OPERATING MANUAL. Mega 17R HIGH SPEED REFRIGERATED CENTRIFUGE HANIL SCIENCE INDUSTRIAL CO., LTD. PRINTED IN KOREA

Similar documents
Ultra 5.0 Manufactured By Hanil Science Industrial Co., Ltd. In Korea

Model : LMHS-40 Please read this manual carefully before using the instrument Labnics Equipment

Operation Manual Supra R22

Operating Manual Cellspin

1580R User s Manual. Multi-Purpose, High Speed Centrifuge. Gyrozen Co., Ltd.

GLO GT416. Benchtop Centrifuge. User s Manual. Glotech Co., Ltd

Supra 22K. Silence Safety Speed. Features. High Speed Centrifuge. Ergonomically designed model of High Speed Centrifuge.

624R User s Manual. Compact Bench-Top Refrigerated Centrifuge

1236R User s Manual. Multi-Purpose, High Speed Centrifuge

CLEMENTS. Orbital 160 Benchtop Centrifuge. CEN V 50Hz. User Manual. Manual No. CEN Issue 4

Bio Lion Table Top Centrifuge XC-H165

Centrifuge. Model XC-1000 Operating Manual

Instruction Manual for the Revolutionary Science RS-200 RevSpin Microcentrifuge

Multi-Purpose, High Speed Centrifuge 1248R/1580R User s Manual

Model : LMHS-60 Please read this manual carefully before using the instrument Labnics Equipment

Centrifuge. Model XC-2000 Operating Manual

CLEMENTS. Orbital 260 Benchtop Centrifuge. CEN V 50/60Hz. User Manual. Manual No. CEN Issue 5

Heal Force Multipurpose Centrifugeses

Rugged steel cabinet construction ensures quiet operation and long term reliability.

Instruction Manual for the Revolutionary Science RS-102 RevSpin Microcentrifuge

L550 Table-Top Low-Speed Large Capacity Centrifuge

BLOOD BANK REFRIGERATED CENTRIFUGE

Operating Manual. Multi Purpose Centrifuge Combi - 514R

Supra R30. High Speed Centrifuges. High Speed Centrifuge. Supra R30 / Supra R22 / Supra R17 / Supra R12

CONTENTS. 3 MAINTENANCE 3.1 Service and Maintenance Maintenance and cleaning Glass breakage Disinfection

INSTRUCTION MANUAL. CAT Place Table Top Centrifuge

Bio Lion Table Top Centrifuge XC-L5

OPERATION MANUAL OM3121

OPERATION & MAINTENANCE MANUAL FOR SCEN-206 LED & MULTI-FUNCTION TYPE CENTRIFUGE PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION.

REFRIGERATED CENTRIFUGE. LRF-A2 Series

XCM12 Mini-Centrifuge

BlueSpin Cryo. (Cat. No. BSD-CRYO)

Centrifuge Operator / Service Manual

STANDARD OPERATING PROCEDURES SOP: 1002 PAGE: 1 of 5 REV: 0.0 DATE: 12/23/96

HIGH SPEED REFRIGERATED CENTRIFUGE.

LABORATORY LOW SPEED REFRIGERATED CENTRIFUGE

Large Centrifuge. Continent 512R I Continent 512R Plus I Combi 514R I Combi 408. Large Centrifuge

Operation Manual Cole-Parmer MPR16 Multipurpose Refrigerated and MP16 Multipurpose Ambient Centrifuges Models , -25, -50, -55

Instruction Manual for Compact Centrifuge Z 206 A

materials and workmanship for 2 years. Should the centrifuge require warranty or Phone: or Fax:

Clinical, Low-Speed Centrifuge. Clinical, Low-Speed Centrifuge. Applications. 59 Order & Inquiry /

LABORATORY LOW SPEED CENTRIFUGE

SECTION 6 5 SERVICE PROCEDURES AND SPECIFICATIONS. Body

Large Capacity, High-Speed Centrifuge

1236R // 1248R LOW. capacity. HIGH G-Force High g-force 25,910 x g for fast centrifugation and efficient separation FAST HIGH

Instruction Manual for the Z 216 MK Refrigerated Microcentrifuge

OPERATIONS MANUAL. Mini Centrifuge Model MCF Certified

3 Liter Programmable Centrifuges

REFRIGERATED CENTRIFUGE. LRF-A1 Series

MODEL 853VES Model 853VES Operator s Manual

Instruction /Service Manual ALC Centrifuge Model PK130R

SECTION 6 5 SERVICE PROCEDURES AND SPECIFICATIONS. Body

SERVICE MANUAL SM3121

Large Capacity, High-Speed Centrifuge

Micro. Centrifuges.

