Setting up SilFlow for BackFlush in your GC

Similar documents
Choose for Your Instrument. GC Supplies - Agilent Instrument Quick Pick Guide

GC Supplies - Varian/Bruker Instrument Quick Pick Guide

Choose for Your Instrument. GC Supplies - Bruker (Varian) Instrument Quick Pick Guide

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

Using the PSD for Backflushing on the Agilent 8890 GC System

Application Note. Abstract. Authors. Environmental Analysis

Analysis of Petroleum Fractions by ASTM D2887

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

Alternative Carrier Gases for ASTM D7213 Simulated Distillation Analysis

AppNote 6/2006. Ultra-Fast Determination of the Hydrocarbon Oil Index by Gas Chromatography using a Modular Accelerated Column Heater (MACH) KEYWORDS

Impact of GC Parameters on The Separation Part 2: Choice of Column Internal Diameter

GAS CHROMATOGRAPHY: INJECTION TECHNIQUES CAPILLARY COLUMNS

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System

Application. Gas Chromatography June 1995

Agilent G2855A Deans Switching System

AZ Chrom s.r.o.

SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)

Deans Switch Microfluidics

STI OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

[ APPLICATION NOTE ] INTRODUCTION APPLICATION BENEFITS WATERS SOLUTIONS KEYWORDS

Method Development for Capillary GC Systems. Slide 1

Fast GC. Dial for e-seminar Audio. Slide 4 SPEEDY GC

Practical Steps in GC Troubleshooting

C2, C3, C4 Monomer Analysis

Installing the 6850 Direct Valve Column Connector Kit

Agilent and ASTM. Update on Recent Activities. Page 1

Refinery Gas. Analysis by Gas Chromatography WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

High-Temperature Simulated Distillation System Based on the 6890N GC Application

Methanol in Biodiesel by EN14110 with the HT3 and Versa Automated Headspace Analyzers. Versa HT3. Application Note. Abstract.

Same Column and Gas Type Try Different/Faster Velocities. Same Column, Switch He to H2 Carrier Then Try Faster Velocities

Stability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas

The Analysis of Hydrocarbon Composition in LPG by Gas Chromatography using the DVLS Liquefied Gas Injector

Optimizing Ultra Fast Simulated Distillation with a Low Thermal Mass (LTM) GC System. Jim McCurry Roger Firor Agilent Technologies Wilmington, DE

High Temperature Simulated Distillation Performance Using the Agilent 8890 Gas Chromatograph

GC Method Compliance and Large Valve Oven Application

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

Agilent Multimode Inlet for Gas Chromatography

Technical Procedure for Gas Chromatography (GC-FID)

PRODUCT SPECIFICATION. Product Specification Crystal 9000 Gas Chromatograph. Chromatec Crystal Laboratory Gas Chromatography System

HIGH PRESSURE SWITCHING VALVES

Agilent 6850 Series II Network GC System G2630A Performance Specifications

DANI Transformer Oil Gas Analyzer

Analysis of Fatty Acid Methyl Esters (FAMES), and Examination of Biodiesel Samples for these Components, by GCxGC-FID

ANALYSIS OF GASOLINE RANGE HYDROCARBONS ON BP1-PONA

Typical Gas Chromatographic System

Proof of Long-Term, Leak-Free Performance for a Novel Self-tightening GC Column Nut

GC Inlets. An Overview. Simon Jones GC Applications Engineer

6890 CHECKOUT PROCEDURE TCD (THERMAL CONDUCTIVITY) DETECTOR

260 Solvent Vapor Exit Accessory

ASTM D2887 Simulated Distillation Calibration Mixture Analysis Using a Differential Acceleration Column

S 1125 HPLC Pump System

performance productivity reliability

Application Note. Authors. Abstract. Energy & Chemicals

TECHNICAL REPORT. Introduction. Agilent 1100 HPLC system. Figure 1: HALO columns exhibit UHPLC-like performance at conventional HPLC pressure

Choose for Your Instrument

IFPAC 2003 Dr. Berthold Andres

Detection of Sulfur Compounds in Natural Gas According to ASTM D5504 with an Agilent Dual Plasma Sulfur Chemiluminescence Detector

