MSD EI and CI Source Cleaning and Installation. Technical Overview. Introduction

Similar documents
GC/MS Analysis of Trace Fatty Acid Methyl Esters (FAME) in Jet Fuel Using Energy Institute Method IP585

Increased sensitivity and reproducibility in the analysis of trace fatty acid methyl esters in jet fuel

Technical Procedure for Gas Chromatography-Mass Spectrometry (GC-MS)

Agilent G3969A Transfer Line Interface for an Agilent 7697A HS Accessory

Agilent Multimode Inlet for Gas Chromatography

Application Note. Abstract. Authors. Environmental Analysis

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

Using a New Gas Phase Micro-Fluidic Deans Switch for the 2-D GC Analysis of Trace Methanol in Crude Oil by ASTM Method D7059 Application

Technical Procedure for Pyrolysis-Gas Chromatography/Mass Spectrometry (Py-GC-MS)

Achieving Higher Sensitivities Using GC-FID with the Agilent Multimode Inlet (MMI)

Agilent 6890 Gas Chromatograph

Using the PSD for Backflushing on the Agilent 8890 GC System

Nickel Catalyst Tube Accessory 19205A

Troubleshooting Tips & Tricks for your GC Analyzer & CFT Application

Fast Simulated Distillation Based on Agilent 6890N Gas Chromatograph Application

Achieving Lower Detection Limits Easily with the Agilent Multimode Inlet (MMI)

SPECIFICATION. Bench top triple quadrupole Mass Spectrometry System coupled to a gas performance Chromatography (GC-MS/MS)

SELERITY TECHNOLOGIES SOLUTIONS FOR YOUR SUPERCRITICAL FLUID NEEDS

Pyrolytic Graphite Platforms

Agilent G2647A (Single) and G2648A (Dual) Flame Photometric Detector Update Kits

Agilent 7693A Automated Liquid Sampler

Dual Channel Simulated Distillation of Carbon and Sulfur with the Agilent 7890A GC and 355 Sulfur Chemiluminescence Detector

6890 Installing ¼ Glass Columns

High Throughput Mineral Oil Analysis (Hydrocarbon Oil Index) by GC-FID Using the Agilent Low Thermal Mass (LTM) System

DETECTORS Photo Ionization Detector - PID. Overview

SIS AP2000 AutoProbe Manual

Agilent Dynamax Analytical HPLC Column Systems 4.6 mm id Data Sheet

INSTRUCTION MANUAL MP7ER/FR Remote Double Inverted Magnetron Cold Cathode Gauge Range: 7ER 10-8 to 10-2 Torr 7FR to 10-2 Torr

High-Temperature Simulated Distillation System Based on the 6890N GC Application

GCMS-QP2010 Series. Consumables. Shimadzu Gas Chromatograph/Mass Spectrometer

Agilent G3348B Flame Photometric Detector

Phenomenex Exclusive! Inlet Base Seals. Agilent 5000, 6000 and. Guaranteed fit for Series GCs. Easy Seals

260 Solvent Vapor Exit Accessory

AZ Chrom s.r.o.

Practical Steps in GC Troubleshooting

Fast and Reliable Trace Gas Analysis Improved Detection Limits for the Agilent 490 Micro GC

G1160A Agilent 1100 Series Multiple Purpose Switching Valve

Beverage Grade Carbon Dioxide

Method Development for Capillary GC Systems. Slide 1

AMS Hardware Updates, Tips and Tricks. Ed Fortner, Bill Brooks Phil Croteau, John Jayne, AMS User s Meeting /26/14

2-D GC/MS and the Deans Switch: Theory and Applications. Fred Feyerherm, Agilent Technologies Inc. Houston, Tx

Application. Gas Chromatography June 1995

Large Volume Injection of Polycyclic Aromatic Hydrocarbons

Customer Information. Service Engineer s Responsibilities. Additional Instruction Notes GC Preventive Maintenance Checklist Standard

GC Inlets. An Overview. Simon Jones GC Applications Engineer

Analysis of biodiesel oil (as per ASTM D6751 & EN 14214) using the Agilent 5100 SVDV ICP-OES

MET-Biodiesel Capillary GC Columns

LABORATORY ZERO AIR GENERATOR MODEL N-GC1500 USER S MANUAL

[ CARE AND USE MANUAL ] WATERS ADVANCED PURIFICATION COLUMNS I. INTRODUCTION CONTENTS I. INTRODUCTION II. USING THE COLUMN III. CARE AND MAINTENANCE

