PolyStat Immersion Circulators

Similar documents
HTC-1200 Recirculating Heater. NESLAB Manual P/N Rev. 04/22/98. Operation and Maintenance Manual

Cryotrol Temperature Controller Instruction and Operation Manual

Operators Manual. Recirculating Chiller /06/08

Operators Manual. Microprocessor Controlled Water Baths /98

LM-100 LM-120. Light Meters. Users Manual. For detailed specifications and ordering info go to

Tracer VMA with AutoReg

Operators Manual. Model 3370 Air Cooled Recirculator rev.8/98


System IV Liquid to Liquid Heat Exchanger. Thermo NESLAB Manual P/N U00682 Rev. 01/18/01. Instruction and Operation Manual

All products subject to this warranty must be returned for examination, repair or replacement to:

IAQ-CALC INDOOR AIR QUALITY METER MODEL 7515

Instructions 3A4793B EN. See page 3 for a complete list of model descriptions and part numbers.

HX-540 Recirculating Chiller. NESLAB P/N Rev. 06/11/97. Instruction and Operation Manual

TX900. Microwave Leakage Detector. 99 Washington Street Melrose, MA Phone Toll Free

Tracer VM with User Interface

USE AND CARE GUIDE TIERED POTS FOUNTAIN

Broadband PowerShield. External Battery Pack 20 Ah 12 V Battery. User Manual

Hybex Microsample Incubator

Instruction Manual for Globe Price Computing Scale Models GS30 and GS30T

Instruction Manual for Globe Price Computing Scale Models GS30 and GS30T

TIME-10 Process Control Stopwatch

OPERATION MANUAL Variable Speed Pump Controller Dated: 06/04/2013 Pump Down Application. Document No.: LMSII_V100_OM Page 1 of 8 Model-V100 LMS II

MM-AVR UPS Series User s Manual

READ THIS MANUAL CAREFULLY BEFORE USING THE PUMP

RollSeal 1733 County Road 68 Bremen, Alabama Part No Rev Owner s Manual RS-Divider Curtain

Valor 1000W Series Instruction Manual. Serie Valor 1000W Manual de Instrucciones. Série Valor 1000W Guide de I utilisateur

User s Manual. ClipperCreek, Inc. Innovative Infrastructure for Electric and Hybrid Vehicles. Model PCS-15

SYSTEM MANUAL. Automated Motorized Hydraulic Traffic Controller. Revision CS72-HTC

StormPro BA Series Sump Pump

INSTRUCTION MANUAL DT173 DT175 DIGITAL THERMOMETERS ALWAYS READ THESE INSTRUCTIONS BEFORE PROCEEDING

Submersible Waterfall Pump 1200 Gallons Per Hour

Model 3400KL / 3401KL / 3400KG / 3400KG-C Digital Wet Diaper / Lap Sponge / Organ Scale

Model 700 Microarray Oven

LH41A. User Manual. Clamp On Ammeter. For detailed specifications and ordering info go to

Illumifalls Patio Pond

Premium Waterfall Pump 1200 Gallons Per Hour

TMA5 Mini Vane Anemometer

Premium Pond Pump 330 Gallons Per Hour

AIR VELOCITY METER ALNOR MODEL AVM410 AIRFLOW INSTRUMENTS MODEL TA410

User Guide. Digital Shipping Scale S150

PRM-4. Phase Sequence and Motor Rotation Tester. Users Manual. For detailed specifications and ordering info go to

18VDC ESB6 Series Cordless Screwdrivers Operation Manual

Patron Call System. Scope s DataPage II Lite & NexCall Coaster. Operators/Installation Manual

OWNER S MANUAL SELF-PRIMING PORTABLE UTILITY PUMP

MICROMANOMETER ALNOR MODEL AXD610 AIRFLOW INSTRUMENTS MODEL PVM610

Type Maxi Mix III Rotary Shaker

SUNTURA SOLAR TRACKER

RS-3 PRO RS-1007 PRO. CAT IV Analog Clamp meter Series. Users Manual. For detailed specifications and ordering info go to

1200+ WITH LVD (LOW VOLTAGE DISCONNECT) USER GUIDE

ECLIPSE Laundry Dispenser Controller

MINISTICK TM UV Germicidal Ultraviolet Light

MaxLite LED Medium Area Light

Thermometer models / 00831A

TIC 300 PRO. Users Manual Mode d emploi Bedienungshandbuch Manual d uso Manual de uso. Non Contact AC Voltage Detector. High Energy Tic Tracer

