CASCADE CHAIR-MOUNTED HANDPIECE CONTROL WITH CASCADE MASTER SERIES TOUCH PAD

Similar documents
INSTALLATION INSTRUCTIONS CASCADE 3181 DUO DELIVERY SYSTEM TO PREFERENCECOLLECTION 5580

Radius Cuspidor (for Cascade 1040 or Decade 1011/1021 Chairs)

A-dec 541 Delivery System and 545 Assistant s Instrumentation

Asepsis 21 Delivery Unit

Cascade 3072 Delivery System INSTALLATION GUIDE

A-dec O Clock Delivery System

axcs Chair Mounted Unit Installation, Operation and Care Manual

Instructions for Use. A-dec 4631 Duo Delivery System and A-dec 4635 Assistant s

DECADE 1021 SOLO CHAIR

Simplicity Chair Mounted Unit

Installation Instructions: Swing Mounted Units BDS-2575

Assistant s Instrumentation T

axcs Magellan CMU Installation, Operation and Care Manual

Owner s Guide PERFORMER III

A-dec 500 Delivery Systems

Simplicity Magellan Unit

Owner's Guide CASCADE 2671 DUO CART AND CASCADE 3171 DUO WALL-MOUNT

Cascade Master Series

Instructions for Use. Cascade 2615 Cart

A-dec 351 Assistant s Instrumentation on an A-dec 311/411/511 Dental Chair INSTALLATION GUIDE

A-dec 542 Side Delivery System I NSTALLATION GUIDE

Mounted Instrument Head

A-dec 2122/2132 Radius System on A-dec 511 Dental Chair I NSTALLATION GUIDE

A-dec 300 Delivery Systems

Installation Instructions: Over-The-Patient Units (& Cuspidors) BDS-2500 BDS-2505 BDS-2510 BDS-2515 BDS-2520

A-dec 371L or 571L Dental Light on an A-dec 361, 362, or 363 Support Center INSTALLATION GUIDE

A-dec 200 Service Guide Contents

INSTRUCTIONS FOR USE CASCADE 2615 CART AND 3175 WALL MOUNT

Service Guide A-DEC SIMULATORS. Mobile (4810) and Stationary (4820)

A-dec Dental Lights on an A-dec 300 Base Mount Post INSTALLATION GUIDE

Procenter Delivery Mounted on a Free Standing Treatment Console (FTC) Installation

A-dec 300 LED Dental Light

Performer Delivery System (Performer I Chair - International, and Performer III Chair)

A-dec 570L Dental Light on a Non-DCS System INSTALLATION GUIDE

Loose Components. VetPro 5000 Wall / Cabinet Mount Installation. Applies to Models:

Procenter Delivery, Console Mounted on an Elevance Chair Installation Guide

Simplicity Hygiene Unit

Installation Instructions: Side Delivery Units BDS-2555

Evogue Dental Unit IMPORTANT

Flex Work Station Delivery Units Installation Guide

A-dec 336 Delivery System on an A-dec 411 or 511 Dental Chair INSTALLATION GUIDE

Installation Instructions for Swing Mount System

INTRODUCTION TO SERVICING THE A-dec Handpiece Control Blocks

Chair Mounted Unit. Installation, Operation and Care Manual. DentalEZ Group 2500 Highway 31 South Bay Minette, AL 36507

DECADE 1021 VAC BACK CHAIR

Installation Operation Maintenance Troubleshooting Version 1, Oct/09. Palm Beach

A-dec Service Guide Edition , 2003

Tubing / Wiring Diagrams: Elevance Delivery Systems (Console / LR Mtd: Elevance Chair)

forest dental australia affordable excellence + designer friendly

A-dec 311, 411, and 511 Dental Chairs Service Reference

Second Stage Regulator - 1/4 Turn

Instructions for Use. Performer. Dental Chair, Delivery System, Assistant s Instrumentation, Support Center with Cuspidor, and Light

Mirage Dental Delivery Unit MSP3500/3550 Installation Instructions

1416 Cabinet Unit. Installation Operation Maintenance Troubleshooting Version Dec/17

JODALE PERRY. Parts List & Mounting Instructions. Jacobsen HR9016 JDP BUILT FOR LIFE

Spirit 2000 Series Installation Instructions

Troubleshooting Guide: Elevance Delivery Systems

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS

U L T I M A T E R A D A R / L A S E R D E F E N S E S Y S T E M

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS

SECTION 4 - FUEL SYSTEMS AND CARBURETION

RANGER 900 POWER STEERING KIT

w w w. h d o n l i n e s h o p. d e CRUISE CONTROL KIT GENERAL INSTALLATION -J04064 REV Kit Number Models Additional Parts Required

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS

C FORD F250 / F L POWERSTROKE DIESEL WITH AUTOMATIC TRANSMISSIONS ONLY

OPERATION AND MAINTENANCE INSTRUCTIONS

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS

PILLOWS DECK-MOUNT BATH FAUCET WITH FLUME SPOUT

In addition to this instruction sheet, the following are included in this kit (some additional parts are included for future maintenance):

INSTALLATION INSTRUCTIONS

DODGE RAM 24V 5.9L CUMMINS

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS TRAILER HITCH MAIN HARNESS KIT

Wall / Casework Mounted Delivery Systems (Procenter)

Installation Instructions

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS

Hydraulics. Part B, Section 1. This section covers the following unit configurations. 3700V 3800V 3900V

DISCONTINUED. Installation. Aerada 900 Series Futura Faucet. With Accu-Zone (AZ) Infrared Control

MODEL NUMBER: MEDIUM DUTY ONBOARD AIR SYSTEM

PRO RATCHET UNIVERSAL SHIFTER

Installation Instructions Unimatic Shifter

I-BFV.KIT Butterfly Valve Gear Operator Replacement

INSTALLATION INSTRUCTIONS

and 1/4 TURN SECOND STAGE REGULATOR TAL 502 (L) Rev. 0 MSA 2005 Prnt. Spec (I) Mat Doc

