HanTron Model HT-3P. Manual. Installation Maintenance Repair Manual

Similar documents
Manual. Economy Glycol Feeder. Installation Maintenance Repair Manual

Analog Condensate Monitor Instruction & Maintenance Manual

MicroTron Series C. Manual. Installation Maintenance Repair. Manual. Chemical Metering Pump

EC1 Pocket Conductivity Tester Instruction Manual

TRACER. POCKETESTER ä ORP CODE 1742

Manual. MicroTron. Series C. Installation Maintenance Repair Manual. Chemical Metering Pump

Tracer ORP PockeTester TM Code 1742

EC20 Pocket Conductivity Tester Instruction Manual

SoilStik ph Meter PRODUCT MANUAL. Item # 2105

ECscan20 Pocket Conductivity Tester Instruction Manual

PC60 Premium Multi-Parameter Tester (ph/ec/tds/salinity/temp.) Instruction Manual. APERA INSTRUMENTS, LLC V6.4

ECscan30 Pocket Conductivity/TDS Tester Instruction Manual

MoistureMatch A next generation grain tester

Corrosion Coupon Rack

Manual. MicroTron. Series B. Installation Maintenance Repair. Manual. Chemical Metering Pump

TDSscan10M Pocket TDS Tester Instruction Manual

Salt 1 Pocket Salinity Tester Instruction Manual

Grain Moisture Meter. Operators manual

Care and use guide Guía de uso y cuidado Guide d utilisation et d entretien.

The function of this Dynamic Active Probe has divided into three preferences on the screen main Menus:

PH20 Value ph Tester Instruction Manual

Breath Alcohol Tester

IAQ-CALC INDOOR AIR QUALITY METER MODEL 7515

ECscan40 Pocket Conductivity/TDS/Salinity Tester Instruction Manual

NORMAL/DIAL BRIGHTNESS: Press the down or up button to adjust dial brightness. Press the center button to save and advance to PEAK PLAYBACK.

Model P772 Conductivity Meter. User Manual

TABLE OF CONTENTS 11.0 WARRANTY

General Information. Installation Tips. Connections

Coating Thickness Gauges

Temperature Programming Quick Reference Guide

PHscan30S Pocket ph Meter Instruction Manual

MOISTURE METER 2155 OPERATION INSTRUCTIONS VERSION 1.0

Manual. Series 2EZ. Installation Maintenance Repair Manual

Digital Resistivity Meter. Part No Instruction Manual. Updated 09/01/2015 Ver. 3.0

IAQ-CALC TM Indoor Air Quality Meter

SB-300 Operator s Manual

PH60 Series Premium ph Testers

HI Portable Waterproof Skin ph Meter INSTRUCTION MANUAL

Manual. Model GF. Digital Glycol Feeder. Installation Maintenance Repair Manual

SPC Series. Digital Scale. Operation Manual

Flue Gas Analyzer IMR 1100F

ORPscan20 Pocket ORP Tester Instruction Manual

AXD 530 MicroManometer

MCL-3014 gauge kit. Optional Readings: Boost Pressure with MBM-09, Front or Rear Air Suspension Pressure with MBM-19

User Guide. Model Insulation Tester / Megohmmeter. Introduction

ECscan10H Pocket Conductivity Tester. Instruction Manual BANTE INSTRUMENTS LIMITED

Bante220 Portable ph Meter Instruction Manual

All products subject to this warranty must be returned for examination, repair or replacement to:

GARNET. SEE Tester II TM INSTRUMENTS LTD. Model 900M Manual PRINTED IN CANADA

Large Display ph Pens , , Instruction Manual

THREE RANGE DS METERS Operation Instructions

AUTO-BLiP. User Manual Porsche INTELLIGENT DOWNSHIFTS. Version 1.2

AUTO-BLiP. User Manual Chevrolet Corvette. Version 1.7

AUTO-BLiP. User Manual Chevrolet Corvette. Version 1.2

AUTO-BLiP. User Manual Lotus INTELLIGENT DOWNSHIFTS. Version 1.0

ITA / ITB - AS / AP / AP SERIES Electronic Table Top Weighing Balance

User's Manual. Model Hot Wire Thermo-Anemometer. Introduction

COMPACT - TRAINING MANUAL

Installation Instructions Model: AB Single-lever Pull-down Spray Faucet

AUTO-BLiP. User Manual Ford Mustang INTELLIGENT DOWNSHIFTS. Version 1.2

User Manual Pocket Pro, Pro + ph

GRAVITY MEASUREMENT, INC.

