***Before starting : view the Virtual tour located at c:/programs/msoffice/powerpointviewer And carry out the tutorials in the manuals supplied***

Force Test Stands Series ES MODELS ESM & ES30. User s Guide

Model 635.XX Axial Extensometers

Model ESMH MOTORIZED FORCE MEASUREMENT TEST STAND. User s Guide

The FMM force tester can be controlled using L1 software for limit, cycling, hold and coefficient of friction testing.

Twin-Column Motorized Tension, Compression & Cycle Tester (M5KNE & M10KNE) User s Manual

C224-E031A. Autograph AGS-J. Shimadzu Table-top Type Universal Tester

Introduction. Installing the TC 2. Unpacking the TC 2

LD Series Operation Manual Dual Column Test Stand. Models LD5, LD10, LD30, LD50, LD100

Model 1100B CHG Terminator. Installation Instructions

2. The engine must be in racing conditions, no combustion chamber cleaning allowed

ESM1500 / ESM750. Data Sheet. Motorized Force Test Stands. Standard Features. Page 1 of 6

Standard G50 90 degree Table with cable to upper machine crosshead and G227 upper grips. G Width 130 mm (5.1 )

ESM1500 / ESM750. Data Sheet. Motorized Force Test Stands. Standard Features. Page 1 of 6

ADHESION/RELEASE TESTER

LTCM-500 Series High Capacity Motorized Tester (550 lbf, 2500N)

Inspekt. Static electromechanical universal testing machines. Competency generates results

UNIMAT BASIC. Material Testing Machine. for the. Quality Control. with. Software (intern) for Stand Alone Usage

Instron Series Clip-On Extensometers

Standard G50 90 degree Table with cable to upper machine crosshead and G227 upper grips

Concrete ASTM COMPRESSION TESTER ACCESSORIES

IMPORTANT! DO NOT THROW AWAY THE SHIPPING CARTON AND PACKING MATERIAL

AutoStream - Model HV. Manual

Static Servohydraulic Universal Testing Machines

Last date for sending comments : 30 November 2011

Experiment P-16 Basic Electromagnetism

Compression/Tension Testing Machines

MPT-200A SPECIFICATIONS AND OPERATING INSTRUCTIONS

KEMKRAFT ENGINEERING, INC. MODEL KEI-234 STEERING WHEEL TORQUE/INCLINOMETER TESTER

Conductor Strength Tests on ACCR 1622TW T13 Pecos Conductor

ELECTRONIC TRACTION CONTROL USER MANUAL

Impact, Torsion, and Crush Tests for 477 kcmil and 795 kcmil 3M Brand Composite Conductor. 3M Company Purchase Order

Model 2500 Horsepower Computer System User Manual

Pendulum Impact Testing Machine PIT452G 450J

CAUTION. Risk of electrical shock. Do not remove cover. No user serviceable parts inside. Refer servicing to qualified service personnel.

What s Included. All items included. Please be familiar with the names of all the accessories listed here. NOTES:

User Manual. Aarhus University School of Engineering. Windtunnel Balance

Part No. NC July LTCM Series Motorized Force Testers. Operator s Manual. LTCM-500 Series. LTCM-100 Series

MANUAL TRANSMISSION CONTENTS

DURHAM GEO SLOPE INDICATOR

UNIVERSAL TESTING MACHINERY STENTOR II CC - ATLAS II CC - T-DRIVE CC

Rotary Motion Sensor

OPERATOR'S MANUAL AND MAINTENANCE INFORMATION MODEL T-2000 TEXTURE TEST SYSTEM

Zwick. Materials testing. Product Information LTM 5/10 electro-dynamic testing machine

AR2000 Rheometer: Instructions

TBI /2012 TRAUMATIC BRAIN INJURY DEVICE

Pet Bicycle Trailer/Jogger. Owner s Manual & Instructions

Universal Testing Machine

Altec LMAP. (Load Moment and Area Protection) Telescopic Boom Cranes. Calibration

Page 1 of 5. United Testing Systems, Inc S. Acacia Avenue Fullerton, CA William Warner, Quality Manager

MTS Exceed. Series 40 Electromechanical Universal Test Systems. Affordable monotonic testing solutions. be certain.