STEERING COLUMN - TILT

Service Manual Air Tech Second Stage

Track Light with Monitor Installation Guide

INSTALLATION INSTRUCTIONS

INSTALLATION INSTRUCTIONS

STEALTH BIG AIR KIT - Yamaha Roadliner/Stratoliner and Raider

BLACKBIRD INSTALLATION SUPPLEMENT

Product Name Chair Conversion Kit (to LR Delivery Unit) Installation [Kit # xx]

Procenter Cuspidor. Common Components Console Mounted Cuspidor. for Ultra Chairs for Elevance Chairs. for European Assistant s Unit

Transcription:

R A-DEC INC. 2601 CRESTVIEW DRIVE POST OFFICE BOX 111 NEWBERG, OREGON 97132 USA TELEPHONE (503)-538-9471 DESIGNATED E.U. REPRESENTATIVE: A-DEC DENTAL U.K., LTD. AUSTIN HOUSE 11 LIBERTY WAY ATTLEBOROUGH FIELDS INDUSTRIAL ESTATE NUNEATON, WARWICKSHIRE ENGLAND CV116RZ PHONE: 0800-ADECUK (233285) WITHIN UK 00 44 24 7635 0901 OUTSIDE UK A-DEC AUSTRALIA 41-43 BOWDEN STREET ALEXANDRIA, N.S.W. 2015 AUSTRALIA TELEPHONE (61) 1.800.225010 CHECK WITH LOCAL CODE AUTHORITIES ABOUT INSTALLATION REQUIREMENTS FOR THIS PRODUCT INSTALLATION INSTRUCTIONS CASCADE CHAIR-MOUNTED HANDPIECE CONTROL WITH CASCADE MASTER SERIES TOUCH PAD Post Box, Assistant s Arm (optional), Handpiece Control, Tray Holder (optional), and Floor Box BEFORE BEGINNING THIS INSTALLATION 1. If the air and water manual shutoff valves have not already been installed, do so now. Refer to the floor box template, A-dec Publication No. 85.0472.00. Purge the air and water supplies of debris. 2. Install the dental chair following the instructions found with the chair. Do not install the upholstery at this time. YOU WILL NEED: Torque wrench, 0-100 ft-lb 15 16" wrench Magnetic bubble level Small, narrow, flat-tip screwdriver Phillips screwdriver Hex keys: 5 64", 3 32", 1 8", and 3 16" Wire nut crimping tool RAISE THE CHAIR TO FULL BASE UP 1. If the chair has a foot switch, raise the chair base by depressing the up arrow on the footswitch 2. If the chair does not have a foot switch but is equipped with a jumper cable (post box end of touch pad wiring harness with caution label), connect the two jumper cable terminals together to raise the chair base. When the chair is full up, remove and discard the jumper cable. 3. If the chair does not include a footswitch or a jumper cable assembly, remove the chair motor/pump and printed circuit board covers. Locate the Base Up (BS UP) label on the chair printed circuit board (see Figure 1). Short the two test points beside the Base Up (BS UP) label to raise the chair base. ELECTRICAL WARNING Hazardous AC voltages are present on the chair printed circuit board, do not touch any part of the printed circuit board except the test points. SW1 ENT BK UP BK DN BS UP DN EX TEST POINTS P1 P2 NOTE: CONNECTOR P1 OMITTED FOR CLARITY. SHORT THESE TWO TEST POINTS TO RAISE THE CHAIR BASE FIGURE 1 Page 1 85.0570.00

INSTALL THE CHAIR SWIVEL STOP NOTE The chair swivel stop must be installed on the chair to meet agency stability requirements. 1. Locate the chair swivel stop (see Figure 2). If the chair has an internal umbilical, the swivel stop is located inside the unit kit. If the chair has an external umbilical, the swivel stop is located inside the chair mount adapter carton. INSTALL THE CHAIR MOUNT ADAPTER Internal umbilical through the Cascade chair, go to INSTALL THE POST BOX on page 4. 1. Locate and remove the chair mount adapter kit from the chair mount adapter carton. 2. Cascade chair: Locate the adapter mounting hole near the center of the chair seat, under the tilt cylinder and springs. Position the adapter pivot stud (see Figure 3) in the mounting hole and install the spring washer (concave [hollowed] side down) onto the mounting stud. Tighten the castle nut. Back the castle nut off one-half turn and install the cotter pin. WARNING Do not allow the adapter to fall from its mounting position before installing the washer and castle nut the adapter is heavy, damage or injury may result. CHAIR SWIVEL CASTING SPRING WASHER CASTLE NUT UP HEX NUT LARGE DISC (1040 Chair only, side stamped UP faces up) OVAL OPENING (for wiring harnesses) SETSCREW CASCADE 1040 CHAIR SEAT CASTING SWIVEL STOP GROOVE ADAPTER PIVOT STUD OUTER STUD FIGURE 3 CASCADE 1040 CHAIR ADAPTER FIGURE 2 2. Install the chair swivel stop on the chair swivel casting (see Figure 2). Position the chair swivel stop with the groove to the front. 3. Tighten the setscrew to secure the stop in place (see Figure 2). 3. Position the adapter for either right or left hand delivery system mounting. Place the large disc (with the side stamped UP facing up, see Figure 3) on the outer stud. Finger tighten the hex nut. Make sure the adapter is positioned fully to the side. Using a torque wrench, firmly tighten the hex nut to 55 foot pounds. Go to INSTALL THE TOUCH PAD AND CUSPIDOR SAFETY SWITCH WIRING HARNESS on page 3. Page 2 85.0570.00

