Installation Guide. Thermal Conductivity Detector on a 6850 GC Accessory G2623A

Similar documents
Installation Guide. Flame Ionization Detector on a 6850 GC Accessory G2621

Installation Guide Thermal Conductivity Detector Accessory 19232E

Installation Guide. Purged Packed Inlet, Electronic Pneumatics Control on 6890 GC Accessory G1551A

Installation Guide. Heater Interface Cable for Mass Sensitive Detector Accessory G1590A

Installing a Purged Packed (PP) Inlet

Installation Guide NPD/ECD Chemical Filters Accessory 19199N

Installation Guide. Auxiliary Pressure Control Manifold on 6890 GC Accessory G1570A

Installation Guide. Flame Ionization Detector on a 6890 GC. Accessories G1561A, G1562A, G1591A, G1598A

Installation Guide. Nitrogen-Phosphorus Detector on 6890 GC Accessories G1575A, G1576A, G1594A, G1599A

Installation Guide. Volatiles Interface for 6890 Gas Chromatograph Accessory G2319A

Installing a PTV Inlet

Nickel Catalyst Tube Accessory 19205A

Installing an Auxiliary Pressure Control Module

Installing a Purged Packed Inlet

Installing a PTV Inlet

Installing a Cool On-Column Inlet

Installing an Auxiliary Pressure Control Module

Installing the CO 2 Cryogenic Oven Cooling Kit

Agilent G1554A On-Column Capillary Inlet with EPC

Installing the Flame Ionization Detector EPC Flow Control Manifold

Installing the G2743A/G2745A Valve Actuator Kits

Agilent G2738A Upstream Capillary Interface

Nickel Catalyst Tube on a 6820 GC, Accessory G4318A

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Valve Box and Valve Accessories G4315A, G4316A, G4317A

Installing the 6850 Direct Valve Column Connector Kit

Agilent G3510A/G3511A Multimode Inlet

Agilent G3348B Flame Photometric Detector

Installation Guide. Agilent 7683 Automatic Liquid Sampler

A g i l e n t A G a s C h r o m a t o g r a p h Service Manual

Deans Switch Microfluidics

29 Installation Step 1. Unpacking the GC Cable diagrams Step 2. Placing the GC system on the benchtop Step 3. Turning the power on

Agilent 210/218 - Isocratic Pump

HP 6890Series Gas Chromatograph. Installation Guide. Flame Ionization Detector. Accessories HP G1561A HP G1562A HP G1591A HP G1598A

Control Panel Interface Upgrade Installation Guide For Model 200i and 250i Motorcycle Dynamometers Serial Number 202xxxx.

Ford 7.3L SuperDuty Cold Air Intake

Agilent 6890 Gas Chromatograph

Agilent 210/218 - Binary Pump

Alcatel-Lucent 7310 DC Metro Cell AWS (B4) Kit MC-DBR1, MC-DBR2, MC-DBR3, MC-DBR4, MC-DBR5. Installation Manual

Forbidden Diesel Performance Ford 6.7L Powerstroke EGR Delete Instructions 6.7L EGR DELETE BY FORBIDDEN DIESEL PERFORMANCE. Page 1

Electrically Held to Mechanically Latched Contactor

Installation Instructions. PowerFlex 700 Drive - Frame 8 Components Replacement

HP BladeSystem c7000 Carrier-Grade Options Installation Guide

Replacing the Batteries in the Fortress LI 660

Agilent SD-1 Purification System ml/min

Installation Instructions

Replacing a Gun RTD. Instruction Sheet P/N B. 1. Introduction. Using this Instruction Sheet

SITRANS F. Ultrasonic Flowmeters. Hi-Precision Mounts. Operating Instructions. Answers for industry.

Upgrading the Sure Coat Modular Gun Control System

DISPLACEMENT PUMP INSTRUCTIONS-PARTS LIST Rev. K. Model , Series A Model , Series B Model , Series A

GPS AutoSteer System Installation Manual

Users Manual. Accessories Motor and Phase Rotation Indicator ships with the following items: 3 test leads 3 test probes 3 alligator clips 9 V battery

5000TOC Sensor Service Manual

INSTALLATION INSTRUCTIONS

Switching DC Power Supply

CAUTION The sensor must always be completely filled with process media in order to measure accurately.

