A GERMAN TORQUE COMPARISON FROM 20 N m TO 200 N m AND ITS RESULTS

Similar documents
INFLUENCE OF CROSS FORCES AND BENDING MOMENTS ON REFERENCE TORQUE SENSORS FOR TORQUE WRENCH CALIBRATION

PERFORMANCE OF THE NEW PRIMARY TORQUE STANDARD MACHINE OF INMETRO, BRAZIL

LEVER OPTIMIZATION FOR TORQUE STANDARD MACHINES

A PORTABLE SYSTEM FOR THE CALIBRATIONOF TRANSDUCERS AND TORQUE WRENCHES: THE CALIBRATION BELL

ANALYTICAL STUDY ON THE UNCERTAINTY OF LOAD CELLS CALIBRATED WITH DEADWEIGHT-FORCE-STANDARD MACHINE AND FORCE-COMPARATOR MACHINE

Hybrid torque standard machine for 1 kn m developed in CENAM

IMGC-CNR Activities to Improve the Force Chain in Italy. C. Ferrero and C. Marinari Istituto di Metrologia G. Colonnetti (IMGC-CNR) Torino - ITALY

TRACEABILITY OF THE CALIBRATION OF TEST CAR FOR ROLL BRAKE TESTER Aimo Pusa Raute-MIKES Finland, by Raute Precision Oy, Force and Mass Laboratory

Tension and Compression Load Cell Model 8435

THE DESIGN OF 1 N M TORQUE STANDARD MACHINE AT NIM

Add-on of the 2-MN force standard machine for testing load cells

K3D300 ±200kN. Description

Multi capacity load cell prototype

K3D300 ±50kN. Description

National comparison on verification of fuel dispensers

Operation Manual for Torque Sensors

Industry 4.0 Is Unthinkable Without Calibration!

RENOVATION OF NMIJ'S 5 MN HYDRAULIC AMPLIFICATION TYPE FORCE STANDARD MACHINE

Evidence of performance Reduction of impact sound pressure level by floor coverings

Metal forming machines: a new market for laser interferometers O. Beltrami STANIMUC Ente Federate UNI, via A. Vespucci 8, Tbrmo,

Operation Manual for Torque Sensors

Publishable Summary for 14IND14 MNm Torque Torque measurement in the MN m range

Torque transducers the complete range for all tasks. Sensors and transducers

Global Flow Measurement Calibration Consistency

A New verification system of mobile scales

CB50X & CB50X-DL load cells Influence factors in weighbridge application

JJG Translated English of Chinese Standard: JJG Wayne Zheng et. al

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF TORQUE METROLOGY (Torque Generating Device Calibration)

Development of Dynamic Calibration Machine for Pressure Transducers

PW29... Data Sheet. Single point load cell. Special Features. Optional: Dimensions (in mm; 1 mm = inches) B en

WEAR PROFILE OF THE CYLINDER LINER IN A MOTOR TRUCK DIESEL ENGINE

KM26z 20N; 50N; 100N, 200N, 500N, 1kN, 2kN, 5kN

Operation Manual for Torque Sensors with Slip Ring Transmission

Optimum design of nozzles tribology systems of a diesel engine fuel injector with high values of rail pressure

Skid against Curb simulation using Abaqus/Explicit

Pressure balance Differential pressure version Model CPB5600DP

Draft Indian Standard PERFORMANCE REQUIREMENTS AND METHODS OF TEST FOR WHEELS FOR PASSENGER CARS/MINI GOODS CARRIERS (First Revision of IS 9436)

Test Report No

Load Cell for Manually Operated Presses Model 8451

Verifying the accuracy of involute gear measuring machines R.C. Frazer and J. Hu Design Unit, Stephenson Building, University ofnewcastle upon Tyne,

ASTM E74-13a Explained

Measuring equipment for the development of efficient drive trains using sensor telemetry in the 200 C range

Operation Manual for Torque Sensors

Certificate of Calibration

Confidence in measurement.