4. Decade chair: Locate the adapter mounting stud near the underside, center of the seat plate. Carefully slip the adapter pivot hole (non-threaded) up onto the mounting stud (see Figure 4). Install the spring washer (concave [hollowed] side up) and castle nut onto the mounting stud. Tighten the castle nut. Back the castle nut off one-half turn and install the cotter pin. MOUNTING STUD (not visible from this view, located underneath the seat plate) SPRING WASHER CASTLE NUT ADAPTER PIVOT HOLE (non-threaded) FIGURE 4 HEX HEAD BOLT (1011/1021 Chair only) DECADE 1011/ 1021 CHAIR SEAT PLATE (located underneath the chair upper structure) DECADE 1011/1021 CHAIR ADAPTER ELECTRICAL WARNING Hazardous AC voltages are present on the printed circuit board. Failure to unplug the chair could result in serious injury. 2. Unplug the chair. 3. Remove the chair motor pump cover and printed circuit board cover. 4. Cascade chair: Route the end of the harness with the ferrite filter (see Figure 5) into the oval opening on the top of the chair mount adapter (see Figure 3 on page 2), route the harness through the adapter, over the chair lift arm cover (see Figure 6), along the left-hand side of the lift cylinder, and into the pump housing. DECADE CHAIR HARNESS THROUGH SWIVEL CASTING P2 ROUTE CASCADE 1040 HARNESS OVER LIFT ARM COVER OPENING LIFT ARM HOUSING PUMP HOUSING 5. Position the adapter for either right or left hand mounting. Align the threaded adapter hole with the hole in the chair seat plate (see Figure 4). Install the hex head bolt through the seat plate and into the adapter. Using a torque wrench, firmly tighten the hex nut to 55 foot pounds. TOUCH PAD CABLE (end with ferrite filter) FIGURE 6 PRINTED CIRCUIT BOARD INSTALL THE TOUCH PAD AND CUSPIDOR SAFETY SWITCH WIRING HARNESS 1. Remove the wiring harness from the unit ship kit (see Figure 5). FERRITE FILTER 5. Decade chair: Route the end of the harness with the ferrite filter (see Figure 5) towards the chair swivel casting (along the top of the chair mount adapter [secure with cable ties]), through the swivel casting into the lift arm housing (see Figure 6), along the left side of the lift cylinder, and into the motor/pump housing. 6. Connect the harness assembly to connector P2 on the chair printed circuit board (see Figure 7). P2 P2 P1 K6 DUPLEX FIGURE 5 S1 P5 BS DN BS UP Bk UP Bk DN K1 MOTOR 9230 Potter & Brumfield T90S1D12-24 COIL: 24VDC 30A, 240VAC 10A, 240VAC 15A, 120VAC MADE IN U.S.A. 4 FIGURE 7 Page 3 85.0570.00

-1989, 91 A-DEC VS 2.2 7. If the unit includes a cuspidor, remove the blue jumper from connector P14 on the chair printed circuit board. Connect the 2-pin connector from the harness assembly to P14 (see Figure 8). P14 P5 P4 U1 P13 INSTALL THE HARNESS SUPPORT Decade 1011/1021 Chairs, go to INSTALL THE POST BOX. 1. Cascade chair with external umbilical. Locate the harness support in the post box ship kit and install it under the seat casting and the swivel stop casting (see Figure 10). P14 P7 P6 FIGURE 8 NOTE Removing the jumper from P14 on the chair printed circuit board and connecting the wiring harness will disable all chair functions, with the exception of base up, until the cuspidor safety switch is connected to the post box end of the wiring harness. To temporarily enable chair functions, install the blue jumper in the post box end of the wiring harness 2-pin cuspidor safety switch connector. CONNECT THE DUPLEX OUTLET (OPTIONAL) 1. Route the duplex outlet power wiring harness over the top to the lift arm cover and down the right side of the lift cylinder into the area of the chair printed circuit board (similar to the routing of the touch pad wiring harness, see Figure 6 on page 3). FIGURE 10 INSTALL THE POST BOX 1. Remove the hex head bolt, thrust washer, and spring washer from the kit. 2. Place the thrust washer on the chair adapter (see Figure 11). HEX BOLT SPRING WASHER 2. Connect the wiring harness to connector P12 on the chair printed circuit board (see Figure 9). F1 1/8 A SLOW BLOW POST BOX THRUST WASHER P12 CHAIR ADAPTER 230 Potter & Brumfield T90S1D12-24 COIL: 24VDC 30A, 240VAC 10A, 240VAC 15A, 120VAC MADE IN U.S.A. FIGURE 9 FIGURE 11 3. For units with an external umbilical, the post box and floor box are connected by the unit umbilical. Position the floor box near enough to the chair adapter to allow the post box to be set in place on the adapter. Page 4 85.0570.00

4. Remove the post box side covers, then carefully lift the post box out of the shipping carton and position it on the chair mount adapter (see Figure 11). The long axis of the post box should be parallel to the chair. 5. Install the spring washer (concave [hollowed] side down) and hex bolt through the post box and into the adapter (see Figure 11). 6. Snug the bolt, but do not tighten at this time. INSTALL THE ASSISTANT S ARM (OPTIONAL) If an assistant s arm is not being installed, go to INSTALL THE HANDPIECE DELIVERY SYSTEM TO THE POST BOX. 1. Unpack and install the assistant s arm on the post box (see Figure 12). INSTALL THE HANDPIECE DELIVERY SYSTEM TO THE POST BOX 1. Unpack the handpiece delivery system. 2. If the rigid arm and the flex arm have been disassembled for packaging, rotate the rigid arm until the short elbow of the rigid arm is aligned with the flex arm knuckle (see Figure 14). Push the rigid arm onto the flex arm knuckle and tighten the setscrews (see Figure 14). RIGID ARM FLEX ARM SETSCREWS FIGURE 14 FLEX ARM KNUCKLE NOTE The trim ring must be in place on the rigid arm before installing the delivery system on the post box (see Figure 15). 3. Using the safety strap found in the shipping carton, secure the flex arm to the rigid arm. FIGURE 12 2. Install the split ring into the groove located on the mounting hub (see Figure 13). Start one end of the ring in the groove and proceed around the mount, spiraling the ring into the groove. 4. Carefully lift the delivery system out of the carton. Guide the tubing and wiring through the post box hub (see Figure 15) and install the delivery system onto the post box. RIGID ARM TRIM RING FLEX ARM POST BOX HUB FIGURE 13 FIGURE 15 Page 5 85.0570.00