MICROGUARD 500 EXTENSION REEL TRAINING MANUAL. Greer Company. Greer Company Crane Systems 1 OF18

SANDNER Messtechnik GmbH

SCHMIDT ManualPress 300 Series

Emergency Power Supply for the OMEGA ACTIVE-System. Operating Instructions

Impulse, Momentum, and Energy Procedure

ZT-1 Tonnage Load Module By Toledo Integrated Systems User s Manual

ASSEMBLY, INSTALLATION, AND REMOVAL OF CONTACTS AND MODULES

2006 Manual. Manual Version: R2.1 Last Update:

Autograph AG-X plus Series

Lab 4 Constant Acceleration by Drew Von Maluski

PORTABLE POWER SUPPLY MODEL NUMBER VMS-1 VERSION 1.0

RESISTIVITY MONITOR/CONTROLLERS

Redox Potentials and the Lead Acid Cell Minneapolis Community and Tech. College v I. Introduction. Part I

Installation Guide. Enhanced Easy-Stop Trailer ABS 2S/2M Standard with PLC Installation Instructions TP Revised 1 Technical Bulletin

Mechanical Testing Solutions for Lithium-Ion batteries in Automotive applications

Development and Analysis of Rapid Prototype Parts for Classroom Applications

ASSEMBLY, INSTALLATION, AND REMOVAL OF CONTACTS AND MODULES

_ ANCHOR LOAD CELLS USER MANUAL

Automotive Application ET01 Software Revision A 12/06

UNIVERSAL TESTING MACHINES STENTOR II CC & ATLAS II CC

SCHMIDT ManualPress. From 1.6 kn to 22 kn

Servohydraulic Testing Machines for Dynamic Testing UFIB Series

Installation Instructions BILLET/PLUS PISTOL-GRIP SHIFTER Fits: 2016 Camaro V6 with Manual Transmission Catalog #

EXPERIMENT 11: FARADAY S LAW OF INDUCTION

Operating Guide. Automated tensile test stand w/ ComTouch Total Control System. User s Manual for: Series 95 TS Series 95 TM Series 95 TL

Model B STRAIN GAGE LOAD CELL. Installation and Operating Manual

INSTALLATION INSTRUCTIONS

Installation Instructions for the EVO3 Height-Adjustable Ultimate Short Shifter

SCHMIDT ManualPress From 1.6 kn to 22 kn / 360 lbs. to 4,945 lbs.

Setting Up an Oscillation Amplitude Sweep Test

SAUTER TVO-S Version /2017 GB

AR2000ex Rheometer Standard Operating Procedures Revised: 5/1/2018

RETRIEVABLE BOREHOLE EXTENSOMETER. Model BOF-EX. Roctest Limited, All rights reserved.

GASENGINES V2 CLASSIC ASSEMBLY PLAN

Torque Module Installation and User Guide for model 250i Motorcycle Dynamometers.

Operating Instructions

Zwick. Materials testing. Product Information Temperature chambers for AllroundLine

INSTALLATION INSTRUCTIONS

Transcription:

OperatingInstructionsforInstronModel4400 Series LoadFrames John D. Williams and David E. Farrow SAFETY FIRST TheInstron 4400loadframegenerateshighforces.Thecapacityoftheframeis100kN(thatis, 22.5kipsor11.3tons).Theframedoesnotknowthedifferencebetweenasteelplateandyourhand. Alwaysfollowinstructions.NEVERWORKALONE. Beawarethatthelargeamountofenergystoredinaspecimenduringelasticdeformationmaybe released suddenly when fracture occurs; small fragments may be launched. SAFETY GLASSES MUSTBEWORNATALLTIMESINTHEAREA. Protect yourself and the equipment always. Leave the mechanical limit stops as set. STAY CLEAROFMOVINGPARTS. Improper use of the equipment may cause damage and IS HAZARDOUS. Use the fixtures as instructedandwithinspecifiedlimits.forexample,bendingasteelbarwellbeyondyieldproduces excessive lateral forces on the bending fixtures. If in doubt, ASK. Report any hazards or injuries immediatelytothepersonnelinchargeofthelab. Figure1.Instron4400 seriesloadframe. Note:Larger,olderversioncontrolconsoleshown 1