Presented By. NicolScales.com Contact Us

Influence of mounting tube diameter on anemometer output

Defining Parameters for Ballistic High Pressure Sensors. Transducer Workshop June 20-21, 2000 Aberdeen, MD

Small Tool Instruments and Data Management

Annex to the Accreditation Certificate D PL according to DIN EN ISO/IEC 17025:2005

FRONTAL OFF SET COLLISION

Load Cell Troubleshooting

DL 01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05

PTB Ex Proficiency Testing Program (PTB Ex PTP) Interlaboratory comparisons in explosion protection

MEASURING CUSTOMER COMFORT OF THE SIDE DOOR SELF-EQUIPMENT FOR THE AUTOMOBILE INDUSTRY

EMS ELONGATION MEASUREMENT SYSTEM. Strain measurement system for wind turbines optimizing the control & condition monitoring

FP 6. EU Proposal SALMAPS

T8 Torque Sensor Operation Manual

Dusty / Dusty Ex. Low-Cost Broken Bag Detection. Operating Instructions. SWR engineering Messtechnik GmbH

LOW SHOCK RELEASE UNIT EASY RESETTABLE AND 100 % REUSABLE. Jens Müller 1, Christoph Zauner 2

KD45 ±20N. Description

76th UNECE GRPE session

Inspekt. Static electromechanical universal testing machines. Competency generates results

Hydraulic Deadweight Tester Model CPB5800

Instruction Manual. Rotating Torque Sensor Type 4501A. Version Q/QA (with standard square connections) Version R (with shaft ends)

Evidence of Performance Airborne sound insulation of flat roofs

THE PEP PARTNERSHIP ON ECODRIVING Goals, achievements and next steps November 2017

Innovative Testing Equipment. Torque sensors Vehicle Applications Actuators

Digital torque transducers. T40 family. For high-precision torque measurement technology

Installation and Operational Instructions for EAS - HTL housed overload clutch Sizes 01 3 Type 490._24.0

CHAPTER 6 MECHANICAL SHOCK TESTS ON DIP-PCB ASSEMBLY

Transmission Error in Screw Compressor Rotors

ABB Instrumentation. Thermal Mass Flowmeter Sensyflow. Intelligent solutions for accurate gas mass flow measurement

Calibration, Fluid Property Effects Reproducibility and Long Term Stability. All in 20 minutes! Dr Gregor Brown Cameron

HLC A1... HLC B1... HLC F1...

Implementation of Rolling Resistance in R117

Load Cells SIWAREX WL230 BB-S SA

Field test for the comparison of LNG static and dynamic mass measurement methods

Operation Manual for Plug and Play USB Torque Sensor

Version 2016 / 07 Stator All-In-One

Customer:... Model Number:... Serial Number:... Factory Reference Number:... Rated Torque (lbf-in):... Torque Overload Capacity (lbf-in):...

STRUCTURAL BEHAVIOUR OF 5000 kn DAMPER

Load cells. Data sheet. Special features. Cable assignment (six wire configuration)

Technical Bulletin. Proposed Pedestrian Grid Procedure - Data Collection. Version 1.0. November 2010 TB 010 TB010-1

CERTIFICATE. EN tested QAL 1 certified TUV approved Annual inspection. The certificate is valid until: 19 July Köln, 17 August 2012

mga research corporation

Research And Development Of Variable-Speed Scroll Compressor

Operating Efficiency of Crankshaft Drive Pumps

Certificate: / 22 March 2013

Theory of Machines II EngM323 Laboratory User's manual Version I

Annex to the Accreditation Certificate D PL according to DIN EN ISO/IEC 17025:2005

Operating Instructions Precision balance

Determination and improvement of bevel gear efficiency by means of loaded TCA

Volkswagen DCC Adaptive Chassis Control - Design and Function DCC Adaptive Chassis Control. Basics of the damping system

CALIBRATION SERVICES HUMIDITY DEW POINT AIR VELOCITY MASS FLOW TEMPERATURE PRESSURE

Product specification

DATAFLEX 42/1000 Torque Measuring Shaft Operating/Assembly instructions DATAFLEX

PRECISION BELLOWS COUPLINGS

CheckStar In line torque transducers with angle measurement option

G.U.N.T. Gerätebau GmbH

Transcription:

IMEKO 22 nd TC3, 12 th TC5 and 3 rd TC22 International Conferences 3 to 5 February, 2014, Cape Town, Republic of South Africa A GERMAN TORQUE COMPARISON FROM 20 N m TO 200 N m AND ITS RESULTS Dirk Röske Physikalisch-Technische Bundesanstalt (PTB), Braunschweig, Germany, dirk.roeske@ptb.de Abstract: In October 2009, the members of the German DKD Torque Committee (FA 10) agreed to carry out a comparison of torque calibrations according to the German torque calibration standard DIN 51309:2005-12 in the range from 20 N m to 200 N m using very stable torque transducers as travelling standards. The pilot laboratory was the torque working group at PTB. Thirteen accredited laboratories from Germany, together with another German laboratory without accreditation and an Austrian laboratory, participated in this comparison. All measurements were carried out in 2010 and 2011, and the first results were discussed in 2012. The report on the comparison was expected to be published later in 2013. This paper summarizes the results for discussion in the international community. In general, there was very good agreement between the results of the participants and the pilot laboratory. Keywords: torque, DKD, DIN 51309, 200 N m, comparison. 1. INTRODUCTION During a meeting of the Torque Committee (FA 10) of the German Calibration Service (DKD) in October 2009, the members decided to carry out a 200 N m torque comparison with torque transducers (not torque wrenches). The aim was to compare the calibration results and issued certificates of laboratories accredited in this range. The comparison was open to participants without accreditation. PTB s torque working group prepared the technical protocol of the comparison, including the timetable. The set of transducers and additional equipment to be sent to the participants were defined. The laboratories had to carry out the measurements and fill in Excel worksheet templates with their measured data, thus allowing a faster data processing by the pilot. The pilot laboratory investigated the stability of the travelling standards, evaluated the data submitted by the participants, applied corrections to this data for the amplifier deviations and calculated final E n values. It is interesting that while the comparison was running, the old DKD (it was dealing with accreditations of calibration laboratories until the end of 2009) became a part of the new German Accreditation Body (DAkkS) in 2010. The DKD was newly founded on 3 rd May, 2011 as a PTB panel for the cooperation between Germany s national metrology institute PTB and accredited laboratories. 2. PARTICIPANTS AND DATES Table 1 shows the 16 laboratories participating in the comparison: one from Austria, all others from Germany. Abbreviation Accred. number HBM D-K-12029-01 Ford D-K-15088-01 Schatz D-K-17572-01 GTM D-K-15106-01 DSM D-K-15152-01 Kopp D-K-15180-01 Kistler D-K-17650-01 Lorenz D-K-17603-01 Atlas Copco D-K-17447-01 DmS Dr. Peschel D-K-15165-01 MPA DA D-K-11048-01 Q-direct D-K-15001-01 Deprag DKD-K-51101 TÜV AT without Porsche without PTB without Company name location Hottinger Baldwin Messtechnik GmbH Darmstadt Ford-Werke GmbH Köln Schatz AG Remscheid Gassmann Testing and Metrology GmbH Bickenbach DSM Messtechnik GmbH Aalen Kalibrierdienst Kopp GmbH Wiesloch Kistler Lorch (formerly Dr. Staiger, Mohilo + Co. GmbH) Lorch Lorenz Messtechnik GmbH Alfdorf Atlas Copco (formerly TBB Industrial Tools Services GmbH) Dingolfing DrehmomentService Dr. Peschel Wendeburg, OT Neubrück Staatliche Materialprüfanstalt Darmstadt Institut für Werkstoffkunde, TU Darmstadt Darmstadt Q-direct GmbH Loiching-Kronwieden DEPRAG SCHULZ GmbH u. CO. Amberg TÜV Austria Services GmbH Vienna / Austria Porsche AG Weissach Physikalisch-Technische Bundesanstalt WG 1.22 Realization of Torque Braunschweig

