Analysis of nitrogen isotope ratios in amino acids by GC/C/IRMS

Similar documents
Molecular Weights and Residue Weights of Protected Amino Acids

Fast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC

Stability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas

Supporting information

Setting up SilFlow for BackFlush in your GC

Method Development for Capillary GC Systems. Slide 1

Agilent 7693A Automated Liquid Sampler

Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems

AppNote 1/2010 KEYWORDS ABSTRACT. Biodiesel, Automation, ASTM D

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

Analysis of Fatty Acid Methyl Esters (FAMES), and Examination of Biodiesel Samples for these Components, by GCxGC-FID

TECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure

Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis

GLYCERIN PACKAGE 100mL $ Glycerin, Ibs/gallon, MONG Material Organic, Non-Glycerin, Ash, Specific Gravity, Moisture by Karl Fisher, Methanol

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

High-Temperature Simulated Distillation System Based on the 6890N GC Application

Application Note. Abstract. Authors. Environmental Analysis

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585

MET-Biodiesel Capillary GC Columns

Technical Procedure for Gas Chromatography (GC-FID)

AppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS

URB '-carbamoylbiphenyl-3-yl cyclohexylcarbamate. DEA Reference Material Collection. Form Chemical Formula Molecular Weight Melting Point ( o C)

Analysis of Petroleum Fractions by ASTM D2887

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System

Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter

Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS)

Application Note. Author. Introduction. Energy and Fuels

ASTM D 6730 Detailed Hydrocarbon Analysis

Fausto Munari e Andrea Cadoppi ThermoFisher - Italy

Determination of Free and Total Glycerin in B100 Biodiesel

Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC

Agilent Solutions for the Analysis of Ethanol- Based Fuels Derived From Biomass. James D. McCurry, Ph.D. Senior Scientist

The Only No Compromise

Agilent 6850 Series II Network GC System G2630A Performance Specifications

Hitachi High-Speed Amino Acid Analyzer

Agilent 7693A Automated Liquid Sampler

Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector

Application Note. Author. Abstract. Energy & Chemicals - Petrochemicals. Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany

INSTALLATION MANUAL for AAA DIRECT

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

6890 CHECKOUT PROCEDURE TCD (THERMAL CONDUCTIVITY) DETECTOR

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

Installing the 6850 Direct Valve Column Connector Kit

HPLC SYSTEM S 500 SERIES

PATROL UPLC Process Analysis System for Production

GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method

Agilent 1290 Infinity Autosampler

Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584

Refinery Support from the R&D Laboratory Perspective using Fast & Micro Gas Chromatography

Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.

Agilent and ASTM. Update on Recent Activities. Page 1

Using the PSD for Backflushing on the Agilent 8890 GC System

Agilent Multimode Inlet for Gas Chromatography

The next level of True Blue Performance

Practical Steps in GC Troubleshooting

Jaap de Zeeuw Restek Middelburg, The Netherlands. Copyrights: Restek Corporation

GC Analysis of PCBs in Fish Oil by Direct Injection Without Any Sample Cleanup. Philip L. Wylie, Ph.D. Sr. Applications Chemist Agilent Technologies

Detection of Volatile Organic Compounds in Gasoline and Diesel Using the znose Edward J. Staples, Electronic Sensor Technology

VERITY Purification Systems Build the perfect system to fit your research

GCMS-QP2010 Series. Consumables. Shimadzu Gas Chromatograph/Mass Spectrometer

S 1125 HPLC Pump System

Large Volume Injection of Polycyclic Aromatic Hydrocarbons

Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE

Technical Overview. Introduction

Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut

Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution

Analysis of Glycerin and Glycerides in Biodiesel (B100) Using ASTM D6584 and EN Application. Author. Abstract. Introduction

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

Phenomenex Exclusive! Inlet Base Seals. Agilent 5000, 6000 and. Guaranteed fit for Series GCs. Easy Seals

HPLC System S 600 Series

Supporting Information. for

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC

Techniques for Making Your GC Analysis More Repeatable, Reproducible and Robust

Study of viscosity - temperature characteristics of rapeseed oil biodiesel and its blends

Optimized Method for Analysis of Commercial and Prepared Biodiesel using UltraPerformance Convergence Chromatography (UPC 2 )

Application Note. Authors. Abstract. Energy & Chemicals

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column

GC Method Compliance and Large Valve Oven Application

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels

Complete Fractionation of Extractable Petroleum Hydrocarbons Using Newly Developed EPH SPE Cartridges

Metathesis Catalysts

GC Inlets. An Overview. Simon Jones GC Applications Engineer

The Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over

Techniques for Making Your GC Analysis More Repeatable, Reproducible and Robust

Agilent 1290 Infinity II Ultra Low Dispersion Kit

For more information about Agilent Technologies services please visit our web site using the following URL

Techniques for Improving the Reproducibility of GC Analysis

Choose for Your Instrument

DETERMINATION OF N-BUTANOL AND ISOBUTANOL IN GASOLINE USING GAS CHROMATOGRAPHY (GC-FID)

HPLC Model 210D. UV detector lamp switch. Rheodyne 7725i injection valve. Prime/purge valve. Column. compartment. thumbscrews. Storage.

