Installing a PTV Inlet

Similar documents
Installing a PTV Inlet

Installing a Purged Packed (PP) Inlet

Installing a Cool On-Column Inlet

Installing the CO 2 Cryogenic Oven Cooling Kit

Installing an Auxiliary Pressure Control Module

Installing a Purged Packed Inlet

Installation Guide. Thermal Conductivity Detector on a 6850 GC Accessory G2623A

Installation Guide. Flame Ionization Detector on a 6850 GC Accessory G2621

Installing an Auxiliary Pressure Control Module

Installation Guide. Flame Ionization Detector on a 6890 GC. Accessories G1561A, G1562A, G1591A, G1598A

Agilent G2738A Upstream Capillary Interface

Installation Guide. Purged Packed Inlet, Electronic Pneumatics Control on 6890 GC Accessory G1551A

Installation Guide. Nitrogen-Phosphorus Detector on 6890 GC Accessories G1575A, G1576A, G1594A, G1599A

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Agilent G1554A On-Column Capillary Inlet with EPC

Nickel Catalyst Tube on a 6820 GC, Accessory G4318A

Installation Guide. Auxiliary Pressure Control Manifold on 6890 GC Accessory G1570A

Installation Guide. Volatiles Interface for 6890 Gas Chromatograph Accessory G2319A

Installation Guide. Heater Interface Cable for Mass Sensitive Detector Accessory G1590A

Agilent G3510A/G3511A Multimode Inlet

Installation Guide Thermal Conductivity Detector Accessory 19232E

1310 Inlets. 1 of 36. Jun Illustrated Parts Breakdown Agilent 6890 Gas Chromatograph Service Manual

Nickel Catalyst Tube Accessory 19205A

Installing the Flame Ionization Detector EPC Flow Control Manifold

Installing the 6850 Direct Valve Column Connector Kit

Installation Guide NPD/ECD Chemical Filters Accessory 19199N

260 Solvent Vapor Exit Accessory

Valve Box and Valve Accessories G4315A, G4316A, G4317A

Agilent 6890 Gas Chromatograph

Agilent G2333B (Single) and G2334B (Dual) Flame Photometric Detectors

1350 Valves. 1 of 14. Jun Illustrated Parts Breakdown Agilent 6890 Gas Chromatograph Service Manual

Alcatel-Lucent 7310 DC Metro Cell AWS (B4) Kit MC-DBR1, MC-DBR2, MC-DBR3, MC-DBR4, MC-DBR5. Installation Manual

Installing the G2743A/G2745A Valve Actuator Kits

29 Installation Step 1. Unpacking the GC Cable diagrams Step 2. Placing the GC system on the benchtop Step 3. Turning the power on

Agilent G3348B Flame Photometric Detector

INSTALLATION OF OPTIC 3 INJECTOR SYSTEM ONTO AN SHIMADZU GC-2010 GAS CHROMATOGRAPH.

Installation Instructions. PowerFlex 700 Drive - Frame 8 Components Replacement

Agilent G2647A (Single) and G2648A (Dual) Flame Photometric Detector Update Kits

Ford 7.3L SuperDuty Cold Air Intake

Technical Tips (April 2017) Product Information

Replacing the Battery in the Patriot SPS 250, SPS 300 and SPS 450

8 Zip Tie Zip Tie 1 Union Fitting 1 ½ ½ Union Reducer Fitting Union 1 5/8 ½ (For Plastic Intake Manifold Vehicles)

Installation Note for Solvent Selection Valve Upgrade Kit (G1360A)

Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

If there is additional labor or other costs above the 4 hrs labor you must call VMAC for approved coverage

Replacing a Gun RTD. Instruction Sheet P/N B. 1. Introduction. Using this Instruction Sheet

Deans Switch Microfluidics

Agilent 7820A Gas Chromatograph

IAG Competition Series Air / Oil Separator (AOS) For 2017 STI

MEDO LA BLOWER INSTRUCTION MANUAL INSTALLATION & MAINTENANCE LA-28B/ LA-45B LA-60B/ LA-80B LA-100/ LA-120

Accessory Fuse Block. Please read this entire manual before proceeding with installation.