Installation Guide and Owner s Manual. For UTCIS-G and UTCIS-V Replacement Warm-up Regulator and Engine Management System

Spray Coater. Standard Operating Procedure. Revision: 2.0 Last Updated: Jul.24/2014, Revised by Nathanael Sieb

Technical Procedure for Gas Chromatography (GC-FID)

Agilent 5975T. Mobile Lab Hardware Installation. Agilent Technologies

Page 1. Alternate Carrier Gas Considerations and Faster GC Analysis

Analysis of Glycerin and Glycerides in Biodiesel (B100) Using ASTM D6584 and EN Application. Author. Abstract. Introduction

URB '-carbamoylbiphenyl-3-yl cyclohexylcarbamate. DEA Reference Material Collection. Form Chemical Formula Molecular Weight Melting Point ( o C)

STI OPENING THE DOORS TO TEMPERATURE PROGRAMMING- A NEW FRONTIER IN HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Agilent 1260 Infinity Manual Injector

MADMax Moisture and Density Interface: Quick Start Guide

Nickel Catalyst Tube on a 6820 GC, Accessory G4318A

BlueFin User's Manual. Version 1.1

Forensic Blood Alcohol Determination with the Intuvo 9000 GC

Agilent G2333B (Single) and G2334B (Dual) Flame Photometric Detectors

Hot Injection Adapter Kit - Installation Guide

Gas Chromatographic Analysis of Diesel Fuel Dilution for In-Service Motor Oil Using ASTM Method D7593

Determination of Free and Total Glycerin in B100 Biodiesel

Setting up SilFlow for BackFlush in your GC

TRACE DSQ TM. Hardware Manual. Xcalibur Data System, 1.4. Revision B PN Analyze Detect Measure Control TM

Dispenser Pan Sensors & Containment Sump Sensors

Determination of Phenolic Antioxidant DBPC and DBP Levels in Electrical Insulating Oil

PICO Cleaning Instructions

Productivity Improvements for Dual Column Applications: USP <467> and Blood Alcohol Analysis

From Helium to Hydrogen: GC-MS Case Study on SVOCs in Water

Alliance HPLC DEFINED BY DEPENDABILITY, TODAY AND IN THE FUTURE

Model 807C Pump Shown

TL-105-D COLUMN HEATER

LIABLE RESULTS AND RELIABLE RESULTS AND EXCEPTIONAL REPRODUCIBILITY

Installation of the 2 position/10 port valve into the 1100 series thermostatted column compartment

Installing the 6850 Direct Valve Column Connector Kit

For more information about Agilent Technologies services please visit our web site using the following URL

Evaluation of Total Petroleum Hydrocarbon in Soil Using LC with Fraction Collector and GC/MS

ASTM D for Denatured Fuel Ethanol Automating Calculations and Reports with Empower 2 Software

Maintenance. Part A, Section 5. This section covers the following unit configurations.

Technical Procedure for the Examination of Arson Evidence

Think GC/MS Couldn t Get Any Better? Think Again!

BBK Liquid-Filled Fuel Pressure Gauge (86-93) - Installation Instructions

Agilent 7696A Sample Prep WorkBench Automated Sample Preparation for the GC Analysis of Biodiesel Using Method EN14105:2011

Dispenser Pan Sensors & Containment Sump Sensors

1 of 7 12/18/2016 9:15 PM

Technical Service Bulletin IOX/DCX Transmitter Maintenance Schedule

GC Analysis of Total Fatty Acid Methyl Esters (FAME) and Methyl Linolenate in Biodiesel Using the Revised EN14103:2011 Method

Oil Free Vacuum Pumps for Mass Spectrometry

DANI Transformer Oil Gas Analyzer

Optimized Method Development of Large Volume Injection for GC/MS/MS of Food Pesticides

Meeting the Requirements of EN12916:2006 (IP391/07) Using Agilent 1200 Series HPLC Systems

Consumables and Supplies Required for Repair

Manual No: Revision: B. CCVP Rain Shield. Installation Guide

HOW TO REPLACE THE ICHEMVELOCITY WASTE FILTER AND PERFORM THE WASTE LINE FLUSH FOR BECKMAN UA CALL CENTER USE

AMS FAQ s and Hardware Tips

ATTACH VACUUM SYSTEM AND LEAK CHECK PROBE C

Transcription:

MSD EI and CI Source Cleaning and Installation Technical Overview Charles Thomson, Max Ruemler, Mickey Freed, Dave Peterson, and Harry Prest Introduction During the course of operation, eventually GC-MSD sources require cleaning. This can be indicated by a loss in analyte response that is not improved by GC inlet and column servicing; or in source tuning by poor calibrant ion peak shapes or escalating repeller or electron multiplier voltages. Consider Figure 1, which shows the difference in response for several pesticides before and after proper source cleaning. Proper cleaning, assembly, and installation are essential to robust and reliable operation, especially for the assembly of the CI ion source. A variety of source cleaning methods exist in laboratories around the world, some advocating polishing compounds, soaps, or harsh agents. These most often work for a very limited application range and never are successful over the long-term performance of the source. For example, most polishing compounds and some soaps contain high-molecular-weight waxes, which produce a raised background and lowered response. Chemical agents can adversely affect the source surface chemistry and increase activity. Based on many years of experience, this overview describes a simple approach to cleaning and reinstalling the 5973 and 5975 ion electron impact (EI) inert and standard sources and the chemical ionization (CI) source. Figure 1. Overlaid RTICs for Azinphos-ethyl (left) and the Demeton isomers (right) before (lower trace) and after (upper trace) mechanical cleaning.

Precautions Like all laboratory work involving flammable or potentially toxic solvents, the appropriate precautions should be taken. All solvents, glassware, and foil must be clean and free of contamination and exchanged as necessary during the cleaning process. Before starting a source cleaning, consumables, such as the filaments, or spare parts, such as repeller ceramics, should be on hand. Cleaning Procedure Source disassembly is described in the hardware manual and associated video (refer to Agilent 5973 and 5975 Series MSD Hardware User Information DVD). It is important that all ceramics, the entrance and ion-focus lenses insulator, the source heater block, all screws, and the filaments are Figure 2. Never expose to solvent Disassembled and separated EI source components. always placed on a clean, lintless material (like solvent-rinsed or fired foil) and are never exposed to any solvent (refer to Figure 2). Separating the metal parts for cleaning makes this easier (Figures 2 and 3). In the past, the green paper was recommended for cleaning sources. While this is still applicable to the Agilent 5971 and 5972 MSD Series, it is not the preferred method of cleaning for the newer sources of the 5973 and 5975 MSD Series. Instead, apply the following approach. 1. Make a slurry or paste of the alumina MICROGRIT supplied with the instrument by adding deionized water to a small amount of the MICROGRIT powder (refer to Figure 4). The consistency should be a little thick. 2. Use the supplied cotton-tipped swabs to apply a small amount of the MICROGRIT-water paste to the metal components. Rubbing will remove enough material to produce a clean and shiny surface. In cleaning the source bodies of the EI and CI sources, it is important to also clean the filament holes. Use a wooden toothpick to apply the MICROGRIT paste to the holes and also to remove or clear MICROGRIT from the filament holes as the last step in cleaning the source bodies. The parts most prone to contamination are the source body, repeller, and drawout lens. These should be cleaned with special attention. 3. Rinse all parts in deionized water. Remove as much MICROGRIT as possible at this stage. 4. Submerge all the rinsed parts in a beaker of deionized water and sonicate for about utes. Steps 4, 5, 6, and 7 refer to Figure 5. EI CI Figure 3. EI and CI Source metal components for cleaning. 2