Agilent Multimode Inlet

Sulfur Detection at ppb Levels in Light Hydrocarbon Streams

Transcription:

Analysis of nitrogen isotope ratios in amino acids by GC/C/IRMS 27 March 2013 Yoshito Chikaraishi The initial version was released by YC in May 2009. The present version reflects feedbacks from Svenja Kruse (Univ. of British Columbia) and Shawn A. Steffan (Univ. of Wisconsin-Madison), Elizabeth Bosak (Univ. of Wisconsin-Madison) in March 2012.

Instrument Instruments in my lab: GC Agilent 6890N Injector Gerstel PTV Column Agilent HP Ultra-2 (50m, 0.32mm, 0.52µm) Flow rate (He) 1.4ml/min Oven program 40ºC (2.5 min) -15ºC/min - 110 (0 min) - 3ºC/min - 150 (0 min) - 6ºC/min - 220ºC (17.3min) Interface GC combustion III (oxidation at 1000ºC, reduction at 550ºC) IRMS Delta plus XP Standard mix of authentic amino acids: after HCl hydrolysis and derivatization, derivatives are stored at 20ºC, in DCM (Currently, I use new standard mix. See page 10 ). Compound Source Known Amino acid Pv/OiPr derivatives δ 15 N MW N(wt%) mg/10ml MW N(wt%) ng/µl ngn/µl nmol/µl L-alanine Ala CERKU-02 22.7 89 16 2.4 215 6.5 579.9 37.7 2.7 glycine Gly CERKU-03 2.2 75 19 2.1 201 7.0 562.9 39.2 2.8 L-valine Val BioGeos 3.5 117 12 3.6 243 5.8 747.8 43.1 3.1 L-leucine Leu CERKU-04 8.8 131 11 3.7 257 5.4 725.9 39.5 2.8 L-norleucine Ile BioGeos 18.0 131 11 4.4 257 5.4 863.3 47.0 3.4 L-aspartic acid Asp BioGeos 3.1 133 11 4.0 301 4.6 905.6 42.1 3.0 L-serine Ser BioGeos 2.4 105 13 3.2 315 4.4 960.2 42.7 3.0 L-glutamic acid Glu BioGeos -5.7 147 10 4.8 315 4.4 1028.9 45.7 3.3 L-phenylalanine Phe BioGeos -5.2 165 8 5.0 291 4.8 882.0 42.4 3.0 L-hydroxyproline Hyp BioGeos -7.1 131 11 4.7 341 4.1 1223.7 50.2 3.6

Procedure for δ 15 N analysis 1. Set up GC/IRMS (e.g., connect GC and IRMS, focusing) 2. Check N 2 leaks for both back-flush and straight mode 3. Check stability by On/OFF of Ref. gas 4. Check reproducibility by measuring δ 15 N-known authentic AAs (amino acid standards), as Pv/OiPr derivatives 5. Make up a sequence file 6. Measure both authentic AAs and samples of interest 7. Check baseline separation on the chromatograms 8. Normalize the measured values (i.e., δ 15 N, vs ref.gas) to international scale (i.e., δ 15 N, vs Air)

1. Focusing Check these 9 boxes To find the best (optimum) position, apply the following steps: 1. After peak center, open auto focus. 2. Enter 50 and 500 for the step width in these two boxes 3. Click OK 4. After peak center, enter 25 and 250 in these two boxes, and click OK 5. After peak center, enter 12 and 120 in these two boxes, and click OK 6. After peak center, 6 and 60 in these two boxes, and click OK 7. After peak center, enter 3 and 30 in these two boxes, and click OK 8. After peak center, enter 2 and 20 in these two boxes, and click OK 9. After peak center, enter 1 and 10 in these to boxes, and click OK (twice)

3. Check stability by On/OFF of Ref. gas Generally, 1σ < 0.03-0.05

4. Check reproducibility by δ 15 N of known authentic AAs Set the isotope value of the ref. gas to 0 on the PC software, measure the authentic AA derivatives, and check the linearity of the relationship between the measured δ 15 N (, vs ref. gas) and known δ 15 N (, vs air) values. 2500 Ref. N 2 Ref. N 2 m/z 28 Intensity (mv) 2000 1500 1000 500 Ala Gly Val Leu Nle Asp Ser Glu Phe Hyp 0 10 15 20 25 30 35 40 45 Retention time (min) 25 If you do not get a good linear relationship, please check the following points: 1. the condition of the ox./red. furnace 2. the condition of GC column 3. dirt in inlet/liner δ 15 N, vs Air 20 15 10 5 0 y = 1.0225 x 4.4523 (R 2 = 0.999) 4. deterioration of capillary connection -5-10 -5 0 5 10 15 20 25 30 δ 15 N, vs refs

