properly. 5. Turn on the pressure and check for leaks. Re-install the Latch-Valve Assembly and perform then inspection and test. 12 Amico Corporation

Model Numbers Flush Mounted Latch-Valve Assembly Rough-in Assembly Complete Assembly Oxygen O-FASC-BSQ-OXY O-BAKBSQ-OXY O-BSQ-REC-OXY Medical Air O-FASC-BSQ-AIR O-BAKBSQ-AIR O-BSQ-REC-AIR Surgical Air O-FASC-BSQ-SAI O-BAKBSQ-SAI O-BSQ-REC-SAI Vacuum O-FASC-BSQ-VAC O-BAKBSQ-VAC O-BSQ-REC-VAC Nitrous Oxide O-FASC-BSQ-N2O O-BAKBSQ-N2O O-BSQ-REC-N2O Anesthetic Gas Scavenging System O-FASC-BSQ-AGS O-BAKBSQ-AGS O-BSQ-REC-AGS Console Mounted Latch-Valve Assembly Rough-in Assembly Complete Assembly Oxygen O-FASC-BSQ-OXY O-BAKCON-BSQ-OXY O-BSQ-CON-OXY Medical Air O-FASC-BSQ-AIR O-BAKCON-BSQ-AIR O-BSQ-CON-AIR Surgical Air O-FASC-BSQ-SAI O-BAKCON-BSQ-SAI O-BSQ-CON-SAI Vacuum O-FASC-BSQ-VAC O-BAKCON-BSQ-VAC O-BSQ-CON-VAC Nitrous Oxide O-FASC-BSQ-N2O O-BAKCON-BSQ-N2O O-BSQ-CON-N2O Anesthetic Gas Scavenging System O-FASC-BSQ-AGS O-BAKCON-BSQ-AGS O-BSQ-CON-AGS Surface Mounted Latch-Valve Assembly Rough-in Assembly Complete Assembly Oxygen O-FASC-BSQ-OXY O-BAKBSQ-OXY O-BSQ-SUR-OXY Medical Air O-FASC-BSQ-AIR O-BAKBSQ-AIR O-BSQ-SUR-AIR Surgical Air O-FASC-BSQ-SAI O-BAKBSQ-SAI O-BSQ-SUR-SAI Vacuum O-FASC-BSQ-VAC O-BAKBSQ-VAC O-BSQ-SUR-VAC Nitrous Oxide O-FASC-BSQ-N2O O-BAKBSQ-N2O O-BSQ-SUR-N2O Anesthetic Gas Scavenging System O-FASC-BSQ-AGS O-BAKBSQ-AGS O-BSQ-SUR-AGS www.amico.com 13

Gas Indexing As seen from the front of the Rough-In Assemblies VACUUM OXYGEN NITROUS OXIDE MEDICAL AIR SURGICAL AIR AGSS 14 Amico Corporation

Replacement Components Latch-Valve Assembly 10 11 9 5 6 7 8 4 1 3 2 Item Description Model Numbers 1 Trim Plate Screws (2 required) H-STPP-6150AB 2 Trim Box Plate O-X-BSQC-BOX625 3 Trim Plate O-X-BSQC-TRIM 4 British Standard Latch-Valve See Above 5 Outlet Mounting Screws (2 required) H-MPP-0616SS 6 Body Seal O-X-OXE-SEAT* 7 Poppet for Body O-X-BSQC-POPET* 8a Spring for All Gases (except Vac.) O-X-LVAPB-SPGAS* 8b Spring for Vacuum O-X-LVAPB-SPVAC* (only for Vacuum) 9 Gas Connector O-X-LVAPB-BDY 10 Retaining Washer O-X-BSQC-WASH 11 Retaining Washer Screws (2 required) H-HLTPP-04037 Above parts with an * are found in repair kit: O-RK-LVA-BSQ www.amico.com 15

Replacement Components AGSS Replacement Components 10 11 12 9 5 6 7 8 4 1 3 2 Item Description Model Numbers 1 Trim Plate Screws (2 required) H-STPP-6150AB 2 Trim Box Plate O-X-BSQC-BOX625 3 Trim Plate O-X-BSQC-TRIM 4 British Standard Latch-Valve See Above 5 Outlet Mounting Screws (2 required) H-MPP-0616SS 6 AGSS Check Valve O-X-BSQC-CHKVLV 7 Flow Adjustment Screw H-HSSS-CPM601-6SS 8 Locking Screw H-HSSS-CPM601-6SS 9 AGSS Check Valve Spring O-X-BSQC-VLV-SPR* 10 Check Valve Body Seal H-ORING-020VTN* 11 Check Valve Body O-X-BSQC-LATCH 12 Check Valve O-Ring O-X-LVAQ-ORING* Above parts with an "*" are found in repair kit: O-RK-LVA-BSQ-AGS 16 Amico Corporation

Replacement Components Rough-in Assembly Item Description Model Numbers 1 Retaining Ring O-X-BAK-RETAI* 2 Washer H-WASH-895-SS* 3 Seat/Seal O-X-BAK-SEAT* 4 Secondary Check Valve O-X-BAK-CHECK* 5 Secondary Check Valve Spring O-X-BAK-SPRING* 6 Console Rough-in Assembly See Above 7 Flush and Surface Mount Rough-in Assembly See Above Note: Suction inlets (Vacuum and AGSS) are not supplied with Secondary Check Valves. The Secondary Check Valve (4) and Secondary Check Valve Spring (5) are not required. www.amico.com 17

www.amico.com Amico Corporation www.amico.com 85 Fulton Way, Richmond Hill Ontario, L4B 2N4, Canada Toll Free Tel: 1.877.462.6426 Toll Free Fax: 1.866.440.4986 Tel: 905.764.0800 Fax: 905.764.0862 Email: info@amico.com Authorized Representative: RMS-UK Limited 28 Trinity Road Nailsea, Somerset BS484NU England Phone & Fax: 01275 858891 APE-INSTAL-MANI-BRIT-STD-OUTLET 10.26.2015