MODEL 853VES. Model 853VES Operator s Manual. Laboratory Centrifuge IVD

High Speed Micro-Centrifuge User manual

Beckman Coulter. GS-6 Series Centrifuges

ISO-9001:2008 CENTRIFUGES CATALOG. Phone: Fax: Web: expotechusa.com

Cassia Siamia BLOOD BANK REFRIGERATED CENTRIFUGE MP 6000 R. Technical Data (MP 6000 R)

PrO-ASTM Centrifuges. Centurion. C1015 Micro Prime Centrifuge. A Micro that offers: Oil testing. Petroleum testing, ASTM methods

MICRO CENTRIFUGES. Experts in Laboratory Equipment

Multi-Purpose Rotor & Micro Rotor can be used in this convenient unit.

Centrifuge 5804 / 5804 R / 5810 / 5810 R. Instruction Manual Mode d'emploi succinct Instrucciones breves


Max. 22,000 rpm / 48,700 xg, full 3 liter capacity

INSTRUCTION MANUAL FOR BENCH TOP CENTRIFUGE Z 200 A

Instructions for Using Micro Plus Multiwell Plate Carriers

HIGH SPEED CENTRIFUGE

Company Profile Hunan Xiang Yi Laboratory Instrument Development Co., Ltd is high technology enterprises mainly engaged in manufacturing centrifuge

Thermo Scientific Heraeus Cryofuge 6000i and Cryofuge 8500i Centrifuge Series

1248/R, 1236R, 1580/R

TD35 VS - TD50 VS Compressor for R134a 12/24V DC Technical specifications

HIGH SPEED CENTRIFUGE

Thermo Scientific Heraeus Primo and Primo R Centrifuges. Flexible and efficient, compact high-performance centrifuges

Trouble Shooting a Jouan C/CR4.22 Centrifuge

Simple design with easy-to-use features

FastPrep FP120 Cell Disrupter INSTRUCTION MANUAL

Neofuge 13/13R. Laboratory Centrifuge. Applications. Heal Force leads you to healthier life

Operating Manual. Operating Manual HYBRIDIZATION INCUBATOR. combi-h12 combi-d24 combi-v12 combi-sv12 combi-sv120

NEOGARD. Preventive Maintenance Manual

Bench Centrifuge or Bench Centrifuge with Trolley

MODEL 755VES Model 755VES Operator s Manual

V1.0. Showven Technologies Co.,Ltd.

Jouan B4i Multifunction Centrifuge Series

2. INSTALLATION 1. MANUFACTURER IMPORTANT: AT LEAST 2 PEOPLE ARE REQUIRED TO LIFT THE CENTRIFUGE FROM THE PALLET AND PLACE IT ON THE BENCH.

SECTION 6 CORROSION PREVENTION AND APPEARANCE CARE. Corrosion prevention and appearance care

SORVALL. Evolution OPERATING INSTRUCTIONS PN

Instruction Manual. Labnet Spectrafuge 16M Microcentrifuge

CENTRIFUGE USER MANUAL 642SYG V.1.0

swing-out rotor MULTIBAS+

Parker AC10 Frequency Inverter (to 22kW) Easy Start Guide

MIKRO R HETTICH DATA SHEET MICROLITRE CENTRIFUGES FASTER RESULTS IN MOLECULAR BIOLOGY FEATURES HIGHLIGHTS

LCD Centrifuge MDCEN-302

Centrifuges and Accessories. Complete Catalogue

Centrifuge 5415 D / Centrifuge 5415 R

Immunity: EN capacity 75 µl Capillaries. 1,5 x 75 mm. Rotor and Lid as evaluation disc incl.

GLASSCO LABORATORY EQUIPMENTS PVT. LTD.