3. TRAVELLING STANDARDS The following equipment was sent to the participants: Torque transducer 1: type TB2 (with fixed adaptors) capacity: 200 N m make: Hottinger Baldwin Messtechnik, Germany serial number: #080830117 dimensions: Ø 103 mm 253 mm (with adaptors) interface: Ø 50h7 mm 80 mm shaft ends mass: 5.7 kg (with adaptors) packaging: in plastic bag inside a transportation box sensor type: strain gauge cable: TB2 LEMO 06-1.22, 6 wires, length 3 m connectors: Lemo FGG6 (transducer) DB-15 Torque transducer 2: type TT1 capacity: 200 N m make: Raute Precision Oy, Finland serial number: 36077-00, bridge M1 dimensions: Ø 57 mm 224 mm interface: Ø 30h7 mm 60 mm shaft ends mass: 1.4 kg packaging: in plastic bag inside a transportation box sensor type: strain gauge cable: K010-1.22, 6 wires, length 3 m connectors: DB-15 (transducer) DB-15 Additional equipment: Bridge calibrator K148 make: Hottinger Baldwin Messtechnik, Germany type: K148 serial number: K148-0100 dimensions: 330 mm 270 mm 75 mm mass: 3.5 kg with power supply unit cable: K148-single channel Data logger make: MSR Electronics, Switzerland type: MSR 145 serial number: 302434 dimensions: 70 mm 40 mm 22 mm mass: 0.1 kg Serial RS-232 cable. The comparison was of a combined ring-star design. A ring consisted of a maximum of two participants and began as well as ended with at least a stability measurement in the pilot laboratory. The pilot end measurement of one loop was the start measurement for the following loop. These loops formed the rays of the star. There were 9 of these loops, and within each loop the measurements were carried out first on the machine with the lower uncertainty. Fig. 1 shows an example of a measurement result in the pilot laboratory: the deviation in parts per million is calculated for each mounting position in relation to the mean result from all mounting positions. It is a good indicator for the reproducibility of the generated torque and the alignment of the machine axes. Rel. dev. from mean in ppm 20 15 10 5 0-5 -10-15 -20 Figure 1: Example of a measurement result in the pilot laboratory 4. STABILITY OF THE STANDARDS The stability of the travelling standards is essential for a good result of the comparison. A much better indicator for the long-term stability of the transducer is the deflection, the signal at nominal torque reduced by the indication at zero torque. Fig. 2 shows the measured deflection of the TT1 transducer at 200 N m over a period of 16 months. The dashed red lines indicate an interval with a (relative) width of 3 10-5. Signal at 200 N m in mv/v 1.32776 1.32774 1.32772 1.32770 1.32768 1.32766 1.32764 1.32762 1.32760 1.1.10 2.7.10 1.1.11 2.7.11 Date Figure 2: Stability of the TT1 transducer over 16 months The bridge calibrator K148 (accuracy lever of 0.0025%) was used by the participants to check their own amplifiers. For comparisons of torque standard machines at a level of a few 10-5, usually a DMP40 is used together with a high-end bridge standard BN100. In this comparison, the best relative measurement uncertainty of a participant s machine is 10-4. For this purpose the K148 was sufficient. The DMP40 amplifier (accuracy lever of 0.0005%) of the pilot laboratory was calibrated each time when the transducers came back. Over the whole period of the comparison, the absolute

deviation between amplifier and bridge standard was between -20 nv/v and +80 nv/v with a maximum span of ±10 nv/v at the maximum deviation. The deviation between the amplifiers used by the pilot and by the participant was taken into account by means of correction factors that were applied to the data. The data logger MSR 145 was used to record the environmental conditions temperature and relative humidity during transportation and measurements in the loops. It was found that nearly all laboratories work under very stable environmental conditions; only one participant had a problem with the air conditioning. The temperature was between 20 C and 23 C, the relative humidity between 25% rh and 50% rh. 5. TYPES OF TORQUE CALIBRATION MACHINES The participants use different kinds of torque calibration machines in their laboratories: deadweight machines, pendulum-weight machines, jockey-weight machines and reference machines; the latter with vertical or horizontal axis. The uncertainties of measurement achievable with these machines depend on the design: deadweight machines are usually more accurate than reference machines utilizing a calibrated reference torque transducer. Air bearings offer a lower uncertainty than other types of bearing or suspension. Four of the participants (pilot not counted) operate a deadweight torque machine similar to that of the pilot laboratory. The principle of operation is shown in Fig. 3. lead to a change of the acting torque due to a displacement of the lever s centre of gravity resulting in an additional countertorque, or due to a shorter length of the lever arm. The best measurement capability (k = 2) for which the four laboratories are accredited is 1 10-4. The second type of machine using a deadweight is the pendulum-weight machine (Fig. 4). Here, a drive is necessary to generate a rotation of the weight around the torque axis. The centre of gravity of the weight will then move along a circular line. The weight force is always directed vertically towards the centre of the Earth and a torque is generated when the centre of gravity of the weight is not exactly below the torque axis (rotational axis of the bearing). For the calculation of the torque value it is necessary to know the horizontal distance between the force acting line going through the centre of gravity of the weight and a line parallel to it going through the torque axis. The torque is then the product of the found distance and the acting weight force. In a real setup, the above distance cannot be found easily, since the real position of the weight s centre of gravity is unknown. The pendulum weight can be calibrated using a horizontally positioned two-armed lever and well-known weights at precisely defined distances in a free setup, where no other parts are connected to the pendulum. In the measurement, the torque is calculated from the rotational angle of the pendulum which itself is measured using an angular encoder. One laboratory is accredited with this type of machine and the best measurement capability (k = 2) is 1 10-4. Figure 4: Pendulum-weight torque machine Figure 3: Deadweight torque machine In the following figures, the same colour scheme is used for similar parts: the weights are blue, the lever is red, the bearing is green, the transducer under test is yellow, the reference transducer is pink and the torque drive is orange. The counterdrive in Fig. 3 is magenta. It has no direct function for torque generation but it is used to ensure the horizontal position of the lever during the calibration. Otherwise the elastic deformation along the torque axis can There is a third type of machine using a dead-weight: the jockey-weight type. Here, a weight is moved along the horizontally positioned lever (Fig. 5). As described above, the torque is determined by the product of the horizontal distance between the weight s centre of gravity and the torque axis with the acting weight force. Again, it is not easy to define the exact position of the centre of gravity. Other influences are caused by the driving mechanism that is necessary for moving the weight. One solution for this problem is the calibration of these machines using torque transfer transducers. Thus there is no direct traceability to mass, length and acceleration. As a result, the

