CONTACTING CONDUCTIVITY GENERAL PURPOSE SENSORS Instruction Manual

Similar documents
ACTIVE ELECTRODELESS SENSORS Instruction Manual

WALCHEM LINEAR POLARIZATION RESISTANCE CORROSION SENSORS Instruction Manual

User s Guide CDCE-90-X SERIES. Conductivity Sensors. Shop online at omega.com SM

Signet Conductivity/Resistivity Electrodes

OPERATING INSTRUCTION MANUAL MS SERIES CONDUCTIVITY CELLS

Signet ph/orp Electrodes

Signet Conductivity/Resistivity Electrodes

Signet ph/orp Electrodes

Signet ph/orp Electrodes

For additional information, please visit our website at SENSOR SPECIFICATIONS

IAQ-CALC TM Indoor Air Quality Meter

IWAKI AMERICA EHC/EHE-FF Series Electronic Metering Pump Instruction Manual

Models P, P, P, , DANGER

Switch-Tek Horizontal and Vertical And Float-Point Float Switches LV20, LV21, LV35, LV36, LH25, LH29, LH35, AV_6 Series 22AUG 08 Rev A

EH Series High Viscosity Electronic Metering Pumps Instruction Manual

Instruction Manual. Series R Peristaltic Pump. For file reference, please record the following data:

User s Guide Shop online at omega.com SM For latest product manuals: CDE-285X SERIES Conductivity Sensors

HIGH PRESSURE AIR COMPRESSOR SYSTEM MODEL #: 091-9HP-220 INPUT:

Cincinnati, OH USA

Model 8340 Intrinsically Safe VELOCICHECK Air Velocity Meter

Indoor Air Quality Meter

IAQ-CALC INDOOR AIR QUALITY METER MODEL 7515

Signet V(D) to V(D) PVDF Conductivity Electrodes

Signet DryLoc ph/orp Electrodes

Hot Dog Roller Grills Instruction Manual Model #8023, Model #8024 and Model #8025 Model #8023SL, Model #8024SL and Model #8025SL

IAQ-CALC TM Indoor Air Quality Meter

Instruction Manual. Fudge Puppy Display Case

INSTRUCTIONS PARTS LIST This manual contains important warnings and information. READ AND RETAIN FOR REFERENCE

Owner s Manual FS-10/FreshSpa

AV16, AV26, AV36, AV46, AV56, LH25, LH29, LH35, LV20, LV21, LV35 & LV36 Series Manual

Rosemount 402 and 402VP

VELOCICALC Air Velocity Meter

OWNER S MANUAL. Model 9880 Intrinsically Safe Air Velocity Meter

AUTO-PUMP 120 VAC AIR COMPRESSOR SYSTEM INSTRUCTION MANUAL. 170 Cherry Avenue West Sayville, NY

Models 396/396P/396R. Retraction/Submersion/Insertion ph/orp Sensor

Air Control Kit INSTRUCTIONS-PARTS LIST. Table of Contents HIGH VOLUME. 250 psi (1.7 MPa, 17.2 bar) Maximum Air Inlet Pressure

Endurance TM Conductivity Sensors

SETRA Operating Instructions Series DPT 230 Pressure Transducers

4&5 RANGE CONDUCTIVITY METERS Operation Instructions

Installation Instructions / Warranty. Citterio 39115XX1

7777D In-Line Mounting for Durafet II ph Electrodes Installation and Maintenance

Wastewater Combination Air Valve. Operation, Maintenance and Installation Manual

OPERATING INSTRUCTION MANUAL

Ventilation Testing. Micromanometer. Alnor Model AXD610 AIRFLOW TM Model PVM610. Operation and Service Manual

