THERMOCOUPLE CALIBRATOR Mode KM-CAL-710

Similar documents
RTD CALIBRATOR Mode KM-CAL-720

4-20mA DC PROCESS CLAMP METER

OPERATION MANUAL MILLIAMP PROCESS CLAMPMETER MODEL - KM 071 LIST OF PRODUCTS

DIGITAL EARTH RESISTANCE Model KM 2030

SOIL RESISTIVITY TESTER

1KV DIGITAL MULTIFUNCTION INSULATION RESISTANCE AND CONTINUITY-VOLTAGE TESTER WITH MOV & GAS ARRESTER TESTING. Model 1154TMF

ANALOG INSULATION RESISTANCE TESTER

TEMPERATURE CALIBRATOR Model KM-CAL-801

DIGITAL INSULATION RESISTANCE TESTER (5KV, 1T )

HIGH VOLTAGE PROXIMITY DETECTOR

DIGITAL 4 WIRE EARTH RESISTANCE TESTER

Volt/mA Calibrator. Manual

INSTRUCTION MANUAL DT173 DT175 DIGITAL THERMOMETERS ALWAYS READ THESE INSTRUCTIONS BEFORE PROCEEDING

712 RTD Calibrator. Instruction Sheet. Introduction

Temperature Calibrator

Instruction Manual PC104/PC105 3-PHASE SOCKET TESTER. Check out what else you can get from Martindale:

INSTRUCTION MANUAL CM95/ CM100 TRUE RMS FLEXIBLE CLAMP METER. 1. SAFETY INFORMATION: Always read before proceeding. REMEMBER: SAFETY IS NO ACCIDENT

362 Clamp Meter. Users Manual

INSTRUCTION MANUAL CM69 TRMS AC EARTH LEAKAGE CLAMP METER. 1. SAFETY INFORMATION: Always read before proceeding. REMEMBER: SAFETY IS NO ACCIDENT

Calibrator Capabilities Function Range Resolution Measure V dc 28 V 1 mv Measure ma dc Source ma dc

Users Manual. Phase Rotation Indicator

323/324/325 Clamp Meter

DT33B/C/D OPERATOR S MANUAL LIMITED WARRANTY AND LIMITATION OF LIABILITY

RS-3 PRO RS-1007 PRO. CAT IV Analog Clamp meter Series. Users Manual. For detailed specifications and ordering info go to

LH41A. User Manual. Clamp On Ammeter. For detailed specifications and ordering info go to

i400s Calibration Information Introduction

362 Clamp Meter. Users Manual

51 & 52 Series II. Users Manual. Thermometer. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

OPERATING INSTRUCTIONS

MY /2-DIGIT DIGITAL MULTIMETER Users Manual

302+/303 Clamp Meter. Users Manual

Calibration Manual. Detachable Jaw True-rms Clamp Meter

Infrared Thermocouple Probe

PRM-4. Phase Sequence and Motor Rotation Tester. Users Manual. For detailed specifications and ordering info go to

Instruction Manual VT4 VOLTAGE TESTER. Check out what else you can get from Martindale:

365 Detachable Jaw True-rms Clamp Meter

TX900. Microwave Leakage Detector. 99 Washington Street Melrose, MA Phone Toll Free

716 Pressure Calibrator

Visit us at

Temperature Probe, Battery (installed), Instruction Sheet, Quick Reference Card and Warranty Card.

User's Manual. High Voltage Megohmmeter. Analog Insulation Tester plus AC Voltage and Continuity Tests. Model

1587 FC/1587/1577. Insulation Multimeter. Safety Information. Warnings

1662/1663/1664 FC. Electrical Installation Tester. Safety Information. Warnings

Tachometer. English. Users Manual

True RMS Autoranging Multimeter

User's Guide. True RMS Multim eter. Extech 430

THWD-3 and TH-3 Relative Humidity and Temperature Meters

Autoranging Multimeter


80PK-2A Type K Immersion Probe

PY-11 Voltage & Continuity Tester

LM-100 LM-120. Light Meters. Users Manual. For detailed specifications and ordering info go to

MM300. INSTRUCTION MANUAL Manual-Ranging Digital Multimeter

KMP Insulation-Continuity Tester. Users Manual

Phase Sequence Tester

373 Clamp Meter. Users Manual

OPERATING INSTRUCTIONS

Dawson DDM190D. Digital Multimeter User s Manual

BT508/BT510/BT520/BT521 Battery Analyzer Safety Sheet

Table of Contents. Title Page UT501/UT502 OPERATING MANUAL. Introduction. Unpacking the Meter. Safety Information. International Electrical Symbols

Mini Multimeter with Non-Contact Voltage Detector (NCV)

233 True-rms Remote Display Digital Multimeter Safety Information

Users Manual. MegOhmMeter

NEW 15KV HIGH VOLTAGE INSULATION RESISTANCE TESTER MODEL KM 7015 IN FEATURES : CAT IV 600V. Short Circuit Current 5mA.