HT8: THE PERFECT PCB COLUMN

ASTM D 6730 Detailed Hydrocarbon Analysis

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC

GC Best Practices & Troubleshooting. Group/Presentation Title Agilent Restricted Month ##, 200X

MET-Biodiesel Capillary GC Columns

Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS)

Operation and Applications of Differential Flow Modulation

GCMS-QP2010 Series. Consumables. Shimadzu Gas Chromatograph/Mass Spectrometer

2-D GC/MS and the Deans Switch: Theory and Applications. Fred Feyerherm, Agilent Technologies Inc. Houston, Tx

GC Injector Liners. Bruker Varian Agilent (HP) Perkin Elmer Finnigan. Phone: Page 4

Jaap de Zeeuw Restek Middelburg, The Netherlands. Copyrights: Restek Corporation

REFINED RESTEK. Detailed Hydrocarbon Analysis. innovative petrochemical solutions. Featuring Rtx -DHA Columns

URB '-carbamoylbiphenyl-3-yl cyclohexylcarbamate. DEA Reference Material Collection. Form Chemical Formula Molecular Weight Melting Point ( o C)

Model 8610C Gas Chromatograph. GC Chassis Types ECD PID FID / DELCD FPD. Mounts up to Six Detectors and Five Injectors

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

Agilent 1290 Infinity Autosampler

Analysis of Glycerin and Glycerides in Biodiesel (B100) Using ASTM D6584 and EN Application. Author. Abstract. Introduction

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

Free and Total Glycerol in B100 Biodiesel by Gas Chromatography According to Methods EN and ASTM D6584

Agilent 7693A Automated Liquid Sampler

Agilent 7693A Automated Liquid Sampler

Phenomenex Exclusive! Inlet Base Seals. Agilent 5000, 6000 and. Guaranteed fit for Series GCs. Easy Seals

Agilent InfinityLab 2D-LC Solution with mass spectrometric detection and diverter valve

The next level of True Blue Performance

Oxygenates in Fuels Analysis Solutions From Trace Levels to Ethanol Fuels

Fausto Munari e Andrea Cadoppi ThermoFisher - Italy

Analysis and. Separation of Oxygenates in Hydrocarbon Matrices. Simon Jones Application Engineer Folsom, CA

Simple Heart Cutting with Deans. Capillary Flow Technolgy. Simon Jones Applications Engineer

SFC-FID ASTM System. SFC-4000 Series SFC. Performance Innovation Reliability

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

Agilent Multimode Inlet

AZ Chrom s.r.o.

Chromatography. Liquid. HPLC Accessories SupelPRO Valves SUPELCO

Agilent Valve Kit G4231A/B and G4232A/B/C

Agilent 6890 Gas Chromatograph

For more information about Agilent Technologies services please visit our web site using the following URL

Simplify your search for. You will find both PEEK-lined stainless steel hardware as well as standard ss column hardware at the below stockingstations.

SPECIFICATION. Bench top triple quadrupole Mass Spectrometry System coupled to a gas performance Chromatography (GC-MS/MS)

Nickel Catalyst Tube Accessory 19205A

Transcription:

Setting up SilFlow for BackFlush in your GC What is backflush and why use it? The BackFlush system eliminates the need to bake heavy sample fractions off the capillary column. Oils, tars and other semivolatile matter can be flushed back out of the injection port while the oven remains at a relatively low temperature. This increases column lifetime dramatically. Run times will also be significantly shorter because the system allows the run to be terminated as soon as the compounds of interest have eluted from the column, with everything else being flushed out the split vent. This results in shorter cycle times, lower maximum oven temperatures, longer column life and no carryover between samples. BackFlushing saves time and keeps detectors clean. Using the SGE SilFlow three or four port microchannel device (Figure 1), for BackFlushing through a precolumn before the midpoint auxillary automated pressure control line, keeps the analytical column in pristine condition as chemicals of low volatility are back flushed before they even reach the analytical column. Silflow TM features SGE s latest chromatography solution SilFlow, is an innovation in design and fabrication resulting in a highly efficient and reliable microchannel device (MCD) for gas chromatography. 1a. 1b. Figure 1a. SilFlow GC 3 Splitter, Figure 1b. SilFlow GC 4 Splitter. Chemically deactivated internal channels result in analysis with enhanced quantitative accuracy and high reproducibility. When incorporated into the GC chromatography system there is no impact on routine chromatography. Low dead volume connections - SilFlow FingerTite metal ferrules result in a reliable zero dead volume connection, giving you optimized peak shapes. Superb operational stability - thermal lag is negligible as the device tracks the oven temperature up to 20 C/min. The design alleviates cold spots and sample condensation. Maximum temperature No practical temperature limit. Limited only by the temperature rating of the GC column being used ( 420 C). Easy to install and leak free, each kit incorporates SilFlow FingerTite fittings that are easy to set up and can be tightened using finger force to achieve a perfect, reliable seal, even for the most sensitive MS systems no wrenches are required. Using SilFlow for Backflush This article demonstrates 3 different configurations that can be used for backflushing using SilFlow. How to program a backflush run will be outlined, then the options for a pre-column, post-column, or dual detector set up will be discussed. Programming a Backflush run Backflushing can be achieved by the following sequence: 1: Calculate the correct midpoint pressure required to give the desired average linear velocity in the analytical column using a GC flow and pressure software application. Enter the analytical column dimensions, temperature and desired linear velocity to calculate the midpoint pressure. 2: Turn off the auxiliary carrier gas supply to the midpoint and adjust the injection port pressure until the midpoint has reached the required pressure to give the desired average linear velocity in the analytical column. 3: With the auxiliary pressure (i.e. midpoint) off, perform a run without backflushing and estimate from the chromatogram the time that you require to backflush - this estimated time is when the last analyte of interest passes the endpoint. 4: To backflush, the auxiliary pressure should be turned down until all required analytes have reached the detector, at which time it must be turned on to provide pressure to the analytical column to backflush to the injector and out the split vent. The head pressure must be switched to the lowest possible value the system is capable of performing when the backflush commences. The injector must be in split mode. Example 1 - Backflush Experiment using a 3 SilFlow MCD pre-column Mass Mass Gas Figure 2. Diagram of backflush system with, Mass and Automated Control.

Specific conditions used for Example 1: Backflush Experiment Using a 3 Silflow MCD Pre Column. Oven: 40 C for 2 min, ramp at 20 C/min to 300 C and hold for 10 min. Run time 25 min Initial temp 280 C 17.4 psi Split ratio: 100:1 : FID: 400 C Hydrogen Flow: 40 ml/min Air flow: 450 ml/min Makeup flow 45.0 ml/min MSD: Solvent delay: 2.6 min Backflush device: 3 Silflow MCD (Micro Channel Device) Analytical Column (MCD to MS detector): x 0.25 µm MCD Midpoint to connection: deactivated fused silica Pre column(inj port to MCD): 10 m x 0.25 x Backflush Mode: Ramp Initial 17.4, Hold for 6 min, then ramp @150 psi/min to 0.01 psi and hold for 19 minutes Agilent control: Initial 10.4, Hold for 20 min No Backflush Mode: Ramp Initial 17.4, Hold for 20 min Agilent control: Initial 10.4, Hold for 20 min Carrier : Alkane mix in pentane The following trace (Figure 4), shows the MS chromatogram without backflushing. Figure 4. Alkane test mix response on MS pre-column backflush. The same test mix was injected with the backflush initiated at 6 minutes (Figure 5). The backflush timing has to be estimated taking into account the type of carrier gas used, the type of analytes being tested, as well as the temperature, dimensions, and pressure of the pre-column and the analytical column. A few analyses will give an experimentally obtained optimum backflush time. In the following example all of the less volatile alkanes have been backflushed out the injection port split vent. Figure 5. Alkane test mix with pre-column backflush initiated at 6 minutes. To Final To Mass Spectometer 10 m x Example 2 - Backflush Experiment using a 3 SILFLOW MCD post column Figure 3. Example of a backflush system with capillary tubing and analytical Figure 3 - This system consists of a midpoint Auxillary EPC (electric pressure control) line to a three port MCD, a 10 meter pre SilFlow column, and a 30 meter post SilFlow analytical column. The objective was to exclude any alkane greater in mass than reaching the analytical column and mass spectrometer. Gas Figure 6. Diagram of backflush system with,, and Automated Control. Figure 7- This system consists of a midpoint auxiliary EPC line, a pre-splitter 30 meter analytical column, and a post splitter restrictor connected to the mass spectrometer.