NOTE The arm safety stop pins must be installed on the chair to meet agency stability requirements. 5. To prevent the flex arm and control head arm from 360 rotation, install the safety stop pins from the ship kit in the rigid arm and control head arm (see Figures 16 and 17). The pins are installed with the smaller diameter end up. NOTE Continental-style control head. The two safety stop pins at the control head can be optionally installed on the two holes nearer each other (see Figure 17). With the stops in these positions, the control has a wider range of movement for trans-thorax positioning. INSTALL THE CASCADE MASTER 6300 DENTAL LIGHT (OPTIONAL) NOTE If a Cascade 6300 Master Dental Light is to be installed to the unit, install it now. Refer to the installation instruction found with the dental light. When installing the dental light to the post box, remove the Master Circuit Board assembly (see Figure 18) by pulling the entire assembly away from the post box frame. Pull the light cable through the space behind the circuit board and above the electrical box. Reinstall the Master Circuit Board assembly. FLEX ARM PHILLIPS SCREWS CAUTION! ELECTRIC SHOCK HAZARD DO NOT REMOVE COVER. REFER SERVICING TO QUALIFIED SERVICE PERSONNEL. WARNING! BEFORE SERVICING UNIT UNPLUG FROM SOURCE OF POWER. THREE SAFETY STOP PINS FIGURE 16 RIGID ARM MASTER CIRCUIT BOARD ASSEMBLY FLEX ARM POST BOX PANEL MASTER DENTAL LIGHT WIRING HARNESS MASTER DENTAL LIGHT AIR-ELECTRIC SWITCH FIGURE 18 CONTROL HEAD MOUNTING ARM (Continental shown) FIGURE 17 TWO SAFETY STOP PINS (see Note above) Remove the light air-electric switch from the light kit. Using the Phillips screws found in the light kit, install the light air-electric switch in the post box (see Figure 18). To ease installation, remove the post box panel.reinstall the panel when installation is complete. Page 6 85.0570.00

LEVEL THE POST BOX 1. Position the control head and, if installed, the dental light in their normal use positions. 2. Place a bubble level on the delivery system rigid arm (see Figure 19). Check and adjust for level. Bring the post into a vertical position by adjusting the leveling setscrews (see Figure 19). RIGID ARM FRICTION ADJUSTMENT 1. The rigid arm includes a friction adjustment to eliminate horizontal drift of the control head (see Figure 20). Turn the adjustment setscrew clockwise to increase friction on the flex arm knuckle. FLEX ARM CHECKING FOR LEVEL PARALLEL TO THE CHAIR RIGID ARM FRICTION ADJUSTMENT CHECKING FOR LEVEL PERPENDICULAR TO THE CHAIR RIGID ARM FIGURE 20 ADJUST TO LEVEL THE POST BOX CHANNEL THE UMBILICAL TUBING AND WIRING INTO THE POST BOX If the unit umbilical is external, go to POST BOX ELECTRICAL CONNECTIONS on page 8. 1. Route the wiring harnesses through the square opening and to the electrical box (see Figure 21). POST BOX MOUNTING BOLT FIGURE 19 ROUTE UMBILICAL TUBING HERE CAUTION! ELECTRIC SHOCK HAZARD DO NOT REMOVE COVER. REFER SERVICING TO QUALIFIED SERVICE PERSONNEL. CONNECTOR WIRING DIAGRAM WARNING! BEFORE SERVICING UNIT UNPLUG FROM SOURCE OF POWER. ELECTRICAL BOX COVER 3. Tighten the post box mounting bolt. Torque to 55 foot pounds using a torque wrench. ROUTE VACUUM AND GRAVITY DRAIN CONVOLUTE TUBING HERE OVAL BRACKET FIGURE 21 ROUTE WIRING HARNESSES HERE 2. Route the umbilical tubing through the square opening and to the air/water side of the post box (see Figure 21). 3. Route the vacuum and gravity drain convolute tubing through the oval hole in the bracket (see Figure 21). Page 7 85.0570.00

CONVOLUTE TUBING CONNECTIONS If the unit is not equipped with optional assistant s instrumentation or optional cuspidor, go to POST BOX ELECTRICAL CONNECTIONS. 1. Cut the convolute tubing to length. 2. Using a waterproof sealant to prevent leakage, push the convolute tubing onto the optional solids separator outlet and the optional cuspidor drain (see Figure 22). POST BOX ELECTRICAL CONNECTIONS 1. Remove the electrical box cover (see Figure 21 on page 7). 2. Connect the wires from the cable coming from the floor box to the terminal strip located inside the electrical box (see Figure 22). The label on the electrical box cover (see Figure 21 on page 7) shows the positions for each wire connection to the terminal strip. SOLIDS SEPARATOR OUTLET CAUTION! ELECTRIC SHOCK HAZARD DO NOT REMOVE COVER. REFER SERVICING TO QUALIFIED SERVICE PERSONNEL. WARNING! BEFORE SERVICING UNIT UNPLUG FROM SOURCE OF POWER. CONTROL HEAD WIRING HARNESS COMING FROM THE CONTROL HEAD CONVOLUTE TUBING TO CUSPIDOR DRAIN CONVOLUTE TUBING TO SOLIDS SEPARATOR OUTLET DENTAL LIGHT WIRING HARNESS COMING FROM THE FLOOR BOX CONTROL HEAD WIRING HARNESS COMING FROM THE FLOOR BOX CONTROL HEAD WIRING HARNESS COMING FROM THE FLOOR BOX TERMINAL STRIP FIGURE 22 Page 8 85.0570.00

