Trouble Shooting a Jouan C/CR4.22 Centrifuge

Similar documents
LG Air Conditioning Universal & Multi Split Fault Codes Sheet. Universal and Multi Split Units

Troubleshooting Guide

BRIVIS DUCTED INVERTER SERVICE MANUAL DRCi

MCD 500 Service Manual. Functional Tests 5 Power Circuit Test 5 Start Performance Test 5 Run Performance Test 6 Trip Messages 6 General Faults 10

BASIC TROUBLE SHOOTING (PERFECTPASS FOR MECHANICAL ENGINES) How PerfectPass Works

Multi-colour LED Status Light - using it as a diagnostic tool:

CORESENSE DIAGNOSTICS FOR STREAM REFRIGERATION COMPRESSORS

POWER DOOR LOCKS 8P - 1 POWER DOOR LOCKS CONTENTS

LG Air conditioning CAC and Multi Split unit Fault code sheet Universal and Multi Split Units

PROBLEM SOLVING GUIDE ELMECO SLUSH MACHINE MODEL FCM

About this Troubleshooting Document

Application Engineering Europe

¼ to 1HP Chiller Service Manual

Main 20 A fuse has blown (2) Connector has come off PCB

ARM V FDBK ENSURE MOTOR IS NOT ROTATING DURING POWER UP STILL FAULTS? YES ENSURE ARMATURE WIRING IS ISOLATED FROM ANY OTHER POWER LEADS STILL FAULTS?

Troubleshooting Guide

Instruction Manual for Microliter Centrifuge Z 233 M-2

LUV INVERTER SERIES. Self-diagnostics and Trouble-shooting. Part 2 of 2 42/38LUV028H 42/38LUV035H 42/38LUV052H 42/38LUV065H 42/38LUV070H 42/38LUV080H

TECHNICAL SERVICE DEPARTMENT Technical Service Bulletin LowNOx Commercial Gas Electronic Spark Ignition Sequence

EB Conveyor Maintenance Guide

A problem with the motor windings. A phase loss on mains terminals L1, L2, or L3 during run mode. Parameter 2-3 Current Imbalance Delay.

STATIM CASSETTE AUTOCLAVE TROUBLESHOOTING GUIDE

SMART DRIVE ELECTRONIC WASHING MACHINE

P-600. Technical Manual. Troubleshooting Repairs Replacements

ZIP Economizer Fault Detection and Diagnostics (FDD) Table

Service Manual Gulf Stream Electronic Full Wall Slide Systems

Bio Lion Table Top Centrifuge XC-H165

PRODUCT INFORMATION BULLETIN #3365 DIGITAL MOTOR CONTROL PLATTER SYSTEMS For Serial Number and After

HGM-MZ Multi-Zone Monitor Annual Maintenance And Troubleshooting Guide

Section 1.0. Section 2.0. Section 3.0. Section 4.0. MCD3000 Series. Contents. Warnings... 3

1580R User s Manual. Multi-Purpose, High Speed Centrifuge. Gyrozen Co., Ltd.

2. Milk does not heat and the heater is not hot. a. Is there power getting to the pasteurizer heater? Check the incoming voltage to the contactor

SUN ELECTRONIC SYSTEMS EC1X HEAT/COOL TROUBLESHOOTING GUIDE

Service Manual. For the SCV2832E, SCV2426, Automatic Scrubbers For: Training Troubleshooting

3. SERVICE TOOLS Inverter checker RSUK09-17 CLIMATE COMFORT. All Seasons. Heating. Air Conditioning. Applied Systems.

Installation Instructions

ALTERNATOR - BOSCH 35/75-AMP & 40/90-AMP


SYNCHRONIZE YNCHRONIZER AUTOMATIC TIC TROUBLESHOOTING GUIDE RELAY ASSEMBLY SOLENOID (MODEL 1750) LIMIT SWITCH ENGINE CABLE CABLE TO

Inclined Platform Lift Installation Instructions 500, 700, and 750 Lb Capacity Lifts

Sure-Lites. Installation Instructions for the Sure-Lites UX Self Powered Exit Sign with Self Diagnostics. IB505037EN Important Safeguards

MAKING MODERN LIVING POSSIBLE. Service Manual VLTp Soft Starter - MCD500 *MG17L402*

Bio Lion Table Top Centrifuge XC-L5

Installation Instructions for Key Switch SNOWRATOR

NOTE: THESE PROCEDURES ARE THE SAME AS BEING IN THE SERVICE MODE. NOTE: ONCE THE UNIT IS POWERED DOWN THE VALUES FOR C4 WILL BE GONE.

