Signet 2760 DryLoc ph/orp Preamplifier

Similar documents
* * Rev. 9 01/18 Product Manual

SIGNET 2760 DryLoc ph/orp Preamplifier

Signet 2760 DryLoc ph/orp Preamplifiers & Connectors

SIGNET 2750 ph/orp Sensor ENGLISH

Signet ph/orp Electrodes

Signet ph/orp Electrodes

Signet ph/orp Electrodes

Signet DryLoc ph/orp Electrodes

Signet DryLoc ph/orp Electrodes

Signet 3719 ph/orp Wet-Tap

Signet 3719 ph/orp Wet-Tap Assembly

* * Signet 2450 Pressure Sensor. Product Manual. English. Description. Dimensions Rev /18

Signet ph/orp Transmitter

Signet Conductivity/Resistivity Electrodes

Signet Conductivity/Resistivity Electrodes

SIGNET 3719 ph/orp Wet-Tap

Signet 2517 Brass Paddlewheel Flow Sensor * *

Pipe fi ttings MUST be installed by a certifi ed welder only. Signet will not assume liability of any kind for improper fi tting installations.

SIGNET Temperature Transmitter

Signet Temperature Transmitter

Pipe fi ttings MUST be installed by a certifi ed welder only. Signet will not assume liability of any kind for improper fi tting installations.

Signet 515/2536 Rotor-X Flow Sensors * * Operating Instructions. English. Description Rev. X 05/17

Signet V(D) to V(D) PVDF Conductivity Electrodes

Signet Pressure Transmitter

Signet 2850 Conductivity/Resistivity Sensor Electronics and Integral Systems with PVDF Sensor

Signet 2552 Metal Magmeter

Signet 2552 Metal Magmeter

Signet 7000/7001 Vortex Flow Sensors * *

User s Guide Shop online at omega.com SM For latest product manuals: CDE-285X SERIES Conductivity Sensors

SIGNET 7002 Vortex Flow Sensor ENGLISH

Signet 515 Rotor-X Paddlewheel Flow Sensors

Signet 515 Rotor-X Paddlewheel Flow Sensors

Signet 4150 Turbidimeter

Signet Pressure Transmitter

Signet 2536 Rotor-X Paddlewheel Flow Sensors

* * Rev. F 03/16 Operating Instructions

* * Rev M 10/14 Operating Instructions

Pipe fittings MUST be installed by a certified welder only. +GF+ SIGNET will not assume liability of any kind for improper fitting installations.

Signet 2552 Magmeter. English * * Specifications. Safe to operate Hot-tap

Measurement and Instrumentation

SIGNET Vortex Flow Sensors ENGLISH

OPERATING INSTRUCTION MANUAL

NEW QUICK CABLE-TO-SENSOR RELEASE,

General Purpose ph/orp Sensors

ph/orp Sensor Model 399 ph/orp

General Purpose ph/orp Sensors

Encapsulated Differential ph and ORP Sensors (available in LCP, foreground, or Ryton body materials)

* * Rev /18 Operating Instructions

FMK-2536, FP-5100, FP-5300,FP-5600, FP-8500, FP-8500A

Instrument Enclosure Assemblies

User s Guide CDCE-90-X SERIES. Conductivity Sensors. Shop online at omega.com SM

GF Signet. Measurement & Control Sensors. Signet Price List 1st Aug 2011 Rev 0.0

Models 396/396P/396R. Retraction/Submersion/Insertion ph/orp Sensor

Chapter Page. 1 Introduction Description 1

+GF+ SIGNET Conductivity Transmitter. Instruction Manual GEORGE FISCHER +GF+

DPU90-WM, FP90UM, FP90IM, PHTX-27G, PHTX-275G, FPM-5000-LTCK, FPM-5000-MB

Accessories Guide ph/orp/conductivity Sensors

OPERATING INSTRUCTION MANUAL MS SERIES CONDUCTIVITY CELLS

Signet 2551 Magmeter Flow Sensor

8000 Series ph/orp Electrode Platform Product Instructions

User s Guide FMG-550 SERIES OMEGAMAG. Shop online at omega.com SM For latest product manuals:

7777 Immersion/In-Line Mounting for Meredian II ph/orp Electrodes Installation and Maintenance Manual

User s Guide FP-5060 SERIES. Micro-Flow Sensors. Shop online at omega.com SM For latest product manuals:

User s Guide FP-2540/FP Flow Sensors. Shop online at omega.com SM For latest product manuals:

Measurement & Instrumentation 2005 Product Catalog on CD. Creating Quality by Connecting Technology to Life. Click on image to begin.

