NOTIS PEMBELIAN PERALATAN /09 (Website : http:// www.medic.usm.my/sebutharga.php) PUSAT PENGAJIAN SAINS PERUBATAN Tarikh Buka : 1 Jun 009 Tarikh Tutup : 15 Jun 009 Bil Nama Alatan Kuantiti 1. A. ITEM DEFIBRILLATOR SET FOR RESUSCITATION The equipment is a life saving device, used to treat patient with cardiac arrest and also can serve as temporary external cardiac pacing. It is portable and can be used during patient transfer. The equipment should come with the following features:- 1. Capabilities for non-invasive pacing.. Upgradeable for multi-parameter monitoring for SpO, blood pressure and CO 3. Fast charging time, 3 seconds or less to 00J, 4. Fast charging time, 5 seconds or less to 70J 5. At least 3 seconds ECG recovery after defibrillation 6. Have protection against sudden capacitor failure 7. Have easy routine checking (user test screen) 8. Child electrode underneath the adult electrode for external hard paddles 9. Colour-coded indicators on paddles for electrode-skin contact quality 10. Paddle change should be made easy using one connector 11. Built-in AED with voice prompt 1. 3-channel built-in thermal recorder 13. Upgradable for telemetry networking 14. Upgradeable for saved data to be transferred to a PC and reviewed with software 15. Able to actively control the shape of the second phase waveform to maintain constant pulse width. 1
16. Waveforms Display - upgradeable in future to 3 waveforms displayed simultaneously for SpO, and CO D. TECHNICAL SPECIFICATION 1. The unit should have biphasic technology.. Built-in external pacing capabilities for recovery from bradycardia after defibrillation. 3. Capability for hands free defibrillation 4. LCD Screen a) At least 5.7" crisp TFT colour b) Should be inclined for easy viewing 5. Operation Panels - should be inclined for easy viewing 6. Paddles and Paddle Holder a) The paddle holder should have inclined design for easy placing and removing of paddles b) Easy paddle change using one connector 7. Review and Reporting At least 8 seconds before and 1 seconds after defibrillation, ECG is stored automatically 8. Back-up battery operation for at least hours E. ESSENTIAL COMPONENTS & ACCESSORIES 1. Power cord : 1 unit each. Recording paper, Z-fold, 50 mm x 30 m : 10 units each 3. ECG connection cable, IEC, 3 electrodes : 1 unit each 4. Disposable electrode, adult, 60 pieces / set : 1 set each 5. Contact gel : 1 unit each 9. Rechargeable battery : 1 unit each 6. Operator's manual : 1 unit each 7. Service manual : 1 unit each 8. CD manual electric defibrillator : 1 unit each 9. Customized epoxy coated trolley, brakeable : 1 unit each F. STANDARD REQUIREMENT & INSTRUCTIONS 1. Power requirement : 40 V, 50 Hz, safety class II. The supplier has to give information regarding their servicing capabilities for the equipment quoted. 3. The equipment should be fully installed and made fully functioning. 4. Equipment supplied should be warranted against manufacturing defects for at least 1 months from date of commissioning.
5. Free preventive maintenance service 1 x 6 monthly during the warranty period. 6. Training should be provided for operation maintenance, troubleshooting and principle study visit 7. The equipments must be a brand new with proper serial number to prove it and must not be a reconditioned instrument. 8. Service and operating manual and original catalogue in English should be provided. 9. Price quoted must include government tax.. A. Item PATIENT TRANSPORT TROLLEY Stretcher to transport patient in the hospital. Simple design of stretcher provides safe and efficient transport for an economical value. 1. Metal frame stretcher complete with bottom tray made of thermo foam plastic.. The stretcher should be supplied with 3 thick polyester mattress 3. Manually operated. 4. Back and headboard can be lowered. 5. Holder for oxygen cylinder. 6. Completely covered fold down safety bedside rails with spring loaded I.V hanger rod with 4 receptacles 7. Four-wheel, steel ring brakes with approximately 0 cm castors provided at dual end. 8. Fifth wheel added for easier traction. 9. Built-in transfer board. 10. Fiberesin pneumatic backrest / stationary foot end litter. 11. Dual pedestal hydraulics. 1. Dual side-mounted foot control hydraulics. 13. Hands-free trendenlenburg. 14. Roller bumpers for protection when pushing the stretcher. 15. Defibrillator tray / footboard / extender. D. TECHNICAL SPECIFICATIONS 1. Height adjustment : low approximately from 500 mm to 900 mm or more in height with back raising system. 3