CAUTION! ELECTRIC SHOCK HAZARD DO NOT REMOVE COVER. BEFORE SERVICING UNIT UNPLUG FROM SOURCE OF POWER. AIR-ELECTRIC SWITCH SINGLE-PIN CONNECTORS TO THE DENTAL LIGHT WIRING HARNESS MASTER POWER SUPPLY WIRING HARNESS (3-pin connector) INTO THE MASTER CIRCUIT BOARD REFER SERVICING TO QUALIFIED SERVICE PERSONNEL. WARNING! CAUTION! ELECTRIC SHOCK HAZARD DO NOT REMOVE COVER. REFER SERVICING TO QUALIFIED SERVICE PERSONNEL. WARNING! BEFORE SERVICING UNIT UNPLUG FROM SOURCE OF POWER. CAUTION! ELECTRIC SHOCK HAZARD DO NOT REMOVE COVER. REFER SERVICING TO QUALIFIED SERVICE PERSONNEL. WARNING! BEFORE SERVICING UNIT UNPLUG FROM SOURCE OF POWER. MASTER POWER SUPPLY WIRING HARNESS (3-PIN CONNECTOR) FROM THE FLOOR BOX TOUCH PAD WIRING HARNESS, CONTROL CONNECTOR (labeled CONTROL ) INTO THE MASTER CIRCUIT BOARD TOUCH PAD WIRING HARNESS, CHAIR CONNECTOR (labeled CHAIR ) CUSPIDOR SAFETY SWITCH (2-pin connector) DENTAL LIGHT WIRING HARNESS CONTROL HEAD WIRING HARNESS FIGURE 23 3. Connect all of the connectors (see Figure 23). 4. If the unit is equipped with a cuspidor, remove the blue tab jumper from the 2-pin cuspidor safety switch connector attached to the touch pad harness coming from the chair. Connect the 2-pin connector coming from the chair to the 2-pin connector coming from the cuspidor (see Figure 23). 6. Locate the master power supply wiring harness (with internal red and black wires) coming from the floor box. Push the 3-pin connector into its mating connector on the master circuit board (see Figure 23). 7. Place all of the connectors in the electrical box and install the cover (see Figure 21 on page 7). 5. Connect the master touch pad harness. Push the master touch pad wiring harness connector, labeled control, into the lower, side connector on the circuit board (see Figure 23). Page 9 85.0570.00

WATER MANIFOLD ROUTE TUBING THROUGH CUTOUTS IN THE POST BOX CASTING FERRULE NUT 3/8" WATER TUBING (blue) FROM FLOOR BOX 1/4" SUPPLY AIR TUBING (yellow) FROM FOOT CONTROL TUBING ASSEMBLY 1/8" REGULATED AIR TUBING (yellow) AIR MANIFOLD CONNECT THE FOLLOWING: 1/8" PILOT AIR TUBING (yellow with short red dashes to same) 1/8" ORAL CAVITY WATER TUBING (red from control to red from floor box or red from self-contained water supply) 1/4" DRIVE AIR TUBING (orange from control head to orange from foot control assembly) 1/8" RETURN TUBING (optional) (purple to purple) 1/8" CHIP BLOWER TUBING (optional) (brown to brown) 1/4" SUPPLY AIR TUBING (yellow) FROM FOOT CONTROL TUBING ASSEMBLY TO AIR MANIFOLD FERRULE NUT 1/4" DRIVE AIR TUBING (orange) FROM FOOT CONTROL TUBING ASSEMBLY TO DRIVE AIR TUBING (orange) FROM CONTROL HEAD 3/8" SUPPLY AIR TUBING (yellow) FROM FLOOR BOX FIGURE 24 INTERNAL FOOT CONTROL TUBING ASSEMBLY POST BOX AIR AND WATER PLUMBING If the unit has an external umbilical, go to step 5. Air (3/8" yellow) and Water (3/8" blue) Tubings 1. Slide a ferrule nut from the ship kit onto the 3/8" blue and 3/8" yellow tubings (see Figure 24). 2. Push the end of the blue tubing onto the connector at the bottom of the water manifold (see Figure 24). 3. Push the end of the yellow tubing onto the connector at the bottom of the air manifold (see Figure 24). 4. Slide the ferrule nuts onto the connectors and finger tighten. Then tighten 1-1/4 turns. Do not overtighten. Air (1/8" yellow) and Water (1/8" red) Tubings Coming from the Control Head 5. Connect the regulated air tubing (yellow) coming from the control head to the air manifold (see Figure 24). 6. Connect the pilot air tubing (yellow with short red dashes) from the control head to the pilot air tubing coming from the floor box (see Figure 24). 7. Connect the oral cavity water tubing (red) from the control head to the inline quick-disconnect (Q.D.) coming from the self-contained water supply or (if no self-contained water) directly to the red tubing from the floor box (see Figure 24). If the system has self-contained water, the red tubing from the floor box will not be connected. 8. If a tank-style water heater is installed in the floor box, connect the return tubing (purple) coming from the control head to the return tubing (purple) coming from the floor box (see Figure 24). This tubing is used for options and is not always connected. Page 10 85.0570.00

