USER MANUAL. White LED Light Meter. Model LT40. Additional User Manual Translations available at

Similar documents
Solar Power Light Meter

USER GUIDE. Model RH10 Hygro Thermometer. Humidity and Temperature Meter

USER MANUAL. 200A AC Mini Clamp Meter. Model MA130. Please visit for user manual translations

Digital Formaldehyde Meter

Non-Contact Phase Rotation Tester

LM-100 LM-120. Light Meters. Users Manual. For detailed specifications and ordering info go to

USER GUIDE Moisture Meter

User Guide. Model Insulation Tester / Megohmmeter

User's Guide. EzFlex Thermo-Anemometer. Model EZ30 UNITS MAX MIN AVG VEL

Hot Wire CFM/CMM Thermo-Anemometer

Model Insulation Tester / Megohmmeter. User Manual

Non-Contact Phase Rotation Tester

User Guide. Model Insulation Tester / Megohmmeter

USER MANUAL. Insulation Tester + DMM. Model MG320

User's Manual. Earth Ground Resistance Tester. Model Introduction

User's Guide. CO 2 Meter. Model CO250

User Guide. Model Insulation Tester / Megohmmeter. Introduction

SL101 Manual_SL101 User Manual 05/10/ :06 Page 1. Instruction Manual

USER MANUAL. 10KV Digital High Voltage Insulation Tester. Model MG500. Additional User Manual Translations available at

User's Guide. Model RPM10 Laser Photo / Contact Tachometer with IR Thermometer. Introduction

Safety design conforming to the following provisions

Mini Multimeter with Non-Contact Voltage Detector (NCV)

Users Manual. Accessories Motor and Phase Rotation Indicator ships with the following items: 3 test leads 3 test probes 3 alligator clips 9 V battery

Phase Sequence Tester

MANUAL DE INSTRUÇÕES DO ANEMÔMETRO DIGITAL MODELO AN-3040

DOUBLE WAYS THERMOCOUPLE METER. Users Manual

MM300. INSTRUCTION MANUAL Manual-Ranging Digital Multimeter

Item ref: UK & UK MTB01 & MTB02 DIGITAL MULTITESTERS. User Manual

DAFM4. Digital Air Flow Meter INSTRUCTION MANUAL Fax: (503) ENGLISH DAFM4 VOL CFM

DT304. Digital Temperature Logger INSTRUCTION MANUAL

INSTRUCTION MANUAL DT173 DT175 DIGITAL THERMOMETERS ALWAYS READ THESE INSTRUCTIONS BEFORE PROCEEDING

Instruction Manual. IM-191 Magnetic Field Meter

Visit us at

Mini Multimeter with Non-Contact Voltage Detector (NCV)

Visit us at

Calibrator Capabilities Function Range Resolution Measure V dc 28 V 1 mv Measure ma dc Source ma dc

PHscan30S Pocket ph Meter Instruction Manual

INSTRUCTION MANUAL MARTINDALE CM52 CLAMP METER. ELECTRIC Trusted by professionals. Other products from Martindale:

OPERATING INSTRUCTIONS

User's Manual. High Voltage Megohmmeter. Analog Insulation Tester plus AC Voltage and Continuity Tests. Model

Quick Guide. brite Advanced Retail scale

OPERATING INSTRUCTIONS

Safety design conforming to the following provisions

ORPscan20 Pocket ORP Tester Instruction Manual

User Manual. CFM/CMM Vane Thermoanemometer. Model with NIST-Traceable Calibration THE STANDARD IN PRECISION MEASUREMENT

THERMOCOUPLE CALIBRATOR Mode KM-CAL-710

Digital Multimeter User's Manual

Model Battery Powered Milliohm Meter

717 Series. Pressure Calibrators. Instruction Sheet. Introduction

GENERAL INFORMATION. H-1649, H-1650, H-1651 H-1653, H-1654 Easy-Count. uline.com. that may hurt accuracy:

Autoranging Industrial Multimeter

4-WIRE EARTH RESISTANCE TESTER C1 ES P1 S. Max. AC 200V MICROPROCESSOR CONTROLLED TEST. Switch ON, TEST, SCROLL RESULTS 20Ω 200Ω 2000Ω OFF

INSTRUCTION MANUAL CM95/ CM100 TRUE RMS FLEXIBLE CLAMP METER. 1. SAFETY INFORMATION: Always read before proceeding. REMEMBER: SAFETY IS NO ACCIDENT