2-ELECTRODE CONDUCTIVITY SENSORS

Insertion/Submersion Flow Through Sensor

AD592/592-10K Non-Room Sensors in BAPI-Box, BAPI-Box 2 or BAPI-Box 4 BA/592-x-(BB, BB2, BB4) Temperature Sensor Instructions & Operation

7777D In-Line Mounting for Durafet II ph Electrodes Installation and Maintenance

ABB MEASUREMENT & ANALYTICS DATA SHEET. TBX5 ph, Redox (ORP) sensors with diagnostics

Analytical Instruments Two-Electrode Conductivity Sensors

Signet Measurement and Instrumentation

User s Guide FP-319. Flow Wet-Tap Assembly. Shop online at omega.com SM For latest product manuals:

Hydro-Flow 2200 Fixed Insertion Vortex Flow Meter

+GF+ SIGNET 2550 Insertion Magmeter

User s Guide. FMG-3000 SERIES Blind Version. Magmeter. Shop online at ISO 9001 ISO 9002

Retractable ph/orp Sensors

Automation Made Easy 2013 Signet Measurement and Control Product Catalog

OPERATING INSTRUCTION MANUAL

Signet Measurement and Instrumentation

Combination ph/redox (ORP) Sensors

User s Guide PH-2720-PA. Preamplifier. Shop online at.

Combination ph/redox (ORP) Sensors

7777 Immersion/In-Line Mounting for Durafet II ph Electrodes Installation and Maintenance

2100 Turbine Flow Sensor

ACTIVE ELECTRODELESS SENSORS Instruction Manual

AD592/592-10K Non-Room Units with a Junction Box, Weatherproof (WP) or Weather Tight (EU) Enclosure

Conductivity Sensors. Model 140, 141, 142. Product Data Sheet series/rev.a June 1998

TUpH Retraction/Submersion/Insertion ph/orp Sensors

Temperature Sensors. Description. Specification

For additional information, please visit our website at SENSOR SPECIFICATIONS

AP300 Series Combination ph/redox (ORP) Sensors. Measurement made easy. A reliable economic solution for a wide range of applications

CombiTemp Temperature Measuring System

Retractable ph/orp Sensors

CPM280. Technical Information. ph/orp measurement Battery-driven pocket meter for ph and ORP measurement

Table of Contents 1. OVERVIEW OPERATION FUNCTION SPECIFICATIONS INSTRUMENT LAYOUT INSTALLATION...

Model 7777DVP Immersion/In-Line Mounting for Durafet III ph Electrodes Installation and Maintenance Rev. 1 3/04

LTX RF LEVEL SENSOR. Instruction Manual

ELECTRICAL INSTALLATION

General Purpose Conductivity Sensors ENDURANCE. Model 400 and 400VP Series. Product Data Sheet PDS /rev.K February 2013

Transcription:

Signet 2760 DryLoc ph/orp Preamplifier *3-2760.090* 3-2760.090 Rev. H 05/11 English English SAFETY INSTRUCTIONS 1. Depressurize and vent system prior to installation or removal. 2. Confi rm chemical compatibility before use. 3. Do not exceed maximum temperature/pressure specifi cations. 4. Wear safety goggles or faceshield during installation/service. 5. Do not alter product construction. 6. When using chemicals or solvents care should be taken and appropriate eye, face, hand, body, and/or respiratory protection should be used. Description The Signet 2760 ph/orp Preamplifi ers with DryLoc connectors are part of Signet's growing line of sensors for industrial fluid measurement and control. There are two styles of Preampliers, and each is available with NPT or ISO threads at the cable outlet. They all connect to four different families of ph and ORP electrodes to supply an amplifi ed mv (analog) signal to Signet instruments. The 2760 family also offers a set of connectors to link the same electrodes to third-party equipment that require a direct millivolt signal for ph measurement without amplifi cation. Technical Data General Compatible Electrodes: Signet DryLoc Electrodes: 2724-2726, 2756-WTx - 2757-WTx, 2764-2767, 2774-2777 Compatible Instruments: Preamplifi er models to Signet 5700, 8750: 2760-1, -2, -11, -21 Connector models to non-signet products: 2760-3, -4, -31, -41 Response Time (includes response time of electrodes): ph: < 6 sec. for 95% of change ORP: application dependent Temperature, τ: 20 sec. Electrical (2760-1, -2,-11, -21 Preamplifiers) Maximum Offset: ± 7 mv Input Impedance, Z: > 10 11 Ω Supply Voltage (V+ to V ): 6 to 12 VDC Current consumption: < 1 ma Environmental Enclosure Rating: 2760-1, -2, -3, -4: NEMA 6P/IP68 2760-11, -21, -31, -41: NEMA 4 with electrode and watertight conduit and/or extension pipe connected Materials: Submersible versions: CPVC (Gray) In-Line versions: PBT (Yellow) Dimensions 3-2760-1 Preamplifi er, Sub, NPT 3-2760-2 Preamplifi er, Sub, ISO 3-2760-3 Connector, Sub, NPT 3-2760-4 Connector, Sub, ISO 3-2760-11 Preamplifi er, In-line, NPT 3-2760-21 Preamplifi er, In-line, ISO 3-2760-31 Connector, In-line, NPT 3-2760-41 Connector, In-line, ISO 31.75 mm (1.25 in.) 70.1 mm (2.76 in.) 101 mm (4.0 in.) 56.6 mm (2.23 in.) 35.0 mm (1.38 in.) 89.4 mm (3.52 in.) Cable: Maximum cable: Shipping Weight: Operating Temperature: Storage Temperature: Relative Humidity: 4.6 m (15 ft) 6-conductor w/shield, 22 AWG 122 m (400 ft) 0.64 kg (1.41 lb) 0 C to 85 C (32 F to 185 F) 20 C to 85 C ( 4 F to 185 F) 0 to 95%, non-condensing (without electrode connected) Standards and Approvals Manufactured under ISO 9001 and ISO 14001 China RoHS (Go to www.gfsignet.com for details) 61.0 mm (2.40 in.) CHEMICAL COMPATIBILITY WARNING The retaining nuts of certain ph and ORP sensors and preamplifiers are not designed for prolonged contact with aggressive substances. Strong acids, caustic substances and solvents or their vapor may lead to failure of the retaining nut, ejection of the sensor and loss of the process fl uid with possibly serious consequences, such as damage to equipment and serious personal injury. Retaining nuts that may have been in contact with such substances, e.g. due to leakage or spilling, must be replaced.