Galaxy VM. Battery Breaker Box Installation 09/

MK7 GTI Electronic Folding Mirror Conversion

Agilent 490 Micro GC Internal and External Mount Genie Filter Brackets

SOS SERIES SOS1 SOS2. Spares On Site Battery Cabinet Installation Guide rEV3

Battery Replacement. Installation Instructions. Summary of Changes. Catalog Number 2711P-RY2032. Original Instructions

Ford 6.7 EGR Delete Kit

Model 4430 Photoionization Detector (PID) Operator s Manual

SITRANS F Coriolis Flowmeters SITRANS FCS400 sensor Quick Start

Agilent G2647A (Single) and G2648A (Dual) Flame Photometric Detector Update Kits

INSTALLATION INSTRUCTIONS

CRECOMZR054B00 CRECOMZR055B00. VERTICAL AND HORIZONTAL ECONOMI$ER2 ACCESSORY MEDIUM ROOFTOP UNITS 15 to 27 1/2 TONS WARNING TABLE OF CONTENTS

Model 1100B CHG Terminator. Installation Instructions

Catalyser Introduction. 3. Hazard Information. 2. Area of Application. 3.1 Symbology

Installation instructions

NPD to FID conversion on SRI 8610C or 310C GC July2012

Altivar 61/71 Adjustable Speed Drives Plastic Kits and Power Terminal Kit

If there is additional labor or other costs above the 4 hrs labor you must call VMAC for approved coverage

In addition to this instruction sheet, the following are included in this kit (some additional parts are included for future maintenance):

Maintenance of the Electronic ( EPC ) Pressure Controllers on SRI GCs

260 Solvent Vapor Exit Accessory

Installation Instructions Electric Heaters 5 20 kw

INSTALLATION GUIDE: EGR Delete Kit Ford 6.7L Powerstroke`

Agilent G2333B (Single) and G2334B (Dual) Flame Photometric Detectors

ELECTRICAL SYSTEM UPGRADE

BATTERY REPLACEMENT GUIDE

Replacing the Battery in the Patriot SPS 250, SPS 300 and SPS 450

Siemens Drives & PLCs

Exchange of rollers from the XTS-Mover

Installation Instructions Electric Heaters 5 20 kw

INSTALLATION & PART S MANUAL

Agilent G4367A Auxiliary Electronic Pressure Control (Aux EPC) Module

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

Model Fixed Rack-Mount Kit

S4M Pulley and Gears Maintenance Kit

SAFETY INSTRUCTIONS Before installation and operation, read these instructions carefully and use this product only in the manner described by the manu

Powerstroke EGR Delete A B C

FlexJet Carriage Circuit Board (PCB) Replacement

Installation Manual. AutoSteer. Gleaner Combine. AutoGuide 2 Steer Ready. Supported Models A66 A76 R66 R76 S67 S77 PN: A

CONSOLE MANAGEMENT CONTROLLER

GPS AutoSteer System Installation Manual

HP Modular Cooling System 200/100 Options Installation Guide

Installation instructions

Installation Note for Solvent Selection Valve Upgrade Kit (G1360A)

Agilent G4236A 2D-LC Valve Installation Guide

Transcription:

Installation Guide Thermal Conductivity Detector on a 6850 GC Accessory G2623A

2

Agilent Technologies 2007 All Rights Reserved. Reproduction, adaptation, or translation without permission is prohibited, except as allowed under the copyright laws. Part number G2623-90108 Replaces G2623-90107 First Edition, JAN 2001 Second Edition, Feb 2007 Replaces Part No. G2623-90100 Operating and Service Manual. HP is a registered trademark of Hewlett-Packard Co. Printed in USA Safety Information The Agilent Technologies Thermal Conductivity Detector meets the following IEC (International Electrotechnical Commission) classifications: Safety Class 1, Transient Overvoltage Category II, and Pollution Degree 2. This unit has been designed and tested in accordance with recognized safety standards and designed for use indoors. If the instrument is used in a manner not specified by the manufacturer, the protection provided by the instrument may be impaired. Whenever the safety protection of the Agilent G2623 has been compromised, disconnect the unit from all power sources and secure the unit against unintended operation. Refer servicing to qualified service personnel. Substituting parts or performing any unauthorized modification to the instrument may result in a safety hazard. Disconnect the AC power cord before removing covers. The customer should not attempt to replace the battery or fuses in this instrument. Safety Symbols Warnings in the manual or on the instrument must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions violates safety standards of design and the intended use of the instrument. Agilent Technologies assumes no liability for the customer s failure to comply with these requirements. WARNING A warning calls attention to a condition or possible situation that could cause injury to the user. CAUTION A caution calls attention to a condition or possible situation that could damage or destroy the product or the user s work. Sound Emission Certification for Federal Republic of Germany Sound pressure Lp < 68 db(a) During normal operation At the operator position According to ISO 7779 (Type Test) Schallemission Schalldruckpegel LP < 68 db(a) Am Arbeitsplatz Normaler Betrieb Nach DIN 45635 T. 19 (Typprüfung) Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USA

4

Installing a Thermal Conductivity Detector on a 6850 Gas Chromatograph These instructions are divided into two parts: Part 1 to prepare the Electronic Pressure Control ("EPC") module for later installation into the Agilent 6850 Gas Chromatograph ("GC") Part 2 to perform the actual Accessory installation into the GC Part 1: Preparation of the TCD EPC module Locate the following item found in your Accessory kit: Covers Kit G2630-60080 consisting of: Qty. Front EPC cover 1 Back EPC cover 1 Hex nuts, 7/16-inch 3 Grommet, 7/16-inch 1 Screw, M4 x12 mm, pan head 3 Copyright 2007 Agilent Technologies Printed in USA Feb 2007 Agilent Part No. G2623-90108 5

Caution Some of these assemblies contain printed circuit boards so standard ESD precautions must be followed: use a static control wrist strap (supplied) connected to a suitable ground in handling the assemblies. 6

The following tools are required for this assembly: Tools Required Torx T-20 driver 7/16-inch open end wrench or nut driver Procedure 1. Prepare a suitable clean, dry, static-free work area. 2. Place the detector system assembly such that its EPC module is centered in your work area with its tubes oriented away from you and its ribbon cable extending to the right. 3. Remove dust caps from the gas connection fittings. 4. Install the back cover onto the three gas fittings making sure the label passes through slot provided. Loosely secure the cover using 7/16-inch hex nuts, one onto each fitting. 7

Location for TCD switching valve cable Install the TCD Swiching Valve cable 1. Locate the TCD Switching Valve cable extending from the EPC module and tie a knot in it as a strain relief approximately 30-35 mm (1.2-1.4 in) from the connector. 2. Locate the supplied small black grommet, cut it vertically on one side, and fit it around the cable wires such that the grommet is located between the connector and your tied knot. 3. Install the grommet and cable into the hole provided in the cover. A small flat-blade screwdriver or similar tool may be useful in properly seating the grommet into its hole. 8

Connector Grommet Strain relief knot Prepare the EPC module cable 1. Reorient the module such that the ribbon cable is upwards and gas fittings are to the left. 2. Fold the cable first to the left and gently crease it at the PC board to maintain the direction. Then fold the cable back on itself to the right and make a gentle crease such that the second fold is just inside the justinstalled back cover. 9

3. Next, fold and gently crease the cable back on itself again, but at an angle such that it now extends at 90-degrees away from you (in the direction of tubing from the module). 10

If correctly done, the cable and its folds will be entirely within the body of the module, and the free end of the cable and its connector will extend from the module in the direction of the tubing. 4. Fit the front cover onto the previously-installed back cover, securing it loosely using three screws provided, two in the gas fittings side and one in the PC board side. Make sure the ribbon cable and its connector exit through the slot provided. 11

