Totally Automated Method for the Determination of Sudan Dyes in Food via On-Line Filtration, SPE and HPLC Analysis

Similar documents
Online sample cleanup on the Agilent 1290 Infinity LC using a built in 2-position/6-port valve

Determination of Sudan Dyes I IV in Curry Paste

High Sensitivity UHPLC-DAD Analysis of Azo Dyes using the Agilent 1290 Infinity LC System and the 60 mm Max-Light High Sensitivity Flow Cell

Impurity Testing of Fixed-Dose Combination Drugs Using the Agilent 1290 Infinity II HDR-DAD Impurity Analyzer Solution

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

Stability Axial Compression Technology Application Ordering Information

Application Note. Determination of Oxygenates in C2, C3, C4 and C5 hydrocarbon Matrices according ASTM D using AC OXYTRACER

Method Detection Limits for EPA Method 8015 Diesel Range Organics using Fully Automated Extraction and Concentration

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

Comparison of LC/MS/MS and GC/MSD Analyses of Pesticide Residues in Food Products when Using the QuEChERS Sample Preparation Technique

HyperShear TM HPLC Mixers

Agilent Multimode Inlet for Gas Chromatography

n multi-residue analysis of pesticides olfgang Schwack, Claudia Oellig niversity of Hohenheim

mono-layer ACR High purity Reinforced 1-10 Excellent Fair mono-layer DD High purity Reinforced Excellent Excellent

4001 Transesterification of castor oil to ricinoleic acid methyl ester

Determination of fuel system icing inhibitor content of aviation turbine kerosine by HPLC

Application Note. Author. Abstract. Energy & Chemicals - Petrochemicals. Edgar Naegele, Agilent Technologies, Inc. Waldbronn, Germany

Simultaneous Determination of Fatty Acid Methyl Esters Contents in the Biodiesel by HPLC-DAD Method

Technical Overview. Introduction

Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve

Analysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES

S 1125 HPLC Pump System

SUPPRESSION IN ANALYSIS OF ARTIFICIAL SWEETENERS WITH

Optimized Method for Analysis of Commercial and Prepared Biodiesel using UltraPerformance Convergence Chromatography (UPC 2 )

HyperShear TM HPLC and UHPLC Mixers

The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector

HyperShear TM HPLC Mixers

Determination of PAH Compounds from Aqueous Samples Using a Non-Halogenated Extraction Solvent and Atlantic C18 Disks

VERITY Purification Systems Build the perfect system to fit your research

Pipetted Liquid Sampling. Automated sample preparation platform for online LC-MS/MS bioanalysis

Supplementary Material

HyperShear TM HPLC and UHPLC Mixers

Accuracy Starts Here. Agilent Captiva Filtration Selection Guide

GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method

Beverage Grade Carbon Dioxide

Application Note. Author. Introduction. Energy and Fuels

PRODUCT MANUAL. CARBONATE REMOVAL DEVICE 200 CRD 200 (4 mm) (P/N ) CRD 200 (2 mm) (P/N )

Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE

The Analysis of Biodiesel for Inorganic Contaminants, including Sulfur, by ICP-OES

Large Volume Injection of Polycyclic Aromatic Hydrocarbons

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585

Mineral Turpentine Adulterant in Lubricating Oil

A Single Method for the Direct Determination of Total Glycerols in All Biodiesels Using Liquid Chromatography and Charged Aerosol Detection

Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE

Dionex CRD Carbonate Removal Device

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

ACQUITY UPLC I-Class System with 2D Technology

Agilent 7693A Automated Liquid Sampler

5 µm Preparative HPLC Columns

Phase Distribution of Ethanol, and Water in Ethyl Esters at K and K

Agilent 7693A Automated Liquid Sampler

Valving Schemes for the BioLogic Duo-Flow Chromatography System

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS

Optimized Method Development of Large Volume Injection for GC/MS/MS of Food Pesticides