SSP-Series Portable Balances Operation Manual

PRM-2 PHASE SEQUENCE TESTER PRM-3 PHASE SEQUENCE & MOTOR ROTATION TESTER User s Manual

SERVICE MANUAL. STARTER Portable Meter ST300 ST300C ST300D ST400D. Ohaus Corporation, 7 Campus Drive, Suite 310, Parsippany, NJ (973)

362 Clamp Meter. Users Manual

l The Battery Tester is designed for measuring the l AC four-terminal method to measure the internal

Table of Contents. Title Page UT501/UT502 OPERATING MANUAL. Introduction. Unpacking the Meter. Safety Information. International Electrical Symbols

POCKET-TACH 100 MONARCH INSTRUMENT Rev 1.4. Instruction Manual. Printed in the U.S.A. Monarch Instrument 2001 all rights reserved

Automatic Pressure Calibrator. 3-Year Limited Warranty. See the Users Manual for the full warranty.

Volt/mA Calibrator. Manual

R&S RT-ZC02 R&S RT-ZC03 AC/DC Current Probe User Manual

INSTRUCTION MANUAL YOKOHAMA QUICK CHARGER WITH LCD DISPLAY. Safety

RECHARGEABLE POWER OUTAGE LED LIGHT INSTRUCTION MANUAL

Users Manual. Airflow Meter

Elcometer 311. Elcomotor Refinishing Gauge. Types F and FNF. Operating Instructions

DIGITAL RCD(ELCB) TESTER

Adam Equipment QT SERIES. ADAM EQUIPMENT CO. LTD. pn Rev. D1, July 2002

ASTROAI USER MANUAL DIGITAL MULTIMETER. NOTE: Fully read and understand this manual before using this instrument.

MAX. AC 600V HOLD V M H.V. LIVE CIRCUIT DIGITAL INSULATION TESTER DO NOT OPERATE IF LIT POWER DATA HOLD TEST ON / OFF 250V 500V 1000V OFF CAT.

True RMS Autoranging Multimeter

Weighing Scale IDS-802. User Manual

323/324/325 Clamp Meter

Autoranging Multimeter

Elcometer 415. Coating Thickness Gauge. Operating Instructions

User's Guide. Vane Anemometer. Model Measures Air Velocity & Temperature. Degrees C/F switchable

Flue Gas Analyzer IMR 1100F

12 VOLT 30 AMP DIGITAL SOLAR CHARGE CONTROLLER

DMG200. Digital Micron Gauge. 99 Washington Street Melrose, MA Phone Toll Free

Digital/Analog Megohmmeter Model 1026

OPERATING MANUAL CMP-200

INSTRUCTION MANUAL CM69 TRMS AC EARTH LEAKAGE CLAMP METER. 1. SAFETY INFORMATION: Always read before proceeding. REMEMBER: SAFETY IS NO ACCIDENT

Megohmmeter Test Probe

8,1,8,1,2,7,2,7,6,3,6,3,4,5,4,5

ECscan40 Pocket Conductivity/TDS/Salinity Tester Instruction Manual

Vol-Con XL (61-086) Voltage/Continuity Tester and Vol-Test (61-085) Voltage Tester Instruction Manual WARNING

ECscan30 Pocket Conductivity/TDS Tester Instruction Manual

LH41A. User Manual. Clamp On Ammeter. For detailed specifications and ordering info go to

ECscan20 Pocket Conductivity Tester Instruction Manual

Short manual testo 610

Adam Equipment DUNE DCT SERIES. (P.N. 9384, Revision B2, June 2013)

40 V LITHIUM-ION BATTERY ATTACH YOUR RECEIPT HERE AB13786C 1. kobalttools.com ITEM # /

Model UT601: OPERATING MANUAL

PHscan10S Pocket ph Tester Instruction Manual

TX900. Microwave Leakage Detector. 99 Washington Street Melrose, MA Phone Toll Free

230VAC Power Inverter 400W Owner s Manual

51 & 52 Series II. Users Manual. Thermometer. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

DCell Suction. Model DM Operating Instructions & Maintenance Manual. Clearing The Airway Is Our #1 Priority

Digital/Analog Megohmmeter Model 1026 USER MANUAL

Transcription:

USER MANUAL White LED Light Meter Model LT40 Additional User Manual Translations available at www.extech.com