Application Guide DISTRIBUTED BY:

PRODUCT CARE INFORMATION

Transcription:

OPERATING MANUAL Mega 17R HIGH SPEED REFRIGERATED CENTRIFUGE HANIL SCIENCE INDUSTRIAL CO., LTD. PRINTED IN KOREA

Mega17R OPERATING SPECIFICATION Mega 17R MAX RPM MAX RCF MAX CAPACITY SPEED DISPLAY 17,000 rpm 33,172 x g 6 x 85ml 1rpm TEMPERATURE RANGE -20 ~ 40 TEMP. LIMIT SET RANGE 0 ~ 30 TIME BRAKE STAGE PROGRAM ROTOR DIGITAL DISPLAY 9HOUR 59MINUTE 59SECOND 10 STAGE 10 memory FIXED ANGLE ROTOR RPM, RCF, TIME, TEMPERATURE, PROGRAM, BRAKE SYSTEM, GENERAL SPECIFICATION POWER SUPPLY DIMENSION WEIGHT DRIVE MOTOR Page 2 SINGLE PHASE, 220V/ 50Hz, 2.0kW 656W x 585D x 865H/mm 127 kg Brushless AC Induction Motor

OVERSPEED All centrifuges have critical limitation of speeds, at vibration occurs. As the speed increases, beyond the critical speed, vibration will cease. The inherent conditions, also occurs during deceleration. Do not continuously operate this centrifuge at observed critical speeds. IMBALANCE TUBE A balanced load is essential for all centrifuges. An unbalanced load produces vibration, and can damage the unit. Always ensures that the rotor is loaded symmetrically, with a full complement of accessories, and a full set of tubes, tube adapters should also be installed symmetrically. Opposites places of the rotors must always be loaded with same accessories and filling. Eg) only one position I not allowed When you insect the tube, you must insect symmetrically If the tubes is not symmetrical in number, You insert to be same weight to use a new tube.. Page 3

SAFETY INSTRUCTION POWER ON/OFF THE SYSTEM AN AUTOMATIC CIRCUIT BREAKER, FOR EMERGENCY SITUATIONS, SUCH AS POWER SURGES, THAT COULD DAMAGE THE UNIT. USE AUTOMATIC CIRCUIT BREAKER TO PROTECT CIRCUIT CONTROLLER WHEN IT OVERPOWERED. DOOR THE ROTOR WILL STOP AUTOMATICALLY WHEN DOOR OPENS WHICH IS SENSING BY LIMIT SWITCH.. CHAMBER SAFETY OTHER OF CHAMBER IS WELL DESIGNED AND INSTALLED WITH AMOR STEEL TO BE TO SUSTAIN UNEXPECTED HAZARDS. SPEED ALL CENTRIFUGES HAVE CRITICAL LIMITATION OF SPEEDS, AT WHICH VIBRATION OCCURS. AS THE SPEED INCREASES, BEYOND THE CRITICAL SPEED, VIBRATION WILL CEASE. THE INHERENT CONDITIONS, ALSO OCCURS DURING DECELERATION. DO NOT CONTINUOUSLY OPERATE THIS CENTRIFUGE AT OBSERVED CRITICAL SPEEDS. Page 4

VIBRATION MOTOR WILL BE VIBRATED WHEN VIBRATING HEAVILY DURING SPINNING OF ROTOR. THIS SYSTEM SENSES VIBRATION OF MOTOR AND GIVES ALARM SOUNDS WITH FLASHING IMBALANCE LED. IT BEGINS TO START DECELERATE BASED ON POSING DECELERATION TIME. THIS SAFEY DEVICE CAN PROTECT ANY ACCIDENT IN THE LACK OF OPERATOR. TEMPERATURE IT GIVES ALALRM SOUNDS WHEN THE UNIT IS OVER TEMPERATURE FROM THE SETTING LIMITATION OF TEMPERATURE. SYSTEM THE UNIT FLASHES LIGHT WHEN PROBLEM OCCURS IN THE UNIT. Page 5

Note for installation LOCATION BEST LOCATION THE RESTINGN WURFACE MUST BE LEVEL. TO ENSURE QUITE VIBRATION-FREE OPERATION. A LOCK AND STABLE LOCATION IS IMPORTANT. IMPROPERLY LOADED CENTRIFUGE MAY VIBRATE OR MOVE. AIR CIRCULATION.ALL ENERGY CONSUMED BY THE CENTRIFUGE IS CONVERTED INTO HEAT AND EMITTED INTO THE AMBIENT AIR. THEREFORE, SUFFICIENT VENTILATION IS VERY IMPORTANT. AS THE AIR-DUCTS IN THE UNIT MUST BE OPEN, KEEP A CLEANRANCE OD AT LEAST 30 CM AROUND CENTRIFUGE. ALSO, CENTRIFUGE SHOULDN T BE POSITIONED NEAR RADIATORS AND SHOULD NOT BE DIRECTLY EXPOSED TO SUNSHINE. TEMPERATURE, HUMIDITY FOR NORMAL OPERATION THE AMBIENT TEMPERATURE SHOULD NOT BE FALL BELOW 10`C AND NOT EXCEED 35`C. THE MAX HUMIDITY OF AIR 80% DURING TRANSPORT FORM COLD TO WARMER PLACES WATER WILL CONDENSATE INSIDE THE CENTRIFUGE. IT IS IMPORTANT THAT THERE IS ENOUGH TIME FOR DRYING BEFORE THE CENTRIFUGE CAN BE STARTED AGAIN. Page 6

