MODEL 454 OZONE MONITOR

Similar documents
MODEL 452 PROCESS OZONE SENSOR

MODEL 452 PROCESS OZONE SENSOR

Flo-Way. Measure Granular or Powdered Material. Stand Alone or Connect to Other Devices for Blending. Used by Successful Producers World-Wide

A Member of the. Integrated Metering Technologies

IAQ-CALC INDOOR AIR QUALITY METER MODEL 7515

BEAMER Tilt Beam Sensor

A Member of the. Integrated Metering Technologies

Active Controlled Cooling System

MPT-200A SPECIFICATIONS AND OPERATING INSTRUCTIONS

Indoor Air Quality Meter

MICROMANOMETER ALNOR MODEL AXD610 AIRFLOW INSTRUMENTS MODEL PVM610

ValveMate 7000 Controller Operating Manual

Laboratory Series. Ultraviolet Ozone Destruct System Installation, Operation and Maintenance Manual. One Source for All Your UV Needs!

INSPECTOR LINE LOAD SIMULATOR INSTRUCTION MANUAL TASCO, INC.

SPC Series. Digital Scale. Operation Manual

IAQ-CALC TM Indoor Air Quality Meter

50407 / DUAL INDEPENDENT FOOTWEAR AND WRIST STRAP TESTER

COMPRESSED AIR LOADER MODEL: AL-1

Ventilation Testing. Micromanometer. Alnor Model AXD610 AIRFLOW TM Model PVM610. Operation and Service Manual

INSTALLATION, CALIBRATION & TROUBLESHOOTING MANUAL

AIR VELOCITY METER ALNOR MODEL AVM410 AIRFLOW INSTRUMENTS MODEL TA410

OPERATING INSTRUCTIONS PLEASE READ CAREFULLY

IAQ-CALC TM Indoor Air Quality Meter

Broadband PowerShield. CP27U Models. User Manual

QB4 PRESSURE CONTROL VALVE INSTALLATION & MAINTENANCE INSTRUCTIONS

MultiCal. Operation Manual for MultiCal mmhga/psia Pressure Module

i400s Calibration Information Introduction

ITA / ITB - AS / AP / AP SERIES Electronic Table Top Weighing Balance

SECULIFE PS100 PATIENT SIMULATOR

Tension your belts at the speed of light!

MPT-250B SPECIFICATIONS AND OPERATING INSTRUCTIONS

TL-105-D COLUMN HEATER

BroadBand PowerShield. 20 AHr Battery. User Manual

RESISTIVITY MONITOR/CONTROLLERS

IV. PROOF OF PURCHASE: A warranty claim must be accompanied by proof of the date of purchase.

PSC-301 Pump Stroke Counter

DP-CALC TM Micromanometer

Dimensions 12/800N 12/1200N D. DC to AC Power Inverters. OWNERS MANUAL for Models: OWNERS MANUAL April ISO 9001:2000 Certified Company

ADI-125/750 ADI-125/1500 ADI-125/2500

For technical questions and replacement parts, please call

& HIGH CURRENT DC POWER SUPPLIES INSTRUCTION MANUAL

DMG200. Digital Micron Gauge. 99 Washington Street Melrose, MA Phone Toll Free

SKY-WEIGH HELICOPTER CARGO HOOK LOAD MONITORING SYSTEM. User Guide Quality Industrial Weighing and Force Measurement Equipment

Document No D1 Revised December Atlantic Ultraviolet Corporation

Calibrator Capabilities Function Range Resolution Measure V dc 28 V 1 mv Measure ma dc Source ma dc

DMG100 Digital Micron Gauge Limited Warranty DMG100 DMG100 Digital Micron Gauge USA CANADA COPYRIGHT 2011 UEi

1 AMP CURRENT SOURCE

NEO-DYN MODEL 100P ENCLOSURE 7 ADJUSTABLE EXPLOSION-PROOF PRESSURE SWITCH

GF1900 SULFURYL FLUORIDE MONITOR

TRANSDUCER INSTRUCTION MANUAL... TYPE SLIM CELL TRANSDUCER. INSTRUCTION NUMBER: AO of 9

This document describes:

OWNERS MANUAL JANUARY 2007 ISO

Web Volume Control Model WV220

Helios. 24 Light 240 Volt Controller. PRODUCT # Instruction Manual.