CONNECT THE MASTER AIR TUBING COMING FROM THE FLOOR BOX (yellow with red stripe) REGULATED AIR TUBING COMING FROM THE AIR MANIFOLD (yellow) MASTER ON/OFF AIR SIGNAL VALVE BOWL RINSE AIR SIGNAL VALVE INPUT SUPPLY 6300 DENTAL LIGHT ON/OFF AIR SIGNAL VALVE 0 1 0 1 0 1 0 1 0 1 0 1 CONNECT THE WATER SIGNAL TUBING COMING FROM THE FOOT CONTROL (green with short dashes) CONNECT THE AIR SIGNAL TUBING COMING FROM THE FOOT CONTROL (green with long dashes) CONNECT THE 40 PSI REGULATED AIR TUBING COMING FROM THE FLOOR BOX (yellow with green dashes) CUP FILL AIR SIGNAL VALVE AIR COOLANT AIR SIGNAL VALVE REGULATED 40 PSI TUBING COMING FROM THE CUSPIDOR (yellow with green dashes) REGULATED 40 PSI TUBING COMING FROM THE SELF- CONTAINED WATER SUPPLY (yellow with green dashes) CONNECT THE MASTER AIR TUBING COMING FROM THE CONTROL HEAD (yellow with red stripe) WATER SIGNAL TUBING COMING FROM THE CUSPIDOR (green with short dashes) OUTPUT SIGNAL FIGURE 25 WATER COOLANT AIR SIGNAL VALVE AIR SIGNAL VALVE MANIFOLD ASSEMBLY AIR SIGNAL TUBING COMING FROM THE CUSPIDOR (green with long dashes) CONNECT THE AIR SIGNAL TUBING COMING FROM THE CONTROL HEAD (green with long dashes) CONNECT THE WATER SIGNAL TUBING COMING FROM THE CONTROL HEAD (green with short dashes) CONNECT THE WHITE TUBING COMING FROM THE AIR ELECTRIC SWITCH IN THE POST BOX Internal Foot Control Tubings Coming from the Chair Mount Adapter If the unit has an external umbilical, go to step 13. 9. Connect the 1/4" yellow tubing coming from the post box end of the foot control tubing assembly to the air manifold (see Figure 24). 10. Connect the drive air tubing (orange) coming from the foot control to the 1/4" drive air tubing (orange) coming from the control head (see Figure 24). 11. Connect the tubing coming from the foot control to the input supply side of the manifold assembly (see Figure 25). Match tubing color and dash pattern to tubing color and dash pattern. Water signal (green with short dashes) tubing, Air signal (green with long dashes) tubing. Air Signal Valve Manifold Assembly Connections If the unit has an external umbilical, go to step 13. 13. Connect the tubing coming from the umbilical to the INPUT SUPPLY side of the air signal valve manifold assembly (see Figure 25). Match tubing color and dash pattern to tubing color and dash pattern. Master air (yellow with red stripe) tubing, Regulated 40 psi (yellow with green dashes) tubing. 14. Connect the tubing coming from the control head to the OUTPUT SIGNAL side of the manifold assembly (see Figure 25). Match tubing color and dash pattern to tubing color and dash pattern. Master air (yellow with red stripe) tubing, Water signal (green with short dashes) tubing, Air signal (green with long dashes) tubing. 12. Connect the brown with white dashes tubing coming from the foot control to the control head only if the foot control has the chip blower option. Page 11 85.0570.00

INSTALL THE FLOOR BOX 1. Remove the floor box cover. The cover on a stainless steel floor box simply lifts off the box. To remove the cover from a Cascade floor box, slide the cover forward, then lift it off the floor box base (see Figure 26). UMBILICAL CLIP CASCADE FLOOR BOX FIGURE 28 FIGURE 26 2. The internal umbilical extends 8" or 3' from the front of the chair. An external umbilical extends 5' from the post box. If the umbilical is too long, it can be shortened to an appropriate length. To shorten the umbilical: a.determine how long the umbilical needs to be and mark it where it will be cut. b.slit the excess umbilical convolute sheathing lengthwise, remove and discard (see Figure 27). SLIT THE EXCESS CONVOLUTE (Be careful not to cut the internal tubing/wiring) UMBILICAL UMBILICAL CLIP LARGE STAINLESS STEEL FLOOR BOX FIGURE 29 CONNECT THE DRAIN TUBING If the unit is not equipped with vacuum tubing and/or gravity drain tubing, go to step 4 on page 13. 1. Attach the gravity drain T connector from the hardware kit to the gravity drain stub using a suitable water resistant adhesive (see Figure 30). The drain air vent prevents air from being trapped in the drain system and inhibiting flow. The drain air vent must not be obstructed. DRAIN AIR VENT INTERNAL TUBING/WIRING FIGURE 27 REMOVE AND DISCARD EXCESS GRAVITY DRAIN c. Install the umbilical in the floor box and secure using the umbilical clip (see Figures 28 and 29). 3. Secure the floor box to the floor. Refer to the floor box installation instructions, A-dec Publication No. 85.0472.00. GRAVITY DRAIN STUB FIGURE 30 1" (25 mm) OR LESS Page 12 85.0570.00

2. Identify the convolute gravity drain tubing. Cut off any excess tubing so the path to the drain is as short and direct as possible, eliminating any low spots. Low spots allow sediment to collect or water to stand, inhibiting drain flow. Push the tubing onto the drain (see Figure 31). CUSPIDOR GRAVITY DRAIN CONVOLUTE TUBING GRAVITY DRAIN If the unit being installed is NOT equipped with a circulating syringe, go to CONNECT THE UTILITIES. 4. Bend and position the copper drip tube to empty into the drain air vent (see Figure 33). Be sure to position the dripper tube at least 1" (25.4 mm) above the vent, then attach it to the gravity drain T connector with the cable ties from the hardware kit. DRIPPER TUBE 1" (25.4 MM) MINIMUM DRAIN AIR VENT SEAL THE CONNECTION WITH A WATERPROOF ADHESIVE FIGURE 31 3. Identify the convolute vacuum tubing and cut it to the proper length, eliminating any low spots. Push it onto the vacuum stub (see Figure 32). CABLE TIES RETURN WATER (purple tubing) FIGURE 33 FIGURE 32 Page 13 85.0570.00