In-line Assembly and Installation 2760 In-Line models used with 2724-2726 series electrodes: 1. The 2724-2726 can use a Signet installation fi tting. Electrodes are sealed by the O-ring when it is installed into a Signet installation fi tting. Lubricate O-rings 2. The lower threaded nut is required to attach the electrode to the installation fi tting when using a Signet fi tting. 3 HAND-TIGHTEN THE THREADED NUT ONTO THE INSTALLATION FITTING. DO NOT USE TOOLS! DO NOT USE THREAD SEALANT OR LUBRICANTS ON THE FITTING THREADS OR THE SENSOR CAP. Unlock 100 mm (4 in.) Lock 3. Lubricate O-rings with a non-petroleum based, viscous lubricant (grease) compatible with the system. 4. Seat the 2760 DryLoc connector onto the top of the electrode. 1 2 Open the DryLoc connector by turning the upper locking ring ¼-turn clockwise. 5. Turn the locking ring ¼-turn counter-clockwise to lock the preamplifi er in place. The mechanism will click when it is locked. When fully assembled the 2760 will stand about 100 mm (4 in.) above the installation fi tting. 4 5 2760 In-Line models used with 2724-2726, 2764-2767 or 2774-2777 series electrodes: BEFORE YOU START: 2764-2767: Install a 1 in. NPT pipe tee into the process line. 2774-2777 and 2724-2726: Install a ¾-in. reducing tee into the process line. 1. Wrap the process threads on the electrode with an appropriate sealant (example: PTFE tape ) to create a water-tight connection. HAND-TIGHTEN the electrode into the installation fi tting. DO NOT USE TOOLS! NOTE: The lower threaded nut may be dropped onto the top of the electrode or it may be discarded. It is not used with these electrodes. 2. Lubricate O-rings with a non-petroleum based, viscous lubricant (grease) compatible with the system. 3. Seat the 2760 DryLoc connector onto the top of the electrode. Open the DryLoc connector by turning the upper locking ring ¼-turn clockwise. Lubricate O-rings 2 4. Turn the locking ring ¼-turn counter-clockwise to lock the preamplifi er in place. The mechanism will click when it is locked. When the 2760 Preamplifi er is fully assembled with one of the threaded electrodes (276X and 277X models) it will stand approximately 18 cm (7 in.) from the process threads to the top of the preamplifi er. Unlock 178 mm (7 in.) 1 3 4 Lock 2760 Submersible models used as in-line system: The Submersible 2760 design can also be used in-line when using the 2724-2764, 2764-2767 series or the 2774-2777 series electrodes. 2760 Preamplifier BEFORE YOU START: 2764-2767: Install a 1 in. NPT pipe tee in the process line. Threaded nut 2774-2777 and 2724-2726: Install a ¾-in. reducing tee in the 1 process line. DO NOT install in a ¾-inch tee or probe will be damaged. 1. Wrap the process threads on the electrode with an Lubricate O-ring appropriate sealant (example: PTFE tape) to create a watertight connection. Electrode 2 3 4 5 6 2. Lubricate O-rings with a non-petroleum based, viscous lubricant (grease) compatible with the system. 3. Remove the threaded nut from the preamplifi er and place it around the electrode from the bottom as illustrated. 4. HAND-TIGHTEN the electrode into the installation fitting. DO NOT USE TOOLS! 5. Seat the 2760 DryLoc connector onto the top of the electrode. 6. Lift the threaded nut up and thread it hand-tight onto the preamplifi er. When the 2760 Preamplifi er is fully assembled with one of the threaded electrodes (276X and 277X models) it will stand approximately 18 cm (7 in.) from the process threads to the top of the preamplifi er. 178 mm 7 in. The 2760 Preamplifi er signal can be routed to a Signet 5700 ProPoint ph Monitor, or to any model of the 3-8750 ph Transmitter. If the preamplifi er cable must be extended beyond the standard 15 ft that is supplied with the 2760, use a junction box to terminate the wires. Note: Do not extend the 2760 cable more than 120 m (400 ft). 2 Signet 2760 DryLoc ph/orp Preamplifi er

RLA 1 Wet-Tap Assembly and Installation The 2760 DryLoc Preamplifi er can be used to connect Signet 3719 Wet-Tap electrodes to a ph or ORP instrument. 1. Select the process fitting. Select a low profi le installation saddle (customer-supplied) with a 1½ in. or 2 in. outlet (depending on 3719 model ordered). The 3719 is available with NPT or ISO thread options. The 3719 can also be adapted to fi t into tee fi ttings. Refer to the 3719 ph Wet-Tap instructions for detailed information. 2. Install the Wet-Tap and the electrode. Install the 3719 into the fi tting, then insert the electrode into the assembly. 3. Attach the preamplifier to the electrode. Attach the 3-2760-11 or -21 Preamplifi er to the DryLoc fitting at the top of the electrode. 4. Connect the 2760 Preamplifier to the instrument. The 2760 Preamplifi er signal can be routed to any 5700 or 8750 ph/orp instrument (see Pg. 5). 1. 2. 3. 4. Lubricate O-rings 2 Lubricate O-rings 3 Signet 2760 DryLoc ph/orp Preamplifi er