INSTRUCTION MANUAL INSTALLATION & MAINTENANCE LA-100/ LA-120

Agilent G4367A Auxiliary Electronic Pressure Control (Aux EPC) Module

Forbidden Diesel Performance Ford 6.7L Powerstroke EGR Delete Instructions 6.7L EGR DELETE BY FORBIDDEN DIESEL PERFORMANCE. Page 1

Agilent 5975T. Mobile Lab Hardware Installation. Agilent Technologies

Hot Injection Adapter Kit - Installation Guide

18SP680Rev3 EPA04 MBE 4000 Car Hauler Low Pressure Fuel Lines

ELECRAFT KX3 Application Note

Replacing the Batteries in the Fortress LI 660

Agilent 6850 Series II Network GC System G2630A Performance Specifications

IAG Street Series Air / Oil Separator (AOS) For 2017 WRX

Lamp House with 2.5 kw lamp

HP Adjustable Depth Fixed Rail Kit Installation

SITRANS F. Ultrasonic Flowmeters. Hi-Precision Mounts. Operating Instructions. Answers for industry.

Agilent 210/218 - Isocratic Pump

Wrenches: ⅞, 8mm, 10mm, 13mm, 19mm P. allen, Other: Electrical Tape

In addition to this instruction sheet, the following are included in this kit (some additional parts are included for future maintenance):

TOYOTA FJ CRUISER / 4RUNNER COLD AIR INTAKE Section I Installation Preparation. 4.0L V6 (1GR) Roller Rocker Part Number(s): PTR

Copper Sleeve, Unit Injector, Replacement

ELECTRIC START KIT KIT P/N

SADDLEBAG AUDIO WIRE HARNESS KIT P/N

2012 JK Vacuum Pump Relocator

INSTALLATION INSTRUCTIONS SEMI-HIDDEN WINCH MOUNT Part Number:70005 Application: Ford Super Duty

Agilent 210/218 - Binary Pump

Installation. Introduction

2015 Ford F150 Rear Bumper w/ LED

Intended Users This manual is meant for Gasboy Authorized Service Contractors (ASCs) who will install this kit on Atlas units.

B-026INT October 6, 1999 FUEL FILTER BRACKET RETROFIT (JAPAN ONLY) GENERAL RETURN THIS TO: INITIAL HERE TECHNICIAN NO.

A g i l e n t A G a s C h r o m a t o g r a p h Service Manual

Installation Instructions

Installation Manual. For ISC, ISL, ISC03 And ISL03 Engines

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

(31 ) ZEON CONTROL PACK RELOCATION KIT

(31 ) ZEON CONTROL PACK RELOCATION KIT

6890 Installing ¼ Glass Columns

cause injury to humans and other physical damage. : Gives information that does not fall in the WARNING or CAUTION categories.

HV angle valve with single acting pneumatic actuator and closing spring (NC)

Backside License Plate Mount for Jeep JK Wrangler

SCION xb PERFORMANCE AIR INTAKE Section I Installation Preparation 2.4L I-4

SOS SERIES SOS1 SOS2. Spares On Site Battery Cabinet Installation Guide rEV3

FOG LAMPS INSTALL KIT

Do not have any open flame or heat sources close to the installation

H280-series Valve Overheat / Relocation Kit Instructions: Standard Oven. Instructions for Applying Kit #

ELECTRICAL SYSTEM UPGRADE

TESLA MODEL S REAR UNDER SPOILER & DIFFUSER SYSTEM

MKVI Jetta Fog Light Kit

Installation Instructions for Chevrolet C5 Corvette Lingenfelter High Flow Fuel Pump Module

IAG Street Series Air / Oil Separator (AOS) For WRX

DURAMAX LBZ EGR DELETE

Transcription:

Agilent 6850 Series II Network GC System Accessories G3345B (Septumless) and G3346B (Septum) These instructions are divided into two parts: Part 1 to prepare the Electronic Pressure Control ("EPC") module for later installation into the Agilent 6850 Gas Chromatograph ("GC") Part 2 to perform the actual Accessory installation into the GC Part 1: Preparation of the PTV EPC module Locate the following item found in your Accessory kit: Cover Kit G2630-60081 consisting of: EPC cover 1 Hex nuts, 7/16-inch 2 Tube, stainless steel (pictured where used) 1 Nut & ferrule set, 1/8-inch (pictured where used) 1 Reducing union, 1/8- to 1/16-inch (pictured where used) Qty. 1