5. Make up a sequence file For example, Line GC method Sample Num. Injection vol. (µl) Sample name IRMS Method 1 Ultra-2 AA 1 1 STD Pv/OiPr Ultra-2_AA_15N 2 Ultra-2 AA 2 1 Sample-A Ultra-2_AA_15N 3 Ultra-2 AA 3 1 Sample-B Ultra-2_AA_15N 4 Ultra-2 AA 4 1 Sample-C Ultra-2_AA_15N 5 Ultra-2 AA 5 1 Sample-D Ultra-2_AA_15N 6 Ultra-2 AA 1 1 STD Pv/OiPr Ultra-2_AA_15N 7 Ultra-2 AA 6 1 Sample-E Ultra-2_AA_15N 8 Ultra-2 AA 7 1 Sample-F Ultra-2_AA_15N 9 Ultra-2 AA 8 1 Sample-G Ultra-2_AA_15N 10 Ultra-2 AA 9 1 Sample-H Ultra-2_AA_15N 11 Ultra-2 AA 10 1 Sample-I Ultra-2_AA_15N 12 Ultra-2 AA 1 1 STD Pv/OiPr Ultra-2_AA_15N

8. Normalization of measured values to the international scale -1 When you set the isotope value of the ref. gas to 0 on the PC software, you have measured values ( vs ref. gas). Known δ 15 N Measured value = δ 15 N, vs ref. gas 1st 2nd 3rd 4th 5th Average 1σ Ala 22.7 26.234 26.755 26.489 26.422 26.667 26.5 0.2 Gly 2.2 6.777 6.101 6.648 6.973 6.207 6.5 0.4 Val 3.5 7.748 7.616 7.624 7.490 6.532 7.4 0.5 Leu 8.8 13.488 13.231 13.254 13.781 13.622 13.5 0.2 Nle 18.0 22.640 22.048 21.853 21.855 21.530 22.0 0.4 Asp 3.1 6.915 7.643 7.299 6.573 7.166 7.1 0.4 Ser 2.4 6.189 6.754 6.454 7.049 6.163 6.5 0.4 Glu -5.7-1.395-1.025-1.331-1.373-1.263-1.3 0.1 Phe -5.2-0.533-0.138-0.836-1.161-0.536-0.6 0.4 Hyp -7.1-2.366-1.654-2.381-2.811-2.357-2.3 0.4 δ 15 N, vs Air 25 20 15 10 5 0-5 y = 1.0225 x 4.4523 (R 2 = 0.999) -10-5 0 5 10 15 20 25 30 δ 15 N, vs refs You can normalize δ 15 N to the international scale (i.e., δ 15 N, vs Air) by using the equation: y = 1.0225x 4.4523 (in this case). The slope 1.0 means that we need this linear normalization instead of an one-point normalization. Known δ 15 N δ 15 N, vs Air 1st 2nd 3rd 4th 5th Average 1σ Δ Ala 22.7 22.4 22.9 22.6 22.6 22.8 22.7 0.2 0.0 Gly 2.2 2.5 1.8 2.3 2.7 1.9 2.2 0.4 0.0 Val 3.5 3.5 3.3 3.3 3.2 2.2 3.1 0.5-0.4 Leu 8.8 9.3 9.1 9.1 9.6 9.5 9.3 0.2 0.5 Nle 18.0 18.7 18.1 17.9 17.9 17.6 18.0 0.4 0.0 Asp 3.1 2.6 3.4 3.0 2.3 2.9 2.8 0.4-0.3 Ser 2.4 1.9 2.5 2.1 2.8 1.8 2.2 0.4-0.2 Glu -5.7-5.9-5.5-5.8-5.9-5.7-5.8 0.2-0.1 Phe -5.2-5.0-4.6-5.3-5.6-5.0-5.1 0.4 0.1 Hyp -7.1-6.9-6.1-6.9-7.3-6.9-6.8 0.4 0.3 Average 0.4 0.0 Std.dev 0.1 0.3 1σ = Precision Δ (= average known) = Trueness In this way, Accuracy (= Mean of Δ) should be Zero ± something.