MIL-24/2600Q MIL-24/3200DQ

INSTRUCTION MANUAL SURFACE RESISTIVITY METER SRM-110

Optidew Vision Optical Dew-Point Meter

OWNER S MANUAL. Model 9880 Intrinsically Safe Air Velocity Meter

DC REGULATED POWER SUPPLY

Fifth Wheel Power Hitch Operations Manual

GC-1. Roof and Gutter De-Icing Control Installation and Operating Instructions FOR EXTERIOR INSTALLATION ONLY

10 Year Limited Warranty

Installation Instructions for: Channel Thermocouple Amplifier

PCS GEAR SELECT MODULE USER GUIDE v4.0

INSTALLATION AND OPERATION MANUAL

Harris IRT Enterprises Digital Resistance Tester Model XP

AXD 530 MicroManometer

SE-400 MANUAL GROUND-FAULT-RELAY TEST UNIT

Model 8340 Intrinsically Safe VELOCICHECK Air Velocity Meter

TMULD-300 ULD-300. Ultrasonic Leak Detector. Users Manual. 99 Washington Street Melrose, MA Phone Toll Free

VELOCICALC Air Velocity Meter

F-4600 INLINE ULTRASONIC FLOW METER Installation and Operation Guide

Gauge Adapter Instruction Manual

RS-3 PRO RS-1007 PRO. CAT IV Analog Clamp meter Series. Users Manual. For detailed specifications and ordering info go to

Model 809 Diaphragm Gauge Controller

INSTRUCTION MANUAL. Rotary Unidirectional Piezoelectric Motor Evaluation Kit. (Model: UPM-28) Made in USA

VANGUARD A EZ INSTALL OPERATOR S MANUAL

DYNAMOMETER CONTROLLER MODEL 5220 INSTRUCTION AND REFERENCE MANUAL

MMV Digital Length Measuring Gauge INSTRUCTION MANUAL

OWNERS MANUAL JANUARY 2007 ISO

DUAL PURPOSE DIAGNOSTIC LEAK DETECTOR

GFM-40ac COMPLIANCE WESTUSA. Ground Circuit Tester. Instruction Manual

6 OFF ROAD LIGHT BAR

Electronic lube meter

Installation and Operation Guide for PD4100 Series Power Control Centers

Model GD-70D4A-XX Supplement to the GD-70D Operator s Manual

Power. On Your Terms.

will result in death or serious injury. could result in death or serious injury. Indicates a hazardous situation which, if not avoided,

ODY-05-2 VOLTMETER FOR REMOTE MONITORING

Installation and Maintenancee Manual

Installation Instructions

OPERATING INSTRUCTIONS

24/3000H-3PH 24/4500H-3PH 24/6000H-3PH

SELECT DIAGNOSTIC GUIDE. INST028 Doc 3.02

SETRA Operating Instructions Series DPT 230 Pressure Transducers

DM1016S INSTALLATION/OWNER'S MANUAL 10" Marine DVC Subwoofer

1000W RADIOMETRIC POWER SUPPLY

LIBERATOR USER MANUAL

RU BMS Power Supply and Battery Management System Owners Guide

QPV1 Electronic Pressure Regulator INSTALLATION AND MAINTENANCE INSTRUCTIONS

Transcription:

MODEL 454 OZONE MONITOR Teledyne Instruments Advanced Pollution Instrumentation Division (T-API) 9480 Carroll Park Drive San Diego, CA 92121-5201 TOLL-FREE: 800-324-5190 TEL: 858-657-9800 FAX: 858-657-9816 E-MAIL: api-sales@teledyne.com WEB SITE: www.teledyne-api.com

SAFETY MESSAGES NOTE The M454 contains no user serviceable parts. CAUTION There is risk of dangerous UV exposure when the cover is removed from the sensor. Take necessary precautions to avoid exposure. CAUTION High voltages exist inside the monitor. Please use caution when sensor cover is removed. Printed 01/05/2009 i

