Nimbus Flex Ambulatory Infusion Pump

Similar documents
Patient s Manual for Using Nimbus Ambulatory Pump for Continuous Peripheral Nerve Block Application

BodyGuardTM 323. BodyGuard 323 Multi-Therapy Ambulatory Infusion Pump. Patient Guide

CADD -Solis VIP Ambulatory Infusion Pump Model 2120

Quick Reference Guide

Accumate 2300 Volumetric Infusion Pump

Ambulatory. Infusion. Pumps

Sentinel Enteral Feeding Pump. Model: S-1000-SI. Operating Manual

CADD-Legacy 1 Ambulatory Infusion Pump Model 6400

NIKI T34 pump. Knowing your way around the NIKI T34 pump

OPERATOR S MANUAL. Ambulatory Infusion Pump Model 6300 PATIENT CONTROLLED ANALGESIA

CADD-TPN. Model 5700 Operator s Manual. Deltec. Ambulatory Infusion System. SIMS Deltec, Inc., St. Paul, MN U.S.A.

Intermittent Delivery Mode

MOOG Medical Devices Group Curlin 6000 Series Ambulatory Infusion Pump PATIENT GUIDE

6 Bay Battery Charger Operation Manual

SurePower TM II Battery Pack Guide

OPERATION AND MAINTENANCE INSTRUCTION MANUAL. AEU-707A & AEU-707AV2 Implant / Surgery Systems

OPERATING MANUAL POWER DRIVE SYSTEM. Cat. No RevH

Instruction for Use (IFU)

Cole-Parmer Pro-Spense

SurePower Battery Pack Guide

Syringe Driver Introduction. Christine Hull & Anita Webb Staff Nurses, Hospice in the Home 2015

Quickdraw. Model 2400 Series. Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

CADD-PCA. Model 5800R Operator s Manual. Deltec. Ambulatory Infusion Pump. SIMS Deltec, Inc., St. Paul, MN U.S.A.

OPERATION AND MAINTENANCE INSTRUCTION MANUAL. AEU-17B & AEU-17BV2 Implant / Surgery Systems

BODYGUARD 595 REGIONAL ANALGESIA INFUSION PUMP

CADD-Prizm PCS II. Operator's Manual. Ambulatory Infusion Pump. Model 6101

Model 2310 Series. Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

Instructions For Use VATEA. Endodontic Irrigation Device

USER MANUAL READ THIS MANUAL BEFORE OPERATING THIS DEVICE. SAVE THIS MANUAL FOR FUTURE REFERENCE. DEALER: THIS MANUAL MUST BE GIVEN TO THE END USER.

CADD-1. Model 5100 HFX Operator s Manual. Note: Colors are Black and Cyan (as second color) Deltec. Ambulatory Infusion Pump

BODYGUARD 323 AMBULATORY INFUSION PUMP

BodyGuard 121 Twins Infusion System Operator Manual

Cordless Rechargeable Saw Instructions for Use

Syringe Infusion Solutions

SmartVent. Instructions for Use Version 4 (Dec 2009)

SR7 Micro-infusion Syringe Pump. Contents

Operator's Manual Tracer III Nerve Locator, Model NL-3

Syringe Infusion Solutions

CADD Ambulatory Infusion Systems

Operator s Manual. spectrum 35700BAX & 35700ABB. Generation 2 Operating System Pump Operating Software v6.05 For Use with MDL Editor v6.2.

Deltec 3000 and Deltec Micro 3100 Large Volume Infusion Pumps Modular Infusion System

Digital Apex Locator. ipex OPERATION MANUAL OM-E0285E

Companion. Enteral Pump. Operating Manual. For Enteral Use Only Not for IV Use

OPERATION AND MAINTENANCE INSTRUCTION MANUAL. AEU-707A & AEU-707AV2 Implant / Surgery Systems

LIST OF FIGURES... ii. LIST OF WARNINGS and CAUTIONS... iii CHAPTER 1 - GENERAL INFORMATION... 1 CHAPTER 2 - DIRECTIONS FOR USE...