Instruction Manual PC104/PC105 3-PHASE SOCKET TESTER. Check out what else you can get from Martindale:

1.4 N/1.4 D. Operating Instructions. 580 Mayer Street, Building #7, Bridgeville, PA phone fax

MaxLite StaxMax LED Flood Lights

18VDC ESB6-X Series Cordless Screwdrivers Operation Manual

EF SERIES MOTORS OPERATION & PARTS MANUAL

FlexChem Model 404 Rotating Oven

Natural Fiber Planter Fountain

Boston Gear Ratiotrol DC Motor Speed Control

StormPro BCV400 Sewage Ejector Pump

IAQ-CALC TM Indoor Air Quality Meter

FCB-450, LCB-600, MCB-800

Cordless Make Up Mirror WITH ILLUMINATED 1X AND 10X MAGNIFICATION PLUS NIGHT LIGHT

SUNTURA HD SOLAR TRACKER

Submersible Pond Pump with UV 800 Gallons Per Hour

Mechanical Filter Pump Kit Up to 500 gallons

Operator s Manual TC-253A. Temperature Converter

INSTRUCTION MANUAL CM95/ CM100 TRUE RMS FLEXIBLE CLAMP METER. 1. SAFETY INFORMATION: Always read before proceeding. REMEMBER: SAFETY IS NO ACCIDENT

PY-11 Voltage & Continuity Tester

Thermometer model 00826

1200+ WITH LVD (LOW VOLTAGE DISCONNECT) V1.00 OWNERS MANUAL

Low-Water Auto Shutoff Pump 420 Gallons Per Hour

USE AND CARE GUIDE TEAL GLAZED POT FOUNTAIN

MODEL TC400 TIRE CART. Installation, Operation & Repair Parts Information REV P/N:

WiFi Tank Level Monitor Installation Instructions

USER MANUAL. NEMA 48C Flange Mounted, Enclosed, Clutch Brake. FMCBE Model 500 FORM NO. L A (i)

MV Series Motors Operation & Parts Manual

S25500 MAXI STIRRER OPERATION MANUAL AND PARTS LIST SERIES 1352

In-Line Electronic Control Handle

Automatic Butter Dispenser Instruction Manual Model #2395NS

Web Volume Control Model WV220

Podiatry Procedures Chair. For Models: Barrier-Free. User s Guide

Operating Instructions for PAC800 Battery Charger

Mechanical Filter Box Kit

Illumifalls Kit REMINDER CALL BEFORE RETURNING TO STORE.

Low-Water Auto Shutoff Pump with Light 300 Gallons Per Hour

MRC 150/300 USER MANUAL. 1 Laird Technologies

PORT-A-WEIGH CRANE SCALES

EQUIPMENT Operation Manual

SERIES 1000 ELECTRIC ACTUATORS

HRS305. Hand Rail Scale

Easy Weigh ADVANCED COUNTING SCALE MODEL: AC-100 OWNER S MANUAL VER 1.00

LIGHTSTICK / LIGHTSTICK PLUS Germicidal Ultraviolet Light

IAQ-CALC TM Indoor Air Quality Meter

Owner s Manual. Features

POWER. Switch opening. Enter/Exit Menu Menu Scroll Increase/Decrease Setpoint GCE SERIES MODEL 10AFE MODEL 20AFE

Container Fountain Kit with LED Light

Transcription:

PolyStat Immersion Circulators Manual P/N U00988 Rev. 06/09/08 Instruction and Operation Manual

PolyStat Immersion Circulator Table of Contents Preface Safety Compliance... 2 Unpacking... 2 Warranty... 2 Warnings... 3 General Information Description... 4 Specifications... 4 Installation Site... 5 Assembly... 5 Filling Requirements... 5 Fluids... 5 Polystat Analogue... 6 Polystat Digital... 7 Safety Features... 8 External Circulation... 8 Polystat Digital Unique Operation Controller... 9 Maintenance & Troubleshooting Cleaning... 12 Algae... 12 Checklist Analogue/Digital... 12 Checklist Digital... 12-1 -