TABLE OF CONTENTS SAFETY MESSAGES... I TABLE OF CONTENTS...II LIST OF FIGURES... IV LIST OF TABLES... IV WARRANTY POLICY...V 1 PRODUCT DESCRIPTIONS... 1-1 1.1 M454 OZONE MONITOR (1/2 DIN PACKAGE)... 1-1 2 SPECIFICATIONS... 2-1 2.1 MECHANICAL SPECIFICATIONS... 2-1 2.2 PERFORMANCE SPECIFICATIONS... 2-2 2.3 OPERATING LIMITS... 2-2 2.4 ELECTRICAL SPECIFICATIONS... 2-2 2.5 CALIBRATION REFERENCE... 2-2 3 THEORY OF OPERATION... 3-1 4 INSTALLATION AND OPERATION... 4-1 4.1 UNPACKING... 4-1 4.2 MECHANICAL INSTALLATION... 4-1 4.2.1 Bench Top Installation... 4-1 4.2.2 Panel Mount Installation... 4-2 4.3 ELECTRICAL CONNECTIONS... 4-3 4.3.1 AC Power Connection... 4-3 4.3.2 Analog Output... 4-4 4.3.3 Control Inputs... 4-4 4.3.4 Status Outputs... 4-5 4.3.5 Serial Digital Interface (RS232/485)... 4-6 4.4 GAS CONNECTIONS... 4-7 4.5 START-UP AND CALIBRATION PROCEDURE... 4-7 ii

4.6 FRONT PANEL DISPLAY AND KEYPAD... 4-8 4.6.1 4 Digit Display... 4-8 4.6.2 Status LED s... 4-9 4.6.3 Zero Calibration Keys... 4-9 4.6.4 Pressure Display Key... 4-9 5 SENSOR AND SYSTEM TROUBLESHOOTING... 5-1 5.1 SENSOR O.K.... 5-1 5.2 INVALID READING... 5-1 5.3 LAMP LOW... 5-2 5.4 CELL DIRTY... 5-2 5.5 STATUS OUTPUT SUMMARY TABLE... 5-2 iii

LIST OF FIGURES FIGURE 2-1: M454 ½ DIN OZONE MONITOR... 2-1 FIGURE 4-1: BENCH TOP INSTALLATION... 4-1 FIGURE 4-2: PANEL MOUNT INSTALLATION... 4-2 FIGURE 4-3: M454 REAR PANEL... 4-3 FIGURE 4-4: DIGITAL OUTPUT CONNECTIONS... 4-5 FIGURE 4-5: FRONT PANEL... 4-8 LIST OF TABLES TABLE 4-1: STATUS OUTPUTS... 4-6 TABLE 5-1: STATUS OUTPUT TRUTH TABLE... 5-3 iv

WARRANTY POLICY ADVANCED POLLUTION INSTRUMENTATION, INC. 02024c Prior to shipment, API equipment is thoroughly inspected and tested. Should equipment failure occur, API assures its customers that prompt service and support will be available. COVERAGE After the warranty period and throughout the equipment lifetime, API stands ready to provide on-site or in-plant service at reasonable rates similar to those of other manufacturers in the industry. All maintenance and the first level of field troubleshooting are to be performed by the customer. NON-API MANUFACTURED EQUIPMENT Equipment provided but not manufactured by API is warranted and will be repaired to the extent and according to the current terms and conditions of the respective equipment manufacturers warranty. GENERAL API warrants each Product manufactured by API to be free from defects in material and workmanship under normal use and service for a period of one year from the date of delivery. All replacement parts and repairs are warranted for 90 days after the purchase. If a Product fails to conform to its specifications within the warranty period, API shall correct such defect by, in API's discretion, repairing or replacing such defective Product or refunding the purchase price of such Product. The warranties set forth in this section shall be of no force or effect with respect to any Product: (i) that has been altered or subjected to misuse, negligence or accident, or (ii) that has been used in any manner other than in accordance with the instruction provided by API or (iii) not properly maintained. THE WARRANTIES SET FORTH IN THIS SECTION AND THE REMEDIES THEREFORE ARE EXCLUSIVE AND IN LIEU OF ANY IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS FOR v