The function of this Dynamic Active Probe has divided into three preferences on the screen main Menus:

Go to to register your product and find more information.

INSTRUCTION MANUAL DM5B Fax: (503)

ATD-5570 Deluxe Automotive Meter with RPM and Temperature Functions Owner s Manual Features:

718 30G/100G. Users Manual. Pressure Calibrator. 99 Washington Street Melrose, MA Phone Toll Free

ATD-5540 Digital Automotive Engine Analyzer/Multimeter Owner s Manual

Instruction Manual ICM A1 AC/DC Clamp Meter ICM A1

User Guide. Model Insulation Tester / Megohmmeter. Introduction

TMD90A Dual Input Digital Thermometer

Digital Multimeter User's Manual

i3000s/i2000 Flex AC Current Probe

PTC8001 Temperature Calibrator

Table of Contents Title Page No.

Mini Multimeter with Non-Contact Voltage Detector (NCV)

USER MANUAL. 200A AC Mini Clamp Meter. Model MA130. Please visit for user manual translations

CM100. Carbon Monoxide Meter. 99 Washington Street Melrose, MA Phone Toll Free

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL PSMR1

KMP7010. Ground Resistance Tester. Users Manual

The following table shows the symbols used on the product and/or in this manual. Hazardous Voltage. Risk of electric shock.

AMB V Insulation Resistance Tester. Users Manual. For detailed specifications and ordering info go to

DT304. Digital Temperature Logger INSTRUCTION MANUAL

Instruction Manual. 99 Washington Street Melrose, MA Fax TestEquipmentDepot.com. Model 308A Digital Insulation & Continuity Meter

TIC 300 PRO. Users Manual Mode d emploi Bedienungshandbuch Manual d uso Manual de uso. Non Contact AC Voltage Detector. High Energy Tic Tracer

l The Battery Tester is designed for measuring the l AC four-terminal method to measure the internal

719 Series. Pressure Calibrator with Electric Pump. Product Overview (E)

TIME-10 Process Control Stopwatch

User's Manual. Earth Ground Resistance Tester. Model Introduction

HIGH VOLTAGE PROVING UNIT

TABLE OF CONTENTS. 8.3 Online Sensor Specifications. Series TS (Models TS1, TSP, TSH, TSL,TSF, TSB1, TSB2) According to Electromatic factory procedure

IR-500 Infrared Thermometer / Voltage Detector / Flashlight

Autoranging Industrial Multimeter

Model 2053 Conductivity Meter Instruction Manual

INSPECTOR LINE LOAD SIMULATOR INSTRUCTION MANUAL TASCO, INC.

Instruction Manual. IM-191 Magnetic Field Meter

TACH-10 Contact / Non-Contact Tachometer

Operator Instructions for Tension Links

Transcription:

An ISO 9001:2008 Company THERMOCOUPLE CALIBRATOR Mode KM-CAL-710 Thermocouple Process Calibrator is a exactitude measurement and source handhold instrument, it can be use to calibrate the Thermocouple instrument. Thermocouple Process Calibrator can measure or simulate 8 types of difference Thermocouple ( C or F), and measure or simulate the millivolt. But is could not use to measurement or source at a same time. SPECIAL FEATURES : Maximum voltage applied between any jac and earth ground or between any two jack : 30V Sources and measures 7 thermocouple type devices (J, K, T, E, R, S, B, N) or mv (-10mV ~ 100mV) Large 5 digit LCD with full angel O High accuracy maximum up to 0.3 C and 0.025% for mv With standard jack for external power supply GENERAL SPECIFICATIONS : Thermocouple Type : J, K, T, E, R, S, B, N Temperature Resolution : Measure or Simulating a Thermocouple J,K,T,E,B,R,S,N 0.1 C or F Ambient temperature compensation and display Temperature error : ± (0.3 + 10uV) Operating Temperature : 0 C ~ 50 C Voltage range : -10mV ~ 75mV Storage Temperature : -40 C ~ 60 C Voltage resolution : 0.01mV Operating altitude : 3000 meters maximum. Voltage accuracy : ± (0.025% + 2 counts) Temperature Coefficient : LCD size : 64 x 42mm ± 0.02% / C on 0 C ~ 18 C and 28 C ~ 50 C Power Supply : 1.5V AAA x 6 batteries. Relative humidity : 95%upto 30 C, 75% upto 40 C, External power option 45% upto 50 C Dimension : 205(L) x 97(W) x 45(H)mm Shock : Random 2g. 5Hz to 500Hz. Weight : Approx. 472g. (including battery) Fuse : F 0.125A / 250V ELECTRICAL SPECIFICATIONS : KM-CAL-710 Measure (input) / Simulate (output) Millivolt specification Preliminary Data Input / Output Range Resolution Accuracy -10mV ~ 100mV 0.01mV ±(0.025% + 2 counts) Maximum Input Voltage : 30Vpp. Measure (input) / Simulate (output) Thermocouple Specification Function Range Resolution Accuracy J TYPE K TYPE T TYPE E TYPE -200 ~ 1200 C / -328 ~2192 F -200 ~ 1370 C / -328 ~2498 F -200 ~ 400 C / -328 ~752 F -200 ~ 950 C / -328 ~1742 F 0.1 C / F ±(0.3 C + 10 V) Function Range Resolution Accuracy R TYPE S TYPE B TYPE N TYPE -20 ~ 1750 C / -4 ~3182 F -20 ~ 1750 C / -4 ~ 3182 F 600 ~ 1800 C / 1112 ~3272 F -250 ~ 1300 C / -418 ~2372 F Maximum Input Voltage : 30Vpp. 1 C / F 0.1 C / F ±(1 C + 10 V) ±(1 C + 10 V) ±(1 C + 10 V) ±(0.3 C + 10 V) Reference Junction Error Reference Junction Error Note: All Specification are Subject to change without prior notice. An ISO 9001:2008 Company G-17, Bharat Industrial Estate, T. J. Road, Sewree (W), Mumbai - 400 015. INDIA. Sales Direct : 022-24156638. Tel.: 022-2 4124540, 24181649 Fax : 022-24149659 Email : kusam_meco@vsnl.net, Website : www.kusamelectrical.com Chhaya Computer/D:/Chhaya/Coreldraw Files/New catlogs Dec 2011/KM-CAL-710.cdr

LIST OF PRODUCTS * Digital Multimeter * Digital AC & AC/DC Clampmeter * AC Clamp Adaptor * AC/DC Current Adaptor * Transistorised Electronic Analog & Digital Insulation Resistance Testers(upto 10 KV) * Digital Sound Level Meter & Sound Level Calibrator * Digital contact & Non-contact Type Tachometer * Digital Non-contact (infrared) Thermometer & Portable Infrared Calibrator * Thermo Hygrometer / Anemometer Digital Absolute pressure meter * Wood, Paper & Grain Moisture Meter * Distance Meter & Network Cable Tester * Digital Hand Held Temperature Indicators * Digital Lux Meter * Thermal Imaging Camera * Power Factor Regulator * Maximum Demand Controller/Digital Power Meter * Earth Resistance Tester * Digital Panel Meters & DC Power Supplies * Digital Storage / Analog Storage Oscilloscope. * Coating Thickness Guage * Process Calibrators & Multifunction Calibrators * Gas Analysers & Waterproof Pen Testers * Frequency Counter / Function Generator * Phasing Sticks & High Voltage Detector * Transducer & Transmitter * Digital Milli Ohm Meter * Solar Power Meter * EMF/ELF Detector / RF Field Strength Meter 17, Bharat Industrial Estate, T. J. Road, Sewree (W), Mumbai-400015. INDIA Sales Direct: 24156638 Tel.: (022)2412 4540, 2418 1649 Fax: 2414 9659 E-mail : kusam_meco@vsnl.net, Website : www.kusamelectrical.com, www.kusam-meco.co.in Thermocouple Process Calibrator MODEL KM 710 OPERATION MANUAL KM-CAL-710 Thermocouple Calibrator