Instruction Manual. Model HC1 Series Electrodless Conductivity Sensors

OPERATING INSTRUCTION MANUAL

Model GX Spray Gun

* * Rev. 9 01/18 Product Manual

Pressure Transmitter. Informational & Installation Manual

AIR VELOCITY METER ALNOR MODEL AVM410 AIRFLOW INSTRUMENTS MODEL TA410

Thermo-Anenometer Model Owner s Manual

INSTRUCTION MANUAL AUTO-PUMP 220/240 VAC MODEL #091-9B-220. INPUT :220/240 volt, 50/60 Hz, 1.9/2.2 AMPS OUTPUT: 100 PSI MAXIMUM 1 YEAR WARRANTY

Chapter Page. 1 Introduction Description 1

DP-CALC TM Micromanometer

Spray Clean 606T-2 Installation Instruction

Switch Pro w/ Compact Junction Box A_1_ Series Owner s Manual

Signet DryLoc ph/orp Electrodes

Air Control Kits INSTRUCTIONS-PARTS LIST. Table of Contents HIGH VOLUME. 250 psi (1.7 MPa, 17.2 bar) Maximum Air Inlet Pressure

Model LA 4100 Time Delay OFF Controller

INSTALLATION, CALIBRATION & TROUBLESHOOTING MANUAL

BRADY SAS Parc Eurocit- Bat 12 45, Ave de l Europe RONCQ Cx Tel : Fax :

GW-120 Bar Glass Washer

Installation Instructions

OPERATING INSTRUCTION MANUAL

Predator. Mobile Heated Adhesive Spray Rig. Parts Identification Manual C-ID June 15, 2005

Defender 3000 Series Base Instruction Manual

ODY-19-1 AIR PRESSURE GAUGE

PT-500 Analog Output Series User Manual

Stealth Electronic Manual ZTM1015 Installation & Maintenance. For Oval Gear Meter Models:

THREE RANGE DS METERS Operation Instructions

CRICKET Alphasonic Level Transmitter Model LA12 Owner s Manual

HanTron Model HT-3P. Manual. Installation Maintenance Repair Manual

Installation Instructions / Warranty

Model C230 Pump Controller

Model NTX7 Series Automatic Battery Charger User s Manual Rev. 1.0 October 17, 2006

2015 CONTROLLER PRICE LIST Products / Spare Parts / Accessories

TECHNICAL ASSISTANCE TOLL FREE TELEPHONE NUMBER TECHNICAL ASSISTANCE FAX:

XM8100F/CP. Satellite Radio Antenna. antenna specialists. Patents Pending

MICROMANOMETER ALNOR MODEL AXD610 AIRFLOW INSTRUMENTS MODEL PVM610

Model GD-K77D4X-XX Supplement to the GD-K77D Operator s Manual

HM860 FLOWMETER OPERATOR S MANUAL DO NOT USE OR OPERATE THIS EQUIPMENT UNTIL THIS MANUAL HAS BEEN READ AND THOROUGHLY UNDERSTOOD

8" - 12" Hydraulic Steel Squeeze Off Tool

Global Water Instrumentation, Inc.

DEMA MODEL 582 FOAM STATION III INSTRUCTION SHEET

VSP SERIES Installation, Operation and Maintenance Manual

Operating manual. ba55321de LRD Conductivity measuring cell to be installed in pipelines. ba55321de01 12/00

ELECTROMAGNETIC FLOWMETERS INSTALLATION, OPERATION AND MAINTENANCE MANUAL

Foam Output 50 gal/min (190 ltr/min) 65 gal/min (246 ltr/min) Foaming Distance 30 feet (9 meters) 35 feet (10 meters) alcohols, ketones.