Introduction Congratulations on your purchase of the Extech LT40 LED Light Meter that measures light intensity from white LED light sources. The LT40 can also measure light from fluorescent, metal halide, highpressure sodium, and incandescent sources, for reference only (unspecified accuracy). The LT40 can measure light up to 40,000 Fc (400,000 Lux). This instrument is shipped fully tested and calibrated and, with proper use, will provide years of reliable service. Please visit our website (www.extech.com) to check for the latest version of this User Guide, Product Updates, and Customer Support. Features Overload Indication: The LCD will show OL. Low battery Indication. Display Update Rate: 2.5 times per second. Cosine Angle corrected. Measures white LED light. Measures the intensity of illumination in Lux or foot candles. Data hold freezes displayed reading. Maximum/Average/Minimum reading Hold. Zero Calibration adjustment. Auto power off with disable function. Auto range. Safety Do not operate the meter in environments where the following are present: explosive gases (or materials), combustible gases (or materials), steam, or dust. Please replace the battery immediately when the low battery symbol appears on the LCD. Do not touch the meter s circuit board for any reason; static electricity or contamination could damage the sensitive components. For Indoor use only. This instrument was designed for pollution degree 2. Operation Altitude: Up to 2000m (7000 ). 2

Meter Description 1. Photo detector (remove protective cover to expose sensor) 2. Display (LCD) 3. Control push buttons 4. Power Button: ON/OFF Battery compartment and tripod mount are located on rear of instrument Push Button Description 1. Lux/Fc units selector 2. Zero Calibration button 3. Maximum/Average/Minimum memory button 4. Data hold button Operation Power ON/OFF Long press the Power button to power the meter. To power the meter OFF, short press the Power button. Taking Measurements 1. Switch the meter ON and remove the sensor s protective cover to expose the light sensor dome. The display should switch ON; if not, check that batteries are installed and fresh. 2. The meter measures the intensity of the light (illuminance) that strikes the sensor dome in foot candles and Lux units (1 fc = 10.76 lux), displaying this measured value on its LCD. 3. Position the meter and light source so that the light strikes the sensor dome straight on (perpendicular) with as little an angle as possible. 4. The meter s display can show a value up to 9999. However, for readings that represent measurements higher than this, the meter uses x10 or x100 feature. For example, to represent a measurement of 11,000 fc the meter will display 1100 with the x10 indicator. 3

Auto Power OFF To save battery life, the meter powers down automatically after approximately 12 minutes of inactivity. Enable/Disable Auto Power Off While the meter is ON, long press the Power button to disable the Auto Power OFF utility (the Clock symbol will switch OFF). To enable the Auto Power OFF utility, repeat this process. LUX/FC Button Press the LUX/FC button to toggle between Lux and FC (foot candles) measurement units. MAX/MIN Button The meter can record the maximum, minimum, and average readings as described below: 1. Short press the MAX/MIN button and the meter will begin to track the maximum/average/minimum measurements; the MAX icon will display at the top of the LCD window indicating that the meter is now showing the maximum reading. The reading will not change until a higher reading is registered. 2. Press the MAX/MIN button again to switch from MAX to AVG, where the meter will show the average measurement value. The AVG icon will be displayed above the displayed value. 3. Press the MAX/MIN button again to change the mode from AVG to MIN, where the meter will show the minimum value measured. The MIN icon will be displayed. 4. Press the MAX/MIN button again to switch from MIN back to MAX. 5. To exit this mode, hold the MAX/MIN button for at least 2 seconds. The MAX/AVG/MIN icons should all be switched OFF when the unit returns to the normal operating mode. HOLD Button Press the HOLD button to freeze the current reading on the LCD. Press the HOLD button again to release the reading. In the Data Hold mode, HOLD is displayed on the LCD. Zero Calibration Procedure 1. Ensure that the protective cap is attached to the light sensor. 2. Power the meter and the LCD should display 0. 3. Short press the CAL ZERO button to perform the zero adjustment (CAL will appear on the display). 4. The CAL display will switch off when the calibration has been completed. 5. If the protective cap is not covering the sensor when the Zero calibration is started the LCD display will read CAP. In this case, please cover the sensor with the cap and restart this procedure. 4