PARTICULAR PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESSED OR IMPLIED. THE REMEDIES SET FORTH IN THIS SECTION ARE THE EXCLUSIVE REMEDIES FOR BREACH OF ANY WARRANTY CONTAINED HEREIN. API SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES ARISING OUT OF OR RELATED TO THIS AGREEMENT OF API'S PERFORMANCE HEREUNDER, WHETHER FOR BREACH OF WARRANTY OR OTHERWISE. TERMS AND CONDITIONS All units or components returned to API should be properly packed for handling and returned freight prepaid to the nearest designated Service Center. After the repair, the equipment will be returned, freight prepaid. vi

1 Product Descriptions 1.1 M454 Ozone Monitor (1/2 DIN Package) The API M454 is a microprocessor-based monitor for measuring the concentration of gaseous ozone in processes such as semiconductor wafer fabrication, water treatment, and ozone research. The M454 can be used as a full flow process sensor or as a sensor to monitor a small flow of gas diverted from a process stream. The ½ DIN package can be panel mounted (panel mount hardware is included) or operated as a bench-top monitor. The M454 features a 4-digit front panel display/keypad as well as a standard 0-5 volt (optional isolated 4-20mA) analog output for reporting ozone concentration. A bi-directional serial interface (RS232 or RS485) is also provided for computer control and data acqusition. A front panel membrane keypad provides buttons for zero calibration and cell pressure display. Four digital status outputs are provided for sensor diagnostics. These are displayed as front panel indicators and are provided as opto-isolated outputs on the rear panel I/O connector. 1-1

INTENTIONALLY BLANK 1-2

2 Specifications Note: All specifications contained herein are subject to change without notice. Please contact API to obtain the current specifications. 2.1 Mechanical Specifications 7.200 [182.9].206 [5.2] 2X 1.167 [29.6] 5.963 [151.5] X.XX = Dimensions in inches [XX.X] = Dimensions in millimeters.500 [12.7] 7.611 [193.3] 3.831 [97.3] 3.558 [90.4].137 [3.5] Figure 2-1: M454 ½ DIN Ozone Monitor Weight: Gas Fittings: Wetted Materials: 4.0 lbs. (1.80 kg) M454 ¼ Gyrolok Type 316 Stainless Steel, Sapphire, Virgin PTFE 2-1

2.2 Performance Specifications Accuracy: ±1% of Full Scale Repeatability: 1% of Full Scale. Response Time: 2 sec. to 95% Zero Drift 1% Full Scale/month (non cumulative) Display: 4 Digit LED 2.3 Operating Limits Measurement Range: 5, 10, 15, 20% w/w 25, 50, 100, 200, 400 g/nm3 Proof pressure: 115 psia Flow: 0.1-5 L/min Temperature range: 5 to 45 C Warm-Up Period: 15 minutes 2.4 Electrical Specifications Power Input: Analog Output: Zero Cal: Digital Outputs: 90-265 VAC, 47-63 Hz, 30 VA 0-5V Full Scale (4-20mA optional) Contact Closure Input Sensor OK, Invalid Reading, Lamp Low, Cell Dirty Serial Data Interface: RS232 or RS485, Half-Duplex, 9600 Baud 2.5 Calibration Reference Span Calibration: Standard Temperature and Pressure (g/nm3 only): Traceable to Buffered KI laboratory calibration 0 C and 760 mmhg 2-2