THE PLACE NOT TO OCCUR CORROSIVE GAS SET UP IN THE PLACE THAT CORROSIVE GAS DON T OCCUR. IF SULFUR DIOXIDE AND CHLORINE GAS, ECT ARE IN THE AIR, THE DAMAGE OF ALL PART OF STEEL CAN OCCUR AND CORROSION IN ROTOR AND AXIS CAN OCCUR. BALANCE BALANCING THE SHAFT SHOULD BE VERTICAL FROM THE GROUND. PUT YOUR BALBNCER ON THE TOP OF MOUNTED ROTOR. YOU HAVE TO MAKE SURE THAT YOU HAVE YOUR BALBNCER TO MAKE HORIZONTAL LEVEL ON THE TOP OF SPRNNING SHAFT OR MOUNTED ROTOR ONLY. DO NOT MEASURE YOUR BALANCE. IT IS NOT MAKE HORIZONTAL WHEN YOU PUT YOUR BALANCER ON THE SURFACE OF MAIN CASE. ROTO HORIZONTAL LEVEL IN CASE OF DIFFICULT TO PUT BALANCER ON THE TOP OF SPINNING SHAFT, INSERT ROTOR INTO SHAFT AND PUT YOUR BALBNCER WITHOUT COVER OF ROTOR AS SAME AS Page 7

ABOVE PICTURE. ELECTRONIC P O W E R G R O U N D I N G You must connect after certificating the table in the back panel of centrifuge, as it can change in accordance with order of user, though the power uses 220V. It s a single phase and you must care that it is not three phases that is used in the factory frequently. If the voltage changes from more than 10% to less than 10% in standard voltage, you can t get a precise reliance rates. In addition, the constant power must continue to be supplied as it can damage in various parts of centrifuge. Grounding method of centrifuge : Grounding line of our centrifuge is a green line of power line, and the power is a single phase power of 220V. If user want to certificate whether external grounding is connected well or not in the power line, user has only to certificate whether between one of black or white line and green line is 110V or not. In installation of product, if the conduct of grounding isn t acted well, as it can cause a wrong function of centrifuge, and electronic system of centrifuge can be damaged by external electronic shock, you have to do the conduct of grounding necessary. Don t connect a grounding line in a water pipe, gas pipe, electronic line, lighting rod, telephone line. Especially, in case of a water pipe, although an exposed part is a coppered pipe, as grounding capacity is impossible to be connected PVC line in the middle of pipe sometimes, you had better avoid. And, in case of using a gas pipe as a grounding line, a fire by a flame can cause. In place of being wet, to institute an electronic leakage cut off is more and more safe. Page 8

OPERATING CONTROLLER PROGRAM SAVE RECALL SPEED REPRESENT THE ROTOR SPEED IN RPM OR FORCE IN RCF. SAVE AND RECALL THE DESIRED SETTING AS STORED PROGRAM 1THROUGH 9 TIME DISPLAYED AS MINUTES SECINDS UP TO 99 MINUTES 59 SECONDS AND CONTINUOUS OPERATING WHEN YOU SELECT 0 TEMPERATURE REFRIGERATED AND POSE ONE DECIMAL POINT DOOR BRAKE NUMERIC USE UP OR DOWN KEY IMBALANCE IMBALANCE LAMP LAMP FLASHES WHEN OPERATE 3 BRAKE STAGES TO POSE rpm/time and FLASHES WHEN IT DOOR OPENS AND PRESS (LOWER, MEDIUM AND enter key to save.. VIBRATE AND START THIS BUTTON TO OPEN HIFHEST). TO DECELERATE Page 9