5/16-inch hex nuts EPC module cover Caution Some of these assemblies contain printed circuit boards so standard ESD precautions must be followed: use a static control wrist strap (supplied) connected to a suitable ground in handling the assemblies. 2

Tools Required The following tools are required for this assembly: Tools Required 7/16-inch open end wrench or nut driver Procedure 1. Prepare a suitable clean, dry, static-free work area. 2. Place the inlet system assembly such that its EPC module is centered in your work area with its ribbon cable extending to the right. 3. Remove any dust caps from gas connection fittings. 4. Fold the cable back on itself and gently crease it about midway along its length. 3

Tools Required 5. Next, fold the cable back on itself again, but at an angle such that it now extends at 90-degrees in the direction of the gas connection fitting. Again, gently crease the cable at the fold. 4

Tools Required 6. Install the cover onto the two gas fittings making sure the ribbon cable and its connector exit through the slot provided. Secure the cover using 7/16-inch hex nuts, one onto each fitting. 7. Replace protective dust caps onto gas connection fittings to maintain cleanliness. 8. Orient the module with its PC boards face-up and remove the two connectors (one yellow, one white) from the small thermocouple board. 9. Then remove two screws to remove the thermocouple board from the EPC module. Do NOT remove the indicated brass standoffs. 5

Tools Required Disconnect Remove Standoffs (do not remove) 10. Install the thermocouple board into its own cover using the screws removed in Step 9. 6

Tools Required Caution For the next steps, to prevent twisting of tubing and/or connected fittings, it is best to perform tightening using two wrenches against each other. 1/8-inch fitting 1/16-inch tubing PTV inlet body 1/8-inch nut / ferrule set 1/8- to 1/16-inch reducing union 11. Finally, at the inlet body, swage the 1/8-inch end of the tube supplied onto the inlet body fitting using the supplied 1/8-inch nut and ferrule set. If present on the inlet body fitting, remove and discard its dust cap. 7

Tools Required 12. At the opposite end of the now-installed tube, swage on the 1/16-inch end of the supplied 1/8- to 1/16-inch reducing union. This completes EPC module and inlet body preparation for this Accessory. 8

Tools Required Part 2: Installation of the PTV Accessory There are kits for installing both septum and septumless PTV inlets. This document describes both installations. Each kit contains: Description Quantity Round grommet, small black 1 Round grommet, large blue 6 M4 machine screw, 12 mm 16 M4 hex nut w/lockwasher, 3.2 mm 1 Cable ties 6 Antistatic pouch 1 Disposable wrist strap 1 Insulation block 1 Extension cable 1 Top cryo bracket 1 Bottom cryo bracket 1 Eyelet 4 Inlet chassis 1 LC0 2 cryo assembly 1 EPC septumless head PTV inlet assembly (G3345B)* or EPC septum head PTV inlet assembly (G3346B)* Ship kit 1 Lowbld septa 5, 11-mm 1 Open-end wrench, 6-mm 1 Open-end wrench, 5-mm 1 1/4 Brass union 1 OEM ship kit 1 1 9

Tools Required Description Quantity Installation sheet (this document) 1 *The inlet is factory-assembled. Do not disassemble it during installation. 10

Tools Required Kit Contents Screws, grommets, eyelets, nut LCO 2 cryo assembly Extension cable Insulation block Inlet chassis Top cryo bracket PTV inlet Covered board assembly Bottom cryo bracket Flow module Required Tools 11

Tools Required T-20 Torx screwdriver 7/16-inch open end wrench Overview Caution Before starting, review the safety information listed at the end of this document. 13. Prepare the GC. 14.. 15. Install the new inlet. 16. Restore the GC to operating condition. Prepare the GC WARNING Note the following warnings: Use extreme caution when working with cryogenic liquids. They can cause very serious burns requiring emergency medical attention. Before separating fittings, always bleed lines containing cryogenic fluids to atmospheric pressure. Wear safety glasses and cover the fittings with a towel when separating them. Before working on an instrument, disconnect the power cord to avoid a potentially dangerous shock hazard. Hazardous voltages are present in the instrument whenever the power cord is connected. Pressurized liquid CO 2 is a hazardous material. Take precautions to protect personnel from high pressures and low temperatures. CO 2 in high concentrations is toxic to humans; take precautions to prevent hazardous concentrations. Do not use copper tubing or thin-wall stainless steel tubing with liquid CO 2. Both harden at stress points and may explode. Consult your local supplier for recommended safety precautions and delivery system design. 12