Measurement Considerations and User Tips For maximum accuracy allow the light being measured to fall directly on the sensor as perpendicular as possible with a minimal angle of incidence. When the meter is not in use, please keep the protective cap in place over the light sensor. This will prolong the life of the sensor. When the meter is to be stored for long periods, please remove the batteries and store them separately. Batteries can leak and cause damage to the meter s components. Avoid areas of high temperature and humidity when using this instrument. Battery Replacement and Maintenance Cleaning and storage 1. The white plastic sensor dome should be cleaned with a damp cloth when necessary. Use only a mild soap if needed. Do not use solvents, abrasives, or harsh detergents to clean the dome. 2. Store the meter in an area with moderate temperature and relative humidity. Battery Replacement When the battery power decreases to a critical level, the low battery symbol will appear on the LCD. Replace the 2 x 1.5V AAA batteries located in the rear battery compartment. 1. Press and hold the battery compartment lock button (see diagram). 2. Slide the battery compartment cover off in the direction of the arrow. Install two (2) 1.5V AAA batteries observing correct polarity and close the battery compartment. Ensure that the compartment cover is securely fastened before using the meter. 1 2 Never dispose of used batteries or rechargeable batteries in household waste. As consumers, users are legally required to take used batteries to appropriate collection sites, the retail store where the batteries were purchased, or wherever batteries are sold. Disposal: Do not dispose of this instrument in household waste. The user is obligated to take end of life devices to a designated collection point for the disposal of electrical and electronic equipment. 5

Specifications Units Range Resolution Accuracy Lux Foot candles 999.9 0.1 9,999 1 99,990* 10 400,000* 100 99.99 0.01 999.9 0.1 9,999 1 40,000* 10 White LED: ±(3%reading + 3 Lux) up to 500 Lux ±(3%reading) >500 Lux Other light sources: Unspecified White LED: ±(3%reading + 0.3 Fc) up to 46 Fc ±(3%reading) >46 Fc Other light sources: Unspecified * Readings above 9999 use a x10 or x100 multiplier (1 Fc = 10.76 Lux) Angle deviation from cosine characteristics 30 ±2% 60 ±6% 80 ±25% General Specifications Sampling rate Photo detector Display 2.5 times per second (digital display) Silicon photo diode with cosine correction 4 digit LCD (maximum display: 9999) with low battery icon, measurement overload, and other function indicators Auto Range Meter automatically ranges the display Operating conditions Temperature: 5 to 40 o C (41 to 104 o F); Humidity: < 80% RH Storage conditions Temperature: 10 to 60 o C (14 to 140 o F); Humidity: < 70% RH Low battery indication appears on the LCD when battery voltage falls critically low LED Type White LED light Power supply 2 x 1.5V AAA batteries Battery life Approximately 200 hours Auto Power OFF Meter switches OFF after 12 minutes of inactivity Dimensions 133 x 48 x 23 mm (5.2 x 1.9 x 0.9 ) Weight 250 g. (8.8 oz.) including batteries 6

Appendices Typical Light Levels (1 Fc = 10.76 Lux) Lux Foot Candles Lux Foot Candles Factories Home 20 75 2 7 Emergency Stairs, Warehouse 100 150 10 15 Washing 75 150 7 15 Exit/Entrance Passages 150 200 15 20 Recreational Activities 150 300 15 30 Packing Work 200 300 20 30 Drawing Room, Table 300 750 30 75 Visual Work: Production Line 300 500 30 50 Makeup 750 1,500 75 150 Typesetting: Inspection Work 500 1,500 50 150 Reading, Study 1,500 3,000 150 300 Electronic Assembly, Drafting 1,000 2,000 100 200 Sewing Office Restaurant 75 100 7 10 Indoor Emergency Stairs 75 150 7 15 Corridor Stairs 100 200 10 20 Corridor Stairs 150 300 15 30 Entrance, Wash Room 200 750 20 75 Conference, Reception Room 300 750 30 75 Cooking Room, Dining Table 750 1,500 75 150 Clerical Work 750 1,500 75 150 Show Window 1,500 2,000 150 2000 Typing, Drafting Store Hospital 75 150 7 15 Indoors 30 75 3 7 Emergency Stairs 150 200 15 20 Corridor/Stairs 75 100 7 10 Stairs 200 300 20 30 Reception 100 150 10 15 Sick Room, Warehouse 300 500 30 50 Display Stand 150 200 15 20 Waiting Room 500 750 50 75 Elevator 200 750 20 75 Medical Exam Room 750 1,500 75 150 Show Window, 750 1,500 75 150 Operating Room Packing Table 1,500 3,000 150 300 Storefront, Show Window 5,000 10,000 500 1000 Eye Inspection Copyright 2013 2017 FLIR Systems, Inc. All rights reserved including the right of reproduction in whole or in part in any form ISO 9001 Certified www.extech.com 7