Instron4400 SeriesLoadFrames Introduction The Instron Model 4483 load frame (Figure 1) applies tensile and compressive forces by means of a moving crosshead. The frame is designedas arigid,stablestructurewithahigh stiffnessvalue. In Figure 2, the protective curtain has been temporarily removed to show one of the counter rotating lead screws that raise and lower the crosshead. In the base, the motor drives the lead screws through a series of belts and pulleys. An encoder senses motor rotation forcrossheadspeedcontrol,crossheadpositioning, and position data acquisition. Do not removetheprotectivecurtains. Load Frame Control The ON OFF switch on the frame panel (Figure3)controlspowertotheframe.DONOT switchthepoweroff.iftheswitchisoff,consult your TA. All lab tests are run with the clutch speedcontrolinthelowposition.inlow,the maximum crosshead speed is 50mm/min (2in/min) and the maximum load range of the frame is 100kN (22.5kips). In HIGH, rates to 500mm/min (20in/min) are possible; however, theloadlimitis25kn(5.6kips). Figure3.Framepanel. Totheleftoftheclutchspeedcontrolisthe emergency stop; press it only when there is danger to personnel or a risk of damage to the frame,specimen,orfixtures. Several machine functions may be controlled manually using the control console (Figure4),includingcrossheadpositioning. Figure2.Instron4400 seriesloadframe withprotectivecurtainremoved. Force is sensed by a 100kN (22.5kip) load cellboltedtothecrosshead.thespecimensare loadedbyfixturesthatarepinnedintotheframe baseplatebelowthespecimentestareaandthe load cell above the specimen test area. An extensometermaybeusedtomeasuredeflection of the specimen. The outputs of the load cell force transducer and extensometer are voltages that are converted to engineering units by the framecontrolleranddisplayedonacomputer. Figure4.Controlconsole. 2

Instron4400 SeriesLoadFrames Machinestatusindicatorsarelocatedonthe controlconsole.atthetopareframepowerand frame readiness status indicators. Just below those, on the right, are the GL RESET (gage length reset) button and status indicator, crosshead FINE POSITION thumbwheel, and crossheadjoggingbuttons.thegreysections,at the middle left to lower left, contain crosshead test direction indicators, a red crosshead STOP buttonandindicator,andacrossheadreturn buttonandindicator.usethejogupandjog DOWN buttons and the FINE POSITION thumbwheeltopositionthecrosshead.pressing GL RESET will define the present crosshead position as zero. Pressing RETURN will return thecrossheadtothezerogagelengthcrosshead position if it is not there already. Obviously, greatcaremustbeexercisedinitsuse. For laboratory experiments, the control functions are performed by Instron Bluehill software running on a computer, which is interfaced to the load frame controller. Realtime extension (crosshead position), load, and strainnumericalreadoutsarelocatedacrossthe topofthecomputerscreen. dimensions are in mm. If not, press the mm/inch button above the display. Hold the specimeninthelefthandandthecalipersinthe right hand. Use the thumb to slide the blades against the specimen. (NOTE: This is a precisioninstrument,notac clamp.)theusual practice is to measure a dimension at three to fivelocationsandusethenumericalaverage. The specimen dimensions, along with other pertinent test information, are entered into the runningbluehilltestmethod(seeappendixb.) Placethespecimeninthelowergrip(Figure 6), but do not tighten. Hold the upper grip awayfromthespecimenandusethejogbutton toraisethecrossheaduntilthespecimencanbe inserted into the upper grip. Locate the specimen so that approximately 5mm of the grip end of the specimen extends out of the uppergrip. Theindicatedloadduringthetestshouldbe the load transmitted to the specimen, thus load cell balancing (zeroing) is performed without thespecimeninstalled.thisisdonebyselecting the load cell icon at the top of the computer screenandthenclickingon Balance.Theload reading(atthetopofthescreen)shouldthenbe nearly zero. It should be necessary to balance theloadcellonlyonceduringanylabperiod. Tensile Test Setup and Conduct The appropriate Bluehill test method (the test type specific routine) is typically started beforesetupandoperatedduringsetup,sothat the on screen directions may be followed to assist in test setup (see Appendix B.) Measure the specimen dimensions as shown in Figure 5. First, slide the blades of the digital calipers together against each other; then press ON/ZERO to zero the display. Check that the Figure5.Measuringspecimen dimensionsusingdigitalcalipers. 3