Preface Compliance Products tested and found to be in compliance with the requirements defined in the EMC standards defined by 89/336/EEC as well as Low Voltage Directive (LVD) 73/23/EEC can be identified by the CE Mark on the rear of the unit. The testing has demonstrated compliance with the following directives: LVD, 73/23/EEC Complies with IEC/EN61010-1 EMC, 89/336/EEC IEC/EN61326-1 For any additional information, refer to the Declaration of Conformity that shipped with the unit. Unpacking Warranty Retain all cartons and packing material until the unit is operated and found to be in good condition. If the unit shows external or internal damage contact the transportation company and file a damage claim. Under ICC regulations, this is your responsibility. The Cole-Parmer circulators distributed by Cole-Parmer, Cole-Parmer warrants to the direct purchaser that the product will be free from defects in material or workmanship for a period of two years from the date of delivery. Cole-Parmer will repair or replace the product or provide credit, as its sole option, upon prompt notification and compliance with its instructions. The Distributor warrants to Customer that upon prompt notification and compliance with Distributor's instructions, that the Distributor will repair or replace, at Distributor's sole option, any Product which is defective in material or workmanship. Distributor expressly disclaims all other warranties, whether expressed, implied or statutory, including the warranties of merchantability, and fitness for a particular purpose. Distributor's sole responsibility and the Customer's exclusive remedy for any claim arising out of the purchase of any Product is repair or replacement, as described above. In no event shall Distributor's liability exceed the purchase price pain therefor; nor shall Distributor be liable for any claims, losses or damage of any third party or for lost profits or any special, indirect, incidental, consequential, or exemplary damages, howsoever arising, even if Distributor has been advised of the possibility of such damages. - 2 -

Warnings Make sure you read and understand all instructions and safety precautions listed in this manual before installing or operating your unit. If you have any questions concerning the operation of your unit or the information in this manual, contact our Sales Department. Performance of installation, operation, or maintenance procedures other than those described in this manual may result in a hazardous situation and may void the manufacturer s warranty. Observe all warning labels. Allow unit surfaces to cool before touching. Never remove warning labels. Always turn off the unit before connecting or disconnecting the power cord or other cables. Always turn off the unit and disconnect the line cord from the power source before performing any service or maintenance procedures, or before moving the unit. Never operate equipment with damaged line cords. Refer service and repairs to a qualified technician. In addition to the safety warnings listed above, warnings are posted throughout the manual. These warnings are designated by an exclamation mark inside an equilateral triangle with text highlighted in bold. Read and follow these important instructions. Failure to observe these instructions can result in permanent damage to the unit, significant property damage, or personal injury or death. - 3 -

General Information Description The PolyStat Immersion Circulators are used to provide precise temperature control. The unit consists of a circulating pump, heater, temperature sensor, controller and a mounting device. The units come supplied with a directional flow nozzle and a mounting clamp. Specifications Model Polystat Analogue Polystat Digital Temperature Range 25 C to 100 C 25 C to 200 C Controller Analog Digital Temperature Stability ±0.04 C ±0.02 C Heater 115V 1000 watts 1200 watts 230V 1500 watts 2000 watts Pump Capacity 14.0 lpm @ 0 feet Weight Pounds 8.1 Kilograms 3.7 Environmental Conditions Operating Temperature 5 C to 40 C Maximum humidity 80% RH - 4 -

Operation Site Assembly Filling Requirements Never place the unit in an area of high humidity, or corrosive or dusty atmosphere. See specifications on page 3. Attach the supplied mounting bracket to the unit. Remove the set of screws and spacers on the rear of the unit then attach the bracket using these screws. Secure the unit to the tank using the bracket. The maximum wall thickness is 26 mm. Fill the tank of the bath to be used with the unit to ensure the pump assembly is submerged. Note the fluid minimum and maximum level marks on the unit (75 and 85mm measured from the bottom of the circulator). Never run the unit in an empty tank. Fluids Never use flammable or corrosive fluids with this unit. Distilled or filtered tap water is the recommended fluid for operation. Normal tap water leads to calcium deposits on the heating element which will then require periodic cleaning. Calcium deposits reduces heating capacity, and shortens the life of the heater. Above +80 C, water vaporization reaches a level requiring the constant fluid replacement. Never use pure ethylene glycol as a fluid. A minimum 80/20 mixture of Ethylene Glycol and filtered tap water is allowed. - 5 -