Analytical Instruments Two-Electrode Conductivity Sensors

Model GD-70D4A-XX Supplement to the GD-70D Operator s Manual

Installation Instructions / Warranty. Starck 10458XX1

Supplemental Manual for All PulsFOG Machines. Dramm pulsfog Manual Supplement 1/14

IMPORTANT OPERATING CONDITIONS. Failure to comply with any of these conditions invalidates the warranty. STANDARD CONFIGURATIONS

DEMA MODEL 692T FOAM STATION I INSTALLATION INSTRUCTION

Lubricator Gun: 10,000 psi (700 bar) Maximum Delivery Pressure when disconnected from Dispenser

Dimensions 12/800N 12/1200N D. DC to AC Power Inverters. OWNERS MANUAL for Models: OWNERS MANUAL April ISO 9001:2000 Certified Company

!"" #$% "!&' ( ( ) *

Analog Condensate Monitor Instruction & Maintenance Manual

Advanced Calibration Designs, Inc. Release II

Care and use guide. Contents. page

Transcription:

CONTACTING CONDUCTIVITY GENERAL PURPOSE SENSORS Instruction Manual

Notice 2018 WALCHEM, Iwaki America Inc. (hereinafter Walchem ) 5 Boynton Road, Holliston, MA 01746 USA (508) 429-1110 All Rights Reserved Printed in USA Proprietary Material The information and descriptions contained herein are the property of WALCHEM. Such information and descriptions may not be copied or reproduced by any means, or disseminated or distributed without the express prior written permission of WALCHEM, 5 Boynton Road, Holliston, MA 01746. This document is for information purposes only and is subject to change without notice. Statement of Limited Warranty WALCHEM warrants equipment of its manufacture, and bearing its identification to be free from defects in workmanship and material for a period of 24 months for electronics and 12 months for mechanical parts and electrodes from date of delivery from the factory or authorized distributor under normal use and service and otherwise when such equipment is used in accordance with instructions furnished by WALCHEM and for the purposes disclosed in writing at the time of purchase, if any. WALCHEM's liability under this warranty shall be limited to replacement or repair, F.O.B. Holliston, MA U.S.A. of any defective equipment or part which, having been returned to WALCHEM, transportation charges prepaid, has been inspected and determined by WALCHEM to be defective. Replaceable elastomeric parts and glass components are expendable and are not covered by any warranty. THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY, EITHER EXPRESS OR IMPLIED, AS TO DESCRIPTION, QUALITY, MERCHANTABILITY, FITNESS FOR ANY PARTICULAR PURPOSE OR USE, OR ANY OTHER MATTER. 180543.E March 2018

Table of Contents 1.0 Introduction..... 4 2.0 Specifications..... 4 3.0 Installation... 4 4.0 Maintenance... 8

1.0 Introduction The contacting conductivity sensors measure a voltage drop between two electrodes which is inversely proportional to the conductivity of the solution. 2.0 Specifications 2.1 Measurement Performance Part Number Cell Constant Range Resolution Accuracy 103903-10 0.01 0-300 µs/cm 0.01 µs/cm ± 1% of reading 103904-10 0.1 0-3,000 µs/cm 0.1 µs/cm ± 1% of reading 103905-10 1.0 0-30,000 µs/cm 1 µs/cm ± 1% of reading 103906-10 10 0-300,000 µs/cm 10 µs/cm ± 1% of reading Note: Conductivity ranges above apply at 25 C. At higher temperatures, the range is reduced per the range multiplier chart. Temperature C Range Multiplier Temperature C Range Multiplier 0 181.3 80 43.5 10 139.9 90 39.2 15 124.2 100 35.7 20 111.1 110 32.8 25 100.0 120 30.4 30 90.6 130 28.5 35 82.5 140 26.9 40 75.5 150 25.5 50 64.3 160 24.4 60 55.6 170 23.6 70 48.9 180 22.9 2.2 Mechanical Part Number 103903-10 103904-10 103905-10 103906-10 Length 5.25 (133 mm) 5.25 (133 mm) 5.25 (133 mm) 9.25 (235 mm) Minimum immersion length 2.25 (57 mm) 0.5 (12.7 mm) 0.5 (12.7 mm) 5.25 (133 mm) Diameter 0.5 (12.7 mm) Pressure (Polypropylene fitting) 0-100 psi (0-6.9 bar) Pressure (316SS fitting) 0-200 psi (0-13.8 bar) Temperature (Polypropylene fitting) 32-212 F (0-100 C) Temperature (316SS fitting) 32-248 F (0-120 C) Fitting Process Connection ½ NPTM Electrode Material 316SS Insulator Material PTFE O-Ring Material EPR Temperature Element Pt1000 RTD Cable length 10 feet (3 m) Maximum cable length 250 feet (76 m) 3.0 Installation 3.1 Mechanical Installation General Guidelines Mount the sensor as close as possible to the controller.

