High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System

Similar documents
Application Note. Abstract. Authors. Environmental Analysis

AppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

Using the PSD for Backflushing on the Agilent 8890 GC System

Analysis of Petroleum Fractions by ASTM D2887

Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585

High-Temperature Simulated Distillation System Based on the 6890N GC Application

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column

Method Development for Capillary GC Systems. Slide 1

Application. Gas Chromatography June 1995

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS

Large Volume Injection of Polycyclic Aromatic Hydrocarbons

Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC

Fausto Munari e Andrea Cadoppi ThermoFisher - Italy

Agilent Multimode Inlet for Gas Chromatography

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

Application Note. Authors. Abstract. Energy & Chemicals

The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector

GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method

Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter

Analysis of Fatty Acid Methyl Esters (FAMES), and Examination of Biodiesel Samples for these Components, by GCxGC-FID

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

Agilent 7693A Automated Liquid Sampler

Agilent Multimode Inlet

White Paper. Improving Accuracy and Precision in Crude Oil Boiling Point Distribution Analysis. Introduction. Background Information

Technical Procedure for Gas Chromatography (GC-FID)

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis

Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584

6890 CHECKOUT PROCEDURE TCD (THERMAL CONDUCTIVITY) DETECTOR

Application Note. Author. Abstract. Energy & Chemicals - Petrochemicals. Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany

GC Inlets. An Overview. Simon Jones GC Applications Engineer

Application Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER

Setting up SilFlow for BackFlush in your GC

Analysis of Glycerin and Glycerides in Biodiesel (B100) Using ASTM D6584 and EN Application. Author. Abstract. Introduction

Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC

Application Note. Author. Introduction. Energy and Fuels

Fast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC

Refinery Gas. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.

Application Note. Authors. Abstract

High Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell

Agilent and ASTM. Update on Recent Activities. Page 1

Agilent 6850 Series II Network GC System G2630A Performance Specifications

STI OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application

Live Crude Oil Volatility

Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution

Complete Fractionation of Extractable Petroleum Hydrocarbons Using Newly Developed EPH SPE Cartridges

One-Minute Essential Oils Characterization by Gas Chromatography through Nanovolume Injection

Live Crude Oil Volatility

GC Method Compliance and Large Valve Oven Application

Liquefied Gas Injector. Solution for the Sampling and Analysis of Liquefied Gases

Jaap de Zeeuw Restek Middelburg, The Netherlands. Copyrights: Restek Corporation

The next level of True Blue Performance

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

Quantification and Characterization of Sulfur in Low-Sulfur Reformulated Gasolines by GC-ICP-MS Application

Method Detection Limits for EPA Method 8015 Diesel Range Organics using Fully Automated Extraction and Concentration

DANI Transformer Oil Gas Analyzer

PRODUCT SPECIFICATION. Product Specification Crystal 9000 Gas Chromatograph. Chromatec Crystal Laboratory Gas Chromatography System

Method-Specific Sample Preparation for EPH Analysis

Determination of Phenolic Antioxidant DBPC and DBP Levels in Electrical Insulating Oil

Restek Corporation 110 Benner Circle, Bellefonte, Pa Barry L. Burger, Neil Johansen, Valerie Gamble, Donald Rhoades

Optimized Method Development of Large Volume Injection for GC/MS/MS of Food Pesticides

Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems

performance productivity reliability

Forensic Blood Alcohol Determination with the Intuvo 9000 GC

Group-Type Analysis (PiPNA) in Diesel and Jet Fuel by Flow Modulated GCxGC FID.

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

Performing ASTM 6584 free and total glycerin in BioDiesel using an SRI Gas Chromatograph and PeakSimple software

Evaluation of Total Petroleum Hydrocarbon in Soil Using LC with Fraction Collector and GC/MS

Agilent 9000 Intuvo GC Applications. Matthew Giardina, Ph.D. James McCurry, Ph.D. Rebecca Veeneman, Ph. D.