1 0 2 0 7 5 6 1 0 2 0 7 5 6 CONNECT THE UTILITIES If the unit being installed has an external umbilical, go to step 2. The filter/regulators are not installed on the umbilical. When installing, remove the short length of identification tubing from the fitting or barb, note its color and dash length, then connect the matching color and dash length tubing from the umbilical. 2. Purge the air and water supplies of debris. 3. Loosen the hex nut on the air and water manual shutoff valves (see Figure 36). AIR FILTER/REGULATOR ASSEMBLY FILTER ELEMENT HOUSING 1. Following the procedure above, connect the filter/regulators to the umbilical (see Figures 34 for standard plumbing or 35 for standard plumbing with self-contained water). WATER FILTER/ REGULATOR ASSEMBLY kg/cm 2 100 psi 10 90 20 80 30 70 50 40 60 3 4 AIR FILTER/REGULATOR ASSEMBLY 3 4 50 40 60 2 5 30 70 20 80 1 10 90 6 0 100 psi 0 kg/cm 2 7 3/8" AIR SUPPLY TUBING (yellow with black dashes) 1/8" REGULATED AIR TUBING (yellow) kg/cm 2 100 psi 10 90 20 80 30 70 50 40 60 3 4 WATER SCREEN (water only) AIR MANUAL SHUTOFF VALVE AIR WATER FILTER/REGULATOR ASSEMBLY WATER 1 2 3 4 50 40 60 30 70 20 80 10 90 0 100 psi 0 kg/cm 2 7 5 6 3/8" WATER TUBING (blue with white dashes) FIGURE 34 (STANDARD PLUMBING) 1/8" MASTER AIR TUBING (yellow with red stripe) 1/8" PILOT AIR TUBING (yellow with red dashes) 1/8" ORAL CAVITY TUBING (red) BLACK VALVE BODY FIGURE 36 WATER MANUAL SHUTOFF VALVE 4. Install the water screen on the water filter/regulator, identified by a black valve body (see Figure 36). 5. Connect the air and water filter/regulators to the manual shutoff valves (see Figure 36). DO NOT overtighten the fittings. The filter/regulators should be positioned at an angle to provide easy access to the filter element on each filter/ regulator manifold. 3 4 50 40 60 2 5 30 70 20 80 1 10 90 6 0 100 psi 0 kg/cm 2 7 3/8" AIR SUPPLY TUBING (yellow with black dashes) 1/8" REGULATED AIR TUBING (yellow) AIR 3 4 50 40 60 2 5 30 70 20 80 1 10 90 6 0 100 psi 0 kg/cm 2 7 1/8" MASTER AIR TUBING (yellow with red stripe) 1/8" PILOT AIR TUBING (yellow with red dashes) 3/8" WATER TUBING (blue with white dashes) WATER 1/8" 40psi REGULATED AIR TUBING (yellow with green dashes, ONLY used with self-contained water) FIGURE 35 (STANDARD PLUMBING WITH SELF-CONTAINED WATER) Page 14 85.0570.00

INSTALL THE POWER SUPPLY Refer to the installation instructions with the power supply. INSTALL THE TRAY HOLDER (Optional on Traditional-style controls, standard on Continental-style controls.) 1. Traditional-style control head: Remove the two setscrews from the control head arm (see Figure 37). SETSCREWS 3. Place the tray pad and tray on the tray holder. 4. To adjust the tray arm tension, insert a 9/64 hex key through the hole in the mounting block. Turn the tray holder and tray arm clockwise to increase the tension or counterclockwise to loosen the tension. LEVEL THE CONTROL HEAD 1. Place a bubble level on the top of the control head and check for level. 2. If leveling is necessary, loosen the two locking screws (see Figure 40) on the underside of the control head. Level the control head, side to side, using the two leveling screws (see Figure 40). After leveling, tighten the two locking screws. I MASTER O 2 3 4 LEVELING SCREWS FIGURE 37 2. Install the tray arm and holder (Traditional-style, see Figure 38; Continental-style see Figure 39). TRAY HOLDER MOUNTING BLOCKS FIGURE 40 LOCKING SCREWS TRAY ARM PUSH 1 2 3 4 TRADITIONAL-STYLE CONTROL HEAD FIGURE 38 MOUNTING SCREWS PUSH 1 2 3 4 TRAY HOLDER TRAY ARM MOUNTING BLOCKS MOUNTING SCREWS CONTINENTAL-STYLE CONTROL HEAD FIGURE 39 Page 15 85.0570.00

MA B L W A C 0 1 0 1 0 1 0 1 0 1 0 1 LEVEL THE ASSISTANT S ARM 1. Place a bubble level of the top of the assistant s arm and check for level. To level the arm, use a 1 8" hex key to adjust the leveling setscrew on the underside of the arm (see Figure 41). INSTALL THE CHAIR ADAPTER COVER 1. Install the chair adapter cover (included in the chair adapter ship kit) as shown in Figure 43. INSTALL THIS WAY WHEN POST BOX IS INLINE WITH THE CHAIR ADAPTER MA B L W A C 0 1 0 1 0 1 0 1 0 1 0 1 BUBBLE LEVEL 1/8" HEX KEY FIGURE 41 TELESCOPING ASSISTANT S ARM NOTE: POST BOX AND CHAIR STRUCTURES NOT SHOWN FOR CLARITY INSTALL THIS WAY WHEN POST BOX IS PARALLEL WITH THE CHAIR ADAPTER INSTALL THE SOLIDS COLLECTOR 1. Install the solids collector screen into the vacuum connector. Be sure to align the round screen opening with the HVE handpiece tubing. Install the vacuum connector and the screen in the post box as shown in Figure 42. FIGURE 43 INSTALL THE CUP FILL SPOUT AND BOWL SCREEN 1. Install the cup fill spout, cup fill spout gasket, and bowl screen on the cuspidor as shown in Figure 44. VACUUM CONNECTOR BOWL SCREEN SOLIDS COLLECTOR SCREEN (cap is installed to screen) CUP FILL SPOUT GASKET (notch to back of cuspidor) CUP FILL SPOUT FIGURE 42 FIGURE 44 Page 16 85.0570.00