Fission. Nano-Protein Skimmer. Instructions for Model#8100

4&5 RANGE CONDUCTIVITY METERS Operation Instructions

Operator's Manual Dielman Rock Island Industrial Dr. St. Louis, MO Tel:(314)

Care and use guide. Contents. page

Model 322 Automated Thermocouple Calibrator Operating Instructions. Product Description. Practical Instrument Electronics

Operating and Service Manual

Care and use guide. Contents. page

American Weigh Scales SG-300. User Manual. SG-300 (300g x 0.1g) Copyright 2012 American Weigh Scales, Inc. All rights reserved. Rev. 2.

PIECAL 322 Automated Thermocouple Calibrator Operating Instructions. Product Description

NEXUS. Introduction SENSOR MODULE &

MODEL MCL-3212 SPEEDOMETER/TACHOMETER for 2012 up Dyna and Softail with 4 gauge

Thermo-Anenometer Model Owner s Manual

Installation Instructions Model: AB Single-Lever Pull-out Kitchen Faucet

Valor 1000W Series Instruction Manual. Serie Valor 1000W Manual de Instrucciones. Série Valor 1000W Guide de I utilisateur

Andatech SOBERPOINT 3. Wall Mounted Breathalyser USER S MANUAL

Installation Instructions for: Channel Thermocouple Amplifier

Care and use guide. Contents. page

Wireless Thermometer model 00782W

DT304. Digital Temperature Logger INSTRUCTION MANUAL

MICROMANOMETER ALNOR MODEL AXD610 AIRFLOW INSTRUMENTS MODEL PVM610

Chek-A-Cell TM. Instruction Manual Rev A 8/12

REMOVAL OF FACTORY GAUGE ULTRA FLHT & FLHX (STREET GLIDE

HI Ethylene Glycol Refractometer INSTRUCTION MANUAL

PLS SLD. Laser Detector PACIFIC LASER SYSTEMS. For use with your PLS pulsed laser alignment tool. The Professional Standard

BBT-205. For 12-volt automotive starting batteries and starting/charging systems INSTRUCTION MANUAL

SERVICE MANUAL. STARTER Portable Meter ST300 ST300C ST300D ST400D. Ohaus Corporation, 7 Campus Drive, Suite 310, Parsippany, NJ (973)

Advanced Calibration Designs, Inc. Release II

Tracer VMA with AutoReg

MODEL MCL /8 SPEEDOMETER/TACHOMETER for 2004 up

Digital Body Weight Scale Model No.: EB4074C. Questions or Concerns? (855)

STUDIO - TRAINING MANUAL

Model FM1000 Series Operating Instructions. Drum Pump Flow Meter

Part Number Analog EGT Gauge

User s Guide HH42A. Thermometer. Shop online at omega.com sm HH42A. For latest product manuals:

Interior Digital Load Scale 202-DDG-01. Installation and Operation Manual Please read carefully before installation

MDX-300 Series. For 12-volt automotive starting batteries and starting/charging systems INSTRUCTION MANUAL

CO95. Carbon Monoxide Detector INSTRUCTION MANUAL ENGLISH Fax: (503)

IAQ-CALC TM Indoor Air Quality Meter

Transcription:

Manual HanTron Model HT-3P Installation Maintenance Repair Manual Advantage Controls P.O. Box 1472 Muskogee, OK 74402 Phone: 800-743-7431 Fax: 888-686-6212 www.advantagecontrols.com email: support@advantagecontrols.com 9/2015 1