Ambulatory Syringe Pump Operation Manual

User s Manual. Versaflo 200 Syringe Pump. For Veterinary Use Only

O p e r a t o r ' s M a n u a l

Ipump Pain Management System SERVICE MANUAL. For use with Ipump devices with hardware revision 2 (HW Rev 2).

USE, CARE & STERILIZATION INSTRUCTIONS. Consumable Accessories Burs Saw Blades K-Wires Steinmann Pins Twist Drills Rasps Ortho Pins/Drills Guide Wires

BodyGuard 323 Infusion Pump System Operator Manual

DCell Suction. Model DM Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

Dedicated, ambulatory nerve block and wound infiltration treatment

Model 605 Portable Aspirator

HEINE NT4 MED V

Vercise TM DBS Charging Handbook

SAFETY ACTION NOTICE By arrangement with the Scottish Executive Health Department

D Dräger Alcotest 3000 Breath Alcohol Measuring Device

s h a p e f i t t i n g t e c h n o l o g y

THE PHOENIX MODEL 1800 KNEE CPM OPERATOR'S MANUAL SERIES Seeds Road Grove City, OH Fax

Multi Therapy Solutions

CADD Ambulatory Infusion Systems

LLS-2054BE. Smart Battery Pack Operator Manual. ST Technologies 6018 Bowdendale Avenue Jacksonville, FL USA

Before installing this unit, read the operating instructions carefully. Strictly adhere to these operating instructions.

Podiatry Procedures Chair. For Models: Barrier-Free. User s Guide

CADD Amulatory Ambulatory Infusion Systems CADD. Ordering Information. CADD Tubing Sets with Flow Stop Free-Flow Protection.

Companion. Enteral Nutrition Pump. Operating Manual. For Enteral Use Only Not for IV Use NOTES. No Purchase No.

MOBILE AIR CHAIR. Operating Instructions. Pressure Relieving System. Tridien Medical. Revision: AO-SM150-03

FLEXIFLO QUANTUM Enteral Pump

Marin Control Digital

WIRELESS TRI-JACK WIRELESS REMOTE KIT COMPONENTS LITERATURE NUMBER REV. C WARNING EXPLOSION WARNING PERSONAL INJURY & PROPERTY DAMAGE

BodyGuard 323 Infusion Pump System Operator Manual

Going home with a CADD Pump - Chemotherapy Medication

SURGICAL POWER & ACCESSORIES. Rx Only PM-X00-710/PM-X IR-12.13

WIRELESS CAMPER JACK LITERATURE NUMBER REV. C CCD WARNING PERSONAL INJURY & PROPERTY DAMAGE WARNING EXPLOSION. Effective NOV 2018

CHEST COMPRESSION SYSTEM

For detailed instructions, specifications, warnings, warranties, and additional information on operating the CADD -Solis ambulatory infusion pump,

1 AMP CURRENT SOURCE

SRK12KX Battery-powered Crimping Tool

Multi Therapy Solutions

CRONO S-PID 50 Ambulatory infusion pump

8,1,8,1,2,7,2,7,6,3,6,3,4,5,4,5

Slimax LED Curing Light System

EMB-901 Built In Fiber Optic Brushless Electric Micromotor Set Operation Manual

RemRest. Service Manual. CAUTION: USA Federal law restricts this device to sale by or on the order of a physician. P/N Rev A

OPERATING INSTRUCTION

Inductive. SeekTech Clamp WARNING! Operator s Manual

Price Computing Scale D Series Operation Manual

INSTRUCTION MANUAL. Tel / Fax SOM-S056S-KE Rev.5

Medi-Temp III FW601 Blood/Fluid Warmer Operator s Manual/Service Manual P/N /04

Sherman + Reilly Battery-powered Pistol Grip Cable Cutters

Pain Management Solutions

Instruction Manual for Safety and Comfort

Keeler Spectra Plus. Instructions. Next

Table of Contents. Title Page UT501/UT502 OPERATING MANUAL. Introduction. Unpacking the Meter. Safety Information. International Electrical Symbols

R06 JH 03/19/13 EN-INTL SimplyGo USER MANUAL

Charging base guide. A Sonova brand

INSTRUCTIONS EVIS EXERA XENON LIGHT SOURCE OLYMPUS CLV-160

Unit 2, Rivington View Business Park, Station Road, Blackrod, Bolton BL6 5BN Tel: +44(0)

portable ac/dc suction pump USER S MANUAL ITEM # 18605

Transcription:

Nimbus Flex Ambulatory Infusion Pump Patient Manual Continuous Mode Infusions (Software Version v5.1.1) Read this entire manual prior to operating the Nimbus Ambulatory Infusion Pump. Failure to follow instructions or heed a warning may result in death or serious injury.