8. Normalization of measured values to the international scale -2 With the equation: (δ 15 N, vs Air) = 1.0225 (measured δ 15 N, vs refs) 4.4523 (in this case), you can calculate normalized δ 15 N (, vs Air) for the samples of interest. Measured value = δ 15 N, vs ref. gas Sample-A Sample-B Sample-C Sample-D Sample-E Sample-F Sample-G Ala 8.344 10.560 6.005 8.550 13.295 12.521 10.345 Gly 7.655 11.150-2.458-3.452 4.410 5.170 9.723 Val 7.498 11.716 7.959 7.724 13.793 10.741 12.925 Leu 4.123 10.024 5.027 3.161 6.955 6.710 11.654 Ile 4.514 9.749 7.650 3.586 11.191 8.898 15.082 Pro 14.228 24.680 8.891 10.548 17.185 15.903 16.704 Ser 5.188 too small 2.288-1.058 5.942 5.843 8.308 Met too small 7.335 too small too small too small too small 2.555 Glu 11.485 15.796 6.897 7.707 14.695 14.040 18.334 Phe 11.388 15.749 13.942 15.757 13.667 15.951 10.302 δ 15 N, vs Air Sample-A Sample-B Sample-C Sample-D Sample-E Sample-F Sample-G Ala 4.1 6.3 1.7 4.3 9.1 8.3 6.1 Gly 3.4 6.9-7.0-8.0 0.1 0.8 5.5 Val 3.2 7.5 3.7 3.4 9.6 6.5 8.8 Leu -0.2 5.8 0.7-1.2 2.7 2.4 7.5 Ile 0.2 5.5 3.4-0.8 7.0 4.6 11.0 Pro 10.1 20.8 4.6 6.3 13.1 11.8 12.6 Ser 0.9 n.d. -2.1-5.5 1.6 1.5 4.0 Met n.d. 3.0 n.d. n.d. n.d. n.d. -1.8 Glu 7.3 11.7 2.6 3.4 10.6 9.9 14.3 Phe 7.2 11.6 9.8 11.7 9.5 11.9 6.1

New standard mix We currently developed wide-range (from 25.9 to +45.6 ) isotopic reference materials of amino acids, which are commercially available from Indiana University (http://php.indiana.edu/~aschimme/hc.html) and SI science (http://www.si-science.co.jp/global/en/). Compoud Source known δ 15 N Amino acid Pv/OiPr derivatives MW N(wt%) mg/10ml MW N(wt%) ng/µl ngn/µl nmol/µl L-alanine Ala Indiana Univ. 43.3 89 16 6.5 215 6.5 652 42 3.0 glycine Gly SI science -25.9 75 19 6.2 201 7.0 618 43 3.1 L-valine Val BioGeos 3.5 117 12 7.3 243 5.8 727 42 3.0 L-leucine Leu SI science 5.6 131 11 7.7 257 5.4 765 42 3.0 L-norleucine Nle SI science 18.2 131 11 7.7 257 5.4 765 42 3.0 L-aspartic acid Asp SI science 35.3 133 11 9.1 301 4.6 906 42 3.0 L-methionine Met SI science -3.9 149 9 9.0 305 4.6 901 41 3.0 L-glutamic acid Glu SI science 45.6 147 10 9.4 315 4.4 943 42 3.0 L-phenylalanine Phe Indiana Univ. 1.7 165 8 8.8 291 4.8 882 42 3.0 L-hydroxyproline Hyp SI science -9.1 131 11 10.2 341 4.1 1015 42 3.0 Known δ 15 N δ 15 N (, vs Air) by GC/IRMS δ 15 N1st δ 15 N2nd δ 15 N3rd δ 15 N4th Average 1σ Δ Ala 43.3 43.0 43.0 43.5 43.7 43.3 0.4 0.0 Gly -25.9-25.9-26.1-26.2-26.9-26.3 0.5-0.4 Val 3.5 3.2 2.7 3.2 3.3 3.1 0.2-0.4 Leu 5.6 5.5 5.8 6.2 5.5 5.7 0.3 0.1 Nle 18.2 19.0 18.4 18.4 19.3 18.8 0.4 0.6 Asp 35.3 34.4 35.1 35.1 34.6 34.8 0.3-0.5 Ser -3.9-3.6-3.8-3.8-4.1-3.8 0.2 0.1 Glu 45.6 45.7 45.4 45.2 45.9 45.6 0.3 0.0 Phe 1.7 2.2 2.3 2.2 2.1 2.2 0.1 0.5 Hyp -9.1-9.4-8.5-9.2-9.1-9.0 0.4 0.1 Average 0.3 0.0 Std.dev 0.1 0.4 δ 15 N ( vs Air) determined by GC/IRMS 50 40 30 20 10 0-10 -20 y = 1.000 R 2 = 0.9998-30 -30-20 -10 0 10 20 30 40 50 Known δ 15 N ( vs Air)