Features Conductivity Accuracy Temperature ± 2 F / 1 C ± 1% 0 to 800 µs/cm ± 1.5% 800 to 5,500 µs/cm ± 2% 5,500 to 20,000 µs/cm Rugged integral probe. Four-electrode cell technology. 5 digit LCD displays full 20,000 µs/cm or PPM. Full auto ranging conductivity measurements from 0 to 20,000 µs/cm. Memory saves up to 20 readings. Calibrations protected from low battery loss (nonvolatile memory). Specifications Display Dimension (LxWxH) Conductivity Cell Capacity Power Battery Life Operating/Storage Temp Protection Rating 5 digit LCD 5.3 x 3.3 x 1.6 in. (136 x 84 x 39 mm). 0.1 oz. (3.2 ml). (2) AAA Alkaline Batteries 10,000 Readings. 32-132 F / 0-55 C IP64 2

Measuring Conductivity and Temperature 1. Rinse conductivity cell three times with sample. 2. Fill cup with sample. All four electrodes must be submerged. Make sure there are no bubbles in cell. 3. Press READ/ENTER: Display shows dashes while reading. Discard this reading and refill cup. 4. READ/ENTER: Measure conductivity of sample. Display shows dashes while reading. Use this reading. 5. : Displays temperature of the sample. 6. : Returns to conductivity of the sample. 7. Power turns off automatically after 30 seconds of inactivity. Storing Sample Measurements to Memory The tester can save up to 20 readings. 1. / SAVE: Displays next available memory location. 2. / SAVE or : To select a specific memory location. 3. READ/ENTER: Stores conductivity and temperature data in location shown. 4. MENU: Exits memory options. CONDUCTIVITY µs/cm MEM NOTE: Display shows -- when memory is full. Data cannot be overwritten, it must be erased via MENU commands. Menu Options MENU is used to: Recall stored measurement records. Erase memory (individual records or all at once). Calibrate conductivity. Calibrate temperature. Change conductivity unit of measure (µs/cm or PPM). Change temperature unit of measure ( C or F). Press MENU at any time to go to reading without saving changes. 3

Recall Measurements Stored in Memory 2. MENU: Displays menu items. (RECALL and MEM flashes). 3. READ/ENTER: Displays memory locations. 4. / SAVE or : To choose a record to view. Display will alternate between conductivity and temperature. Erase Individual Records from Memory 2. MENU: Displays menu items. 3. / SAVE or : To display CLEAR and MEM (flashing). 4. READ/ENTER: Selects clear memory function. 5. / SAVE or : To choose record to clear. 6. READ/ENTER: Displays data that is to be cleared. 7. READ/ENTER: Clears data from selected memory location and returns to measurement mode. Erase All Records from Memory 2. MENU: Displays menu items. 3. / SAVE or : To display CLEAR and MEM (flashing). 4. READ/ENTER: -- MEM flashes to indicate all records are about to be deleted. 5. READ/ENTER: Clears data from all memory locations and returns to measurement mode. Calibrate Conductivity Conductivity calibration should be performed monthly, or upon changing batteries. There are three internal scales that each hold a different calibration value. Each of these scales can only be calibrated with a solution value within an allowable range. Solutions used outside of the allowable value for each scale will result in an error. Allowable Values for Calibration Operating Range Low 120 to 700 µs/cm ~ 0 to 800 Mid 1,200 to 4,000 µs/cm ~ 800 to 5,500 High 12,000 to 25,000 µs/cm ~ 5,500 to 25,000 4

1. Rinse conductivity cell three times with proper conductivity standard solution. 2. Refill conductivity cell with the same standard solution. 3. Press READ/ENTER: Display shows dashes while reading. Discard this reading. 4. Refill cell and press READ/ENTER: Use this reading. 5. MENU: Displays menu items. 6. / SAVE or : To select CAL and CONDUCTIVITY (flashing). 7. READ/ENTER: Displays the conductivity of the standard solution. CAL CONDUCTIVITY µs/cm 8. / SAVE or : Change the conductivity reading to match the value of sample. Hold down key for rapid scrolling. 9. READ/ENTER: Stores new calibration factor and returns to the measurement mode. Call and ask us about our calibration kits. 148 Low 3,482 Mid 12,880 High Calibrate Temperature Note: Temperature calibration is not recommended. 2. MENU: Displays menu items. 3. / SAVE or : To select CAL TEMP (flashing) 4. READ/ENTER: Display current temperature. 5. / SAVE or : Change temperature to desired value (± 5 F, ± 5 C from current reading) 6. READ/ENTER: Stores calibrated temperature and returns to measurement mode. 5

