Standard Operating Procedure for Detecting MEA Triazine in a Crude Oil Matrix using LONESTAR TM

Similar documents
Lonestar- Tips for Aqueous Based Liquid and Powder Sample Analysis Using the v2.x Lonestar and v2.x ATLAS

DxC 700 AU Job Aid Booklet

CC PR 28 April Operational and Cleanliness verification of a Lonestar 3.0 with ATLAS Sampling Module 2.x

Autoplex Specifications

Atomic Absorption Spectroscopy

Ion PGM System with Reagent and Wash Bottles attached

MADMax Moisture and Density Interface: Quick Start Guide

VOC Duct and Rough Service Sensor

breakthrough versatility

Touch plate serial number. Please save this info here for use later:

Test Stand Preparation Data Collection

Performing ASTM 6584 free and total glycerin in BioDiesel using an SRI Gas Chromatograph and PeakSimple software

TSHR 7000 series TN-TS-TX analyzer

Dilutor Module for TriPlus RSH

ECscan40 Pocket Conductivity/TDS/Salinity Tester Instruction Manual

ECscan30 Pocket Conductivity/TDS Tester Instruction Manual

X117E. X117E Contact-Less Valve Position Transmitter INSTALLATION / OPERATION / MAINTENANCE MODEL. X117E Contact-Less Valve Position Transmitter

multi Mid. vol. Max. vol. Min. vol. < 3.0% 1) < 1.0% < 0.6% < 0.6% < ± 1.5% < ± 0.8% < ± 0.7% < ± 0.8% < ± 1.5% < ± 0.8% < ± 0.7% < 1.0% < 0.

GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585

Mitos P-Pump Accessories

Deans Switch Microfluidics

ECscan20 Pocket Conductivity Tester Instruction Manual

Quick-start Guide. 1 Enter Airlock Locate pull tab on the back of the instrument. Pull to remove it and activate the battery. Wait 15 seconds.

economics and performance Spark Holland introduces OPTIMAS The optimum balance of BETTER SAMPLE CARE

FULL AUTOMATIC ASPHALTENE ANALYZER APD-500A USER S GUIDE

Online Capacity Tester MK70 User and PC-Software Manual

Troubleshooting. CryoSmart 1 - CryoSmart 20

ECscan10H Pocket Conductivity Tester. Instruction Manual BANTE INSTRUMENTS LIMITED

Technical Procedure for the Examination of Arson Evidence

Chem(Bio) Week 10 Bomb Calorimetry of Biodiesel

Instruction of connection and programming of the VECTOR controller

Packing Procedure for 4 to 7.5 mm Liquid Samples. Sample Volumes of 46 to 450 µl Pub. No , Rev. C 0907

INSTRUCTION MANUAL. Mi490 Peroxide value. Milwaukee Lab Photometer. Test Equipment Depot Washington Street Melrose, MA 02176

TDSscan10M Pocket TDS Tester Instruction Manual

BRAKE TESTER DECELEROMETER

MET-Biodiesel Capillary GC Columns

4025 Synthesis of 2-iodopropane from 2-propanol

Workshop Automated Sample Preparation and Introduction for the Analysis of Unknowns. Ray Perkins Anatune Ltd

Callisto 100. Cold Filter Plugging Point Tester

air-cooled 18,000 rpm 25,356 xg 24,000 rpm 45,078 xg refrigerated Table-Top High-Speed Centrifuges High performance instrument for the daily routine

CLA-VAL e-drive-34. User Manual. Motorised Pilots. CLA-VAL Europe LIN072UE - 04/16

Extremely quiet due to unique flexible drive system. Super High-Speed Centrifuge. Refrigerated - 30,000 rpm / 70,434 xg

APPLICATION VAPODEST ALCOHOL IN BEVERAGES AND INTERMEDIATES

Power, Safety and Versatility for Multiple Applications. Thermo Scientific IEC CL40 and FL40 Centrifuge Series

MB3600-CH30 Laboratory FT-NIR analyzer for biodiesel applications Suitable for production optimization and product quality assessment

USV Ultra Shear Viscometer

PHscan10S Pocket ph Tester Instruction Manual

Scout II Operation Instructions

Stability, Linearity and Repeatability of Nitrogen Determination by Flash Combustion using Argon as Carrier Gas

Contents. Purpose. 1.Measurement Condition. 2.Sample. 3.Standard Requirement. 4.Contents of evaluation. 5.Result. 6.Conclusion. 7.Notices.