Take care to shield the cable properly. Maximum cable length is 250 feet. Position the sensor such that a fresh, representative sample of the solution is available. Position the sensor such that air bubbles will not be trapped within the sensing area. Position the sensor where sediment or oil will not accumulate within the sensing area. If cable is installed in metal conduit (recommended), either flexible conduit should be used or some other provision made for removal of sensor from the process for maintenance. Submersion Installations The submersion sensor requires a standard 1/2 NPTF coupling (user supplied) for connection of the sensor fitting to a standard compatible pipe. The sensor should be immersed away from the walls and floor of the tank by a minimum of 2 inches. The support pipe must be long enough to be above solution level. It should be sealed at the top, with a user supplied cable clamp, to prevent moisture from filling the pipe. See Figure 1. This pipe will usually be suspended from a bracket attached to the lip of the tank. If the cable run will be exposed to moisture (rain, hosing, etc.), it must be protected with flexible conduit (preferably metal). In-Line Installations Thread the fitting into the end port of a 1/2 NPTF tee as shown in Figure 2. NOTE: It is important that the flow direction is in the end port and out the side port (as shown in Figure 2) to provide maximum cleaning of the sensor. If the sensor will be exposed to moisture (rain, hosing, etc.) the cable end must be protected.

½ NPT STRAIN RELIEF ½ NPTF TO 1/2 SOCKET TEFLON TAPE SEAL 3/8 x 3 W POLYPRO MTG BRACKET ½ PIPE 1/2 NPTM TO 1/2 SOCKET 1/2" NPTF COUPLING TEFLON TAPE SEAL O-RING FITTING NUT FITTING BODY PLASTIC WASHER (FLAT SIDE UP) METAL WASHER 103907 Gland SENSOR Figure 1 Submersion Installation

103907 Gland 191669 Gland FITTING NUT METAL WASHER FITTING NUT O-RING PLASTIC WASHER (FLAT SIDE DOWN) O-RING METAL WASHER FITTING BODY FITTING BODY TEFLON TAPE SEAL TEFLON TAPE SEAL SENSOR 103903-10 through 103906-10 SENSOR 103903-10 through 103906-10 ½ NPT TEE FLOW OUT ½ NPT TEE FLOW OUT FLOW IN FLOW IN Low Pressure High Pressure Figure 2 In-Line Installation

3.3 Electrical Installation Route the cable through one of the water tight cable glands on the W100, W600 or W900 series controller, and connect the wires matching the wire label with the terminal block label. 4.0 Maintenance 4.1 Cleaning the sensor Note: the controller must be recalibrated after cleaning the probe. The probe should be cleaned periodically. The frequency required will vary by installation. In a new installation, it is recommended that the probe be cleaned after two weeks of service. To determine how often the probe must be cleaned, follow the procedure below: Read and record the conductivity. Remove, clean and replace the conductivity probe. Read conductivity and compare with the reading in step 1 above. If the variance in readings is greater than 5%, increase the frequency of probe cleaning. If there is less than 1% change in the reading, the probe was not dirty and can be cleaned less often. Cleaning Procedure An accumulation of dirt or debris on the sensor can affect the accuracy and the thermal time constant. This accumulation should be removed periodically. This can be accomplished by scrubbing with a toothbrush or stiff bottle brush. Detergent or isopropyl alcohol cleaner may help remove oils. A mild acid will remove calcium scale. Harsh abrasives should be avoided. Rinse the sensor thoroughly before returning to service. 5 Boynton Road Hopping Brook Park Holliston, MA 01746 TEL: 508-429-1110 Web: www.walchem.com