2760 Submersible Sensor Assembly and Installation The submersible 2760 can be installed in a tank or vessel if conditions are within pressure and temperature specifi cations. The 2760 is able to withstand fl uid temperatures up to 85 C, but the electrode tolerances vary. Check the specifi cations of each component before installation. 1. Insert any DryLoc electrode into base of the 2760. Lubricate O-rings with a non-petroleum based, viscous lubricant (grease) compatible with the system. 2. Thread retaining cap over electrode and hand-tighten onto 2760. Keep the DryLoc electrical interconnection between electrode and preamplifi er dry and clean at all times. 3. Attach 3 /4 in. watertight pipe to the top of the 2760. Secure the threaded connection to prevent any leakage. Fill the lower 75-100 mm (3 to 4 in.) of conduit or extension pipe with a fl exible sealant such as silicone to prevent liquid accumulation on the back seal of the 2760. 3 1 Fill with 75-100 mm (3 to 4 in.) of sealant 2 Lubricate O-ring 55 mm/ 2.16 in. 55 mm/ 2.16 in. Use standard installation hardware to connect the submersible 2760 directly to external equipment. OR The 8052-1 NPT Mount Junction box connects to 3 /4 in. pipe or conduit and provides convenient wiring termination. OR The 8050-1 Universal Mount junction box mounts fl at onto a wall or can be strapped to a post or pipe. ½ in. NPT threads Fill with 3 to 4 in. of sealant Fill with 3 to 4 in. of sealant Fill with 3 to 4 in. of sealant Signet 2760 DryLoc ph/orp Preamplifi er 4

Wiring to Signet Instruments Any of these electrodes can be connected to the Signet ProPoint and ProcessPro ph instruments using the 3-2760-1(X) or -2(X) Preamplifi ers. Wiring diagrams illustrate the actual terminals on these products for convenience. Signet 8750-X ProcessPro ph/orp Transmitter RLA 1 2 Signet 5700 ProPoint ph/orp Monitor Standard output, Signet installation fittings 3-2724-10, -11 3-2725-60, -61 3-2726-10, -11 3-2726-HF-10, -11 3-2726-LC-10, -11 3-2756-WT-1 3-2756-WTP-1 3-2757-WT 3-2757-WTP Standard output, ¾ in. process connector. 3-2774 3-2775 3-2776 3-2777 3-5700 ph/orp Monitor Differential output, 1 in. process connector. 3-2764-1 3-2765-1 3-2766-1 3-2767-1 mv Input Brown wire Iso. Gnd Blue wire T + Green wire T - White wire V + Red Wire V - Black wire Brown mv Blue Iso. Gnd Green T+ White T- Red V+ Black V- 3-2760-1 3-2760-2 3-2760-11 3-2760-21 + - 4-20 ma ProcessPro ph/orp Transmitter 3-8750-3 3-8750-2 3-8750-1 16 15 14 13 12 11 10 9 18 17 16 15 14 13 12 11 12 11 10 9 8 7 6 5 White (Temp) Green (Temp) Blue (Iso. Gnd) Brown (mv input) Shield Black (V-) Red (V+) 3-2760-1 3-2760-2 3-2760-11 3-2760-21 Connect this terminal to Earth GND to reduce electrical interference. 5 Signet 2760 DryLoc ph/orp Preamplifi er

Wiring from 2760 DryLoc Preamplifier to Other Manufacturer's Equipment The 2760-1/-2 Preamplifi er and electrode options may be used with third party equipment if the equipment can process the AMPLIFIED mv signal for ph and is compatible with one of the temperature compensation options. (3 K, PT1000, or 300 ). Refer to the manufacturer's manual for wiring details. Other products may use different terms to identify the function of each conductor. Signet Combination ph Electrodes 3-2724-10 (3 KΩ, NPT) 3-2724-11 (3 KΩ, ISO) 3-2726-10 (3 KΩ, NPT) 3-2726-11 (3 KΩ, ISO) 3-2726-HF-10 (HF Resist, 3 KΩ, NPT) 3-2726-HF-11 (HF Resist, 3 KΩ, ISO) 3-2726-LC-10 (Low Cond, 3 KΩ, NPT) 3-2726-LC-11 (Low Cond, 3 KΩ, ISO) 3-2774 (3 KΩ TC) 3-2776 (3 KΩ TC) Combination ORP (Redox) Electrodes 3-2725-60 (10 KΩ ID, NPT) 3-2725-61 (10 KΩ ID, ISO) 3-2757-WT (10 KΩ ID) 3-2775 (10 KΩ ID) 3-2775-1 (no Temp Comp) 3-2777 (10 KΩ ID) 3-2777-1 (no Temp Comp) DryLoc Preamplifier, Submersible 3-2760-1 3-2760-2 DryLoc Preamplifier, In-Line 3-2760-11 3-2760-21 White (Temperature) Green (Temperature) Blue (mv Reference) Shield (Earth Ground) Black (V-) Red (V+) Brown (Active mv, ph/orp) Differential ph Electrodes 3-2764-1 (3 KΩ RTD) 3-2764-3 (300 Ω RTD) 3-2766-1 (3 KΩ RTD) 3-2766-3 (300 Ω RTD) Differential ORP (Redox) Electrodes 3-2765-1 (10 KΩ ID) 3-2765-3 (300 Ω RTD) 3-2767-1 (10 KΩ ID) 3-2767-3 (300 Ω RTD) Signet 2760 DryLoc ph/orp Preamplifi er 6