Mass To Final analytical column 0.650 x 0.11 mm Column Figure 7. Example of a Backflush system with capillary tubing and analytical Specific conditions used for Example 2: Backflush Experiment Using a 3 Silflow MCD Post Column. Oven: 40 C for 2 min, ramp at 20 C/min to 300 C and hold for 10 min. Run time 25 min Initial temp 280 C 17.4 psi : Solvent delay: 2.6 min Backflush device: 3 Silflow MCD (Micro Channel Device) Analytical Column: x 0.25 µm MCD Midpoint to connection: deactivated fused silica MCD to MS detector Connection restrictor: 0.650 m x 0.11 mm deactivated fused silica column Carrier Backflush: Mode: Ramp Initial 17.4, Hold for 6 min, then ramp @150 psi/min to 0.01 psi and hold for 19 minutes Agilent control: Set midpoint EPC pressure to Off (at start of run) Agilent Runtime Midpoint Control: Set runtime event to start midpoint EPC pressure setting at 17.4psi No Backflush: Leave midpoint pressure off and injection port pressure on throughout whole run : Alkane mix in pentane The same objective of excluding any alkane greater in mass than reaching the analytical column and mass spectrometer was tested. Figure 8 trace shows the MS chromatogram without backflushing. The same test mix was injected with the backflush initiated at 6 minutes (Figure 9). The backflush timing has to be estimated taking into account the type of carrier gas used, the type of analytes being tested, as well as the temperature, dimensions, and pressure of the precolumn and the analytical column. A few analyses will give an experimentally obtained optimum backflush time. In the following example all of the less volatile alkanes have been backflushed out the injection port split vent. Figure 9. Alkane test mix response on MS backflush initiated at 6 minutes. Example 3 - Backflush Experiment with 4 SilFlow MCD two detector backflush Figure 10. Diagram of backflush system with,,, and Automated Control. To final To Final Ionisation deactivated fused silica Ionisation Figure 11. Four SilFlow set up. 0.650m x 0.11mm deactivated fused silica restrictor Ionisation C28 10 m x C34 Figure 8. MS Chromatogram without backflushing on Alkane test mix post-column backflush. Figure 12. Example of a backflush system with capillary tubing and analytical