uncertainty is slightly higher, but on the other hand, this method is much simpler and cheaper. Two laboratories are accredited with this type of machine and the best measurement capability (k = 2) is 2 10-4. The principle of operation is quite simple: reference transducer and transducer under test are mounted in line. A drive (orange part in the figures) generates a torque in the system. The torque is measured by the reference transducer and this value is usually used for the control of the machine, and it is recorded together with the indication of the transducer under test. The accredited expanded uncertainties (k = 2) with this type of machine are 2 10-4 (one), 1 10-3 (five) and 2 10-3 (two laboratories). 6. FURTHER DETAILS OF THE COMPARISON Figure 5: Jockey-weight torque machine The remaining eight laboratories operate reference torque calibration machines: six of them with a vertical axis (Fig. 6), another two with a horizontal axis (Fig. 7). Figure 6: Reference torque machine (vertical axis) The measurements were carried out according to DIN 51309 (three mounting positions at each 120, one repeated incremental series at 0, both clockwise and anticlockwise torque directions) between February 2010 and June 2011. One participant measured at 90 positions due to the square drive adaptation. The pilot laboratory performed 12 measurements in total on each of the two transducers; most of them were short stability measurements. According to the first version of the protocol, the comparison was to be finished at the end of 2010. One participant could not perform the measurements according to this first schedule because of problems with unstable environmental conditions in the laboratory. The measurements were therefore conducted in 2011. Two other laboratories joined the comparison after the first schedule was agreed upon. These participants carried out their measurements also in 2011. After all data had been received by the pilot, a first version of the evaluation was discussed in October 2011 during a meeting of the DKD s Torque Committee FA 10. The data showed that there is obviously an influence of the machine s misalignment on the results. At the next meeting it was decided to also measure the sensitivity of both transducers to cross forces and bending moments by applying a welldefined force (deadweight) acting perpendicular to the axis of the transducers at a known distance (measured along the axis) from its centre. The transducers were rotated about their axes in order to obtain the dependence of the output signal on the direction of cross force and bending moment. According to the setup of this measurement, both quantities were applied in superposition; it was not possible to apply a pure bending moment or a pure cross force. One of the participants (D-K-15165-01) agreed to carry out these measurements. Last but not least, not only were the measurement results compared but also the calibration certificates issued by the participating laboratories were evaluated. These certificates were checked for completeness and formal requirements. It was also checked whether correct results and uncertainties were calculated. 7. RESULTS Figure 7: Reference torque machine (horizontal axis) The E n value calculated according to the common formula (1) = Lab Ref (1) Lab + Ref