Tools Required Caution Do not use liquid CO 2 as a coolant for temperatures below 40 C. The expanding liquid may form solid CO 2 dry ice in the GC oven. If dry ice builds up in the oven, it can seriously damage the GC. To prepare the GC: 1. Turn off the GC and unplug the power cord. Allow time for all heated zones to cool. 2. Open the lid. If a column is installed, disconnect it at the detector end. Remove the nut warmer, insulation, and capillary adapter, if present. Close the lid. 3. Turn all gases off at their sources. Disconnect the carrier and detector gas tubing from the back panel of the instrument. Disconnect gas supply tubing from these fittings 13

4. Remove the lid top cover. Remove vent tubing, if present Lid top cover, without valve box accessory Remove T-20 Torx screws (8 places) Remove detector cover Caution Caution Caution As part of this step and those that follow, you must bend various sections of tubing. Make the bends gradual and avoid kinks. It is neither necessary nor advisable to separate the inlet from its pneumatics module. Doing so can create leaks. Although handling the inlet and pneumatics module as a unit is awkward, it can be managed. This procedure requires precautions against electrostatic discharge. Use the grounded wrist strap (part number 9300-1408) and connect it to a bare metal surface of the GC. Failure to heed this caution may result in damage to the instrument or to the PTV assembly. 1. Loosen the screws holding the connector cover plate next to the inlet flow module. 2. Slide the cover plate off and disconnect the cable from the flow module. 3. Disconnect the gas fitting on the rear of the flow module. 14

4. Remove the three screws on top of the flow module. (2) Slide (3) Disconnect (4) Remove (1) Loosen 5. Complete the appropriate following steps (based on the type of inlet you are removing) to finish removing the inlet. Cool On-Column inlet a. Use a T-20 Torx screwdriver to loosen the three captive screws that attach the inlet weldment to the top of the inlet carrier. b. Slide the inlet up and out of the carrier. If necessary, you can also slide the insulation sleeve off of the bottom of the inlet. Purged Packed inlet a. Trace the heater/sensor cable from the inlet to the wiring harness connector. Disconnect it. b. Remove the three screws holding the inlet. c. Lift the inlet and flow module out of the lid. Remove the insulation from the hole under the inlet. 15

Programmable Temperature Vaporization inlet a. Lift up the PTV inlet. b. Disconnect the power cable, thermocouple connector, and cryo connector from the flow module and board. Disconnect the module ribbon cable from the pneumatics board. c. Slide the flow module out of the chassis, remove the chemical trap assembly from the mounting bracket, and remove the PTV assembly from the GC. Split/Splitless inlet a. Use the inlet wrench to release the large nut on top of the inlet. Large nut Insert weldment (includes nut) b. If you do not have a gas or liquid sample valve Remove the insert weldment and the flow module. c. If you do have a gas or liquid sample valve The insert weldment is part of the valve assembly. Move it out of the way. Disconnect the gang block fitting (also part of the valve assembly), from the flow module. Remove the flow module. d. Use forceps to remove the liner and O-ring. e. Disconnect the split vent trap. Disconnect split vent trap here Top of liner 16

f. Trace the heater/sensor cable to the wiring harness connector. Disconnect it. g. Remove the three screws holding the inlet body. h. Lift the inlet assembly out of the lid. Remove the insulation from the hole under the inlet. Install the new inlet 1. Insert the inlet chassis into the inlet opening in the top of the GC. As you face the front of the GC, the raised edge of the chassis is on the left. 2. Fit the insulation cup into position on the inlet body. Be sure the slot in the insulation cup is properly fitted with the inlet. Flow module Slot Inlet with insulation cup 3. Place the inlet flow module into position. Be sure that the lip at the bottom of the flow module fits into the slot in the bottom of the flow module opening. See the picture with Step 5. 17