Instron4400 SeriesLoadFrames grip should be tightened firmly without forcing. Figure6.Placingtensiletestspecimen inlowergrip. Figure8.Tighteningthelowergrip. Tighten the lower grip (Figure 8). A small load may be introduced when the grip is tightened; however, this load (typically 0.5 to 1.0kN for metal specimens) is small compared with the testing loads. If the load is not small, loosenthegripandtightenitagain. Press GL RESET to initialize the crosshead position and set the extension display on the computer screen to zero. The green LED lights toindicatezeroextensionposition. If an extensometer is not used, proceed to the final check. Otherwise, continue with extensometerinstallation. Figure7.Tighteningtheuppergrip. Tighten the upper grip first (Figure 7). (NOTE: The upper grip swivels on a U joint; therefore,thespecimenmaybebentifthelower grip is tightened first.) Using the jog button, lower the crosshead to center the specimen betweenthegrips. Extensometer Installation Slide the extensometer (Figure 9) off the yellow hanger on the test frame. Press the knobstoengagetheconestosetthegagelength of the extensometer. (NOTE: Hold the extensometer with the knobs at the top and the signalwireatthebottom.) The grip inserts are shaped to provide wedge action so that they grip more firmly as the test progresses. The gripped ends of the specimen should be free of scale and oil. The 4

Instron4400 SeriesLoadFrames the running Bluehill test method.the strain indicatoronthecomputerscreenshoulddisplay nearly zero. (NOTE: The displayed strain is usually not exactly zero. The extensometer is extremely sensitive and even very small changes, including those caused by vibrations, aredetected.). Figure9.Extensometer,showingknobs tosetgagelength. Continue to hold the knobs and open the spring clips only enough to slide the extensometer knife edges onto the specimen (Figure10).Thebodyoftheextensometermust be oriented to the front (or back) of the test frame to avoid contact with the grips. Center theextensometerverticallywithinthespecimen gage section. Ascertain that the clips seat the extensometer firmly on the specimen. Release theknobs. Final Check Perform a final check of the entire setup: Verify that the load, extension and strain readouts on the computer screen are displayingreasonablenumbers. Verifythatthefixturesandthespecimenare properlyinstalled. Verify that the test section is clear of tools, cables,bodyparts,etc. Verifythatthesafetyshieldsareinplace. If these conditions are satisfied, then the Instron 4400 is ready to run the test. Click STARTonthecomputerscreentobeginthetest. (SeeAppendixB.) After the Test The appearance of the specimen fracture surfaces is important for the analysis of test results.becarefultoprotectthesesurfaces. Figure10.Placingextensometeron tensiletestspecimen. Thestrainindicatoronthecomputerscreen shoulddisplayavaluewithintherangeof±5%. Ifitisoutsidethisrange,consultyourTA. Allthephysicalsetupisnowcomplete.On the computer, click on BALANCE STRAIN in Be certain that the machine has stopped. If not, press the red STOP button on the controlconsole. Carefullyremovetheextensometerfromthe specimen and place it back on its hanger. Usetheholeinthebodyoftheextensometer tohangitup. Loosen the lower grip (using the opposite rotationfromfigure8). Swing the upper grip out of line and carefully slide the lower portion of the specimenupfromthelowergrip. Support the upper portion of the specimen, and then loosen the upper grip (using the opposite rotation from Figure 7). Remove thespecimenupperportion. 5

Instron4400 SeriesLoadFrames Measurethefinalspecimendimensionsand enterthedataontoyourdatasheetsandinto the running Bluehill test method on the computer. Finish the test sample in the Bluehill software(seeappendixb). Theframeisnowreadyforanothertest. Revised 7/12 6