Safety Information To avoid possible electric shock or personal injury: Never apply more than 30V between any two jacks, or between any jack and earth ground. Make sure the battery door is closed and latched before you operate the calibrator Remove test leads from the calibrator before you open the battery door. Do not operate calibrator if it is damaged. Do not operate the calibrator around explosive gas, vapor, or dust. To avoid possible damage the calibrator Make sure to choose the right jack and range, before using the calibrator to measurement or calibraton Introduction Thermocouple Process Calibrator is a pecision measurement and source handheld instrument, it can be used to calibrate the Thermocouple instrument. Thermocouple Process Calibrator can measure or simulate 8 types of difference Thermocouples ( C of F) and measure or simulate the millivolt. But it can not be used for measurement & source at a same time. The accessories: 2 pcs Thermocouple plugs (no wire), 6* AAA 1.5V battery, user's manual. If the Calibrator is broken or short of some accessories, please contact the supplier. The following table has showed the technical parameter and function of the calibrator. 01 Specification All the specification will under 1 year calibration cycle and temperature between 18~28 C, except if mentioned otherwise.l Measure (input)/simulate (output) Millivolt specification Input / Output Resolution Accuracy Range -10mV~100mV 0.01mV ± (0.025%+2counts) Maximum input voltage : 30Vpp. Measurement (input)/simulate (output) Thermocouple specification Function Range Resolution Accuracy Reference Junction Error ±(0.3 C+10 V) -200~1200 C/ -328~2192 F -200~1370 C/ -328~2498 F -200~400 C/ -328~752 F -200~950 C/ -328~1742 F -20~1750 C/ -4~3182 F -20~1750 C/ -4~3182 F 600~1800 C/ 1112~3272 F -250~1300 C/ -418~2372 F J TYPE ±(0.3 C+10 V) K TYPE ±(0.3 C+10 V) T TYPE ±(0.3 C+10 V) E TYPE ±(1 C+10 V) 1 C/ F R TYPE Maximum input voltage : 30Vpp. 02 ±(1 C+10 V) 1 C/ F S TYPE ±(1 C+10 V) 1 C/ F B TYPE ±(0.3 C +10 V) N TYPE

INPUT OUTPUT TC TYPE C F Thermocouple Type K J T B E N R S 30V MAX General Specifications : Maximum voltage applied between any jack and earth ground or between any two jack : 30V Storage temperature : -40 C ~ 60 C Operating temperature : 0 C ~ 55 C Operating altitude : 3000 meters maximum Temperature coefficient : ± 0.02% / C on 0 C~18 C and 28 C~50 C Relative humidity : 95% up to 30 C, 75% up to 40 C, 45% up to 50 C Shock : Random 2g, 5Hz to 500Hz. Safety : 1 meter drop test Fuse : F 0.125A/250V Power requirement : 6 Pcs AAA 1.5V Battery Size : 205mm x 98mm x 46mm Weight : approx. 472 g (include battery) Symbol INTERNATIONAL SYMBOLS Earth ground Meanings Explanation on Front Panel The front panel is show as in figure: 1) Input / Output jack 2) Power key 3) Mode key 02 4) C / F key 06 5) Input / Output key 05 6) Increase set value key 07 7) Reduce set value key 01 8) Increase set value key 9) Reduce set value key Understanding Display Screen LCD screen is shown as in following figure POWER DC CAL 8.8.8.8.8 INPUT OUTPUT INPUT/OUTPUT + 8.8.8.8mA mv ma% Thermocouple Calibrator TERMINAL 04 03 08 09 Conforms to European Union directives Refer to this instruction sheet for information about this feature. Battery Double insulation 10) Input state indication 11) Output state indication 12) Indicating AUTO POWER OFF function is enabled. 13) Result value 14) Calibration mode indication 15) Low power indication 16) Unit indication 17) Type indication 18) Ambient temperature display 03 04

INPUT/OUTPUT Operation Instructions. Thermocouple or Millivolt measurement / input 1) Press the power switch 2, turn on the calibrator. 2) Press the input/output key 5, to set the input mode. 3) Press mode key 3, on the measure type you want 4) Put the measure thermocouple or millivolt source into the input jack 1 5) Get the reading value 13. In the thermocouple measurement, ambient temperature value 18 is displayed on LCD. In voltage measurement, there is no temperature display. The number in the, referring to the Explanation on Front panel (Page 4 ) and the Understanding Display Screen (Page 4). INPUT TC TYPE K Tamb 20.0 C 100.0 C INPUT/OUTPUT + TERMINAL 30V MAX Thermocouple Type K J T B E N R S Thermocouple or Millivolt simulate / output 1) Press the power switch 2, turn on the calibrator. 2) Press the input/output key 5, to set the input mode. 3) Press mode key 3, on the measure type you want 4) Press the adjust key 6 7 8 9 to set the value as required. 5) Put the thermocouple instrument or voltage meter into the output jack 1 5) If you want to change the output value, then press the adjust key 6 7 8 9 or change to other thermocouple type use the mode key 3. INPUT TC TYPE K Tamb 20.0 C 1000.0 C + TERMINAL 30V MAX Thermocouple Type K J T B E N R S Auto power OFF Auto power off default setting is 30 min. setting Autopower off option: 1) Keep press 4 C / F key, then turn on the power. 2) Release 4 C / F key, press 6 increase more value key or 7 Reduce more value key to adjust the time. (off, 15min. ~ 60min.) 3) Then press 4 C / F key to finish setting autopower off option. After change battery the autopower off setting is set to default setting If change battery and found that meter can not turn on power, please take off the battery, and wait 3min, then try again. Display all symbol Setting display all symbol: 1) Keep press 3 mode key, then turn on the power. 2) It will display all symbols on LCD. 3) Press any key to exit and enter into measurement / source mode. 05 06