Analysis of Petroleum Fractions by ASTM D2887

EGC II K 2 CO 3. Cartridge Electrolytic ph Modifier EGC Carbonate Mixer

SiliCycle MiniBlock XT

Analysis of Petroleum Samples Using the Teledyne Leeman Labs

Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels

Gilson, Inc. Table of Contents. Solid Phase Extraction Systems HPLC: Systems 2 9. Liquid Handlers HPLC: Liquid Handlers/Injectors 10 14

Agilent G5632 Valve Kit Instructions

TSHR 6000 series TN/TS/TX analyzer

Application Note. Abstract. Authors. Environmental Analysis

Application Note. Authors. Abstract. Energy & Chemicals

Universal Vacuum Manifold Assembly UCT. Patent No.: US 7,438,862 B2 INNOVATION THROUGH CHEMISTRY

A Rapid imethodtm Test for Analysis of Acid Herbicides in Water

C2, C3, C4 Monomer Analysis

2014 Buick Encore. Fig. 2: Replacing Measured Compressor Oil Courtesy of GENERAL MOTORS COMPANY

UCT VACUUM EXTRACTION MANIFOLD

Fixed Flow Splitter *

High-throughput protein aggregate analysis of monoclonal antibodies using a novel dual-channel UHPLC instrument

AppNote 1/2010 KEYWORDS ABSTRACT. Biodiesel, Automation, ASTM D

The Agilent 1200 Series high perfor- mance autosampler SL: Area precision, injection volume linearity, minimum accessible volume, carry-over

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.

Technical Procedure for Gas Chromatography (GC-FID)

Complete Fractionation of Extractable Petroleum Hydrocarbons Using Newly Developed EPH SPE Cartridges

DIONEX ELUENT GENERATOR CARTRIDGE DIONEX EGC 500 K 2 CO 3 HIGH PRESSURE ELECTROLYTIC PH MODIFIER (DIONEX EPM 500) Quick Start

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

InertSustainSwift. Achieving separations faster than ever before

Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584

> Boosting the performance of LC/MS/MS analysis of mycotoxins and their transformation products by use of innovative separation techniques

Determination of Free and Total Glycerin in Pure Biodiesel (B100) by GC in Compliance with EN 14105

Gilson, Inc. Table of Contents. HPLC: Liquid Handlers/Injectors GX-271 Analytical Liquid Handler...43 ISO 9001 certified.

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

NEW HPLC - SYSTEM LINE

Validation Report 11

Liquefied Gas Injector. Solution for the Sampling and Analysis of Liquefied Gases

Analyte Focusing in Elevated Temperature HPLC

Prodigy ICP Application Note: # 1039

Reliable. Efficient. Economical. Distillation Technology ENGINEERING - EQUIPMENT - TURNKEY SYSTEMS

HIGH PRESSURE SWITCHING VALVES

Agilent Multimode Inlet

Characterization of Tiki Torch Fuels

HPLC SYSTEM S 500 SERIES

HPLC System S 600 Series

SPE Manifolds and Accessories

Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D

The Vanquish Platform: Improved Throughput and Resolution of Xanthones in Mangosteen Pericarp

Transcription:

Totally Automated Method for the Determination of Sudan Dyes in Food via On-Line Filtration, SPE and HPLC Analysis Joan M Stevens, Ph.D., Mark Crawford and Greg Robinson Single Presenter Name

Abstract: Sudan dyes are categorized as Class 3 carcinogens by the International Agency for Research on Cancer (IARC), and, therefore, are illegal as food additives according to both FDA and the EU. The European commission requires products to have documentation confirming the absence of Sudan dyes. Since 2003, European nations have required random product testing and testing of suspected adulterated products. Items found to contain Sudan dyes must be disposed of as hazardous waste. The system presented offers a totally automated approach to the extraction of Sudan dyes from food products. The product is solubilized, filtered and SPE is performed prior to HPLC analysis. Detection limits set by the EU of 0.5-1 mg/kg are obtained from this method without manual intervention or human exposure. Data presented shows the efficacy of this approach relative to preexisting methodology.