3 Theory of Operation The detection of ozone molecules is based on absorption of 254 nm UV light due to an internal electronic resonance of the O 3 molecule. The Model 454 uses a mercury lamp constructed so that a large majority of the light emitted is at the 254 nm wavelength. Light from the lamp shines through an absorption cell through which the sample gas being measured is passed. The ratio of the intensity of light passing through the gas to a reference measurement, which does not pass through the gas, forms the ratio I/I o. This ratio forms the basis for the calculation of the ozone concentration. The Beer-Lambert equation, shown below, calculates the concentration of ozone from the ratio of light intensities. C O 1 Τ 14.695 psi = ln o α 273 Κ Ρ 3 l Ι Ι o Where: I I o α l C O3 T P = Intensity of light passed through the sample = Intensity of light through sample free of ozone = Absorption coefficient = Path length = Concentration of ozone = Sample temperature in degrees Kelvin = Pressure in pounds per square inch (absolute) As can be seen the concentration of ozone depends on more than the intensity ratio. Temperature and pressure influence the density of the sample. The density changes the number of ozone molecules in the absorption cell, which impacts the amount of light, removed from the light beam. These effects are addressed by directly measuring temperature and pressure and including their actual values in the calculation. The absorption coefficient is a number that reflects the inherent ability of ozone to absorb 254 nm light. Lastly, the absorption path length determines how many molecules are present in the column of gas in the absorption cell. 3-1

The intensity of light is converted into a voltage by the detector/preamp module. The voltage is converted into a number by a voltage-to-frequency (V/F) converter capable of 80,000 count resolution. The digitized signal, along with the other variables, is used by the CPU to compute the concentration of ozone using the above formula. 3-2

4 Installation and Operation 4.1 Unpacking Upon receiving the M454 please verify that no apparent shipping damage has occurred. If damage has occurred please advise shipper first, then API. 4.2 Mechanical Installation 4.2.1 Bench Top Installation Affix the four(4) self-adhesive rubber feet to the bottom of the chassis as shown in Figure 4-1 below. 4X SELF-ADHESIVE RUBBER FOOT (SUPPLIED WITH UNIT) Figure 4-1: Bench Top Installation 4-1

4.2.2 Panel Mount Installation The M454 can be mounted in a standard ½ DIN panel cut-out as shown below in Figure 4-2. Simply slide the M454 into the cut-out from the front. Next, slide the Panel Mount Clamps into the grooves on either side of the chassis. Next, thread the Clamp Screws into the rear openings of the grooves until the Panel Mount Clamps are snug against the rear of the panel. Thread locking compound can be used on the screws to prevent them from vibrating loose. 2X PANEL MOUNTING AND ADJUSTMENT CLAMP SCREW (SUPPLIED WITH UNIT) 2X PANEL MOUNTING AND ADJUSTMENT CLAMP (SUPPLIED WITH UNIT) MOUNTING PANEL (CUSTOMER SUPPLIED) Figure 4-2: Panel Mount Installation 4-2

4.3 Electrical Connections All electrical connections to the M454 are accessible on the rear panel of the instrument. ADVANCED POLLUTION INSTRUMENTATION, INC MODEL 454 POWER IN OUT ANLG CTRL OUT IN + - 1 2 STATUS SERIAL OUT OUT AUX 1 2 3 4 C A B 1 2 3 90-265 VAC 47-63 Hz 30 VA Figure 4-3: M454 Rear Panel 4.3.1 AC Power Connection AC power should be connected to the IEC320 receptacle on the rear panel. Any AC voltage from 90-265 VAC may be used to power the M454. CAUTION Make sure that the power cord is connected to a properly grounded AC outlet. 4-3

4.3.2 Analog Output The analog output is a 0-5 volt signal representing the ozone concentration measured by the sensor. This output is provided on the 16-pin modular connector and is labeled ANLG OUT (See Figure 4-3). The output is scaled to the concentration range that the sensor has been configured to measure. Check the serial number label on the M454 to determine the concentration range. For best performance, the analog output should be connected to a voltmeter or A/D converter with a differential input and a minimum input impedance of 2KΩ. If the isolated current output has been installed, this output will be a 4-20mA current source. 4.3.3 Control Inputs Two control inputs are provided on the 16-pin modular connector and are labeled CTRL IN (See Figure 4-3). Control Input #1 is used for remote zero calibration. To zero the M454, Control Input #1 should be connected to the ground pin provided for at least 1 second. This can be accomplished using a Normally Open switch or relay. Control Input #2 is used to activate the pressure display on the front panel. Connecting Input #2 to the ground pin will switch the front panel readout to display the cell pressure. When this input is released, the normal concentration display will resume. This function does not affect the analog output. The analog output will continue to reflect the ozone concentration while the pressure is being displayed on the front panel. 4-4