TEST THE UNIT If the unit being tested does not include a self-contained water system, go to step 3. 1. Remove a bottle from the self-contained water system kit. Fill it with the water specified by the doctor. MASTER TOUCH PAD FUNCTIONS 5. Turn the system on by pressing the master on/off control on the Master Touch Pad (see Figure 45). The LED above the master on/off control will illuminate when the system is ON. NOTE The doctor may object to putting office water through the unit. Check with the doctor before filling the bottle. 2. Align the full bottle with the self-contained water cap in the post box. Be sure the 1 4" clear tubing from the cap extends straight down into the water bottle. Screw the bottle onto the cap until it is snug. COOLANT WATER ON/OFF CONTROL COOLANT AIR ON/OFF CONTROL CUSPIDOR BOWL RINSE ACTUATOR MASTER ON/OFF LED LIGHT 1 2 0 3 MASTER ON/OFF CONTROL 6300 DENTAL LIGHT ON/OFF CONTROL CUSPIDOR CUP FILL ACTUATOR 3. Fully open both the air and water manual shutoff valves. Check plumbing connections for leakage. 4. Plug in the power supplies. FIGURE 45 6. Operate the syringe while checking all plumbing connections for leakage. 7. Press the coolant air on/off and coolant water on/off controls (see Figure 45) to test the handpiece coolant. The LEDs above the control will illuminate when the functions are ON. 8. If a 6300 Dental Light is installed with the unit, check the function of the light by pressing the dental light on/off control on the Master Touch Pad (see Figure 45). The dental light will illuminate when the dental light on/off control is pressed. 9. For complete Cascade Master Touch Pad operation instructions, refer to the Cascade Master Touch Pad Owner s Guide, (Publication No. 85.2627.00). Page 17 85.0570.00

TOUCH PAD FUNCTIONS FOR THE CHAIR 10. The touch pad arrowhead buttons (see Figure 46) are used to manually position the chair. Back Down: Press and hold the back down (left) arrowhead to lower the chair back. Back Up: Press and hold the back up (right) arrowhead to raise the chair back. Base Up: Press and hold the base up (top) arrowhead to raise the chair base. Base Down: Press and hold the base down (bottom) arrowhead to lower the chair base. CUSPIDOR BOWL RINSE ACTUATOR CUSPIDOR CUP FILL ACTUATOR Pre-Position adjustment 12. Locate the program button and arrowhead buttons on the touch pad (see Figure 46). a.using the touch pad arrowhead buttons, set the chair to the doctor s preferred position. b.press and release the program button. You will hear a single audible tone. Within four seconds, press pre-position 0, 1, 2, or 3. You will hear multiple audible tones confirming that the new pre-position function for that button has been programmed. c. Check the new button function by moving the chair to another position and then pressing the pre-position button. The chair should move to the position programmed. 13. For complete touch pad chair operation instructions, refer to the Chair Owner s Guide. If the unit has an optional cuspidor, go to TEST THE (OPTIONAL) CUSPIDOR. 1 2 0 3 PRE-POSITION AUTO-RETURN BACK DOWN CUSPIDOR/ RETURN BASE UP BACK UP BASE DOWN PROGRAM BUTTON FIGURE 46 11. Pressing a pre-position button (see Figure 46) will send the chair s base and back to a pre-set position. Button 0 (auto-return): Programmed at the factory to send the chair to an entry/exit position (base down, back up). Buttons 1 and 2: Programmed at the factory to send the chair to the same position when pressed. The doctor may want these positions programmed. Button 3 (cuspidor/return): Programmed at the factory to send the chair to a cuspidor position. When pressed a second time, the chair will return to its previous position. Page 18 85.0570.00

TEST THE (OPTIONAL) CUSPIDOR 14. Place a cup under the cup fill spout and press the cup fill actuator button on the cuspidor (see Figure 47). CUP FILL SPOUT CUP FILL ACTUATOR BUTTON BOWL RINSE ACTUATOR BUTTON BOWL RINSE SPOUT 15. Press the cup fill actuator on the Master Touch Pad (see Figure 46 on page 18). When activated, water flow from the cup fill spout should be similar to water flow when the cuspidor cup fill actuator button was pressed in Step 14. 16. Press the bowl rinse actuator button on the cuspidor (see Figure 47). If bowl rinse flow adjustment is necessary, locate the bowl rinse flow adjustment on the water manifold in the post box (see Figure 47). Notice the location of the indicator dot showing the position of the current setting (see NOTE). Using a 1 8" hex key, turn the key clockwise to increase the bowl rinse volume or counterclockwise to decrease the volume. VOLUME ARROWS BOWL RINSE ADJUSTMENT 17. Press the bowl rinse actuator on the Master Touch Pad (see Figure 46 on page 18). When activated, water flow from the bowl rinse spout should be similar to water flow when the cuspidor bowl rinse actuator button was pressed in Step 16. INDICATOR DOT CUP FILL ADJUSTMENT BEFORE LEAVING THE FACILITY 1. Replace all covers. 2. Install the chair seat upholstery. 3. Distribute the Owner s Guides and instruct the dental team in the operation and maintenance of the equipment. FIGURE 47 (Shown with optional assistant s instrumentation without the master air signal valve manifold for clarity) If cup fill flow adjustment is necessary, locate the cup fill flow adjustment on the water manifold in the post box (see Figure 47). Notice the location of the indicator dot showing the position of the current setting (see NOTE). Using a 1 8" hex key, turn the key clockwise to increase the cup fill volume or counterclockwise to decrease the volume. NOTE When making adjustments, the indicator dots on the volume adjustments should remain within the volume arrows. Page 19 85.0570.00

1998-12 Rev E (00412) Recycle Conserve Made with 10% post-consumer fiber Printed in U.S.A. Copyright 1998 All Rights Reserved Page 20 85.0570.00