Wiring from 2760 DryLoc Connectors to Other Manufacturer's Equipment The 2760-3/4 Connectors can be used to interface other manufacturer's instruments to these Signet DryLoc ph and ORP electrodes. The instrument must be able to process a UNAMPLIFIED mv signal from the ph or ORP electrode. Several different temperature devices are available. Refer to the manufacturer's manual for wiring details. Other products may use different terms to identify the function of each conductor. Signet Combination ph Electrodes 3-2724-10 (3 KΩ, NPT) 3-2724-11 (3 KΩ, ISO) 3-2726-10 (3 KΩ, NPT) 3-2726-11 (3 KΩ, ISO) 3-2726-HF-10 (HF Resist, 3 KΩ, NPT) 3-2726-HF-11 (HF Resist, 3 KΩ, ISO) 3-2726-LC-10 (Low Cond, 3 KΩ, NPT) 3-2726-LC-11 (Low Cond, 3 KΩ, ISO) White RTD (Temperature) Green RTD (Temperature) Blue Solution Ground Shield Earth Ground Black Electrode Reference White RTD (Temperature) Green RTD (Temperature) Blue Solution Ground Shield Earth Ground Black Electrode Reference Red (Active mv, ph/orp) Brown (no connection) To any instrument with internal amplifi er and matching temp device. 3-2774 (3 KΩ TC) Red (Active mv, ph/orp) Brown (no connection) 3-2776 (3 KΩ TC) Combination ORP (Redox) Electrodes 3-2725-60 (10 KΩ ID, NPT) 3-2725-61 (10 KΩ ID, ISO) 3-2757-WT (10 KΩ ID) 3-2775 (10 KΩ ID) 3-2775-1 (no Temp Comp) 3-2777 (10 KΩ ID) 3-2777-1 (no Temp Comp) DryLoc Connector, Submersible 3-2760-3 3-2760-4 DryLoc Connector, In-Line 3-2760-31 3-2760-41 Signet 2760 DryLoc ph/orp Preamplifi er 7