4. With the flow module in place, insert the inlet and insulation cup into the inlet chassis. Fasten the inlet and chassis into position with three T-20 Torx screws. Screws Inlet chassis Inlet tubing 5. Route the inlet tubing as shown. At the rear of the GC, follow the "C" (carrier) tubing. Using three or four cable ties, tie the inlet tubing with the other tubing found under the lid at evenly spaced intervals. Be careful to avoid bending the tubing sharply. Close the lid. 18

Flow module Lid Inlet tubing tied together four places (circled) 6. Place the covered board assembly into position as shown. Fasten the assembly down with two screws. On top of the lid, insert the captive screw into one of the vent holes in the covered board assembly. Beneath the lid, fasten the hex nut onto the captive screw. 19

Screw Captive screw (along side the cover in a vent hole) Covered board assembly Screw Captive screw and hex nut 7. Close the lid. 20

8. Attach the red cryo extension cable connector to the covered board assembly as shown. Run the loose end of the cabling behind the inlet flow module and to the rear of the inlet flow module. Be sure to plug the center connector s positive and negative sides correctly. Covered board assembly Center connector negative Center connector positive Red cryo extension cable connector 21

9. Loosen the screws on the detector module and move it forward far enough to lower the cryo extension cable connector through the top of the lid. Tie the red cable in with the other wiring, leaving as much slack as possible to avoid straining the soon-to-be-made connection. Red cable 10. If you WILL NOT be installing the Cryo kit, prepare the bottom bracket by inserting the six blue grommets into the holes. Insert the four brass eyelets into the top four grommets as shown. Grommets (4) with eyelets Install PTV valve here Grommets (2) only If you ALREADY have installed the Cryo kit, disconnect the coolant outlet union (see the picture with Step 12) and remove the screws on the bracket to pull it far enough away from the GC to proceed with Step 11. 22

11. Attach the cryo valve onto the lower-left position of the cryo bracket, including the screws to hold it in place. Screws 23

12. Prepare to fasten the cryo bracket into place by first pulling the cryo valve tubing and wires through the hole on the side of the GC. You may not have the upper valve pictured below if you do not have the Cryo kit installed. Coolant outlet union 24

13. Place the cryo bracket over the four mounting holes on the left side of the GC. Insert screws through the brass eyelets and secure them loosely. Reattach the coolant outlet union. Mounting holes w/bracket in place 25

14. Open the lid and connect the valve wiring to the previously-routed red cryo extension cable connector inside the left panel of the GC. PTV valve wire/ red cable connect Cryo valve tubing Extension cable 26

15. Route the cryo valve tubing around the counterbalance cam to the back of the GC to meet the inlet tubing you routed in Step 5. Use a couple of cable ties to tie the tubing together. Connect the tubing. Lid Counterbalance cam Cryo tubing Inlet tubing fitting 16. Continue installing the Cryo kit, if applicable. See the documentation that came with that kit for more information. 17. Close the lid and put the top cryo bracket on. Restore the GC to operating condition 1. Install the capillary adapter, if used. 2. Restore the column connection. 3. Install the lid top cover. 4. Restore carrier and other gases to the instrument. 5. Restore power. 6. Apply your normal operating pressures. Leak-check the manifold, back panel, and column fittings. 27

6 Agilent Technologies, Inc. 2007 All Rights Reserved. Reproduction, adaptation, or translation without permission is prohibited, except as allowed under the copyright laws. Part number G3345-90008 First Edition, July 2004 Second Edition, March 2007 Printed in USA Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 Acknowledgements Torx is a U.S. registered trademark of Textron, Inc. Safety Symbols Warnings in the manual or on the instrument must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions violates safety standards of design and the intended use of the instrument. Agilent Technologies assumes no liability for the customer s failure to comply with these requirements. In the manual A warning calls attention to a condition or possible situation that could cause injury to the user. A caution calls attention to a condition or possible situation that could damage or destroy the product or the user s work. On the instrument See accompanying instructions for more information. Indicates a hot surface. Indicates hazardous voltages. Indicates earth (ground) terminal. Indicates explosion hazard. Indicates radioactivity hazard. Indicates electrostatic discharge hazard. Pinch hazard.