Specific conditions used for Example 3: Backflush Experiment Using a 4 Silflow MCD - Two detector backflush. Oven: 40 C for 2 min, ramp at 20 C/min to 300 C and hold for 5 min. Run time 20 min 100:1 Temp: 280 C Initial 17.4 psi : FID: Temp: 400 C Hydrogen Flow: 40.0 ml/min Air Flow: 450 ml/min Nitrogen make up Flow: 45 ml/min MSD: Solvent delay: 2.6 min Backflush device: 4 Silflow MCD (Micro Channel Device) Analytical Column: x 0.25 µm MCD Midpoint to connection: deactivated fused silica MCD to FID detector Connection restrictor: 0.650 m x 0.11 mm deactivated fused silica column to MCD: 10 m x x 0.25 µm Carrier Backflush Mode: Ramp Initial 17.4, Hold for 6 min, then ramp @150 psi/min to 0.01 psi and hold for 19 minutes Agilent control: Initial 10.4, Hold for 20 min No Backflush Mode: Ramp Initial 17.4, Hold for 20 min Agilent control: Initial 10.4, Hold for 20 min : Alkane mix in pentane This system uses two detectors. A detector (FID) and a mass spectrometer (MS). A 30 meter analytical column is connected to the midpoint four port SilFlow and the MS (see Figure 12). A 10 meter pre-column is connected to the injection port and the midpoint SilFlow. The auxiliary EPC is connected to the midpoint. The flow rate (ml/min) from the midpoint is balanced between the FID and the MS detectors. A restrictor that has the appropriate dimensions to balance the flow is installed between the midpoint and the FID. This will ensure that good signals will be recorded on both detectors. Prior to the backflush experiment a sample of alkanes was injected to establish the observed time for the backflush event. The objective was to exclude any alkane greater in mass than reaching the analytical column and mass spectrometer. Figure 13 trace shows the MS chromatogram on top and the midpoint FID chromatogram underneath. While the most volatile alkane, octane, appears on the MS trace, the octane peak is missing from the FID trace due to the FID data acquisition only commencing once the MS solvent delay had finished. Based on the observed elution time at the FID, the backflush event was timed for 6 minutes. Figure 13. Alkane test mix response on MS and FID The same test mix was injected with the backflush initiated at 6 minutes. In Figure 14, the top MS trace shows that the backflush was successful and no alkane greater in mass than made it onto the analytical column all of the less volatile alkanes have been backflushed out the injection port split vent. MS trace C28 FID chromatogram MS trace FID chromatogram Figure 14. Alkane test mix response on MS and FID with backflush initiated at 6 minutes. Taking care of your SILFLOW MicroChannel Device Tubing to the SilFlow Microchannel Device (MCD) can be disconnected and reconnected many times without removing the pre-swaged ferrules. However, it is very important to inspect the capillary end carefully before reconnecting to the SilFlow MCD. Make sure the end of the capillary is intact. If the capillary does not have a clean square end, the column end has to be cut again and a new ferrule to be pre-swaged. When it is not connected, protect the SilFlow MCD from particulates and dust getting into the internal channels. Make sure to block the bosses using either self-sealing nuts or pre-swaged ferrule to pieces of metal wire and appropriate FingerTite nuts.

Part Number Kits Part Description 123722 SilFlow 3 MCD Kit 123732 SilFlow 4 MCD Kit Tubing 123751 2 m x 100 µm/363 µm VSD Tubing 0624431 2 m x 250 µm/363 µm VSD Tubing 0624459 2 m x 110 µm/310 µm VSD Tubing Ferrule 123713 SilFlow FingerTite Ferrules 0.35 mm ID pk10 (for use with 310 µm OD tubing) GC Capillary Column 054101 30 m ID x 0.25 µm Table 1. Part numbers for ordering SilFlow 3 and 4 MCD Kits for backflushing. Initial installation kits have all components required to get you started. Tubing and GC capillary column reference in this article are also listed. For more information contact our technical customer support team on: techsupport@sge.com AUSTRALIA & PACIFIC REGION SGE Analytical Science Pty Ltd Toll Free: 1800 800 167 Tel: +61 (0) 3 9837 4200 Fax: +61 (0) 3 9874 5672 Email: support@sge.com CHINA SGE Shanghai Representative Office Tel: +86 21 6407 9382 Fax: +86 21 6407 9386 Email: china@sge.com MIDDLE EAST SGE Gulf Tel: +971 6 557 3341 Fax: +971 6 557 3541 Email: gulfsupport@sge.com TA-0144-A SGE Analytical Science Pty Ltd 12/2011 EUROPE SGE Europe Ltd European Head Office Toll Free: 00800 2790 8999 Toll Free Fax: 00800 2626 2609 Tel: +44 1908 568 844 Fax: +44 1908 566 790 Tel France: +33 1 69 29 80 90 Fax France: +33 1 69 29 09 25 Tel Germany: +49 (0) 6155 / 60746 0 Fax Germany: +49 (0) 6155 / 60746 50 Email: europe@sge.com INDIA SGE Laboratory Accessories Pvt Ltd Tel: +91 22 24715896 Fax: +91 22 24716592 Email: sgeindia@vsnl.com JAPAN SGE Japan Inc Tel: +81 45 222 2885 Fax: +81 45 222 2887 Email: japan@sge.com UNITED STATES OF AMERICA SGE Incorporated Toll Free: (800) 945 6154 Tel: +1 512 837 7190 Fax: +1 512 836 9159 Email: usa@sge.com