. Overall length and width of stretcher a) Length : approximately,000 mm b) Width : approximately 800 mm 3. Patient surface approximately : 660 mm x 1,900 mm 4. Side rails approximately : 330 mm x 1,400 mm 5. The stretcher should come with pop-up push handles 6. Weight capacity more than 00 kg. E. ESSENTIALS COMPONENTS & ACCESSORIES Each trolley supplied should come complete with the followings: 1. I.V Pole : unit each. Defibrillator tray and footboard extender : 1 unit each 3. Operator / service manual : 1 set each F. STANDARD REQUIREMENTS & INSTRUCTIONS 1. The equipment should be supplied fully installed and made fully functioning.. All equipments must be brand new with proper serial number to prove it and must not be used or reconditioned instrument. 3. Service and operating manual and original catalogue should be provided. 4. The equipment should be warranted against manufacturing defects for 1 year from the date of commissioning. 5. Training should be provided for operation maintenance, troubleshooting and principle study visit 6. Free preventive maintenance service 1 x 6 monthly during warranty period 7. The supplier has to give information regarding their servicing capabilities for the equipment quoted. 8. Price quoted must include government tax. 3. A. ITEM RESUSCITATION TROLLEY (pediatric)) 1 The equipment is used to occupies less room but maximizes space available. It also provides more than 40 personalised solutions to all hospital needs and is the most efficient solution to your medicine distribution and logistic needs also allows the cart to be easily moved and turned around with a fingertip. The equipment should come with the following features - 4
1. The material allows washing with water at a temperature up to 100*. The materials also self- extinguishing accounting to class UL VO. 3. The Trolley is 100% recyclable once disposed. 4. It has no metallic parts which may interfere or damage medical devices 5. The trolley is well rounded, no angled part. 6. Color of trolley is non fade and the surface is washable. 7. The trolley is composed of a single piece body in order to allow immediate disassembly for washing, disinfection and can be configured with other accessories. 8. Handle for pushing / pulling is moulded into the body to allow mobility, ease of use and durability 9. Has a perimeter bumper protection preventing damage to the trolley. 10. The trolley top has a 4-sided moulded curved rim to avoid any spilling of material and to prevent material from falling out of the trolley. 11. It a specific drawer for dangerous medication the interior of this drawer permits visibility for emergency needs. 1. The drawers and the side pocket has a central locking device. 13. Each drawer has an individual handle and an individual tag for labeling items kept in the drawer. 14. Optional removable dividers. 15. Drawers are accessible without hindrance at all times with the cardiac board at the rear. 16. There is specific pocket side space for catheters and endoscopes. 17. The trolley has 1 independent IV rod, there is a medication basin is made of stainless steel fully autoclave able. 18. There is a waste container with a top lid, built in to the structure. 19. There is an ampoule vial opener that is detachable and cleaned easily 0. There is mounded utility hooks at the rear. 1. High quality wheels : wheels with lock, 1 with directional lock and 1 anti static D. TECHNICAL SPECIFICATION 1. Material : Baydur Polyurethane. Dimension : 90 cm x 83 cm x 65cm 3. Baydur Polyurethane materials allow deep disinfection and, it is non-reactive to chemicals such as diluted acids, alkali and solvents. 4. The cardiac massage board is placed at the rear without perimeter bumper interference 5. The oxygen tank holder is built in. 6. The defibrillator holder 1 kg weigh.it is swivel movement and rotate by 360 degrees. E. ESSENTIAL COMPONENTS & ACCESSORIES 1. Basic body beige with seal lock. 100mm drawer 3. 150MM drawer 4. IV pole holder 5. Monitor /Defibrillator holder tray 6. Cardiac CPR message Board 7. Ampoule and Vial opener 8. Plastic swinging pockets 9. Catheters space 10. Used needles bucket 11. Holder for additional used needle 1. Bucket 5
13. Life saving drugs Drawer 14. Centralized Locking Mechanism 15. Water bottle holder 16. Front opening space 17. Waste bin with lid 18. Sliding writing / working surface 19. Stainless steel 18 18/10 medication 0. Basin (into drawer) 1. Label holder. Drawer for medication basin 3. Utility hooks 4. Oxygen tank holder 5. Shelf for suction pump 6. Macintosh Laryngoscope (pediatric) 1 set 7. Reusable resuscitator bag (pediatric ) 1 pc 8. Geudal Airway 1 set F. STANDARD REQUIREMENT & INSTRUCTIONS 1. The tenderer has to give information regarding their technical and servicing capabilities for the equipment quoted - list of service Engineers and technical application support staff must be provided.. The equipment must be brand new with proper serial number to prove it and must not be a reconditioned instrument. 3. The equipment should be installed and made fully functioning upon commissioning with operating and service manuals provided. 4. All installation charges inclusive of electrical and builders works must be come by supplier. 5. The equipment should be warranted against manufacturing defects for at least 1 months from date of commissioning. 6. Free preventive maintenance service 1 x 6 monthly during the warranty period 7. Maintenance and servicing agreement- optional service contracts after warranty period must be offered and quoted separately 8. Comprehensive training should be provided for operation, maintenance and troubleshooting. If training cost is to be incurred, this should be indicated. 4. A. ITEM Examination couches (resuscitation room) Examination couch use to examine the patient during clinical examination - Finishing : Epoxy Powder Coated - Cushion Top - Dark colored D. Technical Specifications - Tubular Steel Framework - 4 Legs Mounted On Rubber Studs 6