********SERVICE MANUAL********* MODELS:

LaSpaziale Vivaldi Progressive Preinfusion Installation

Furnace Web Site FAQs. Pro Press 100 / Pro 100 / Pro 100 plus. Lift. Belt Noise

ELECTRICAL SYSTEM RP-7

THIS GUIDE IS INTENDED FOR DEALERS AND SOLAR COMFORT TECHNICIANS ONLY AND IS NOT MEANT OR INTENDED TO BE REPRODUCED OR DISTRIBUTED TO THE CONSUMER

BIGLA30-T/BIELA14-T Event Codes Quick Reference EXPLANATION CORRECTIVE ACTION PARTS TO CARRY ON SERVICE CALL

RAAS Troubleshooting Guide. Confirm that the power switch to the unit is in the on position.

Repair instructions UNIVERSAL 320 UNIVERSAL 320 R Andreas Hettich GmbH & Co. KG AR1401EN

Motor. Document # Vari-Green Motor and Controls. Table of Contents. Features, Operation, Wiring and Troubleshooting

REFRIGERATION COOLERS

6. Troubleshooting. 6-1 Basic items for trouble shooting. Samsung Electronics 17

C960, C962, C964 Treadmill

Note: Please read through the entire guide before attempting any kind of installation.

WALL MOUNTED type SPLIT TYPE ROOM AIR CONDITIONER. Models

SERVICE MANUAL. Standard Horizon PF-P310 Paper Folder

MICROGUARD 500 EXTENSION REEL TRAINING MANUAL. Greer Company. Greer Company Crane Systems 1 OF18

KEYLESS ENTRY SYSTEM & TIRE PRESSURE MONITOR ACCESSORIES & EQUIPMENT General Motors Corp. - Remote Keyless Entry System

Technical Information and Diagnostic Guide RestStar Use this guide with 5700XE RestStar Unit. Western Star 5700XE.

WARNING This manual should only be used by a qualified Service Technician. FinishPro 390/395 Airless/Air-Assisted Sprayer Repair Electrical Manual

Auto-Level Troubleshooting (Old Platform) Electronic Control- Prior to 2009, Pressure Switch Control panel #s 2057, 2058, 2795, 2795B

3200NT Timer Service Manual

12. Troubleshooting Items to be checked first. Samsung Electronics

Electronic Refrigerator Diagnostics

TROUBLESHOOTING TP. Index

ELECTRIC FENCE ENERGIZER SERVICE MANUAL MODEL 950 SERVICE MANUAL FOR OLLI 950 FENCE ENERGIZERS


MICRO P-711 Exerciser Instruction Book

Solstice Electric Fryers SE Series Service Manual

Wolverine Turn Signal / Horn Kit 2102

PREMIER POWER PACK INSTRUCTION MANUAL EN54-4 POWER SUPPLY UNIT INSTRUCTION MANUAL. GLT.MAN-138 Issue: /05/2016 N.R.P.J.

XR Conveyor Maintenance Guide

2005 and 09 Mustang install instructions Sequential / Chase Unit Partial Plug-N-Play Kit Meter4it Eng. Updated: 3/28/09

INSTALLATION MANUAL SPECTRUM BRAKE CONTROL

TROUBLESHOOTING GENERATOR UNITS

Quickie Rhapsody Service Manual

624R User s Manual. Compact Bench-Top Refrigerated Centrifuge

USERS GUIDE LO-21U LOCKOUT RELAY

SERVICE MANUAL. Kysor Instrumentation Troubleshooting Guide

C944 Treadmill. C944 Treadmill

ILQ TROUBLESHOOTING GUIDE

Trouble Shooting 1 The power light doesn't illuminate and the scooter doesn't run The power light is on and the scooter doesn't run.