To use Adapter (Only apply to AC power adapter version mentioned in manual). Connecting the power adapter: 1) Connect the AC power cord to the AC-DC converter 2) Plug the AC power cord into an electrical outlet (100V-240V). 3) Plug the DC power plug of the convertor into DC power socket of the meter DC power Socket DC power Plug AC-DC Converter. AC power Cord AC/DC adapter information: Input : 100V-240VAC, 50-60Hz 1.8A Output : DC 12V 2A MAX Polarity : + - WARNING: 1) Please use the original AC power adapter, using other AC power adapter may damage your instrument. 2) The AC power adapter can only be used indoors. 3) Please plug the AC power cord into an electrical outlet first and then firmly insert DC plug into DC input end in the right side of the meter. When unplugged, first pull out the DC plug perpendicular to DC input end and then unplug the AC plug from the electrical outlet. 4) Do not use the AC power adapter in other equipment except this instrument. 5) In use, it is a normal phenomenon that the AC power adapter will be hot. 6) Do not demolish the AC power adapter. Otherwise, it may be dangerous 07 7) Do not use the AC power adapter in a high temperature or wet place. 8) Please make the AC power adapter avoid a strong bump. 9) It is normal when the AC power adapter make some noise in use. MAINTENANCE Cleaning :- Periodically wipe the case with a damp cloth and deterdent; do not use abrasives or solvents. Calibration :- Calibrate your calibrator once a year to ensure that it performs according to its specifications. Replacing the Battery :- Please change the battery when the LCD indicates Turn off the power of the Calibrator, when you change the battery, and screw off the breechblock on the battery cabinet cover, then take off it and replace it with a fresh battery. Replacing a Fuse Warning To avoid personal injury or damage to the calibrator, use only a 0.125A 250V fast fuse. In the thermocouple input mode, if 'OL' does not appears on LCD with no thermocouple input, the fuse is probably blown. A new fuse should be used. Connect wire. Use the accessories thermocouple plug to make the difference plug connect wire which you want. 08

MUMBAI TEST CERTIFICATE THERMOCOUPLE PROCESS CALIBRATOR This Test Certificate warrantees that the product has been inspected and tested in accordance with the published specifications The instrument has been calibrated by using equipment which has already been calibrated to standards traceable to national standards. MODEL NO. KM-CAL-710 SERIAL NO. DATE: WARRANTY Each KUSAM-MECO product is warranted to be free from defects in material and workmanship under normal use & service. The warranty period is one year (12 months) and begins from the date of despatch of goods. In case any defect occurs in functioning of the instrument, under proper use, within the warranty period, the same will be rectified by us free of charges, provided the to and fro freight charges are borne by you. This warranty extends only to the original buyer or end-user customer of a KUSAM-MECO authorized dealer. This warranty does not apply for damaged Ic s, burn't PCB's, fuses, disposable batteries, carrying case, electrodes probes, cables or to any product which in KUSAM-MECO s opinion, has been misused, altered, neglected, contaminated or damaged by accident or abnormal conditions of operation or handling. KUSAM-MECO authorized dealer shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of KUSAM-MECO. KUSAM-MECO s warranty obligation is limited, at option, free of charge repair, or replacement of a defective product which is returned to a KUSAM-MECO authorized service center within the warranty period. ISO 9001 REGISTERED QC KUSAM-MECO PASS THIS WARRANTY IS BUYER S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. KUSAM-MECO SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE WHATSOEVER. All transaction are subject to Mumbai Jurisdiction. 09 10