CONSTITUTION OF COAL TAR PITCH AND ITS EFFECT ON PROPERTIES

Detailed Hydrocarbon Analysis Featuring Rtx -1 PONA Columns

ANALYSIS OF GASOLINE RANGE HYDROCARBONS ON BP1-PONA

Determination of Sudan Dyes I IV in Curry Paste

Set Your Sights on Superior Performance

Characterization of Tiki Torch Fuels

Technical Overview. Introduction

SPECIFICATION. Bench top triple quadrupole Mass Spectrometry System coupled to a gas performance Chromatography (GC-MS/MS)

MET-Biodiesel Capillary GC Columns

Detection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology

Techniques for Improving the Reproducibility of GC Analysis

AppNote 1/2010 KEYWORDS ABSTRACT. Biodiesel, Automation, ASTM D

Operation and Applications of Differential Flow Modulation

REFINED RESTEK. Detailed Hydrocarbon Analysis. innovative petrochemical solutions. Featuring Rtx -DHA Columns

Sulfur Detection at ppb Levels in Light Hydrocarbon Streams

Same Column and Gas Type Try Different/Faster Velocities. Same Column, Switch He to H2 Carrier Then Try Faster Velocities

Optimized Supercritical Fluid Chromatographic Instrumentation for the Analysis of Petroleum Fractions

Agilent 7693A Automated Liquid Sampler

Optimized Method for Analysis of Commercial and Prepared Biodiesel using UltraPerformance Convergence Chromatography (UPC 2 )

Transcription:

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System Application Note Authors Frank David Research Institute for Chromatography, Pres. Kennedypark 26, B-5 Kortrijk, Belgium Matthew S. Klee Agilent Technologies, 25 Centerville Rd., Wilmington, DE 19 Abstract Cycle time for gas chromatography-flame ionization detection analysis of mineral oil in environmental samples was dramatically reduced and the sensitivity increased by fast oven temperature programming using a low thermal mass system. Regulated method requirements are met for the environmental analysis of the C 1 hydrocarbon fraction in soil and water extracts using splitless injection with an analysis time less than three minutes. Cool-down time is less than two minutes, resulting in an injectioninjection cycle time of five minutes. Method performance criteria, including repeatability, linearity and solute discrimination are presented.

