Agilent G4239C Valve Kit - Instructions

Similar documents
Agilent InfinityLab Quick Change Valve G4237A

Agilent G4236A 2D-LC Valve Installation Guide

Agilent Online-SPE Direct Inject Kit G4744A

Agilent G5632 Valve Kit Instructions

Agilent Valve Kit G4231A/B and G4232A/B/C

Agilent InfinityLab Quick Change Valve G4235A Instructions

Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D

Agilent Valve Kit G4231A/B/C and G4232B/C/D

Installation of the 2 position/10 port valve into the 1100 series thermostatted column compartment

1260 Infinity II Multicolumn Thermostat

Agilent InfinityLab LC Series 1260 Infinity II Fraction Collectors

Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve

Dual-Needle Infinity II Upgrade Kit

Agilent InfinityLab LC Series 1260 Infinity II Delay Coil Organizer

Agilent 1290 Infinity II Ultra Low Dispersion Kit

Agilent InfinityLab LC Series Multi-wash Upgrade Kit

Agilent 1290 Infinity Autosampler

Agilent 210/218 - Isocratic Pump

Agilent SD-1 Purification System ml/min

Please exercise care when using solvents close to the instrument. It could damage the vapor sensor present in the ELSD.

Agilent 210/218 - Binary Pump

G1160A Agilent 1100 Series Multiple Purpose Switching Valve

Agilent 1200 Series Valves

Chromatography. Liquid. HPLC Accessories SupelPRO Valves SUPELCO

Agilent 1200 Series Thermostatted Column Compartment G1316A/G1316B/G1316C

HIGH PRESSURE SWITCHING VALVES

10 Maintenance. Agilent Technologies. This chapter describes the maintenance of the module Infinity Binary Pump User Manual

Agilent 1260 Infinity Standard Autosampler

Agilent 1100 Series Thermostatted Column Compartment. Reference Manual

Agilent 1200 Series Manual Injector

Agilent 1260 Infinity Manual Injector

Installation Note for Solvent Selection Valve Upgrade Kit (G1360A)

Agilent 1260 Infinity Bio-inert Manual Injector

1260 Infinity II Degasser User Manual. 5 Maintenance

Agilent 218 Purification Solution

CHROMATOGRAPHY. SYKAM S 6000 Valve Series

FITTINGS, LOOPS, AND VALVE ACCESSORIES

Agilent 1200 Series High Performance Autosamplers & Micro Well Plate Autosampler

Agilent InfinityLab LC Purification Solutions SELECTION GUIDE

1350 Valves. 1 of 14. Jun Illustrated Parts Breakdown Agilent 6890 Gas Chromatograph Service Manual

TECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure

Quick Reference Guide. Supplies for Waters HPLC Systems

Agilent 1100 Series Capillary LC System. System Manual

System Manual. Agilent 1260 Infinity Capillary LC System

Replacing the Column Stabilizer Tube with Finger-tight Fittings

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

Agilent 1260 Infinity Binary Pump VL

Installation Guide. Volatiles Interface for 6890 Gas Chromatograph Accessory G2319A

Fittings SECTION 1 OF 4 SECTIONS ON FITTINGS

Your Essential Resource for INFINITYLAB LC SUPPLIES

Agilent 1200 Series Standard and Preparative Autosamplers

Installation Guide Thermal Conductivity Detector Accessory 19232E

Application Note. Author. Abstract. Energy & Chemicals - Petrochemicals. Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany

High Pressure Valves. MX Series II Modules Manual Sample Injection Switching

Agilent 1260 Infinity Capillary Pump

78 동방하이테크상사 T.(02) F.(02)

Agilent 1200 Series Nano Pump

Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC

Ultra High Performance Fittings. Fittings Connectors Kits

Agilent 1290 Infinity Quaternary Pump

Simplify your search for. You will find both PEEK-lined stainless steel hardware as well as standard ss column hardware at the below stockingstations.

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

LC20 CHROMATOGRAPHY ENCLOSURE OPERATOR S MANUAL Dionex Corporation

Agilent 1290 Infinity Quaternary Pump

1110 Valves. Introduction

Agilent 1290 Infinity Binary Pump

Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution

One-Piece UPchurch Fingertight Fittings 1,2. ChromTRAC. TOP SELLER see starred products in price block. Fingertight Fittings.