Table of Contents Section 1: General Description... 3 Indications for use... 3 Symbols... 3 Warnings... 4 Device Diagrams... 7 Keypad Diagram... 8 LCD Screen Symbols... 9 Section 2: Operate the Pump... 10 Start Infusion... 10 Stop Infusion... 11 Power off Pump... 11 Pack and Wear Pouch... 11 Section 3: Troubleshooting... 13 Section 4: Maintenance... 15 Cleaning and Disinfection... 15 Cleaning... 15 Disinfection... 15 2 Nimbus TM Flex Patient Manual - 511-IFU-IT1029, Rev. A

Section 1: General Description Indications for use The Nimbus TM Flex Ambulatory Infusion System is intended to deliver medications and/or fluids to a patient under the direction or supervision of a physician or other certified healthcare professional. The device is indicated for subcutaneous, epidural, perineural, and intravenous infusion. The device is intended for 240 hours or 1500 ml of infusion, whichever limit is reached first. Symbols Consult Instructions for Use Serial Number Warning Federal (U.S.A) law restricts this device to sale by or on the order of a physician Keep Dry Use by YYYY-MM-DD Keep Away From Heat Temperature Limitations Do Not Re-Sterilize IP22 Ingress Protection Level Do Not Re-Use Type BF Applied Part CE certificate Nimbus TM Flex Patient Manual - 511-IFU-IT1029, Rev. A 3

Warnings A Warning ( ) alerts you to a potentially hazardous condition, safety hazard or equipment damage. Failure to heed a warning may result in death or serious injury. Read the entire Patient Manual prior to operating the Nimbus TM Flex ambulatory infusion pump. InfuTronix assumes no responsibility for incidents that may occur if its product is not used in accordance with its product labeling. If a patient or caregiver is to operate the Nimbus TM Flex ambulatory infusion pump without the direct supervision of a medical professional, the patient or caregiver must be trained by a medical professional to operate the pump. This training should include all necessary operation of the pump as well as pump placement, alert/signal recognition, and when/how to contact a medical professional regarding pump use or status. The patient or caregiver should have the means to contact the patient s health care provider at all times when not under direct supervision. If the Nimbus TM Flex Ambulatory Infusion pump is used to provide medication or fluid therapy to any patient under the age of 18, that patient should, during therapy and while having physical access to the pump and administration set, be under the direct supervision of an adult caregiver trained to operate the pump. If the mental or physical condition of a patient or caregiver could interfere with or preclude the proper operation or safe use of the Nimbus TM Flex Ambulatory Pump, an alternate means of therapy should be utilized. Only use pump as directed by physician or other certified health care professional. If you have any questions regarding use of the pump or experience any problems using the pump, contact your physician or other certified health care provider for further instructions. The pump is not to be used for delivery of blood or cellular blood products. The pump is not intended for delivery of life-sustaining medication. Do not use the pump or cassette/administration set to administer any infusion to the epidural space unless the medication and/or fluid infused is indicated for epidural administration. If the pump is used to deliver critical medication, a backup pump should be available. Do not use the pump in or near a MRI (Magnetic Resonance Imaging) device. 4 Nimbus TM Flex Patient Manual - 511-IFU-IT1029, Rev. A