4.3.4 Status Outputs Four digital status outputs are provided on the 16-pin modular connector and are labeled STATUS OUT (See Figure 4-3). The outputs are labeled 1-4 and the common ground pin for the outputs is labeled C. The four outputs are used for indicating error status and when operational parameters have moved out of normal limits. These outputs are in the form of opto-isolated open-collector transistors. They can be used to drive status LED s on a display panel or interface to a digital device such as a Programmable Logic Controller (PLC). Figure 4-4 below shows the most common way of connecting the digital outputs to an external device such as PLC. Note: Most devices, such as PLC s, have internal provision for limiting the current that the input will draw from an external device. When connecting to a unit that does not have this feature, external dropping resistors must be used to limit the current through the transistor output to 50mA or less. M452 Programmable Logic Controller or other device +5V Digital Output #1-4 (Collector) Digital Output Common (Emmiter) Digital Input Opto-Isolator Ground Provided by PLC Figure 4-4: Digital Output Connections 4-5

Table 4-1 below describes the function of the status outputs. Table 4-1: Status Outputs Output # Name On State Off state 1 Sensor O.K. 2 Invalid Reading 3 Lamp Low Normal State Pressure > 45 psia, Negative Ozone Concentration Full-scale concentration range exceeded Reference Detector<600mV 4 Cell Dirty Measure/Reference ratio < 0.5 (zero gas) Reference or Measure > 4995mV; Reference < 400mV Normal State Normal State Normal State See Section 5 for details on using the Status Outputs for diagnosing sensor and system-level malfunctions. 4.3.5 Serial Digital Interface (RS232/485) The M454 features a bi-directional digital serial interface that can be used for sensor control and data acquisition. The interface can be configured for RS232 or RS485 operation. Please contact API for documentation on the use of the RS485 interface. 4-6

4.4 Gas Connections Gas connections to the M454 are made using ¼ compression tube fittings. The ¼ compression fittings can be used with ¼ O.D. Stainless Steel or Teflon tubing. Connect the ozone supply tubing to the inlet fitting labeled IN. Connect the exhaust tubing to the outlet fitting labeled OUT. Note that an external flow control device must be used to ensure that the flow rate into the monitor is maintained in the range of 0.1 to 5.0 L/min. CAUTION The exhaust gas from the monitor may contain dangerous levels of ozone. Make sure that the gas connections are leak-tight and the exhaust line is connected to a suitable ozone destruct unit. To avoid contamination of the optical cell in the M454, ensure that all tubing upstream of the M454 is properly cleaned and purged before the M454 is installed. In order to achieve an acceptable response time and to avoid sample degradation, the system should be set up so that a minimum flow rate of 0.1 LPM is established through the M454. If long tubing runs are used between the measurement point and the M452, then higher flow rates should be used to avoid sample degredation. Appropriate tests should be conducted to determine minimum flow requirements. 4.5 Start-Up and Calibration Procedure 1. Verify that the proper electrical connections have been made (See Section 4.3) and apply power to the M454. Allow the M454 to warm up for at least 15 minutes. 2. Purge the M454 with zero gas (usually oxygen) at a minimum flow rate of 0.5 LPM for at least 2 minutes. 3. Check that the Status Outputs (See Section 4.3.4) or Status LED s (See Section 4.6.2) are in their normal states and no errors are indicated. 4-7

4. Close the zero calibration input (See Section 4.3.3) or press the zero keys (See Section 4.6.3) on the front panel to perform the automatic zero calibration. 5. Re-check the Status Outputs or Status LED s to ensure that no errors are indicated. 6. Check the voltage on the analog output (See Section 4.3.2) and verify that it reads 0.000 ± 0.010 volts. 7. The M454 is now ready for operation. 4.6 Front Panel Display and Keypad The M454 front panel features a 4-digit display, four Status LED s, and push-button keys for performing zero calibration and displaying cell pressure. Figure 4-5 below shows the front panel of the M454. SENSOR OK % INVALID READING LAMP LOW CELL DIRTY ZERO PRESSURE OZONE MONITOR - MODEL 454 Figure 4-5: Front Panel 4.6.1 4 Digit Display The 4-digit display gives a continuous, real-time display of ozone concentration in units of weight percent (%), grams per cubic meter (g/m3), or grams per Normal (or Standard) cubic meter (g/nm3). The Standard conditions for the grams per Normal cubic meter units are 0 C (273 K), and 760 mmhg. The selection of measurement units is a factory configuration. 4-8