1. 2. Operating procedure Power supply Turn on the main power switch on the back of unit. Turn key switch to On position and lamps will be on and ready to use There will be sounds and all displays will be on when you turn to Power on Position, the unit displays last saved on the control board. Open door and install rotor into the shaft Door will be opened when power on. Door will be closed automatically when the rotor starts to spin, door will be opened when rotor stops. Door will be open when press the door button on the controller during the spinning stage. (Don t press door button when rotor spins). 3. Pose desired setting parameters. Resetting desired data for RPM, Time, Temp. Brake system Press control button to change for speed, time, temperature and so on. Flashes previous setting data Input desired parameters Press save button to save it..4. Press start button to run the unit Press start button to run the unit without changing any parameters such as rpm, time, temperature. The refrigeration system will run to calibrate temperature and rotor will not run when actual temperature is not in the temperature limit range. Page 10

The unit start to decelerate when it reach the time limit. Door will be opened when It stops and displays setting parameters. Press the stop button to stop operation. Control functions RPM Provide very precise rpm (every 4 rpm) and control system in this model 4 rpm, user can select desired rpm (every 1 rpm). Refrigeration system will only operate and rotor does not spin during posing operation time when posting o in rpm window. It s very efficient way to adjust chamber temperature as same as sample evenly. Speed display unit : 1 rpm Speed setting unit : 1 rpm Speed accuracy : Maximum 0.5% Maximum 10 rpm (Under 2,000rpm ) Setup Press RPM button on the controller Previous setting data displays on displayed board and gives sound. RPM data will be returned to the main beginning point and flashing data will stop after pressing enter button. Use numeric button to pose desired setting parameters and press enter button to save. RCF ( X g) Page 11

Displays current RCF in speed window when pressing RCF button on the control and xg lamp will be on simultaneously. RPM/RCF are calculated together. The unit calculate it s RCF automatically when user pose RPM, and also the unit displays its RPM when posting RCF, and the unit displays its RPM when posting RCF data. It can be changed its data when the unit operates. RCF display unit RCF setting unit Setup Press PCF button on the control panel. Previous setting data displays and flashes lamp on displayed board and gives sound. RCF data will return to the main beginning point and flashing data will stop after pressing enter button. Use numeric button to pose desired setting parameters and press enter-button to save posing data will be flashing during posing desired data. Time Display range of operation time is from 0 to 9 hours 59 minutes, 59 seconds, displays 5 digits on the displayed board and it displays down counting of operation time from setting point and displays every seconds. Setup Pressing TIME button on the control panel. Previous setting data displays and flashes lamp on displayed board and gives sound. Time data will return to the main beginning point and flashing data will stop after pressing enter button or temperature button. Use numeric button to pose desired setting parameters with shift lefting method and press enter - button to save posing data will be flashing during posing desired data. Free run when press o in the temperature window and decelerate and down counting until the end of operation time. Use FREE RUN function to operate more than 10 hours of centrifugation. Pose only 0 in the time Page 12

window to use free run function for this unit. Acceleration & Deceleration Acceleration time / Deceleration time control stages : 10 stages 0 is the slowest stages, 9 is the fastest stages The unit will accelerate very quickly when posting 9 and will decelerate very quickly in case of 9 stage. It gives very high effective results of separation Setup Press Brake button on the control panel Previous setting data displays and flashes lamp on displayed board and gives sound. Time data will return to the main beginning point and flashing data will stop after pressing enterbutton. Use numeric button to pose desired setting parameters and press-enter button to save posing data will be flashing during posing desired data. Temperatuare Display temperature in 3 digits and input one decimal point. Precise temperature measurement and control (0.1 0 C) Temp. setting range : -20 0 C ~ + 40 0 C Temp. display unit : 0.1 0 C Temp. setting unit : 0.1 0 C Temp. accuracy : 0.5 0 C (horizontal temp) Setup Press Temperature button on the control panel. Previous setting data displays and flashes lamp on displayed board and gives sound. Temperature data will return to the main beginning point and flashing data will Page 13 Note