Ordering Information Mfr. Part No. Code Description 3-2760-1 159 000 939 Submersible Preamplifi er with ¾ in. NPT threads and 4.6 m (15 ft) cable 3-2760-2 159 000 940 Submersible Preamplifi er with ¾ in. ISO 7/1R threads and 4.6 m (15 ft) cable 3-2760-3 159 000 941 Submersible Connector with 4.6 m (15 ft) cable and ¾ in. NPT threads 3-2760-4 159 000 942 Submersible Connector with 4.6 m (15 ft) cable and ISO 7/1R ¾ in. threads 3-2760-11 159 001 367 In-line Preamplifi er with ¾ in. NPT threads and 4.6 m (15 ft) cable 3-2760-21 159 001 368 In-line Preamplifi er with ¾ in. ISO 7/1R threads and 4.6 m (15 ft) cable 3-2760-31 159 001 369 In-line Connector with 4.6 m (15 ft) cable and ¾ in. NPT threads 3-2760-41 159 001 370 In-line Connector with 4.6 m (15 ft) cable and ISO 7/1R ¾ thread Accessories and Replacement Parts Calibration Tools 3-2759 159 000 762 ph/orp System Tester (adapter cable sold separately) 3-2759.391 159 000 764 2759 Adapter Cable for use w/2760 DryLoc Preamplifi er 3-0700.390 198 864 403 ph Buffer Kit (1 each 4, 7, 10 ph buffer in powder form, makes 50 ml) 5523-0624 159 000 636 Cable, 6 conductor, 24 AWG shielded 3822-7004 159 001 581 ph 4.01 buffer solution, 1 pint (473 ml) bottle 3822-7007 159 001 582 ph 7.00 buffer solution, 1 pint (473 ml) bottle 3822-7010 159 001 583 ph 10.00 buffer solution, 1 pint (473 ml) bottle 3822-7115 159 001 606 20 gram bottle Quinhydrone for ORP calibration 3-2700.395 159 001 605 Calibration kit: included 3 Polypropylene cups, box as cup stand, 1 pint ph 4.01, 1 pint ph 7.00 Special Request NIST Traceable Certifi cate Electrodes DryLoc ph Electrodes 3-2724-10 159 001 547 Combo fl at ph Electrode, 3 KΩ RTD, ¾ in. NPT or Signet fi ttings 3-2724-11 159 001 548 Combo fl at ph Electrode, 3 KΩ RTD, ISO 7/1 R¾ or Signet fi ttings 3-2726-10 159 001 555 Combo bulb ph Electrode, 3 KΩ RTD, ¾ in. NPT or Signet fi ttings 3-2726-11 159 001 556 Combo bulb ph Electrode, 3 KΩ RTD, ISO 7/1 R¾ or Signet fi ttings 3-2726-HF-10 159 001 551 Combo bulb ph Electrode, 3 KΩ RTD, HF resistant, ¾ in. NPT or Signet fi ttings 3-2726-HF-11 159 001 552 Combo bulb ph Electrode, 3 KΩ RTD, HF resistant, ISO 7/1 R¾ or Signet fi ttings 3-2726-LC-10 159 001 559 Combo bulb ph Electrode, 3 KΩ RTD, Low conductivity, ¾ in. NPT or Signet fittings 3-2726-LC-11 159 001 560 Combo bulb ph Electrode, 3 KΩ RTD, Low conduct., ISO 7/1 R¾ or Signet fi ttings 3-2764-1 159 000 943 Differential ph Electrode, 3 KΩ RTD, 1 in. NPT process connection 3-2764-3 159 000 945 Differential ph Electrode, 300 Ω RTD, 1 in. NPT process connection 3-2766-1 159 000 949 Differential ph Electrode, 3 KΩ RTD, 1 in. NPT process connection 3-2766-3 159 000 951 Differential ph Electrode, 300 Ω RTD, 1 in. NPT process connection 3-2774 159 000 955 Combo fl at ph Electrode, 3 KΩ RTD, ¾ in. NPT process connection 3-2776 159 000 959 Combo fl at ph Electrode, 3 KΩ RTD, ¾ in. NPT process connection DryLoc ORP (Redox) Electrodes 3-2725-60 159 001 561 Flat ORP Electrode, 10KΩ ID, ¾ in. NPT or Signet installation fi ttings 3-2725-61 159 001 562 Flat ORP Electrode, 10KΩ ID, ¾ in. ISO 7/1 R¾ or Signet installation fi ttings 3-2765-1 159 000 946 Differential ORP (Redox) Electrode, 10 KΩ ID, 1 in. NPT process connection 3-2765-3 159 000 948 Differential ORP (Redox) Electrode, 300 Ω RTD, 1 in. NPT process connection 3-2767-1 159 000 952 Differential ORP (Redox) Electrode, 10 KΩ ID, 1 in. NPT process connection 3-2767-3 159 000 954 Differential ORP (Redox) Electrode, 300 Ω TC, 1 in. NPT process connection 3-2775 159 000 957 Combo fl at ORP (Redox) Electrode, 10 KΩ ID, ¾ in. NPT process connection 3-2775-1 159 000 958 Combo fl at ORP (Redox) Electrode, no Temp, ¾ in. NPT process connection 3-2777 159 000 961 Combo bulb ORP (Redox) Electrode, 10 KΩ ID, ¾ in. NPT process connection 3-2777-1 159 000 962 Combo bulb ORP (Redox) Electrode, no Temp, ¾ in. NPT process connection Georg Fischer Signet LLC, 3401 Aerojet Avenue, El Monte, CA 91731-2882 U.S.A. Tel. (626) 571-2770 Fax (626) 573-2057 For Worldwide Sales and Service, visit our website: www.gfsignet.com Or call (in the U.S.): (800) 854-4090 For the most up-to-date information, please refer to our website at www.gfsignet.com 3-2760.090 Rev. H 05/11 English Georg Fischer Signet LLC 2011