ILT - Series. Troubleshooting: ILT Twinfold Dual Cylinder Folding Unit

LS8.0T Service Manual

Convertible Relay Replacement System

VAGABOND S HANDBOOK TRANSMISSION

Subject Underhood G System Error Codes and Symptoms System or Parts affected

70001 and Clutch Rebuild Instructions

Phone: Fax: ILD Series. Troubleshooting: ILD Column Lift

OPERATION MANUAL Common for All models Bobbin Sensor (Weft End Detection) DP150 - L2 BP150 - L2-12V BP150 - L2-24V (PNP/NPN)

OPERATING MANUAL. Mega 17R HIGH SPEED REFRIGERATED CENTRIFUGE HANIL SCIENCE INDUSTRIAL CO., LTD. PRINTED IN KOREA

KMIR. Multi outdoor units. troubleshooting Multi zone CONDENSING UNITS MKMIR218-H221, KMIR327-H217, KMIR436-H217, KMIR545-H219.

LLP-UL-M-KIT Series Motorized Electric Latch Retraction

Aftermarket Interface Module

INSTALLATION & OPERATION MANUAL

Transcription:

The world leader in serving science Trouble Shooting a Jouan C/CR4.22 Centrifuge No Spin Condition Erratic Speed Control No Cooling General Parts List Wiring Diagram

No Spin Condition (Removing the Front Panel) Using a 3mm Allen Wrench, remove screws at the base of the Front Panel Using a Flat blade screwdriver, insert the tip of the screwdriver between the top deck of the machine, and the front panel Using the screwdriver, pry the front panel pins out of the top deck to completely detach the front panel from the centrifuge. Carefully allow the panel to fold down, allowing the string to tighten and hold the panel in place. 2

No Spin Condition (Removing the Left Side Panel) Using a 3mm Allen Wrench, remove screws at the base of the Left Side Panel Using a Flat blade screwdriver, insert the tip of the screwdriver between the top deck of the machine, and the left side panel Using the screwdriver, pry the panel pins out of the top deck to completely detach the front panel from the centrifuge. 3

No Spin Condition Error Codes The C/CR4.22 centrifuge will alarm if an alarm condition is met. The alarms that the unit will give are: Chamber Overtemperature Motor Overtemperature Imbalance Lid Open If the unit does not give any errors and does not spin: Click Here 4

No Spin Condition Chamber Overtemperature If a Chamber Overtemperature alarm occurs: Verify the actual temperature in the chamber (at the sensor) Verify the temperature reading on the display To see the temperature read by the centrifuge, when in standby press the F key. The chamber temperature will flash on the display If the Displayed temp. value matches the Actual: Fault in the Compressor Circuit If the Displayed temp and Actual temperature do not match: Faulty Temperature Sensor Faulty CPU assembly 5

No Spin Condition Motor Overtemperature If a Motor Overtemperature error occurs: Locate connector J5 on the Main PCB With connector still connected place a jumper across Pins 4 & 6 (Brown or Blue wires) If unit spins: Check that the Overtemperature wires are not cut or disconnected from the connector on the PCB. If the Overtemperature wires are OK, then replace the motor If unit fails to spin: Replace CPU Click Here for a General Parts List 6

No Spin Condition Imbalance The first thing to check on an imbalance alarm: Unit is positioned on a sturdy countertop Buckets are properly Loaded. These units are capable of running with up to 20 grams of imbalance. Check that the unit s trunnions are properly lubricated. If the above checks are OK, proceed to next slide 7

No Spin Condition Imbalance (Continued) If an Imbalance Condition occurs: If an Imbalance occurs right after start is pressed: Imbalance sensor is somehow shorted Imbalance sensor is shorted during shipping to prevent damage to the Pendulum. Check that sensor is not physically shorted Unplug the imbalance sensor from Main CPU Connector J5 Pins 7 & 8 If Imbalance still occurs right after start is pressed with sensor unplugged: Replace CPU PCB Click Here for a General Parts List 8