Technical Procedure for Gas Chromatography (GC-FID)

The Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over

Nickel Catalyst Tube Accessory 19205A

Installing a Cool On-Column Inlet

Customer Information. Service Engineer s Responsibilities. Agilent Cary 7000 UMS/UMA/Solids Autosampler

Agilent 1260 Infinity Quaternary Pump VL

Installing a PTV Inlet

Installation Guide. Nitrogen-Phosphorus Detector on 6890 GC Accessories G1575A, G1576A, G1594A, G1599A

For more information about Agilent Technologies services please visit our web site using the following URL

Installing a Purged Packed (PP) Inlet

Application Note. General Use Inlet Solvent Filters. Inlet Solvent Filters. Filters and Degassers. New!

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

Agilent 7820A Gas Chromatograph

Technical Overview. Introduction

High Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell

Liquid Chromatography LC Consumables SSI-LC

Agilent 1260 Infinity Analytical SFC System

seal, needle seat, needle assembly, seat cap kit Includes: rotor seal, needle assembly, needle seat 1100 G kit 25271

Agilent 7693A Automated Liquid Sampler

Installation Guide. Auxiliary Pressure Control Manifold on 6890 GC Accessory G1570A

1260 Infinity II SFC System

2017 HPLC Product Catalog 1

Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application

Bruker BioSpin MATCH. Multiple Adjustable Tube Clamp Holder User Manual - Version 005. NMR Solutions. think forward

Choose for Your Instrument. GC Supplies - Agilent Instrument Quick Pick Guide

Fixed Flow Splitter *

98 동방하이테크상사 T.(02) F.(02)

Agilent 7693A Automated Liquid Sampler

Physiological Calibration Standard II Calibration standard physiological fluid CALIBRATION STANDARD PROTEIN

Using the PSD for Backflushing on the Agilent 8890 GC System

MPP 910 IDS MPP 930 IDS

HPLC Tubing and Fittings Hints & Tips for the Chromatographer

Transcription:

Agilent G4239C Valve Kit - Instructions Agilent G4239C Valve Kit - Instructions Technical Information about the Agilent Valve Kit G4239C. Contents Typical Applications 2 Multicolumn Selection 2 Method Development 3 Delivery Checklist 4 Install the Valve Heads 7 Remove the Transportation Lock and the Valve Dummy 7 8 Install the Capillaries 11 Setup Examples 13 Valve Specifications 15 Valve Parts 15 Agilent Technologies

Typical Applications Multicolumn Selection Typical Applications Multicolumn Selection Advantages Increase productivity Higher instrument up-time The valve facilitating quick changes allows the selection between up to eight different stationary phases for a variety of applications, or the usage of identical stationary phases in columns with different dimensions for either faster run-times (short columns) or higher resolution (long columns) or for loading studies with different internal diameters. Figure 1 The G7116B 1290 Infintiy II Multicolumn Thermostat equipped with an InfinityLab Quick-Change 8 Column Selector Valve 2

Typical Applications Method Development Method Development Advantages: Faster method development Automated method development possible Figure 2 Totally different chromatographic results by using the same sample but three different stationary phases 3