5. Finally, firmly secure all cover screws and hex nuts on the gas connection fittings. Then replace protective dust caps onto the gas connection fittings to maintain cleanliness. 12

Add the TCD Switching Valve extender cable 1. Locate the TCD Switching Valve extender cable and two large grommets. 2. Connect the extender cable to the TCD Switching Valve cable from the assembled EPC module making sure connectors lock together. 3. Loop the free end of the combined cable through the grommets to hold them in place. This completes EPC module preparation for this Accessory. 13

Part 2: Installation of the TCD Accessory This kit contains: Kit G2623-64000 Thermal conductivity detector system 1 Performance evaluation sample kit 1 Bottom insulation 1 Seal insulation 1 Plastic grommet 1 Screw, M4 x 12 mm, pan head 5 Wrist strap, disposable, large 1 Installation sheet (this document) 1 The TCD assembly is a factory-assembled, tested, and calibrated unit. Disassembly of the unit is not required for installation. Qty. 14

Heater/sensor cable PRT and filament wires TCD assembly Switching valve extension cable and grommet Flow manifold cable Screws Bottom insulation Plastic grommet Seal insulation Signal board Tubing plugs Tools Required Torx T-20 driver Flat blade screwdriver Open end wrenches Safety Information Before continuing, read the safety information on your Agilent 6850 Gas Chromatograph User Information CD-ROM. 15

BNC BNC To Base - T To Base - T Local Talk Local Talk Installing a Thermal Conductivity Detector on a 6850 Gas Chromatograph Prepare for Installation WARNING Before proceeding, turn off the oven and any heated zones and let them cool down. Turn off all detector gases at their supplies, then turn off the main power switch and unplug the power cord. 1. Open the lid. If a column is installed, disconnect it at the detector end. Remove the nut warmer, insulation, and capillary adapter, if present. Close the lid. 2. Turn all gases off at their sources. Disconnect the carrier and detector gas tubing from the back panel of the instrument. Disconnect gas supply tubing from these fittings 16

Remove the Lid Top Cover Remove vent lines, if present Lid top cover, without valve box accessory Remove T-20 Torx screws (8 places) Remove detector cover, if desired Remove the lid top cover. Disconnect the Heater/Sensor Cable 1. Trace the detector heater/sensor cable to the wiring harness near the injection port cooling fan. 2. Disconnect the cable. Raise the Lid to the Service Position 1. Open the lid. Locate the counterbalance cam in the left rear corner under the lid. Loosen the screw on the right side of the cam. This allows the stop plate to drop down. 17

Cam follower Stop plate Screw Counterbalance cam 2. Raise the lid until it is stopped by the safety cable. 3. Raise the stop plate and tighten the screw to lock the lid in the upright service position. WARNING The lid is heavy. Always lock the lid when it is in the service position. 18

Remove the Detector Electronics Cover 1. Loosen the two bottom screws on the detector electronics cover. Remove screw Detector electronics cover Loosen screws 2. Remove the top screw on the cover. 3. Lift and remove the detector electronics cover. 19

4. Loosen the signal board clamp. Ribbon cable Signal board (FID shown) Signal board clamp A/D converter board 5. Unplug the signal board from the A/D converter board. 6. Disconnect the wiring from the signal board. FID Disconnect the ribbon cable. Unplug the ignitor cable. Pass both cables through the hole to the top side of the lid. The ribbon cable connector will not pass through the rubber grommet in the hole. Use a small screwdriver to push the grommet to the top of the lid; the cables will now fit through the hole. 20

Ribbon cable connector Ignitor cable connector TCD Unplug the two-wire switching valve cable at the connector. To release the PRT and filament wires, press on the orange tabs. Pass the disconnected cables/wires through the hole to the top side of the lid. Switching valve cable connector PRT wires Filament wires 7. Lower the stop plate on the lid cam. Pull the lid forward until the cam follower rests on the curved surface of the cam. Leave the stop plate in the "down" position. 8. Close the lid. 21