- Steel Sheet Constructed Cabinet And Drawers - At least 75mm Thick Vinyl PVC with Padded P.U. Foam - Manually Adjustable Backrest - 4 Reclining Positions - Dimension approximately 1850mm(L) x 600 mm(w) x 750mm(H) - 1 Retractable Step Stool E. Other Requirements and Instructions 1. The equipment should be installed and made fully functioning.. All equipments must be brand new with proper serial 3. Service and operating manual should be provided. 4. The equipment should be warranted against manufacturing defects for a period of at least 1 year from the date of commissioning. 5. The price quoted must be including government tax. 5. A. Item PORTABLE SUCTION PUMP A lightweight, compact portable suction unit for emergency use in hospital. 1 1. This portable suction unit is silent and vibration free.. No internal overflow or cross contamination D. TECHNICAL SPECIFICATIONS 1. Body material highly heat resistant, electrically insulated ABS plastic material. Non collapsible, high pressure at least 1 meter long autoclavable silicone tubing.. Jar material polycarbonate (Autoclavable) 3. Jar capacity at least liters 4. Vacuum pressure at least 500mm Hg 5. Suction capacity at least 0 liters/minute 6. Type of pump piston -Cylinder pump, Oil free 7. Weight less than 5 kg E. ESSENTIALS COMPONENTS & ACCESSORIES i. Complete suction unit. i. Disposable sac/liner unit. F. STANDARD REQUIREMENTS & INSTRUCTIONS 1. The equipment should be supplied fully installed and made fully functioning.. All equipments must be brand new with proper serial number to prove it and must not be used or reconditioned instrument. 3. Service and operating manual and original catalogue should be provided. 4. The equipment should be warranted against manufacturing defects for 1 year from the date of commissioning. 5. Training should be provided for operation maintenance, troubleshooting and principle study visit 7
6. Free preventive maintenance service 1 x 6 monthly during warranty period 7. The supplier has to give information regarding their servicing capabilities for the equipment quoted. 8. Price quoted must include government tax. 6. A. ITEM RESUSCITATION TROLLEY (adult) The equipment is used for medicine distribution and logistic needs to be easily moved. 1. The material allows washing with water at a temperature up to 100 o C.. The materials also self- extinguishing accounting to class UL VO. 3. The trolley is recyclable once disposed. 4. It has no metallic parts which may interfere or damage medical devices. 5. The trolley is well rounded, no angled part. 6. Color of trolley is non fade and the surface is washable. 7. The trolley is composed of a single piece body in order to allow immediate disassembly for washing, disinfection and can be configured with other accessories. 8. Handle for pushing / pulling is moulded into the body. 9. A perimeter bumper. 10. 4-sided moulded curved rim. 11. Specific drawer for dangerous medication. 1. The drawers and the side pocket has a central locking device. 13. Each drawer has an individual handle and tag. 14. Removable dividers. 15. Pocket side space for catheters and endoscopes. 16. One IV rod, stainless steel medication basin and autoclavable. 17. Built in waste container with a top lid. 18. Detachable an ampoule vial opener. 19. mounded utility hooks at the rear. 0. High quality wheels : wheels with lock, 1 with directional lock and anti static. D. TECHNICAL SPECIFICATION 1. Material : Polyurethane or equivalent.. Dimension approximately 90 cm x 83 cm x 65cm 3. Built in oxygen tank holder. 4. The defibrillator holder movable and rotatable by 360 degrees. F. STANDARD REQUIREMENT & INSTRUCTIONS 1. The tenderer has to give information regarding their technical and servicing capabilities for the equipment quoted - list of Service engineers and technical application support staff must be provided.. The equipment must be brand new with proper serial number to prove it and must not be a reconditioned instrument. 3. The equipment should be installed and made fully functioning upon commissioning 8
with operating and service manuals provided. 4. All installation charges inclusive of electrical and builders works must be come by supplier. 5. The equipment should be warranted against manufacturing defects for at least 1 months from date of commissioning. 6. Free preventive maintenance service 1 x 6 monthly during the warranty period 7. Maintenance and servicing agreement- optional service contracts after warranty period must be offered and quoted separately 8. Comprehensive training should be provided for operation, maintenance and troubleshooting. If training cost is to be incurred, this should be indicated. 1) Pembekal yang berminat diminta menghantar sebutharga bersama butir-butir berikut kepada : Ketua Penolong Pendaftar Bahagian Pentadbiran, Pusat Pengajian Sains Perubatan Universiti Sains Malaysia, 16150 Kubang Kerian, Kelantan ) Tempoh sah laku sebutharga : 90 Hari 3) Tempoh penyerahan : 4) Jenis/Buatan : 5) Salinan Sijil Pendaftaran dengan Kementerian Kewangan dan Akuan Pendaftaran Pembekal Bumiputera 6) Pihak Universiti berhak menerima kesemua atau sebahagian dari sebutharga-sebutharga tersebut dan tidak terikat untuk menerima sebutharga yang termurah atau mana-mana sebutharga pun Untuk sebarang pertanyaan sila hubungi : Pejabat Pentadbiran PPSP 09-7664007 / 09-766414 9