Select Fahrenheit or Celsius 2. MENU: Displays menu items. 3. / SAVE or : To select TEMP F (or C) (flashing) 4. READ/ENTER: Enables unit of measure selection. 5. / SAVE or : Select F or C. 6. READ/ENTER: Stores unit of measure and returns to measurement mode. Replace Batteries BATT will display constantly when batteries are low. 1. Dry the instrument thoroughly. 2. Remove the single battery cover screw from bottom of tester. 3. Lift battery cover starting at screw end. 4. Replace with two of the same kind AAA alkaline batteries. 5. Replace battery cover starting at non-screw end. 6. Re-install battery cover screw. Note: Battery panel needs to be flush for a water-resistant seal. NOTE: Stored measurement records will be lost without battery power. Removing the batteries will not affect the calibration values and other menu settings. However, conductivity calibration is recommended at this time. Troubleshooting To reset all menu items to the factory settings: 2. Press and hold all four buttons for five seconds. 3. Display shows the firmware version and then shows reset 4. READ/ENTER: Reset to factory defaults. NOTE: Press MENU to exit without resetting the meter. The conductivity cup must be empty and dry before resetting the device to factory settings. If the cup is not dry, the reset will indicate a fail. 5. Factory defaults are not calibrated. Follow Calibrate Conductivity instructions in this manual to recalibrate. 6

Errors Codes (unit reads Err0r ) 01 - Thermistor shorted or temperature higher than expected. 02 - Thermistor broken or temperature lower than expected. 03 - Conductivity lower than expected. 04 - Conductivity higher than expected. 05 - Calibration solution reading out of range. 06 - Unable to calibrate, reading is unstable. 07 - Unable to zero, reset with a clear dry cell. 08 - Conductivity reading is unstable. 09 - Test station unexpected range. Care and Maintenance Keep conductivity cell as clean as possible. Flush conductivity cell with clean water after use to prevent buildup on electrodes. Clean oily films or organic material from conductivity cell with foaming non-abrasive household cleaner. Do not scrub inside cell. Avoid solvents. Take care not to drop the tester. Shock may damage components. Do not place solutions hotter than 160 F (71 C) in conductivity cell. Do not exceed rated operating temperature. Take care not to leave tester in a vehicle or unairconditioned building on a hot day as temperatures may surpass 150 F. 7

Manufacturer s Product Warranty Advantage Controls warrants units of its manufacture to be free of defects in material or workmanship. Liability under this policy extends for 24 months from date of purchase. Liability is limited to repair or replacement of any failed equipment or part proven defective in material or workmanship upon manufacturer s examination. Removal and installation costs are not included under this warranty. Manufacturer s liability shall never exceed the selling price of equipment or part in question. Advantage disclaims all liability for damage caused by its products by improper installation, maintenance, use or attempts to operate products beyond their intended functionality, intentionally or otherwise, or any unauthorized repair. Advantage is not responsible for damages, injuries or expense incurred through the use of its products. The above warranty is in lieu of other warranties, either expressed or implied. No agent of ours is authorized to provide any warranty other than the above. Opening the meter is discouraged. Doing so may increase your chance of damaging the meter and may void your warranty. 30 Day Billing Memo Policy Advantage Controls maintains a unique factory exchange program to ensure uninterrupted service with minimum downtime. If your unit malfunctions, call 1-800-743-7431, and provide our technician with Model and Serial Number information. If we are unable to diagnose and solve your problem over the phone, a fully warranted replacement unit will be shipped, usually within 48 hours, on a 30 Day Billing Memo. This service requires a purchase order and the replacement unit is billed to your regular account for payment. The replacement unit will be billed at current list price for that model less any applicable resale discount. Upon return of your old unit, credit will be issued to your account if the unit is in warranty. If the unit is out of warranty or the damage not covered, a partial credit will be applied based upon a prorated replacement price schedule dependent on the age of the unit. Any exchange covers only the pen tester electronics. Electrode tips are consumable and are not included under the factory warranty. 8