Polystat Analogue ON LED Temperature fine adjust Heating LED Temperature setting dial Mains switch Fault lamp Thermometer clamp mounting screw Pump outlet (screw supplied angled nozzle into pump outlet) NOTE: High Temperature Cutout is located on rear of unit. Pump outlet (ensure hexagon plug installed) Start Up Set the High Temperature Cutout on the rear of the unit, see Safety Features on next page. Place the I/O switch to the I position. The pump will start. The ON LED will illuminate and the HEAT LED will illuminate as needed. Temperature Adjustment Turn the dial to the desired temperature. Use a small screwdriver in the unit's fine adjust potentiometer for exact tuning. Use a thermometer (not supplied) to check the temperature. - 6 -

Polystat Digital Heating control display Value arrows Reset Menu display High Temperature Cutout Temperature/ Setpoint C display Enter Menu button Mains switch Pump outlet (screw supplied angled nozzle into pump outlet) Pump outlet (ensure hexagon plug installed) Start Up Set the High Temperature Cutout on the front of the unit, see Safety Features on next page. Place the I/O switch to the I position. The pump will start. The C display will run a self-test. After the test the C display will indicate the temperature of the fluid near the unit's control sensor. The menu display will illuminate. The heating control display will illuminate as needed. Temperature Adjustment To display the temperature setpoint keep pressing the menu button until S is shown in the menu display. The C display will indicate the setpoint. If desired, change the setpoint using the value arrows until the desired setpoint is displayed. Press the enter button to accept the new setpoint value or the unit will default to the old setpoint. - 7 -

Safety Features All units are equipped with automatic thermally-triggered fuses located on the rear of the unit. Wait at least five seconds before resetting the fuses. Press the fuses to reset them. If the fuse trips again the unit has a defect and should be returned for servicing. All units have a motor alarm High Temperature Cutout (HTC). If the HTC temperature limit is exceeded the unit will shut down and, on Analogue units, the FAULT light will illuminate. On Digital units, an alarm will sound and a fault message will be displayed. Ensure the HTC is set lower than the heat transfer fluid's flash point. Use the dial to adjust the approximate desired limit. To set an exact limit, run the bath to the desired limit and then slowly turn the dial counterclockwise until the unit cuts off. Without moving the dial, on Analogue units press the outer ring around the dial to reset the HTC. On Digital units, press the RESET button on the front of the unit. NOTE: The fluid may have to cool before the unit will reset. In the event that the above fault condition occurs, remove power to the unit and corrected the fault condition prior to restarting. External Circulation The pump outlet is located on the front-bottom of the unit. Screw the angled nozzle into the outlet. Ensure the hexagon plug is installed in the other outlet. Secure your hose to the angled nozzle. The nozzle will accept 12mm I.D. hose. NOTE: The required hose material depends on the operating temperature and the fluid used. - 8 -

Polystat Digital Unique Operation Heating control display Value arrows Temperature/ Setpoint C display Enter Reset Menu button Menu display High Temperature Cutout Mains switch Controller Displaying the software version Press the menu button until FU is shown in the menu display. Press the value arrows until con is displayed. Press the enter button. Press the menu button until bs is shown in the menu display. The software version is temporarily displayed in the C display. Adjusting setpoint temperature Press the menu key until the letter S (for setpoint value) is shown on the menu display. Increase or decrease the value shown on the display using the value arrows. If you keep one of the value arrows depressed, the temperature changes slowly in tenths of a degree. Within a few seconds the temperature change rate will accelerate. Once the desired value is displayed press the enter button. The new value is not saved and permanently stored unless the enter button is pressed. If not, the circulator continues to use the old setpoint value. After a short time the C display automatically indicates the fluid temperature. Heating control indicator This small circular display lights up when heating is switched on (temperature setpoint is higher than the current temperature). The display lights up constantly during the heating up phase. The display flashes during the control phase. - 9 -

The display does not light up if heating is not activated (set temperature setpoint is lower than the current temperature). Correction factor C The C display shows the actual temperature at the unit's control sensor. This temperature does not correspond directly to the temperature in the circulator s bath and even less to the temperature of any external connected system. The temperature difference is determined by measuring the actual current temperature using a suitable measuring device (calibrated or gauged thermometer). The difference is a correction factor (c) which can be entered and stored into the unit. The resolution of the correction factor is 0.01 C, with a range of ±9.9 C. Press the menu button until FU is shown in the menu display. Press the value arrows until rta is displayed. Press the enter button. Press the menu button until cs is shown in the menu display. Increase or decrease the value shown on the C display using the value arrows. Once the desired value is displayed press the enter button. The new value is not saved and permanently stored unless the enter button is pressed. If not, the circulator continues to use the old correction factor. After a short time the C display automatically indicates the fluid temperature. The correction factor may have to be changed if the temperature setpoint is changed. Example: Setpoint value programmed at the circulator = 70.7 C Actual measured temperature in bath/system = 70.5 C Delta T = 0.2 C Enter the correction factor = +0.20 C The temperature displayed at the circulator will now match the external system (70.5 C). Temperature limitation values If required by the application or the flash point of the selected heat transfer liquid, the circulator's temperature range can be limited. Setting a limit helps avoid faults when operating the unit. The highest temperature limit possible is 200 C, the lowest limit is -50 C. This is not a safety feature but an aid to avoid faults. The High Temperature Cutout must also be set. Press the menu button until FU is shown in the menu display. Press the value arrows until H-L is displayed. Press the enter button. Press the menu button - 10 -