No Spin Condition Lid Open If a Lid Open error Occurs: Faulty Lid Switches When the Lid is closed you should see a short across the micro- Switches To test the switches unplug connector J4 from the Main CPU. The Lid switches connect to J4 Pins 5 & 6 (Gray Wires) With the lid closed check that the switches are shorted. Replace switches if they are open Faulty Main CPU If the above check is OK Replace Main CPU. Click Here for a General Parts List 9

No Spin Condition Checking Motor Voltage Locate the Power PCB on the Left Side of the unit. You will notice a four pin connector right above the Motor Relay. Check the DC Voltage when Start is pressed on the Red & Black wire. Voltage applied should be approximately 30vdc at start, and ramp up to as high as 110vdc at maximum speed. 10

No Spin Condition Checking Motor Voltage (continued) If voltage is present: Brushes may be hung up Brushes may be to short Must be at least 0.50 Long Faulty Motor If no Voltage is applied to the motor proceed to next slide Click Here for a General Parts List 11

No Spin Condition, No Motor Voltage - PCB Checks The two possible causes left on a No Spin Condition are: Main CPU Power PCB The best way to determine Which PCB is faulty is to run the integrated tests. Specifically Test # 7. Continue to the next slide for instructions on how to run the test. 12

Running the Integrated Tests (View of Front Panel) 13

Running the Integrated Tests Test # 7 *Note: It is necessary to have a program with a DELTA TEMP = 0, before access to test procedure. To call the test Procedure push the following keys in sequence: ENTER, STOP, 7 Press ENTER, then ENTER again. The following Message appears FULL=0, ONE=1 Keying 0 will cause all the tests to run. Keying 1 requires the operator to select which test to run, by keying the test number. In this case, press 1 then ENTER The following Message appears Test NO.? Press 7, then ENTER 14

Running the Integrated Tests Test # 7 (Continued) The message on the display is now: Test No. 7: Motor Test Press Enter The displayed message now is: Close the Lid Even if the lid is closed, this message will appear Close the lid and press ENTER The display now says: Press Start Press START to run the test 15

Running the Integrated Tests Test # 7 (Continued) The display will show alternately: 220v = 5.5A 110v = 11A If the unit spins, the RPM value will also be displayed If the unit spins, the display will show: TEST NO 7 OK If the unit does not spin, the display will show: TEST NO 7 BAD 16

Test # 7 Results If the unit fails Test # 7: Failure of the SCR Bridge, or Relay on Power PCB is likely. Replacement of the Power Control PCB is needed. If unit passes Test # 7: Replacement of the Main CPU is needed. It is possible to reset the processor on the MAIN CPU, and sometimes, temporarily fix the no spin condition. Proceed to the next slide for details. Click here for a General Parts List 17

Resetting the Processor To reset the Processor: Short pins 35 & 36 together on the chip (see next slide) for 3 seconds. Unit will beep The display will come up in French It will be necessary to reprogram unit Unit may spin now, which can help your customer to continue their work, but PCB still needs to be replaced. See next slide for a view of CPU PCB and pins 35 & 36 of EPROM 18

View of CPU PCB. 19

Erratic Speed Control (Removing the Front Panel) Using a 3mm Allen Wrench, remove screws at the base of the Front Panel Using a Flat blade screwdriver, insert the tip of the screwdriver between the top deck of the machine, and the front panel Using the screwdriver, pry the front panel pins out of the top deck to completely detach the front panel from the centrifuge. Carefully allow the panel to fold down, allowing the string to tighten and hold the panel in place. 20

Erratic Speed Control (Removing the Left Side Panel) Using a 3mm Allen Wrench, remove screws at the base of the Left Side Panel Using a Flat blade screwdriver, insert the tip of the screwdriver between the top deck of the machine, and the left side panel Using the screwdriver, pry the panel pins out of the top deck to completely detach the front panel from the centrifuge. 21