Introduction Environmental contamination by hydrocarbon fractions, such as diesel or motor oil, is currently measured using gas chromatography with flame ionization detection (GC-FID). This method, also called hydrocarbon oil index (HOI), mineral oil, or total petroleum hydrocarbon (TPH) determination is one of the most important applications in environmental analysis because it represents the highest sample loads in many laboratories. For analysis, samples (water, soil, sediment) are extracted by an apolar (hydrocarbon) solvent with a boiling point between 36 C and 69 C (for example, hexane). The extract is cleaned by passing over Florisil (to retain more polar solutes such as lipids), concentrated by N 2 blow down (or Kuderna-Danish) and analyzing by GC-FID [1]. The fraction eluting on an apolar (HP-1, HP-5) column between decane (C 1 ) and tetracosane (C 4 ) is defined as mineral oil or HOI, over which the area is summed for quantitation. The ISO 9377 method specifies the use of a column with a high phase ratio (thin film) to facilitate elution of C 4. GC oven program conditions should also allow the separation of the extraction solvent from the first peak (decane). Therefore, low initial temperatures (35 C to 4 C) are often necessary. Minimal solute discrimination is an important requirement of the injection method. The method specifies that the relative mass response for tetracosane and eicosane (C 2 ) should be higher than.. Typically, the analysis is performed using a 1 to 3-m column using splitless, programmable temperature vaporizing (PTV) or cool on-column injection and oven programming from 4 C to 34 C at 1 to 2 C/min, resulting in analyses times of 2 to 3 min [2]. Oven cool-down time to the low initial temperature typically requires an additional 5 min or more, resulting in a total cycle time of 3 min or longer. Currently, environmental laboratories are seeking ways to improve throughput and decrease cost per sample. Significantly improved cycle times can be achieved by adding an LTM module to an Agilent 79A GC system with a split/splitless inlet (SSI), fast automatic liquid sampler (763B or 7693A), and FID. In this application note, optimized method conditions and performance metrics are presented. Cycle times of 5 min are demonstrated, while still meeting all method requirements. Experimental Conditions Solute and Sample Preparation An alkane standard containing even numbered n-alkanes from C 1 to C 4 was purchased from Restek (cat. no 37, Restek, Bellefonte, USA). The sample was diluted to 5 ng/µl in hexane. Mineral oil calibration was done using a 1:1 mixture of diesel and motor oil (cat. no 33, Restek). Calibration samples were prepared at concentrations between 4 and 1 mg/l in hexane. In addition, a reference sample from RIVM (NMI, The Netherlands) containing diesel and motor oil was used. This sample was diluted at 1 mg/l in hexane. GC-FID Conditions Analyses were performed on an Agilent 79A GC System equipped with an SSl, FID and an LTM column module containing a 1 m.32 mm,.1 µm DB-5HT column (to order, use p/n 1-2LTM for custom LTM column, and state 123-571 with long legs in the description) connected directly to an SSl and FID. The fast GC conditions are listed in Table 1. Table 1. GC-FID Setpoints for Fast Mineral Oil Analysis Using a Low Thermal Mass Oven Injection 1 µl, splitless (.4 min purge delay), 35 C Inlet liner Split/splitless, p/n 513-4647 (bottom taper, glass wool near top, 4 mm id) SSl Inlet He pressure [emulates constant flow mode] Program 12 kpa (.5 min) to kpa @ 4 kpa/min to 1 kpa (.5 min) @ 2 kpa/min Standard oven program 4 C (.5 min) to 34 C (.5 min) @ 15 C/min [total run time = 21 min] GC oven temp when 34 C isothermal using LTM LTM column program 4 C (.5 min), 2 C/min to C, 1 C/min to 34 C (.5 min) [total time = 3 min] FID 34 C, H 2 = 4 ml/min, Air = 4 ml/min, N 2 = 2 ml/min 2

Results and Discussion System suitability was checked using a C 1 alkane test mixture. The obtained chromatogram is shown in Figure 1. Decane elutes at. min and is well separated from the solvent. Tetracosane elutes at 2.6 min. Retention time repeatability and peak area repeatability were determined for six runs and the results are summarized in Table 2. The repeatability of retention times was excellent, with a standard deviation <.3 min (<.1% RSD). This effectively demonstrates that the LTM system reproducibly heats the capillary column. Peak area repeatability was also excellent (<1% RSD). This is due to the combination of fast autoinjection with an Agilent 763B or 7693A ALS and appropriate liner selection. Discrimination was checked against requirements by measuring the peak area ratio of C 4 /C 2. In addition, discrimination for C 1 /C 2 and C 4 /C 1 were also determined and are presented in Table 2. The C 4 /C 2 ratio was.99 (±.), well above the method criterion (>.). This provides a nice margin for maintaining compliance and demonstrates that the standard SSl inlet can meet method requirements. Splitless injection is also the most robust injection method and it is applicable to extracts from both clean samples such as surface water, and contaminated samples, such as soil or sediment. Table 2. Figures of Merit Solutes/sample t R (min) t R Peak Area Peak Area mean (n=6) s (min) (RSD%) mean (n=6) RSD (%) C 1.12.1 (.6%) 4265.6 C 2 1.51 <.1 (<.1%) 4579 1.5 C 4 2.633.3 (.1%) 4117.5 Ratio C 4 /C 2 (*).99 1.31 Ratio C 1 /C 2.935.97 Ratio C 4 /C 1.965.96 Mineral oil (**) 1.711.1 (.3%) 42.64 RIVM C 1 -C 2 9.45 RIVM C 2 1119. RIVM ratio (***) 1.356.31 (*) ratio of C4/C2 should be >. (**) Restek cat. no. 33, at 4 mg/l (***): ratio of area sums C 2 /C 1 should be between 1.25 and 1.4 1 5 2 2 C 1 Response 12 C 2 4 C 4. 1. 2. Time Figure 1. n-alkane test mixture run with fast oven program on LTM oven module. 3