has a disadvantage: a poor misalignment of a machine that has to be compared with other machines of the same type and better alignment usually yields a higher expanded measurement uncertainty U Lab associated with the measurement result X Lab. This will reduce the absolute value of E n, thus improving the result of this laboratory. In order to minimize this influence, a stricter version of (1) was used in this comparison. Instead of using the expanded measurement uncertainties U Lab found in the calibration, the best measurement capability (U Lab,bmc) of the participating machines was taken for the calculation according to: = Lab Ref Lab,bmc (2) + Ref,bmc The misalignment was investigated for each participant by evaluating the results of each mounting position similar to the result shown in Fig. 1 for the pilot laboratory. E n E n 1.0 0.8 0.6 0.4 0.2 0.0-0.2-0.4-0.6-0.8-1.0 Figure 8: Example: E n values for a measurement with TT1, deadweight machine, b.m.c.: 1 10-4, (full symbols: incremental torque, empty symbols: decremental torque) 1.0 0.8 0.6 0.4 0.2 0.0-0.2-0.4-0.6-0.8-1.0 Figure 9: Example: E n values for a measurement with TB2, pendulum-weight machine, b.m.c.: 1 10-4, (full symbols: incremental torque, empty symbols: decremental torque) In general, the vast majority of the E n values calculated according to (2) lies in the range between -1 and 1. Some examples are given in Figs. 8 through 10. One participant had a problem with one of the two transducers; the results of the second one was excellent. This one measurement was repeated later on and the problem did not occur anymore. Three participants had a very small number (one up to three out of 60, i.e. 2% to 5%) of E n values that fall slightly outside the above range, often values calculated for decremental torque. The reason can be friction in the driving components. One participant used square drive connectors which were not covered by the scope of this comparison. In this case 12, out of 60 E n values were below the interval with a minimum of -1.4. This participant increased the b.m.c. value after the comparison. E n 1.0 0.8 0.6 0.4 0.2 0.0-0.2-0.4-0.6-0.8-1.0 Figure 10: Example: E n values for a measurement with TT1, jockey-weight machine, b.m.c.: 2 10-4, (full symbols: incremental torque, empty symbols: decremental torque) Figs. 11 and 12 show the results of the cross force/bending moment test of the two transducers. Rel. signal change in % 0.04% 0.03% 0.02% 0.01% 0.00% -0.01% -0.02% -0.03% -0.04% 0 60 120 180 240 300 360 Angle in Figure 11: Relative change in the output signal of the TT1 torque measuring bridge related to the signal at full load, where a cross force of 102 N and a bending moment of 11.2 N m are applied and the transducer is rotated around its horizontally directed axis (symbols: measured values, line: fitting function)

As expected, the dependence of the output signal on the rotational angle follows a sine function. Due to the different mechanical design, the TB2 s cross force/bending moment sensitivity is higher than that of the TT1. But this cannot explain all of the results since some of the participants got better results with their TT1 measurement, others with their TB2 calibration. It is possible that the transducers and the machines must be investigated separately with pure bending moment and pure cross force to reveal the underlying effects. The examination of the calibration certificates revealed some minor formal mistakes. Most of them were corrected immediately after they became known to the participants or on the occasion of laboratory assessments. All results will be published in the Report [2]. Rel. signal change in % 0.04% 0.03% 0.02% 0.01% 0.00% -0.01% -0.02% -0.03% -0.04% 0 60 120 180 240 300 360 Angle in Figure 12: Relative change in the output signal of the TB2 torque measuring bridge related to the signal at full load, where a cross force of 158 N and a bending moment of 18.2 N m are applied and the transducer is rotated around its horizontally directed axis (symbols: measured values, line: fitting function) 8. CONCLUSIONS The comparison showed very good agreement between the results of the 15 participants and the pilot laboratory. It supports the claimed uncertainties of measurement for the calibration laboratories. The examination of the calibration certificates was helpful for improving these documents from both formal and objective points of view. 9. REFERENCES [1] D. Röske, Technisches Protokoll des DKD-Ringvergleiches nach DIN 51309:2005-12, 20 N m bis 200 N m (Technical Protocol of the DKD round robin comparison according to DIN 51309:2005-12: 20 N m to 200 N m), (DKD FA10 document, not published, in German) [2] D. Röske, DKD-E 10.1 Bericht über den DKD-Ringvergleich nach DIN 51309:2005-12: statische Drehmomente von 20 N m bis 200 N m (Report on the DKD round robin comparison according to DIN 51309:2005-12: static torques from 20 N m to 200 N m), Draft (in German)