Delivery Checklist Method Development Delivery Checklist Delivery Checklist p/n Description 5067-4233 8 Column Selector, 1300 bar 5067-4248 Capillary Kit, 0.12 mm (OPTIONAL) 5067-6697 Capillary Kit, 0.12 mm, 8-column sel, long columns, PCHex (OPTIONAL) 5067-6723 Capillary kit 0.17 mm, 8-column sel, long columns, QC-HE (OPTIONAL) 5067-6724 Capillary kit 0.17 mm, 8-column sel, short columns, QC-HE (OPTIONAL) Capillary Kit (5067-4248) # p/n Description 1 5500-1202 Capillary ST 0.12 mm x 500 mm M4-SL PS-PS Autosampler to valve 8 5500-1199 Capillary ST 0.12 mm x 130 mm M4-SL PS-PS Valve to heat exchanger 1 5500-1200 Capillary ST 0.12 mm x 130 mm SL/M Column to valve 1 5063-6591 PEEK Fittings 10/PK Column outlet 8 5500-1201 Capillary ST 0.12 mm x 105 mm SL-- PS-LS Heat exchanger (ps-sl) to column (long socket - use G1314-68703) 8 G1314-68703 Cap fitting kit special Column inlet connection fitting 1 5500-1203 Capillary ST 0.12 mm x 280 mm M4-SL PS-PS Valve (ps-m4) to detector (ps-sl) 1 5500-1204 Capillary ST 0.12 mm x 150 mm M4-M4 PS-PS Valve to valve (column bypass) 1 5023-2504 Hex driver SW-4 slitted Tool for M4 fittings 8 G7116-60015 Heat Exchanger Assembly 1.6 µl-z Quick Connect Heatexchanger Standard Flow Low dispersion heat exchangers 1 G1375-87326 Waste tube Waste tube incl. M4 PEEK fitting 3 5067-6141 M4 Blank nut for plugging unused valve ports 8 G7116-68003 Column Holder Clips (2/Pk) 4

Delivery Checklist Method Development Capillary Kit, 0.12 mm, 8-column sel, long columns, PCHex (5067-6697) # p/n Description 1 5023-2504 Hex driver SW-4 slitted 8 5500-1201 Capillary ST 0.12 mm x 105 mm SL-- PS-LS heat exchanger to column 8 G1314-68703 Cap fitting kit special Fitting Column Inlet 3 5067-6141 M4 Blank nut 16 5500-1286 Capillary ST 0.12 mm x 340 mm SL/M 16 valve to heat exchanger / column to valve 1 5500-1272 Capillary ST 0.12 mm x 800 mm M4-SL 1 ALS to valve 1 5500-1203 Capillary ST 0.12 mm x 280 mm M4-SL PS-PS valve to detector 1 5500-1204 Capillary ST 0.12 mm x 150 mm M4-M4 PS-PS valve to valve bypass 8 G7116-60015 Heat Exchanger Assembly 1.6 µl-z Quick Connect Heatexchanger Standard Flow 8 G7116-68003 Column Holder Clips (2/Pk) 1 G1375-87326 Waste tube valve to waste 1 5062-8541 Fingertight fitting long 10/pk 1 5965-0050 CARTON-CORRUGATED Capillary kit 0.17 mm, 8-column sel, long columns, QC-HE (5067-6723) # p/n Description 1 5023-2504 Hex driver SW-4 slitted 8 5500-1193 Capillary ST 0.17 mm x 105 mm, long socket Heat exchanger to column 8 G1314-68703 Cap fitting kit special Fitting column inlet 3 5067-6141 M4 Blank nut 16 5067-5108 SST-Capillary 340x0.17 mm ps-ns SW-M4 Valve to heat exchanger / column to valve 1 5067-5120 SST Cap. 0.17 mm ID 700 mm SW-1/16/M4 Autosampler to valve 1 5067-5108 SST-Capillary 340x0.17 mm ps-ns SW-M4 Valve to detector 1 5067-4737 Capillary ST 0.17 mm x 150 mm M/M Valve to valve bypass 8 G7116-60051 Quick-Connect Heat Exchanger Large ID 8 G7116-68003 Column Holder Clips (2/Pk) 1 G1375-87326 Waste tube Valve to waste 1 5062-8541 Fingertight fitting long 10/pk 1 5065-4454 Fitting screw long 10/pk 1 5067-6654 Number Kit 1-8 colored (3x) 5