Starrsed RS Instrument (formerly Auto Compact)

SmarTire TPMS Maintenance Hand Tool. Revision User Manual

electronic micropipettes

Finnpipette Novus. Single Channel & Multichannel. User Manual. Instructions for Use

CALIBRATION PROCEDURE NI Channel, ±10 V, 16-Bit Analog Voltage Output Module

prepare CALIBRAtIon PHOTO MANUAL FOR BACTIQUANT -WATER HIGH VOLUME PACKAGING 3. Use a black cuvette and a standard 4. Press CAL 5.

electronic micropipettes

GC-101. Testing of Biodiesel Using Gas Chromatography. by Bob Armantrout. Presented at: Local Biodiesel: A Biodiesel Coop Conference

PHscan30S Pocket ph Meter Instruction Manual

RAPID OIL TEST GAUGE

Agilent 7693A Automated Liquid Sampler

Foss Electric Milko-Scope II Meter Manual

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

AssayMAP 96AM Wash Station. Maintenance Guide

Introduction. Figure 1: Labeled picture of the Instron 3367 load frame.

Stabinger Viscometer Series. SVM Series

DCI-11 Corrosion Inhibitor for Gasoline-Alcohol fuels

ALL INDIA INSTITUTE OF MEDICAL SCIENCES, JODHPUR ******** CORRIGENDUM. Rate Contract for Supply of Glassware and Plastics

URB '-carbamoylbiphenyl-3-yl cyclohexylcarbamate. DEA Reference Material Collection. Form Chemical Formula Molecular Weight Melting Point ( o C)

Agilent 7693A Automated Liquid Sampler

WHEATON. SOCOREX Acura electro Pipettes Your Electronic Choice. A World of Pipetting Expertise with SOCOREX.

ARE/AREX-6 Series Hot Plate Stirrers

OOCC Tank Sampling Guidance Document

Stirring Motors from CAT. Ingenieurbüro M. Zipperer GmbH. Our development makes your success. High power output up to 600 Ncm. Fuzzylogic control

TSHR 6000 series TN/TS/TX analyzer

Cannondale Diagnostic Tool Manual

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

OPERATION MANUAL MiniStirrus MAGNETIC STIRRER/HEATER VP 706B-4H

Productivity Improvements for Dual Column Applications: USP <467> and Blood Alcohol Analysis

Characterization of Tiki Torch Fuels

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

PS-20 and PS-22 Test Kits Instructions

MiR Hook. Technical Documentation

JJG Translated English of Chinese Standard: JJG

4001 Transesterification of castor oil to ricinoleic acid methyl ester

HI Ethylene Glycol Refractometer INSTRUCTION MANUAL

LUX Assure. OMMICA q. Onsite Analysis of Methanol and MEG in Produced Fluids

TORQUE CALIBRATION USING THE PROEDIT PROCEDURE WIZARD AND CALIBRATING USING PROCAL

EPAS Desktop Pro Software User Manual

Hot Injection Adapter Kit - Installation Guide

MODEL 905V OPERATING INSTRUCTIONS

Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector

MiniLab EL Series for High Performance Engine Development and Maintenance

Warning! Before continuing further, please ensure that you have NOT mounted the propellers on the MultiRotor.

SmarTire TPMS Maintenance Hand Tool. Revision User Manual

DynoTune Wideband Gauge

SmartChip MyDesign Kit User Manual

User Manual Electronic Titrator

HyperShear TM HPLC and UHPLC Mixers

Newcastle Weighing Services Pty Ltd

Transcription:

Standard Operating Procedure for Detecting MEA Triazine in a Crude Oil Matrix using LONESTAR TM Contact us Owlstone Nanotech Inc 2012 The Owlstone logo, OWLSTONE, LONESTAR, OVG-4 and OHG are trademarks of Owlstone Inc