Cover plate screws Remove the Existing Detector and Flow Manifold 1. On the top of the lid, loosen the connector cover plate and remove the plates. Gas manifold (FID shown) Manifold mounting screws Gas fittings 2. Disconnect the gas fittings. The figure above shows the FID manifold; the TCD manifold is similar but has only one gas connection. 3. Trace the ribbon cable from the flow manifold and disconnect it from the harness. 4. Remove the three manifold mounting screws. 5. Remove the screws holding the detector assembly in place. 22

FID Remove four mounting screws. FID mounting screws (one concealed) TCD Remove two mounting screws. Mounting screws 6. Lift the detector and flow manifold out of the lid as a unit. WARNING The fibrous insulation material used in the Agilent 6850 can cause irritation to the skin, eyes, and mucous membranes. Always wear gloves when working with the insulation. If the insulation is flaky/crumbly, wear protective eyewear and a respirator. 7. Remove any insulation that may be in the hole under the detector. Install the Thermal Conductivity Detector and Flow Manifold 1. Place the bottom insulation in the detector hole. Align the flat sides parallel to the sides of the lid. 2. Slide the seal insulation onto the bottom of the detector. Align it to cover the detector bottom. 23

3. Align the thermal conductivity detector as shown and lower it into place. Position the flow manifold. Check that no wiring or tubing is trapped under either assembly. Heater/sensor cable connection Screws Switching valve cable connection PRT and filament leads 4. Secure the detector assembly to the lid with two screws. 5. Secure the manifold to the lid with three screws. Manifold cable connector is under this plate Connect gas line M here 6. Plug the manifold ribbon cable into its connector under the connector cover plate. 7. Install the two connector cover plates. 8. Connect the gas supply line labeled M to the center fitting on the flow manifold. 24

WARNING The gas line labelled H is plumbed to the hydrogen supply. Be sure to plug this line to prevent accidental release of hydrogen gas into the instrument. 9. The A and H gas lines are not used with the TC detector. If they are not already plugged, plug them with the brass plugs and push them down into the space under the manifold fittings. Avoid kinking the tubing. Connect the Heater/Sensor Cable Connect the detector heater/sensor cable to the wiring harness near the injection port cooling fan. Install the Switching Valve Extension Cable and Grommet 1. Locate the oblong hole for the switching valve extension cable. If there is a grommet in the hole, remove it. 2. Pass the end of the extension cable through the oblong hole into the signal board space. Use a small flat blade screwdriver to install the grommet in the oblong hole. Connect the other Leads and the Signal Board 1. Raise the lid to the upright service position and lock it in place by raising the stop plate and tightening the screw. 2. Locate the round hole for the PRT and filament leads. If there is a grommet in the hole, remove it and install the one from the kit. Pass the two PRT leads and the two filament leads through the grommet into the signal board space. 3. Plug the signal board into the A/D converter board and secure it with the signal board clamp. 4. Connect the PRT and filament leads to the signal board by pressing the orange tabs and inserting the appropriate wires. Polarity does not matter, but the wires must go to the correct pair of connectors. PRT leads have light-colored insulation Filament leads have thick, black insulation 25

Switching valve cable connector PRT wires Filament wires Caution Double check the PRT and filament connections before proceeding. An incorrect connection can damage the detector when power is restored. 5. Plug the switching valve extension cable into the signal board. 6. Install the detector electronics cover and tighten all three screws. 7. Lower the stop plate on the lid cam. Pull the lid forward until the cam follower rests on the curved surface of the cam. Raise the stop plate behind the cam follower and tighten the screw. 8. Close the lid. 9. Install the lid top cover. Finishing Up 1. If not already present, make an external T-connection between the carrier gas supply and the Reference/Makeup gas fitting on the back of the instrument. 2. Install the capillary adapter, if used. 3. Restore the column connection. 4. Restore carrier gas to the instrument. 5. Restore power. 6. Apply your normal operating pressures. Leak-check the manifold, back panel, and column fittings. 26

Printed on recycled paper. This product is recyclable. Agilent Technologies, Inc. Printed in USA April 2007