Next, a calibration mixture of diesel and motor oil was analyzed. The chromatogram for a 4 mg/l calibration sample is shown in Figure 2. The two humps corresponding to diesel and motor oil fractions can easily be detected. Calibration is normally done with a synthetic mineral oil made from a composite of diesel and motor oil in the concentration range from 1 to 1 mg/l. The linearity of the fast GC-FID LTM method was tested from 4 mg/l to 1 mg/l. The obtained calibration curve of the peak area (sum of peak area from end of decane peak to start of tetracosane peak) as a function of concentration is shown in Figure 3. The linearity was excellent (r² >.999). The repeatability of the peak area (sum C 1 ) of the calibration mixture at 4 mg/l was better than 1% RSD and the limit of detection was below 25 mg/l. As an additional benefit of fast oven programming, the hydrocarbon fraction is compressed into a narrower and higher hump. Therefore, method sensitivity is higher than standard methods that use slower temperature programming. A comparison of a standard oven program with the fast LTM program (3-min run time) is shown in Figure 4 for a 4 mg/l sample. The compression and higher signal gained from the seven times faster LTM run is illustrated. Additional discrimination testing was done using a popular reference sample from RIVM. Peak areas were measured for the C 1 -C 2 and C 2 fractions. The mean peak areas (n=6) and corresponding RSDs are included in Table 2. The repeatability of peak area was again excellent with a relative standard deviation < 1%. The ratio of the peak areas of the C 2 fraction ( motor oil ) versus the C 1 -C 2 fraction ( diesel ) was 1.36 (.31% RSD). These results are well within the specifications, since methods require a value to be between 1.25 and 1.4. This test also clearly shows that solute discrimination was minimal using a split/splitless inlet and fast autoinjection, and that the fast temperature program easily meets method performance criteria. 1 5 2 Diesel 2 Response 12 Motor oil 4 1. 2. Time Figure 2. 4 mg/l calibration sample of diesel plus motor oil. 4

12 1 R 2 =.9991 Area (x1-6 ) 6 4 2 2 4 6 1 mg/l Figure 3. Mineral oil calibration. Calibration from 4 1 mg/l. Linearity: r² >.999; repeatability at 4 mg/l: RSD on peak area =.64%; LOD: < 25 mg/l. 1 5 LTM, 3 min run Response. 2. 4. 6.. 1. 12. 14.. 1. 2. Standard, 21 min run. 2. 4. 6.. 1. 12. 14.. 1. 2. Time min Figure 4. Comparison of sensitivity gain from fast LTM program to standard oven program. Diesel plus motor oil standard at 4 mg/l. 5

As can be deduced from the repeatability data, automatic integration of areas using fixed integration event times will remain valid and accurate for the whole C 1 range, as well as for specific regions. These include smaller fractions such as C 1 -C 2, C 2 -C 3, and so forth, and reduces or eliminates the need for manual integration. Conclusion Great improvements can be achieved in the GC-FID analysis of mineral oil in environmental samples by adding an LTM module to an Agilent 79A GC. Analysis time for the separation of C 1 to C 4 alkanes is below 3 min, and cool-down time to 4 C was also very short (2 min). This results in a total injection-injection cycle time of 5 min. Excellent repeatability (retention times and peak areas), linearity and low LOD were achieved. The fast autoinjection allowed method criteria to be met using a standard hot SSl inlet. References 1. International standard ISO 9377-2, Water Quality, Determination of hydrocarbon oil index, part 2: Method using solvent extraction and gas chromatography, 2. 2. B. Wuest, Agilent Technologies application note 59-621EN, 2. www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 29 Printed in the USA October 23, 29 599-451EN