Delivery Checklist Method Development Capillary kit 0.17 mm, 8-column sel, short columns, QC-HE (5067-6724) # p/n Description 8 G7116-60051 Quick-Connect Heat Exchanger Large ID 8 G7116-68003 Column Holder Clips (2/Pk) 1 5067-6188 Capillary ST 0.17 mm x 500 mm SL-M4 PS-PS Autosampler to valve 8 5067-5109 Capillary ST 0.17 mm x 90 mm SL/M Valve to heat exchanger 8 5067-4746 Capillary ST 0.12 mm x 250 mm SV/M long column to valve 8 5500-1200 Capillary ST 0.12 mm x 130 mm SL/M Short column to valve 1 5063-6591 PEEK Fittings 10/PK Column outlet 8 5500-1193 Capillary ST 0.17 mm x 105 mm, long socket Heat exchanger to column 1 5500-1203 Capillary ST 0.12 mm x 280 mm M4-SL PS-PS Valve to detector 1 5500-1204 Capillary ST 0.12 mm x 150 mm M4-M4 PS-PS Valve to valve (column bypass) 2 5065-4454 Long Fittings and Ferrules 10/pk 1 5023-2504 Hex driver SW-4 slitted 1 5067-6141 M4 Blank nut 1 G1375-87326 Waste tube 1 5965-0050 CARTON-CORRUGATED 1 5067-6654 Number Kit 1-8 colored (3x) Capillary & Fitting Information M4 = fitting thread size for micro valve ports PS = pre-swaged fitting NS = non-swaged fitting SL = fitting screw long LS = long capillary socket (required for special fittings as the A-Line Quick Turn Fittings) 6

Install the Valve Heads Remove the Transportation Lock and the Valve Dummy Install the Valve Heads The valve drives are factory-installed in the Multicolumn Thermostat. The valve heads are interchangeable and can be easily mounted. At the first installation, the transportation lock and the dummy valve have to be removed, see Remove the Transportation Lock and the Valve Dummy on page 7. The valve heads can be installed by mounting the valve heads onto the valve drives and fastening the nut manually (do not use any tools). Be sure that the guide pin snaps into the groove of the valve drive thread. The valves are mounted on pull-out rails to allow easy installation of capillaries. Push the valve gently into its housing until it snaps into the inner position, push it again and it slides out. When all capillaries are installed, push the valve back into its housing, see Install the Valve Head and Connect Capillaries on page 8. Remove the Transportation Lock and the Valve Dummy The following procedure demonstrates the necessary steps for installing the valve head to the valve drive of a Multicolumn Thermostat (MCT). For the installation of a valve head to a G1170A Valve Drive you can ignore the steps that describe the MCT features of the transportation lock and spring loaded valve drive. 1 Switch off the module. 2 When unscrewing the transportation lock, push it back until the last screw is removed - the valve rail is spring-loaded. 7

Install the Valve Heads 3 Press on the valve dummy (1.) to release it (2.) (spring-loaded valve rail). 4 5 For bio-inert modules use bio-inert parts only! CAUTION The valve actuator contains sensitive optical parts, which need to be protected from dust and other pollutions. Pollution of these parts can impair the accurate selection of valve ports and therefore bias measurement results. Always install a valve head for operation and storage. For protecting the actuator, a dummy valve head can be used instead of a functional valve. Do not touch parts inside the actuator. CAUTION Column Damage or Bias Measurement Results Switching the valve to a wrong position can damage the column or bias measurement results. Fit the lobe to the groove to make sure the valve is switched to the correct position. 8

Install the Valve Heads CAUTION Valve Damage Using a low pressure valve on the high pressure side can damage the valve. When using multiple column compartments as part of a method development solution, make sure that the high pressure valve head is connected to the autosampler and the low pressure valve head is connected to the detector. CAUTION Sample degradation and contamination of the instrument Metal parts in the flow path can interact with the bio-molecules in the sample leading to sample degradation and contamination. For bio-inert applications, always use dedicated bio-inert parts, which can be identified by the bio-inert symbol or other markers described in this manual. Do not mix bio-inert and non-inert modules or parts in a bio-inert system. For a correct installation of the valve head, the outside pin (red) must completely fit into the outside groove on the valve drive s shaft (red). A correct installation is only possible if the two pins (green and blue) on the valve head fit into their corresponding grooves on the valve drive s actuator axis. Their match depends on the diameter of the pin and groove. The tag reader reads the valve head properties from the valve head RFID tag during initialization of the module. Valve properties will not be updated, if the valve head is replaced while the module is on. Selection of valve port positions can fail, if the instrument does not know the properties of the installed valve. To allow correct valve identification, power off the valve drive for at least 10 s. For firmware requirements see Information on new RFID Tag Assembly Version Technical Note (01200-90133) which is included to each valve head. 9