Erratic Speed Control Verifying Values Create a program with an Acceleration Rate of 9, and a Set Speed of 4000 RPM. Allow centrifuge to either stabilize at set speed, or stabilize at the maximum speed it will make. Remove the Black viewport cap. Using a Photo-Tachometer, measure the actual speed of the unit through the view port in the center of the centrifuge lid. If the speed is accurate compared to the set point, try again at 2000RPM s. 22

Erratic Speed Control Possible Failures If the speed display and the actual speed match: Check Brushes If the speed display is different from the actual speed, check: Tachometer Sensor Motor Magnets Main CPU 23

Erratic Speed Control Checking Brushes You can check the brushes two different ways: Remove the back panel (refrigerated units only) and remove the brushes Turn the unit on it s back or side (if refrigerated, refrigeration side must go down) and the bottom of the motor is exposed. Click Here for a General Parts List 24

Erratic Speed Control Tachometer Sensor Check Locate Connector J5 on Main CPU Verify that there is a 12vdc supply on J5 Pins 1 & 3 (Pink wire & White/Black wire) If no 12vdc supply is present: Replace Main CPU Check for 4 pulses per 1 revolution of the rotor on J5 Pins 2 & 3 (Orange Wire & White/Black wire). If one or more pulses are missing: Check the Motor Magnets Check the frequency at Pins 2 & 3 while the motor is spinning. Multiply your reading by 15 to obtain the correct tach reading Click Here for a General Parts List 25

Erratic Speed Control Checking the Magnets in the Motor Remove rotor from the centrifuge. Disconnect Motor wires, Tachometer wires, and Imbalance wires. Turn centrifuge on it s side (refrigeration side down) Remove the three 10mm bolts from the motor mounts located under the chamber. Rotate the motor so that it can be pulled straight out. Once motor is pulled, remove tach sensor from motor by unscrewing the two Phillips head screws or 7mm bolts from Tach tray. Look inside the motor towards the top. You should see four circular magnets. If you are missing one or more of these, replace the motor. Click Here for a General Parts List 26

Erratic Speed Control Checking the CPU Check the following: Brushes Tachometer Sensor Motor Magnets If the above checks are OK: Then the problem is with the CPU Click Here for a General Parts List 27

No Cooling In a NO-COOL situation there are several possibilities: Temperature Sensor CPU Triac Compressor 28

No Cooling Checking Temperature Sensor Measure Temperature Sensor resistance: 30KΩ @ 25 C across Brown & Green Wires 6KΩ @ 25 C across Brown & Yellow Wires If sensor does not read correctly replace sensor Click Here for General Parts List 29

No Cooling Checking CPU Checking the Gate Signal Locate connector J3 Connect meter leads to pins 9 & 11 When CPU commands the Compressor to be OFF, voltage should read line voltage. When CPU commands the compressor to be ON, voltage should drop to approximately 1-2VAC. If signal is not applied correctly, replace the CPU 30

No Cooling If sensor and PCB are correct: Verify proper operation of cooling Triac Verify proper operation of Compressor and starting components. Check that start cap is not corroded If CPU, Temperature Sensor, Triac and starting components are OK, then replace the compressor Click Here for General Parts List 31

General Parts List Bushes (sold individually) ALC Motor 89000214 Non ALC Motor 89000056 Condensing Unit (Embraco) 51245458 Keypad 25176008 Lid Switches Old style (lever actuator) 25445041 New Style (push button) 86001406 Locking Assy. 89003307 Motor 25509559 PCB s Main CPU- C4.22 89000191 CR4.22 81000312 CT4.22 81000314 PCB s Power 25153152 Low Voltage Supply 26005001 Sensors Tachometer 25153151 Imbalance 85240244 Temperature - 86001274 Starting Components (Embraco) Capacitor 81001062 Relay 81001063 Triac (cooling) 26897009 32

Wiring Diagram (1 of 3) 33

Wiring Diagram (2 of 3) 34

Wiring Diagram (3 of 3) 35