Warnings, Continued Do not expose the pump to X-Rays, Gamma Rays, or other Radiation. Operating the pump near equipment that radiates high-energy radio frequencies (electrosurgical/cauterizing equipment, etc.) may cause false alarm conditions. If this happens, reposition the device away from the source of interference or use an appropriate clinical alternative. Do not operate this pump in the presence of flammable anesthetics mixture with air, oxygen, or nitrous oxide. Do not operate the pump at temperatures below 5 C or above 40 C. Incorrect device handling may lead to allergen exposure and patient allergic response. This may result in death or serious injury. The pump is not to be used if variance within the specified fluid delivery accuracy range could cause harm to the patient. For those patients who are likely to be adversely affected by unintended operations and failures, including interrupted medication or fluid delivery from the device, close supervision and provision for immediate corrective action should be provided by their healthcare provider. Only use InfuTronix approved cassettes/administration sets with the pump. Always close the slide clamp before removing the cassette from the pump. Do not operate the pump outside of the environmental limits detailed in the pump specifications. Always disconnect the administration set from the patient before removing the cassette from the pump. Do not put excessive pressure on the pump that may cause the device, bag or tubing to compress (i.e. leaning against pouch, sleeping on pouch, etc.). Do not leave the pump or cassette/administration set unattended at home or other places where young children or pets may come in contact with the device. Do not store the pump with a cassette/administration set attached. Do not use a Nimbus TM Flex pump or cassette/administration set beyond its shown use by date. Do not place objects over the Nimbus TM Flex pump that may muffle the pump. System delivery inaccuracies may occur as a result of excessive pressure on the bag, back pressure or fluid resistance, which depends upon drug viscosity, catheter size, and extension set tubing. Nimbus TM Flex Patient Manual - 511-IFU-IT1029, Rev. A 5

Warnings, Continued If the pump is dropped or hit, inspect the pump for damage. Immediately stop using the pump if it is damaged or not functioning properly. Contact Customer Service to return the pump for service. If the pump is not functioning as expected, the pump should not be used, and should be returned for inspection and repair. Follow instructions specified in Section 4: Maintenance in this Patient Manual for cleaning and disinfection of the pump. There are no user serviceable parts inside. Refer all service, repair, and calibration to qualified technical personnel. Do not make unauthorized modifications. Avoid spills on the Nimbus TM Flex ambulatory pump and do not provide a patient with a wet device. To avoid mechanical or electronic damage, do not steam, autoclave, or immerse the pump in any fluids or cleaning solutions, and do not spray such fluids directly on the pump. Failure to properly follow cleaning instructions may result in an electrical hazard, damage to the pump, and voided warranty coverage. Do not use this device for neonates, as risk assessment of leachables and extractables has not been performed to characterize additional risk associated with this population. Do not store the pump beyond the expiration date. Battery leakage could occur. Do not expose the pump to humidity levels below 15% or above 93% relative humidity. Do not expose the pump to atmospheric pressure of less than 700hPa or more than 1060hPa. Avoid using and/or storing of the Nimbus pump where it will come in contact with excessive lint, dust, direct sunlight, heat and moisture. Do not use a single Nimbus TM Flex ambulatory infusion pump for longer than 240 hours. 6 Nimbus TM Flex Patient Manual - 511-IFU-IT1029, Rev. A

Device Diagrams Front View Rear View Nimbus TM Flex Patient Manual - 511-IFU-IT1029, Rev. A 7

Keypad Diagram Key Name Description RUN/STOP Press to start, resume, or stop the infusion. ON/OFF Press and hold the ON/OFF key for 3 seconds to power on or off the device. Note: The pump has additional functions beyond those shown above. These functions are intended to be used by medical professionals only. Do not attempt to use these functions unless specifically instructed to do so and trained to do so by a physician or other certified health care professional. 8 Nimbus TM Flex Patient Manual - 511-IFU-IT1029, Rev. A

LCD Screen Symbols Item # 1 Therapy Mode 2 3 4 Name Symbol Description Configuration Mode Programming Lockout Battery Level Indicator 5 Alert The symbol flashes when the pump is infusing. The pump is in configuration mode. Clinician code is required to change parameter values. No clinician code needed to change parameter values. The symbol indicates a full battery level. The symbol indicates a medium battery level. The symbol indicates a low battery level. The symbol indicates an empty battery level. The symbol will flash if an alert condition is present. 6 Setting The pump is in the setting mode. 7 Pause The pump is in the infusion mode. Note: This table includes references to functions that are intended to be used by medical professionals only. Do not attempt to use these functions unless specifically instructed and trained to do so by a physician or other certified health care professional. Nimbus TM Flex Patient Manual - 511-IFU-IT1029, Rev. A 9