Install the Valve Heads The following procedure shows the valve head installation with an G7116 (MCT) module as an example. For other modules it is similar. 1 Insert the valve head into the valve shaft. 2 When the outer pin is locked into the groove, manually screw the nut onto the valve head. OR If the outside pin does not fit into the outside groove, you have to turn the valve head until you feel that the two pins snap into the grooves. Now you should feel additional resistance from the valve drive while continuously turning the valve head until the pin fits into the groove. Fasten the nut manually. Do not use any tools. 10

Install the Capillaries 3 Install all required capillary connections to the valve. 4 Push the valve head until it snaps in and stays in the rear position. 5 Power on or power-cycle your module, so the valve head gets recognized during module initialization. Install the Capillaries CAUTION Damage to the rotor seal Instant pressure release within the valve will lead to water jet effects that can harm internal parts of the valve. This pressure release typically happens if the valve gets switched under high pressure over unused or open channels. Block all unused channels properly with the M4 blank nut. To minimize valve movement over open connections it is recommended to plumb the column connected channels in one row. e.g.: channel 1 column 1 channel 2 column 2 channel 3 column 3 channel 4 column 4 channel 5 blocked channel 6 blocked channel 7 waste channel 8 bypass The blank nuts are only required for the ports on the inner circle that connect the valve with the column inlet. 11

Install the Capillaries The capillary connections depend on the capillary kit that is in use. For details, see Delivery Checklist on page 4. 1 Install the in and out connectors. from sampler to the valve from valve to the detector The In port is hydraulically connected to the column inlet ports 1-8 on the inner ring while the Out port connects to the column outlet ports 1`-8` on the outer ring. 2 Install the column inlet and outlet connections. ports 1-8 for connections from valve to the heat exchanger ports 1`-8` for connections from column outlet to valve 12

Setup Examples Setup Examples 1 8 Column-selection Figure 3 Hydraulic flow path schematics for an 8-column selection setup 2 Six column selection with purge line and valve- bypass Figure 4 Hydraulic flow path schematics for 6-column selection setup with purge and valve bypass line. 13

Setup Examples 3 The Kit 5067-6697 and 5067-6723 are for column selection of 8 long columns, i.e. longer than 150 mm. For this set-up a second MCT is required. In each MCT four columns are placed. Figure 5 Example setup based on a 1290 Infinity II LC (with two MCT), two stack configuration 4 The plumbing is exactly the same as for the kit 5067-4248. 14

Valve Specifications Valve Specifications Table 1 Type G4239C (5067-4233), 8-column selector, 1300 bar Specification Liquid contacts Port size Flow passage diameters Port to Port Volume Maximum pressure PEEK, Stainless Steel Accepts M4 male threaded fittings Stator: 0.25 mm (0.010 in) Rotor Seal: 0.30 mm (0.012 in) 1.46 µl Pre-column (inlet side of the valve) 1.52 µl Post-column (outlet side of the valve) 1300 bar ph Range 0 14 Maximum Temperature 110 C Valve Parts Valve Rotor Seal Stator Head Bearing Ring Stator Screws (Pack of 10) Stator Ring 8 Column Selector, 1300 bar (5067-4233) Rotor Seal (PEEK) (5068-0200) Stator Head (5068-0199) Isolation seal (1535-4045) Stator screws (5068-0089) Stator ring (5068-0120) 15

Valve Parts Valve Head Parts Valve Head Parts The figure below illustrates replacement parts for the valve heads, with the 12ps/13pt Selector valve as an example. The valves can vary in their appearance and do not necessarily include all of the illustrated parts. Neither, every spare part is available for each flavor of the valve. Figure 6 Valve Head Parts (example) 1 Stator screws 2 Stator head assembly 3 Stator ring screws (not available) 4 Stator ring (available for service only) 5 Rotor seal 6 Bearing ring 7 Spanner nut (available for service only) *G4239-90002* *G4239-90002* G4239-90002 Rev. B Edition: 12/2017 Printed in Germany Agilent Technologies, Inc 2016-2017 Agilent Technologies, Inc Hewlett-Packard-Strasse 8 76337 Waldbronn Germany