Sudan Dye Structures:

Outline: Background Information on Sudan dyes Sample Preparation System and Procedure Data and Results Summary and Conclusion

Background Information on Sudan Dyes: Sudan dyes are red dyes that are used for coloring solvents, oils, waxes, petrol, shoe and floor polishes They have also been found in some chili powder imported from India, in a number of relishes, chutneys and seasonings containing this chili powder Sudan dyes are not allowed to be added to food in the USA, UK and the rest of the EU and Japan They are azo dyes, which when eaten, may be split into carcinogenic amines

Sample Preparation: 4 gms of chili powder was mixed with 20 ml of acetone and vigorously shaken on an orbital shaker, 600 RPM The chili solution was filtered off-line with a 0.45 um PTFE filter 1 ml of the chili solution was diluted to 5 ml with aqueous NaOH (ph 11) The diluted chili solution was spiked with 100 ppb standard solutions of Sudan I, II, III, IV in acetone

The diluted chili solutions were filtered on the SPE-215 liquid handler via a syringe barrel type filter, 0.45 um The collected filtered solutions were subjected to automated SPE on the SPE-215 SPE eluent was evaporated to dryness and reconstituted in 200 ul acetonitrile/water 90:10 The SPE eluent was analyzed via analytical HPLC at 504 nm

SPE Methodology: 215-SPE, 4 probe automated SPE and liquid handler Oasis MCX, 60 mgs, 3 ml, SPE cartridges Condition Column: 2 ml ethyl acetate 2 ml methanol 1 ml 0.1 NaOH 2 ml water - Load 5 ml of diluted sample, in two 2.5 ml additions to the column - Wash column: 2 ml 70% methanol in water 1 ml 1.0 M NaOH 2 ml methanol, 1 ml ethyl acetate - Elute sample: 2 ml 89:9:2 ethyl acetate: methanol: formic acid - Dry down eluent, reconstitute with 200 ul 90:10 ACN, water

215-SPE: Automated SPE Instrument The SPE-215 uses positive pressure to push the solvents and sample through the SPE cartridges, can accommodate 1, 3, 96 well formats COLLECT WASTE

HPLC Methodology: Solvent A: Water, 0.1% TFA Solvent B: ACN, 0.1% TFA Flow: 1.5 ml/min Gradient conditions: 75% (B) 0-1 min 75 95% (B) 1 10 min 95% (B) 10-12 min 95-75% (B) 12 12.5 min 75% (B) 12.5 13.5 min Injection volume: 25 ul, Column: Thermo BetaBasic C-18, 4.6 x 150 mm, 3 um UV/Vis Detector 504 nm, sensitivity 0.1 AUFS

Data and Results: Graph 1: Chromatogram of McCormick s chili powders, all three types (chili powder, hot Mexican, and chipotle) gave the same results, no detection of Sudan I or its analogues.

I II III IV Graph 2: Chromatogram representing the results from a spiked chili powder solution with Sudan I, II, III, IV, 7-11 ug on column

I II III IV Graph 3: Chromatogram of a mixture of Sudan I and its analogues in acetone/naoh dilution mix, not chili powder extract

Recovery: Sudan I 95% (4% RSD) Sudan II 85% (2% RSD) Sudan III 93% (3% RSD) Sudan IV 75% (3% RSD) Recovery (n = 10, 7 11 ug on column) Limit of detection have been determined to be in the 0.1 0.5 ppm range without LC/MS

Conclusion: A versatile method was developed, which would determine Sudan I, Sudan II, Sudan III and Sudan IV in food stuffs, within the same chromatogram A chromatographic analysis time of 10 minutes, and total SPE time of less than 5 minutes per sample was achieved, thereby increasing sample throughput The recovery and LOD proved that LC/MS was not required to test food stuffs for Sudan I and its analogues We could continue to use our chili powders from the US, safe in the knowledge that there had been no trace of Sudan I or its analogues, found in off the shelf chili powders...phew!