Stop after pressing enter button or temperature button. Rotor is not working one minute to protect compressor after pressing start button to stabilize power supplying with power on. Please just wait and will start to operate normally, compressor will stop when posing 100 in the TEMP window. It s very convenient to stop operation compressor. Refrigeration system will only operate without spinning rotor in the chamber after when posing o in rpm window and pressing start button. It s very efficient way to adjust chamber temperature as same as sample evenly. User can keep their sample with rotor in the centrifuges under this condition. Procedure Posing 4 in the temperature and posing 5 in the Temp setting range. Compressor will operate without spinning rotor when posing 0 in the RPM. Begin to operate when temperature displays and maintain 4 degree in the chamber. Time Display range of operating time is from 0 to 9 hours 59 minutes, 59 seconds, display 5 digits on the displayed board and it displays down counting of operation time from setting point and displays every second. Setup Pressing TIME button on the control panel. Previous setting data displays and flashes lamp on displayed board and gives sound. Time data will return to the main beginning point and flashing data will stop after pressing enter button or temperature button. Use numeric button to pose desired setting parameters with shift lefting method and press enterbutton to save posing data will be flashing during posing desired data. Free run when press 0 in the temperature window and decelerate and down counting until the end of operating time. Use FREE RUN function to operate more than Page 14

10 hours of centrifugation. Pose only 0 in the time window to use free run function for this unit. ACCELERATION & DECELERATION Acceleration time / Deceleration time control stages : 10 stages 0 is the slowest stages, 9 is the fastest stages. The unit will accelerate very quickly when posting 9 and will decelerate very quickly in case of 9 stage. It gives very high effective results of separation. Setup Press Brake button on the control panel. Previous setting data displays and flashes lamp on displayed board and gives sound. Time data will be returned to the main beginning point and flashing data will stop after pressing enter button to save posing data will be flashing during posing desired data. Program (Save) Capabilities of saving 10 programs from 0 to 9. How to save Can be change with new desired operating parameters. Setup How to save Press the program button to recall saved program. It displays previous saved number and gives sounds. Press the enter button and stop flashing. Pose desired number in the program window and press the enter button after posing new operation parameters to save desired parameters, it will change and displays new data on the displayed board. Program (Recall) Capabilities of recalling 10 programs from 0 to 9. Page 15

Setup How to recall Press the program button to recall saved programs. It displays previous saved number and gives sounds. Press the enter button and stop flashing. Pose desired number in the program window and press the enter button after posing new operating parameters to save desired parameters. It will change and displays new data on the displayed board. Imbalance The unit give alarm sounds with flashing lamp and starts to decelerate according in time limit at the time of heavy vibration of the unit and rotor imbalance situation. Door Press the foot pedal on the right hand side to open and pull up the door. The unit will not be operated while door opened. Press start button to operate with closed door. Power supply will be stop when door open. Please do not try to open door during centrifugation. The button is located in the right hand side of the unit. Operation Press Start button to operate the unit with posing desired parameters. Stop The unit will decelerate when reaching setting operating time limit. And the unit will stop operating when pressing stop button on the control panel. Care and Maintenance Page 16

This chapter explains how to keep your unit in good operating order. It includes instructions for cleaning, decontaminating and storing. This chapter also covers the cover interlock by pass. See the end of the chapter, for information on service and warranties. Care and cleaning Keep your centrifuge cleans, to ensure good operation, and to extend it s life. Clean the sample chamber, rotor and lid at the end of each works day and immediately after any spill. To clean the chamber, use a damp sponge, warm water, and a mild liquid detergent, suitable for washing dishes by hand. Don t use caustic detergents or detergents that contain chlorine ions. These attack metals. Remove stubborn stains with a plastic scrub pad. Don t use steel wool, wire brushes, abrasives, or sandpaper. They create corrosion sites. Never pour water directly into the rotor chamber. Scrub the rotor s tube cavities with a stiff test tube brush that has end bristles and a non-metallic tip. Dry each part, after cleaning, with a clean, absorbent towel. If glass breakage occurs, remove all broken glass embed in the plastic or rubber accessories. Glass particles can come in contact with new glass tubes, creating pressure points that may results in breakage recurring. Glass particles, in the chamber, grind to a fine gray dust, during centrifugation. This dust can coat the inside of the centrifuge. Storage Store parts on a soft surface, to avoid damage. Rotors and other parts should be clean and dry. Store them open to the air, not in a plastic bag, so that any residual moisture evaporates. Face the parts upward to avoid moisture retention in the cavities. Decontamination If tube breakage occurs, releasing toxic, infections, pathogenic, or radioactive material into the unit, Page 17

decontaminate the chamber. Rotors have sealed continers that provide aerosol containment and, if used as directed, keep sillage confined. If breakage occurs, it may be sufficient to only decontaminate the sealed carriers. Cover door lock The cover will remain locked, if power fails. If you need to remove samples from the unit, before power is restored, use the cover door-lock, after the rotor has come to a stop. Page 18