until either LL (low limit) or HL (high limit) is shown in the menu display. Increase or decrease the value shown on the C display using the value arrows. Once the desired value is displayed press the enter button. The new value is not saved and permanently stored unless the enter button is pressed. If not, the circulator continues to use the old value. After a short time the C display automatically indicates the fluid temperature. Secure mode The secure mode determines how the unit will react after a power failure. With the mode on, once power is restored the unit will display 8888 and the reset button has to be depressed to start the unit. With the mode off, once power is restored the unit will automatically start. Press the menu button until FU is shown in the menu display. Press the value arrows until con is displayed. Press the enter button. Press the menu button until 88 is shown in the menu display. Use the value arrows to select on or OFF. Once the desired mode is displayed press the enter button. The new mode is not saved and permanently stored unless the enter button is pressed. If not, the circulator continues to use the old mode. After a short time the C display automatically indicates the fluid temperature. Adjusting LED display contrast Press the menu button until FU is shown in the menu display. Press the value arrows until con is displayed. Press the enter button. Press the menu button until In is shown in the menu display. Use the value arrows to change the display contrast. Once the desired contrast is displayed press the enter button. The new contrast is not saved and permanently stored unless the enter button is pressed. If not, the circulator continues to use the old contrast. After a short time the C display automatically indicates the fluid temperature. Adjusting LED display resolution Press the menu button until FU is shown in the menu display. Press the value arrows until con is displayed. Press the enter button. Press the menu button until ta is shown in the menu display. Use the value arrows to change the resolution, 0.1 C or 0.01 C. Once the desired resolution is displayed press the enter button. The new resolution is not saved and permanently stored unless the enter button is pressed. If not, the circulator continues to use the old resolution. After a short time the C display automatically indicates the fluid temperature. - 11 -

Maintenance & Troubleshooting Cleaning Periodically clean the unit using a mild soapy solution and a soft, nonabrasive cloth. Do not use steel wool. It is abrasive and will lead to rusting. Algae We recommend the use of Chloramine-T, 1 gram per 3.5 liters. Checklist (Analogue/Digital) Unit will not start Check for fault conditions. Unit may need a reset. Check the fuses. Check power source for correct voltage output. Check line cord plug wiring, see electrical requirements. No external circulation Check for obstructions in external system lines. Circulation will cease when pump head is exceeded. Review specifications. Controller ceases operation Ensure the fluid viscosity is 1.6 centipoise or less. Checklist (Digital) Power failure If the unit shuts down because of a power failure and the display reads 8888 once power is restored, press the reset key. Fault Displays If a fault occurs an alarm will sound and the C display will indicate the fault condition. The menu display will indicate AL. The heating element and pump will automatically shut down. The following faults are possible: -- C Excess temperature The fluid temperature is higher than the temperature set on the HTC. If this is not the case return the unit for servicing. - 12 -

-- -n Low liquid level There is not enough liquid in the bath. Fluid may have evaporated but also check for leaks. Top off if necessary. --P Pump or motor overloading The motor or pump is blocked. It can take 10 minutes or longer until the motor temperature cools enough so that the unit can be switched on again by pressing the reset button. If the circulator switches off again return the unit for servicing. -- -F Sensor breakage or short circuit The sensor is damaged. Please return the unit for repairs. -- -r Undefined fault This can be caused by a momentary fault, i.e. a fluctuating bath level when the filling level is very close to minimum. Top off the unit. In all other cases this unit must be checked by qualified service personnel. Fault eliminated? After the fault has been eliminated, the cause of the fault is shown on the C display (e.g. C). The preceding three zeros mean that the fault has been eliminated. The reset button must be pressed in order to restart the unit. - 13 -