4.6.2 Status LED s Four status LED s are displayed on the front panel. The green Sensor OK LED should be on during normal operation, and the other three red error LED s should be off. The function of these LED s is identical to the Status Outputs described in Section 4.3.4 and Section 5. 4.6.3 Zero Calibration Keys To perform a zero calibration, press both Zero keys simultaneously. The display should momentarily show four dashes (----), indicating that the zero calibration has been performed. See Section 4.5 for details on all the steps required to successfully perform a zero calibration. 4.6.4 Pressure Display Key Pressing the Pressure key will display the cell pressure in psia (pounds per square inch absolute). This is a momentary function and the real-time ozone concentration display will resume when the key is released. This function does not affect the analog output. The analog output will continue to reflect the ozone concentration while the pressure is being displayed on the front panel. 4-9

INTENTIONALLY BLANK 4-10

5 Sensor and System Troubleshooting This chapter gives guidelines for diagnosing system and sensor malfunctions using the four digital Status Outputs or front panel Status LED s. All troubleshooting should be done after the M454 has been turned on and allowed to warm up for at least 15 minutes. 5.1 Sensor O.K. The normal state for the Sensor O.K. output in ON. During the warm-up period on start-up this output will stay off until the UV lamp reaches a minimum intensity. If this output remains off after the 15 minute warm-up period, or goes off during normal operation, then the M454 is in need of servicing. If the Sensor O.K. output turns off AND the Lamp Low output is on, this indicates that the lamp intensity has below the minimum level required for proper operation. If the Sensor O.K. output turns off and the Lamp Low output is also off, then one of the analog voltages in the sensor has exceeded the range of the internal A/D converter. Adjustment by qualified service personnel is required. 5.2 Invalid Reading The normal state for the Invalid Reading output is OFF. If this output turns on, this indicates that the M454 is still operational, but a system fault or calibration fault exists that may make the current ozone reading invalid. The Invalid Reading output is turned on when the measured pressure in the M454 exceeds 45 psia, the sensor is indicating an excessive negative reading, or the concentration has exceeded the full-scale range of the sensor. 5-1

5.3 Lamp Low The normal state for the Lamp Low output is OFF. If this output turns on, this indicates that the UV lamp intensity as measured by the reference detector has dropped below 600mV. If the Lamp Low output turns ON and the Sensor O.K. output is ON, this indicates that the lamp intensity is still adequate for measurement, but adjustment should be made when possible. If the Lamp Low output turns ON and the Sensor O.K. output is OFF, this indicates a failure condition and measurement is no longer possible. 5.4 Cell Dirty The normal state for the Cell Dirty output is OFF. If this output turns on, then the ratio of the measure detector to the reference detector (at zero) is < 0.5. This value is calculated when the zero calibration is performed. When this output is on, it indicates a loss of optical transmission through the windows in the absorption cell or a calibration fault. 5.5 Status Output Summary Table Table 5-1 is a logic truth table summarizing the recommended actions based on the states of the four status outputs. A 1 indicates the output is ON, a 0 indicates the output is OFF, and X indicates the output is in either state. 5-2

Table 5-1: Status Output Truth Table Sensor OK Invalid Reading Lamp Low Cell Dirty 1 0 0 0 Actions Normal operation, no action required 0 X X X Service required 1 1 X X 1 X 1 X 1 X X 1 Check Pressure > 45 psia Calibrate at Zero Check for concentration over-range Lamp adjustment useful, though not required Calibrate at zero Clean Cell 5-3

INTENTIONALLY BLANK 5-4