Lonestar Method BHM1 MEA Triazine H2S Scavenger Detection in a Crude Oil Matrix Principles 5g of the retrospective drawn sample is then weighed out into a sampling vial accurately using an analytical balance. The sample is then loaded into a pre-calibrated Lonestar system which continually flushes the headspace with clean dry air. Within the sampling module (GBH) of the Lonestar the air is split and then diluted for the higher concentrations allowing a more dynamic range. The lower concentrations (0ppm 600ppm) do not contain a split as the monoethanolamine (MEA) triazine has not reached saturation. As the headspace replenishment rate is temperature dependent the sample is temperature controlled at 35C. A measurement is taken and converted on screen to an equivalent concentration. Apparatus Plastic Pasteur pipette ( Part number Fisher Scientific FB 55345) Analytical balance (calibrated) 10ml GCMS glass vial (Part Number Fisher Scientific VGA-100-070K) O-ring (Part number Owlstone 50-0510) Directions Load an empty GC glass vial into the silver insert into the bottle holder of the GBH attached to the Lonestar as pictured below in Figure 1. Fit bottle holder onto interface on the Lonestar and spin the securing nut until snug. (Please ensure the silver GBH insert has an O-Ring attached) Figure 1 Loading sample bottle The Lonestar analyser screen appears as shown in Figure 2 2

Figure 2 On the pull down menu labelled Method select the HHTT content 50-800ppm 6 mins v4a. Before any sample can be ran on the system it is essential to ensure the system is clean. This is done by clicking the Check Clean button (shown in figure 2). The button will only be selectable if the top banner is green. If it is red, one or more parameters for example system flow or pressure does not match the method file. To solve this issue either consult troubleshooting guide in the Lonestar manual or contact us directly via Zendesk. Figure 2 Lonestar analyser Screen Click the OK button to continue. The software will then check that the system is clean and ready to begin a measurement. (shown in figure 3) Figure 3 clean down sample required and acquisition screen shots If the bottle and system is clean, the light by the Check button will switch from red to green and a time stamped Cleanliness Confirmed message will appear. The system is now ready to receive a sample (it can be left in this state almost indefinitely but if the time exceeds 4 hours it is recommended that the check is rerun before taking the sample). The time to carry out a cleanliness check is approximately 2 minutes. The light will be green and the Measure button will be available to select. Unload the sampling vial by releasing the securing nut on the GBH 3

Using a plastic Pasteur pipette weigh 5g of a representative crude oil sample into the glass sampling vessel. Place sampling vial back into the GBH, drop vial into silver insert inside the GBH. Fit bottle holder onto interface on the Lonestar and spin the securing nut until snug Select the Measure button on the analyser screen (shown in figure 4) During analysis of the sample the orange banner should turn to green if all parameters are met. Again refer to the Lonestar manual or contact Zendesk if this does not happen. Figure 4 - Select measure button Analysis time is approximately 6 minutes. Upon completion the screen is updated with the concentrations and a time stamp. The cleanliness of the system will switch to Unconfirmed and the indicator will turn red waiting for another blank bottle Figure 5 Results screen Remove sample and swap in a fresh clean bottle to allow the system to blank Select the Check Clean button on the analyser screen The system is then ready for the next sample 4

Lonestar Method BCM2 Calibration check standard Principles In order to check the system functionality a test sample is created at a mid-range concentration for the application. The analyser output for the sample should be within the 10% of the expected signal (the specified accuracy for the method), if it fails then a calibration adjustment will be required. Reagents Hexahydro-1,3,5-tris (hydroxyethyl) triazine CAS No:-4719-04-4 Crude oil Apparatus 10-100µl positive displacement pipette & pipette tips GCMS glass Vials (New vial used for every sample) O-ring (Replace on silver insert once per week) Analytical balance Stirring plate Fridge to store samples Glass double ended magnetic spinner 40mm (Part number Fisher Scientific FB55628) 100ml Duran bottle with lid (Part number Fisher scientific FB 33144) Directions (200ppm check standard) This assumes that the Mea-Triazine content in the sample is 100%, additional calculations may have to be applied if the content is lower The density used in the calculations is 1.16g/ml Place magnetic stirrer into 100ml Duran bottle Using an analytical balance weigh out 100g of crude oil into 100ml Duran bottle using a plastic Pasteur pipette Using the 10-100µl positive displacement pipette, pipette 17.2µl of MEA triazine into the 100g of crude oil Place lid on top of Duran bottle & label Invert sample 3 times and place on stirring plate Set plate to 400RPM and leave for 30 minutes Load sample into Lonestar and run sample using HHTT content 50-800ppm 6 mins v4a Analyser output should be 200ppm +/-10%. If outside of this range, first repeat sampling on a fresh sample and then refer to the troubleshooting guide in Lonestar manual or via Zendesk. 5

Re-stir samples at 400RPM for 10 minutes if sample is left longer than 24hours. 6