Section 2: Operate the Pump The Nimbus Ambulatory Infusion pump should be set up and programmed by a medical professional. Do not attempt to perform any of these functions unless specifically instructed to do so and trained to do so by a physician or other certified health care professional. Start Infusion Press the RUN/ STOP key to execute infusion. The LCD screen will display the infusion rate in ml/hr and display the mode or prescription name of the current infusion. The therapy mode symbol will keep flashing. The LED will flash green indicating the process of infusion. The LED will turn off after 10 minutes of infusion, and will only be on again if the user has pressed any key or an alert condition occurs. Parameters VTBI (Volume to Be Infused) and VINF (Volume Infused during the current infusion) will also be alternatively displayed below the rate value. Note: There will be an infusion delay upon powering on the pump and/ or loading the cassette. Upon pump power on and/or loading of cassette, the pump will need a 10-40 second delay on starting/resuming infusion to obtain upstream and downstream pressure baseline. Acknowledge Alert During an ongoing infusion, the pump can indicate a condition by showing an alert message on the LCD screen. The LED will flash red light. Press the RUN/STOP key to acknowledge the aler. Contact your health care provider for further instruction. 10 Nimbus TM Flex Patient Manual - 511-IFU-IT1029, Rev. A

Stop Infusion Press the RUN/STOP key for 3 seconds any time during infusion execution to stop the ongoing infusion. The LED will turn yellow. The therapy mode sign disappears from the display screen, indicating that the infusion has been stopped. The current infusion will be paused and the pump will enter the programming screen. Power off Pump Press and hold ON/OFF key for 2 seconds to power off the pump. Pack and Wear Pouch Warning: Before packaging the drug/fluid bag and pump into the pouch, check the fluid path for kinks, closed clamp or other occlusions. Open the pump pouch and place it flat on a table. Place the drug/fluid bag in the bag chamber. Open the Velcro strap, position the pump against the clear window with the display screen facing out, and then close the Velcro straps tightly to hold the pump securely in the window. Place the drug/fluid bag in the bag chamber with the bag spike pointing left. Fold the bag spike over the top of the medicine bag and route the upstream end of the tubing between drug/fluid bag and pump carefully so that the tubing does not have kinks or occlusions. Route the upstream end of the tubing between the drug/fluid bag and the pump carefully so that the tubing does not have kinks or occlusions. Route the downstream end of the tubing carrying fluid to the patient through the tubing outlet. Close the pouch cover and zip closed from both directions. Secure the pouch around the patient s waist or over the patient s shoulder. The carry strap may be adjusted for size. Nimbus TM Flex Patient Manual - 511-IFU-IT1029, Rev. A 11

Caution: Leave the pump inside the carry pouch at all times except for programming, loading and unloading the cassette/administration set, or troubleshooting. Fluid Bag Carrying Pouch Cover Clear Viewing Window Pump Holder Patient Luer Lock Carrying Strap 12 Nimbus TM Flex Patient Manual - 511-IFU-IT1029, Rev. A

Section 3: Troubleshooting When a potentially unsafe condition is detected by the pump, the pump will exhibit an audio signal and a visual signal that indicate the presence of a condition that may interrupt the infusion. Note: The following table is included for reference only and includes references to functions that are intended to be used by medical professionals only. The user may press RUN/STOP key to acknowledge an alert message. Do not attempt to use these functions unless specifically instructed to do so and trained to do so by a physician or other certified health care professional. Alert Message Indication Signal The cassette may not be properly loaded or has become dislodged. Cassette Loading Error System Error Firmware Error Upstream Occlusion Note: The infusion will be stopped. The pumping mechanism may not work properly. Note: The infusion will be stopped. The system fails to operate in a controlled fashion. Note: The infusion will be stopped. A decrease in pressure is sensed in the supply line between the pump and IV bag. Note: The infusion will be stopped. The screen displays Cassette Loading Error ; The LED flashes red light; The audio alarm sounds. The screen displays System Error ; The LED flashes red light; The audio alarm sounds. The screen displays Firmware Error ; The LED flashes red light; The audio alarm sounds. The screen displays Upstream Occlusion; The LED flashes red light; The audio alarm will sound after 1 minute if the upstream occlusion is not resolved. Nimbus TM Flex Patient Manual - 511-IFU-IT1029, Rev. A 13