5. Use metal tweezers to remove the parts carefully from the beaker of water and submerge them in a beaker of methanol (pesticide or HPLC grade). for about utes. 6. Use metal tweezers to remove the parts carefully from the beaker of methanol and submerge them in a beaker of acetone (pesticide or HPLC grade). for about utes. 7. Use metal tweezers to remove the parts carefully from the beaker of acetone and submerge them in a beaker of hexane (pesticide or HPLC grade). for about utes. 8. Use metal tweezers to remove the parts carefully from the beaker of hexane and place them on clean foil or lint-free tissue. Carefully Figure 4. Microgrit and slurry. reassemble the source as instructed in the manual. Inspect the filaments and replace them if they look worn. Be especially careful with the ceramics for the repeller to ensure that they are not cracked, which usually happens if they have been overtightened. 9. Immediately reinstall the source in the MSD. Do not bother baking the source in an oven as the source should be very dry after the acetone rinse and free of organics after the hexane rinse, which rapidly evaporates. Only residues in your solvents will remain, hence the requirement for very clean solvents. Reinstallation Procedure If the GC column is being installed, it should be properly conditioned outside the MSD first and then installed through the MSD transfer line. After extending through the transfer line, it should be trimmed and pulled back until just 1 mm extends past the tip of the transfer-line interface (Figure 6). Tighten the transfer-line ferrule until Water Methanol Acetone Hexane Removes salts Removes polar agents Removes non-polar agents Solvents must be very clean. Especially the hexane! Figure 5. Ion Source cleaning sequence. two squeaks are heard. Remember that after the transfer line has been heated, the ferrule will have to be tightened again. It should be checked for tightness consistently as this is a major source of leaks when the transfer line is temperature cycled. Another approach is to use the SilTite metal ferrules (part number 5184-3569 for 0.2- to 0.25-mm id columns). These ferrules must be carefully tightened. When properly installed, they form a stable seal that does not require constant readjustment after temperature cycling. An ion source should never be heated until it is established that there is a good vacuum. This means checking that there are no leaks and that the water has been lowered. Heating a source to operating temperatures in the presence of a leak or high-water background can activate it, essentially undoing the effort of the cleaning process. A better procedure is to pump down the system with the source, quad, and GC zone temperatures cold (i.e., set to 0 C), wait 20 minutes, and examine. 1. Does the vacuum gauge indicate the manifold pressure is coming down? (Or is the foreline pressure dropping toward operating values?) 2. Then, if pressures look good, enter the Manual Tune panel and examine the air and water values (Figure 7). If nitrogen (28 m/z) has dropped to < 25,000 counts absolute, then the system is mainly airtight. Water will drop more slowly and requires system baking. 3. Load the analytical method. The philosophy for operating in this manner is that if there is a leak, the instrument zones are still cold, so the user can troubleshoot immediately without waiting for the zones to cool. 3

Figure 6. Installed and conditioned column in MSD. Pull back column until 1 mm extends beyond the interface tip. The macro Bake.mac is called as macro "bake.mac",go <enter> from the command line. The user is prompted for the baking temperatures and the duration of the bake (refer to Figure 8). Bake.mac should be loaded or located under the \msdchem\msexe directory. This macro requires MSD ChemStation G1701DA rev D03 or higher. If the macro is not located under the directory users may obtain it from the Agilent Web site if they login as users and have a registration number for their software. More detailed instructions are given below. Analyzer Baking A useful macro allows baking of the source and quadrupole to lower water background and improve performance in tuning and acquisition. This requires that the tune parameters allow higher source temperatures and these should be set to the maximums allowed: 300 C for the ion source and 200 C for the quad (in the Manual Tune Panel or from the Source Temperature Panel) and saved (Figure 8). Automated Baking and Performance Checkout An automated approach uses the macro Bake.mac. Here the user has pumped the system down, checked that the system is leak-free, and loaded their method and this checkout sequence. Typically this Checkout.M method is used to confirm the performance of the instrument after a servicing. This macro operates in a sequence such as Figure 7. Pumpdown background checking. 4

Can not exceed TUNE file setpoints Bake temps Bake time Final Temps Figure 8. This is for manual bakeout of the source and quad. Add the file C:\MSD- CHEM\MSEXE\Bake.mac Call as macro bake.mac,go <enter>. that shown in Figure 9. Line by line, this checkout sequence: 1. Calls the bake.mac and sets the baking temperatures of the ion source and quadrupole and the hours of bake time as BAKE SourceTemp, QuadrupoleTemp, Hours. Here the figure shows the source is baked at 300 C and the quadrupole at 200 C for 8 hours. 2. Source and quadrupole return to the temperatures of the loaded method tune file and then executes an AutoTune. 3. Injects a method performance standard twice. The user can then examine the tune report for air and water and the datafiles of the performance standard to see that the instrument is operating within the required specifications of the user s analysis. Instructions for Downloading the Macros from the Agilent Web Site Go to http://www.chem.agilent.com/cag/servsup/ softdocs/ssbmain.html?anch=ms. You must Register and acquire a username and a password. You also must have a valid MSD Productivity ChemStation registration number. Then go to the MSD 1. Calls Bake Macro Bake SourceTemp, QuadTemp, Hours This example sets source to 300 and quad to 200 for 8 hours 2. Runs AUTOTune 3-4. Checks Standards using the method CHECKOUT.M Figure 9. Pumpdown sequence: Checkout.S. 5

www.agilent.com/chem For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc. 2006 Printed in the USA December 11, 2006 5989-5974EN