Downstream Occlusion Battery Depleted Pump End-oflife Invalid Infusion Parameter Pump Unattended Infusion Complete Approaching End-of-Life An increase in pressure is detected in the line between the pump and the patient. Note: The infusion will be stopped. The infusion will continue for 30 minutes prior to shut down. The pump has reached the end of its service life. Note: The infusion will be stopped. An infusion parameter has been entered that exceeds the remaining pump life. The pump is standing by and not infusing for more than 10 minutes. The infusion is completed. The pump is approaching its end of life. The remaining infusion volume is less than 100 ml or the remaining infusion time is less than 24 hours. The screen displays Downstream Occlusion ; The LED flashes red light; the audio alarm will sound after 1 minute if the downstream occlusion is not resolved. The screen displays Battery Depleted for 5 seconds and repeats every 20 seconds; The empty battery level indicator flashes on the screen; The LED shows solid yellow light; the audio alarm sounds. The screen displays Pump End-of-Life ; The LED flashes red light; The audio alarm sounds. The screen displays on Invalid infusion Parameter ; The LED flashes red light; The audio alarm sounds. The screen displays Pump Unattended ; The LED shows solid yellow light; The audio alarm sounds. The screen displays Infusion Complete ; The LED flashes red light; The audio alarm sounds. The screen displays Approaching End-of-life for 3 seconds; The battery level indicator symbol shows 1 bar and flashes on the screen. 14 Nimbus TM Flex Patient Manual - 511-IFU-IT1029, Rev. A

Section 4: Maintenance Cleaning and Disinfection The pump should be cleaned and disinfected daily, whenever the pump becomes visibly soiled, or following facility protocols. Turn the pump off before cleaning. DO NOT spray cleaning fluids directly onto the pump. DO NOT steam autoclave, EtO sterilize, immerse the pump, or allow fluid to enter the pump. DO NOT use hard or pointed objects to clean any part of the pump. DO NOT use acetone, other plastic solvents, or abrasive cleaner to clean the pump, as damage to the pump may occur. Acceptable cleaning and disinfection solution is: 10% bleach solution (1 part household bleach to 9 parts water) Cleaning 1. Keep the pump keypad facing up and do not allow any part of the pump to become saturated with or submersed in fluid during the cleaning operation. 2. Use soft cloth dampened with 10% bleach solution to clean all exposed surfaces until visually clean. 3. After application, wipe all surfaces with a water-dampened soft cloth. Disinfection 1. Wipe all exposed surfaces with a new piece of soft cloth dampened with 10% bleach solution. 2. Allow the pump surfaces to remain visibly wet for a minimum of 20 minutes 3. Remove bleach residue by using a water-dampened soft cloth Failure to follow these instructions may result in an electrical hazard, damage to the pump, and / or voided warranty coverage. Nimbus TM Flex Patient Manual - 511-IFU-IT1029, Rev. A 15

Note: Please do not attempt to disassemble or alter the device in any means. If you have any problems or questions regarding the pump, please contact your health care provider. 16 Nimbus TM Flex Patient Manual - 511-IFU-IT1029, Rev. A

(This page is intentionally left blank.) Nimbus TM Flex Patient Manual - 511-IFU-IT1029, Rev. A 17

(This page is intentionally left blank.) 18 Nimbus TM Flex Patient Manual - 511-IFU-IT1029, Rev. A

(This page is intentionally left blank.) Nimbus TM Flex Patient Manual - 511-IFU-IT1029, Rev. A 19

InfuTronix, LLC 177 Pine Street Natick, MA 01760 Customer Service: Phone: 855-923-2273 Email: Support@infutronix.com www.infutronix.com 20 Nimbus TM Flex Patient Manual - 511-IFU-IT1029, Rev. A